K Number
K141488

Validate with FDA (Live)

Device Name
CAREVIEW 1800R
Date Cleared
2015-09-08

(460 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CareView 1800R X-ray Flat Panel Detectors is indicated for digital imaging solution designed for providing general radiographic system in all general purpose diagnostic procedures. CareView 1800R X-ray Flat Panel Detectors is a component of a digital imaging system. Properly integrated into a completed X-ray system, the digital X-ray imaging intended for medical application.

This product is not intended for mammography applications and dental X-ray applications.

Device Description

CareView 1800R X-ray Flat Panel Detectors supports the single frame mode, with the key component of TFT/PD image sensor flat panel of active area: 43cm×43cm. The sensor plate of CareView 1800R X-ray Flat Panel Detectors is direct-deposited with Csl scintillator to achieve the conversion from X-ray to visible photon. The visible photons are transformed to electron signals by diode capacitor array within TFT panel, which are composed and processed by connecting to scanning and readout electronics, consequently to form a panel image by transmitting to PC through the user interface.

The major function of the CareView 1800R X-ray Flat Panel Detectors detector is to convert the X-ray to digital image, with the application of high resolution X-ray imaging. This detector is the key component of DR system, enables to complete the digitalization of the medical X-ray imaging with the DR system software.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the CareView 1800R X-ray Flat Panel Detector, based on the provided text:

Acceptance Criteria and Device Performance

The acceptance criteria are not explicitly stated, but rather performance characteristics are compared to a predicate device (Canon CXDI-40EG). The reported device performance for CareView 1800R is presented alongside the predicate device's performance in the table below. The underlying assumption for acceptance is that the CareView 1800R performs comparably or better than the predicate.

CharacteristicCareView 1800R X-ray Flat Panel Detectors (Proposed Device)Canon CXDI-40EG (Predicate Device)
Modulation Transfer Function (MTF)
1 lp/mm(70±3)%68%
2 lp/mm(42±3)%38%
3 lp/mm(23±3)%18%
Detective Quantum Efficiency (DQE) at 5 µGy
0 lp/mm(63±3)%37% (at 5 mR)
1 lp/mm(45±3)%34% (at 5 mR)
2 lp/mm(33±3)%24% (at 5 mR)
3 lp/mm(20±3)%12% (at 5 mR)
ADC Digitization16 bit14 bit
Detector Size492 x 492 x 33.5 mm550 x 550 x 67.5 mm
Detector Weight13 kg20 kg
Pixel Array2816x28162688 x2688
Pixel Pitch154µm160µm
X-ray AbsorberCsI ScintillatorGOS
Installation TypeStationary, permanently installedStationary, permanently installed
Readout MechanismThin Film TransistorThin Film Transistor
Power Consumption~36 WMax.200 VA
Power SupplyVoltage: 100-250 VAC, Frequency: 50/60 HzVoltage: 100V, 120V, 230/240 V, Frequency: 50/60 Hz
Operating ConditionTemperature: 5 ~ 35°C, Relative humidity: 3075% RH, Air pressure: 700hPa1060hPaTemperature: 535°C, Relative humidity: 3075%, Air pressure: 700hPa ~1060hPa
Storage ConditionTemperature: -2055 °C, Relative humidity: 1090%RH, Air pressure: 700hPa~1060hPaTemperature: -3060 °C, Relative humidity: 1060%, Air pressure: 700hPa ~1060hPa

Study Information

2. Sample size used for the test set and the data provenance:

  • Sample Size: 30 clinical images.
  • Data Provenance: Not specified (e.g., country of origin). The study is described as a "concurrence study," which implies a retrospective comparison of images.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This information is not provided in the document. The document states "There was no significant difference between the images of the CareView 1800R and those of the predicate device," but does not detail how this "no significant difference" was determined or by whom.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • This information is not provided in the document.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • A multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly performed or described in this document. The clinical study was a "concurrence study of 30 clinical images... to compare the performance... to that of the predicate device." This suggests a direct comparison of image sets rather than assessing human reader performance with or without AI assistance.
  • The device is an X-ray flat panel detector, not an AI-powered diagnostic system, hence "human readers improve with AI vs without AI assistance" is not relevant for this type of device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • While the device itself is a standalone component (the detector), the "concurrence study" involved clinical images, which are implicitly interpreted by humans. However, the performance metrics (MTF, DQE) are standalone algorithm/device performance. Therefore, yes, standalone performance was assessed through non-clinical laboratory testing and then clinically supported by the concurrence study.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • For the non-clinical laboratory tests (MTF, DQE, etc.), the "ground truth" would be objective physical measurements based on established standards.
  • For the clinical concurrence study, the "ground truth" implicitly relies on the assessment that there was no significant difference between images from the proposed device and the predicate device. The method or criteria for this assessment (e.g., comparison against established radiographic quality, diagnostic findings) is not explicitly detailed.

