K Number
K142084

Validate with FDA (Live)

Date Cleared
2014-10-28

(89 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EPi-Sense Guided Coagulation System with VisiTrax is intended for the coagulation of cardiac tissue using Radiofrequency (RF) energy using thoracoscopic, endoscopic, and laparoscopic surgical techniques.

The EPi-Sense Guided Coagulation System with VisiTrax may be used for temporary cardiac signal sensing and recording during surgery when connected to an external recording device.

Device Description

The EPi-Sense Guided Coagulation System with VisiTrax consists of a sterile, single-use, disposable coagulation electrode device (1cm, 2cm, & 3cm sizes provided) intended to be used to coagulate cardiac tissue. The flexible, cooled electrode device, with a suction stabilizer feature, transmits radiofrequency (RF) energy from an Electrosurgical Generator (non-sterile, re-useable) connected through an Instrument Cable (sterile). A temporary sensing electrode feature may be used with an off-the-shelf electrogram recording system with the use of an additional instrument cable (non-sterile). An accessory Cannula may be used to facilitate coagulation device access and visibility of the heart. The Cannula accessory changes are the subject of this submission.

AI/ML Overview

This document describes a 510(k) premarket notification for the nContact Surgical, Inc. EPi-Sense Guided Coagulation System with VisiTrax. The submission focuses on minor changes to the Cannula accessory, not the core coagulation device or an AI component. Therefore, the request for information related to AI/algorithm performance (acceptance criteria, sample sizes for test/training sets, experts, MRMC studies, standalone performance, ground truth establishment) is not applicable to this submission.

The "device" in this context refers to the EPi-Sense Guided Coagulation System with VisiTrax, specifically the Cannula accessory.

Here's the information based on the provided document, addressing the non-AI aspects:

1. A table of acceptance criteria and the reported device performance

The document does not provide a table of precise acceptance criteria with numerical targets. Instead, it states that "Performance bench tests were executed to ensure that the Cannula accessory performed as intended and met design specifications." The reported device performance is that the tests were "successfully completed to evaluate equivalence."

The successful completion of the following tests implies they met pre-defined acceptance criteria, although the specific criteria are not detailed in this summary:

Test CategoryDescriptionReported Performance
BiocompatibilityPer ISO 10993 for cannula accessory materials.Successfully completed
Sterilization ValidationPer ISO 11137-2, Sterilization of Health Care Products - Radiation - Method VD Max for cannula accessory.Successfully completed
Reliability TestingSuch as shipping and accelerated aging of packaged units.Successfully completed
Tensile TestingOf critical bonds and joints.Successfully completed
Flexion Fatigue TestingSuccessfully completed
Electrical Integrity TestingFor the cannula accessory used with the coagulation device to pertinent sections of IEC 60601-1-2.Successfully completed

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes used for each of the nonclinical tests. It only states that "Performance bench tests were executed." There is no mention of data provenance (e.g., country of origin, retrospective/prospective) which is typically relevant for clinical studies, not bench testing of a device accessory.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable. The testing described is bench testing of physical properties and regulatory compliance (biocompatibility, sterilization), not a diagnostic or AI performance study requiring expert ground truth.

4. Adjudication method for the test set

This question is not applicable for the reasons stated above.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable as there is no AI component mentioned in this 510(k) submission.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This question is not applicable as there is no AI component mentioned in this 510(k) submission.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

This question is not applicable. The "ground truth" for these tests would be established by validated test methods and industry standards (e.g., ISO for biocompatibility and sterilization, IEC for electrical safety), not expert consensus, pathology, or outcomes data.

8. The sample size for the training set

This question is not applicable as there is no AI component mentioned in this 510(k) submission.

9. How the ground truth for the training set was established

This question is not applicable as there is no AI component mentioned in this 510(k) submission.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure composed of three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 28, 2014

nContact Surgical, Inc. % Jane Ricupero VP of Regulatory and Quality 1001 Aviation Parkway, Suite 400 Morrisville, North Carolina 27560

Re: K142084

Trade/Device Name: Epi-Sense Guided Coagulation Device With VisiTrax Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: OCL Dated: September 4, 2014 Received: September 8, 2014

Dear Jane Ricupero,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Melissa A. Torres -S

For

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known): K

Device Name: EPi-Sense® Guided Coagulation System with VisiTrax®

The EPi-Sense Guided Coagulation System with VisiTrax is intended for the coagulation of cardiac tissue using Radiofrequency (RF) energy using thoracoscopic, endoscopic, and laparoscopic surgical techniques.

