K Number
K141879

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-06-17

(341 days)

Product Code
Regulation Number
880.6880
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Belimed Steam Sterilizers MST-H, models 9-6-12, 9-6-15, and 9-6-18 are designed for sterilization of non-porous and porous heat and moisture-stable materials used in healthcare facilities.

Device Description

The Belimed Steam Sterilizers MST-H, models 9-6-12, 9-6-15, and 9-6-18 are available in a single door (HS1) or double door (HS2) version. The sterilizers are equipped with Pre-vacuum, Gravity, Liquid, Air Leak Test and Bowie-Dick Test cycles.

AI/ML Overview

The provided document describes the Belimed Steam Sterilizer MST-H, not an AI-powered device. Therefore, the questions related to AI-specific criteria (such as effect size of human readers with AI assistance, standalone algorithm performance, number of experts for ground truth, training set information, or a multi-reader, multi-case study) are not applicable.

The document details the acceptance criteria (performance requirements) for the steam sterilizer and summarizes the studies proving it meets these criteria through non-clinical validation.


1. A table of acceptance criteria and the reported device performance

The document presents performance criteria and verified results implicitly throughout the "EFFECTIVENESS" and "NONCLINICAL COMPARISON TO PREDICATE DEVICE" sections by referencing compliance with ANSI/AAMI ST8:2013 and specific outcomes. Here's a structured table:

Acceptance Criteria / Performance RequirementReported Device Performance (Belimed Steam Sterilizer MST-H)
Biological Performance (Sterility Assurance Level - SAL)SAL of 10⁻⁶ reduction (through achievement of no growth at half cycle with validation loads), complying with AAMI ST8:2013. Verified for PreVac 270 cycles (fabric), PreVac and Gravity instrument cycles, and liquid cycle.
Moisture Retention (Fabrics)<3% increase in pre-sterilization textiles test pack weight, and no wet spots. Complies with ANSI/AAMI ST8:2013 section 5.7.1.4.
Moisture Retention (Instruments)<20% increase in pre-sterilization towel weight of the test load, and no wet spots on the outer wrapper. Complies with ANSI/AAMI ST8:2013 section 5.7.2.4.
Empty Chamber Testing (Thermal Conditions)Demonstrated capability to provide steady-state thermal conditions within the chamber corresponding with the predicted SAL in the load. Meets temperature control requirements of ANSI/AAMI ST8:2013 Section 4.4.2.5 for PreVac, Gravity, and Liquid cycles.
Bowie-Dick Test Cycle PerformanceDemonstrated a uniform color change throughout the test sheet, and load temperature devices attained exposure temperature within 10 seconds of progressing into the exposure phase. Qualified according to ANSI/AAMI ST8:2013 section 5.6.1.
Liquid Cycle Performance (Temperature & Loss)Achieved sterility assurance level of 10⁻⁶ (no growth). Temperature above 121°C for at least 12 minutes. Loss of liquid did not exceed 70ml in a 1000ml bottle (unloading temperature 90°C), including with full load liquids. Verified according to ANSI/AAMI ST8:2013 section 5.5.3. Also note: Liquid cycle is not intended to sterilize liquids that are used for direct patient contact (stated in factory programmed cycle details, page 3).
Software ValidationPerformed according to FDA's moderate level of concern recommendations in "Guidance for the Content for Premarket Submissions for Software Contained in Medical Devices (May 2005)".
Compliance with StandardsDesigned, constructed, and tested to meet safety and performance requirements of standards including ANSI/AAMI ST8:2013, ANSI/AAMI ST79:2010 (+A1:2010 +A2:2011 +A3:2012 +A4:2013), IEC EN 62304:2006, EN ISO 14971:2012, EN ISO 13485:2003+AC2009, IEC EN 60601-1-2:2007, UL 61010-1:2004, IEC EN 61010-2-040:2005, and ASME Section VIII, Division 1 for unfired pressure vessels Ed. 2013.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The information provided does not specify a "test set" in the context of individual patient data or medical images, as this is a steam sterilizer for heat and moisture-stable materials, not a diagnostic or AI device.

