The Sofia® Strep A+ FIA detects Group A Streptococcal antigens from patients with signs and symptoms of pharyngitis, such as sore throat. All negative test results should be confirmed by bacterial culture because negative results do not preclude Group A Strep infection and should not be used as the sole basis for treatment. The test is intended for professional and laboratory use as an aid in the diagnosis of Group A Streptococcal infection.

Device Description

The Sofia Strep A FIA employs immunofluorescence technology that is used with the Sofia analyzer (Sofia) to detect Group A Streptococcal antigen. The Sofia Strep A FIA involves the extraction of the antigenic components of the Group A Streptococcus (GAS) bacteria. The patient's swab sample is placed in the Reagent Tube containing the Reagent Solution, during which time the bacterial antigens are extracted, making them more accessible to the specific antibodies. An aliquot of the extracted sample is dispensed into the Cassette sample well. From the sample well, the sample migrates through a test strip containing various unique chemical environments. If Group A Streptococcal antigens are present, they will be bound by antibodies coupled to fluorescent microparticles that migrate through the test strip. The fluorescent microparticles containing bound antigen will be captured by antibodies at a defined location on the test strip where they are detected by Sofia. If antigens are not present, the fluorescent microparticles will not be trapped by the capture antibodies nor detected by Sofia. Note: The Cassette, now containing the sample, is placed directly inside Sofia for automatically timed development (WALK AWAY Mode). Sofia scans, measures, and interprets the immunofluorescent signal using method-specific algorithms. Sofia will display the test results (Positive, Negative, or Invalid) on the screen. The results can also be automatically printed on an integrated printer if this option is selected, or transmitted via an LIS connection.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Device: Sofia® Strep A+ FIA

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" numerically for sensitivity and specificity. However, based on the performance study results, the implied acceptance criteria would be the statistically significant ranges demonstrated by the device.

MetricAcceptance Criteria (Implied from performance)Reported Device Performance (95% CI)
Sensitivity> 89.1% (lower bound of 95% CI)93.7% (89.1%-96.5%)
Specificity> 92.4% (lower bound of 95% CI)94.4% (92.4%-95.9%)
PPVNot explicitly defined, but reported81.2%
NPVNot explicitly defined, but reported98.3%
Reproducibility Inter-laboratory (Negative)> 95.9% (lower bound of 95% CI)100% (95.9-100.0%)
Reproducibility Inter-laboratory (High Negative)> 82.1% (lower bound of 95% CI)90% (82.1-94.7%)
Reproducibility Inter-laboratory (Low Positive)> 78.1% (lower bound of 95% CI)87% (78.1-92.2%)
Reproducibility Inter-laboratory (Mod Positive)> 95.9% (lower bound of 95% CI)100% (95.9-100.0%)
Reproducibility Intra-laboratory (Overall Site 1)> 89.5% (lower bound of 95% CI)95% (89.5-97.7%)
Reproducibility Intra-laboratory (Overall Site 2)> 87.4% (lower bound of 95% CI)93% (87.4-96.6%)
Reproducibility Intra-laboratory (Overall Site 3)> 88.5% (lower bound of 95% CI)94% (88.5-97.2%)
Limit of Detection (LoD)Not explicitly defined as a single criterion, but values must be providedRanged from 2.76E+03 to 8.13E+03 cfu/test (for 3 strains)
Analytical Reactivity100% detection of tested strains at specified concentrationAll 21 tested Streptococcus pyogenes strains detected at 1.74E+04 cfu/test.
Analytical Specificity (Cross-Reactivity)No cross-reactivity with non-GAS organisms/viruses at specified concentrationsNone of the 61 non-Group A Streptococcus bacterial and fungal microorganisms, and 26 viral isolates showed any sign of cross-reactivity.
Interfering SubstancesNo interference at specified concentrationsMost substances did not interfere. Nacho Flavor Doritos interfered at 25% w/v and Fresh Whole Blood interfered at 100 µL/swab. Bovine submaxillary mucin interfered at 28.7 mg/mL.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Clinical Performance Study: 851 patients.
  • Data Provenance: Retrospective, collected during 2014 from 7 distinct CLIA-waived sites in various geographical regions within the United States.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the number of experts or their qualifications for interpreting the ground truth (bacterial culture and PCR). However, it mentions:

  • "A central Reference Laboratory" conducted the bacterial culture and PCR. This implies qualified laboratory personnel.

