K Number
K141655

Validate with FDA (Live)

Date Cleared
2015-03-05

(258 days)

Product Code
Regulation Number
866.5100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

QUANTA Flash® Ro52 is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-Ro52 autoantibodies in human serum. The presence of anti-Ro52 autoantibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of Systemic Lupus Erythematosus, Systemic Sclerosis, Idiopathic Inflammatory Myopathies.

QUANTA Flash Ro52 Controls are intended for use with the QUANTA Flash Ro52 Reagents for quality control in the determination of IgG anti-Ro52 autoantibodies in human serum.

QUANTA Flash Ro52 Calibrators are intended for use with the OUANTA Flash Ro52 Reagents for the determination of Ig G anti-Ro52 autoantibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

Device Description

The QUANTA Flash Ro52 assay is designed to run on the BIO-FLASH® instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash Ro52 assay utilizes a reagent cartridge format, which is compatible with the BIO-FLASH instrument.

Purified recombinant Ro52 antigen is coated onto paramagnetic beads. The bead suspension is lyophilized and stored in the bead tube. Prior to use in the BIO-FLASH system, the sealed reagent tubes are pierced with the reagent cartridge lid and the beads are rehydrated and resuspended using resuspension buffer by pipetting up and down with a transfer pipette. The reagent cartridge is then loaded onto the BIO-FLASH instrument. Samples are also loaded onto the instrument in sample racks. Serum samples are prediluted by the BIO-FLASH with system rinse in a small disposable plastic cuvette. Small amounts of the diluted patient serum, the beads, and assay buffer are all combined into a second cuvette, and mixed. This cuvette is then incubated at 37°C. The beads are magnetized and washed several times. Isoluminol conjugated anti-human IgG antibodies are then added to the cuvette, and again incubated at 37°C. The beads are magnetized and washed repeatedly. The isoluminol conjugate is oxidized when Trigger 1 (Fe(II)coproporphyrin in sodium hydroxide solution) and Trigger 2 (ureahydrogen peroxide in sodium chloride solution) are added to the cuvette, and the flash of light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. The RLU are proportional to the amount of isoluminol conjugate that is bound to the human IgG, which is in turn proportional to the amount of anti-Ro52 antibodies bound to the corresponding beads.

For quantitation, the QUANTA Flash Ro52 assay utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. Every new lot number of reagent cartridge must be calibrated before first use, with the QUANTA Flash Ro52 Calibrators. Based on the results obtained with the two Calibrators included in the Calibrator Set (sold separately), an instrument specific Working Curve is created, which is used to calculate chemiluminescent units (CU) from the instrument signal (RLU) obtained for each sample.

The QUANTA Flash Ro52 kit contains the following materials:

One (1) QUANTA Flash Ro52 Reagent Cartridge
One (1) vial of Resuspension buffer
One (1) Transfer pipette

The QUANTA Flash Ro52 reagent cartridge contains the following reagents for 50 determinations:

a. Ro52 antigen coated paramagnetic beads, lyophilized.
b. Assay buffer colored pink, containing Tris-buffered saline, Tween 20, protein stabilizers and preservatives.
c. Tracer IgG - Isoluminol labeled anti-human IgG antibodies in buffer, containing protein stabilizers and preservative.

The QUANTA Flash Ro52 Calibrators kit contains two vials of Calibrator 1 and two vials of Calibrator 2:

QUANTA Flash Ro52 Calibrator 1: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human antibodies to Ro52 in buffer, protein stabilizer and preservative.
QUANTA Flash Ro52 Calibrator 2: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human antibodies to Ro52 in buffer, protein stabilizer and preservative.

The QUANTA Flash Ro52 Controls kit contains two vials of Negative Control and two vials of Positive Control:

QUANTA Flash Ro52 Negative Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to Ro52 in buffer, protein stabilizer and preservative.
QUANTA Flash Ro52 Positive Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to Ro52 in buffer, protein stabilizer and preservative.

AI/ML Overview

This document describes the QUANTA Flash® Ro52, QUANTA Flash® Ro52 Calibrators, and QUANTA Flash® Ro52 Controls, which are chemiluminescent immunoassays for the semi-quantitative determination of IgG anti-Ro52 autoantibodies in human serum. The presence of these autoantibodies aids in the diagnosis of Systemic Lupus Erythematosus, Sjögren's Syndrome, Systemic Sclerosis, and Idiopathic Inflammatory Myopathies.

Here's an analysis of the acceptance criteria and the studies performed:

1. Table of Acceptance Criteria and Reported Device Performance:

The document presents several acceptance criteria and results for various analytical performance characteristics. Below is a compilation of these, focusing on the ones directly stating "Acceptance criteria".