Instead, the "test set" refers to the tested load configurations and biological indicator runs:

  • Sample Size: Not explicitly quantified as a single number, but refers to "validation loads," "single fabric test pack," "full load fabrics," "full load instruments trays (25lbs each)," "Bowie-Dick Test Pack," and "liquids in 1000ml in vented containers" (including "3 bottles" and "full load liquids"). The number of cycles performed for each validation is not specified in detail (e.g., "All PreVac 270 cycles were verified...").
  • Data Provenance: The studies were part of the device's premarket notification, implying they were conducted by or on behalf of the manufacturer (Belimed Sauter AG, based in Switzerland). The testing was done to meet U.S. FDA requirements and U.S. standards (ANSI/AAMI). The nature of the testing is prospective validation of the device's performance against established standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

This question is not applicable. For a steam sterilizer, "ground truth" is established by direct measurement of physical parameters (temperature, pressure, time) and the observed biological outcome (kill of biological indicators) as defined by established standards (e.g., AAMI ST8:2013). There is no "expert consensus" on diagnostic interpretations or image analysis. The "experts" involved would be microbiologists, engineers, and quality assurance personnel following standardized protocols.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This question is not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in human interpretive tasks (e.g., medical image reading) where disagreement among experts is possible. For a steam sterilizer, compliance is determined by objective physical and biological measurements against predefined criteria in standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. The device is a steam sterilizer and does not involve human readers or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable. The device is a steam sterilizer, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the steam sterilizer's performance is established by:

  • Biological Indicators: Achievement of a Sterility Assurance Level (SAL) of 10⁻⁶ reduction, meaning no growth from biological indicators after exposure to the sterilization cycle. This is a direct measure of microbial inactivation.
  • Physical Measurements: Verification of appropriate temperature and pressure profiles throughout the sterilization chamber during cycles, ensuring steady-state thermal conditions and proper air removal (Bowie-Dick test).
  • Moisture Retention: Quantitative measurement of weight increase in test loads and visual inspection for wet spots to ensure adequate drying.
  • Compliance with Standards: The ultimate "ground truth" is adherence to the specified performance requirements laid out in recognized national and international standards, primarily ANSI/AAMI ST8:2013.

8. The sample size for the training set

This question is not applicable. The device is a physical product (steam sterilizer), not a machine learning model; therefore, it does not have a "training set" in the AI sense. Its design and performance are based on engineering principles and validated through physical and biological testing.

9. How the ground truth for the training set was established

This question is not applicable, as there is no "training set" for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three human profiles facing right, connected by a flowing ribbon-like design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 17, 2015

Belimed Sauter AG Michael Hari Head of Medical Engineering Zelgstrasse 8 CH8583 Sulgen, Switzerland

Re: K141879

Trade/Device Name: Belimed Steam Sterilizer MST-H Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: II Product Code: FLE Dated: May 18, 2015 Received: May 20, 2015

Dear Mr. Hari,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Hari

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number:

K141879

Device Name: Belimed Steam Sterilizer MST-H

Indications for Use

The Belimed Steam Sterilizers MST-H, models 9-6-12, 9-6-15, and 9-6-18 are designed for sterilization of non-porous and porous heat and moisture-stable materials used in healthcare facilities.

The Belimed Steam Sterilizers MST-H, models 9-6-15, and 9-6-18 are available in a single door (HS1) or double door (HS2) version.

Type of Use (Select one or both, as applicable) □ Prescription Use (Part 21 CFR 801 Subpart D) 区 Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995. DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

The Belimed Steam Sterilizer MST-H is offered with the following factory-programmed predefined sterilization cycles:

