K Number
K141698

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-12-17

(176 days)

Product Code
Regulation Number
880.2800
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The STERIZONE® CI+ Chemical Indicator is a single-use process indicator intended to distinguish between processed and unprocessed packaged medical devices to be sterilized using the STERIZONE® VP4 Sterilizer. The device is intended for use only with the STERIZONE® VP4 Sterilizer, which has a single sterilization cycle ("Cycle 1"). Critical process parameters for Cycle 1 are summarized in Table 1.

The red stripe chemical indicator is located above the peach-colored stripe labeled as the "REFERENCE". After exposure to the Cycle 1 of the STERIZONE® VP4 Sterilizer, the chemical indicator color changes from red to peach-Reference color (or lighter).

Device Description

The STERIZONE® CI+ Chemical Indicator (CI+) is a Class 1 process indicator that conforms to ANSI/AAMI/ISO 11140-1: 2005(R)2010, and is intended to only be used with the STERIZONE® VP4 Sterilizer. The single pre-set cycle (Cycle 1) of the STERIZONE® VP4 Sterilizer utilizes vaporized hydrogen peroxide and ozone in a multiphase process, to rapidly sterilize a variety of reusable medical devices.

The CI+ chemical indicator consists of a polymeric material strip on which an indicator ink, a reference color and a varnish coating have been deposited. The CI+ provides a visual indication that a sterilization load has been exposed to the STERIZONE® VP4 Sterilizer Cycle 1. The indicator works by means of a chemical reaction, which results is a recognizable color change from red to peach (or lighter).

AI/ML Overview

The provided text describes the STERIZONE® CI+ Chemical Indicator and its assessment for substantial equivalence to a predicate device. It explicitly states that "Performance testing was conducted to demonstrate the functionality of the STERIZONE® CI+ Chemical Indicator and general conformance with the requirements for Class 1 vaporized hydrogen peroxide sterilization indicators as detailed in ANSI/AAMI/ISO 11140-1:2005(R)2010."

However, the document does not describe a study involving an AI device or algorithm. The "device" in question is a chemical indicator that changes color. Therefore, most of the requested information regarding AI device-specific criteria (like sample sizes for test/training sets, data provenance, number/qualification of experts, adjudication methods, MRMC studies, or standalone algorithm performance) is not applicable or cannot be extracted from this text.

I will provide the acceptance criteria and performance as described for the chemical indicator, and note where the requested information is not relevant to this type of device.


1. Table of Acceptance Criteria and Reported Device Performance:

Requirement/Test from ANSI/AAMI/ISO 11140-1:2005(R)2010Acceptance Criteria (Implicit from "Passed")Reported Device Performance
CI+ FunctionalityDevice functions as intendedPassed
Shelf-lifeMaintains performance over shelf-lifePassed
Endpoint color stabilitySustains distinctive color changePassed
ISO 11140-1 complianceMeets all requirements of the standardPassed
BiocompatibilityNo adverse biological effectsPassed

2. Sample Sizes and Data Provenance for Test Set:
The document does not specify specific sample sizes (e.g., number of indicators tested) for each non-clinical test.
Data provenance (country of origin, retrospective/prospective) is not mentioned for these tests, which are laboratory-based functionality tests for a chemical indicator, not typically tied to patient data.

3. Number and Qualifications of Experts for Ground Truth (Test Set):
Not applicable. This device is a chemical indicator with a physical color change. Its performance is assessed by physical and chemical tests against a standard, not by human expert interpretation in the same way an AI diagnostic tool would be.

4. Adjudication Method for Test Set:
Not applicable for the reasons stated above. The "ground truth" for a chemical indicator is its physical response to sterilization parameters as defined by industry standards.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
Not applicable. This is a chemical indicator, not an AI system. There are no "human readers" to improve with "AI assistance" in this context.

6. Standalone (Algorithm Only) Performance Study:
Not applicable. This device is a passive chemical indicator, not an algorithm.

7. Type of Ground Truth Used:
The ground truth used for evaluating the STERIZONE® CI+ Chemical Indicator is based on industry standards and physical/chemical properties. Specifically, performance is assessed against the requirements for Class 1 vaporized hydrogen peroxide sterilization indicators as detailed in ANSI/AAMI/ISO 11140-1:2005(R)2010. The color change (red to peach or lighter) triggered by exposure to the STERIZONE® VP4 Sterilizer's Cycle 1 is the intended functional ground truth.

