K Number
K141826

Validate with FDA (Live)

Date Cleared
2015-04-02

(269 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

FlashPak® is a reusable rigid container system to be used during immediate use steam sterilization (IUSS or flash) by hospitals and healthcare facilities. It is intended to enable sterilization of the enclosed devices and prevent recontamination during immediate transport to the point of use. The container is compatible with gravity-displacement steam sterilization using a 10 minute cycle at 132° C for porous items with lumens or a 3 minute cycle at 132° C for nonporous items like routine metal instruments. The container is also compatible with pre-vacuum steam sterilization using a 4 minute cycle at 132° C for porous items with lumens or a 3 minute cycle for nonporous items like routine metal instruments. FlashPak is recommended for sterilization of lumens with the following limits: gravity-displacement (5.5mm inner diameter or larger and up to 184mm in length), pre-vacuum (1mm inner diameter or larger and up to 203mm in length).

FlashPak® Model/Maximum Instrument Load Weight Recommendation:

9020 w/ 9020-08 Basket - Gravity 3 pounds (1.36 Kg) - Pre-Vacuum 3 pounds (1.36 Kg)

9030 w/ 9030-08 Basket - Gravity 10 pounds (4.55 Kg) - Pre-Vacuum 10 pounds (4.55 Kg)

9040 w/ 9040-08 Basket - Gravity 14 pounds (6.35 Kg) - Pre-Vacuum 14 pounds (6.35 Kg)

9050 w/ 9050-08 Basket - Gravity 16 pounds (7.27 Kg) - Pre-Vacuum 16 pounds (7.27 Kg)

Device Description

The FlashPak Container System consists of a Radel® lid and base that together fully enclose a removable metal wire basket that holds items to be sterilized. A silicone gasket creates an air tight seal between the lid and base by means of metal latches. Two pressure actuated valves; one in the center of the lid and one in the center of the base, allow ingress and egress of steam sterilant into the container during the sterilization cycle and when closed form an airtight enclosure against separate silicone seals. The container comes in four sizes and is indicated for use with gravity-displacement and dynamic-air-removal sterilization modes recommended in ANSI/AAMI ST79 for immediate use steam sterilization (IUSS). Approved chemical indicators are recommended for use during each sterilization cycle.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the FlashPak® Sterilization Container System, based on the provided document:

This document is a 510(k) premarket notification for a medical device. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel effectiveness. Therefore, the "study" described is a validation of additional sterilization parameters rather than a comparative effectiveness study in the typical sense of AI/human reader performance.


1. Table of Acceptance Criteria and Reported Device Performance:

The primary acceptance criteria for this submission are based on the successful demonstration of sterilization efficacy for the newly added sterilization cycles. The existing acceptance criteria for the device's physical and chemical properties, material compatibility, and microbial barrier properties are implicitly met because the device itself is physically identical to the previously cleared predicate.

Acceptance Criteria (for added sterilization cycles)Reported Device Performance
Sterilization Efficacy (Gravity-displacement Cycle - 3 min):No microbial growth observed in three half-cycle runs. Achieved SAL of 10⁻⁶.
Sterilization Efficacy (Dynamic-air-removal Cycle - 3 min):No microbial growth observed in three half-cycle runs. Achieved SAL of 10⁻⁶.
Sterilant Penetration:Three thermal mapping studies for each of the two sterilization modes demonstrating a precise match (within seconds) between the chamber and container temperature profiles.
Microbial Barrier (Packaging Integrity):The current container design and its performance were deemed acceptable, as it's the same as the predicate device which passed a microbial aerosol challenge test.

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set: Biological indicators (BIs) with Geobacillus Stearothermophilus were used for efficacy testing. The document states "Half cycles were run three times for each sterilization mode, gravity displacement and dynamic-air removal." This means a total of 6 half-cycle runs were performed (3 for gravity-displacement, 3 for dynamic-air-removal). The number of BIs per run is not explicitly stated, but they were placed "throughout the container among the devices and four devices were directly inoculated."
  • Data Provenance: The study was conducted by Symmetry Surgical, Inc. (the manufacturer) for their device. The provenance is internal, prospective testing conducted to validate the new sterilization parameters. No country of origin is specified for the data itself, but the company is located in Antioch, TN, USA.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

  • Number of Experts: Not applicable in the context of this device and study. The "ground truth" for sterilization efficacy is the sterility assurance level (SAL) achieved, which is a microbiological outcome determined through standardized laboratory testing, not expert consensus.
  • Qualifications of Experts: N/A.