8. The sample size for the training set:

  • The CareView 1800R is an X-ray flat panel detector, which is hardware, not a machine learning algorithm. Therefore, there is no training set in the context of AI/ML.

9. How the ground truth for the training set was established:

  • As there is no training set (see point 8), this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 8, 2015

CareRay Digital Medical System Co., Ltd. % Jamie Ku Official Correspondent Compass Innovations, Inc. 3001 Winchester Blvd, Suite 3 CAMPBELL CA 95008

Re: K141488

Trade/Device Name: CareView 1800R X-ray Flat Panel Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: August 13, 2015 Received: August 17, 2015

Dear Jamie Ku:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

For

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141488

Device Name CareView 1800R X-ray Flat Panel Detectors

Indications for Use (Describe)

CareView 1800R X-ray Flat Panel Detectors is indicated for digital imaging solution designed for providing general radiographic system in all general purpose diagnostic procedures. CareView 1800R X-ray Flat Panel Detectors is a component of a digital imaging system. Properly integrated into a completed X-ray system, the digital X-ray imaging intended for medical application.

This product is not intended for mammography applications and dental X-ray applications.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

[As required by 21 CFR 807.92]

1. Date Prepared [21 CFR807.92 (a) (1)]

August 13, 2015

2. Submitter's Information [21 CFR807.92 (a) (1)]

Company Name:CareRay Digital Medical System Co., Ltd.
Company Address:Suite 201, Building A2, BIOBAY, 218 Xing Hu Street,SuZhou Industrial Park, SuZhou 215123, P.R.China
Contact Person:Ms. Ding
Phone Number:(86) 512-86860288
Fax Number:(86) 512-86860388
E-mail:yl.ding@careray.com

3. Trade Name, Common Name, Classification [21 CFR807.92(a)(2)]

Trade Name:X-ray Flat Panel Detectors
Model Name:CareView 1800R
Classification Name:Stationary X-ray system
Regulation Number:21 CFR 892.1680
Regulatory Class:Class II
Product Code:MQB

4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)[

The identified predicates within this submission are as follows:

Manufacturer:Canon Inc.
Trade Name:Canon
Model Name:CXDI-40EG
Classification Name:Stationary X-ray system
Regulation Number:21 CFR 892.1680
Regulatory Class:Class II

{4}------------------------------------------------

Product Code:MQB
FDA 510(k) #:K050987

5. Description of the Device [21 CFR 807.92(a)(4)]

CareView 1800R X-ray Flat Panel Detectors supports the single frame mode, with the key component of TFT/PD image sensor flat panel of active area: 43cm×43cm. The sensor plate of CareView 1800R X-ray Flat Panel Detectors is direct-deposited with Csl scintillator to achieve the conversion from X-ray to visible photon. The visible photons are transformed to electron signals by diode capacitor array within TFT panel, which are composed and processed by connecting to scanning and readout electronics, consequently to form a panel image by transmitting to PC through the user interface.

The major function of the CareView 1800R X-ray Flat Panel Detectors detector is to convert the X-ray to digital image, with the application of high resolution X-ray imaging. This detector is the key component of DR system, enables to complete the digitalization of the medical X-ray imaging with the DR system software.

6. Intended Use [21 CFR 807.92(a)(5)]

CareView 1800R X-ray Flat Panel Detectors is indicated for digital imaging solution designed for providing general radiographic system in all general purpose diagnostic procedures. CareView 1800R X-ray Flat Panel Detectors is a component of a digital imaging system. Properly integrated into a completed X-ray system, the detector enables the digital X-ray imaging intended for medical application.

This product is not intended for mammography applications and dental X-ray applications.