CyclePre-treatmentSterilizing temperatureSterilizing Time 1)Drying time 2)Recommended load 3)
PreVac 2704S/30D5 pulses270 °F4 min30 minDouble wrapped instrumenttrays, max. weight of 25 lbsper tray, Fabric packs
PreVac 2704S/5D5 pulses270 °F4 min5 minFabric packs4)
PreVac 2753S/30D5 pulses275 °F3 min30 minDouble wrapped instrumenttrays, max. weight of 25 lbsper tray, Fabric packs
Bowie-DickTest5 pulses273 °F3.5 min1 minOne Bowie-Dick-Test-Pack
Gravity 27015S/30DPurge270 °F15 min30 minDouble wrapped instrumenttrays, with non-porousinstruments, max. weight of25 lbs per tray
Liquid 250 45SPurge250 °F45 min0 minLiquids, up to 1000 ml invented or open containers 5)
Air leak TestVacuum: 65mbarTest time: 15 min---Empty chamber
Warm up & AirLeak TestVacuum: 65mbarTest time: 15 min270 °F3 min3 minEmpty chamber
Warm up2 pulses270 °F3 min3 minEmpty chamber

Table 1: Factory programmed Sterilization cycles

Notes for Table 1:

    1. Only factory set sterilizing time should be utilized with medical devices intended for patient contact.
    1. Factory set drying time. These are the minimum validated drying times, which may be extended depending on load conditions and wraps.
    1. Recommended load: Refer to table 2 and 3. The load configurations listed in table 2 and 3 are those used during performance testing of the sterilizer models.
    1. Fabric load should be preconditioned between 68°F and at a relative humidity of at least 35% to 60% for at least 2 hours.
    1. Liquid cycle is not intended to sterilize liquids that are used for direct patient contact.

{4}------------------------------------------------

ModelInstrument trays,max. 25 Ibs eachFabric Packs,max. 3.3 lbs each
9-6-121632
9-6-152040
9-6-182448

The following tables show the recommended loads for the Belimed Steam Sterilizer MST-H:

Table 2: Recommended loads for non-porous and porous load cycles

ModelVolume of liquid in one containerNumber of containers
9-6-121000 ml168
9-6-151000 ml210
9-6-181000 ml252

Table 3: Recommended loads for liquid cycle 250 °F

The Belimed Steam Sterilizer MST-H is offered in the following models:

ModelConfigurationNet loadingcapacity (L)Chamber size(H x W x D) (mm)Overall Dimensions(H x W x D) (mm)
9-6-12 HS11 door6911080 x 660 x 13981970 x 1700 x 1714
9-6-12 HS22 doors6911080 x 660 x 13981970 x 1700 x 1714
9-6-15 HS11 door8641080 x 660 x 17061970 x 1700 x 2022
9-6-15 HS22 doors8641080 x 660 x 17061970 x 1700 x 2022
9-6-18 HS11 door9771080 x 660 x 20141970 x 1700 x 2330
9-6-18 HS22 doors9771080 x 660 x 20141970 x 1700 x 2330

Table 4: Models and dimensions

{5}------------------------------------------------

510(k) Summary

Date:June 10, 2015
Submitter's Name / Address:Belimed Sauter AGZelgstrasse 8CH- 8583 Sulgen / Switzerland
Contact Person:Michael HariEmail: michael.hari@belimed.chPhone: +41 71 644 86 37Fax: +41 71 644 85 11
Trade Name:Belimed Steam Sterilizer MST-H
Models:9-6-12, 9-6-15 and 9-6-18
Classification:Steam Sterilizer – Class II, as listed per 21 CFR 880.6880Product Code: FLE
Predicate Device:Belimed Steam Sterilizer TOP 5000Series 9-18 (K021223)

INTENDED USE:

The Belimed Steam Sterilizers MST-H, models 9-6-12, 9-6-15, and 9-6-18 are designed for sterilization of non-porous and porous heat and moisture-stable materials used in healthcare facilities.

DESCRIPTION OF DEVICE:

The Belimed Steam Sterilizers MST-H, models 9-6-12, 9-6-15, and 9-6-18 are available in a single door (HS1) or double door (HS2) version.

The sterilizers are equipped with Pre-vacuum, Gravity, Liquid, Air Leak Test and Bowie-Dick Test cycles.

