K Number
K141673
Date Cleared
2014-09-04

(73 days)

Product Code
Regulation Number
876.1075
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stryker IVAS Bone Biopsy Kit can be used as a biopsy tool to remove a sample of bone tissue from a vertebral body for diagnostic purposes using a coring, cutting or aspiration technique.

Device Description

The Stryker® Inflatable Vertebral Augmentation System (iVAS®) 8 Gauge Bone Biopsy Kit may be used as a biopsy tool to remove a sample of bone tissue from a vertebral body for diagnostic purposes using a coring, cutting or aspiration technique.

The Stryker® iVAS® 8 gauge bone biopsy kit is a disposable device, intended for single-use, individually packaged and provided sterile. The biopsy device will be sold in a kit consisting of a coaxial biopsy needle, obturator and syringe. The biopsy kit is used with access cannula and stylet accessories of compatible size. The access cannula and stylet are sold separately and are Class I exempt under 21 CFR 888.4200, product code OAR (Injector, vertebroplasty (does not contain cement).

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Non-Clinical TestingAcceptance CriteriaReported Device Performance
Impact testingBiopsy needle to withstand impact loads of at least 25 cycles at 2-in-lb.Not explicitly stated, but implied to have met criteria. The document states, "Stryker has determined that the device modification to the predicate device(s) raises no new questions of safety or effectiveness."
Tensile force testingBiopsy needle and obturator to withstand a force measuring less than 15-lbs.Not explicitly stated, but implied to have met criteria.
Non-axial torque testingBiopsy needle handle to withstand a force measuring less than 25-in-lb.Not explicitly stated, but implied to have met criteria.
Axial torque testingBiopsy needle handle to withstand a force measuring less than 3.25-in-lb.Not explicitly stated, but implied to have met criteria.
Functional testingNon-swaged biopsy needle must retain an equivalent or greater average subjective size measurement than the swaged biopsy needle.Not explicitly stated, but implied to have met criteria. The document states, "Verification and validation testing have been conducted to ensure there are no new concerns of safety and effectiveness."
Sterilization testingSterilization method is VDmax25, in compliance with the ISO 11137-1 and ISO11137-2 standards, acceptance criteria includes:Bioburden less than 1000 CFU Less than 2 positive tests of sterility in product verification dose experiment SAL 10-6Not explicitly stated, but implied to have met criteria.

Important Note: The document consistently states that testing was performed to mitigate risk and ensure no new safety or effectiveness concerns, indicating that all acceptance criteria were met, even if the specific numerical results are not listed.

2. Sample Size Used for the Test Set and Data Provenance

The document does not provide specific sample sizes for each non-clinical test. It generally refers to "verification and validation testing" without detailing the number of units tested for each criterion.

  • Test Set Sample Size: Not specified.
  • Data Provenance: The testing was non-clinical (laboratory/bench testing) performed by Stryker Instruments. The country of origin for the data is implicitly the USA (Kalamazoo, Michigan), where Stryker Instruments is located. The tests were prospective as they were performed to evaluate the modified device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Given that this is a 510(k) submission for a physical medical device (bone biopsy kit) and involves non-clinical bench testing, the concept of "experts establishing ground truth for a test set" in the context of image analysis or diagnostic performance is not directly applicable.

The "ground truth" for the non-clinical tests is established by the engineering specifications and regulatory standards (e.g., ISO 14971, ISO 11137-1/2) that define the performance metrics and their acceptance limits. The experts involved would be the design engineers, quality assurance personnel, and regulatory affairs specialists within Stryker who designed the tests, conducted them, and interpreted the results against these established standards. Their qualifications would be in engineering, materials science, and medical device regulations. The document does not specify the exact number of such individuals involved.

4. Adjudication Method for the Test Set

Again, for non-clinical bench testing of a physical device, an "adjudication method" as typically understood in studies involving human interpretation (e.g., 2+1 consensus) is not directly relevant.

The "adjudication" in this context would involve:

  • Verification and validation testing protocols: These define how each test is to be performed and how results are to be interpreted.
  • Quality control procedures: Internal company procedures ensure that testing is conducted correctly and documented accurately.
  • Review and approval processes: Engineering and quality management teams would review the test results against the acceptance criteria. Any discrepancies or failures would lead to further investigation or device redesign.

