K Number
K142156

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-02-13

(191 days)

Product Code
Regulation Number
866.3305
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AnyplexTM II HSV-1/2 Assay is a real-time polymerase chain reaction (PCR)-based in vitro diagnostic test intended for the qualitative detection and differentiation of Herpes Simplex Virus Type-1 (HSV-1) and Herpes Simplex Virus Type-2 (HSV-2) DNA from female skin lesions from anogenital sites. The test is intended for use as and in the diagnosis of anogenital HSV infection in symptomatic patients.

WARNING: The AnyplexTM II HSV-1/2 Assay is not indicated for use with cerebrospinal fluid (CSF). The assay is not intended to be used for prenatal screening.

Device Description

The Anyplex™ II HSV-1/2 Assay uses PCR to generate amplified product from HSV-1 and HSV-2 present in clinical specimens. The presence of HSV-1 and/or HSV-2 target DNA is indicated by the fluorescent signal generated through the use of fluorescently-labeled oligonucleotide probes (duplex Catcher) on the Cepheid SmartCycler® II Dx instrument. The probes do not generate a signal unless they are specifically bound to the amplified product. A preparation of HSV-1 and HSV-2 plasmids is included as the positive control in the Anyplex™ II HSV-1/2 Assay. Run as a separate control, the positive control serves to demonstrate that the HSV-1/2 PCR reagents are functional, and discriminate the validity of the run. In addition, the positive control functions as a process control, demonstrating that sample preparation has proceeded correctly during the run. An internal control (IC) is also included in the assay kit. The IC is added to each sample specimen during sample preparation, and is also used to create a Blank Negative Control by adding a set amount to viral transport media to serve as an extraction control. In addition, the RNase-free water is used to create the Negative Control by adding a set volume to the prepared master mix. Users are instructed to include all three controls, Positive, Negative, and Blank Negative Control with each test run.

AI/ML Overview

The document describes the performance of the Anyplex™ II HSV-1/2 Assay, a real-time PCR-based in vitro diagnostic test for the qualitative detection and differentiation of Herpes Simplex Virus Type-1 (HSV-1) and Herpes Simplex Virus Type-2 (HSV-2) DNA from female skin lesions from anogenital sites.

Here's an analysis of the acceptance criteria and study details based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria (e.g., "The device must achieve sensitivity of X% and specificity of Y%"). Instead, it presents the results of its performance studies, which are implicitly understood to demonstrate adequate performance for regulatory approval based on comparison to a predicate device. The clinical performance metrics are the most relevant in this context.

Implicit Acceptance Criteria (based on reported clinical performance): The device is expected to demonstrate high sensitivity and specificity for both HSV-1 and HSV-2 detection when compared to a reference method in symptomatic patients with anogenital lesions.

Metric (HSV-1)Reported Performance (Anyplex™ II HSV-1/2 Assay)
Sensitivity (95% CI)98.9% (91/92); [94.1%-99.8%]
Specificity (95% CI)93.7% (429/458); [91.0%-95.6%]
Metric (HSV-2)Reported Performance (Anyplex™ II HSV-1/2 Assay)
Sensitivity (95% CI)97.2% (103/106); [92.0%-99.0%]
Specificity (95% CI)93.6% (515/550); [91.3%-95.4%]

Analytical Performance (examples from the document):

  • Precision/Repeatability: Achieved 100% agreement with expected results for HSV-1 Low Positive, HSV-1 High Positive, HSV-2 Low Positive, HSV-2 High Positive, and Negative samples in an in-house study. Ct %CV values were low (e.g., 1.34% - 2.56%).
  • Reproducibility: Across three sites, overall agreement for positive controls (3X LoD and 1X LoD) was high (e.g., 100% for 3X LoD for both HSV-1 and HSV-2; 96.67% for HSV-1 1X LoD; 97.8% for HSV-2 1X LoD). Negative controls showed 99.4% agreement.
  • Limit of Detection (LoD): Varied by strain (e.g., HSV-1 MacIntyre: 3.75x10^2 TCID50/mL; HSV-2 MS: 3.75x10^1 TCID50/mL).
  • Cross-Reactivity/Microbial Interference: No cross-reactivity or interference observed with a panel of 50 organisms.
  • Interfering Substances: None of 22 tested substances showed detectable interference.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: A total of 656 valid prospective specimens were included in the final data set for clinical performance analysis.
    • For HSV-1 performance calculation, 550 specimens were used (specimens positive for HSV-2 by the reference method were removed due to the reference method's inability to detect co-infections).
    • For HSV-2 performance calculation, all 656 specimens were used.
  • Data Provenance: The study was conducted at three geographically diverse locations within the United States from 2013-2014. The data is prospective, as indicated by "656 prospective specimens included in the study."