Test TypeAcceptance Criteria (Set by Manufacturer)Reported Device Performance
PrecisionTotal %CV: < 10%All samples (9 tested) showed Total %CV ranging from 4.3% (Sample ID 1, Reproducibility) to 8.5% (Sample ID 110689-50, Precision), and 7.1% (Sample ID 110686-6, Precision) or 8.0% (Sample ID 110689-15, Precision). All observed Total %CV values were < 10%.
ReproducibilityTotal %CV: < 10%All samples (6 tested) showed Total %CV ranging from 4.3% (Sample 1) to 6.7% (Sample 3). All observed Total %CV values were < 10%.
Linearity (Recovery)Recovery is between 80-120%, or ± 4 CU, whichever is greater.All four specimens showed dilution linearity individually (implied meeting acceptance criteria, though specific recovery ranges not tabled).
Linearity (Regression)Slope is between 0.9-1.1, and R2 is ≥ 0.95.Individual samples: Slope ranged from 0.92 to 1.02; R2 from 0.99 to 1.00. Combined data: Slope 0.97 (0.96 to 0.97), R2 1.00. All results met the criteria.
Interference85% - 115% recovery, or ± 4 CU difference, whichever is greater.Bilirubin (up to 10 mg/dL): 87% to 107%. Hemoglobin (up to 200 mg/dL): 89% to 107%. Triglycerides (up to 1000 mg/dL): 91% to 112%. Cholesterol (up to 224.3 mg/dL): 91% to 112%. RF IgM (up to 500 IU/mL): 92% to 108%. No interference was detected, indicating all recoveries were within the acceptance range.
Lot to Lot ComparisonWeighted r: ≥0.975 for linear regression0.999, 0.997, 0.997 (for the three pair-wise comparisons) - Met.
Intercept of the regression line (constant bias): ± 15% of cut-off (3 CU)1.3, 0.7, -0.6 (for the three pair-wise comparisons) - Met (all values within ±3 CU).
Slope of the regression line (proportional bias): 0.9-1.10.97, 0.94, 0.96 (for the three pair-wise comparisons) - Met.
Weighted S y/x: ≤ 0.50.043, 0.06, 0.06 (for the three pair-wise comparisons) - Met.
Predicted bias (difference) at cut-off: ±15% (3 CU)0.8, -0.6, -1.3 (for the three pair-wise comparisons) - Met (all values within ±3 CU).
Shelf Life (Beads)Lower 95% CI of regression line ≥ 85% at 2 weeks AND no individual data point ≤ 75% recovery at 2 weeks.All three lots of beads retained > 85% reactivity (considering the 95% CI) after two weeks at 37 ± 3°C, and therefore pass the acceptance criteria for one year expiration date.
Shelf Life (Cal/Controls)Lower 95% CI of regression line ≥ 90% at 2 weeks AND no individual data point ≤ 80% recovery at 2 weeks.All Calibrators and Controls maintained > 90% reactivity (considering the 95% CI) when stored at 37 ± 3°C for 2 weeks, and therefore pass the acceptance criteria for one year expiration dating.
In-use Stability (Calibrators)All five calibrations successful; Average Calibrator RLU recovery 90-110% compared to first use.5 successful calibrations over 8.5 hours. Calibrator RLU values remained within 90-110% range. Controls and patient panel samples ran within expected range, supporting 4 calibrations over 8 hours.
In-use Stability (Controls)All values run within established range; linear regression line for %recovery between 85-115% at run 15.All controls ran within acceptable ranges for all runs. Regression line remained between 85-115% at run 15 for both Controls, supporting up to 15 uses at 10 minutes per use.
Real-time Stability (Controls)Results fall within their acceptable ranges as established at release.All results were within the acceptance limits.
Real-time Stability (Calibrators)% recovery of average of triplicates 85-115% AND %CV of triplicates < 10%.All results were within the acceptance limits.
Real-time Stability (Reagent Cartridge)Results fall within their respective QC ranges.All results were within the acceptance limits.
Clinical Sensitivity (SS)Not explicitly stated as an "acceptance criteria" but presented as validation results.44.0% (95% CI: 33.6-54.8%)
Clinical Specificity (SS)Not explicitly stated as an "acceptance criteria" but presented as validation results.96.8% (95% CI: 93.5-98.7%)
Clinical Sensitivity (SLE)Not explicitly stated as an "acceptance criteria" but presented as validation results.35.9% (95% CI: 27.7-44.7%)
Clinical Specificity (SLE)Not explicitly stated as an "acceptance criteria" but presented as validation results.96.8% (95% CI: 93.5-98.7%)
Clinical Sensitivity (SSc)Not explicitly stated as an "acceptance criteria" but presented as validation results.16.3% (95% CI: 8.9-26.2%)
Clinical Specificity (SSc)Not explicitly stated as an "acceptance criteria" but presented as validation results.96.8% (95% CI: 93.5-98.7%)
Clinical Sensitivity (IIM)Not explicitly stated as an "acceptance criteria" but presented as validation results.40.0% (95% CI: 28.0-52.9%)
Clinical Specificity (IIM)Not explicitly stated as an "acceptance criteria" but presented as validation results.96.8% (95% CI: 93.5-98.7%)
Method Comparison (Total Agree)Not explicitly stated as an "acceptance criteria" but presented as validation results.89.9% (95% CI: 86.2 – 92.9%) vs predicate
Method Comparison (Pos. Agree)Not explicitly stated as an "acceptance criteria" but presented as validation results.80.3% (95% CI: 72.3 – 86.8%) vs predicate
Method Comparison (Neg. Agree)Not explicitly stated as an "acceptance criteria" but presented as validation results.95.4% (95% CI: 91.8 – 97.8%) vs predicate

2. Sample Sizes Used for the Test Set and Data Provenance:

  • Precision and Reproducibility:
    • Precision: 9 samples, each run in duplicates, twice a day for 21 days (84 replicates per sample).
    • Reproducibility: 3 samples, each run in quadruplicates, two times a day for 10 days (80 data points per sample).
    • Provenance: Not explicitly stated whether retrospective or prospective, or country of origin for these specific samples. Implied to be clinical samples ("samples containing various concentrations of Ro52 antibodies").
  • Limit of Blank (LoB) and Limit of Detection (LoD):
    • LoB: 4 blank samples (System Rinse) from two different lots, run in replicates of five on two reagent lots, once per day, for 3 days (60 data points per reagent lot).
    • LoD: 4 low level samples (prepared by diluting anti-Ro52 positive samples with System Rinse), run in replicates of five on two reagent lots, once per day, for 3 days (60 data points per reagent lot).
    • Provenance: Not specified.
  • Linearity: 4 serum samples with various Ro52 antibody concentrations, diluted in 10% increments. Assayed in duplicates.
    • Provenance: Not specified.
  • Interference: 3 specimens (negative, at cutoff, positive). Interfering substances spiked (bilirubin, hemoglobin, triglycerides, cholesterol, RF IgM). Assessed in triplicates.
    • Provenance: Not specified.
  • Cross-reactivity: 199 samples from the total 233 control samples of the clinical validation study. These were from patients with autoimmune diseases (Graves', Hashimoto, Celiac, Crohn's, Ulcerative Colitis, Primary Antiphospholipid Syndrome, Vasculitis, Rheumatoid arthritis, Behçet's disease) or positive infectious disease serology (HCV, HBV, CMV, EBV, HIV, Syphilis).
    • Provenance: Not explicitly stated, but these are referred to as "patient samples" which suggests they are clinical samples.
  • Lot to Lot Comparison: 20 unique samples and the Positive and Negative Controls (total 22 specimens), tested in triplicates.
    • Provenance: Not specified.
  • Shelf Life and In-use Stability: Involved characterized samples (up to 6 for beads, typically patient specimens for cartridge) and Calibrators/Controls.
    • Provenance: Not specified.
  • Cut-off / Reference Range Establishment: 155 subjects: 115 apparently healthy blood donors, 8 viral hepatitis positive samples, 5 HIV positive samples, 5 syphilis positive samples, 22 rheumatoid arthritis patients. All serum samples.
    • Provenance: Not explicitly stated, but implies mixed sources representing different conditions.
  • Clinical Performance (Sensitivity/Specificity - Validation Set): 600 characterized samples.
    • Patient Groups: 91 Sjögren's Syndrome (SS), 131 Systemic Lupus Erythematosus (SLE), 80 Systemic Sclerosis (SSc), 65 Idiopathic Inflammatory Myopathies (IIM).
    • Control Groups: 233 control samples (including Graves' Disease, Hashimoto Thyroiditis, Celiac Disease, Crohn's Disease, Ulcerative Colitis, HCV, HBV, CMV, EBV, HIV, Syphilis, Primary Antiphospholipid Syndrome, Secondary Antiphospholipid Syndrome*, Vasculitis, Rheumatoid arthritis, Osteoarthritis, Behçet's disease).
    • Provenance: Not explicitly stated, but these are "characterized samples" suggesting confirmed diagnoses from clinical settings.
  • Expected Values: 111 apparently healthy blood donors (90 females, 21 males, ages 17-60, average 32.6 years). This group was different from the one used to establish the cutoff.
    • Provenance: Not explicitly stated, but implies a normal healthy population.
  • Method Comparison: 319 samples from the clinical validation study + 27 additional samples having a speckled pattern on HEp-2 by ANA IIF (total 346 samples). 283 samples were within the reportable range.
    • Provenance: Clinical samples.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

