K Number
K142123

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-09-24

(51 days)

Product Code
Regulation Number
868.1890
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The REACH Access Telemedicine Dose Calculator is designed for used by trained clinicians to calculate any individual patient's dose for a given agent based on a weight determined by the clinician. Drug dosing for a patient must be made only after careful consideration of the full clinical status of the patient by the ordering clinician and the software provides no validation as to the appropriateness of the entered weight-based dose. The REACH Access Telemedicine Dose Calculator is a convenience feature for trained clinicians based upon accurately entered data. No medical decision should be based solely upon the results provided by this software program. The REACH Access Telemedicine Dose Calculator is intended to be used as an aid in the consultative process and does not overrule or replace physician judgment or diagnosis.

Device Description

REACH Access is a highly integrated Telemedicine solution that combines workflow, video conferencing, electronic clinical documentation, and medical imaging into a comprehensive and secure internet-based service. It enables healthcare organizations to form virtual collaborative care teams which are created based on patient need rather than availability of specialists. With a flexible, template-based design tool, REACH Access can build template sets which support clinician workflow and documentation relevant to numerous protocols. REACH Access templates support a specific point-in-time consult. This means the workflow processes and clinical content is captured and coordinated with the integrated audio/video conference to allow a remote provider to collaborate with caregivers at a different location in the care of a patient.

The normal process is the patient is assessed and treated in the Emergency Department (ED) by ED staff and the video conference support of a specialist consultant. After reviewing key patient data and assessing the patient, the remote consultant renders an impression, recommends the treatment order as indicated, and helps determine the appropriate disposition. This usually concludes the consultant's part in the session. The ED continues to render care by completing the recommended treatment order and monitoring patient vital signs and other assessments. The disposition commonly consists of transferring the originating hospital to the Hub with which the consultant is associated. However, if there is adequate local support, the decision may be made to admit the patient to the local facility.

This process/workflow is similar to what may be required for other point-in-time consults needed to provide additional remote assistance to caregivers for other patient problems or conditions at the patient point of care. A protocol/template can be part of a coordinated collection of toolsets. The flexible template design process in REACH Access addresses the need for multiple protocols to support the various phases of care. The flexibility of the tool allows for the unique presentation of data for a specific condition; e.g., stroke, in the format most relevant to the patient's current need/condition. This flexibility can be replicated to address similar capability. This can be handled through a combination of both individual and a series of protocols arranged in clinically logical groups.

AI/ML Overview

This document is a 510(k) premarket notification decision letter from the FDA for the "REACH Access Telemedicine Dose Calculator." It focuses on establishing substantial equivalence to a predicate device rather than presenting a performance study with acceptance criteria and results from an independent test set.

Therefore, most of the information requested in your prompt regarding acceptance criteria, study design, sample sizes, ground truth establishment, expert involvement, and MRMC studies is not present in this document. This typically wouldn't be included in a 510(k) summary for devices like a dose calculator where the primary demonstration of safety and effectiveness relies on software verification and validation, and comparison to a predicate device with similar operational principles.

Here's what can be extracted and what cannot be from the provided text:

1. A table of acceptance criteria and the reported device performance:

  • Acceptance Criteria: Not explicitly stated as formal, quantitative acceptance criteria for a performance study. The core "acceptance" for this device is "substantial equivalence" to the predicate, based on its intended use, technological characteristics, and principles of operation, and that it raises "no issues of safety or effectiveness."
  • Reported Device Performance: Instead of performance metrics from a clinical study, the document reports "Validation and Verification" was done. The conclusion states "The performance of the REACH Access Telemedicine Dose Calculator is substantially equivalent to that of the Picis Weight Based Dose Converter and raises no safety or effectiveness issues and performs as well or better than the predicate device."
    • Implication: For a dose calculator, "performance" would primarily relate to the accuracy of its mathematical calculations. The mention of "Validation and Verification" suggests that the software was tested to ensure it correctly computes doses based on the input weight and dose rate, as per its simple mathematical equation. However, no specific numerical performance results (e.g., error rates, precision) are provided.

2. Sample sizes used for the test set and the data provenance:

  • Sample Size (Test Set): Not mentioned. "Validation and Verification" for a software like a dose calculator typically involves testing a defined set of inputs and expected outputs, not a "sample size" in the sense of patient data in a clinical trial.
  • Data Provenance: Not applicable in the context of this device type and the information provided. The "data" being processed is clinician-entered weight.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Number of Experts: Not applicable. Ground truth for a mathematical calculation is the correct mathematical result, not expert interpretation.
  • Qualifications of Experts: N/A

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Adjudication Method: Not applicable.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • MRMC Study: No. This is not an AI-assisted diagnostic imaging device. It's a simple dose calculator.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Standalone Performance: For a dose calculator, the "standalone performance" refers to its mathematical accuracy. While "Validation and Verification" implies such testing was done, no specific results or methods are detailed in this summary. The device's function is purely algorithmic: "Uses simple mathematical equation to calculate dosing based on patient weight."