4. Adjudication Method for the Test Set

The primary comparison was against standard bacterial culture. For discordant results between the Sofia Strep A+ FIA and culture, an "FDA-cleared molecular device" (RT-PCR assay) was used for resolution.

  • Method: Initial comparison of Sofia Strep A+ FIA to bacterial culture.
  • Discordant Resolution: For specimens where Sofia Strep A+ FIA and culture disagreed, an FDA-cleared molecular device (PCR) was used to resolve the discrepancy. This is a 2-step adjudication approach (initial culture, then PCR for discrepancies).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not done. This study is for a diagnostic device (FIA) that measures antigens, not an AI-powered image analysis system that would assist human readers in interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the clinical performance study directly evaluated the "rapid FIA test result" from the Sofia Strep A+ FIA against the ground truth. The device itself (Sofia Strep A+ FIA with the Sofia analyzer) provides the result (Positive, Negative, or Invalid) and automatically interprets the immunofluorescent signal using method-specific algorithms, thus representing a standalone performance evaluation.

7. The Type of Ground Truth Used

The ground truth for the clinical performance study consisted of:

  • Bacterial Culture: Throat swabs were streaked on a sheep blood agar plate (SBA) and cultured for up to 48 hours.
  • Molecular (PCR): An FDA-cleared Group A Streptococcus RT-PCR assay was used to resolve discordant results between the device and primary culture. This is a combination of culture and a highly sensitive molecular method.

8. The Sample Size for the Training Set

The document does not specify a separate "training set" sample size. This type of device (immunofluorescence assay) doesn't typically undergo a machine learning training phase in the same way an AI-powered diagnostic algorithm would. The development of its "method-specific algorithms" would be based on analytical studies (LoD, cross-reactivity, precision) to define positive/negative thresholds rather than a distinct training dataset of patient samples.

9. How the Ground Truth for the Training Set Was Established

As noted above, a traditional "training set" with ground truth in the context of machine learning is not applicable here. The "method-specific algorithms" in the Sofia analyzer would have been developed and validated based on extensive analytical studies to establish appropriate thresholds for signal detection and interpretation. These analytical studies involved:

  • Serial dilutions of known concentrations of Streptococcus pyogenes strains (for LoD).
  • Testing against known non-Group A Streptococcus organisms and various viruses at specific concentrations (for analytical specificity/cross-reactivity).
  • Testing with potential interfering substances at defined concentrations.

These analytical studies establish the performance characteristics that dictate how the device's inherent algorithms classify a result, rather than a separate "training set" of patient data.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often associated with medicine and healthcare. The logo also includes the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the caduceus symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

QUIDEL CORPORATION December 16, 2014 JOHN TAMERIUS SENIOR VICE PRESIDENT, CLINICAL AND REGULATORY 12544 HIGH BLUFF (ADMINISTRATIVE OFFICE LOCATION) SAN DIEGO CA 92130

Re: K141775

Trade/Device Name: Sofia® Strep A+ FIA Regulation Number: 21 CFR 866.3740 Regulation Name: Streptococcus spp. serological reagents Regulatory Class: I Product Code: GTY Dated: June 30, 2014 Received: July 1, 2014

Dear Dr. Tamerius:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Image /page/1/Picture/7 description: The image shows the name "Uwe Scherf - S for" in a large, bold font. The text is black and is set against a white background. The text is horizontally aligned and appears to be the title or heading of a document or presentation.

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K141775

Device Name Sofia® Strep A+FIA Assay

Indications for Use (Describe)

The Sofia® Strep A+ FIA detects Group A Streptococcal antigens from patients with signs and symptoms of pharyngitis, such as sore throat. All negative test results should be confirmed by bacterial culture because negative results do not preclude Group A Strep infection and should not be used as the sole basis for treatment. The test is intended for professional and laboratory use as an aid in the diagnosis of Group A Streptococcal infection.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