3. Number of Experts and their Qualifications for Ground Truth

The document does not explicitly state the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience") used to establish the ground truth.

4. Adjudication Method for the Test Set

The primary adjudication method for discordant results in the clinical performance study was bidirectional sequencing.

  • For HSV-1:
    • 29 samples positive by Anyplex but negative by reference method: HSV-1 was detected in 18 cases by bidirectional sequencing. The remaining 11 remained discordant.
    • 1 sample negative by Anyplex but positive by reference method: HSV-1 was detected in this sample by bidirectional sequencing.
  • For HSV-2:
    • 35 samples positive by Anyplex but negative by reference method: HSV-2 was detected in 21 cases by bidirectional sequencing. The remaining 14 remained discordant.
    • 3 samples negative by Anyplex but positive by reference method: HSV-2 was not detected in any of these by bidirectional sequencing (meaning the reference method's positive call was likely a false positive, and Anyplex was correct).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an in-vitro diagnostic assay (laboratory test), not an imaging or interpretation device that typically involves human readers. The clinical performance study compares the device's results against a laboratory-based reference method, not against human interpretation with or without AI assistance.

6. Standalone Performance

Yes, a standalone performance study was done. The entire clinical performance section (Tables 6 and 7) and the analytical performance sections (precision, reproducibility, LoD, cross-reactivity, interfering substances) describe the performance of the Anyplex™ II HSV-1/2 Assay algorithm/device only, without human intervention in the interpretation of the output. The assay provides a qualitative "POS" or "NEG" result for HSV-1 and HSV-2.

7. Type of Ground Truth Used

The ground truth for the clinical performance study was established using the ELVIS® (Enzyme Linked Virus Inducible System) HSV ID and D Typing Test System (Diagnostic Hybrids, Athens, OH), which is described as the "reference ELVIS viral culture method." For discordant results, bidirectional sequencing was used as a confirmatory method to resolve discrepancies.

8. Sample Size for the Training Set

The document does not specify the sample size used for a training set. This is typical for a traditional IVD device using PCR technology, where "training" in the machine learning sense isn't applicable. The development of such assays involves designing primers and probes, and then validating their performance across various analytical and clinical studies.

9. How the Ground Truth for the Training Set Was Established

As noted above, the concept of a "training set" with ground truth in the AI/machine learning sense is not applicable to this type of device. The assay design (e.g., selection of primers and probes) is based on known genetic sequences of HSV-1 and HSV-2. The extensive analytical performance studies (precision, LoD, cross-reactivity) ensure the assay's fundamental ability to detect targets accurately, and the clinical performance study validates this in human samples against a recognized reference method.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol consists of a stylized caduceus, with three human profiles facing right, representing health and well-being. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 17, 2015

SEEGENE C/O FRAN WHITE, REGULATORY CONSULTANT MDC ASSOCIATES, LLC 180 CABOT STREET BEVERLY MA 01915

Re: K142156

Trade/Device Name: Anyplex™ II HSV-1/2 Assay Regulation Number: 21 CFR 866.3305 Regulation Name: Herpes Simplex Virus Nucleic Acid Amplification Assay Regulatory Class: II Product Code: OQO Dated: February 10, 2015 Received: February 11, 2015

Dear Dr. White:

This letter corrects our substantially equivalent letter of February 13, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{1}------------------------------------------------

medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Stephen J. Lovell -S for

Uwe Scherf. M. Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K142156

Device Name AnyplexTM II HSV-1/2 Assay

Indications for Use (Describe)

The AnyplexTM II HSV-1/2 Assay is a real-time polymerase chain reaction (PCR)-based in vitro diagnostic test intended for the qualitative detection and differentiation of Herpes Simplex Virus Type-1 (HSV-1) and Herpes Simplex Virus Type-2 (HSV-2) DNA from female skin lesions from anogenital sites. The test is intended for use as and in the diagnosis of anogenital HSV infection in symptomatic patients.

