K Number
K141484

Validate with FDA (Live)

Device Name
OPEN FMSEALER
Date Cleared
2014-07-30

(55 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Open FMsealer is intended for use in general and gynecological surgery and other open surgical procedures where ligation of vessels, including lymph vessels, is desired. The system creates a vessel ligation (seal) by the application of heat (coagulation) to vessels interposed between the jaws of the device. The Open FMsealer can be used to seal and ligate vessels up to 7mm and tissue bundles as large as will fit in the jaws of the instrument.

The Open FMsealer has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these purposes.

Device Description

An electrosurgical cutting and coagulation device (and accessories) is a device intended to remove tissue and control bleeding by use of high-frequency electrical current (see 21 CFR §878.4400). It is classified as a Class II (510(k)) device.

The Open FMsealer is a device intended for sealing of vessels and for soft tissue cutting and coagulation. The device is designed to be used only with the FMwand Generator (K130606). The two-piece device consists of a sterile, single-use patient-contacting shears section (Open FMsealer) and an accessory cable (Power Module or PM) that connects the FMsealer to the generator. The dual footpedal connects to the rear of the generator and can be used to activate the FMsealer if desired by the user.

The Open FMsealer is a hand-held surgical instrument intended for use in open surgical procedures where ligation of vessels or cutting and coagulation of soft tissue is desired. The Open FMsealer consists of a scissor-like handle with two actuation buttons and jaws to grasp the tissue to be affected.

The Open FMsealer is connected to the FMwand Generator (K130606) by the Power Module. The connectors on both ends of the Power Module are designed to fit only the FMwand Generator and FMwand sterile, disposable accessories. The power module and FMwand Generator were previously cleared as part of K130606.

This system creates sealing, cutting and coagulation by the application of heat and compression to tissue bundles and vessels interposed between the iaws of the instrument. Like the predicate device, the Open FMsealer includes actuation buttons that can be used to activate the device.

The Open FMsealer has been shown (in animal studies) to cut and seal vessels up to and including 7mm in diameter, lymphatics and tissue bundles as large as will fit into the jaws of the device.

AI/ML Overview

This 510(k) summary describes the "Open FMsealer" intended for vessel sealing and soft tissue cutting and coagulation in surgical procedures. The device uses ferromagnetic induction to generate heat, similar to its predicate device, the FMwand Ferromagnetic Surgical System (K130606). It also claims substantial equivalence to the LigaSure Vessel Sealing System (K043273).

Here's an analysis of the provided information according to your requested criteria:


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in the format of pass/fail thresholds for specific quantitative metrics. Instead, it describes performance characteristics and states that the device "passed all bench tests" and "is at least as effective as the predicate device." The performance testing focused on demonstrating substantial equivalence to the predicate devices.

Performance Feature / Acceptance Criteria (Implied)Reported Device Performance
Vessel Sealing CapabilityCuts and seals vessels up to and including 7mm in diameter.
Tissue Sealing/Cutting CapabilitySeals and ligates tissue bundles as large as will fit into the jaws of the device. Cuts and coagulates soft tissue.
Electrical Safety Standards ComplianceComplies with IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, and IEC 60601-2-2.
BiocompatibilityMaterials in patient-contacting portions are known to be biocompatible or passed testing according to ISO 10993-5 (Cytotoxicity), 10993-10 (Acute Systemic Toxicity), and 10993-11 (Sensitization and Irritation).
SterilizationProvided sterile, single-patient use. Sterilization method is Ethylene Oxide: SAL 10-6.
Functional Equivalence"At least as effective as the predicate device" for vessel sealing strength, tissue histological analysis, and acute/chronic animal studies.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document mentions "animal studies" for performance testing, including "acute and chronic animal studies" and "testing on animals and animal tissues." However, specific sample sizes (number of animals, number of vessels/tissue bundles, etc.) are not provided in the summary.
  • Data Provenance: The studies were prospective animal studies. The country of origin for the data is not specified but implicitly US-based, given the FDA submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document does not mention the use of human experts to establish ground truth for the animal studies. It refers to "tissue histological analysis" which would be performed by pathologists, but no details are given regarding the number or qualifications of such experts.


4. Adjudication Method for the Test Set

Since human experts are not mentioned as establishing ground truth in a consensus-based manner, no adjudication method (e.g., 2+1, 3+1) is described or applicable in the context of this summary. The results of the animal studies and histological analyses would likely be interpreted by the researchers performing the studies and potentially reviewed by veterinary pathologists.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The clearance of this device was "not based on human clinical testing." The studies focused on animal models and bench testing to demonstrate technical and performance equivalence to predicate devices.


6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

This question is not applicable to the Open FMsealer. This is a physical surgical device, not an AI algorithm. Its performance is inherent in its design and operation, not through an algorithm operating independently.