The document does not specify the explicit number or qualifications of "experts" used for establishing ground truth for the test set in the clinical evaluation. The samples are described as "characterized samples," "patients with autoimmune diseases," and "healthy blood donors." This implies that the diagnosis (ground truth) for these samples was established through standard clinical diagnostic procedures, which would involve medical specialists (e.g., rheumatologists) and established laboratory tests, rather than a panel of independent experts reviewing the data solely for this study.

4. Adjudication Method for the Test Set:

Not applicable or not described. The document refers to "characterized samples" for clinical evaluation, indicating pre-existing diagnoses or classifications rather than a prospective adjudication process for this specific device study.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and Effect Size:

No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not performed. This device is an automated in vitro diagnostic (IVD) assay (chemiluminescent immunoassay) for direct measurement of autoantibodies in serum, not an imaging analysis tool typically subject to MRMC studies. The comparison is between the automated assay and a predicate ELISA device, or with clinical diagnoses.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Yes, the studies described are standalone performance evaluations of the QUANTA Flash® Ro52 assay. The device is a fully automated system that processes samples, runs the assay, and reports results without real-time human intervention in the result determination. The "performance" described throughout the document (precision, linearity, clinical sensitivity/specificity) reflects the algorithm's (assay's) performance independent of direct human interpretive input during routine operation.

7. The Type of Ground Truth Used:

The ground truth for clinical performance evaluation (sensitivity and specificity) appears to be clinical diagnosis and established laboratory markers/conditions. For example:

  • Clinical Diagnosis: Patients characterized with Sjögren's Syndrome (SS), Systemic Lupus Erythematosus (SLE), Systemic Sclerosis (SSc), Idiopathic Inflammatory Myopathies (IIM) were used as "disease" groups. Control groups were comprised of patients with other autoimmune diseases, infectious diseases, and apparently healthy individuals.
  • Predicate Device Comparison: The predicate device (QUANTA Lite® SS-A 52 ELISA) served as a comparator for method comparison studies.
  • CDC ANA Reference Sera: Used for value assignment and traceability, but not as the primary ground truth for clinical sensitivity/specificity.

8. The Sample Size for the Training Set:

The document does not explicitly use the term "training set" in the context of machine learning. However, for the development and optimization of the assay, the following samples effectively serve as "development" or "internal validation" data points:

  • Master Curve Standards: 6 standards (2.3 CU to 1685.3 CU) were established.
  • Calibrators and Controls: Human serum with high titer anti-Ro52 antibodies, diluted to achieve target CU values, and then tested on at least two instruments and two reagent lots in replicates of 10.
  • LoB/LoD: 4 blank and 4 low-level samples, run over several days and lots.
  • Linearity/Interference/Lot-to-lot: Various characterized samples (e.g., 4 for linearity, 3 for interference, 22 for lot-to-lot).

The specific "training set" for the fundamental algorithm/reagents themselves is not quantified in the typical machine learning sense, but the assay's performance characteristics were developed and internally validated using various sets of characterized samples and standards as outlined above.

9. How the Ground Truth for the Training Set was Established:

Again, assuming "training set" refers to the samples used during the development and establishment of the assay's analytical characteristics:

  • Master Curve Standards: Generated from in-house standards, traceable internally.
  • Calibrators and Controls: Manufactured by diluting human serum containing high-titer anti-Ro52 antibodies with buffer. "Target CU is achieved through trial dilutions on small scale." These are then "bulked, tested, and adjusted." Final value assignment is determined by testing on multiple instruments and reagent lots.
  • Other Analytical Studies (Precision, Linearity, Interference, Lot-to-Lot): The ground truth for these samples would be their known concentrations or characteristics (e.g., a sample known to have a certain level of Ro52 antibodies, or a known interfering substance concentration). "Characterized samples" imply that their status was determined by other validated methods prior to their use in these studies.
  • Cut-off/Reference Range: Based on the 97th percentile of results from a population of 155 subjects (healthy blood donors and patients with various conditions). This is a statistical determination from measured values, rather than an expert-adjudicated ground truth.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 5. 2015

INOVA DIAGNOSTICS, INC. DR. GABRIELLA LAKOS DIRECTOR, RHEUMATOLOGY RESEARCH 9900 OLD GROVE ROAD SAN DIEGO, CA 92131-1638

Re: K141655 Trade/Device Name: Quanta Flash® Ro52 Ouanta Flash® Ro52 Calibrators Quanta Flash® Ro52 Controls Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear antibody immunological test system Regulatory Class: II Product Code: OBE, JIT, JJX Dated: January 15, 2015 Received: January 22, 2015

Dear Dr. Lakos:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Leonthena R. Carrington -A

Leonthena R. Carrington, MS, MBA, MT(ASCP) Director (Acting) Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141655

Device Name

QUANTA Flash® Ro52, QUANTA Flash® Ro52 Controls, QUANTA Flash® Ro52 Calibrators,

Indications for Use (Describe)

QUANTA Flash Ro52 is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-Ro52 autoantibodies in human serum. The presence of anti-Ro52 autoantibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of Systemic Lupus Erythematosus, Systemic Sclerosis, Idiopathic Inflammatory Myopathies.

QUANTA Flash Ro52 Controls are intended for use with the QUANTA Flash Ro52 Reagents for quality control in the determination of IgG anti-Ro52 autoantibodies in human serum.

QUANTA Flash Ro52 Calibrators are intended for use with the OUANTA Flash Ro52 Reagents for the determination of Ig G anti-Ro52 autoantibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Inova Diagnostics, a Werfen Company. The logo consists of two stylized shapes, one red and one blue, resembling checkmarks or arrows pointing in opposite directions. To the right of the shapes is the text "Inova Diagnostics" in a sans-serif font, with the words stacked on top of each other. Below that, in a smaller font, is the text "A Werfen Company."