8. Sample Size for Training Set:
Not applicable. This is a manufactured chemical indicator, not an AI model that requires a training set.

9. How Ground Truth for Training Set Was Established:
Not applicable. As above, no training set or AI model is involved.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, composed of three human profiles facing right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 17, 2014

TSO3 Inc. C/O Thomas Richard, Ph.D. Consultant IM3. Inc. 512F NE 81st Street. Suite 101 Vancouver, WA 98665

Re: K141698

Trade/Device Name: STERIZONE® CI+ Chemical Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Chemical Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: November 14, 2014 Received: November 17, 2014

Dear Dr. Richard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D. Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141698

Device Name STERIZONE® CI+ Chemical Indicator

Indications for Use (Describe)

The STERIZONE® CI+ Chemical Indicator is a single-use process indicator intended to distinguish between processed and unprocessed packaged medical devices to be sterilized using the STERIZONE® VP4 Sterilizer. The device is intended for use only with the STERIZONE® VP4 Sterilizer, which has a single sterilization cycle ("Cycle 1"). Critical process parameters for Cycle 1 are summarized in Table 1.

The red stripe chemical indicator is located above the peach-colored stripe labeled as the "REFERENCE". After exposure to the Cycle 1 of the STERIZONE® VP4 Sterilizer, the chemical indicator color changes from red to peach-Reference color (or lighter).

Table 1. STERIZONE® VP4 Sterilizer - Cycle 1 process parameters

Hydrogen peroxide exposureOzone exposure
Hydrogenperoxide solutionChamberdifferentialpressureset pointTimeSterilantinjectedVaporizer setpoint /ChambertemperatureO3injectionO3dwellNb ofphases
125-280 Solution™(H2O2 50 wt%)19 Torr210-600sec*8.4-24 g*120°C /41 ±3°C2 mg/L5 min2
  • Vaporized hydrogen peroxide injection/exposure time (Dynamic HoOs exposure step) varies with load composition and conditions. The quantity of vaporized hydrogen peroxide injected is directly related to the time required to reach a pressure differential of 19 Torr in the chamber, for load temperature ranging from 20°C to 26°C. If the H₂O₂ injection time is less than 210 seconds, or greater than 600 seconds, the cycle will abort.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a chemical structure. The structure contains the letters T and S. There is also a subscript number 3 in the bottom right corner of the image.

5.0 510(k) Summary

5.1. Applicant's Name and Address and submission date

Applicant's Name and Address

TSO3 Inc., 2505, avenue Dalton, Québec (Quebec) Canada G1P 3S5

TSO3 Contact Person, Telephone, FAX

Marc Chaunet, Director of Regulatory Affairs Tel : (418) 651-0003 ext.242 FAX : (418) 653-5726 E-mail: mchaunet@tso3.com

U.S. Contact

Contact: Thomas Richards, Ph.D. of IM3, Inc. Phone: 503-415-0250 Email: tomami20x@gmail.com

Submission Date

December 10, 2014

5.2. Name of the device

Trade Name

STERIZONE® CI+ Chemical Indicator

Common Name

Chemical sterilization process indicator

Classification Name (if known)

Indicator, Physical/Chemical Sterilization Process

Regulatory Class

Class II under Sterilization Process Indicator in 21 CFR 880.2800 (b) by the General Hospital and Personal Use Devices Panel. Product code: JOJ

5.3. Legally Marketed Equivalent Device Name(s)

Verify® V-PRO Chemical Indicator (Version 1A) (K091174)

{5}------------------------------------------------

Description of device 5.4.

The STERIZONE® CI+ Chemical Indicator (CI+) is a Class 1 process indicator that conforms to ANSI/AAMI/ISO 11140-1: 2005(R)2010, and is intended to only be used with the STERIZONE® VP4 Sterilizer. The single pre-set cycle (Cycle 1) of the STERIZONE® VP4 Sterilizer utilizes vaporized hydrogen peroxide and ozone in a multiphase process, to rapidly sterilize a variety of reusable medical devices.

The CI+ chemical indicator consists of a polymeric material strip on which an indicator ink, a reference color and a varnish coating have been deposited. The CI+ provides a visual indication that a sterilization load has been exposed to the STERIZONE® VP4 Sterilizer Cycle 1. The indicator works by means of a chemical reaction, which results is a recognizable color change from red to peach (or lighter).