{6}------------------------------------------------

The Belimed Steam Sterilizer MST-H is offered with the following factory-programmed predefined sterilization cycles:

CyclePre-treatmentSterilizingtemperatureSterilizingTime 1)Dryingtime 2)Recommended load3)
PreVac 2704S/30D5 pulses270 °F4 min30 minDouble wrapped instrumenttrays, max. weight of 25 lbsper tray, Fabric packs
PreVac 2704S/5D5 pulses270 °F4 min5 minFabric packs4)
PreVac 2753S/30D5 pulses275 °F3 min30 minDouble wrapped instrumenttrays, max. weight of 25 lbsper tray, Fabric packs
Bowie-DickTest5 pulses273 °F3.5 min1 minOne Bowie-Dick-Test-Pack
Gravity 27015S/30DPurge270 °F15 min30 minDouble wrapped instrumenttrays, with non-porousinstruments, max. weight of25 lbs per tray
Liquid 250 45SPurge250 °F45 min0 minLiquids, up to 1000 ml invented or open containers 5)
Air leak TestVacuum: 65mbarTest time: 15 min---Empty chamber
Warm up & AirLeak TestVacuum: 65mbarTest time: 15 min270 °F3 min3 minEmpty chamber
Warm up2 pulses270°F3 min3 minEmpty chamber

Table 1: Factory programmed Sterilization cycles

Notes for Table 1:

    1. Only factory set sterilizing time should be utilized with medical devices intended for patient contact.
    1. Factory set drying time. These are the minimum validated drying times, which may be extended depending on load conditions and wraps.
    1. Recommended load: Refer to table 2 and 3. The load configurations listed in table 2 and 3 are those used during performance testing of the sterilizer models.
    1. Fabric load should be preconditioned between 68°F and at a relative humidity of at least 35% to 60% for at least 2 hours.
    1. Liquid cycle is not intended to sterilize liquids that are used for direct patient contact.
ModelInstrument trays,max. 25 lbs eachFabric Packs,max. 3.3 lbs each
9-6-121632
9-6-152040
9-6-182448

The following tables show the recommended loads for the Belimed Steam Sterilizer MST-H:

Table 2: Recommended loads for non-porous and porous load cycles

{7}------------------------------------------------

ModelVolume of liquid in onecontainerNumber of containers
9-6-121000 ml168
9-6-151000 ml210
9-6-181000 ml252

Table 3: Recommended loads for liquid cycle 250 °F

The Belimed Steam Sterilizer MST-H is offered in the following models:

ModelConfigurationNet loadingcapacity (L)Chamber size(H x W x D) (mm)Overall Dimensions(H x W x D) (mm)
9-6-12 HS11 door6911080 x 660 x 13981970 x 1700 x 1714
9-6-12 HS22 doors6911080 x 660 x 13981970 x 1700 x 1714
9-6-15 HS11 door8641080 x 660 x 17061970 x 1700 x 2022
9-6-15 HS22 doors8641080 x 660 x 17061970 x 1700 x 2022
9-6-18 HS11 door9771080 x 660 x 20141970 x 1700 x 2330
9-6-18 HS22 doors9771080 x 660 x 20141970 x 1700 x 2330

Table 4: Models and dimensions

{8}------------------------------------------------

NONCLINICAL COMPARISON TO PREDICATE DEVICE:

The Belimed Steam Sterilizers MST-H, models 9-6-12, 9-6-15, and 9-6-18 are very similar to the predicate device.

A summary of the differences to the predicate device is included in Table 5 :