7. The type of ground truth used:

  • Ground Truth: For a drug dose calculator, the ground truth would be the quantitatively correct mathematical dose calculation based on the entered weight and the specified drug dose per unit weight (which the clinician determines). This ground truth is derived from the fundamental mathematical operations the calculator performs.

8. The sample size for the training set:

  • Training Set Sample Size: Not applicable. This device is not an AI/ML model that requires a training set. It is a rule-based software performing a direct calculation.

9. How the ground truth for the training set was established:

  • Ground Truth for Training Set: Not applicable as there is no training set for this type of device.

Summary of what is present in the document relevant to your query:

  • The device is a "REACH Access Telemedicine Dose Calculator."
  • Its purpose is to "calculate any individual patient's dose for a given agent based on a weight determined by the clinician."
  • The calculation "Uses simple mathematical equation to calculate dosing based on patient weight."
  • The study done was "Validation and Verification" (which for such a device implies testing the accuracy of the calculation).
  • The primary method of demonstrating legitimacy was to show "Substantial Equivalence" to a predicate device (Picis Weight Based Dose Converter, K121542), meaning it has the "same intended uses, technological characteristics and principles of operation."

This document does not describe the kind of performance study one would expect for an AI/ML diagnostic or prognostic device, but rather a software validation process for a simple calculation tool.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized human figure or a series of faces in profile, often interpreted as representing the department's focus on people and health.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 24, 2014

REACH Health, Incorporated Ms. Beth DeStasio Director, Regulatory Affairs & Strategy 10745 Westside Way Alpharetta, GA 30009

Re: K142123

Trade/Device Name: REACH Access Telemedicine Dose Calculator Regulation Number: 21 CFR 868.1890 Regulation Name: Drug Dose Calculator Regulatory Class: II Product Code: NDC, LLZ, OUG Dated: August 4, 2014 Received: August 5, 2014

Dear Ms. DeStasio:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Ms. DeStasio

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

of Surveillance and Biometrics/Division of Postmarket Surveillance.

Sincerely yours,

Mary S. Runner -S

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K142123

Device Name REACH Access Telemedicine Dose Calculator

Indications for Use (Describe)

The REACH Access Telemedicine Dose Calculator is designed for used by trained clinicians to calculate any individual patient's dose for a given agent based on a weight determined by the clinician. Drug dosing for a patient must be made only after careful consideration of the full clinical status of the ordering clinician and the software provides no validation as to the appropriateness of the entered weight-based dose. The REACH Access Telemedicine Dose Calculator is a convenience feature for trained clinicians based upon accurately entered data. No medical decision should be based solely upon the results provided by this software program. The REACH Access Telemedicine is intended to be used as an aid in the consultative process and does not overrule or replace physician judgmosis.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a logo for Reach Health, a telemedicine company. The logo features a blue circle with a gray plus sign in the center, followed by the word "REACHHEALTH" in black and blue letters. Below the company name is the tagline "POWERING TELEMEDICINE" in gray letters. The logo is simple and modern, and it conveys the company's focus on healthcare and technology.

5. 510(k) Sumn

Submitter:REACH Health, Inc.
Address:10745 Westside WayAlpharetta, GA 30009
Phone number:(678) 436-8214
Fax number:(678) 710-9064
Contact person:Beth DeStasio
Phone number:(678) 436-8214
Fax number:(678) 710-9064
Date prepared:8/1/2014
Trade name:REACH Access Telemedicine Dose Calculator
Common name:Drug Dose Calculator
Primary Product Code:NDC
Classification Number:21 CFR 868.1890
Class:II
Secondary Product Codes:LLZ (Picture archiving and communications system), OUG (MedicalDevice Data System)
510(k) Number:K142123

Substantial equivalence claimed to: PICIS WEIGHT BASED DOSE CONVERTER, K121542

Description:

REACH Access is a highly integrated Telemedicine solution that combines workflow, video conferencing, electronic clinical documentation, and medical imaging into a comprehensive and secure internet-based service. It enables healthcare organizations to form virtual collaborative care teams which are created based on patient need rather than availability of specialists. With a flexible, template-based design tool, REACH Access can build template sets which support clinician workflow and documentation relevant to numerous protocols. REACH Access templates support a specific point-in-time consult. This means the workflow

processes and clinical content is captured and coordinated with the integrated audio/video conference to allow a remote provider to collaborate with caregivers at a different location in the care of a patient.

