K Number
K141803

Validate with FDA (Live)

Date Cleared
2015-02-12

(224 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Immunalysis Tramadol Urine Enzyme Immunoassay:

The Immunalysis Tramadol Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 200ng/ mL. The assay is intended for use in laboratories for the qualitative and semi-quantitative analysis of Tramadol in human urine with automated clinical chemistry analyzers. This assay is calibrated against Tramadol. This in-vitro diagnostic device is for prescription use only.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GC-MS or permitting laboratories to establish quality control procedures. The test is not intended to differentiate between drugs of abuse and prescription use of Tramadol. There are no uniformly recognized drug levels for Tramadol in urine.

The Immunalysis Tramadol Urine Enzyme Immunoassay Kit provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography / Mass Spectroscopy (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Immunalysis Tramadol Urine Controls:

The Immunalysis Tramadol Urine Controls are used as control materials in the Immunalysis Tramadol Urine Enzyme Immunoassay.

Immunalysis Tramadol Urine Calibrators:

The Immunalysis Tramadol Urine Calibrators are used as calibrators in the Immunalysis Tramadol Urine Enzyme Immunoassay for the qualitative and semi-quantitative determination of Tramadol in urine on automated clinical chemistry analyzers.

Device Description

The assay consists of antibody/ substrate reagent and enzyme conjugate reagent. The antibody/ substrate reagent includes goat antibodies to Tramadol, glucose-6-phosphate (G6P) and nicotinamide adenine dinucleotide (NAD) in Tris buffer with Sodium Azide as a preservative. The enzyme conjuqate reagent includes tramadol derivative labeled with qlucose-6-phosphate dehydrogenase (G6PDH) in Tris buffer with Sodium Azide as a preservative. Calibrators and controls are sold separately. Reagents are liquid, ready to use

The tramadol calibrator and controls consists of a single calibrator at 200ng/mL, a control set containing a LOW control at 150ng/mL and a HIGH control at 250ng/mL and a calibrator set containing a negative calibrator, a Level 1 calibrator at 100ng/mL, a Level 2 calibrator at 200ng/mL, a Level 3 calibrator at 500ng/mL and a Level 4 calibrator at 1000nq/mL.

AI/ML Overview

The document describes the Immunalysis Tramadol Urine Enzyme Immunoassay and its performance characteristics. The study aimed to demonstrate substantial equivalence to a predicate device.

Here's the breakdown of the acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined "acceptance criteria" in a separate table. Instead, it presents study results which implicitly serve as performance metrics that demonstrate the device's acceptable function. The "Result" column in the tables below indicates the device's performance against expected outcomes at various concentrations/conditions.