WARNING: The AnyplexTM II HSV-1/2 Assay is not indicated for use with cerebrospinal fluid (CSF). The assay is not intended to be used for prenatal screening.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date of Summary:February 9, 2015
Product NameAnyplex™ II HSV-1/2 Assay
SponsorSeegeneTaewon Building, 91 Ogeum-ro Songpa-GuSeoul, South Korea, 138-050
CorrespondentMDC Associates, LLCFran White, Regulatory Consultant180 Cabot StreetBeverly, MA 01915
Device Identification
Trade or Proprietary Name:Anyplex™ II HSV-1/2 Assay
Common or Usual Name:HSV-1/2 Assay
Product Code:OQO
Regulation Section:21 CFR 866.3305
Product Classification:Class II

Intended Use

The Anyplex 110 II HSV-1/2 Assay is a real-time polymerase chain reaction (PCR)-based in vitro diagnostic test intended for the qualitative detection and differentiation of Herpes Simplex Virus Type-1 (HSV-1) and Herpes Simplex Virus Type-2 (HSV-2) DNA from female skin lesions from anogenital sites. The test is intended for use as an aid in the diagnosis of anogenital HSV infection in symptomatic patients.

WARNING: The Anyplex™ II HSV-1/2 Assay is not indicated for use with cerebrospinal fluid (CSF). The assay is not intended to be used for prenatal screening.

Device Description

The Anyplex™ II HSV-1/2 Assay uses PCR to generate amplified product from HSV-1 and HSV-2 present in clinical specimens. The presence of HSV-1 and/or HSV-2 target DNA is indicated by the fluorescent signal generated through the use of fluorescently-labeled oligonucleotide probes (duplex Catcher) on the Cepheid SmartCycler® II Dx instrument. The

{4}------------------------------------------------

probes do not generate a signal unless they are specifically bound to the amplified product. A preparation of HSV-1 and HSV-2 plasmids is included as the positive control in the Anyplex™ II HSV-1/2 Assay. Run as a separate control, the positive control serves to demonstrate that the HSV-1/2 PCR reagents are functional, and discriminate the validity of the run. In addition, the positive control functions as a process control, demonstrating that sample preparation has proceeded correctly during the run. An internal control (IC) is also included in the assay kit. The IC is added to each sample specimen during sample preparation, and is also used to create a Blank Negative Control by adding a set amount to viral transport media to serve as an extraction control. In addition, the RNase-free water is used to create the Negative Control by adding a set volume to the prepared master mix. Users are instructed to include all three controls, Positive, Negative, and Blank Negative Control with each test run.

Substantial Equivalency

The Anyplex™ II HSV-1/2 Assay is substantially equivalent to the IMDx HSV-1/2 for Abbott m2000 assay. Table 1 compares the characteristics of the Anyplex™ II HSV-1/2 Assay (New Device) and the IMDx HSV-1/2 for Abbott m2000 assay (Predicate Device).

Similarities
CharacteristicIMDx HSV-1/2 forAbbott m2000 Assay(Predicate Device)Anyplex™ II HSV-1/2 Assay(New Device)
510(k)K140198K142156
Regulation21 CFR 866.330521 CFR 866.3305
Product CodeOQOOQO
Device ClassClass IIClass II
Intended useThe IMDx HSV-1/2 for Abbott m2000 assay is an in vitro diagnostic test for the direct, qualitative detection and differentiation of herpes simplex viruses type 1 (HSV-1) and type 2 (HSV-2) DNA from male and female skin lesions from anogenital or oral sites. The test is intended for use as an aid in the diagnosis of HSV infection in symptomatic patients. The assay is intended to be run on the Abbott m2000 instrument system.Warning: The IMDx HSV-1/2 for Abbott m2000 assay is not FDA cleared for use with cerebral spinalThe Anyplex™ II HSV-1/2 Assay is a real-time polymerase chain reaction (PCR)-based in vitro diagnostic test intended for the qualitative detection and differentiation of Herpes Simplex Virus Type-1 (HSV1) and Herpes Simplex Virus Type-2 (HSV2) DNA from female skin lesions from anogenital sites. The test is intended for use as an aid in the diagnosis of anogenital HSV infection in symptomatic patients.WARNING: The Anyplex™ II HSV-1/2 Assay is not indicated for use with cerebrospinal fluid (CSF). The

