K Number
K141931

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-10-09

(85 days)

Product Code
Regulation Number
866.3980
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Lyra™ Adenovirus Assay is a real-time polymerase chain reaction (PCR) in vitro diagnostic test for the qualitative detection of human adenovirus (HAdV) viral DNA isolated from nasal and nasopharyngeal swab specimens obtained from individuals exhibiting signs and symptoms of acute respiratory infections. The intended use of this test is to aid in the diagnosis of HAdV in conjunction with other clinical and laboratory findings.

The test detects, but does not differentiate. HAdV species (A, B, C, D, E, and F) or serotypes (HAdV 1-52).

Negative results do not preclude HAdV infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

Device Description

The Lyra™ Adenovirus Assay is a real-time polymerase chain reaction (PCR) in vitro diagnostic test for the qualitative detection of human adenovirus (HAdV) viral DNA isolated from nasal and nasopharyngeal swab specimens obtained from individuals exhibiting signs and symptoms of acute respiratory infections. The assay detects viral nucleic acids that have been extracted from a patient sample. A real-time PCR reaction is carried out using the ABI 7500 Fast Dx Real-Time PCR Instrument ("ABI Fast Dx Instrument") under optimized conditions in a single tube generating amplicons for adenovirus and the Process Control (PRC). Identification of human adenovirus (HAdV) and the PRC occurs by the use of target-specific primers and fluorescent-labeled probes that hybridize to conserved regions in the genomes of adenovirus and the PRC.

AI/ML Overview

The Lyra™ Adenovirus Assay is a real-time Polymerase Chain Reaction (PCR) in vitro diagnostic test for the qualitative detection of human adenovirus (HAdV) viral DNA from nasal and nasopharyngeal swab specimens. Its intended use is to aid in the diagnosis of HAdV in individuals showing symptoms of acute respiratory infections. The device detects HAdV species (A, B, C, D, E, and F) or serotypes (HAdV 1-52) but does not differentiate them.

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not explicitly state "acceptance criteria" with numerical thresholds. However, it presents performance data that would implicitly serve as the criteria for substantial equivalence. Based on the provided clinical study results, the implied acceptance criteria would likely be high sensitivity and specificity against the reference methods.

MetricAcceptance Criteria (Implied)Reported Device Performance (Prospective Study)Reported Device Performance (Retrospective Study)
SensitivityHigh100% (95% CI: 90.1% to 100%)N/A (Positive Percent Agreement reported instead)
SpecificityHigh95.5% (95% CI: 94.2% to 96.5%)N/A (Negative Percent Agreement reported instead)
Positive Percent AgreementHighN/A100% (95% CI: 87.5% to 100%)
Negative Percent AgreementHighN/A98.7% (95% CI: 93.1% to 99.8%)
LoD (e.g., HAdV A/31)Low concentration detected$8.00 x10^{-2}$ TCID50/mLN/A
InclusivityDetection of HAdV serotypes49 out of 52 serotypes detected experimentally; 3 serotypes by in silico analysisN/A
Cross-Reactivity/InterferenceNo interference/cross-reactivityNo interference or cross-reactivity observed with 57 organisms and 11 substancesN/A

2. Sample size used for the test set and the data provenance:

Prospective Study (Test Set 1):

  • Sample Size: 1241 fresh upper respiratory tract specimens initially, reduced to 1239 after excluding two invalid specimens.
  • Data Provenance: Multi-center, prospective study. The document does not specify the exact countries of origin, but generally, FDA submissions imply US-based clinical studies unless otherwise stated. Specimens were collected and transported to labs daily and tested within 72 hours.

