K Number
K142088

Validate with FDA (Live)

Date Cleared
2014-10-31

(91 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is for use by medical professional or home users. It is intended to measure the systolic and diastolic blood pressure of an adult individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the wrist.

Device Description

Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series is a non-invasive blood pressure measurement system for use by medical professional or at home. It is designed to measure the systolic and diastolic blood pressure, and pulse rate of an individual in each measurement and then display the readings on a digital panel. The device utilizes the oscillometric methodology, in which an inflatable cuff is wrapped around the wrist of an individual, for blood pressure measurement. This means the monitor detects your blood's movement through your brachial artery and converts the movement into a digital reading.

AI/ML Overview

The provided text describes the regulatory clearance of a medical device, the "Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series," based on its substantial equivalence to a predicate device. While it mentions clinical and non-clinical tests, it does not provide detailed acceptance criteria or the specific results in a format that directly addresses all the requested points for a "study that proves the device meets the acceptance criteria."

However, I can extract the relevant information and present it in the closest possible format based on the given text.

1. A table of acceptance criteria and the reported device performance

The document references ISO 81060-2:2009 for clinical accuracy. This standard defines the accuracy requirements for non-invasive sphygmomanometers. While the specific numerical acceptance criteria (e.g., mean difference and standard deviation of blood pressure measurements) for ISO 81060-2:2009 are not explicitly stated in this document, the study reports that it was conducted in accordance with this standard. The "reported device performance" is implicitly that the device met these standards, as it received clearance.

For completeness, I will include the blood pressure measurement accuracy range for the device itself, as stated in the comparison table, which serves as a self-declared performance specification.

Acceptance Criteria / Performance MetricReported Device Performance (or Standard Met)
Clinical Accuracy (Systolic/Diastolic Blood Pressure Measurement)Met requirements of ISO 81060-2:2009
Blood Pressure Measurement Accuracy (Device Specification)± 3 mmHg
Pulse Rate Measurement Accuracy (Device Specification)± 5% of the reading
Blood Pressure Measurement RangeCuff Pressure: 0 - 300 mmHg, Systolic: 50 - 250 mmHg, Diastolic: 30 - 200 mmHg
Pulse Rate Measurement Range30 - 180 beats/min
Compliance with Electrical Safety and EMCIEC 60601-1:2012, IEC 60601-1-2:2007
BiocompatibilityISO 10993-5:2009, ISO 10993-10:2010
Software Life Cycle ProcessesIEC 62304:2006
Non-invasive Sphygmomanometers - Part 1: General requirementsEN 1060-1:1995+A2:2009
Non-invasive Sphygmomanometers - Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systemsEN 1060-3:1997+A2:2009
Performance of automatic non-invasive sphygmomanometersIEC 80601-2-30 (bench testing conducted)

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Sample Size: 100 patients (46 males and 54 females)
  • Data Provenance: Not specified, but generally, clinical studies for medical devices are prospective. The text indicates "were invited for the study," which suggests a prospective clinical trial. The country of origin of the data is not mentioned within the provided text.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Number of Experts: Not explicitly stated. The "Standard auscultation method was used as the reference blood pressure monitor" implies that a trained healthcare professional performed the auscultation, which serves as the "ground truth" reference. The number of such professionals is not specified, nor are their specific qualifications (e.g., "physician with X years of experience").

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Adjudication Method: Not applicable or not specified. In a blood pressure accuracy study using auscultation as the reference, the auscultatory measurement itself is considered the reference. There isn't typically an "adjudication" process in the sense of reconciling multiple expert interpretations; rather, the accuracy of the device is assessed against this direct reference.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, an MRMC comparative effectiveness study was not done. This type of study design is typically used for AI/software-as-a-medical-device (SaMD) products where human readers (e.g., radiologists) are assisted by AI. The device in question is a standalone blood pressure monitor, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: Yes, the clinical test described details the standalone performance of the device (Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series) against the reference auscultation method. This is not an "algorithm only" study in the context of AI, but rather the performance of the complete automated device. The device itself operates without continuous human input during the measurement cycle.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Type of Ground Truth: The ground truth for blood pressure measurements was established using the standard auscultation method, which is performed by trained medical professionals (human experts) and is considered the gold standard for non-invasive blood pressure measurement in such studies.

8. The sample size for the training set

  • Training Set Sample Size: The document does not describe a "training set." The device is cleared as substantially equivalent to a predicate device, and the mentioned clinical study is a validation or verification test, not a training process for an algorithm. If the device contained a machine learning component, a training set would be relevant, but this is not suggested.