510(k) Summary

QUANTA Flash® Ro52 QUANTA Flash® Ro52 Calibrators QUANTA Flash® Ro52 Controls

Table of Contents

Administrative data
Device description
Intended use(s)
Substantial equivalence
Comparison to predicate device
Analytical performance characteristics
Value assignment and traceability of Calibrators and Controls
Precision
Reproducibility
Limit of Blank (LoB) and Limit of Detection (LoD)
Analytical Measuring Range (AMR)
Auto-rerun function and reportable results
High concentration hook effect
Linearity
Interference
Cross-reactivity
Lot to lot comparison
Stability
Cut-off, reference range
Clinical performance characteristics
Clinical sensitivity, specificity
Expected values
Comparison with predicate device

{4}------------------------------------------------

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Administrative data

Submitter:Inova Diagnostics, Inc9900 Old Grove Road,San Diego, CA, 92131
Purpose of submission:New device(s)
Devices in the submission:QUANTA Flash® Ro52QUANTA Flash® Ro52 CalibratorsQUANTA Flash® Ro52 Controls
Scientific contact:Gabriella Lakos, Director of Research and DevelopmentInova Diagnostics, Inc.9900 Old Grove Road, San Diego, CA, 92131Phone: 858-586-9900/1393Fax: 858-863-0025email: glakos@inovadx.com
Quality Systems contact:Ronda Elliott, VP, Quality Systems and RAInova Diagnostics, Inc9900 Old Grove Road, San Diego, CA, 92131Phone: 858-586-9900/1381Fax: 858-863-0025email: relliot@inovadx.com
Preparation date:06/09/2014
Device name (assay kit):Proprietary name: QUANTA Flash® Ro52Common name: Anti-Ro52 Chemiluminescent ImmunoassayClassification name: anti-SS-A 52 antibody, antigen and control
Regulation DescriptionAntinuclear antibody immunological test system
Regulation Medical SpecialtyImmunology
Review PanelImmunology
Product CodeOBE, Anti-SS-A 52 autoantibodies

{5}------------------------------------------------

510(k) Summary QUANTA Flash® Ro52

Regulation Number866.5100
Device Class2
Device name (Calibrators):Proprietary name:QUANTA Flash® Ro52 Calibrators
Common name:Ro52 Calibrators
Classification name:Calibrator, secondary
Regulation DescriptionCalibrator
Regulation Medical SpecialtyClinical Chemistry
Product CodeJIT
Regulation Number862.1150
Device Class2
Device name (Controls):Proprietary name:QUANTA Flash® Ro52 Controls
Common name:Ro52 Controls
Classification name:single (specified) analyte controls (assayed and unassayed)
Regulation DescriptionQuality control material (assayed and unassayed)
Regulation Medical SpecialtyClinical Chemistry
Product CodeJJX
Regulation Number862.1660
Device Class1 (reserved)
Predicate device:Device descriptionQUANTA Lite® SS-A 52 ELISA, 510(k) number: K063565

The QUANTA Flash Ro52 assay is designed to run on the BIO-FLASH® instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash Ro52 assay utilizes a reagent cartridge format, which is compatible with the BIO-FLASH instrument.

Purified recombinant Ro52 antigen is coated onto paramagnetic beads. The bead suspension is lyophilized and stored in the bead tube. Prior to use in the BIO-FLASH system, the sealed reagent tubes are pierced with the reagent cartridge lid and the beads are rehydrated and resuspended using

{6}------------------------------------------------

resuspension buffer by pipetting up and down with a transfer pipette. The reagent cartridge is then loaded onto the BIO-FLASH instrument. Samples are also loaded onto the instrument in sample racks. Serum samples are prediluted by the BIO-FLASH with system rinse in a small disposable plastic cuvette. Small amounts of the diluted patient serum, the beads, and assay buffer are all combined into a second cuvette, and mixed. This cuvette is then incubated at 37°C. The beads are magnetized and washed several times. Isoluminol conjugated anti-human IgG antibodies are then added to the cuvette, and again incubated at 37°C. The beads are magnetized and washed repeatedly. The isoluminol conjugate is oxidized when Trigger 1 (Fe(II)coproporphyrin in sodium hydroxide solution) and Trigger 2 (ureahydrogen peroxide in sodium chloride solution) are added to the cuvette, and the flash of light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. The RLU are proportional to the amount of isoluminol conjugate that is bound to the human IgG, which is in turn proportional to the amount of anti-Ro52 antibodies bound to the corresponding beads.

For quantitation, the QUANTA Flash Ro52 assay utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. Every new lot number of reagent cartridge must be calibrated before first use, with the QUANTA Flash Ro52 Calibrators. Based on the results obtained with the two Calibrators included in the Calibrator Set (sold separately), an instrument specific Working Curve is created, which is used to calculate chemiluminescent units (CU) from the instrument signal (RLU) obtained for each sample.

The QUANTA Flash Ro52 kit contains the following materials:

One (1) QUANTA Flash Ro52 Reagent Cartridge One (1) vial of Resuspension buffer

One (1) Transfer pipette

The QUANTA Flash Ro52 reagent cartridge contains the following reagents for 50 determinations:

  • a. Ro52 antigen coated paramagnetic beads, lyophilized.
  • Assay buffer colored pink, containing Tris-buffered saline, Tween 20, protein b. stabilizers and preservatives.
  • C. Tracer IgG - Isoluminol labeled anti-human IgG antibodies in buffer, containing protein stabilizers and preservative.

The QUANTA Flash Ro52 Calibrators kit contains two vials of Calibrator 1 and two vials of Calibrator 2:

QUANTA Flash Ro52 Calibrators:

  • । QUANTA Flash Ro52 Calibrator 1: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human antibodies to Ro52 in buffer, protein stabilizer and preservative.
  • QUANTA Flash Ro52 Calibrator 2: Two (2) barcode labeled tubes containing 0.3 mL -

{7}------------------------------------------------

prediluted, ready to use reagent. Calibrators contain human antibodies to Ro52 in buffer, protein stabilizer and preservative.

The QUANTA Flash Ro52 Controls kit contains two vials of Negative Control and two vials of Positive Control:

QUANTA Flash Ro52 Controls:

  • । QUANTA Flash Ro52 Negative Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to Ro52 in buffer, protein stabilizer and preservative.
  • , QUANTA Flash Ro52 Positive Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to Ro52 in buffer, protein stabilizer and preservative.

Intended use(s)

QUANTA Flash Ro52 is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-Ro52 autoantibodies in human serum. The presence of anti-Ro52 autoantibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of Systemic Lupus Erythematosus, Sjögren's Syndrome, Systemic Sclerosis, Idiopathic Inflammatory Myopathies.