Retrospective Study (Test Set 2):

  • Sample Size: 105 frozen upper respiratory tract specimens.
  • Data Provenance: Retrospective study conducted at Site 1 (a pediatric hospital in the Southwest United States).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

The document does not specify the number or qualifications of experts used to establish the ground truth for the clinical studies.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Prospective Study:

  • The ground truth (reference method) for the prospective study was a "composite method of direct fluorescent antibody staining (CDFA) followed by direct fluorescent antibody staining of the specimen (SDFA)."
  • A specimen was considered positive if either CDFA or SDFA was positive.
  • A specimen was considered negative if both CDFA and SDFA were negative.
  • This method effectively acts as an adjudication by combining two established diagnostic methods. There isn't a stated number of human readers for each method, but CDFA and SDFA are laboratory techniques with their own interpretation protocols, likely performed by trained laboratory personnel.

Retrospective Study:

  • The reference method for the retrospective study was an "additional FDA-cleared molecular device" (a comparator PCR assay). There is no mention of adjudication for this comparative analysis.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This device is an in vitro diagnostic (PCR-based assay) and not an AI-powered diagnostic tool for human readers. Therefore, an MRMC comparative effectiveness study involving human readers with/without AI assistance was not performed and is not relevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

The performance reported for the Lyra™ Adenovirus Assay (sensitivity, specificity, positive percent agreement, negative percent agreement) in both the prospective and retrospective studies represents the standalone performance of the device (algorithm/assay only) compared to the respective reference methods. This indicates the device's ability to detect HAdV without human interpretation affecting the outcome of the assay itself, although humans are involved in performing the test and interpreting the final qualitative result generated by the instrument.

7. The type of ground truth used:

  • Prospective Study: A "composite method" of direct fluorescent antibody staining (CDFA) of viral culture and direct fluorescent antibody staining of the specimen (SDFA). This is a laboratory-based reference method, essentially a combination of viral culture and expert interpretation of fluorescent antibody staining.
  • Retrospective Study: An "additional FDA-cleared molecular device" (a comparator FDA-cleared PCR assay).

8. The sample size for the training set:

The document does not explicitly describe a separate "training set" in the context of machine learning. For in vitro diagnostic assays like this, method development and optimization phases typically involve extensive internal testing and reagent optimization, which would be analogous to "training" certain parameters. However, the document provided focuses on the analytical and clinical validation studies.

9. How the ground truth for the training set was established:

As noted in point 8, a distinct "training set" with ground truth in the machine learning sense is not detailed. The development process of the Lyra™ Adenovirus Assay would have involved establishing its limits of detection (LoD) using quantified viral stocks (e.g., TCID50/mL), assessing inclusivity against known serotypes, and testing for cross-reactivity and interference against quantified organisms. These are established through recognized laboratory standards and methodologies rather than "ground truth" derived from expert consensus on patient cases in a training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is simple, clean, and easily recognizable.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 9, 2014

Quidel Corporation Ronald H. Lollar, Senior Director Clinical and Quality Affairs Diagnostic Hybrids, Inc. (Wholly Owned Subsidiary of Quidel) 2005 East State Street, Suite 100 Athens, OH 45701

Re: K141931

Trade/Device Name: Lyra™ Adenovirus Assay Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory viral panel multiplex nucleic acid assay Regulatory Class: II Product Code: OCC Dated: July 15, 2014 Received: July 16, 2014

Dear Mr. Lollar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tamara V. Feldblyum -S for

Sally Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141931

Device Name Lyra™ Adenovirus Assay

Indications for Use (Describe)

The Lyra™ Adenovirus Assay is a real-time polymerase chain reaction (PCR) in vitro diagnostic test for the qualitative detection of human adenovirus (HAdV) viral DNA isolated from nasal and nasopharyngeal swab specimens obtained from individuals exhibiting signs and symptoms of acute respiratory infections. The intended use of this test is to aid in the diagnosis of HAdV in conjunction with other clinical and laboratory findings.

The test detects, but does not differentiate. HAdV species (A, B, C, D, E, and F) or serotypes (HAdV 1-52).