StudyAcceptance Criteria (Implicit from Study Design)Reported Device Performance (Summary)
Qualitative Analysis (200ng/mL cutoff)Correct classification of samples relative to the 200ng/mL cutoff.At -100% to -25% of cutoff (0-150 ng/mL): 80/80 Negative. At cutoff (200 ng/mL): 44 Negative / 36 Positive (demonstrating boundary function). At +25% to +100% of cutoff (250-400 ng/mL): 80/80 Positive.
Semi-Quantitative Analysis (200ng/mL cutoff)Correct classification and semi-quantitative results relative to the 200ng/mL cutoff.At -100% to -25% of cutoff (0-150 ng/mL): 80/80 Negative. At cutoff (200 ng/mL): 47 Negative / 33 Positive (demonstrating boundary function). At +25% to +100% of cutoff (250-400 ng/mL): 80/80 Positive.
Specificity and Cross-Reactivity (Qualitative)Correct classification of structurally similar and non-similar compounds at specified concentrations.Tramadol (200ng/mL): 100% Cross-Reactivity. n-Desmethyl Tramadol (450ng/mL): POS, 44.4% Cross-Reactivity. o-Desmethyl Tramadol (25000ng/mL): POS, 0.8% Cross-Reactivity. Venlafaxine/o-Desmethyl Venlafaxine (100,000ng/mL): NEG, N.D.
Specificity and Cross-Reactivity (Semi-Quantitative)Correct classification and semi-quantitative results for structurally similar and non-similar compounds.Tramadol (200ng/mL): 100% Cross-Reactivity. n-Desmethyl Tramadol (450ng/mL): 44.4% Cross-Reactivity. o-Desmethyl Tramadol (25000ng/mL): 0.8% Cross-Reactivity. Venlafaxine/o-Desmethyl Venlafaxine (100,000ng/mL): N.D.
Interference (Structurally Non-Similar Compounds)No interference (Negative results for -25% cutoff, Positive for +25% cutoff).For most compounds tested (e.g., 6-Acetylcodeine, Alprazolam, Caffeine, Cocaine, Codeine, etc.): Negative at -25% cutoff (150ng/mL) and Positive at +25% cutoff (250ng/mL), indicating no interference.
Interference (Endogenous Compounds)No interference (Negative results for -25% cutoff, Positive for +25% cutoff).For most compounds tested (e.g., Acetone, Ascorbic Acid, Creatinine, Ethanol, Glucose, Urea, etc.): Negative at -25% cutoff (150ng/mL) and Positive at +25% cutoff (250ng/mL), indicating no interference. Boric Acid (1% w/v): Negative at both -25% and +25% cutoff, indicating interference. (Note: This interference was acknowledged and added to labeling).
Interference (Effect of pH)No interference within the physiological pH range.pH 3.0 to 11.0: Negative at -25% cutoff (150ng/mL) and Positive at +25% cutoff (250ng/mL), indicating no interference.
Interference (Effect of Specific Gravity)No interference across a range of specific gravities.Specific Gravity 1.000 to 1.030: Negative at -25% cutoff (150ng/mL) and Positive at +25% cutoff (250ng/mL), indicating no interference.
Linearity/RecoveryRecovery percentage within an acceptable range across the expected concentration range.Recovery %: Ranged from 94% to 110% across concentrations of 50 ng/mL to 1100 ng/mL.
Method Comparison (Qualitative)High agreement with LC/MS confirmation for positive and negative samples.Positive: 100% agreement (100 samples). Negative: 100% agreement (50 samples). Overall Agreement (%): 100% for both positive and negative categories at various concentration ranges.
Method Comparison (Semi-Quantitative)High agreement with LC/MS confirmation for positive and negative samples.Positive: 100% agreement (100 samples). Negative: 100% agreement (50 samples). Overall Agreement (%): 100% for both positive and negative categories at various concentration ranges.
Stability (Closed Accelerated Reagents)Continued performance within specifications over time.Qualitative & Semi-Quantitative: Levels 0 & 150 ng/mL consistently negative; 250 ng/mL consistently positive up to 40 days (accelerated). Supports 1-year expiration.
Stability (Open/On-board Reagents)Continued performance within specifications when opened and on-board.Qualitative & Semi-Quantitative: Levels 150 ng/mL consistently negative; 250 ng/mL consistently positive up to 28 days. Supports 28-day open vial expiration.
Stability (Specimen and Storage Handling)Stable performance of specimens over time under storage conditions.MS data: Specimen below cutoff consistently negative; specimen above cutoff consistently positive for Day 0, Week 1, 2, 3, and 4. Supports 1-month storage at 2-8°C.
Stability (Calibrator and Control - Open Accelerated)Stable performance of calibrators and controls over time.All calibrator and control levels: Within specifications for Day 0, 3, 7, 10, and 13. Supports 6-month open vial expiration.
Calibrator and Control Traceability & Value AssignmentTraceability to commercial standards and accurate value assignment via mass spectrometry.All components traced to Cerilliant Chemicals standard. Calibrators/controls adjusted and retested by mass spectrometry until within acceptable ranges.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Precision/Cutoff Characterization Study: N=80 for each concentration level (0, 50, 100, 150, 200, 250, 300, 350, 400 ng/mL). This involved 20 days, 2 runs per day, in duplicate.

  • Specificity and Cross-Reactivity: Not explicitly stated but implies testing of individual compounds at specified concentrations.

  • Interference (Structurally Non-Similar Compounds, Endogenous Compounds, pH, Specific Gravity): Not explicitly stated, but each potential interferent was tested at two concentrations (-25% and +25% of cutoff). This is typically performed on multiple replicates.

  • Linearity/Recovery: 13 different concentrations were prepared and tested.

  • Method Comparison: 100 positive samples and 50 negative samples were analyzed.

  • Stability Studies:

    • Closed Accelerated Stability: Tested at various time points (Day 0, 2, 8, 16, 24, 32, 40) for 0, 150, 250 ng/mL levels.
    • Open/On-board Stability: Tested at various time points (Day 0, 7, 14, 21, 28) for 150, 250 ng/mL levels (all replicates).
    • Specimen and Storage Handling: Tested at various time points (Day 0, Week 1, 2, 3, 4) for one specimen below cutoff and one above cutoff.
    • Calibrator and Control Stability: Tested at various time points (Day 0, 3, 7, 10, 13) for all 4 calibrator levels and 2 control levels.
  • Data Provenance: The document states that for the Method Comparison study, "Unaltered, anonymous and discarded clinical urine samples obtained from clinical testing laboratories were analyzed with the test device." This indicates the data is retrospective and derived from a clinical laboratory setting, but the specific country of origin is not mentioned. For other studies, samples were prepared in the laboratory (e.g., drug-free urine spiked with analytes).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • Ground Truth Establishment: For the method comparison study, "Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography / Mass Spectroscopy (LC/MS) is the preferred confirmatory method." The document explicitly states that the device's performance was "verified by Mass Spectrometry" and LC/MS was used. LC/MS is an analytical gold standard, performed by trained laboratory personnel.
  • Number of Experts/Qualifications: The document does not specify human experts or their qualifications for establishing ground truth; instead, it relies on the results from the confirmatory analytical method (LC/MS).