QUANTA Flash Ro52 Calibrators are intended for use with the QUANTA Flash Ro52 Reagents for the determination of IgG anti-Ro52 autoantibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

QUANTA Flash Ro52 Controls are intended for use with the QUANTA Flash Ro52 Reagents for quality control in the determination of IgG anti-Ro52 autoantibodies in human serum.

Substantial equivalence

The QUANTA Flash Ro52 Reagent, the QUANTA Flash Ro52 Calibrators and the QUANTA Flash Ro52 Controls have the same intended use and assay principle as the predicate device.

Comparison to predicate device

Similarities
ItemQUANTA Flash Ro52Predicate Device
Intended useSemi-quantitative determination of anti-Ro52 antibodies in human serumSemi-quantitative detection of anti-Ro52 antibodies in human serum

QUANTA Flash Ro52 reagent kit

{8}------------------------------------------------

Similarities
ItemQUANTA Flash Ro52Predicate Device
Assay methodologySolid phase (heterogeneous)immunoassaySolid phase (heterogeneous)immunoassay
AntigenPurified recombinant Ro52 antigenPurified recombinant Ro52 antigen
TraceabilityInternational Reference Preparationis not availableResults are traceable to in-houseStandardsInternational Reference Preparation isnot available
Sample typeSerumSerum
Shelf lifeOne yearOne year
Differences
ItemQUANTA Flash Ro52Predicate Device
Detection/Operating principleChemiluminescent immunoassayEnzyme-linked immunosorbent assay
Solid phaseParamagnetic microparticles (beads)96-well plate
ConjugateIsoluminol conjugated anti-human IgGHRP conjugated anti-human IgG
CalibrationLot specific Master Curve + twocalibrators (sold separately)SS-A 52 ELISA Low Positive(Included in the kit)

QUANTA Flash Ro52 Calibrators

ItemQUANTA Flash Ro52 CalibratorsPredicate Device
Intended useQUANTA Flash Ro52 Calibrators areintended for use with the QUANTAFlash Ro52 Reagents for thedetermination of IgG anti-Ro52autoantibodies in human serum. Eachcalibrator establishes a point ofreference for the working curve thatis used to calculate unit values.No separate intended use; calibrator ispart of the kit.
AnalyteAnti-Ro52 antibodiesAnti-Ro52 antibodies
MethodQUANTA Flash Ro52chemiluminescent immunoassayQUANTA Lite SS-A 52 ELISA
MatrixHuman serum, buffer, proteinstabilizer, and preservativeHuman serum, buffer, proteinstabilizer, and preservative
UnitCU (Chemiluminescent units)(arbitrary)units (arbitrary)

{9}------------------------------------------------

ItemQUANTA Flash Ro52 CalibratorsPredicate Device
Physico-chemicalcharacteristicsLiquid, prediluted, ready to useLiquid, prediluted, ready to use
Storage2-8 °C2-8 °C
Shelf lifeOne yearOne year

QUANTA Flash Ro52 Controls

ItemQUANTA Flash Ro52 ControlsPredicate Device
Intended useQUANTA Flash Ro52 Controls areintended for use with the QUANTAFlash Ro52 reagents for qualitycontrol in the determination of IgGanti-Ro52 autoantibodies in humanserum.No separate intended use; controls arepart of the kit.
AnalyteAnti-Ro52 antibodiesAnti-Ro52 antibodies
MethodQUANTA Flash Ro52chemiluminescent immunoassayQUANTA Lite SS-A 52 ELISA
MatrixHuman serum, stabilizers, andpreservativeHuman serum with preservative
UnitCU (Chemiluminescent units)(arbitrary)units (arbitrary)
Physico-chemicalcharacteristicsLiquid, ready to useLiquid, prediluted, ready to use
Levels2 (negative and positive)2 (ELISA negative and High positive)
Storage2-8 °C2-8 °C
Shelf lifeOne yearOne year

Analytical performance characteristics

Value assignment and traceability of Calibrators and Controls

There is currently no recognized international standard for the measurement of Ro52 antibodies. The Centers for Disease Control and Prevention ANA reference serum #2 (IS2073 - IIF ANA [speckled pattern]; anti-SS-B/La), #7 (IS2105 - Anti-SS-A/Ro), and #10 (IS2187 - Anti-Jo1) were tested for Ro52 and produced the following results: CDC ANA #2: 40.7 CU CDC ANA #7: 32.4 CU CDC ANA #10: 443.7 CU

{10}------------------------------------------------

The QUANTA Flash Ro52 Calibrators and Controls are manufactured by diluting human serum that contains high titer of anti-Ro52 antibodies with buffer, containing protein stabilizer and preservative. The human serum is obtained from commercial sources and it is tested for markers of infectious substances.

The target CU is achieved through trial dilutions on small scale. Once a dilution is selected, the Calibrators and Control are bulked, tested, and adjusted. Upon completion of the manufacturing process, the Calibrators and Controls are tested on at least two instruments, on at least two lots of reagent cartridge, in replicates of 10 to determine final value assignment.

Calibrator and Control values are directly traceable to in-house Standards that are used to create the Master Curves for the QUANTA Flash Ro52 assay.

MaterialAssigned Value
Ro52 Master Curve Standard 12.3 CU
Ro52 Master Curve Standard 210.1 CU
Ro52 Master Curve Standard 352.4 CU
Ro52 Master Curve Standard 498.0 CU
Ro52 Master Curve Standard 5386.4 CU
Ro52 Master Curve Standard 61685.3 CU

List of Ro52 Standards, Calibrators and Controls:

MaterialManufacturingTarget ValueManufacturingTarget Range
Ro52 Calibrator 110 CU8 – 12 CU
Ro52 Calibrator 2400 CU360 – 440 CU
Ro52 Negative Control10 CU8 – 12 CU
Ro52 Positive Control50 CU40 – 60 CU

Precision

The precision of the QUANTA Flash Ro52 assay was evaluated on 9 samples containing various concentrations of Ro52 antibodies in accordance with CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Procedures - Approved Guideline: samples were run in duplicates, twice a day, for 21 days. Production reagent lot 131006 was used for the studies.

Data were analyzed with the Analyse-it for Excel method evaluation software, and within run, between run, between day and total precision were calculated.

Acceptance criteria: Total %CV: < 10%

Results are summarized in the Table below.