The EPi-Sense Guided Coagulation System with VisiTrax may be used for temporary cardiac signal sensing and recording during surgery when connected to an external recording device.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

510(k) Summary:

Application Date:July 30, 2014
Sponsor:nContact Surgical, Inc.1001 Aviation Parkway, Suite 400,Morrisville, NC 27560
Establishment RegistrationNumber:3006142617
Correspondent:Jane RicuperoVP of Regulatory & Quality1001 Aviation Parkway, Suite 400,Morrisville, NC 27560
Contact Numbers:Phone: 919 655-1355Fax: 919 655-1690E-mail: jricupero@ncontactinc.com
Device Proprietary Name(s):EPi-Sense® Guided CoagulationSystem with VisiTrax®Model numbers: CDK-1411; CDK-1412; CDK-1413
Device Common Name:Electrosurgical device andaccessories
Device Classification:21 CFR 878.4400Class II
Product Code:OCL
Classification Name:Electrosurgical cutting andcoagulation device and accessories
Predicate DeviceEPi-Sense Guided CoagulationSystem with VisiTrax, Modelnumbers: CDK-1411; CDK-1412;CDK-1413
Legally Marketed unmodifieddevice 510k numberK120857 cleared Nov. 13, 2012

{4}------------------------------------------------

Device Description:

The EPi-Sense Guided Coagulation System with VisiTrax consists of a sterile, single-use, disposable coagulation electrode device (1cm, 2cm, & 3cm sizes provided) intended to be used to coagulate cardiac tissue. The flexible, cooled electrode device, with a suction stabilizer feature, transmits radiofrequency (RF) energy from an Electrosurgical Generator (non-sterile, re-useable) connected through an Instrument Cable (sterile). A temporary sensing electrode feature may be used with an off-the-shelf electrogram recording system with the use of an additional instrument cable (non-sterile). An accessory Cannula may be used to facilitate coagulation device access and visibility of the heart. The Cannula accessory changes are the subject of this submission.

Indications for Use:

Indications for Use: The EPi-Sense® Guided Coagulation System with VisiTrax® is intended for the coagulation of cardiac tissue using Radiofrequency (RF) energy using thoracoscopic, endoscopic, and laparoscopic surgical techniques.

The EPi-Sense® Guided Coagulation System with VisiTrax® may be used for temporary cardiac signal sensing and recording during surgery when connected to a temporary external recording device.

Description of Change for this Submission:

The following changes are being made to the Cannula Accessory for the EPi-Sense Guided Coaqulation System with VisiTrax.

  • Manufacturing process change for the cannula tip from an extrusion ● process to a molded process
  • Material/part vendor changes ●
  • Vacuum tubing wall thickness increase
  • Addition of blue lumen marker ●
  • Sterile barrier pouch sealing operation change of location .
  • Addition of two Cannula models

Comparison to Predicate Device and Summary of Technological Characteristics:

The nContact EPi-Sense System described in this submission is identical to the predicate nContact EPi-Sense System described in K120857 and has the identical intended use and technological characteristics.

  • · Coagulation Device functions, such as perfusion, suction, the RF electrode, and sensing electrodes are identical, with use of the same RF Generator and connecting cables.

{5}------------------------------------------------

  • The Cannula accessory used with the EPi-Sense device has minor ● changes as described above with design features that are the same or slightly modified from the previous Cannula design.

Summary of Nonclinical Testing for Accessory Cannula:

Performance bench tests were executed to ensure that the Cannula accessory performed as intended and met design specifications in conjunction with the EPi-Sense Guided Coagulation System with VisiTrax.

The EPi-Sense device incorporates design features and specifications identical to those specified by the predicate device, there were no changes made to the coagulation device. The Cannula accessory with minor changes has completed testing to verify that no safety or effectiveness issues arise when using the coagulation device in conjunction with the Cannula.

The following tests were successfully completed to evaluate equivalence.

    1. Biocompatibility testing per ISO 10993 for cannula accessory materials.
    1. Sterilization validation per ISO 11137-2, Sterilization of Health Care Products - Radiation - Method VD Max for cannula accessory.
    1. Reliability testing such as shipping and accelerated aging of packaged units.
    1. Tensile testing of critical bonds and joints.
    1. Flexion fatigue testing.
    1. Electrical integrity testing for the cannula accessory used with the coagulation device to pertinent sections of IEC 60601-1-2.

Substantial Equivalence Conclusion:

nContact concludes that the modifications for the Cannula Accessory used with the EPi-Sense Guided Coagulation System with VisiTrax may be considered substantially equivalent or identical to the legally marketed predicate devices based on the results of design verification and validation. The indications for use, principle of operation, technology, performance specifications (as re-verified through design controls), and labeling and sterilization parameters have no substantial changes or modifications that significantly affect the safety or efficacy of the devices. The Epi-Sense Guided Coagulation System with VisiTrax with Accessory Cannula is substantially equivalent to the stated predicate device.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.