7. The Type of Ground Truth Used

The ground truth for the performance evaluations appears to be based on:

  • Animal Outcomes Data: Specifically, successful vessel sealing and tissue coagulation, indicated by "vessel sealing strength" and outcomes from "acute and chronic animal studies."
  • Pathology/Histology: "Tissue histological analysis" was performed, which relies on microscopic examination of tissue by pathologists to confirm the effects of the device (e.g., cell necrosis, collagen denaturation, seal integrity).
  • Bench Test Measurements: Quantitative measurements of physical parameters (e.g., thermal performance, seal strength).

8. The Sample Size for the Training Set

Not applicable. This device is a physical surgical instrument and does not use a "training set" in the context of machine learning.


9. How the Ground Truth for the Training Set Was Established

Not applicable. As above, there is no "training set" for this device.

{0}------------------------------------------------

510(k) SUMMARY

Open FMsealer Date of Summary: June 3, 2014 JUL 3 0 2014

General Provisions

510(k) Owner's Name:Domain Surgical, Inc.
Address:1370 South 2100 EastSalt Lake City, Utah 84108
Contact Person:Curtis Jensen, Vice President of Quality and Regulatory Affairs
Phone Number:(801) 924-4958
Fax Number:(801) 924-4951
Classification Name:Electrosurgical cutting and coagulation device and accessories (21 CFR§878.4400, Product Code GEI)
Proprietary Name:Open FMsealer
Common Name:Electrosurgical cutting and coagulation device and accessories

Name of Predicate Devices

  • FMwand Ferromagnetic Surgical System (Product Code GEI) #K130606 .
  • . LigaSure Vessel Sealing System (Product Code GEI) 510(k) #K043273

Device Description

An electrosurgical cutting and coagulation device (and accessories) is a device intended to remove tissue and control bleeding by use of high-frequency electrical current (see 21 CFR §878.4400). It is classified as a Class II (510(k)) device.

The Open FMsealer is a device intended for sealing of vessels and for soft tissue cutting and coagulation. The device is designed to be used only with the FMwand Generator (K130606). The two-piece device consists of a sterile, single-use patient-contacting shears section (Open FMsealer) and an accessory cable (Power Module or PM) that connects the FMsealer to the generator. The dual footpedal connects to the rear of the generator and can be used to activate the FMsealer if desired by the user.

The Open FMsealer is a hand-held surgical instrument intended for use in open surgical procedures where ligation of vessels or cutting and coagulation of soft tissue is desired. The Open FMsealer consists of a scissor-like handle with two actuation buttons and jaws to grasp the tissue to be affected.

The Open FMsealer is connected to the FMwand Generator (K130606) by the Power Module. The connectors on both ends of the Power Module are designed to fit only the FMwand Generator and FMwand sterile, disposable accessories. The power module and FMwand Generator were previously cleared as part of K130606.

This system creates sealing, cutting and coagulation by the application of heat and compression to tissue bundles and vessels interposed between the iaws of the instrument. Like the predicate device, the Open FMsealer includes actuation buttons that can be used to activate the device.

The Open FMsealer has been shown (in animal studies) to cut and seal vessels up to and including 7mm in diameter, lymphatics and tissue bundles as large as will fit into the jaws of the device.

Technological Comparison

The Open FMsealer employs ferromagnetic induction to generate heat in the tip. This method of heat generation is identical technology as the FMwand Ferromagnetic Surgical System (K130606), and differs from the LigaSure System which uses the resistance of the target tissue and electrical current to produce the heat necessary for cutting and sealing (bipolar electrosurgery).

{1}------------------------------------------------

PerformanceFeatureOpen FMsealerLigaSureVessel Sealing SystemFMwand Ferromagnetic SurgicalSystem
ManufacturerDomain Surgical, Inc.Covidian AGDomain Surgical, Inc.
510(k) NumberTo be assignedK043273K130606
Prescription/OTCPrescription OnlyPrescription OnlyPrescription Only
Product CodeGEIGEIGEI
ClassificationRegulation21 CFR 878.440021 CFR 878.440021 CFR 878.4400
Basis Intended UseSealing, cutting andcoagulation of soft tissueSealing, cutting and coagulation ofsoft tissueCutting and coagulation of softtissue
Heat GenerationMethodFerromagnetic inductionprovides an elevatedtemperature surface whichwill cauterize soft tissue asit seals and cuts.Application of bipolar electrosurgicalenergy to tissue interposed betweenthe jaws of the instrument.Ferromagnetic induction provides anelevated temperature blade whichwill cauterize tissue as it cuts.
OperationalControl MethodControlled power deliveredto tip of connectedaccessoryControlled power delivered to tip ofconnected accessoryControlled power delivered to tip ofconnected accessory
Mode of OperationIntermittent OperationIntermittent OperationIntermittent Operation
Output TypeType CFType CFType CF
TreatmentModalityVessel sealing, cuttingand/or coagulation usingheat combined withcompression of the targettissues between the jaws ofthe instrument.Vessel sealing, cutting and/orcoagulation using heat combinedwith compression of the targettissues between the jaws of theinstrument.Cutting and/or coagulation usingconducted heat from an elevatedtemperature handpiece tip.
Bench TestingThe Open FMsealer passedall bench tests performedincluding testing on animalsand animal tissues to showvessel sealing strength.Preclinical studies were performedto show vessel sealing strength.The FMwand Ferromagnetic SurgicalSystem passed all bench testsperformed
Meets applicablesections of IEC60601-2-2Yes. Applicable electricalsafety testing has beensuccessfully performed.Some testing was performed toshow compliance to IEC 60601-2-2(Alarm frequency and Audio Level).It is unknown if any other testingwas performed.Yes. Applicable electrical safetytesting has been successfullyperformed.
BiocompatibilityTestingMaterials used in thepatient-contacting portionsof the Open FMsealer iseither known to bebiocompatible or havepassed testing performedaccording to ISO 10993-5(Cytotoxicity), 10993-10(Acute Systemic Toxicity)and 10993-11(Sensitizationand Irritation).Details arefound in Section 12 of thissubmission.There is no mention in the deviceliterature, website or 510(k)summary as to whether theGenerator or shears were tested forbiocompatibility.Materials used in the patient-contacting portions of the FMwandFerromagnetic Surgical System areeither known to be biocompatible orhave passed testing performedaccording to ISO 10993-5(Cytotoxicity), 10993-10 (AcuteSystemic Toxicity) and 10993-11(Sensitization and Irritation).
SterilizationMethodOpen FMsealer is for single-patient use and providedsterile. Sterilization methodis Ethylene Oxide: SAL 10-6.Shears are single-patient use andprovided sterile. Sterilizationmethod is Ethylene Oxide: SAL 10-6*.FMwand handpiece is for single-patient use and is provided sterile.Sterilization method is EthyleneOxide: SAL 10-6.