Negative results do not preclude HAdV infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 201 Subpart D)
Over-The-Counter Use (21 CFR 330 Subpart D)

|X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Applicant:

Quidel Corporation 12544 High Bluff Drive, Suite 200 San Diego, California 92130 Telephone: 858-552-7910 Fax: 858-646-8045

Contact Information:

Ronald H. Lollar, Senior Director Clinical and Regulatory Affairs 2005 East State Street Suite 100 Athens, Ohio 45701 740-589-3300 – Corporate number 740-589-3373 – Desk phone 740-593-8437 – Fax lollar@dhiusa.com

Date of preparation of 510(k) summary:

July 15, 2014

Device Name:

Trade name - Lyra™ Adenovirus Assay Classification name - Respiratory viral panel multiplex nucleic acid assay Product Code - OCC Regulation Section - 21CFR 866.3980

Substantial Equivalency

The Lyra™ Adenovirus Assay used an FDA-cleared composite method of direct fluorescent antibody (DFA) staining of viral culture followed by DFA staining of the specimen as the reference comparator. The predicate device for the assay is the Adenovirus R-gene® US. The characteristics of the Lyra™ Adenovirus Assay ("Subject Device") and the legally marketed device are described in the Table below:

{4}------------------------------------------------

ItemSubject DeviceLyraTM AdenovirusAssayPredicate Device(K121942)Adenovirus R-geneUS Assay
Intended UseThe LyraTM AdenovirusAssay is a real-timepolymerase chainreaction (PCR) in vitrodiagnostic test for thequalitative detection ofhuman adenovirus(HAdV) viral DNAisolated from nasal andnasopharyngeal swabspecimens obtained fromindividuals exhibitingsigns and symptoms ofacute respiratoryinfection. The intendeduse of this test is to aidin the diagnosis ofHAdV in conjunctionwith other clinical andlaboratory findings.The test detects, but doesnot differentiate, HAdVspecies (A, B, C, D, E,and F) or serotypes(HAdV 1-52).Negative results do notpreclude HAdVinfection and should notbe used as the sole basisfor treatment or otherpatient managementdecisions.Adenovirus R-gene USis a Real Time PCR invitro diagnostic test forthe rapid andqualitative detection ofAdenovirus viral DNAisolated and purifiedfrom nasopharyngealspecimens (swab orwash/aspirate) obtainedfrom individualsexhibiting signs andsymptoms of acuterespiratory infection.The intended use forthis test is to aid in thediagnosis ofAdenovirus infectionsin humans.Negative results do notpreclude Adenovirusinfection and shouldnot be used as the solebasis for treatment orother managementdecisions.
DNAAmplificationTechnologyReal-time polymerasechain reaction (PCR)Same

{5}------------------------------------------------

ItemSubject DeviceLyra™ AdenovirusAssayPredicate Device(K121942)Adenovirus R-geneUS Assay
Assay ResultsQualitativeSame
Viral TargetSequencePenton base geneHexon gene
Sample TypesNasal andNasopharyngeal swabsNasopharyngeal swabsand Nasopharyngealaspirates/washes
AmplificationSelf-contained andautomatedSame
DetectionTechniquesSelf-contained andautomatedSame
Nucleic AcidExtractionMethodbioMérieux NucliSENSeasyMAG SystemSame
Collection andTransport MediaUniversal TransportMedium (Copan/DHI),MicroTest M4, M4-RT,M5, or M6 TransportMedium (Remel)Universal TransportMedium (DHI)MicroTest M4RTTransport (Remel)
Instrument/AssayPlatformApplied Biosystems(ABI) 7500 Fast DxReal-Time PCRInstrumentCepheid SmartCyclerII
ControlsIncluded withAssavInternal process control(MS2 Bacteriophage;Zeptometrix )Positive control(Plasmid DNA)Negative Control(molecular gradewater)Internal control (IC2)phage particle

{6}------------------------------------------------

Intended Use

The Lyra™ Adenovirus Assay is a real-time polymerase chain reaction (PCR) in vitro diagnostic test for the qualitative detection of human adenovirus (HAdV) viral DNA isolated from nasal and nasopharyngeal swab specimens obtained from individuals exhibiting signs and symptoms of acute respiratory infections. The intended use of this test is to aid in the diagnosis of HAdV in conjunction with other clinical and laboratory findings.