4. Adjudication Method for the Test Set

  • Given that the ground truth for the method comparison study was established by LC/MS, an objective analytical method, there was no need for a human adjudication method (e.g., 2+1, 3+1). The LC/MS result is considered the definitive truth.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an in-vitro diagnostic device (immunoassay) intended for use with automated clinical chemistry analyzers, not for human interpretation of medical images or other subjective assessments. Therefore, the concept of "human readers improving with AI vs without AI assistance" is not applicable here.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Yes, the entire study effectively demonstrates the standalone performance of the Immunalysis Tramadol Urine Enzyme Immunoassay. It is an automated assay run on a Beckman Coulter AU 400e (an automated clinical chemistry analyzer). The results presented are the output of the algorithm/assay itself, with LC/MS serving as the independent gold standard for comparison. The device is intended to provide a preliminary analytical test result.

7. The Type of Ground Truth Used

  • The ground truth used for the method comparison study was confirmatory analytical testing, specifically Liquid Chromatography / Mass Spectroscopy (LC/MS). For other performance studies (e.g., precision, cross-reactivity, interference), ground truth was established by precise preparation and spiking of samples with known concentrations of analytes.

8. The Sample Size for the Training Set

  • This is an in-vitro diagnostic (IVD) immunoassay kit. IVD assays are developed through chemical and biological formulation and optimization, not typically through machine learning or AI algorithms that require "training sets" in the conventional sense. Therefore, the concept of a "training set" as understood in AI/ML is not applicable here. The reagent formulation and assay parameters are chemically and biochemically defined and validated, not "trained" on a data set.

9. How the Ground Truth for the Training Set Was Established

  • As explained in point 8, there isn't a "training set" in the AI/ML sense for this type of device. The ground truth for the development and optimization of the immunoassay itself would come from standard analytical chemistry principles, characterization of antibody binding, enzyme kinetics, and comparison to established reference standards during the formulation process. For example, calibrators are prepared and validated against commercially available standard solutions (Cerilliant Chemicals, as mentioned in the document).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is a stylized image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 12, 2015

IMMUNALYSIS CORPORATION JOSEPH GINETE REGULATORY AFFAIRS SPECIALIST 829 TOWNE CENTER DR. POMONA CA 91767

Re: K141803

Trade/Device Name: Immunalysis Tramadol Urine Enzyme Immunoassay, Immunalysis Tramadol Urine Controls, Immunalysis Tramadol Urine Calibrators Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: II Product Code: DJG, DLJ, LAS Dated: December 30, 2014 Received: January 2, 2015

Dear Mr. Joseph Ginete:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -A

For : Courtney H. Lias, Ph.D.

Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141803

Device Name

Immunalysis Tramadol Urine Enzyme Immunalysis Tramadol Urine Controls and Calibrators

Indications for Use (Describe)

Immunalysis Tramadol Urine Enzyme Immunoassay:

The Immunalysis Tramadol Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 200ng/ mL. The assay is intended for use in laboratories for the qualitative analysis of Tramadol in human urine with automated clinical chemistry analyzers. This assay is calibrated against Tramadol. This in-vitro diagnostic device is for prescription use only.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GC-MS or permitting laboratories to establish quality control procedures. The test is not intended to differentiate between drugs of abuse and prescription use of Tramadol. There are no uniformly recognized drug levels for Tramadol in urine.

The Immunalysis Tramadol Urine Enzyne Immunoassay Kit provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography / Mass Spectroscopy (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Immunalysis Tramadol Urine Controls:

The Immunalysis Tramadol Urine Controls are used as control materials in the Immunalysis Tramadol Urine Enzyme Immunoassay.

Immunalysis Tramadol Urine Calibrators:

The Immunalysis Tramadol Urine Calibrators are used as calibrators in the Immunalysis Tramadol Urine Enzyme Immunoassay for the qualitative and semi-quantitative determination of Tramadol in urine on automated clinical chemistry analyzers.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The letters are bold and slightly blurred, giving them a three-dimensional appearance. The red background is a solid color, and the word is centered in the image.

A. Contact Information

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(c).

    1. Manufacturer: Immunalysis Corporation 2. Contact Name: Joseph Ginete 3. Contact Title: Regulatory Affairs Specialist 4. Address: 829 Towne Center Drive Pomona, CA 91767 (909) 482-0840 5. Phone: (909) 482-0850 6. Fax: 7. Email: jginete@immunalysis.com 8. Summary prepared on: February 02, 2015 B. Device Information 1. Trade Name: Immunalysis Tramadol Urine Enzyme Immunoassay Immunalysis Tramadol Urine Controls Immunalysis Tramadol Urine Calibrators 2. Common Name: Immunalysis Tramadol Urine Enzyme Immunoassay Immunalysis Tramadol Urine Controls Immunalysis Tramadol Urine Calibrators C. Requlatory Information 1. Device Classification: Class II Class I, reserved 2. Requlation Number: CFR 862.3650 Opiate Test System CFR 862.3200 Clinical Toxicology Calibrator CFR 862.3280 Clinical Toxicology Control Materials 3. Panel: Toxicology(91) 4. Product Code: DJG DLJ LAS D. Legally Marketed Device to Which We are Claiming Equivalence (807.92(A)(3)) LZI Opiate 2000 Enzyme Immunoassay 1. Predicate Device: LZI Opiate 2000 Enzyme Controls
    • LZI Opiate 2000 Enzyme Calibrators 2. Predicate Company: Lin-Zhi International Inc. 3. Predicate K Number: K120761

