I-125 RSLN [RADIOACTIVE SEED AND LOCALIZATION NEEDLE]
Applicant
Isoaid, LLC
Product Code
KXK · Radiology
Decision Date
Dec 5, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5730
Device Class
Class 2
Indications for Use
The I-125 Radioactive Seed Localization Needle is intended as a temporary implant to aid in localization and excision of suspicious tissues It is intended to be used with or without absorbable strands.
Device Story
Device consists of pre-sterilized 18-gauge stainless steel needle containing low-activity I-125 Iodine seed; needle tip occluded with bone wax. Seed provided loose or stranded, with or without trailing spacers. Used by clinicians to facilitate introduction of radionuclide seed into non-palpable suspicious tissue for localization prior to surgical excision. Seed acts as point-source marker rather than therapeutic brachytherapy source. Device is single-use, EtO sterilized.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and design similarity to predicate.
Technological Characteristics
18-gauge stainless steel needle; lengths 5cm, 7cm, 12cm; I-125 radionuclide seed; bone wax tip occlusion; EtO sterilization. Meets ISO 2919, ISO 9978, and Type A packaging requirements.
Indications for Use
Indicated for the localization of suspicious tissues (non-palpable lesions) for excision using radioactive seeds.
Regulatory Classification
Identification
A radionuclide brachytherapy source is a device that consists of a radionuclide which may be enclosed in a sealed container made of gold, titanium, stainless steel, or platinum and intended for medical purposes to be placed onto a body surface or into a body cavity or tissue as a source of nuclear radiation for therapy.
Special Controls
*Classification.* Class II (special controls). A prostate seeding kit intended for use with a radionuclide brachytherapy source only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
BrachySciences (Biocompatibles) Radioactive Seed Localization Needle with Anchor Seed (K111979)
K111979 — RADIOACTIVE SEED LOCALIZATION NEEDLE · Biocompatibles, Inc. · Oct 18, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a wavy line below them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 5, 2014
IsoAid, LLC % Mary Ann Greenawalt Regulatory & Quality Director 7824 Clark Moody Blvd. PORT RICHEY FL 34668
Re: K141701 Trade/Device Name: I-125 Radioactive Seed And Localization Needle Regulation Number: 21 CFR 892.5730 Regulation Name: Radionuclide brachytherapy source Regulatory Class: II Product Code: KXK Dated: October 29, 2014 Received: October 30, 2014
Dear Ms. Greenawalt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K141701
Device Name
Radioactive Seed Localization Needle [RSLN]
Indications for Use (Describe)
The IsoAid RSLN is a device intended for the localization of suscipcious tissues (non-palpable lesions) for excision with the use of radioactive seeds.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/8 description: The image shows the word "ISOAID" in large, green letters. The letters are slightly faded, giving them a textured appearance. Below the word is a curved, orange line that adds a touch of color to the image. The overall design is simple and eye-catching.
K 141701
510 (k) SUMMARY (As required by 807.92(c))
| Submitter of 510(k): | IsoAid, LLC<br>7824 Clark Moody Blvd., Port Richey, FL 34668<br>Phone: 727-815-3262<br>Fax: 727-815-1973 | |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Contact Person: | Mary Ann Greenawalt<br>IsoAid Regulatory and Quality | |
| Date of Summary: | 23 JUNE 2014 | |
| Trade Name: | I-125 Radioactive Seed Localization Needle [IA-RSLN] | |
| Common Name: | Source, Brachytherapy, Radionuclide | |
| Classification: | Class II [21 CFR 892.5730, Product Code KXK] | |
| Classification Name: | Radionuclide Brachytherapy Source | |
| Predicate Devices: | Device | 510(k) |
| | BrachySciences (Biocompatibles) Radioactive Seed<br>Localization Needle with Anchor Seed | K111979 |
| Device Description: | The IsoAid I-125 Radioactive Seed Localization Needle [IA-RSLN] is a<br>pre-sterilized 18 gauge needle containing a low-activity I-125 Iodine Seed<br>[K011205]. The needle tip is occluded with bone wax to contain the<br>contents prior to implant. The iodine seed is loaded loose or stranded and is<br>provided with or without trailing spacers. The stainless steel needles are<br>provided in lengths: 5cm, 7cm, and 12cm. The device is used to facilitate<br>the introduction of the radionuclide seed into the suspicious tissue. The<br>seed is used singly as a point-source for localization rather than for<br>therapeutic use as in brachytherapy. The device is packaged and labeled and<br>EtO sterilized. | |
| Intended Use: | The I-125 Radioactive Seed Localization Needle is intended as a temporary<br>implant to aid in localization and excision of suspicious tissues It is<br>intended to be used with or without absorbable strands. | |
| Indications for Use: | The I-125 Radioactive Seed Localization Needle is intended for the<br>localization of suspicious tissues [non-palpable lesions] for excision with<br>the use of radioactive seeds. | |
IsoAid 510k Summary for I-125 Radioactive Seed Localization Needle
5-1 of 2
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for ISOAID. The text "ISOAID" is written in large, green letters. Below the text is an orange, curved line.
## Summary of Technological Characteristics.
The IsoAid I-125 RSLN device has the same intended use and fundamental scientific technology as the predicate device. As with the predicate device, the I-125 RSLN device is an iodine radionuclide seed provided loose or stranded, with or without a trailing spacer. The RSLN device is provided sterile for one-time use only, and meets ISO 2919, ISO 9978, and Type A packaging requirements.
### Conclusion.
Based on the above, IsoAid LLC believes the RSLN device is substantially equivalent to the predicate device, and is safe and effective for its intended use.
-end-
IsoAid 510k Summary for I-125 Radioactive Seed Localization Needle
5-2 of 2
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.