The document states that a "Risk Management File (RMF), in compliance with EN ISO 14971:2012" was completed, indicating a structured process for risk assessment and verification activities.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical testing was deemed necessary for this 510(k)." Therefore, there is no information on how much human readers improve with AI vs. without AI assistance, as AI is not part of this device or its evaluation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, a standalone algorithm performance study was not done. This device is a physical bone biopsy kit, not an algorithm or AI-powered system.

7. The Type of Ground Truth Used

For the non-clinical tests, the "ground truth" is based on:

  • Engineering specifications: Design requirements for physical strength, torque resistance, impact resistance, and functional performance (e.g., sample size retention).
  • Regulatory standards: Compliance with international standards for sterilization (ISO 11137-1 and ISO 11137-2 for VDmax25, SAL 10-6) and risk management (EN ISO 14971:2012).
  • Predicate device characteristics: The modified device's performance is compared to the established performance and safety profile of the predicate device (Stryker Bone and Vertebral Body Biopsy Kit K032943) to ensure substantial equivalence.

8. The Sample Size for the Training Set

This is not applicable. This is a physical medical device, not an AI/ML algorithm that requires a "training set."

9. How the Ground Truth for the Training Set Was Established

This is not applicable, as there is no training set for this physical device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three faces in profile, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 4, 2014

Stryker Instruments Ms. Brittney M. Larsen Senior Regulatory Affairs Representative 4100 E. Milham Avenue Kalamazoo, Michigan 49001

Re: K141673

Trade/Device Name: Stryker iVAS 8 Gauge Bone Biopsy Kit Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology-biopsy instrument Regulatory Class: Class II Product Code: KNW Dated: July 7, 2014 Received: August 8, 2014

Dear Ms. Larsen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

for

Sincerely yours,

David Krause -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S.

Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141673

Device Name Stryker® iVAS® 8 Gauge Bone Biopsy Kit

Indications for Use (Describe)

The Stryker IVAS Bone Biopsy Kit can be used as a biopsy tool to remove a sample of bone tissue from a vertebral body for diagnostic purposes using a coring, cutting or aspiration technique.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

4100 E. Milham Avenue Kalamazoo, MI 49001
t: 269 323 7700 f: 269 324 5412 www.stryker.com

Image /page/3/Picture/1 description: The image shows the Stryker Instruments logo. The word "stryker" is in bold, sans-serif font, with a registered trademark symbol to the right. Below the word "stryker" is a horizontal line, and below that is the word "Instruments" in a smaller, sans-serif font.

510(k) Summary

I. Contact Details

a.510(k) Owner:Stryker Instruments4100 E. Milham AvenueKalamazoo, Michigan 49001USAPh: +1-269-323-7700Fax: +1-269-324-5412
b.FDA EstablishmentRegistration Number:1811755
c.Contact Person:Brittney M. LarsenPh: +1-269-389-5293Fax: +1-269-389-5412Brittney.Larsen@Stryker.com
d.Date Submitted:June 20, 2014
II.Device Name
a.Trade Name:Stryker® iVAS® 8 Gauge Bone Biopsy Kit
b.Common Name:Vertebral Bone Biopsy Needles
c.ClassificationName:Gastroenterology-urology biopsy instrument(21 CFR 876.1075, Product code KNW)
d.Classification:II

III. Legally Marketed Predicate Device(s)

510(k) NumberProduct CodeTrade NameManufacturer
K032943KNWStryker Bone and VertebralBody Biopsy KitStrykerInstruments

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Stryker Instruments. The word "stryker" is in bold, sans-serif font, with a registered trademark symbol to the right of the word. Below the word "stryker" is a horizontal line, and below that is the word "Instruments" in a smaller, sans-serif font.

4100 E. Milham Avenue Kalamazoo. MI 49001 t: 269 323 7700 f: 269 324 5412 www.stryker.com

-Purpose of the Special 510(k) Submission

Stryker submits this Special 510(k): Device Modification to request clearance for a modification to our Stryker Bone and Vertebral Body Biopsy Kit(s). The predicate device(s) are currently cleared for 10-13gauge (G) sizes. The subject modification is to add an 8-G size product offering. which is outside the currently cleared size range. This device modification does not change the intended use, indications for use or the fundamental scientific technology of the predicate device(s).