General SterilizerFeaturesBelimed Steam SterilizerTOP5000Series 9-18 (K021223)Belimed Steam SterilizerMST-HSubstantiallyEquivalent orDifferent
Intended UseTerminal Sterilization of non-porous and porous heat andmoisture-stabile materials used inhealthcare facilitiesTerminal Sterilization of non-porous and porous heat andmoisture-stabile materials usedin healthcare facilitiesSame
Operating PrincipleThe sterilizing agent of theBelimed Sterilizers MST-H issteam. Steam is made fromdemineralized water and istherefore non-toxic. Steam is thepreferred method to sterilizeporous an non-porous heatresistant materials and textiles inhospitals for more than 100 years.Preferred sterilizationtemperatures are 250 °F, 270 °Fand 275 °F.The sterilizing agent of theBelimed Sterilizers MST-H issteam. Steam is made fromdemineralized water and istherefore non-toxic. Steam is thepreferred method to sterilizeporous an non-porous heatresistant materials and textiles inhospitals for more than 100years. Preferred sterilizationtemperatures are 250 °F, 270 °Fand 275 °F.Same
Product codeFLEFLE
C.F.R: code880.6880 Steam Sterilizer880.6880 Steam Sterilizer
ClassIIII
Built according toStandardANSI/AAMI ST8:1994Hospital Steam SterilizersANSI/AAMI ST8:2013Hospital Steam SterilizersSubstantiallyEquivalent,(ST8:2013 usesheavier instrumentload and refers tonew standardversions)
Electrical safetystandardUL 3101-1IEC 61010-1 ed1, am1IEC 61010-2-041 de1UL IEC 61010-1: 2004,IEC 61010-2-040:2005SubstantiallyEquivalent
EMC conformityIEC / EN 61326:2001 / 2002IEC / EN 61326:2001 / 2002Same
Electrical supply3 Phase 208V, 60Hz3 Phase 208V, 60Hz
9-6-9 &9-6-12Max. current: 13AMax. current: 13ASame
9-6-15 & 9-6-18Max. current: 20AMax. current: 20A
Utility supply pressurerequirementSteamWaterCompressed air2.5 - 3 bar g2 – 5 bar g5 - 7 bar g2.5 - 3.5 bar g2 – 5 bar g5 – 7 bar gSubstantiallyEquivalent
Steam capacityrequirement:Model 9-6-9Model 9-6-12Model 9-6-15Model 9-6-18120 kg/h160 kg/h200 kg/h240 kg/hn/a100 kg/h120 kg/h170 kg/hLower,SubstantiallyEquivalent
Compressed aircapacity requirement:10 Nm³/h10 Nm³/hSame
Cooling waterrequirement:Model 9-6-9Model 9-6-12Model 9-6-15Model 9-6-182 m³/h2.5 m³/h3 m³/h3.5 m³/hn/a2.4 m³/h2.8 m³/h3.2 m³/hLower,SubstantiallyEquivalent
Control devices
Vacuum deviceModel 9-6-9, 9-6-12Model 9-6-15, 9-6-18Water-ring vacuum pump90 m³/h125 m³/hWater-ring vacuum pump90 m³/h125 m³/hSame
Additional gas ejector /sterilizer runs withmax. cooling watertemperaturesNo/80 °FYes, if optional cooling waterloop is not installed90 °FSubstantiallyEquivalent
Steam condenserShell-and-tube heat exchanger,stainless steelShell-and-tube heat exchanger,stainless steelSubstantiallyEquivalent
Air Filtration0,2µm, efficiency >99.5%>99.97% for 0.3-micron particles0,2µm, efficiency >99.5%>99.97% for 0.3-micron particlesSame
Control valvesfor steam, vacuum andcondensateStainless steel grade 316 Lpneumatically driven piston-valvesStainless steel grade 316 Lpneumatically driven piston-valvesSubstantiallyEquivalent
Control valves forcooling waterBrassElectric solenoid valvesFiber-reinforced plastic materialElectric solenoid valvesSubstantiallyEquivalent
Piping material gradeAISI 316LAISI 316LSame
Pressure vessel
Chamber and jacketmaterial gradeAISI 316LAISI 316LSame