5.5. Indication for use

The STERIZONE® CI+ Chemical Indicator is a Class 1 process indicator intended to distinguish between processed and unprocessed packaged medical devices to be sterilized using the STERIZONE® VP4 Sterilizer. The device is intended for use only with the STERIZONE® VP4 Sterilizer, which has a single sterilization cycle ("Cycle 1"). Critical process parameters for Cycle 1 are summarized in Table 1.

The red stripe chemical indicator is located above the peach-colored stripe labeled as the "REFERENCE". After exposure to the Cycle 1 of the STERIZONE® VP4 Sterilizer, the chemical indicator color changes from red to peach-Reference color (or lighter).

Hydrogen peroxide exposureOzone exposure
Hydrogenperoxide solutionChamberdifferentialpressureset pointTimeSterilantinjectedVaporizer setpoint /ChambertemperatureO3injectionO3dwellNb ofphases
125-280 Solution™(H2O2 50 wt%)19 Torr210-600sec*8.4-24 g*120°C /41 ±3°C2 mg/L5 min2

Table 1. STERIZONE® VP4 Sterilizer – Cycle 1 process parameters

  • Vaporized hydrogen peroxide injection/exposure time (Dynamic H2O2 exposure step) varies with load composition and conditions. The quantity of vaporized hydrogen peroxide injected is directly related to the time required to reach a pressure differential of 19 Torr in the chamber, for load temperature ranging from 20°C to 26°C. If the H₂O₂ injection time is less than 210 seconds, or greater than 600 seconds, the cycle will abort.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows a chemical structure diagram. The diagram depicts a molecule with the letters 'TS' on the left, followed by a central sulfur atom (S) bonded to three oxygen atoms (O). The subscript '3' indicates that there are three oxygen atoms in the molecule. The diagram is rendered in black and gray, with the 'TS' and the sulfur atom in black, and the oxygen atoms in gray.

5.6. Assessment of performances data

5.6.1. Summary of nonclinical performance tests

Performance testing was conducted to demonstrate the functionality of the STERIZONE® CI+ Chemical Indicator and general conformance with the requirements for Class 1 vaporized hydrogen peroxide sterilization indicators as detailed in ANSI/AAMI/ISO 11140-1:2005(R)2010.

Table 2. Summary of nonclinical tests performed to demonstrate Safety and Effectiveness of the STERIZONE® CI+ Chemical Indicator

TestResult
1CI+ FunctionalityPassed
2Shelf-lifePassed
3Endpoint color stabilityPassed
4ISO 11140-1 compliancePassed
5BiocompatibilityPassed

5.7. Substantial equivalence

The STERIZONE® CI+ Chemical Indicator is substantially equivalent to the Verify® V-PRO Chemical Indicator (K091174). The CI+ and predicate device are both single-use process indicators intended to monitor sterilization cycles that utilize vaporized hydrogen peroxide as a primary sterilant. Furthermore, the CI+ and predicate device are identical in design, both utilizing a dye, which when exposed to hydrogen peroxide, changes color. The difference in endpoint color change between exposed and non-exposed indicators is minor, and does not raise different questions of safety and effectiveness.

Table 3. Comparison between the STERIZONE® CI+ Chemical Indicator and the Verify® V-PRO Chemical Indicator (K091174)