II. Device Description

The Stryker® Inflatable Vertebral Augmentation System (iVAS®) 8 Gauge Bone Biopsy Kit mav be used as a biopsy tool to remove a sample of bone tissue from a vertebral body for diagnostic purposes using a coring, cutting or aspiration technique.

The Stryker® iVAS® 8 qauge bone biopsy kit is a disposable device, intended for single-use, individually packaged and provided sterile. The biopsy device will be sold in a kit consisting of a coaxial biopsy needle, obturator and syringe. The biopsy kit is used with access cannula and stylet accessories of compatible size. The access cannula and stylet are sold separately and are Class I exempt under 21 CFR 888.4200, product code OAR (Injector, vertebroplasty (does not contain cement).

III. Indications for use

The Stryker iVAS Bone Biopsy Kit can be used as a biopsy tool to remove a sample of bone tissue from a vertebral body for diagnostic purposes using a coring, cutting or aspiration technique.

IV. Substantial Equivalence Comparison

The Stryker® iVAS® 8 Gauge Bone Biopsy Kit is substantially equivalent to the Stryker Bone and Vertebral Body Biopsy Kit predicate device(s) (K032943). Stryker claims this equivalence because the Stryker® iVAS® 8 Gauge Bone Biopsy Kit has an equivalent intended use, mechanism for use, and mode of action as compared to the predicate device(s). The differences between the Stryker® iVAS® 8 Gauge Bone Biopsy Kit and the predicate device(s) have been identified and explained in Table 2, the substantial equivalence comparison matrix.

{5}------------------------------------------------

4100 E. Milham Avenue
Kalamazoo, MI 49001
t: 269 323 7700 f: 269 324 5412
www.stryker.com

Image /page/5/Picture/1 description: The image shows the Stryker Instruments logo. The word "stryker" is written vertically in a bold, sans-serif font. To the right of the word "stryker" is the word "Instruments" written vertically in a smaller, sans-serif font.

Table 2. Substantial equivalence comparison matrix

Table 2. Substantial equivalence comparison matrix
DescriptionStryker Bone and Vertebral Body Biopsy Kit(s) (Predicate)Stryker® IVAS® 8 Gauge Bone Biopsy Kit (Subject)Explanation of Differences
General Information
ClassificationClass IIClass IIIdentical
Regulation21 CFR 876.1075, Gastroenterology-urology biopsy instrument21 CFR 876.1075, Gastroenterology-urology biopsy instrumentIdentical
Product codeKNW, Instrument, BiopsyKNW, Instrument, BiopsyIdentical
Review panel78 Gastroenterology/Urology78 Gastroenterology/UrologyIdentical
Indications for useStryker's Bone and Vertebral Body Biopsy Kits can be used as a biopsy tool to remove a sample of bone tissue from a vertebral body for diagnostic purposes using a coring, cutting or aspiration technique.The Stryker iVAS Bone Biopsy Kit can be used as a biopsy tool to remove a sample of bone tissue from a vertebral body for diagnostic purposes using a coring, cutting or aspiration technique.Identical
UseSingle-useSingle-useIdentical
Anatomical sitesBone and vertebral bodyBone and vertebral bodyIdentical
Visualization techniqueFluoroscopyFluoroscopyIdentical
Method of placementPercutaneousPercutaneousIdentical
Mechanics of actionManualManualIdentical
Mode of actionSuctionSuctionIdentical
Shelf life2 years1 yearSimilar – the intent is to extend the shelf life for the subject device upon successful completion of aging studies.
DescriptionStryker Bone and Vertebral BodyBiopsy Kit(s)(Predicate)Stryker® IVAS® 8 Gauge BoneBiopsy Kit(Subject)Difference
Size Information
Penetrationdepth5 - 7.5-in.5 - 7.5-in.Identical
Biopsy needleand ObturatorCompatible with 10, 11 and 13-gaugeaccess cannulae/styletsCompatible with 8-gauge accesscannula/styletThe subject device biopsy needlediameter increased to accommodateuser preference for an 8-gauge biopsysystem. Verification and validationtesting have been conducted to ensurethere are no new concerns of safetyand effectiveness.
Patient Contacting Material Information
Biopsy needleStainless steelStainless steelIdentical
Performance Information
Tensile force(Pull test)Withstand < 15-lbs. of forceWithstand < 15-lbs. of forceIdentical
Biopsy needletipSwagedNot swagedSubject biopsy needle does not exhibita swaged, tapered, distal tip in order toincrease sample size. Verification andvalidation testing have been conductedto ensure there are no new concerns ofsafety and effectiveness.
Sterilization Information
SterilizationmethodGamma irradiation, VDmax25Gamma irradiation, VDmax25Identical
Sterilizationassurance level(SAL)10-610-6Identical
Minimumsterilizationdose25-kilograys (kGy)25-kilograys (kGy)Identical