Chamber constructionCubic welded structure with all-round heating ductsCubic welded structure with all-round heating ductsSame
Chamber designpressure39psi40psiSubstantiallyEquivalent
Safety valvesBronce, ASME approvedBronce, ASME approvedSame
Chamber designstandardASME Section VIII, Division 1, forunfired pressure vesselsASME Section VIII, Division 1, forunfired pressure vesselsSame
Door material gradeAISI 321AISI 321Same
Door openingSingle or double doorHorizontal door operationMotor drivenSingle or double doorHorizontal door operationMotor drivenSubstantiallyEquivalent
Door safety:Door safety barAdditional door lockingfixtureNumber of door limitswitchesYesNoTwo 'door closed' switches perdoorYesYesone 'door closed' switch per doorand one 'door closed &mechanically locked' switch perdoorSubstantiallyEquivalent
Door sealCompressed air activatedCompressed air activatedSame
Chamber sizesModel:9-6-99-6-129-6-159-6-18Control(H x W x D) in mm:1080x660x10401080x660x14001080x660x17001080x660x2000(H x W x D) in mm:Not available1080x660x13981080x660x17061080x660x2014SubstantiallyEquivalent
Control TechnologyPLCPLCsame
HMI Technologyloading sideLCD color / 5,7"TFT color 10"Different, largerdisplay
HMI Technologyunloading sideBT2-2004 LCD display /h x w = 35mm x 68mmLSC display /h x w=27mmx100mmSubstantiallyEquivalent
Printer TechnologyMatrix Dot technology42 char /line, paper width 57mmThermal technology42 char /line, paper width 57mmSubstantiallyEquivalent(same Informationon printout)
Process controlAutomatic through all phases of acycleAutomatic through all phases of acycleSame
Sterilization processcontrolPressure controlled by anabsolute pressure transmitterPressure controlled by anabsolute pressure transmitterSame
Sterilization processrecordingIndependent temperature andpressure sensors for recordingIndependent temperature andpressure sensors for recordingSame
Sterilization processmonitoringTemperature sensor for exposuretemperature monitoringTemperature sensor for exposuretemperature monitoringSame
Process alarmsDetected errors cause an acousticand visual alarm. Alarms arerecorded on the printer.At cycle end the door on theunloading side (double doormodels) can't be opened. Alarmsmust be manually acknowledgedon the load side operating panel.Detected errors cause anacoustic and visual alarm. Alarms arerecorded on the printer.At cycle end the door on theunloading side (double doormodels) can't be opened. Alarmsmust be manually acknowledgedon the load side operating panel.Same
Cycle phasesAll cycle phases and sterilizer status are monitored on display. Display of:-Chamber temperature & pressure-Process status-Door status-Alarms, if pendingAll cycle phases and sterilizer status are monitored on display. Display of:-Chamber temperature & pressure-Process status-Door status-Alarms, if pendingAdditional process status display for cycle state above the chamber (LED bar display)SubstantiallyEquivalent(Additionalinformation)
Cycle endAcoustic signal at cycle endCycle end Indication on operating panel display.Acoustic signal at cycle endCycle end Indication on operating panel display.Same
Performance:
Biological performanceSAL of 10-6 reductionSAL of 10-6 reduction
(half cycle)Through achievement of no growth at half cycle with validation loads.AAMI ST8:1994Through achievement of no growth at half cycle withvalidation loads.AAMI ST8:2013Same
Comply with