FeatureSTERIZONE®CI+ Chemical IndicatorVerify® V-PRO ChemicalIndicator (Version 1A)(K091174)
Intended UseSingle use sterilization processindicatorSingle use sterilization processindicator
Indications for useThe STERIZONE® CI+ ChemicalIndicator is a single-use processA process indicator intended todistinguish between processed and
FeatureSTERIZONE®CI+ Chemical IndicatorVerify® V-PRO ChemicalIndicator (Version 1A)(K091174)
indicator intended to distinguishbetween processed and unprocessedpackaged medical devices to besterilized using Cycle 1 of theSTERIZONE® VP4 Sterilizer.unprocessed units when placedwithin sterilization wraps, trays orpouches to indicate, through avisible change from magenta toyellow, when the device has beenexposed to a V-PRO 1 LowTemperature Sterilization process(Lumen Cycle) or V-PRO 1 PlusLow Temperature sterilizationprocess (Lumen and Non-Lumencycle).
Endpointspecifications -Distinctive ColorChangeRed stripe changes to 'peach'(reference) stripe or a lighter colorwhen exposed to the STERIZONE®VP4 Sterilizer.Magenta spot changes to yellow(reference) when exposed to theAmsco V-PRO 1 or V-PRO 1 PlusLow Temperature Sterilizationsystem
Indicator agentRosolic acid baseImage: Rosolic acid base chemical structurePararosaniline base1Image: Pararosaniline base chemical structure
SubstratePolymer (Polypropylene)Polymer (Polypropylene)
TechnologyfeaturesThe molecule structure comprisesthree conjugated rings.Chromophore functional group isassociated with the molecule. Theenergy difference between twodifferent molecular orbitals in thechromophore is within the visiblespectrum. If the bond is broken dueto H2O2 exposure, the colorchanges.Same
FeatureSTERIZONE®CI+ Chemical IndicatorVerify® V-PRO ChemicalIndicator (Version 1A)(K091174)
Principles ofoperationRedox Reaction. C=C and C=O arethe two double bonds present in thechromophore of the molecule. Theinitial color is affected by the C=Ogroup, which is part of the overallconjugated system. When exposedto H2O2, the C=C is broken due tothe action of HO2- (H2O2 anion).This results in a color change.Redox Reaction. C=C and C=N arethe two double bonds present in thechromophore of the molecule. Theinitial color is affected by the C=Ngroup, which is part of the overallconjugated system. When exposedto H2O2, the C=C is broken due tothe action of HO2- (H2O2 anion).This results in a color change.
Sterilizationmethod and cyclesSTERIZONE® VP4 Sterilizerutilizes both vaporized hydrogenperoxide and ozone in a single pre-set cycle (Cycle 1), which isintended for general instruments,single-channel flexible endoscopes,and rigid-channel devices includingsingle-channel and double-channelrigid endoscopes.V-PRO 1 Low TemperatureSterilization process utilizesvaporized hydrogen peroxide inmultiple cycle configurations(Lumen Cycle and Non-LumenCycle), which are intended fordifferent medical devices.
Dedicated UseYes, in the STERIZONE® VP4SterilizerYes, in the V-PRO® Sterilizationsystem
CI can be used bothinternally (withinpackage) andexternallyYesYes
Ease of ResultsInterpretationReference color for end point is astripe identified “REFERENCE”.No external color reference isrequired.Reference color for end point is ayellow colored spot printed aside ofthe indicator ink spot. No externalcolor reference is required.
Storage ConditionsDry area, ambient temperature of15° to 30° CTemperature: 6° to 30° CRH: 30 to 60%
Picture of thedeviceUnprocessedImage: STERIZONE CI+ Chemical Indicator UnprocessedImage: Verify V-PRO Chemical Indicator Unprocessed
ProcessedImage: STERIZONE CI+ Chemical Indicator ProcessedImage: Verify V-PRO Chemical Indicator Processed

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a chemical structure diagram. The diagram depicts a molecule with the letters 'T' and 'S' connected to a circle with the number '3' as a subscript. The circle is connected to the 'S' in the diagram. The diagram is a simplified representation of a chemical compound.

SECTION 5 – 510(k) SUMMARY K141698

¹ US Patent 2009/0023217, assigned to Steris, describes the use of pararosaniline base (a triphenylmethane dye) as an indicator agent for exposure to hydrogen peroxide.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows a chemical structure diagram. The diagram includes the letters 'T' and 'S', with 'O' and '3' also visible. The diagram appears to represent a chemical compound or molecule, with the letters and numbers indicating the elements and their quantities within the structure.

SECTION 5 – 510(k) SUMMARY

{9}------------------------------------------------

$$\mathbb{S}^{\mathbb{S}}$$

SECTION 5 – 510(k) SUMMARY K141698

Overall Performance Conclusions 5.8.

Performance tests demonstrate that the STERIZONE® CI+ Chemical Indicator is suitable to distinguish between processed and unprocessed packaged medical devices to be sterilized using Cycle 1 of the STERIZONE® VP4 Sterilizer. The STERIZONE® CI+ Chemical Indicator is as safe and as effective as the Verify® V-PRO Chemical Indicator (Version 1A) (K091174).

Both the subject and predicate indicators have the same intended use, technical characteristics and performance. Based on the non-clinical performance testing data the subject device has proven to be as safe and effective as the predicate device (V-PRO Chemical Indicator (Version 1A) (K091174), when monitoring the Cycle 1 sterilization process for the STERIZONE VP4 Sterilizer.

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).