4. Adjudication Method for the Test Set:

  • Adjudication Method: Not applicable. Sterilization efficacy is an objective outcome based on microbial growth, not a subjective interpretation requiring adjudication. Positive and negative controls were concurrently processed and passed, indicating proper test execution.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • MRMC Study: No. This device is a sterilization container, not a diagnostic imaging device or an AI-powered tool that assists human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Standalone Study: Not applicable. This is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used:

  • Ground Truth: Microbiological outcome data, specifically the absence of microbial growth on biological indicators after sterilization and incubation, demonstrating a Sterility Assurance Level (SAL) of 10⁻⁶. Thermal mapping data provided evidence of sterilant penetration.

8. The Sample Size for the Training Set:

  • Training Set Sample Size: Not applicable. This is a physical device; there is no "training set" in the context of machine learning. The device was developed and validated through engineering design, material science, and microbiological testing, not machine learning.

9. How the Ground Truth for the Training Set Was Established:

  • Ground Truth Establishment for Training Set: Not applicable. As there is no training set, there is no corresponding ground truth to establish in that context.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three faces in profile, stacked on top of each other. The faces are black and are oriented to the right. The logo is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in black.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 2, 2015

Symmetry Surgical, Inc. Mr. Christopher Smith, RAC Senior Director of QA/RA 3034 Owen Drive Antioch, TN 37013

Re: K141826

Trade/Device Name: FlashPak® Sterilization Container System Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: February 18, 2015 Received: February 23, 2015

Dear Mr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Smith

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141826

Device Name FlashPak® Sterilization Container System

Indications for Use (Describe)

FlashPak® is a reusable rigid container system to be used during immediate use steam sterilization (IUSS or flash) by hospitals and healthcare facilities. It is intended to enable sterilization of the enclosed devices and prevent recontamination during immediate transport to the point of use. The container is compatible with gravity-displacement steam sterilization using a 10 minute cycle at 132° C for porous items with lumens or a 3 minute cycle at 132° C for nonporous items like routine metal instruments. The containle with pre-vacum steam sterilization using a 4 minute cycle at 132° C for porous items with lumens or a 3 minute cycle for nonporous items like routine metal instruments. FlashPak® is recommended for sterilization of lumens with the following limits: gravity-displacement (5.5mm inner diameter or larger and up to 184mm in length), pre-vacuum (1mm inner diameter or larger and up to 203mm in length).

FlashPak® Model/Maximum Instrument Load Weight Recommendation:

9020 w/ 9020-08 Basket - Gravity 3 pounds (1.36 Kg) - Pre-Vacuum 3 pounds (1.36 Kg)

9030 w/ 9030-08 Basket - Gravity 10 pounds (4.55 Kg) - Pre-Vacuum 10 pounds (4.55 Kg)

9040 w/ 9040-08 Basket - Gravity 14 pounds (6.35 Kg) - Pre-Vacuum 14 pounds (6.35 Kg)

9050 w/ 9050-08 Basket - Gravity 16 pounds (7.27 Kg) - Pre-Vacuum 16 pounds (7.27 Kg)

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

区 Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

រៀត

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Symmetry Surgical. The logo consists of a stylized letter "S" made up of red and gray shapes, followed by the text "symmetry surgical" in a gray, sans-serif font. The logo is simple and modern, with a focus on the company name.

510K Summary of Safety and Effectiveness (Per 21 CFR 807.92)

General Company Information

Company Name:Symmetry Surgical
Company Address:3034 Owen Drive
Antioch, TN 37013
Company Telephone:615-964-5392
Contact:Christopher M. Smith
Sr. Director of QA/RA
Contact Address & Telephone:3034 Owen Drive
Antioch, TN 37013
615-964-5392
Date:April 2, 2015
Device Trade Name:FlashPak® Sterilization Container System
Common Name:Wrap, Sterilization
Classification, Regulation # andProduct Code:Class II - 880.6850 - FRG
Predicate Device:FlashPak® Sterilization Container System, K113776

Image /page/3/Picture/5 description: The image is a solid red color. The shape is a rectangle that is slightly distorted. The right side of the rectangle is curved inward.

3034 Owen Drive, Antioch, TN 37013 Office 800.251.3000 615,964,5567 Fax = symmetrysurgical.com

Image /page/3/Picture/7 description: The image is a close-up shot of a dark, textured surface. The surface appears to be made of a rough material, possibly stone or concrete. The lighting is dim, which accentuates the texture and creates shadows. The left side of the image has a curved edge, while the rest of the image is more rectangular.