Submitted By:Quidel Corporation12544 High Bluff Drive, Suite 200(Administrative Offices)San Diego, California 92130Telephone: 858-552-7908Fax: 858-646-8045
Submission Contact:John D. Tamerius, Ph.D.
Date Prepared:June 30, 2014
Device Trade Name:Sofia® Strep A+ FIA and Sofia
Common Name:Strep A immunological test system and Fluorometer
Predicate Devices:Sofia Strep A FIA for use with Sofia, K123793
Device Classification/Name:21 CFR 866.3740 / Streptococcus Group Aserological reagents
Intended Use:The Sofia Strep A+ FIA detects Group AStreptococcal antigens from throat swabs frompatients with signs and symptoms of pharyngitis, suchas sore throat. All negative test results should beconfirmed by bacterial culture because negativeresults do not preclude Group A Strep infection andshould not be used as the sole basis for treatment.The test is intended for professional and laboratoryuse as an aid in the diagnosis of Group AStreptococcal infection.
Physiologic Basis of theTest:Group A Streptococcus is one of the most commoncauses of acute upper respiratory tract infection. Earlydiagnosis and treatment of Group A Streptococcalpharyngitis has been shown to reduce the severity ofsymptoms and serious complications such asrheumatic fever and glomerulonephritis.Conventional procedures for identification of Group AStreptococcus from throat swabs involve the culture,isolation, and subsequent identification of viablepathogen at 24 to 48 hours or longer for results.

{5}------------------------------------------------

Device Description: The Sofia Strep A FIA employs immunofluorescence technology that is used with the Sofia analyzer (Sofia) to detect Group A Streptococcal antigen.

The Sofia Strep A FIA involves the extraction of the antigenic components of the Group A Streptococcus (GAS) bacteria. The patient's swab sample is placed in the Reagent Tube containing the Reagent Solution, during which time the bacterial antigens are extracted, making them more accessible to the specific antibodies. An aliquot of the extracted sample is dispensed into the Cassette sample well. From the sample well, the sample migrates through a test strip containing various unique chemical environments. If Group A Streptococcal antigens are present, they will be bound by antibodies coupled to fluorescent microparticles that migrate through the test strip. The fluorescent microparticles containing bound antigen will be captured by antibodies at a defined location on the test strip where they are detected by Sofia. If antigens are not present, the fluorescent microparticles will not be trapped by the capture antibodies nor detected by Sofia.

Note: The Cassette, now containing the sample, is placed directly inside Sofia for automatically timed development (WALK AWAY Mode).

Sofia scans, measures, and interprets the immunofluorescent signal using method-specific algorithms. Sofia will display the test results (Positive, Negative, or Invalid) on the screen. The results can also be automatically printed on an integrated printer if this option is selected, or transmitted via an LIS connection.

{6}------------------------------------------------

Device Comparison:

Note: The shaded cells in Table 1 below identify where there are differences between the proposed and predicate devices.

ItemProposed DevicePredicate Device
FeaturesSofia Strep A+ FIA with SofiaSofia Strep A FIA with Sofia
Intended UseThe Sofia® Strep A+ FIA detectsGroup A Streptococcal antigensfrom throat swabs from patientswith signs and symptoms ofpharyngitis, such as sore throat.All negative test results should beconfirmed by bacterial culturebecause negative results do notpreclude Group A Strep infectionand should not be used as thesole basis for treatment. The testis intended for professional andlaboratory use as an aid in thediagnosis of Group AStreptococcal infection.The Sofia Strep A FIA employsimmunofluorescence technology todetect Group A Streptococcalantigens from throat swabs ofsymptomatic patients. All negativetest results should be confirmed bybacterial culture because negativeresults do not preclude Group AStrep infection and should not beused as the sole basis fortreatment. The test is intended forprofessional and laboratory use asan aid in the diagnosis of Group AStreptococcal infection.
FDA FileNumberK141775K123793
ManufacturerQuidel Corporation and LREQuidel Corporation and LRE
RegulationNumber21 CFR 866.3740 and21 CFR 866.256021 CFR 866.3740 and21 CFR 866.2560
ClassificationProduct CodeGTY and KHOGTY and KHO
InstrumentSofiaSofia
AnalyteGroup A StreptococcusGroup A Streptococcus
AutomatedAnalysisYesYes
Mode(s)Walk-Away onlyRead-Now and Walk-Away
Read ResultsRead results on instrument screenor print with optional printerRead results on instrument screenor print with optional printer
CalibratorYes - Calibration Cassette andQC Card providedYes - Calibration Cassette andQC Card provided
Read ResultTime5 Minutes5 Minutes
SpecimenTypesThroat swabThroat swab
QualitativeYesYes
Test PrincipleImmunofluorescence DeviceImmunofluorescence Device
ItemProposed DevicePredicate Device
FeaturesSofia Strep A+ FIA with SofiaSofia Strep A FIA with Sofia
FormatLateral-flow Test CassetteLateral-flow Test Cassette
AntibodiesUsedPolyclonal rabbit antibodies thatare specific to Group AStreptococcusPolyclonal rabbit antibodies thatare specific to Group AStreptococcus
DetectionParticlePolystyrene microparticles dyedwith Europium chelatePolystyrene microparticles dyedwith Europium chelate
StorageRoom TemperatureRoom Temperature
ReagentSolutionOne reagent bottle containingsodium nitrite and hydrochloricacid in glass ampouleOne reagent bottle containingsodium nitrite and acetic acid inglass ampoule
TransferDeviceFixed volume pipette used totransfer patient sample mixed withreagent into Test CassetteFixed volume pipette used totransfer patient sample mixed withreagent into Test Cassette
ExternalControlsTest kit contains Positive andNegative Control SwabsTest kit contains Positive andNegative Control Swabs
Quality ControlFeaturesBuilt-in features include:• Built-in procedural controlzone scanned by theanalyzer to determinewhether adequate flowoccurred• Analyzer prevents used orexpired cartridge from beingread by the reader• Cassette properly insertedBuilt-in features include:• Built-in procedural controlzone scanned by theanalyzer to determinewhether adequate flowoccurred• Analyzer prevents used orexpired cartridge from beingread by the reader• Cassette properly inserted