{6}------------------------------------------------

4100 E. Milham Avenue
Kalamazoo, Ml 49001
t: 269 323 7700 f: 269 324 5412
www.stryker.com

Instrument

Table 2 continued. Substantial equivalence comparison matr

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Stryker Instruments. The word "Stryker" is in a bold, sans-serif font, with a registered trademark symbol to the right of the word. Below the word "Stryker" is a horizontal line, and below that is the word "Instruments" in a smaller, sans-serif font.

4100 E. Milham Avenue Kalamazoo. MI 49001 t: 269 323 7700 f: 269 324 5412 www.stryker.com

V. Non-clinical Testing

A risk Management File (RMF), in compliance with EN ISO 14971:2012, Medical devices - Application of risk management to medical devices, was completed to assess the impact of the subject device modification. In accordance with the risk assessment, verification and validation testing was performed in an effort to mitigate risk where possible. Testing performed includes: functional, packaging integrity and sterilization testing. See Table 3, for a summary of the non-clinical testing and acceptance criteria used to evaluate the subject device modification. Stryker has determined that the device modification to the predicate device(s) raises no new questions of safety or effectiveness.

Non-Clinical TestingAcceptance Criteria
Impact testingBiopsy needle to withstand impact loads of at least 25 cycles at 2-in-lb.
Tensile force testingBiopsy needle and obturator to withstand a force measuring less than 15-lbs.
Non-axial torque testingBiopsy needle handle to withstand a force measuring less than 25-in-lb.
Axial torque testingBiopsy needle handle to withstand a force measuring less than 3.25-in-lb.
Functional testingNon-swaged biopsy needle must retain an equivalent or greater average subjective size measurement than the swaged biopsy needle.
Sterilization testingSterilization method is VDmax25, in compliance with the ISO 11137-1 and ISO11137-2 standards, acceptance criteria includes:Bioburden less than 1000 CFU Less than 2 positive tests of sterility in product verification dose experiment SAL 10-6

Table 3. Non-clinical testing and acceptance criteria summary

VI. Clinical Testing

No clinical testing was deemed necessary for this 510(k).

VII. Conclusion / Substantial Equivalence (SE) Rationale

The Stryker® iVAS® 8 Gauge Bone Biopsy Kit is substantially equivalent in intended use, indications for use, technological characteristics, safety and effectiveness to the previously cleared Stryker Bone and Vertebral Body Biopsy

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Stryker Instruments. The word "stryker" is in bold, black letters, with a registered trademark symbol to the right of the word. Below the word "stryker" is a horizontal line, and below that is the word "Instruments" in a smaller font.

4100 E. Milham Avenue Kalamazoo, MI 49001 t: 269 323 7700 f: 269 324 5412 www.stryker.com

Conclusion/Substantial Equivalence (SE) Rationale continued.

Kit. The predicate device(s) have the same fundamental scientific technology, basic design, functional characteristics and applications. The device modification does not introduce any new concerns of safety or effectiveness. Therefore, the Stryker® iVAS® 8 Gauge Bone Biopsy Kit is substantially equivalent to the existing predicate device(s).

§ 876.1075 Gastroenterology-urology biopsy instrument.

(a)
Identification. A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification. (1) Class II (performance standards).(2) Class I for the biopsy forceps cover and the non-electric biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.