{4}------------------------------------------------

E. Device Description

The assay consists of antibody/ substrate reagent and enzyme conjugate reagent. The antibody/ substrate reagent includes goat antibodies to Tramadol, glucose-6-phosphate (G6P) and nicotinamide adenine dinucleotide (NAD) in Tris buffer with Sodium Azide as a preservative. The enzyme conjuqate reagent includes tramadol derivative labeled with qlucose-6-phosphate dehydrogenase (G6PDH) in Tris buffer with Sodium Azide as a preservative. Calibrators and controls are sold separately. Reagents are liquid, ready to use

The tramadol calibrator and controls consists of a single calibrator at 200ng/mL, a control set containing a LOW control at 150ng/mL and a HIGH control at 250ng/mL and a calibrator set containing a negative calibrator, a Level 1 calibrator at 100ng/mL, a Level 2 calibrator at 200ng/mL, a Level 3 calibrator at 500ng/mL and a Level 4 calibrator at 1000nq/mL.

  • F. Intended Use
    Immunalysis Tramadol Urine Enzyme Immunoassay:

The Immunalysis Tramadol Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 200ng/mL. The assay is intended for use in laboratories for the qualitative and semi-quantitative analysis of Tramadol in human urine with automated clinical chemistry analyzers. This assay is calibrated against Tramadol. This in-vitro diagnostic device is for prescription use only.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GC-MS or permitting laboratories to establish quality control procedures. The test is not intended to differentiate between drugs of abuse and prescription use of Tramadol. There are no uniformly recognized drug levels for Tramadol in urine.

The Immunalysis Tramadol Urine Enzyme Immunoassay Kit provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liguid Chromatography / Mass Spectroscopy (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Immunalysis Tramadol Urine Controls:

The Immunalysis Tramadol Urine Controls are used as control materials in the Immunalysis Tramadol Urine Enzyme Immunoassay.

Immunalysis Tramadol Urine Calibrators:

The Immunalysis Tramadol Urine Calibrators are used as calibrators in the Immunalysis Tramadol Urine Enzyme Immunoassay for the qualitative and semi-quantitative determination of Tramadol in urine on automated clinical chemistry analyzers.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image. The image is simple and easy to read.

ItemPredicate Device (K120761)Test Device
Intended UseFor the qualitative and semi-quantitative determination of the presence of opiates in human urine at a cutoff of 2000ng/mlFor the qualitative and semi-quantitative determination of the presence of tramadol in human urine at a cutoff of 200ng/ml
Type of ProductAnalytical ReagentsAnalytical Reagents
Measured AnalytesOpiatesTramadol
Test MatrixUrineUrine
Cutoff Levels2000ng/mL of Opiates200ng/mL of Tramadol
Test SystemEnzyme ImmunoassayHomogenous Enzyme Immunoassay
MaterialsR1 antibody reagent and R2 enzyme reagentAntibody/ Substrate Reagents and Enzyme Labeled Conjugate
Mass Spectroscopy ConfirmationRequired for preliminary positive analytical resultsRequired for preliminary positive analytical results
AntibodyMouse monoclonal anti-morphine derivativeGoat Polyclonal Antibody to Tramadol
Storage2 – 8°C until expiration date2 – 8°C until expiration date
Calibrator FormLiquidLiquid
Calibrator LevelsOne (1) Level (2000ng/mL)One (1) Level (200ng/mL)
Control Set LevelsTwo (2) Levels (1500ng/mL and 2500ng/mL)Two (2) Levels (150ng/mL and 250ng/mL)
Calibrator Set LevelsFive (5) Levels (0, 1000, 2000, 4000 and 6000 ng/mL)Five (5) Levels (0, 100, 200, 500 and 1000 ng/mL)
  • G. Comparison of the new device with the predicate device

H. Test Principle

    1. Test Principle and Procedure:
      This assay uses a Tramadol specific antibody. The assay is based on the competition of Tramadol labeled enzyme glucose-6-phosphate dehydrogenase (G6PDH) and the free drug in the urine sample for the fixed amount of antibody binding sites. In the absence of the free drug in the sample, the antibody binds the drug enzyme conjugate and enzyme activity is inhibited. This creates a dose response relationship between drug concentration in the urine sample and enzyme activity. The enzyme G6PDH activity is determined at 340 nm spectrophotometrically by the conversion of NAD to NADH.
  • l. The following laboratory performance studies were performed to determine substantial equivalence of the Immunalysis Tramadol Urine Enzyme Immunoassay to the predicate

      1. Precision/ Cutoff Characterization Study was performed for 20 days, 2 runs per day in duplicate (N=80) on concentration of ±25%, ±50%, ±75% and ±100% of the cutoff. The study verified that the cutoff serves as a boundary between a negative and positive interpretation of a qualitative result. In addition, it also verified the product performance relative to the ability of the device to produce the same value during repeated measurements. The instrument used for this test was a Beckman Coulter AU 400e.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The letters are bold and have a slight shadow effect, making them stand out from the background. The red background is a solid color, and the word is centered in the image.

a.The following is a summary table of the Qualitative Analysis for the 200nq/mL cutoff test data results.