Table 1

{7}------------------------------------------------

Summary of Performance Data:

Sofia Strep A+ FIA Performance vs. Cell Culture and vs. Cell Culture Resolved by PCR

The performance of the Sofia Strep A+ FIA was compared to standard bacterial culture and identification and an FDA-cleared Group A Streptococcus RT-PCR assay in a multi-center clinical field study. This study was conducted by untrained health care personnel during 2014 at 7 distinct CLIA-waived sites in various geographical regions within the United States. In this multicenter, point-of-care (POC) field trial, two (2) throat swabs were collected from eight hundred fiftyone (851) patients with symptoms suggestive of bacterial pharyngitis.

One throat swab was tested fresh at the CLIA-waived site in the Sofia Strep A+ FIA. A second swab was placed into transport medium and transported on cold ice packs to a central Reference Laboratory. The swab was streaked on a sheep blood agar plate (SBA) and cultured for up to 48 hours. A portion of the transport medium was subsequently tested in the PCR assay. The performance of the Sofia Strep A+ FIA was determined by comparison of the rapid FIA test result to the corresponding culture result (Table 2).

{8}------------------------------------------------

Table 2 Sofia Strep A+ FIA Performance Compared to Culture

Culture
PosNegSensitivity
Sofia Pos16438*93.7% (164/175)(95%CI=89.1%-96.5%)
Sofia Neg11**638Specificity
94.4% (638/676)(95% CI=92.4%-95.9%)
Total:175676PPV
81.2% (164/202)
NPV
98.3% (638/649)

*Of the 38 discordant specimens, 24 of these specimens were positive for GAS when tested with an FDA-cleared molecular device, 14 were negative.

**Of the 11 discordant specimens, 3 were negative when tested with an FDA-cleared molecular device, 8 were positive.

Reproducibility Studies

The reproducibility of the Sofia Strep A+ FIA was evaluated at 3 different laboratories. Two different operators at each site tested a series of coded, contrived samples, prepared in negative clinical matrix, ranging from negative (no bacteria) to moderate positive (3 x LoD) Group A Streptococcus. The inter-laboratory agreement (Table 3) for negative samples was 90-100% and 87-100% for positive samples. The intra-laboratory agreement (Table 4) for all samples ranged from 93-95%.