{11}------------------------------------------------

QUANTA Flash Ro52Within RunBetween RunsBetween DaysTotal
Sample IDNumber ofMeanSDCVSDCVSDCVSDCV
replicates(CU)(CU)(%)(CU)(%)(CU)(%)(CU)(%)
000674-108414.00.53.4%0.43.2%0.43.0%0.85.5%
120814-5508420.50.73.4%0.63.1%1.25.8%1.57.4%
110688-558427.21.03.6%0.72.6%1.55.5%1.97.1%
110685-408418.80.63.4%0.42.1%0.94.6%1.16.1%
110689-5084140.58.35.9%2.01.5%8.46.0%12.08.5%
110684-1884343.38.62.5%6.21.8%17.04.9%20.05.8%
120815-1484638.229.14.6%0.00.0%38.76.1%48.57.6%
110686-6841081.348.74.5%30.62.8%50.74.7%76.77.1%
110689-15841537.568.24.4%65.34.2%79.45.2%123.48.0%

Reproducibility

Three samples were tested on two different reagent lots, using two different lots of Calibrators, by two operators. Samples were run in quadruplicates, two times a day, for 10 days, to generate 80 data points per sample. Production reagent lots 131006 and 141007, and Calibrator and Control lots 131005 and 131006 were used for the studies. Data were analyzed with the Analyse-it for Excel method evaluation software, and within run, between reagent lots, between calibrator lots, between operators and total precision were calculated.

Acceptance criteria: Total %CV: < 10%

Results are summarized in the Table below.

BetweenBetweenBetween
QUANTA Flash Ro52Within RunReagent LotsCalibrator LotsOperatorsTotal
SampleNumber ofreplicatesMean(CU)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)
1809.80.33.30.44.50.44.20.22.50.44.3
28022.40.62.81.14.71.35.80.73.01.14.9
38096.92.22.38.58.76.16.33.83.96.56.7
C80847.628.63.451.76.121.42.523.92.838.84.6
A801140.844.83.956.65.056.14.957.85.162.45.5
B801472.261.94.254.23.757.43.957.83.966.94.5

Limit of Blank (LoB) and Limit of Detection (LoD)

The LoD of the QUANTA Flash Ro52 assay is 402 RLU, which is below the analytical measuring range of the assay. It was determined consistent with CLSI EP17-A2 guideline with proportions of false positives

{12}------------------------------------------------

(alpha) less than 5% and false negatives (beta) less than 5%; based on 120 determinations, with 60 measurements on blank samples and 60 measurements of low level samples, per reagent lot.

For determining the LoB, 4 blank samples (System Rinse) from two different lots were run in replicates of five on two reagent lots, once per day, for 3 days. Production reagent lots 131006 and 141007 were used for the studies. Sixty data points were generated on each lot.

The LoB was determined on each lot separately with the Analyse-it for Excel software's Reference Interval function, at the 950 percentile, using the non-parametric method, as the dataset showed nonnormal distribution (p values at 0.0068 and < 0.0001, for each lot, respectively). The LoB for lot 131006 was determined as 308 RLU, and for lot 141007 as 319 RLU. The final LoB value is 319 RLU.

For determining the LoD, 4 low level samples (prepared by diluting anti-Ro52 positive samples with System Rinse) were run in replicates of five on two reagent lots, once per day, for 3 days. Production reagent lots 131006 and 141007 were used for the studies. Sixty data points were generated on each lot.

The LoD was determined separately on each lot according to CLSI EP17-A2 guideline. The limit of detection for lot 131006 was determined as 400 RLU, and for lot 141007 as 402 RLU. The final LoD value is 402 RLU.

These values are below the value of the lowest QUANTA Flash Ro52 Master Curve standard, i.e. below the Analytical Measuring Range.

Analytical Measuring Range (AMR)

QUANTA Flash Ro52:2.3 CU – 1685.3 CU
The AMR is defined by the values of the lowest and highest Master Curve Standards.

Auto-rerun function and reportable results

The BIO-FLASH software has an Auto-rerun option available. If this option is selected, the instrument will automatically rerun any sample that has a result of >1685.3 by further diluting it by 35 fold, thereby bringing the measured value within the AMR. The final result will be calculated by the software by taking into account the additional dilution factor. As the highest value that can be measured is 1685.3 CU, the highest value that can be reported is 58985 CU.

High concentration hook effect

To assess hook effect, measurement signal (relative light units, RLU) was examined for five high positive samples (results above the AMR) before and after automatic or manual dilution. All sera produced significantly higher RLU values (above the AMR) when used "as is" compared to the manually or automatically diluted ones (that were within the AMR), thereby confirming that high positive specimens above the analytical measuring range do not show hook effect up to 23005 CU in the Ro52 assay (the highest concentration that was tested).

{13}------------------------------------------------

Linearity

The linearity of the AMR was evaluated by a study according to CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approved Guideline. Production reagent lot 131006 was used for the study. Four serum samples with various Ro52 antibody concentrations were diluted in 10% increments (from 0% to 90% diluent) to obtain values that cover the AMR. The dilutions were assayed in duplicates. Percent recovery of obtained results was calculated compared to the expected results (based on the dilution factor). Moreover, obtained values were plotted against expected values, and linear regression analysis was performed. Only results within the AMR were included in the regression analysis.

Acceptance criteria:

  • Recovery is between 80-120%, or ± 4 CU, whichever is greater.
  • For linear regression analysis, slope is between 0.9-1.1, and R-is ≥ 0.95.

All four specimens showed dilution linearity individually.

SampleTest Range (CU)Slope (95% CI)R2
1239.2 to 1618.20.97 (0.93 to 1.01)1.00
237.2 to 370.41.01 (0.99 to 1.04)1.00
311.0 to 104.21.02 (0.99 to 1.04)1.00
43.8 to 17.00.92 (0.86 to 0.98)0.99

The combined data yielded the following results with linear regression:

SampleTest Range (CU)Slope (95% CI)R2
All3.8 to 1618.20.97 (0.96 to 0.97)1.00

Interference

The interference study was performed according to CLSI EPO7-A2, Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition. Three specimens were tested (negative: 16.5 CU; at the cutoff: 21.9 CU; positive: 64.9 CU). Interfering substances were spiked into every specimen at three different concentrations in 10% of total specimen volume, and the resulting samples were assessed in triplicates with the Ro52 assay. Recovery of the unit values was calculated compared to control samples spiked with the same volume of diluents (10% of total). Acceptance criteria for the interference studies were 85% - 115% recovery, or ± 4 CU difference, whichever is greater.

No interference was detected with bilirubin up to 10 mg/dL (recovery: 87% to 107%), hemoglobin up to 200 mg/dL (recovery: 89% to 107%), triglycerides up to 1000 mg/dL (recovery: 91% to 112%),

{14}------------------------------------------------

cholesterol up to 224.3 mg/dL (recovery: 91% to 112%), and RF IgM up to 500 IU/mL (recovery: 92% to 108%).