{4}------------------------------------------------

ymmetry surgical

Device Description:

The FlashPak Container System consists of a Radel® lid and base that together fully enclose a removable metal wire basket that holds items to be sterilized. A silicone gasket creates an air tight seal between the lid and base by means of metal latches. Two pressure actuated valves; one in the center of the lid and one in the center of the base, allow ingress and egress of steam sterilant into the container during the sterilization cycle and when closed form an airtight enclosure against separate silicone seals. The container comes in four sizes and is indicated for use with gravity-displacement and dynamic-air-removal sterilization modes recommended in ANSI/AAMI ST79 for immediate use steam sterilization (IUSS). Approved chemical indicators are recommended for use during each sterilization cycle.

Intended Use:

The intended use of the predicate device was changed to add additional steam sterilizations. The 3 minute gravity displacement cycle at 132 ºC and 3 minute dynamic-air removal cycle at 132 ºC were validated and added to the following intended use statement.

"FlashPak® is a reusable rigid container system to be used during immediate use steam sterilization (IUSS or flash) by hospitals and healthcare facilities. It is intended to enable sterilization of the enclosed devices and prevent recontamination during immediate transport to the point of use. The container is compatible with gravitydisplacement steam sterilization using a 10 minute cycle at 132° C for porous items with lumens or a 3 minute cycle at 132° C for nonporous items like routine metal instruments. The container is also compatible with pre-vacuum steam sterilization using a 4 minute cycle at 132° C for porous items with lumens or a 3 minute cycle for nonporous items like routine metal instruments. FlashPak is recommended for sterilization of lumens with the following limits: gravity-displacement (5.5mm inner diameter or larger and up to 184mm in length), pre-vacuum (1mm inner diameter or larger and up to 203mm in length).

FlashPak® Model/Maximum Instrument Load Weight Recommendation:

9020 w/9020-08 Basket - Gravity 3 pounds (1.36 Kg) - Pre-Vacuum 3 pounds (1.36 Kg)
9030 w/9030-08 Basket - Gravity 10 pounds (4.55 Kg) - Pre-Vacuum 10 pounds (4.55Kg)
9040 w/9040-08 Basket - Gravity 14 pounds (6.35 Kg) - Pre-Vacuum 14 pounds (6.35 Kg)
9050 w/9050-08 Basket - Gravity 16 pounds (7.27 Kg) - Pre-Vacuum 16 pounds (7.27 Kg)

Substantial Equivalence Discussion

The predicate device is the FlashPak Sterilization Container System cleared under 510(k) #K113776. The predicate device and the subject device are physically identical. They are the same device. The differences between the predicate and the subject device involve the steam sterilization parameters recommended in the instructions for use (IFU) and intended use statement. The subject device has additional immediate use steam sterilization parameters recommended for its use that were taken from ANSI/AAMI ST79 and were validated for efficacy.

Image /page/4/Picture/10 description: The image is a solid red color. The shape is rectangular, but the right side is slightly curved. The red color is uniform throughout the image.

34 Owen Drive, Antioch, TN 37013 615 964 5567 symmetry surgical.com

Image /page/4/Picture/12 description: The image is a close-up shot of a dark, textured surface. The surface appears to be slightly curved on the left side, creating a rounded edge. The texture is grainy and uneven, with subtle variations in shading that suggest depth and dimension. The overall impression is one of a solid, unyielding material with a rough, natural finish.

{5}------------------------------------------------

etry surgical"

A comparison of the similarities and differences between the predicate device and the subject device is provided in the table below. The table follows the elements of interest defined in the guidance for Industry and FDA—Premarket Notification [510(k)] Submissions for Medical Sterilization Packaging Systems in Health Care Facilities of March 7, 2002.