SiteNegative*(C₀)High Negative*(C₅)Low Positive**(C₉₅)Mod Positive**(C₁₀₀)
130/3027/3027/3030/30
230/3029/3023/3030/30
330/3025/3028/3030/30
Total90/9081/9078/9090/90
% OverallAgreement(95% CI)100%(95.9-100.0%)90%(82.1-94.7%)87%(78.1-92.2%)100%(95.9-100.0%)

Table 3 Sofia Strep A+ FIA Reproducibility Study Inter-laboratory Agreement

*Bacteria not detected/total

**Bacteria detected/total

{9}------------------------------------------------

Table 4 Sofia Strep A+ FIA Reproducibility Study Intra-laboratory Agreement

SiteNegative*(Co)High Negative*(Cis)Low Positive**(C95)Mod. Positive**(C100)% OverallAgreement(95% CI)
130/3027/3027/3030/3095% (114/120)(89.5-97.7%)
230/3029/3023/3030/3093% (112/120)(87.4-96.6%)
330/3025/3028/3030/3094% (113/120)(88.5-97.2%)

*Bacteria not detected/total

**Bacteria detected/total

Limit of Detection

The limit of detection (LoD) for the Sofia Strep A+ FIA was determined using 3 strains of Group A Streptococcus pyogenes. The LoD ranged from 2.76E+03 to 8.13E+03 colony forming units (cfu)/test (Table 5).

Table 5 Sofia Strep A+ FIA Limits of Detection

StrainMinimum Detectable Level*
Bruno [CIP 104226]4.00E+03 cfu/test
CDC-SS-14028.13E+03 cfu/test
CDC-SS-14602.76E+03 cfu/test

cfu/test = colony forming units/test

*The levels of bacteria were determined by limiting dilution, bacterial culture, and colony counting to give cfu/test.

{10}------------------------------------------------

Analytical Reactivity

Analytical reactivity for the Sofia Strep A+ FIA was demonstrated using 21 strains of Group A Streptococcus pyogenes tested at 1.74E+04 colony forming units (cfu)/test (Table 6).

Streptococcus pyogenes Strain
Strain #1 (ATCC-19615)
Strain #2 (ATCC-700942)
Strain #3 (ATCC-700952)
Strain #4 (Clinical Isolate-52123)
Strain #5 (Clinical Isolate-52120)
Strain #6 (Clinical Isolate-62055)
Strain #7 (Clinical Isolate-52152)
Strain #8 (Clinical Isolate-62092)
Strain #9 (Clinical Isolate-52151)
Strain #10 (ATCC-700482)
Strain #11 (ATCC-BAA-1315)
Strain #12 (ATCC-700459)
Strain #13 (ATCC-12203)
Strain #14 (ATCC-700944)
Strain #15 (Clinical Isolate-52154)
Strain #16 (Clinical Isolate-5036)
Strain #17 (Clinical Isolate-5095)
Strain #18 (Clinical Isolate-5017)
Strain #19 (Clinical Isolate-5060)
Strain #20 (Clinical Isolate-5112)
Strain #21 (Clinical Isolate-5008)

Table 6 Analytical Reactivity

Analytical Specificity

Cross Reactivity

The cross reactivity of the Sofia Strep A+ FIA was evaluated with a total of 61 non-Group A Streptococcus bacterial and fungal microorganisms, and 26 viral isolates. None of the microorganisms or viruses listed below in Table 7 showed any sign of cross reactivity in the assay. The same microorganisms and viruses in Table 7 were pre-mixed with Group A Strep and tested in the Sofia Strep A+ FIA.