9. How the ground truth for the training set was established

  • Ground Truth for Training Set: Not applicable, as there is no mention of a training set for a machine learning algorithm. The device's operation is based on established oscillometric methodology and engineering principles rather than a trained AI model.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of a bird-like figure with three human profiles forming the body.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 31, 2014

Grandway Technology (Shenzhen) Limited Patrick Chow General Manager Block 6 and 7. Zhu Keng Industrial Zone. Ping Shan, Long Gang District, Shenzhen, Guang Dong, China

Re: K142088

Trade/Device Name: Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: September 30, 2014 Received: September 30, 2014

Dear Patrick Chow,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Melissa A. Torres -S

Bram D. Zuckerman, M.D. For Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K142088

Device Name

Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series, MD2200, MD2210

Page 1 of 1

Indications for Use (Describe)

This device is for use by medical professional or home users. It is intended to measure the systolic and diastolic blood pressure of an adult individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the wrist.

Type of Use (Select one or both, as applicable)

  • Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SubmitterGRANDWAY TECHNOLOGY (SHENZHEN) LIMITED
AddressBlock 6 and 7, Zhu Keng Industrial Zone, Ping Shan, Long Gang
District, Shenzhen, Guang Dong, People's Republic of China
Phone Number(00852)-2851-6789
Fax Number(00852)-2851-6278
Contact PersonMr. Patrick Chow
Date of Submission30th July, 2014

Submitter Identification 1.

Device Identification 2.

Trade NameDigital Automatic Wrist Blood Pressure Monitor SWBPM22 Series[Model No.: MD22x0]x --- The first character (0 and 1) is for the identification of outlook.
Common NameNon-invasive Blood Pressure Measurement System
Classification NameNon-invasive Blood Pressure Measurement System(CFR 870.1130, Class II, Product Code DXN)

3. Predicate Device

Predicate DeviceDigital Automatic Wrist Blood Pressure Monitor WBPM22 Series
ManufacturerGRANDWAY TECHNOLOGY (SHENZHEN) LIMITED
510(k) NumberK133618

Device Description 4.

Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series is a non-invasive blood pressure measurement system for use by medical professional or at home. It is designed to measure the systolic and diastolic blood pressure, and pulse rate of an individual in each measurement and then display the readings on a digital panel.

The device utilizes the oscillometric methodology, in which an inflatable cuff is wrapped around the wrist of an individual, for blood pressure measurement. This means the monitor detects your blood's movement through your brachial artery and converts the movement into a digital reading.

The table below illustrate the feature presence in Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series.

ModelBlood PressureMeasurementPulse RateMeasurementWHOClassificationIrregularHeartbeatLCD TypeUser ×Memory
MD2200PositiveReflective2 × 120
MD2210PositiveReflective2 × 120

{4}------------------------------------------------

5. Indication for Use

This device is for use by medical professional or home users. It is intended to measure the systolic and diastolic blood pressure of an adult individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the wrist.

Comparison of Technological Characteristics between New Device and 6. Predicate Devices

Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series is compared to the predicate device, WBPM22 Series (K133618) in the device comparison table below.