ELEMENTPREDICATESUBJECT DEVICE
Intended UseFlashPak® is a reusable rigidcontainer system to be used duringflash sterilization by hospitals andhealthcare facilities. It is intendedto enable flash sterilization of theenclosed devices and preventrecontamination during immediatetransport to the point of use. Thecontainer is compatible withgravity-displacement flashsterilization using a 10 minute cycleat 132° C and with pre-vacuumflash sterilization using a 4 minutecycle at 132° C. FlashPak isrecommended for surfacesterilization of stainless steelinstruments and for lumens withthe following limits: gravity-displacement (5.5mm innerdiameter or larger and up to184mm in length), pre-vacuum(1mm inner diameter or larger andup to 203mm in length).FlashPak® is a reusable rigidcontainer system to be used duringimmediate use steam sterilization(IUSS or flash) by hospitals andhealthcare facilities. It is intendedto enable sterilization of theenclosed devices and preventrecontamination during immediatetransport to the point of use. Thecontainer is compatible withgravity-displacement steamsterilization using a 10 minute cycleat 132° C for porous items anditems with lumens or a 3 minutecycle at 132° C for nonporous itemslike routine metal instruments. Thecontainer is also compatible withpre-vacuum steam sterilizationusing a 4 minute cycle at 132° C forporous items and items withlumens or a 3 minute cycle fornonporous items like routine metalinstruments. FlashPak isrecommended for sterilization oflumens with the following limits:gravity-displacement (5.5mm innerdiameter or larger and up to184mm in length), pre-vacuum(1mm inner diameter or larger andup to 203mm in length).
Container:Radel R5000 series base and lid;SS (304) Instrument Basket;
ELEMENTPREDICATESUBJECT DEVICE
Material CompositionSS (304) Latches;Same as predicate
Clear Medical Grade Silicone Gasketand Valve Seal;
SS (300) Hardware;
Sintered SS Filter Disk;
Silicone Lid Vent Button
Valve:
SS (304) Bellows;
SS (316) Bracket
Physical PropertiesMaterials used in the FlashPak arecommonly used in many otherlegally marketed sterilization traysand cases. Physical properties ofthe system materials andconstruction have a demonstratedlong history of safe and effectiveuse for this application.Same as predicate
Chemical PropertiesMaterials used in the FlashPak arecommonly used in many otherlegally marketed sterilization traysand cases. Chemical properties ofthe system materials andconstruction have a demonstratedlong history of safe and effectiveuse for this application.Same as predicate
Configurations/DimensionsContainer Sizes:(11" x 11" x 7") #9020(20" x 11" x 8"), #9030(24" x 11" x 9") #9040(24" x 13" x 9") #9050Same as predicate
ELEMENTPREDICATESUBJECT DEVICE
Air PermeanceN/AContainer operates with valves, nofilters, wraps or perforationsN/A
Percent of SurfacePerforationsN/AContainer is not perforatedN/A
PERFORMANCEPREDICATENEW DEVICE
Sterilant PenetrationSterilization EfficacyGravity-displacement Cycle:10 minutes at 132°C(270°F) for porous and cannulated devicesDynamic-air-removal Cycle4 minutes at 132°C(270°F) for porous and cannulated devicesSame as predicate with the added indications for:Gravity-displacement Cycle:3 minutes at 132°C(270°F) for nonporous and routine metal devicesDynamic-air-removal Cycle3 minutes at 132°C(270°F) for nonporous and routine metal devices
Sterilant PenetrationThree thermal mapping studies for each of the two sterilization modes demonstrating a precise match (within seconds) between the chamber and container temperature profiles.
Microbial Barrier Properties(Packaging Integrity)The container completely encapsulates the items sterilized inside with the use of a silicone lid gasket, a silicone lid vent, sintered SS microfilter and silicone valve seals to provide an effective microbial barrier during transport to the point of use immediatelySame as predicate
ELEMENTPREDICATESUBJECT DEVICE
after sterilization.
Container passed a microbial aerosol challenge test by preventing microbial ingress during a one hour exposure to aerosolized microbes in a closed chamber.
Material CompatibilityMaterials used in the FlashPak are commonly used in many other legally marketed sterilization trays and cases. The system materials and construction have a demonstrated long history of safe and effective use for this application.Same as predicate
Toxicological Properties(Biocompatibility, includingSterilant Residue Limits)Materials used are either medical grade polymers or stainless steel. Instruments placed in the container only come in contact with the Stainless Steel wire basket. Cytotoxicity and Intracutaneous Reactivity studies were completed on the final configuration.Same as predicate
N/A
Shelf LifeFlashPak is intended for immediate use steam sterilizationSame as predicate
N/A
Drying TimeFlashPak is intended for immediate use steam sterilizationSame as predicate
Aeration Time & EtOResidualsN/AFlashPak is only recommended for steam sterilizationN/A

Image /page/5/Picture/4 description: The image is a close-up shot of a red parallelogram. The parallelogram is a solid color, with no visible texture or pattern. The shape is slightly distorted, with the right side curving inward. The lighting is even, and there are no shadows or highlights.

3034 Owen Drive, Antioch, TN 37013 Office 800.251.3000 615 964.5567 Fax symmetrysurgical.com

Image /page/5/Picture/6 description: The image is a close-up shot of a dark, textured surface. The surface appears to be slightly curved on the left side, and the texture is rough and uneven. The lighting is dim, which makes it difficult to discern the details of the surface. The overall impression is one of a weathered, aged material.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Symmetry Surgical. The logo consists of a stylized "S" made up of gray and red shapes, followed by the words "symmetry surgical" in gray sans-serif font. The word "surgical" is slightly smaller than the word "symmetry".