Table 7
Cross Reactivity
Organism/VirusTest Concentration**
Arcanobacterium haemolyticum3.00E+05 cfu/test
Bacteroides fragilis3.00E+07 cfu/test
Bordetella pertussis3.00E+07 cfu/test
Candida albicans3.00E+04 cfu/test
Corynebacterium diphtheriae3.00E+05 cfu/test
Corynebacterium pseudodiphtheriticum*3.00E+06 cfu/test
Enterococcus faecalis*1.40E+06 cfu/test
Organism/VirusTest Concentration**
Enterococcus faecium3.00E+06 cfu/test
Escherichia coli1.50E+07 cfu/test
Fusobacterium necrophorum3.00E+06 cfu/test
Haemophilus influenzae3.00E+07 cfu/test
Haemophilus parahaemolyticus3.00E+06 cfu/test
Klebsielle pneumoniae3.00E+07 cfu/test
Moraxella catarrhalis3.00E+06 cfu/test
Neisseria gonorrhoeae3.00E+06 cfu/test
Neisseria lactamica3.00E+06 cfu/test
Neisseria meningitidis3.00E+06 cfu/test
Neisseria sicca3.00E+07 cfu/test
Neisseria subflava3.00E+07 cfu/test
Proteus vulgaris3.00E+07 cfu/test
Pseudomonas aeruginosa3.00E+06 cfu/test
Serratia marcescens3.00E+07 cfu/test
Staphylococcus aureus*3.00E+06 cfu/test
Staphylococcus epidermidis3.00E+06 cfu/test
Staphylococcus haemolyticus3.00E+05 cfu/test
Staphylococcus intermedius3.00E+05 cfu/test
Staphylococcus saprophyticus3.00E+06 cfu/test
Streptococcus anginosus3.00E+06 cfu/test
Streptococcus gordonii3.00E+04 cfu/test
Streptococcus mitis3.00E+04 cfu/test
Streptococcus mutans*3.00E+06 cfu/test
Streptococcus oralis3.00E+06 cfu/test
Streptococcus parasanginis*3.00E+06 cfu/test
Streptococcus pneumoniae3.00E+06 cfu/test
Streptococcus salivaris3.00E+05 cfu/test
Streptococcus sanguinis3.00E+06 cfu/test
Streptococcus Group B Strain #1:Streptococcus agalactiae3.00E+06 cfu/test
Streptococcus Group B Strain #23.00E+06 cfu/test
Streptococcus Group B Strain #33.00E+06 cfu/test
Streptococcus Group B Strain #43.00E+06 cfu/test
Streptococcus Group B Strain #53.00E+06 cfu/test
Streptococcus Group C Strain #13.00E+06 cfu/test
Streptococcus Group C Strain #23.00E+06 cfu/test
Streptococcus Group C Strain #33.00E+06 cfu/test
Streptococcus Group C Strain #4:Streptococcus dysgalactiae*3.00E+06 cfu/test
Organism/VirusTest Concentration**
Streptococcus Group C Strain #53.00E+05 cfu/test
Streptococcus Group D Strain #1:Enterococcus casseliflavus3.00E+06 cfu/test
Streptococcus Group D Strain #23.00E+06 cfu/test
Streptococcus Group D Strain #3*3.00E+06 cfu/test
Streptococcus Group D strain #4:Enterococcus faecalis3.00E+06 cfu/test
Streptococcus Group D strain #5:Enterococcus faecalis3.00E+06 cfu/test
Streptococcus Group F Strain #11.00E+05 cfu/test
Streptococcus Group F Strain #23.00E+06 cfu/test
Streptococcus Group F Strain #31.00E+06 cfu/test
Streptococcus Group F Strain #4*3.00E+05 cfu/test
Streptococcus Group F Strain #53.00E+05 cfu/test
Streptococcus Group G strain #1:Streptococcus dysgalactiae3.00E+07 cfu/test
Streptococcus Group G Strain #23.00E+06 cfu/test
Streptococcus Group G Strain #33.00E+06 cfu/test
Streptococcus Group G Strain #43.00E+06 cfu/test
Streptococcus Group G Strain #53.00E+06 cfu/test
Adenovirus Type 1*3.00E+11 TCID50/test
Adenovirus Type 3*3.00E+05 TCID50/test
Adenovirus Type 47.50E+03 TCID50/test
Adenovirus Type 53.00E+05 TCID50/test
Adenovirus Type 113.00E+04 TCID50/test
Coronavirus 229E3.00E+04 TCID50/test
Coronavirus OC433.00E+04 TCID50/test
Coxsackievirus B5 (Faulkner)3.00E+06 TCID50/test
Cytomegalovirus (Towne)3.00E+03 TCID50/test
Echovirus Type 31.50E+04 TCID50/test
Epstein Barr Virus (EBV)*3.00E+07 genomecopies/test
Herpes Simplex Virus 13.00E+04 TCID50/test
Herpes Simplex Virus 23.00E+04 TCID50/test
Influenza A/New Jersey/8/76 (H1N1)3.00E+04 TCID50/test
Influenza A/Victoria/3/75 (H3N2)3.00E+04 TCID50/test
Influenza B/Hong Kong/5/723.00E+04 TCID50/test
Influenza B/Panama/45/901.50E+04 TCID50/test
Influenza C/Taylor/1233/471.50E+04 TCID50/test
Measles (Edmonston)3.00E+04 TCID50/test
Mumps (Enders)*3.00E+03 TCID50/test
Organism/VirusTest Concentration**
Parainfluenza virus 13.00E+04 TCID50/test
Parainfluenza virus 21.10E+05 TCID50/test
Parainfluenza virus 36.80E+05 TCID50/test
Parainfluenza virus 4A3.00E+04 TCID50/test
Rhinovirus Type 23.00E+03 TCID50/test
Rhinovirus Type 153.00E+04 TCID50/test

{11}------------------------------------------------

{12}------------------------------------------------

{13}------------------------------------------------

cfu/test = colony forming units/test TCID50/test = 50% tissue culture infectious dose

*This organism/virus may interfere with this assay.