Qualitative Analysis (for 200ng/mL cutoff)
Concentration (ng/mL)% of cutoff# of determinationsResult
0-100%8080 Negative
50-75%8080 Negative
100-50%8080 Negative
150-25%8080 Negative
200Cutoff8044 Negative/36 Positive
250+25%8080 Positive
300+50%8080 Positive
350+75%8080 Positive
400+100%8080 Positive

b.The following is a summary table of the Semi-Quantitative Analysis for the 200ng/mL cutoff test data results.

Semi-Quantitative Analysis (for 200ng/mL cutoff)
Concentration (ng/mL)% of cutoff# of determinationsResult
0-100%8080 Negative
50-75%8080 Negative
100-50%8080 Negative
150-25%8080 Negative
200Cutoff8047 Negative/ 33 Positive
250+25%8080 Positive
300+50%8080 Positive
350+75%8080 Positive
400+100%8080 Positive

Specificity and Cross-Reactivity - Structurally similar compounds were spiked 2. into drug free urine at levels that will vield a result that is equivalent to the cutoff. The study verified assay performance relative to the ability of the device to exclusively determine certain drugs. The instrument used for this test was a Beckman Coulter AU 400e.

a. The qualitative result summary table is outlined below:

Structurally Related Compounds – Qualitative
CompoundConcentration Tested (ng/mL)ResultCross-Reactivity (%)
Tramadol200N/A100.00
n-Desmethyl Tramadol450POS44.4
o-Desmethyl Tramadol25,000POS0.8
Venlafaxine100,000NEGN.D.
o-Desmethyl Venlafaxine100,000NEGN.D.

N.D. = Not Detected (<0.05%)

b. The semi-quantitative result summary table is outlined below:

Structurally Related Compounds – Semi-Quantitative
CompoundConcentration Tested (ng/mL)Cross-Reactivity (%)
Tramadol200100.00
n-Desmethyl Tramadol45044.4
o-Desmethyl Tramadol25,0000.8
Venlafaxine100,000N.D.
o-Desmethyl Venlafaxine100,000N.D.
  1. Interference - Structurally non-similar compounds, endogenous compounds, the effect of pH and the effect of specific gravity was evaluated by spiking the potential interferent into drug free urine containing the target analyte at ±25% of the cutoff. Boric Acid caused a false negative response at the concentration

{7}------------------------------------------------

tested. All other potential interferents analyzed verified that assay performance is unaffected by externally ingested compounds or an internally existing physiological condition. The instrument used for this test was a Beckman Coulter AU 400e.