Table 1: Substantial Equivalence

{5}------------------------------------------------

Similarities
CharacteristicIMDx HSV-1/2 forAbbott m2000 Assay(Predicate Device)Anyplex™ II HSV-1/2 Assay(New Device)
fluid (CSF) or for pre-natal screening.assay is not intended to be used forprenatal screening.
Test PrincipleReal-time PCR DNA amplificationReal-time PCR DNA amplification
Assay ResultsQualitative detection anddifferentiation of HSV-1 and HSV-2Qualitative detection anddifferentiation of HSV-1 and HSV-2
Differences
CharacteristicIMDx HSV-1/2 forAbbott m2000 Assay(Predicate Device)Anyplex™ II HSV-1/2 Assay(New Device)
InstrumentationSample extraction and real-time PCRamplification/detection using theAbbott m2000 system.Real-time PCRamplification/detection using theCepheid SmartCycler II DX system.
ExtractionMethodAutomated on Abbott m2000 systemManual extraction using the QIAGENQIAamp® DNA Mini Kit
DetectionMethodDouble-labeled (fluorophore andquencher) hydrolysis probes.Measures increase in assayfluorescence with each PCR cycle.Double-labeled (fluorophore andquencher) duplex Catcher. Measuresincrease in assay fluorescence witheach PCR cycle.
Sample typeMale and female skin lesions fromanogenital or oral sitesFemale skin lesions from anogenitalsites only

Performance Characteristics

Analytical Performance

Precision/Repeatability:

A precision study was conducted in-house by testing 5 different panels consisting of HSV-1/2 Negative, HSV-1 Low Positive (1X LoD), HSV-1 High Positive (10X LoD), HSV-2 Low Positive (1X LoD) and HSV-2 High Positive (10X LoD). All panels were tested twice per day for twenty days by one operator. Samples were tested in triplicates for each run (for a total of 600 data points for the 40 runs).

{6}------------------------------------------------

Panel MemberLevelAgreementwith expectedresults95%ConfidenceIntervalAvg. CtSD Ct%CV Ct
HSV-1 LowPositive1X LoD100.00%(120/120)96.90% -100.00%42.930.751.74%
HSV-1 HighPositive10X LoD100.00%(120/120)96.90% -100.00%40.320.711.76%
HSV-2 LowPositive1X LoD100.00%(120/120)96.90% -100.00%41.311.062.56%
HSV-2 HighPositive10X LoD100.00%(120/120)96.90% -100.00%38.190.511.34%
NegativeN/A100.00%(120/120)96.90% -100.00%N/AN/AN/A

Table 2: Precision Study Average Ct Values

Mean: Average Ct of positive results only.

Precision/Reproducibility Results

Reproducibility studies were performed at three sites (one internal and two external clinical sites) using 3 lots of the reagent kits. Each test site tested one lot of the reagent kit. The test panel of seven members was prepared blinded and randomized and tested for five (5) days nonconsecutive days by two operators with each operator running the panel, in triplicate, once a day. Variables in the reproducibility study included between run-to-run, operator-to-operator, site-tosite and lot-to-lot.

Table 3 below shows the Percent (%) Agreement and Average of Ct for each site as well as the Total (%) agreement and Average of Ct for all 3 sites combined.