Image /page/6/Picture/5 description: The image is a close-up shot of a solid red color on the left side, gradually transitioning to white on the right. The red portion occupies a significant part of the image, while the white area appears as a curved shape on the right edge. The transition between the two colors is smooth, creating a visually simple yet striking composition.

3034 Owen Drive, Antioch, TN 37013 Office 800.251.3000
Fax symmetrysurgical.com

Image /page/6/Picture/4 description: The image is a close-up shot of a dark, solid surface. The surface appears to be slightly textured, with subtle variations in tone and shade. The lighting is somewhat diffused, creating a soft, even illumination across the surface. The overall impression is one of simplicity and understated texture.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Symmetry Surgical. The logo consists of a stylized "S" symbol on the left, followed by the text "symmetry surgical" in a sans-serif font. The "S" symbol is made up of red and gray shapes.

Image /page/7/Picture/2 description: The image is a solid red color. The shape is a rectangle that is slightly curved on the right side. The red color is consistent throughout the image.

3034 Owen Drive, Antioch, TN 37013 Qffice 800 251.3000
Fax symmetrysurgical.com

Image /page/7/Picture/4 description: The image is a close-up shot of a dark, textured surface. The surface appears to be slightly curved on the left side, and the texture is somewhat rough and uneven. The overall tone of the image is dark and muted, with subtle variations in shading across the surface.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for Symmetry Surgical. The logo consists of two stylized letter 'S' shapes, one in red and one in gray, positioned next to each other. To the right of the logo is the text "symmetry surgical" in a gray, sans-serif font.

Image /page/8/Picture/2 description: The image is a solid red rectangle. The rectangle is slightly distorted, with the top and bottom edges not perfectly parallel. The color is a bright, saturated red, and there are no other colors or objects in the image.

3034 Owen Drive, Antioch, TN 37013 Office 800.251.3000
Fax symmetrysurgical.com

Image /page/8/Picture/4 description: The image is a close-up shot of a dark gray or black surface. The surface appears to be textured or slightly rough, with subtle variations in tone and shading. The lighting is soft and diffused, creating a smooth and even illumination across the surface. The overall impression is one of simplicity and understated elegance.

6 Page

{9}------------------------------------------------

ymmetry surgical

Technological Characteristics:

The technological characteristics of the subject device are identical to the predicate since there have been no physical changes made to the device. The materials, manufacturing and fundamental scientific technology are the same.

Performance Data:

The sterilization parameters tested were taken from ANSI/AAMI ST79 for IUSS. Testing consisted of sterilization efficacy studies for the additional sterilizations. The over kill test method described in ISO 17665-1 was applied. A worst case loading configuration was utilized; i.e. maximum recommended weight of routine metal instruments.

FlashPak ModelMaximum Instrument Load Weight Recommendation
GravityPre-Vacuum
9020w/ 9020-08 Basket3 pounds (1.36 Kg)3 pounds (1.36 Kg)
9030w/ 9030-08 Basket10 pounds (4.55 Kg)10 pounds (4.55 Kg)
9040w/ 9040-08 Basket14 pounds (6.35 Kg)14 pounds (6.35 Kg)
9050w/ 9050-08 Basket16 pounds (7.27 Kg)16 pounds (7.27 Kg)

Biologic indicators (Bl) with resistant Geobacillus Stearothermophilus were placed throughout the container among the devices and four devices were directly inoculated. Half cycles were run three times for each sterilization mode, gravity displacement and dynamic-air removal. Following sterilization the Bls were aseptically removed and processed to culture media then incubated to observe for any microbial growth. In every run there was no observed microbial growth indicating an SAL of 10* was achieved. Positive and negative controls were concurrently processed and passed as well.

Conclusion: Performance testing data demonstrates the the FlashPak® Sterilization Container is substantially equivalent to the predicate device.

Image /page/9/Picture/9 description: The image is a solid red color. The shape is a quadrilateral, but the right side is curved. The red color is consistent throughout the image.

034 Owen Drive, Antioch, TN 37013 800 251 3000 615 964 5567 symmetrysurgical.com

Image /page/9/Picture/11 description: The image is a close-up shot of a dark, textured surface. The surface appears to be slightly curved on the left side, creating a concave shape. The texture is rough and uneven, with variations in color and tone throughout the surface. The overall impression is that of a solid, unyielding material with a natural or weathered appearance.

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).