Comparison between SWBPM22 Series and Predicate device
ItemPredicate DeviceWBPM22 SeriesComment
Indication for UseDigital Automatic Wrist Blood PressureThis device is for use by medicalEquivalent
Monitor WBPM22 Series is for use byprofessional or home users. It is intended
medical professional or home user. Theto measure the systolic and diastolic
WBPM22 Series is intended to measureblood pressure of an adult individual by
the systolic and diastolic blood pressure,using a non-invasive technique, in which
and pulse rate of an adult individual byan inflatable cuff is wrapped around the
using a non-invasive technique, in whichwrist.
an inflatable cuff is wrapped around the
wrist.
MeasurementNon-invasive, OscillometricNon-invasive, OscillometricIdentical
Method
IHB DetectionYesYesIdentical
Patient PopulationAdultAdultIdentical
BP MeasurementCuff Pressure: 0 - 300 mmHgCuff Pressure: 0 - 300 mmHgIdentical
RangeSystolic Pressure: 50 - 250 mmHgSystolic Pressure: 50 - 250 mmHg
Diastolic Pressure: 30 - 200 mmHgDiastolic Pressure: 30 - 200 mmHg
Number of User2 independent users2 independent usersIdentical
Memory Space2 users × 60 memory space;2 users × 120 memory spaceEquivalent
or 2 users × 120 memory space
Resolution ofBlood Pressure: 1 mmHg or 0.1kPaBlood Pressure: 1 mmHg or 0.1 kPaIdentical
MeasurementPulse Rate: 1 beat/ minPulse Rate: 1 beat/ min
Blood Pressure± 3 mmHg or 2% of reading± 3 mmHgEquivalent
Measurement
Accuracy
Pulse Rate30 - 180 beats/min30 - 180 beats/minIdentical
Measurement
Range
Pulse Rate± 5 % of the reading± 5 % of the readingIdentical
Measurement
Accuracy
Display TypeLCDLCDIdentical
Power Source2 × 1.5 V AAA-batteries2 × 1.5 V AAA-batteriesIdentical
PressurizationAutomatic InflationAutomatic InflationIdentical
Mode
Deflation ModeAutomatic Exhaust/ DeflationAutomatic Exhaust/ DeflationIdentical
OperatingTemperature: +5 to +40 °CTemperature: +5 to +40 °CIdentical
ConditionHumidity: 15 to 93 % R.H. maxHumidity: 15 to 93 % R.H. max
Atmospheric Pressure: 700 - 1060 hPaAtmospheric Pressure: 700 - 1060 hPa
Storage andTemperature: -25 to +70 °CTemperature: -25 to +70 °CIdentical
TransportationHumidity: up to 93% R.H. maxHumidity: up to 93% R.H. max
ConditionAtmospheric Pressure: 700 - 1060 hPaAtmospheric Pressure: 700 - 1060 hPa

{5}------------------------------------------------

Comparison between SWBPM22 Series and Predicate device
ItemPredicate DeviceWBPM22 SeriesComment
MaterialResistances, capacitance, transistors, amplifiers, pressure sensor, CPU, PCB, cuff ABS button, ABS cabinet, batteries and packagingResistances, capacitance, transistors, amplifiers, pressure sensor, CPU, PCB, cuff ABS button, ABS cabinet, batteries and packagingIdentical
Compatibility with Environment and Other DevicesNo influence with environment and other deviceNo influence with environment and other deviceIdentical
Applicable Standard- EN 1060-1:1995+A2:2009- EN 1060-3:1997+A2:2009- IEC 60601-1:2012- EN 60601-1-2:2007- FCC Part 15 Subpart B- ISO 10993-5:2009- ISO 10993-10:2010- IEC 62304:2006- IEC 81060-2:2009- EN 1060-1:1995+A2:2009- EN 1060-3:1997+A2:2009- IEC 60601-1:2012- IEC 60601-1-2:2007- FCC Part 15 Subpart B- ISO 10993-5:2009- ISO 10993-10:2010- IEC 62304:2006- IEC 81060-2:2009Equivalent

Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series is a non-invasive measuring device and utilizes the oscillometric methodology to measure the blood pressure reading. The key components of device are: a pressure sensor, a electric valve and an electronic control module together with an electric pump. The electric pump inflate (and deflate) the inflatable cuff automatically according to our designed architecture. The predicate device adopts exactly same methodology and key components for measuring blood pressure.

Clinical and Non-clinical Tests 7.

Clinical Test Summary

Testing to insure clinical accuracy of the device in accordance with ISO 81060-2:2009 as documented in Clinical Test report.

One hundred patients (46 males and 54 females) were invited for the study. Standard auscultation method was used as the reference blood pressure monitor measuring in the left wrist. Blood pressure measurements were repeated alternatively with the device and auscultation in the same arm according to the sequence in ISO 81060-2:2009.

Non-Clinical Test Summary

Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series has performed several nonclinical tests to show that all requirement specifications and standard requirements are met. The tests includes the follows:

  • EN 1060-1:1995+A2:2009 <>
  • EN 1060-3:1997+A2:2009

  • ◆ IEC 60601-1:2012
  • や IEC 60601-1-2:2007
  • や FCC Part 15 Subpart B
  • ◆ ISO 10993-5:2009
  • ISO 10993-10:2010

  • IEC 62304:2006

{6}------------------------------------------------

All of the clinical and non-clinical testing performed on Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series are same as the predicate device.

Also, bench testing, IEC 80601-2-30, is conducted to show the performance of Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series is equivalent to the predicate device.

Conclusion 8.

Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series has the same intended use and same technological characteristics as the predicate device, Digital Automatic Wrist Blood Pressure Monitor WBPM22 Series (K133618). Moreover both clinical and non- clinical testing has demonstrated that no differences in the technological characteristics and questioning on safety or effectiveness to be raised. Thus, Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series is substantially equivalent to the predicate device.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).