The test detects, but does not differentiate, HAdV species (A, B, C, D, E, and F) or serotypes (HAdV 1-52).

Negative results do not preclude HAdV infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions

Methodology

The assay detects viral nucleic acids that have been extracted from a patient sample. A real-time PCR reaction is carried out using the ABI 7500 Fast Dx Real-Time PCR Instrument ("ABI Fast Dx Instrument") under optimized conditions in a single tube generating amplicons for adenovirus and the Process Control (PRC). Identification of human adenovirus (HAdV) and the PRC occurs by the use of target-specific primers and fluorescent-labeled probes that hybridize to conserved regions in the genomes of adenovirus and the PRC.

Table 1. Lyra™ Probe Labels
TargetDye
HAdVFAM
PRCCy5

Performance Data

Precision/Reproducibility

Precision

The Precision/Within Laboratory Repeatability study was assessed using a panel of four (4) simulated samples that include moderate positive and low positive, high negative and negative. The four (4) member panel consisting of HAdV and negative samples were tested by two (2) operators, in triplicate for twelve (12) days.

Table 2. Results Summary – Cts and Positivity
PlatformTargetPos. ControlCt Values
5X LoD2X LoD0.5X LoDNeg. MatrixNeg. Control
ABI 7500Fast DxOperator 1 Avg Ct20.825.526.629.7NEGNEG
Operator 2 Avg Ct20.425.426.729.3NEGNEG

{7}------------------------------------------------

Quidel Corporation

Table 2.Results Summary – Cts and Positivity
PlatformTargetCt Values
Pos.Control5X LoD2X LoD0.5XLoDNeg.MatrixNeg.Control
Positivity100%100%100%100%0%0%

Section 05, 510(k) Summary

The Lyra™ Adenovirus Assay produces results that are highly reproducible.

Reproducibility

The reproducibility of the Lyra™ Adenovirus Assay was evaluated at three (3) laboratory sites (two external, one in-house). Reproducibility was assessed using a panel of four (4) simulated samples that include moderate positive and low positive, high negative and negative adenovirus samples. The panels and controls were processed and tested on the ABI 7500 Fast Dx Instrument. Panels and controls were tested at each site by 2 operators for 5 non-consecutive days (triplicate testing x 2 operators x 3 replicates x 5 days x 3 sites = 90 results per level for each virus). The LoD values were based on the values obtained in the LoD study.

Table 3.Reproducibility Results
PanelSite 1Site 2Site 3Combined Site Data
MemberIDRate ofDetectionAVECt%CVRate ofDetectionAVECt%CVRate ofDetectionAVECt%CVRate ofDetectionAVECt%CV
HAdVHighNegative(0.5x LoD)30/3029.04.930/3028.94.226/29*27.88.886/89*28.66.2
HAdVLowPositive(2x LoD)30/3026.72.330/3026.23.030/3026.19.390/9026.45.9
HAdVModeratePositive(5x LoD)30/3025.21.730/3024.83.030/3024.16.890/9024.84.8
HAdVNegativeNasalMatrix0/30N/AN/A0/30N/AN/A0/30N/AN/A0/90N/AN/A
HAdVPositiveControl30/3020.52.530/3018.23.930/3017.13.790/9018.68.5
HAdVNegativeControl0/30N/AN/A0/30N/AN/A0/30N/AN/A0/90N/AN/A

{8}------------------------------------------------

  • One aliquot was removed from the analysis because it had an invalid PRC result. Three aliquots (from the same replicate) tested negative.

The data from the combined sites indicates that the Lyra™ Adenovirus Assay, on the ABI 7500 Fast Dx Instrument, generates reproducible results for the detection of adenovirus and the internal control.