Cross-reactivity

To test potential cross-reactivity with autoantibodies and infection-induced antibodies, results obtained on 199 of the total 233 control samples that were included in the clinical validation study were assessed. These samples were from patients with autoimmune diseases that are characterized with disease specific autoantibodies, or from patients with positive infectious disease serology. The composition of the cohort and the anti-Ro52 positivity rate is shown in the Table below:

DiagnosisNumber of samples# pos% pos
Graves' Disease1000.0%
Hashimoto Thyroiditis1000.0%
Celiac Disease1100.0%
Crohn's Disease2000.0%
Ulcerative Colitis2000.0%
HCV6116.7%
HBV600.0%
CMV1119.1%
EBV (with or without other infection)12216.7%
HIV500.0%
Syphilis500.0%
Primary Antiphospholipid Syndrome1500.0%
Vasculitis1700.0%
Rheumatoid arthritis5024.0%
Behçet's disease100.0%
Total controls19963.0%

Based on the results, the QUANTA Flash Ro52 assay does not show cross-reactivity with autoantibodies that are present in various autoimmune diseases, or antibodies against infectious agents.

Lot to lot comparison

Twenty unique samples and the Positive and Negative Controls (altogether 22 specimens) with various reactivity levels were tested in triplicates with three different reagent lots: 121005, 131006 and 141007. The samples covered the total analytical measuring range of the assay. Results were processed by linear

{15}------------------------------------------------

510(k) Summary QUANTA Flash® Ro52

regression analysis and bias calculation according to CLSI EP09-A2, Method Comparison and Bias Calculation Using Patient Samples; Approved Guideline - Second Edition.

Pair-wise comparisons were performed between lot 121005 vs 131006, lot 121005 vs 141007 and lot 131006 vs 141007, considering individual replicates instead of the mean of replicates.

Acceptance criteria and results are in the Table below. All results were within the acceptance limits.

121005 vs131006121005 vs141007131006 vs141007
Acceptance criteria
Weighted r: ≥0.975 for linear regression0.9990.9970.997
Intercept of the regression line (constant bias):± 15% of cut-off (3 CU)1.30.7-0.6
Slope of the regression line (proportional bias): 0.9-1.10.970.940.96
Weighted S y/x: ≤ 0.50.0430.060.06
Predicted bias (difference) at cut-off: ±15% (3 CU)0.8-0.6-1.3

Stability

Shelf life

To establish the initial claim for shelf life, accelerated stability studies were performed for 4 weeks at 37°C ± 3°C, where one week is equal to six months at 5 ± 3°C.

Accelerated stability testing was performed on each of the following sealed components of the QUANTA Flash Ro52 to establish initial stability claim: the beads, the two Calibrators, and the negative and positive Controls. Each week a new sealed component was placed in the incubator, and all components were tested at the end of the experiment together with the one that was stored at 5 ± 3℃. The recovery of the measured values was calculated for each time point (compared to those obtained with 5 ± 3°C stored reagent). All calculations were performed by comparing results of sealed components stored at 5 ± 3°C (control) to those stored at 37 ± 3°C (test) for 1, 2, 3, and 4 weeks, where one week is equal to six months at 5 ± 3℃. Linear regression analysis was performed between recovery values and the number of days.

Acceptance criteria for one year preliminary expiration dating:

  • Beads:

With regression analysis, the lower 95% Cl interval of the regression line is ≥ 85% at 2 weeks, and no individual data point has ≤ 75% recovery at 2 weeks.

  • Controls and Calibrators:

With regression analysis, the lower 95% Cl interval of the regression line is ≥ 90% at 2 weeks, and no individual data point has ≤ 80% recovery at 2 weeks.

Beads

{16}------------------------------------------------

Testing was performed on three lots of Ro52 coupled beads using up to 6 characterized samples with various reactivity levels.

All three lots of beads retained > 85% reactivity (considering the 95% Cl) after two weeks at 37 ± 3°C, and therefore pass the acceptance criteria for one year expiration date.

Calibrators and Controls

Testing was performed on three lots of Ro52 Calibrators and Controls. All Calibrators and Controls maintained > 90% reactivity (considering the 95% CI) when sored at 37 ± 3°C for 2 weeks, and therefore pass the acceptance criteria for one year expiration dating.

In-use (onboard) stability

Calibrators

Onboard stability claim: 4 calibrations, or 8 hours onboard

During assessing on-board stability, Calibrators were placed uncapped, onboard the instrument, and calibration was performed altogether five times over 8.5 hours. Controls and a panel of characterized patient specimens were run on each calibration curve.

Calibrators are considered stable if all five calibrations performed in the 8.5 hour period are successful, and average Calibrator RLU recovery values are between 90% and 110% compared to the first use.

A total of 5 successful calibrations were performed over an 8.5 hour period. Calibrator RLU values remained within the 90-110% range. Moreover, all Controls and patient panel samples ran within their expected range. This supports the claim that calibrators can be used for up to 4 calibrations over an 8 hour period.

Controls

Onboard stability claim: up to 15 uses, at 10 minutes onboard per use

During assessing on-board stability, 2 vials of each Control were assayed twice a day for a total of 20 runs. The first run (each vial run in duplicate) was used to establish baseline value, and then additional 19 runs (each vial run in singleton) were performed. During runs, the Controls were left uncapped, onboard the instrument for 15 minutes per run. When not in use, the controls were capped, and stored at 5 ± 3°C.

Percent recovery of each value was calculated against the mean of duplicates of each vial from the first run. Controls are considered stable when all values run within their established range, and the linear regression line obtained by plotting %recovery values against the number of runs stays between 85% and 115% at run 15.

All controls ran within their respective acceptable ranges for all runs. Moreover, the regression line remained between 85% and 115% at run 15 for both Controls. These results support the claim that controls can be used for up to 15 times, at 10 minutes per use.

Reagent Cartridge

{17}------------------------------------------------

To establish the in-use stability of the QUANTA Flash Ro52 reagent cartridge, three lots of cartridges were tested with up to 4 serum specimens (with different reactivity levels) along with the Negative and Positive Controls. The specimens were tested periodically up to 49 days. Percent recoveries were calculated compared to the day zero average values, and linear regression analysis was performed by plotting %recovery against the number of days. The claim was established using the following criteria (using the one that is fulfilled first):

  • The stability claim is established at the actual measurement day preceding the day when the 95% confidence interval of the regression line reaches 85% or 115% recovery, or

  • At the actual measurement day preceding the day when 2 data points or ≥2% of the recovery data (whichever is greater) is ≤ 75% or ≥ 125% recovery.

The onboard stability results of the three lots are as follows:

RP0002: 36 days

121005: 36 days

141007: 42 days

Using these criteria, the in-use (onboard) stability of Ro52 reagent cartridge was set at 36 days.