{7}------------------------------------------------

Table 3: Reproducibility Study Summary

Site 1Site 2Site 3All 3 sites Combined
Panel MemberConcentration% AgreementAgreementwith expectedresultAvg. Ct(%CV)% AgreementAgreement withexpected resultAvg. Ct(%CV)% AgreementAgreementwith expectedresultAvg. Ct(%CV)% AgreementAgreementwith expectedresultAvg. Ct(%CV)
HSV-1 High-3X LoD100.0%30/3036.4(2.0%)100.0%30/3035.9(2.6%)100.0%30/3036.1(2.9%)100.0%90/9036.1(2.6%)
Positive*
HSV-1 Low-1X LoD100.0%30/3039.9(2.2%)90.0%27/3039.8(3.3%)100.0%30/3039.2(1.7%)96.67%87/9039.6(2.6%)
Positive*
HSV-1 High-<1X LoD6.67%2/3042.9(3.2%)43.33%13/3042.6(3.9%)3.33%1/3042.4(2.4%)17.78%16/9042.6(3.1%)
Negative**
HSV-2 High-3X LoD100.0%30/3036.5(1.6%)100.0%30/3035.7(2.2%)100.0%30/3036.4(1.7%)100.0%90/9036.2(2.0%)
Positive*
HSV-2 Low-1X LoD100.0%30/3039.8(2.3%)93.3%28/3039.6(2.4%)100.030/3039.9(2.4%)97.8%88/9039.8(2.3%)
Positive*
HSV-2 High-<1X LoD30.0%9/3042.3(2.8%)46.67%14/3042.1(4.0%)16.67%5/3042.2(2.5%)31.1%28/9042.2(3.0%)
Negative**
HSV Negative**N/A100.0%60/600.098.33%59/6042.7100.0%60/600.099.4%179/18042.7

*Expected result is positive. %CV calculated from results with non-zero Ct values.

** Expected result is negative. %CV calculated from results with non-zero Ct values.

{8}------------------------------------------------

Analytical Sensitivity (Limit of Detection)

The LoD is determined as the HSV-1/2 titer (TCIDs(mL) detected with a positivity rate of 95% or greater. The LoD of the Anyplex™ II HSV-1/2 Assay was determined for two strains of HSV-1 and two strains of HSV-2. The results, representative of the analytical sensitivity of the AnyplexTM II HSV-1/2 Assay, are summarized in Table 4.

StrainLimit of Detection (95% CI)
HSV-1 MacIntyre3.75X102
HSV-1 HF1.88X102
HSV-2 MS3.75x101
HSV-2 G3.75x101

Table 4: Limit of Detection

Cross-Reactivity and Microbial Interference

A panel of 50 organisms was tested for cross-reactivity and interference with the Anyplex™ II HSV-1/2 Assay. Intermediate stocks of bacteria, yeast, viruses, and organism genomic DNA were prepared from quantitated stocks and then diluted to their final test concentration. All samples were prepared by diluting organisms or DNA into M4 viral transport medium. No strains tested were positive for HSV-1 or HSV-2 using the Anyplex™ II HSV-1/2 Assay.

Table 5: Cross-Reactivity and Microbial Interference Panel

OrganismOrganism
Atopobium vaginaeGardnerella vaginalis
Bacteroides fragilisHuman Herpes 6B virus (Z29 strain)
Candida albicansHuman Herpes 7 virus (SB strain)
Candida glabrata Z007Human Papilloma virus-16 (Caski)
Candida guilliemondii Z008Human Papilloma virus-18 (Hela)
Candida krusei Z009Klebsiella pneumoniae Z026
Candida lusitaniae Z010Lactobacillus acidophlus
Candida parapsilosis Z011Lactobacillus crispatus
Candida tropicalis Z012Lactobacillus gasseri
Chlamydia trachomatis (D-UW3)Lactobacillus jensenii
Chlamydia trachomatis (serotype E)Moraxella catarrhalis Ne 11
Chlamydia trachomatis (serotype F)Mycoplasma hominis
Chlamydia trachomatis (serotype G)Neisseria gonorrhoeae
Chlamydia trachomatis (serotype H)Rubella virus
Chlamydia trachomatis (serotype I)Serratia marcescens
Chlamydia trachomatis (serotype J)Staphylococcus saprophyticus
Chlamydia trachomatis (serotype K)Streptococcus mitis
Cytomegalovirus (AD169)Streptococcus mutans Z072

{9}------------------------------------------------

OrganismOrganism
Enterococcus casseliflavusStreptococcus oralis
Enterococcus faecalisStreptococcus pneumoniae 19F
Enterococcus faeciumStreptococcus pyogenes Rosenbach
Enterococcus gallinarumToxoplasma gondii
Enterovirus (Type 71)Trichomonas vaginalis
Epstein-Barr virus (B95-8 strain)Ureaplasma urealyticum
Escherichia coliVaricella Zoster virus

Interfering Substances

The interference study was conducted with the Anyplex™ II HSV-1/2 assay using a panel of twenty two (22) interfering substances that could be present in the female anogenital swab lesion specimens and interfere with the performance of the Anyplex " II HSV-1/2 assay.