Limit of Detection (LoD)

The analytical sensitivity (limit of detection or LoD) of the Lyra™ Adenovirus Assay was determined using quantified (TCID50/mL) stocks of representatives of the six (6) species of adenovirus diluted in a negative matrix. Testing was performed with three manufactured device lots. Analytical sensitivity (LoD) is defined as the lowest concentration at which 95% of all replicates tested positive.

Table 4. Summary of LoD Study
Species/ SerotypeLoD(TCID50/mL)Avg. Ct Value
A/31$8.00 x10^{-2}$28.5
B/3$8.00 x10^{-2}$28.3
C/1$8.00 x10^{-2}$28.1
D/19$1.61 x10^{1}$28.6
E/4$1.00 x10^{0}$27.7
F/41$3.20 x10^{-2}$28.9

Analytical Reactivity (Inclusivity)

To verify the Lyra™ Adenovirus Assay detects the fifty-two (52) known serotypes of HAdV, 49 serotypes were initially tested at or near the species LoD (see above). Higher concentrations were tested if the organism was not detected at the LoD. Each of the 49 serotypes tested was detected by the Lyra Adenovirus Assay. Three (3) of the serotypes (38, 42, and 52) were unavailable for testing with the device. These serotypes were evaluated using in silico analysis.

SpeciesSerotypeConcentrationTested(TCID50/mL)Multiple ofLoDDetected
AHAdV-121.60 x10-12x LoD
AHAdV-181.60 x10-12x LoD
BHAdV-71.60 x10-12x LoD
SpeciesSerotypeConcentrationTested(TCID50/mL)MultipleLoDDetected
HAdV-111.60 x10-12x LoD
HAdV-141.60 x10-12x LoD
HAdV-161.60 x10-12x LoD
HAdV-212.00 x101250x LoD
HAdV-341.60 x10-12x LoD
HAdV-351.60 x10-12x LoD
HAdV-501.60 x10-12x LoD
HAdV-21.60 x10-12x LoD
CHAdV-51.60 x10-12x LoD
HAdV-61.60 x10-12x LoD
FHAdV-406.40 x10-22x LoD
HAdV-83.22 x10-12x LoD
HAdV-93.22 x10-12x LoD
HAdV-103.22 x10-12x LoD
HAdV-133.22 x10-12x LoD
HAdV-159.66 x1016x LoDb
HAdV-179.66 x1016x LoDc
HAdV-203.22 x10-12x LoD
HAdV-223.22 x10-12x LoD
HAdV-233.22 x10-12x LoD
HAdV-243.22 x10-12x LoD
DHAdV-253.22 x10-12x LoD
HAdV-263.22 x10-12x LoD
HAdV-273.22 x10-12x LoD
HAdV-283.22 x10-12x LoD
HAdV-293.22 x10-12x LoD
HAdV-303.22 x10-12x LoD
HAdV-323.22 x10-12x LoD
HAdV-333.22 x10-12x LoD
HAdV-363.22 x10-12x LoD
HAdV-373.22 x10-12x LoD
Table 5. HAdV Serotypes Reactivity Summary
SpeciesSerotypeConcentration(TCID50/mL)Multiple ofLoDDetected
HAdV-433.22 x10-12x LoD
HAdV-443.22 x10-12x LoD
HAdV-453.22 x10-12x LoD
HAdV-463.22 x10-12x LoD
HAdV-473.22 x10-12x LoD
HAdV-483.22 x10-12x LoD
HAdV-493.22 x10-12x LoD
HAdV-513.22 x10-12x LoD

{9}------------------------------------------------

{10}------------------------------------------------

4 HAdV-21 required repeat testing and was ultimately found to be detected at 250X LoD.

b HAdV-15 required repeat testing and was ultimately found to be detected at 6X LoD.

C HAdV-17 required repeat testing and was ultimately found to be detected at 6X LoD.

