K Number
K142223

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-09-12

(30 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pilot Pro Diode Laser is indicated for the procedures of removal of lesions, excision, incision, vaporization, hemostasis, and photocoagulation on soft tissue in the fields of otolaryngology (ear, nose, and throat), dentistry and oral surgery, arthroscopy, dermatology, podiatry, plastic surgery, urology, and gynecology.

Device Description

The Pilot Pro Diode Laser is a device for delivering laser energy to human soft tissue for a variety of surgical procedures and treatments. This energy is generated by solid-state diodes, which provide a consistent and reliable generation of laser energy at 810 ± 20nm for a maximum of 5 watts of energy output. The laser energy is delivered to the surgical site by means of an optical fiber system, which allows for the safe transmission of laser energy to the site without creating undue risk to the patient or operatory staff by errant or collateral laser emissions. The device features some user definable settings, including a switchable 630nm aiming beam, adjustable power output for both the working beam and aiming beam, and continuous emission or pulse emission options. The working end of the delivery fiber is contained within a metal handpiece with a disposable single-use tip. This handpiece system is removable from the optical fiber system for cleaning and sterilization in between uses. The activation of the working beam diodes is completed by use of a foot-actuated switch.

AI/ML Overview

The provided text is a 510(k) Summary of Safety and Effectiveness for the Pilot Pro Diode Laser. It explicitly states that the device is substantially equivalent to a legally marketed predicate device (DenLaser 800 Plus, K062619). The primary method used to demonstrate this substantial equivalence is via bench testing to confirm the device meets design criteria for essential performance and satisfies performance requirements indicated in 21 CFR 1010 and 21 CFR 1040.

Here's a breakdown of the requested information based on the provided text. It's important to note that since this is a 510(k) for substantial equivalence and not a de novo premarket approval, the level of detailed clinical study information (like MRMC studies, specific expert qualifications, or ground truth establishment methods for large datasets) is often less comprehensive or simply not required in the same way.


1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't present a formal table of specific, quantifiable acceptance criteria with corresponding performance data in the typical sense of a clinical trial. Instead, it relies on demonstrating equivalence to a predicate device and adherence to regulatory standards and design specifications.

Acceptance Criteria (Implied / Stated)Reported Device Performance
Substantial Equivalence to Predicate Device (DenLaser 800 Plus, K062619)The Pilot Pro Diode Laser has "identical indications for use," "identical performance features including wavelength, power output, energy type, operating controls, and laser delivery method," and "identical methods of disinfection and sterilization," "identical methods of control systems, safety features, and performance monitoring."
Compliance with Design Criteria for Essential Performance"Bench testing on an evaluation sample of the current device revealed that the device met the design criteria for essential performance."
Satisfaction of Performance Requirements (21 CFR 1010 & 1040)"Bench testing on an evaluation sample of the current device revealed that the device... satisfied the performance requirements indicated in 21 CFR 1010 and 21 CFR 1040."
Device Outputs within Performance Requirements"Device outputs were within performance requirements." (Specific numerical requirements for output are not detailed in this summary but would be part of the full 510(k) submission).
Safety Features and Functions Operating Correctly"All safety features and functions were operating correctly."
Compliance with Recognized StandardsThe device is "designed to comply with the performance requirements of 21 CFR 1040.10 and 1040.11, with permissible deviations relative to Laser Notice 50, dated June 24, 2007." It also "complies with the recognized standards of IEC 60601-2-22 Edition 2 and IEC 60825-1 Edition 1.2." The device is "designed in compliance to the entirety of IEC 60601-1: 2nd Edition, IEC 60601-1-2, and IEC 60601-1-4."

2. Sample Size used for the test set and data provenance:

  • Sample Size: The document mentions "an evaluation sample of the current device." It does not specify the exact number of devices tested. This implies a limited number of devices used for bench testing, sufficient to demonstrate compliance with design and regulatory requirements.
  • Data Provenance: The testing was "bench testing," meaning it was conducted in a laboratory setting. There is no information about country of origin of data or whether it was retrospective or prospective, as it pertains to device performance testing, not patient data.

3. Number of experts used to establish the ground truth for the test set and their qualifications:

Not applicable. The study described is bench testing of a laser device's physical performance and safety features, not a diagnostic or clinical study requiring expert-established ground truth on patient data.


4. Adjudication method for the test set:

Not applicable. This was bench testing, not a study requiring adjudication of expert interpretations.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No. This document describes a laser surgical instrument, not an AI-assisted diagnostic device. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant or described.


6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Not applicable. This is a hardware device (diode laser) and not an algorithm.


7. The type of ground truth used:

The "ground truth" for this device's performance is based on:

  • Engineering specifications and design criteria: The device was tested against its own design parameters for essential performance.
  • Regulatory standards: It was evaluated against the requirements of 21 CFR 1010, 21 CFR 1040, IEC 60601 series, and IEC 60825-1.
  • Predicate device characteristics: The predicate device's established performance and safety served as a key benchmark for substantial equivalence.

8. The sample size for the training set:

Not applicable. This is not a machine learning or AI-driven device that requires a training set.