ItemProposed Device:CareView 1800R X-ray Flat PanelDetectorsPredicate Device: CXDI-40EG
510(K)NumberTo be assignedK050987
Intended UseThe CareView 1800R X-ray FlatPanel Detectors is indicated fordigital imaging solution designedfor providing general radiographicsystem in all general purposediagnostic procedures. CareView1800R X-ray Flat Panel Detectorsis a component of a digital imagingsystem. Properly integrated into aDIGITAL RADIOGRAPHYCXDI-40EG provides digitalimage capture for conventionalfilm/screen radiographicexaminations. The device isintended to replace radiographicfilm/screen systems in all generalpurpose diagnostic procedures.
completed X-ray system, thedetector enables the digital X-rayimaging intended for medicalapplication.This product is not intended formammography applications anddental X-ray applications.
ClassificationNameStationary X-ray systemStationary X-ray system
ProductCodeMQBMQB
RegulationNumber21 CFR 892.168021 CFR 892.1680
PanelRadiologyRadiology
ClassIIII
X-rayAbsorberCsI ScintillatorGOS
InstallationTypeStationary, permanently installedStationary, permanently installed
ReadoutMechanismThin Film TransistorThin Film Transistor
DetectorSize492 x 492 x 33.5 mm550 x 550 x 67.5 mm
DetectorWeight13 kg20 kg
Pixel Array2816x28162688 x2688
Active Area43cm×43cm43cm×43cm
Pixel Pitch154µm160µm
ADCDigitization16 bit14 bit
PowerConsumption~36 WMax.200 VA
ModulationTransferFunction(MTF)1 lp/mm (70±3)%2 lp/mm (42±3)%3 lp/mm (23±3)%1 lp/mm 68%2 lp/mm 38%3 lp/mm 18%
DetectiveQuantumEfficiency(DQE)DQE at 5 µGy0 lp/mm (63±3)%1 lp/mm (45±3)%2 lp/mm (33±3)%3 lp/mm (20±3)%DQE at 5 mR0 lp/mm 37%I lp/mm 34%2 lp/mm 24%3 lp/mm 12%
PowerSupplyVoltage: 100-250 VACFrequency: 50/60 HzVoltage: 100V, 120V, 230/240 VFrequency: 50/60 Hz
OperatingConditionTemperature: 5 ~ 35°CRelative humidity: 3075% RHAir pressure: 700hPa1060hPaTemperature: 535°CRelative humidity: 3075%Air pressure: 700hPa ~1060hPa
StorageConditionTemperature: -2055 °CRelative humidity: 1090%RHAir pressure: 700hPa~1060hPaTemperature: -3060 °CRelative humidity: 1060%Air pressure: 700hPa ~1060hPa

7. Technological Characteristics [21 CFR 807.92(a)(6)]

{5}------------------------------------------------

{6}------------------------------------------------

8. System requirements to operate with other radiographic system components

    1. Recommended Generator Specification:
      Energy range: 40~150kVp

mA range: 10~1000mA (depending on the generator power)

ms range: 106300ms to produce 0.11000mAs (depending on the generator power) Note: To our best knowledge, the detector is compatible with the X-ray generators with the specifications described above. If you have questions regarding the compatibility issue for other generators, please contact your distributor or CareRay.

    1. Application Program Interface (API) for system integration manufacturer
      Peripheral hardware: CareView detector connected via wired communication.

CPU: Intel (R) Core (TM) 2 Duo, 2.93GHz or above

RAM: 2 GB or higher

Hard disk: 160 GB or higher

Monitor: 1280 x 1024 or higher

OS: Windows XP or Windows 7

Development environment: MS Visual Studio 2005

    1. X-ray exposure mode
      The synchronous connection mode is the signal transfer mode between the X-ray generator which sends the X-ray and the detector which receives the X-ray.

CareView 1800R supported typical sync mode contains external sync, manual sync mode.

The detector can't provide feedback to the generator to terminate the x-ray exposure.

9. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92(b)(2)]

  • Electrical safety and EMC testing
    Electrical, mechanical, environmental safety and performance testing according to IEC/ES 60601-1 was performed, and EMC testing was also conducted in accordance with IEC/EN 60601-1-2. All test results are satisfactory. Section 201.7.2.7, 201.7.9.1, 201.7.9.2.1.103 and 203.4.1 of IEC 60601-2-54 were also conducted.

  • A Nonclinical considerations
    The following non-clinical studies have been performed and the results have shown that the CareView 1800R X-ray flat panel detector is substantially equivalent to the predicate devices on the Market (CXDI-40EG).

{7}------------------------------------------------

-Detective quantum efficiency (DQE), Quantum limited performance, Modulation transfer function (MTF), Effects of aliasing, Sensitivity linearity, Lag(Erasure thoroughness), Change in detection sensitivity, Dose requirement and reciprocity changes, Stability of device characteristics with time, Uniformity of device characteristic, Noise power spectrum(NPS), Spatial resolution, Image Acquisition time, & Black level

  • A Clinical considerations
    A concurrence study of 30 clinical images was conducted to compare the performance of the CareView 1800R to that of the predicate device (K050987).

Clinical images were provided; these images were not necessary to establish substantial equivalence based on the superior performance in comparison to the predicate based on laboratory testing and based on the use of a detector scintillator material that is already on the marketplace; however, they provide further evidence in addition to the laboratory performance data to show that the complete system works as intended.

There was no significant difference between the images of the CareView 1800R and those of the predicate device.

10. Conclusion [21 CFR 807.92(b) (3)]

In accordance with the Federal Food, Druq and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, the CareView 1800R X-ray flat panel detector is substantially equivalent to predicate device CXDI-40EG (K050987). Both propose and predicate devices are same or very similar in the intended use, the design principle, the applicable standards and specification. Some characteristics, for example, the detector size, pixel pitch, weight and ADC digitization, X-ray absorber, storage condition are different. However the test reports in this submission documents provide demonstration that these differences do not raise any new questions of safety and effectiveness. Therefore, CareRay Digital Medical System Co., Ltd. concludes the CareView 1800R X-ray flat panel detector is substantially equivalent with the predicate device CXDI-40EG (K050987) of Canon Inc.

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.