HAdV Species D serotypes 38 and 42 and Species G serotype 52 (often associated with conjunctivitis and gastroenteritis) were not evaluated for inclusivity as the sponsor could not obtain these strains. Instead, the manufacturer conducted an in silico analysis where they aligned their probe and primers with the target regions from a single sequence (obtained from GenBank) for each of these adenovirus types. The alignments showed a 100% agreement of the forward primer with types 38, 42, and 52. For the reverse primer, there were 3 mismatches detected on the sequences for types 38 and 42 (84% homology) and 2 mismatches with the sequence for type 52 (89.5% homology). For the probe, there was 100% agreement with types 38 and 42, while three mismatches were observed with the sequence for type 52 (85% homology).

Microbial Cross-Reactivity and Interference

The analytical specificity of the Lyra Adenovirus Assay was evaluated by testing a panel consisting of 30 viral, 26 bacterial, and 1 yeast strain representing respiratory pathogens or flora commonly present in the nasopharynx. The organisms were tested in the presence and absence of 2X LoD HAdV to determine if there was interference or cross-reactivity, respectively. Samples were extracted using the NucliSENS easyMAG instrument and tested in triplicate. All HAdV positive samples remained positive and all HAdV negative samples remained negative. indicating that there was no interference or cross-reactivity in the presence of the organisms tested. Each microorganism was tested three times, once in the presence of HAdV type 4, once in the presence of HAdV type 31, and once in the presence of negative nasal matrix.

{11}------------------------------------------------

Table 6. Organisms used for the study were the following:Concentration tested (TCID50/mL or CFU/mL)Lyra Adenovirus Assay Result
OrganismInterference (+ HAdV)Cross-reactivity (- HAdV)
Bordetella pertussis9.08E+08PositiveNegative
Bordetella bronchiseptica5.40E+08PositiveNegative
Chlamydophila pneumonia0.11 ng/ µLPositiveNegative
Chlamydia trachomatis2.10E+06PositiveNegative
Legionella pneumophila1.42E+09PositiveNegative
Mycobacterium intracellualre1.53E+09PositiveNegative
Mycobacterium tuberculosis9.30E+06PositiveNegative
Mycobacterium avium3.18E+09PositiveNegative
Mycoplasma pneumoniae3.16E+06PositiveNegative
Haemophilus influenzae4.00E+08PositiveNegative
Pseudomonas aeruginosa1.32E+09PositiveNegative
Proteus vulgaris6.53E+08PositiveNegative
Proteus mirabilis1.19E+09PositiveNegative
Neisseria gonorrhoeae1.40E+09PositiveNegative
Neisseria meningitidis1.29E+08PositiveNegative
Neisseria mucosa1.61E+08PositiveNegative
Klebsiella pneumoniae9.75E+08PositiveNegative
Escherichia coli1.13E+09PositiveNegative
Moraxella catarrhalis1.26E+09PositiveNegative
Corynebacterium diptheriae3.44E+08PositiveNegative
Lactobacillus plantarum3.18E+08PositiveNegative
Streptococcus pneumoniae1.43E+08PositiveNegative
Streptococcus pyogenes6.38E+08PositiveNegative
Streptococcus salivarius5.40E+08PositiveNegative
Staphylococcus epidermidis9.23E+08PositiveNegative
Staphylococcus aureus6.08E+08PositiveNegative
Candida albican9.70E+07PositiveNegative
Coronavirus 229E2.46E+07PositiveNegative
Coronavirus NL631.41E+04PositiveNegative
Coronavirus OC432.42E+07PositiveNegative
Coxsackievirus B42.00E+07PositiveNegative
Coxsackievirus B5/10/20063.62E+05PositiveNegative
Cytomegalovirus2.14E+06PositiveNegative
Table 6. Organisms used for the study were the following:
Lyra Adenovirus Assay Result
OrganismConcentration tested(TCID50/mL orCFU/mL)Interference(+ HAdV)Cross-reactivity(- HAdV)
Echovirus 67.60E+08PositiveNegative
Echovirus 74.45E+06PositiveNegative
Echovirus 92.17E+07PositiveNegative
Echovirus 112.17E+05PositiveNegative
Enterovirus 702.41E+06PositiveNegative
Enterovirus 712.03E+05PositiveNegative
Epstein Barr Virus9.27E+07PositiveNegative
HSV Type 1 MacIntyre Strain5.89E+06PositiveNegative
HSV Type 2 Strain G1.96E+06PositiveNegative
Human Metapneumovirus (A1)3.66E+05PositiveNegative
Human Rhinovirus 452.94E+04PositiveNegative
Human Rhinovirus 522.63E+04PositiveNegative
Influenza A/Mexico/4108/20094.08E+05PositiveNegative
Influenza A/Port Chalmers3.55E+08PositiveNegative
Influenza B/Florida/04/20061.54E+06PositiveNegative
Measles1.95E+06PositiveNegative
Mumps Virus2.75E+08PositiveNegative
Parainfluenza Type 13.97E+06PositiveNegative
Parainfluenza Type 23.15E+08PositiveNegative
Parainfluenza Type 32.36E+07PositiveNegative
Parainfluenza Type 4A1.04E+05PositiveNegative
RSV A (Long)4.36E+04PositiveNegative
RSV B (Wash/18537/62)3.43E+05PositiveNegative
Varicella Zoster Virus1.11E+04PositiveNegative