Structurally Non-Similar Compounds (for 200ng/mL cutoff)
CompoundConcentration Tested (ng/mL)-25% Cutoff (150ng/mL)+25% Cutoff (250ng/mL)
ResultInterference?ResultInterference?
6-Acetylcodeine100,000NegativeNoPositiveNo
6-Acetylmorphine100,000NegativeNoPositiveNo
7-Aminoclonazepam100,000NegativeNoPositiveNo
7-Aminoflunitrazepam100,000NegativeNoPositiveNo
7-Aminonitrazepam100,000NegativeNoPositiveNo
Acetaminophen500,000NegativeNoPositiveNo
Acetylsalicyclic Acid500,000NegativeNoPositiveNo
Alprazolam50,000NegativeNoPositiveNo
Amitriptyline100,000NegativeNoPositiveNo
Amobarbital100,000NegativeNoPositiveNo
S-(+) Amphetamine100,000NegativeNoPositiveNo
Benzoylecgonine500,000NegativeNoPositiveNo
Benzylpiperazine100,000NegativeNoPositiveNo
Bromazepam100,000NegativeNoPositiveNo
4-Bromo-2,5,Dimethoxyphenethylamine100,000NegativeNoPositiveNo
Buprenorphine100,000NegativeNoPositiveNo
Bupropion25,000NegativeNoPositiveNo
Butabarbital100,000NegativeNoPositiveNo
Caffeine500,000NegativeNoPositiveNo
Cannabidiol100,000NegativeNoPositiveNo
Cannabinol100,000NegativeNoPositiveNo
Carbamazeprine100,000NegativeNoPositiveNo
Carisoprodol100,000NegativeNoPositiveNo
Chlordiazepoxide100,000NegativeNoPositiveNo
Chlorpromazine100,000NegativeNoPositiveNo
Clobazam100,000NegativeNoPositiveNo
Clomipramine100,000NegativeNoPositiveNo
Clonazepam100,000NegativeNoPositiveNo
Cocaine100,000NegativeNoPositiveNo
Codeine100,000NegativeNoPositiveNo
Cotinine100,000NegativeNoPositiveNo
Cyclobenzaprine100,000NegativeNoPositiveNo
Delta-9-THC100,000NegativeNoPositiveNo
Demoxepam100,000NegativeNoPositiveNo
Desakylflurazepam100,000NegativeNoPositiveNo
Desipramine100,000NegativeNoPositiveNo
Dextromethorphan100,000NegativeNoPositiveNo
Diazepam50,000NegativeNoPositiveNo
Dihydrocodeine100,000NegativeNoPositiveNo
Diphenhydramine500,000NegativeNoPositiveNo
Structurally Non-Similar Compounds (for 200ng/mL cutoff)
CompoundConcentrationTested (ng/mL)-25% Cutoff (150ng/mL)+25% Cutoff (250ng/mL)
ResultInterference?ResultInterference?
Doxepin100,000NegativeNoPositiveNo
Ecgonine100,000NegativeNoPositiveNo
Ecgonine methyl ester100,000NegativeNoPositiveNo
EDDP100,000NegativeNoPositiveNo
1R,2S(-)-Ephedrine100,000NegativeNoPositiveNo
1S,2R(+)-Ephedrine100,000NegativeNoPositiveNo
EtG100,000NegativeNoPositiveNo
Ethylmorphine100,000NegativeNoPositiveNo
Fenfluramine100,000NegativeNoPositiveNo
Fentanyl100,000NegativeNoPositiveNo
Flunitrazepam100,000NegativeNoPositiveNo
Fluoxetine100,000NegativeNoPositiveNo
Flurazepam100,000NegativeNoPositiveNo
Heroin100,000NegativeNoPositiveNo
Hexobarbital100,000NegativeNoPositiveNo
Hydrocodone100,000NegativeNoPositiveNo
Hydromorphone100,000NegativeNoPositiveNo
11-hydroxy-delta-9-THC100,000NegativeNoPositiveNo
Ibuprofen100,000NegativeNoPositiveNo
Imipramine100,000NegativeNoPositiveNo
Ketamine100,000NegativeNoPositiveNo
Lamotrigine100,000NegativeNoPositiveNo
Levorphanol100,000NegativeNoPositiveNo
Lidocaine100,000NegativeNoPositiveNo
Lorazepam100,000NegativeNoPositiveNo
Lorazepam Glucuronide50,000NegativeNoPositiveNo
Lormetazepam100,000NegativeNoPositiveNo
LSD100,000NegativeNoPositiveNo
Maprotiline100,000NegativeNoPositiveNo
S(+)-MDA100,000NegativeNoPositiveNo
MDEA100,000NegativeNoPositiveNo
MDMA100,000NegativeNoPositiveNo
Meperidine100,000NegativeNoPositiveNo
Meprobamate100,000NegativeNoPositiveNo
Methadone500,000NegativeNoPositiveNo
S(+)-Methamphetamine500,000NegativeNoPositiveNo
Methaquolone100,000NegativeNoPositiveNo
Methylphenidate100,000NegativeNoPositiveNo
Midazolam100,000NegativeNoPositiveNo
Morphine100,000NegativeNoPositiveNo
Morphine-3 -glucuronide100,000NegativeNoPositiveNo
Morphine-6 -glucuronide100,000NegativeNoPositiveNo
Nalorphine100,000NegativeNoPositiveNo
Naloxone100,000NegativeNoPositiveNo
Naltrexone100,000NegativeNoPositiveNo
Naproxen100,000NegativeNoPositiveNo
Structurally Non-Similar Compounds (for 200ng/mL cutoff)
CompoundConcentrationTested (ng/mL)-25% Cutoff (150ng/mL)+25% Cutoff (250ng/mL)
ResultInterference?ResultInterference?
N-desmethyltapentadol100,000NegativeNoPositiveNo
Nitrazepam100,000NegativeNoPositiveNo
Norbuprenorphine100,000NegativeNoPositiveNo
Norcodeine100,000NegativeNoPositiveNo
Nordiazepam100,000NegativeNoPositiveNo
Normorphine100,000NegativeNoPositiveNo
Norpropoxyphene100,000NegativeNoPositiveNo
Norpseudoephedrine100,000NegativeNoPositiveNo
Nortriptyline100,000NegativeNoPositiveNo
Oxazepam100,000NegativeNoPositiveNo
Oxazepam Glucuronide10,000NegativeNoPositiveNo
Oxycodone100,000NegativeNoPositiveNo
Oxymorphone100,000NegativeNoPositiveNo
PCP100,000NegativeNoPositiveNo
Pentazocine100,000NegativeNoPositiveNo
Pentobarbital100,000NegativeNoPositiveNo
Phenobarbital100,000NegativeNoPositiveNo
Phentermine100,000NegativeNoPositiveNo
Phenylephrine100,000NegativeNoPositiveNo
Phenylpropanolamine100,000NegativeNoPositiveNo
Phenytoin100,000NegativeNoPositiveNo
PMA100,000NegativeNoPositiveNo
Propoxyphene100,000NegativeNoPositiveNo
Propranolol100,000NegativeNoPositiveNo
Protriptyline100,000NegativeNoPositiveNo
R,R(-)-Pseudoephedrine100,000NegativeNoPositiveNo
S,S(+)-Pseudoephedrine100,000NegativeNoPositiveNo
Ranitidine100,000NegativeNoPositiveNo
Ritalinic Acid100,000NegativeNoPositiveNo
Salicylic Acid100,000NegativeNoPositiveNo
Secobarbital100,000NegativeNoPositiveNo
Sertraline100,000NegativeNoPositiveNo
Sufentanil Citrate100,000NegativeNoPositiveNo
Temazepam100,000NegativeNoPositiveNo
11-nor-9 carboxy THC100,000NegativeNoPositiveNo
Theophylline100,000NegativeNoPositiveNo
Thioridazine100,000NegativeNoPositiveNo
Trazodone100,000NegativeNoPositiveNo
Triazolam100,000NegativeNoPositiveNo
Trifluoromethylphenyl-piperazine100,000NegativeNoPositiveNo
Trimipramine100,000NegativeNoPositiveNo
Zolpidem Tartrate100,000NegativeNoPositiveNo

a. The following is a summary table of the structurally non-similar compounds for the 200ng/mL cutoff