Device Comparison Table (Shaded entries are identical)

The differences between the Open FMsealer and the predicate devices are discussed in detail in the appropriate sections of this submission. None of the differences raise new questions of safety and effectiveness.

{2}------------------------------------------------

Indications for Use

The Open FMsealer is intended for use in general and gynecological surgery and other open surgical procedures where ligation of vessels, including lymph vessels, is desired. The system creates a vessel ligation (seal) by the application of heat (coagulation) to vessels interposed between the jaws of the device. The Open FMsealer can be used to seal and ligate vessels up to 7mm and tissue bundles as large as will fit in the jaws of the instrument.

The Open FMsealer has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these purposes.

Performance Testing Data Summary

Questions of safety and effectiveness are the same for this device as they are for the predicate device and other similar devices on the market. All applicable bench testing was performed with the Open FMsealer to assure that it functioned as intended. The clearance of this device was not based on human clinical testing.

Performance testing that was successfully completed includes thermal testing, seal strength testing, tissue histological analysis, and acute and chronic animal studies. These studies showed that the Open FMsealer is at least as effective as the predicate device, which supports the claim of substantial equivalence.

Safety and Biocompatibility Summary

Questions of safety and effectiveness are the same for this device as they are for the predicate devices and other similar devices on the market. Bench testing was performed with the Open FMsealer to assure that it functioned as intended. The system was also tested by an accredited independent testing laboratory to assure that it complies with the applicable electrical safety standards for medical electrosurgical devices, including the applicable sections of IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, and IEC 60601-2-2.

The patient contacting materials used in the Open FMsealer were chosen for their biocompatibility, function and suitability for the intended use of this device. Successful biocompatibility testing of applicable parts of the system was completed by accredited independent testing laboratories according to ISO 10993-1 and 510(k) Memorandum G95-1. Invivo and in-vitro tissue testing was performed that demonstrates that the Open FMsealer is as effective at ligation and sealing of vessels or cutting and coagulation of soft tissue as the LigaSure System.

Conclusion

The Open FMsealer is substantially equivalent to the predicate devices. The intended use of the Open FMsealer is the same as the predicate devices with minor differences that have no effect on safety or effectiveness. The Open FMsealer differs from the LigaSure Shears in technological characteristics; however the differences do not raise different types of questions of safety and effectiveness. The information presented in the 510(k), including the bench, laboratory and animal testing, demonstrates that the Open FMsealer is as safe and effective as the predicate device for its intended use.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 30, 2014

Domain Surgical Incorporated Mr. Curtis Jensen Vice President of Quality and Regulatory Affairs 1370 South 2100 East Salt Lake City, Utah 84108

Re: K141484

Trade/Device Name: Open FMsealer Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: June 3, 2014 Received: June 5, 2014

Dear Mr. Jensen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2 l CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Mr. Curtis Jensen

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

David Krause -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. for

Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

K141484 510(k) Number (if known):

Device Name: Open FMsealer

Indications for Use:

The Open FMsealer is intended for use in general and gynecological surgery and other open surgical procedures where ligation of vessels, including lymph vessels, is desired. The system creates a vessel ligation (seal) by the application of heat (coagulation) to vessels interposed between the jaws of the device. The Open FMsealer can be used to seal and ligate vessels up to 7mm and tissue bundles as large as will fit in the jaws of the instrument.

The Open FMsealer has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these purposes.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use: _ (21 CFR 801 Subpart C)

(Please do not write below this line - Continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Joshua C.Nipper-S

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.