Moisture retentionFabrics:<3% increase in pre-sterilization textiles test pack weightFabrics:<3% increase in pre-sterilization textiles test pack weight
Comply withInstruments<20% increase in pre-sterilization towel weight of the test loadAAMI ST8: 1994Instruments<20% increase in pre-sterilization towel weight of the test loadAAMI ST8:2013Same
Factory predefinedcycles
Prevac 270°FYesYes
Sterilize temperature132°C / 270°F132°C / 270°F
Sterilize time4 minutes4 minutes
Dry time20 minutes30 minutes
No. of pre-vacuums45 (improved air removal)
Qualified according toAAMI ST8: 1994AAMI ST8:2013SubstantiallyEquivalent
Test LoadInstrument load in double wrapped trays, full loadInstrument load in double wrapped trays, full load
17lbs gross weight per tray25 lbs gross weight per tray
Load capacity (trays)model 9-6-9model 9-6-12model 9-6-15model 9-6-189121518n/a162024
Prevac 270°FYesYesSubstantiallyEquivalent
Sterilize temperature132°C / 270°F132°C / 270°F
Sterilize time4 minutes4 minutes
Dry time5 minutes5 minutes
No. of pre-vacuums45 (improved air removal)
Qualified according toAAMI ST8: 1994AAMI ST8:2013
Test LoadSingle fabric packSingle fabric pack and full load
Load capacity (packs)model 9-6-9model 9-6-12model 9-6-15model 9-6-1818303642n/a324048
Liquid 250°FYesYes
Sterilize temperature121°C / 250°F121°C / 250°F
Sterilize time45 minutes45 minutes
Qualified according toAAMI ST8:1994AAMI ST8:2013
Test LoadLiquids in 1000ml in ventedcontainers, 3 bottlesLiquids in 1000ml in ventedcontainers, 3 bottles and full load
Load capacitymodel 9-6-9model 9-6-12model 9-6-15model 9-6-18Number of containers126168210252Number of containersn/a168210252
Prevac 275°Fn/aYesDifferent
Sterilize temperature135°C / 275°F
Sterilize time3 minutes
Dry time30 minutes
No. of pre-vacuums5 (improved air removal)
Qualified according toAAMI ST8:2013
Test LoadInstrument load in doublewrapped, full load25 lbs gross weight per tray
Load capacitymodel 9-6-12model 9-6-15model 9-6-18162024
Gravity 270°Fn/aYes
Sterilize temperature132°C / 270°F
Sterilize time15 minutes
Dry time30 minutes
Qualified according toAAMI ST8:2013
Test LoadMetal load in double wrappedtrays, 25 lbs gross weight pertrayDifferent
Load capacitymodel 9-6-12model 9-6-15model 9-6-18162024
BOWIE-DICK TestYesYes
Sterilize temperature134°C134°C
Sterilize time3.5 minutes3.5 minutesSubstantially
Dry time3 minutes3 minutesEquivalent
No. of pre-vacuums45
Qualified according toAAMI ST8: 1994AAMI ST8:2013
Air leak test
Leak testYesYesSame
Qualified according toAAMI ST8: 1994AAMI ST8:2013
Warm-Up and leaktestYesYesSame
Qualified according toAAMI ST8: 1994 (air leak test)AAMI ST8:2013 (air leak test)
Express 270°FYes, Prevac cyclen/a
Sterilize temperature132°C / 270°F
Sterilize time4 minutes
Dry time3 minutesDifferent
Qualified according toAAMI ST8:1994
Test LoadSingle wrapped instrument traywith a single instrument
Gravity 270°FYes Flash cyclen/a
Sterilize temperature132°C / 270°F
Sterilize time3 minutes
Dry time1 minuteDifferent
Qualified according toAAMI-ST37:1996
Test LoadUnwrapped instrument tray with asingle instrument
Gravity 270°FYes Flash cyclen/a
Sterilize132°C / 270°F
temperature10 minutes
Sterilize timeminuteDifferent
Dry timeAAMI-ST37-1996
Qualifiedaccording toUnwrapped instrument tray with anon-porous multiple instruments,
Test Loadmax. weight of 17lbs