The interfering substances were tested at concentrations at or above physiological levels or typical usage levels with HSV strains (HSV-1 MacIntyre and HSV-2 MS) at 3X LoD. None of the 22 substances showed detectable effect of interference, resulting in all samples being positive for HSV1 or HSV2.

Clinical Performance Characteristics

The performance of the Anyplex™ II HSV-1/2 Assay was evaluated at three geographicallydiverse locations within the United States from 2013-2014. A total of 656 valid specimens was included in the final data set and analyzed for product performance as compared to results obtained from the ELVIS® (Enzyme Linked Virus Inducible System) HSV ID and D Typing Test System (Diagnostic Hybrids, Athens, OH). The reference ELVIS viral culture method used in this study is unable to detect co-infected specimens and cannot identify HSV-1 if HSV-2 is identified first. Consequently, if a specimen was positive for HSV-2, it was removed from the calculation of the HSV-1 clinical performance.

Assay performance for anogenital specimens is shown below for the 656 prospective specimens included in the study.

HSV-1 PerformanceReference Method
POSNEGTotal
Anyplex™ II HSV-1/2 AssayPOS9129a120
NEG1b429430
Total92458550
Sensitivity; 95% CI98.9% (91/92);
95% CI [94.1%-99.8%]
Specificity; 95% CI93.7% (429/458);
95% CI [91.0%-95.6%]

Table 6: HSV-1 Anogenital Results

{10}------------------------------------------------

ª Discordant analysis (bidirectional sequencing) was performed on all 29 samples identified as HSV-1 positive by the Anyplex™ II HSV-1/2 Assay. HSV-1 was detected in 18 of the 29 samples. The remaining 11 specimens remained discordant (HSV-1 was not detected).

b Discordant analysis (bidirectional sequencing) was performed on the one sample identified as negative by the Anyplex™ II HSV-1/2 Assay. HSV-1 was detected in this sample.

HSV-2 PerformanceReference Method
POSNEGTotal
Anyplex TM II HSV-1/2 AssayPOS10335c138
NEG3d515518
Total106550656
Sensitivity; 95% CI97.2% (103/106);95% CI [92.0%-99.0%]
Specificity; 95% CI93.6% (515/550);95% CI [91.3%-95.4%]

Table 7: HSV-2 Anogenital Results

^ Discordant analysis (bidirectional sequencing) was performed on all 35 discordant specimens identified as HSV-2 positive by the Anyplex™ II HSV-1/2 Assay. HSV-2 was detected in 21 of the 35 specimens. The remaining 14 remained discordant (HSV-2 was not detected).

d Discordant analysis (bidirectional sequencing) was conducted for the three specimens identified as HSV-2 negative by the Anyplex™ II HSV-1/2 Assay. HSV-2 was not detected in any specimen.

Conclusions

The submitted information in this premarket notification is complete and supports a substantial equvalence decision.

§ 866.3305 Herpes simplex virus serological assays.

(a)
Identification. Herpes simplex virus serological assays are devices that consist of antigens and antisera used in various serological tests to identify antibodies to herpes simplex virus in serum. Additionally, some of the assays consist of herpes simplex virus antisera conjugated with a fluorescent dye (immunofluorescent assays) used to identify herpes simplex virus directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by herpes simplex viruses and provides epidemiological information on these diseases. Herpes simplex viral infections range from common and mild lesions of the skin and mucous membranes to a severe form of encephalitis (inflammation of the brain). Neonatal herpes virus infections range from a mild infection to a severe generalized disease with a fatal outcome.(b)
Classification. Class II (special controls). The device is classified as class II (special controls). The special control for the device is FDA's revised guidance document entitled “Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 and 2 Serological Assays.” For availability of the guidance revised document, see § 866.1(e).