{12}------------------------------------------------

Interfering Substances

A study was performed on the ABI 7500 Fast Dx Instrument to evaluate the performance of the Lyra™ Adenovirus Assay in the presence of eleven (11) of potentially interfering/cross-reactive substances, at clinically relevant levels, that might be present in specimens. Each substance was tested in the presence of 2X LoD HAdV and with negative matrix.

{13}------------------------------------------------

Table 7.Interfering/Cross-reactive Substances Summary
HAdV
Substance NameSpecies ESerotype 4Species ASerotype 31No AnalyteInhibition(yes/no)
Ct Avg.SDCt Avg.SDCt Avg.SD
Controls26.00.627.60.9NEGN/AN/A
Mucin (Bovine SubmaxillaryGland, type I-S)26.70.827.10.2NEGN/ANo
Blood (human), EDTAanticoagulated28.01.227.60.6NEGN/ANo
Neo-Synephrine27.21.026.90.2NEGN/ANo
Afrin Nasal Spray26.20.826.70.1NEGN/ANo
Zicam Homeopathic Non-Drowsy Allergy Relief NoDrip Liquid Nasal Gel25.60.326.90.3NEGN/ANo
Saline Nasal Spray25.80.528.21.3NEGN/ANo
OTC Throat Lozenges:Ricola Action Cherry26.00.326.31.3NEGN/ANo
Zanamivir25.80.826.50.3NEGN/ANo
Tobramycin26.60.226.80.1NEGN/ANo
Mupirocin26.40.427.70.8NEGN/ANo
Oseltamivir phosphate26.00.027.20.2NEGN/ANo

None of the eleven (11) of substances tested cross-reacts with the Lyra™ Adenovirus Assay.

Carry-Over and Cross-Contamination

Studies were performed on the ABI 7500 Fast Dx Instrument using a 96-sample panel consisting of 48 high positive (1.0 x 10 TCID50 mL HAdV type 4) and 48 negative specimens (negative nasal matrix). The high positive samples were extracted and added to the plate in a checkerboard pattern that alternated with the negative samples. This testing was repeated over a 5-day period.

Over the course of 5 days, cross-contamination and amplicon carry-over did not occur with the Lyra™ Adenovirus Assay when extracted using the NucliSens easyMAG automated nucleic acid extraction instrument and analyzed on the ABI 7500 Fast Dx Instrument.