9. How the ground truth for the training set was established:

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's logo, which consists of a stylized caduceus with three intertwined snakes and a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the logo in a circular fashion.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 12, 2014

CAO Group Incorporated Mr. Robert K. Larsen Regulatory Affairs Manager 4628 West Skyhawk Drive West Jordan, Utah 84084

Re: K142223

Trade/Device Name: Pilot Pro Diode Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: August 4, 2014 Received: August 13, 2014

Dear Mr. Larsen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

for

Sincerely yours,

David Krause -S

  • Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K142223

Device Name Pilot Pro Diode Laser

Indications for Use (Describe)
The Pilot Pro Diode Laser is indicated for the procedures of

removal of lesions, excision, incision, vaporization, hemostasis, and photocoagulation

on soft tissue in the fields of

otolaryngology (ear, nose, and throat), dentistry and oral surgery, arthroscopy, dermatology, podiatry, plastic surgery, urology, and gynecology.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary of Safety and Effectiveness

CAO Group, Inc. 4628 West Skyhawk Drive West Jordan. UT 84084 Tel: 801-256-9282 Fax: 801-256-9287

Prepared By: Robert K. Larsen, Preparation Date: August 5, 2014

Device Name:

Trade Name:Pilot Pro Diode Laser
Common Name:Soft Tissue Diode Laser
Product Classification:Powered Laser Surgical Instrument

Legally Marketed Predicate Devices for Substantial Equivalence:

DenLaser 800 Plus, manufactured by CAO Group, Inc. (K062619)

Rationale for Substantial Equivalence:

The aforementioned device has identical indications for use with that of the present device for removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation on soft tissue in the fields of otolaryngology (ear, nose, and throat), dentistry and oral surgery, arthroscopy, gastroenterology, dermatology, podiatry, plastic surgery, urology, and gynecology.

The predicate device and submitted device have identical performance features including wavelength, power output, energy type, operating controls, and laser delivery method. The devices have identical methods of disinfection and sterilization. The devices have identical methods of control systems, safety features, and performance monitoring.

Description of Submitted Device:

The Pilot Pro Diode Laser is a device for delivering laser energy to human soft tissue for a variety of surgical procedures and treatments. This energy is generated by solid-state diodes, which provide a consistent and reliable generation of laser energy at 810 ± 20nm for a maximum of 5 watts of energy output. The laser energy is delivered to the surgical site by means of an optical fiber system, which allows for the safe transmission of laser energy to the site without

{4}------------------------------------------------

creating undue risk to the patient or operatory staff by errant or collateral laser emissions. The device features some user definable settings, including a switchable 630nm aiming beam, adjustable power output for both the working beam and aiming beam, and continuous emission or pulse emission options.

The working end of the delivery fiber is contained within a metal handpiece with a disposable single-use tip. This handpiece system is removable from the optical fiber system for cleaning and sterilization in between uses. The activation of the working beam diodes is completed by use of a foot-actuated switch.

Intended Uses of the Submitted Device:

The Pilot Pro Diode Laser is indicated for the procedures of

removal of lesions, excision, incision, vaporization, ablation, hemostasis, and photocoagulation

on soft tissue in the fields of

otolaryngology (ear, nose, and throat), dentistry and oral surgery, arthroscopy, gastroenterology, dermatology, podiatry, plastic surgery, urology, and gynecology.

Technological Characteristics and Substantial Equivalence:

The DenLaser 800 Plus is a device for delivering laser energy to human soft tissue for a variety of surgical procedures and treatments. This energy is generated by solid-state diodes, which provide a consistent and reliable generation of laser energy at 810 ± 20nm for a maximum of 5 watts of energy output. The laser energy is delivered to surgical site by means of an optical fiber system, which allows for the safe transmission of laser energy to the site without creating undue risk to the patient or operatory staff by errant or collateral laser emissions. The device features some user definable settings, including a switchable 630nm aiming beam, adjustable power output for both the working beam and aiming beam, and continuous emission or pulse emission options.

The working end of the delivery fiber is contained within a metal handpiece with a disposable single-use tip. This handpiece system is removable from the optical fiber system for cleaning and sterilization in between uses. The activation of the working beam diodes is completed by use of a foot-actuated switch.

Conformity to Standards:

The Precise SHP Diode Laser is designed to comply with the performance requirements of 21 CFR 1040.10 and 1040.11, with permissible deviations relative to Laser Notice 50, dated June 24, 2007. The device also complies with the recognized standards of IEC 60601-2-22

{5}------------------------------------------------

Edition 2 and IEC 60825-1 Edition 1.2. The device is designed in compliance to the entirety of IEC 60601-1: 2nd Edition, IEC 60601-1-2, and IEC 60601-1-4.

Performance Data

Bench testing on an evaluation sample of the current device revealed that the device met the design criteria for essential performance, and satisfied the performance requirements indicated in 21 CFR 1010 and 21 CFR 1040. Device outputs were within performance requirements and all safety features and functions were operating correctly.

Conclusion

The Pilot Pro Diode Laser is substantially equivalent to the listed predicate device without raising any new issues of safety or effectiveness. This device shares similar intended uses, operating principles, design features, and functional and performance characteristics. The device is designed to comply with relevant federal and international safety and performance standards.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.