**The levels of bacteria were determined by limiting dilution, bacterial culture, and colony counting to give cfu/test. Virus concentrations were determined by standard virology methods, Reed-Muench.

Interfering Substances

Several over-the-counter (OTC) products, whole blood, mucin and blood agar were evaluated with the Sofia Strep A+ FIA at the levels tested (Table 8).

Table 8 Interference Testing

SubstanceConcentration
Crest Pro-Health Deep Clean Mint Mouth wash (Cetylpyridinium chloride)24% v/v
Listerine Original Antiseptic Mouth wash (Eucalyptol, Menthol, Methyl salicylate, and Thymol)24% v/v
Listerine Cool Mint Antiseptic Mouth wash (Eucalyptol, Menthol, Methyl salicylate, and Thymol)24% v/v
RiteAid Sore throat relief (Benzocaine and Menthol)24% v/v
Chloraseptic Max Sore Throat (Phenol and Glycerin)24% v/v
Dimetapp Children's Cold & Cough (Brompheniramine maleate, Dextromethorphan HBr, and Phenylephrine HCI)24% v/v
RiteAid Children's Cold & Allergy (Brompheniramine maleate and Phenylephrine HCI)24% v/v
CVS Children's Cold & Cough DM (Brompheniramine maleate, Dextromethorphan HBr, and Phenylephrine HCI)24% v/v
RiteAid tussin cough&cold mucus relief CF (Dextromethorphan HBr, Guaifenesin, and Phenylephrine HCI)24% v/v
Robitussin Max Strength Multi-Symptom CF Max (Dextromethorphan HBr, Guaifenesin, and Phenylephrine HCI)24% v/v
Robitussin Night Time Multi-Symptom Cold CF (Acetaminophen, Diphenhydramine HCl, and Phenylephrine HCI)24% v/v
Cepacol Sore Throat Cherry (Benzocaine and Menthol)24% w/v
Halls Triple Soothing Action Cherry (Menthol)24% w/v
Halls Triple Soothing Action Menthol-lyptus (Menthol)24% w/v
Ricola Natural Herb Cough Drops (Menthol)24% w/v
Sucrets Complete Vapor Cherry (Dyclonine Hydrochloride and Menthol)24% w/v
Chloraseptic Sore Throat Cherry (Phenol and Glycerin)24% w/v
BreathSavers Spearmint (Cetylpyridinium chloride)24% w/v

{14}------------------------------------------------

SubstanceConcentration
Tic Tac freshmints (Eucalyptol, Menthol, Methyl salicylate, and Thymol)24% w/v
Cheetos, Flaming Hot12% w/v
Doritos, Nacho Flavor12% w/v*
Fresh Whole Blood75 µL/swab**
Mucin4.3% w/v***
Sheep Blood Agar (5% Sheep Blood)24% w/v
Horse Blood Agar (5% Horse Blood)24% w/v

*Nacho Flavor Doritos interfered at 25% w/v

** Fresh Whole Blood interfered at 100 µL/swab

14 100n Whole Blood Intonered at 100 perowas
*** Bovine submaxillary mucin interfered at 28.7 mg/mL

Conclusion:

These studies demonstrated the substantial equivalence of the Sofia Strep A+ FIA with the Sofia to the existing product Sofia Strep A FIA with Sofia (K123793).

§ 866.3740

Streptococcus spp. serological reagents.(a)
Identification. Streptococcus spp. serological reagents are devices that consist of antigens and antisera (excluding streptococcal exoenzyme reagents made from enzymes secreted by streptococci) used in serological tests to identifyStreptococcus spp. from cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genusStreptococcus and provides epidemiological information on these diseases. Pathogenic streptococci are associated with infections, such as sore throat, impetigo (an infection characterized by small pustules on the skin), urinary tract infections, rheumatic fever, and kidney disease.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.