{14}------------------------------------------------

Method Comparison

Prospective Study

The evaluation of the Lyra™ Adenovirus Assay occurred in two separate studies: a prospective multi-center study using one thousand two hundred and forty-one (1241) fresh specimens from the upper respiratory tract; and a retrospective study using one hundred five (105) frozen specimens from the upper respiratory tract. In both studies the specimens were processed the bioMérieux NucliSENS® easyMag® at all sites for the extraction of nucleic acids from the clinical specimens. The ABI 7500 Fast Dx Instrument was used with the Quidel Lyra™ Adenovirus Assay for the amplification and detection of the target nucleic acids. The prospective specimens were also processed and tested with viral culture with an FDA-cleared composite method of direct fluorescent antibody staining (CDFA) followed by direct fluorescent antibody staining of the specimen (SDFA). A specimen was called positive if either the CDFA or the SDFA were positive. For a specimen to be called negative, it must be negative for both CDFA and SDFA. The retrospective specimens were extracted and tested with an additional FDAcleared molecular assay.

Age# TestedLyraAdenovirusAssay PositiveExpectedValueTotal Positiveby ReferenceMethodObservedPrevalence
< 173811.0%11.4%
1 to 54526013.3%245.3%
6 to 10157106.4%42.5%
11 to 156746.0%23.0%
16 to 215535.5%11.8%
> 2143740.9%30.7%
Total1241897.2%352.8%

Expected Values for the Winter of 2013 and 2014 (Combined)

One thousand two hundred forty-one (1241) fresh specimens were collected and transported to each laboratory for testing with the Lyra™ Adenovirus Assay. Two (2) specimens were invalid when initially tested with the Lyra™ Adenovirus Assay. The specimens were re-tested according to the instructions for use and were invalid upon repeat testing. The specimens have been removed from the data presented below. The specimens shipped daily with cold packs for CDFA and SDFA to the central location and were tested within 72-hours of collection. The table below details the HAdV results for the remaining one thousand two hundred thirty-nine (1239) specimens.

{15}------------------------------------------------

Combined Site Data
Comparator: CDFA with SDFA
LyraTM AdenovirusAssayPositiveNegativeTotal
Positive3554*89
Negative011501150
Total3512041239
95% CI
Sensitivity35/35100%90.1% to 100%
Specificity1150/120495.5%94.2 % to 96.5%
  • Forty-five (45) of the fifty-four (54) positives were positive by an additional PCR assay. Four (4) of the fifty-four (54) positives were negative by an additional PCR assay. Two (2) of the fifty-four (54) positives were invalid by an additional PCR assay. Three (3) of the fifty-four (54) positives had insufficient volume for testing by an additional PCR assay.

Retrospective Study

Due to the low prevalence of HAdV at the clinical sites during the study period, a retrospective study was conducted at Site 1 with specimens obtained from a pediatric hospital in the Southwest United States. One hundred five (105) frozen specimens from the upper respiratory tract were tested concurrently with the Lyra™ Adenovirus Assay and an additional FDA-cleared molecular device.

Adenovirus
Comparator: FDA-Cleared PCR Assay
Lyra™ Adenovirus AssayPositiveNegativeTotal
Positive271*28
Negative07777
Total2778105
95% CI
Positive Percent Agreement27/27100%87.5% to 100%
Negative Percent Agreement77/7898.7%93.1% to 99.8%
  • One (1) of one (1) positive was positive by a third FDA-cleared PCR assay.

Statement of Safety and Effectiveness

In prospectively collected clinical specimens, when performed on the ABI 7500 Fast Dx Instrument, the Lyra™ Adenovirus Assay vielded good sensitivity and specificity when compared to the composite reference method of direct specimen fluorescent antibody (SDFA) and viral culture with DFA (CDFA).

{16}------------------------------------------------

In retrospectively collected clinical specimens, when performed on the ABI 7500 Fast Dx Instrument, the Lyra™ Adenovirus Assay yielded good positive and negative percent agreement when compared to a FDA-cleared molecular device.

§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.

(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.