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The letters are bold and have a slight shadow effect, making them stand out from the background. The word is centered and fills most of the frame. The red background is a solid color, providing a strong contrast to the white letters.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters and is centered in the image. The red background is a solid color and there are no other objects in the image.

b. The following is a summary table of the endogenous compounds results for the 200ng/mL cutoff

Endogenous Compounds (for 200ng/mL cutoff)
CompoundConcentration-25% Cutoff (150ng/mL+25% Cutoff (250ng/mL)
Tested (ng/mL)ResultInterference?ResultInterference?
Acetone1.0 g/dLNegativeNoPositiveNo
Ascorbic Acid1.5 g/dLNegativeNoPositiveNo
Bilirubin0.002 g/dLNegativeNoPositiveNo
Boric Acid1% w/vNegativeNoNegativeYes
Creatinine0.5 g/dLNegativeNoPositiveNo
Ethanol1.0 g/dLNegativeNoPositiveNo
Galactose0.01 g/dLNegativeNoPositiveNo
y-Globulin0.5 g/dLNegativeNoPositiveNo
Glucose2.0 g/dLNegativeNoPositiveNo
Hemoglobin0.300 g/dLNegativeNoPositiveNo
Human Serum Albumin0.5 g/dLNegativeNoPositiveNo
Oxalic Acid0.1 g/dLNegativeNoPositiveNo
Riboflavin0.0075 g/dLNegativeNoPositiveNo
Sodium Azide1% W/vNegativeNoPositiveNo
Sodium Chloride6.0 g/dLNegativeNoPositiveNo
Sodium Flouride1% w/vNegativeNoPositiveNo
Urea6.0 g/dLNegativeNoPositiveNo

Boric Acid interferes with the assay and the limitation has been added C. to the labeling regarding this compound.

d. The following is a summary table of the effect of pH results for the 200ng/mL cutoff

Effect of pH (for 200ng/mL cutoff)
Test ParameterValue-25% Cutoff (150ng/mL)+25% Cutoff (250ng/mL)
ResultInterference?ResultInterference?
pH3.0NegativeNoPositiveNo
pH4.0NegativeNoPositiveNo
pH5.0NegativeNoPositiveNo
pH6.0NegativeNoPositiveNo
pH7.0NegativeNoPositiveNo
pH8.0NegativeNoPositiveNo
pH9.0NegativeNoPositiveNo
pH10.0NegativeNoPositiveNo
pH11.0NegativeNoPositiveNo

e. The following is a summary table of the effect of specific gravity result for the 200ng/mL cutoff:

Effect of Specific Gravity (for 200ng/mL cutoff)
Test ParameterValue-25% Cutoff (150ng/mL)+25% Cutoff (250ng/mL)
Specific Gravity1.000NegativeNoPositiveNo
Specific Gravity1.002NegativeNoPositiveNo
Specific Gravity1.005NegativeNoPositiveNo
Specific Gravity1.010NegativeNoPositiveNo
Specific Gravity1.015NegativeNoPositiveNo
Specific Gravity1.020NegativeNoPositiveNo
Specific Gravity1.025NegativeNoPositiveNo
Specific Gravity1.030NegativeNoPositiveNo

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The letters are bold and have a slight shadow effect, making them stand out from the background. The word is centered and fills most of the frame, making it the focal point of the image. The overall impression is one of clarity and emphasis, with the bright colors and bold font drawing the viewer's attention to the word.

    1. Linearity/ Recovery A drug free urine pool was spiked with a high concentration of the target analyte as a high value specimen. Additional pools were made by serially diluting the high value specimen. The study verified assay linearity in the semi-quantitative mode. The instrument used for this test was a Beckman Coulter AU 400e.
a. Summary results are listed in the following table:
---------------------------------------------------------
Linearity/ Recovery
Expected Concentration (ng/mL)Mean Concentration (ng/mL)Recovery (%)
01N/A
5051101
1009595
200201100
300330110
400426107
500512102
600647108
700770110
800853107
900911101
100094494
1100105996
  1. Method Comparison - Unaltered, anonymous and discarded clinical urine samples obtained from clinical testing laboratories were analyzed with the test device. The study verified that the product performance can be verified by Mass Spectrometry. The instrument used for this test was a Beckman Coulter AU 400e and an Agilent 6430 Liquid Chromatography Tandem Mass Spectrometry.
  • a. The following is a comparison table of qualitative assay performance for the 200ng/mL cutoff
LC/MS Confirmation
(+)(-)
Test Device(+)1000
(-)050
  • b. The following is a summary table of qualitative assay performance for the 200ng/mL cutoff
Assay Performance verified by LC/MS – 200ng/mL Cutoff
Tramadol Concentration
Type< 100ng/mL100 ~ 199 ng/mL200 ~ 300 ng/mL> 300 ng/mLAgreement (%)
Qualitative/ Positive001090100%
Qualitative/ Negative45500100%

c. The following is a comparison table of semi-quantitative assay performance for the 200ng/mL cutoff