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

Table 5: Comparison to predicate device

CLINICAL DATA:

No clinical data is required for this device classification submission.

EFFECTIVENESS:

Efficacy of sterilizer function and exposure time recommendations are ultimately shown by complete kill of biological indicators and verifying an appropriate safety factor or sterility assurance level (SAL) of 10° reduction. Belimed Sauter AG validates its sterilization cvcles by recommended practices, standards and guidelines developed by various independent organizations such as the Association for Advancement of Medical Instrumentation (AAMI). Prior to release, Belimed Steam Sterilizer MST-H, models 9-6-12, 9-6-15, and 9-6-18 were validated to meet the requirements of ANSI/AAMI ST8:2013.

The results of the Belimed Steam Sterilizer MST-H, models 9-6-12, 9-6-15, and 9-6-18 verification studies demonstrate that the sterilizers perform as intended and are summarized as follows:

  • Empty chamber testing was performed as described in Section 5.4.2.5 of ANSI/AMMI ST8:2013, for the PreVac, Gravity and Liquid cycles. These cycles demonstrated the sterilizer's capability to provide steady state thermal conditions within the chamber that are corresponding with the predicted sterility assurance level (SAL) in the load. The sterilizer meets the temperature control requirements of Section 4.4.2.5 of ANSI/AAMI ST8.
  • All PreVac 270 cycles were verified using a single fabric test pack for biological ● performance, as described in Section 5.5.2 ANSI/AAMI ST8:2013 were qualified according to section 5.5.2.5 of ANSI/AAMI ST8:2013. These cycles demonstrated a sterility assurance level (SAL) of 10th through achievement of no growth at half cycle with validation loads. Moisture retention of less than 3 % increase in pre-sterilization test pack weight was demonstrated with a single fabric pack and with full load fabrics, and no wet spots were examined, according to section 5.7.1.4 of ANSI/AAMI ST8:2013.
  • . All PreVac and Gravity instrument cycles were qualified according to section 5.5.4 of ANSI/AAMI ST8:2013. The tests were performed with full load instruments travs. 25lbs each. Instead of cotton wrappers, FDA cleared wrappers (double wrapped) were used. These cycles demonstrated a sterility assurance level of 10° through achievement of no growth at half cycle with validation loads, moisture retention of less than 20 % increase in pre-sterilization weight of the towel, and exhibited no wet spots on the outer wrapper according to section 5.7.2.4 of ANSI/AAMI ST8:2013.
  • . The Bowie Dick Test cycle was verified using the Bowie-Dick Test Pack was qualified according to section 5.6.1 of ANSI/AAMI ST8:2013 and demonstrated a uniform color change throughout the test sheet (FDA cleared test indicator) and the load temperature devices attained the exposure temperature within 10 seconds of progressing into the exposure phase.

{15}------------------------------------------------

  • The liquid cycle was verified according to section 5.5.3 of ANSI/AAMI ST8:2013. . This cycle demonstrated a sterility assurance level of 10th through achievement of no growth with validation loads and the temperature was above 121°C for at least 12 minutes. The loss of liquid did not exceed 70ml in a 1000ml bottle (unloading temperature 90°C). Additional tests with full load liquids showed the same results.
  • The software validation for the cycle operation was performed according to FDA's moderate level of concern recommendations provided in the document "Guidance for the Content for Premarket Submissions for Software Contained in Medical Devices (May 2005)".

{16}------------------------------------------------

Standards:

Belimed Sauter AG's sterilizers including the Belimed Steam Sterilizers MST-H, models 9-6-12, 9-6-15, and 9-6-18 have been designed, constructed and tested to meet the safety and performance requirements of various national safety codes and standards. The Belimed Steam Sterilizer MST-H complies with the following safety standards:

    1. ANSI/AAMI ST8:2013
    1. ANSI/AAMI ST79:2010 +A1:2010 +A2:2011 +A3:2012 +A4:2013
    1. IEC EN 62304:2006
    1. EN ISO 14971:2012
    1. EN ISO 13485:2003+AC2009 Medical Devices - Quality management systems. Requirements for requlatory purposes
  • IEC EN 60601-1-2:2007 6.
  • UL 61010-1:2004 7.
    1. IEC EN 61010-2-040:2005
  • American Society of Mechanical Engineers (ASME), Section VIII, Division 1 for unfired 9. pressure vessels Ed. 2013

USER INFORMATION:

Belimed Sauter AG provides information to the user that is intended to insure safe and effective use of steam sterilization in its detailed Operating Instructions, Technical Manual and other labeling. Belimed Sauter AG also recommends the use and periodic review of the AAM steam sterilization standards to ensure further assurance of the safe and effective use of steam sterilization equipment in health care facilities.

CONCLUSION:

The Belimed Steam Sterilizers MST-H, models 9-6-12, 9-6-15, and 9-6-18 are substantially equivalent devices to that of the predicate device. There have been no substantial changes in technology and intended use to the predicate device series 9-18 (K021223). This steam sterilizer MST-H, models 9-6-12, 9-6-15, and 9-6-18 meet the applicable requirements of the applicable standards.

Based on the provided information in this premarket notification, it can be concluded that the subject device is substantial equivalent to the predicate device.

§ 880.6880 Steam sterilizer.

(a)
Identification. A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.(b)
Classification. Class II (performance standards).