The normal process is the patient is assessed and treated in the Emergency Department (ED) by ED staff and the video conference support of a specialist consultant. After reviewing key patient data and assessing the patient, the remote consultant renders an impression, recommends the treatment order as indicated, and helps determine the appropriate disposition. This usually concludes the consultant's part in the session. The ED continues to render care by completing the recommended treatment order and monitoring patient vital signs and other assessments. The disposition commonly consists of transferring the originating hospital to the Hub with which the consultant is associated. However, if there is adequate local support, the decision may be made to admit the patient to the local facility.

{4}------------------------------------------------

This process/workflow is similar to what may be required for other point-in-time consults needed to provide additional remote assistance to caregivers for other patient problems or conditions at the patient point of care. A protocol/template can be part of a coordinated collection of toolsets. The flexible template design process in REACH Access addresses the need for multiple protocols to support the various phases of care. The flexibility of the tool allows for the unique presentation of data for a specific condition; e.g., stroke, in the format most relevant to the patient's current need/condition. This flexibility can be replicated to address similar capability. This can be handled through a combination of both individual and a series of protocols arranged in clinically logical groups.

Indications for Use:

The REACH Access Telemedicine Dose Calculator is designed for used by trained clinicians to calculate any individual patient's dose for a given agent based on a weight determined by the clinician. Drug dosing for a patient must be made only after careful consideration of the full clinical status of the patient by the ordering clinician and the software provides no validation as to the appropriateness of the entered weight-based dose. The REACH Access Telemedicine Dose Calculator is a convenience feature for trained clinicians based upon accurately entered data. No medical decision should be based solely upon the results provided by this software program. The REACH Access Telemedicine Dose Calculator is intended to be used as an aid in the consultative process and does not overrule or replace physician judgment or diagnosis.

{5}------------------------------------------------

Substantial Equivalence:

The REACH Access Telemedicine Dose Calculator is substantially equivalent to the predicate Device, the Picis Weight Based Dose Converter.

REACH AccessPicis (K121542)
ManufacturerREACH Health, Inc.Picis Incorporated
510(k)NAK121542
Indication foruseThe REACH Access TelemedicineDose Calculator is designed forused by trained clinicians tocalculate any individual patient'sdose for a given agent based on aweight determined by the clinician.Drug dosing for a patient must bemade only after carefulconsideration of the full clinicalstatus of the patient by theordering clinician and the softwareprovides no validation as to theappropriateness of the enteredweight-based dose. The REACHAccess Telemedicine DoseCalculator is a conveniencefeature for trained clinicians basedupon accurately entered data. Nomedical decision should be basedsolely upon the results provided bythis software program. TheREACH Access TelemedicineDose Calculator is intended to beused as an aid in the consultativeprocess and does not overrule orreplace physician judgment ordiagnosis.The Picis Weight Based DoseConverter is designed for use bytrained clinicians to calculate anyindividual patients dose for a givenagent based on a weight baseddose determined by the clinician.Drug dosing for a patientmust be made only after carefulconsideration of the full clinicalstatus of the patient by theordering clinician and thesoftware provides no validationas to the appropriateness of theentered weight based dose.The Picis Weight Based DoseConverter is not a substitute forclinical reasoning. The PicisWeight Based Dose Converteris a convenience feature fortrained clinicians based uponaccurately entered data. Nomedical decision should be basedsolely upon the results providedby this software program.
TargetPopulationIn-hospital patientsIn-hospital patients
User PopulationTrained healthcare professionalsTrained healthcare professionals
DoseComputationSoftware based application usedto compute total drug doseSoftware based application usedto compute total drug dose
AlgorithmsUses simple mathematicalequation to calculate dosing basedon patient weightUses simple mathematicalequation to calculate dosingbased on patient weight andnumerical dose and units
TestingValidation and VerificationValidation and Verification

{6}------------------------------------------------

Substantial Equivalence Discussion

The REACH Access Telemedicine Dose Calculator is substantially equivalent to the predicate Device, the Picis Weight Based Dose Converter. The REACH Access Telemedicine Dose Calculator has the same intended uses, technological characteristics and principles of operation as the predicate device. The minor technological differences between the REACH Access Telemedicine Dose Calculator and the predicate device raise no issues of safety or effectiveness. Both devices calculate an individual patient's dose for a given agent based on a weight based dose determined by the clinician.

Conclusion:

The performance of the REACH Access Telemedicine Dose Calculator is substantially equivalent to that of the Picis Weight Based Dose Converter and raises no safety or effectiveness issues and performs as well or better than the predicate device.

§ 868.1890 Predictive pulmonary-function value calculator.

(a)
Identification. A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.(b)
Classification. Class II (performance standards).