LC/MS Confirmation
Test Device(+)(-)
(+)1000
(-)050
  • d. The following is a summary table of semi-quantitative assay nerformance for the 200ng/ml_cutoff
Assay Performance verified by LC/MS – 200ng/mL Cutoff
TypeTramadol ConcentrationAgreement (%)
< 100ng/mL100 ~ 199 ng/mL200 ~ 300 ng/mL> 300 ng/mL
Semi-Quantitative/ Positive001090100%
Semi-Quantitative / Negative45500100%
  1. Stability -

a. A closed accelerated stability study was performed on reagents.

calibrators and controls at 25°C to establish the initial expiration dating. The stability study supported an initial expiration date of 1 year for

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image contains the word "IMMUNALYSIS" in white letters on a red background. The letters are bold and have a slight shadow effect, making them stand out from the background. The word is centered and fills most of the image.

reagents. This stability study supported an initial expiration date of 12 months for calibrators and controls. The instrument used for this test was a Beckman Coulter AU 400e.

    1. The following is a summary of the qualitative stability data. The 0 and 150ng/mL levels were negative in comparison to the 200ng/mL cutoff for Day 0. 2. 8. 16. 24. 32 and 40. The 250ng/mL level was positive in comparison to the 200ng/mL cutoff for Day 0, 2, 8, 16, 24, 32 and 40. This accelerated stability study was performed to establish initial expiration dating. Real time stability studies are ongoing.
    1. The following is a summary of the semi-quantitative stabililty data for the 200ng/mL cutoff. The 150ng/mL level was negative in comparison to the 200ng/mL cutoff for Day 0, 2, 8, 16, 24, 32 and 40. The 250ng/mL level was positive in comparison to the 200ng/mL cutoff for Day 0, 2, 8, 16, 24, 32 and 40. This accelerated stability study was performed to establish initial expiration dating. Real time stability studies are ongoing.
  • b. An open/ on-board stability study was performed on reagents to establish expiration dating when reagents are opened and stored on board the instrument at 2℃ to 8℃. The stability study supported an initial open vial expiration date of 28 days. The instrument used for this test was a Beckman Coulter AU 400e.
      1. The following is a summary of the qualitative open/ on-board stability data for the 200ng/mL cutoff. All replicates for the 150ng/mL level were negative in comparison to the 200ng/mL cutoff for Day 0, 7, 14, 21 and 28. All replicates of the 250ng/mL level were positive in comparison to the 200ng/mL cutoff for Day 0, 7, 14, 21 and 28.
      1. The following is a summary of the semi-quantitative open/ onboard stability data for the 200ng/mL cutoff. The mean of the replicates for the 150ng/mL level were negative in comparison to the 200ng/mL cutoff for Day 0, 7, 14, 21 and 28. The mean of the replicates of the 250ng/mL level were positive in comparison to the 200ng/mL cutoff for Day 0, 7, 14, 21 and 28.
  • c. A Specimen and Storage Handling study was performed on a specimen below the cutoff near the bracketing control and a specimen above the cutoff near the bracketing control to establish specimen storage and handling stability at 2 – 8°C. The stability study supported a 1 month specimen storage and handling recommendation at 2 - 8°C.
      1. The following is a summary of the Mass Spectrometry data. The specimen below the cutoff was negative in comparison to the 200ng/mL cutoff for Day 0, Week 1, 2, 3 and 4. The specimen above the cutoff was positive in comparison to the 200ng/mL cutoff for Day 0, Week 1, 2, 3 and 4.
    1. Calibrator and Control Traceability all components of the calibrator and controls have been traced to a commercially available standard solution from Cerilliant Chemicals.
    1. Calibrator and Control Stability An open accelerated stability study was performed at 37°C to establish the initial open vial expiration dating. The stability study supported an initial open vial expiration date of 6 months. The instrument used for this test was a Beckman Coulter AU 400e. All calibrator levels (100, 200, 500 and 1000ng/mL) and control levels (150 and 250ng/mL) were within specifications for Day 0, 3, 7, 10 and 13. This accelerated stability study was

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The letters are bold and sans-serif. The word is centered in the image and takes up most of the space. The red background is a solid color.

performed to establish initial expiration dating. Real time stability studies are onqoing.

    1. Calibrator and Control Value Assignment calibrators and controls are manufactured and are tested by mass spectrometry. If any of the analytes are out of the acceptable range, then the calibrator or control is adjusted and retested. Values are assigned to the calibrator and controls once the Mass spectrometry results are within the acceptable ranges.
  • J. Conclusion

The information provided in this pre-market notification demonstrates that the Immunalysis Tramadol Urine Enzyme Immunoassay is substantially equivalent to the legally marketed predicate device for its general intended use.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).