Real time stability

Real time stability testing was performed at 3, 6, 9 and 12 and 18 months on the Calibrators and Controls, to support the one year expiration, and at 3, 6, 9 and 12 months on one lot of reagent cartridge, and at 0 and 19 months on another lot of reagent cartridge.

Each control was tested in triplicates at each time point.

  • Acceptance criteria: results should fall within their acceptable ranges as it was established at the release of the controls.

Calibrators were tested in triplicates at each time point as it is done during calibration. Averages of the triplicates were compared to the value that was assigned to the Calibrators at release.

  • Acceptance criteria: % recovery of the average of the triplicates is between 85% and 115%, and %CV of the triplicates is < 10%

For reagent cartridge, QC panel samples (with target results) were tested at each time point. This QC panel is used by the QC Department for reagent release and QC.

  • Acceptance criteria: results should fall within their respective QC ranges.

All results were within the acceptance limits.

Cut-off, reference range

QUANTA Flash Ro52: Negative <20 CU

{18}------------------------------------------------

Positive ≥20 CU

The reference population for establishing the reference interval for the Ro52 assay consisted of 155 subjects:

Sample GroupN
Apparently healthy blood donors115
Viral hepatitis positive samples8
HIV positive samples5
Syphilis positive samples5
Rheumatoid arthritis patients22

All specimens were the same matrix (serum) as specified in the Intended Use. All specimens were unaltered. The cut-off was established in accordance to CLSI C28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition. The Analyseit for Excel software was used to make the calculations. The results was non-normal (Saphiro-Wilk p<0.0001), so the non-parametric percentile method was used. The 97th percentile of the obtained values calculated as 9655.6 RLU.

The cutoff was set to 10000 RLU. Three apparently healthy blood donors and one rheumatoid arthritis sample tested positive at this cutoff level.

Clinical performance characteristics

Clinical sensitivity, specificity

A cohort of characterized samples, none of which were used for establishing the reference range, was used to validate the clinical performance of the QUANTA Flash Ro52. A total of 600 characterized samples were included in the Validation Set for the QUANTA Flash Ro52. All samples were run on the QUANTA Flash Ro52. The distribution of the Ro52 positivity rate is in the Table below:

Patient groupNNumber positive% positive
Graves' Disease1000.0%
Hashimoto Thyroiditis1000.0%
Celiac Disease1100.0%
Crohn's Disease2000.0%
Ulcerative Colitis2000.0%
HCV6116.7%
HBV600.0%
CMV1119.1%
EBV (with or without other infection)12216.7%
HIV500.0%

{19}------------------------------------------------

Patient groupNNumber positive% positive
Syphilis500.0%
Primary Antiphospholipid Syndrome1500.0%
Secondary Antiphospholipid Syndrome*14642.9%
Vasculitis1700.0%
Rheumatoid arthritis5024.0%
Osteoarthritis2015.0%
Behçet's disease100.0%
Total controls233135.6%
Sjögren's Syndrome (SS)914044.0%
Systemic Lupus Erythematosus (SLE)1314735.9%
Systemic Sclerosis (SSc)801316.3%
Idiopathic Inflammatory Myopathies (IIM)652640.0%
Total600
* Patients may have SLE

The results were analyzed to calculate sensitivity and specificity for SS (n=91), SLE (n=80), and IIM (n=65) separately, using 219 disease controls. The secondary APS group was excluded from all calculations.

Clinical sensitivity and specificity of the QUANTA Flash Ro52 in SS

n=310DiagnosisAnalysis(95% confidence)
SSControlsTotal
QUANTA FlashRo52Positive40747Sensitivity = 44.0% (33.6-54.8%)
Negative51212310Specificity = 96.8% (93.5-98.7%)
Total91219310

Clinical sensitivity and specificity of the QUANTA Flash Ro52 in SLE

n=350DiagnosisAnalysis(95% confidence)
SLEControlsTotal
QUANTA FlashRo52Positive47754Sensitivity = 35.9% (27.7-44.7%)
Negative84212296Specificity = 96.8% (93.5-98.7%)
Total131219350

Clinical sensitivity and specificity of the QUANTA Flash Ro52 in SSc

n=299DiagnosisAnalysis
QUANTA FlashPositiveSScControlsTotal(95% confidence)
13720Sensitivity = 16.3% (8.9-26.2%)

{20}------------------------------------------------

Ro52Negative67212279Specificity = 96.8% (93.5-98.7%)
Total80219299

Clinical sensitivity and specificity of the QUANTA Flash Ro52 in IIM

n=283DiagnosisAnalysis
IIMControlsTotal(95% confidence)
QUANTA FlashRo52Positive26733Sensitivity = 40.0% (28.0-52.9%)
Negative39211250Specificity = 96.8% (93.5-98.7%)
Total65218283

Expected values

The expected value in the normal population is "negative". Anti-Ro52 autoantibody levels were analyzed in a cohort of 111 apparently healthy blood donors (90 females and 21 males, ages 17 to 60 years, with an average age of 32.6 years and median age of 31 years) using the QUANTA Flash Ro52. This patient population was different from the one that was used to establish the cutoff, and was only used to assess expected values. With the cut-off of 20 CU, none of the samples were positive on the QUANTA Flash Ro52. The mean concentration was 8 CU, and the values ranged from <2.3 to 19.3 CU.

Comparison with predicate device

Samples for method comparison analysis included 319 samples from the clinical validation study, along with 27 additional samples characterized as having a speckled pattern on HEp-2 by ANA IIF. From the total sample size of 346, results for 283 samples were within the reportable range of the assay. These samples were tested on both the QUANTA Flash Ro52 and on the predicate ELISA. The data are presented in the following tables:

Method Comparison, all samples:

Method Comparison (N=346)Ro52 ELISAPercent Agreement(95% Confidence)
NegativePositiveTotal
QUANTA Flash ®Ro52 CIANegative20925234Pos. Agree = 80.3% (72.3 – 86.8%)
Positive10102112Neg. Agree = 95.4% (91.8 – 97.8%)
Total219127346Total Agree = 89.9% (86.2 – 92.9%)

Method Comparison, samples within reportable range:

Method Comparison (N=283)Ro52 ELISAPercent Agreement
NegativePositiveTotal(95% Confidence)

{21}------------------------------------------------

510(k) Summary QUANTA Flash® Ro52

Page 19 of 19

QUANTA Flash®Ro52 CIANegative15122173Pos. Agree = 82.0% (74.0 – 88.3%)
Positive10100110Neg. Agree = 93.8% (88.9 – 97.0%)
Total161122283Total Agree = 88.7% (84.4 - 92.1%)

§ 866.5100 Antinuclear antibody immunological test system.

(a)
Identification. An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).(b)
Classification. Class II (performance standards).