K Number
K141460

Validate with FDA (Live)

Date Cleared
2015-03-20

(291 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Finapres NOVA is intended to be used with patients who have a need for a noninvasive blood pressure and hemodynamic monitor. The noninvasive blood pressure waveform is measured on the subject's finger. The Finapres NOVA provides a noninvasive characterization of the arterial circulation and its beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals. Cardiac output derived from the flow signal requires a calibration with thermal dilution.

The Finapres NOVA has the option to include additional modules to extend its functionality with ECG and SpO2 measurements and blood pressure calibration.

When the SpO2 module is present, the Finapres NOVA can additionally monitor the functional oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate.

When the ECG module is present, the Finapes NOVA can additionally monitor the ECG parameters of a patient and their pulse rate. Alarms concerning the pulse rate will be available from the monitor.

When the blood pressure calibration module is present, the Finapres NOVA can additionally provide an upper arm non-invasive blood pressure measurement to determine the blood pressure value for calibration.

The Finapres NOVA is intended to be used for subjects above 18 years of age.

The Finapres NOVA is intended for use in a professional medical environment.

Device Description

The Finapres NOVA is an instrument to noninvasively monitor blood pressure and hemodynamic parameters. The Finapres NOVA provides a characterization of the arterial circulation and its beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals. The Finapres NOVA has the option to include four additional modules to extend its functionality with ECG and SpO2 measurements, blood pressure calibration and data transfer to and from the device.

The embedded software in the device provides computation of real-time and beat-tobeat blood pressure as well as hemodynamic parameters from the non-invasely measured blood pressure waveform. Hemodynamic parameters include cardiac output based on the modelflow method and total peripheral resistance.

The measurement of blood pressure in a finger is based on the arterial volume-clamp method of the Czech physiologist J. Peñáz, and the Physiocal - physiological calibration criteria for the proper unloading of the finger arteries of K.H. Wesseling. With this method, finger arterial pressure is measured using a finger cuff and an inflatable bladder in combination with an infrared plethysmograph, which consists of an infrared light source and detector.

The SpO2, upper arm calibration and ECG modules used in the Finapres NOVA are commercially available OEM modules that are used in FDA approved systems. The finger blood pressure measurement module used in the system is similar to other Finapres Medical Systems B.V. devices available on the market. The analog input/output module has been developed by Finapres during the Finapres NOVA development.

AI/ML Overview

The provided text describes the Finapres NOVA device and its substantial equivalence determination. However, it does not contain the specific information required to complete all sections of your request, particularly regarding detailed acceptance criteria testing and clinical performance studies as you've outlined.

The document explicitly states: "Clinical performance data was not required to demonstrate substantial equivalence." This means that a specific clinical study with acceptance criteria for the Finapres NOVA's performance (e.g., accuracy against a gold standard in a human population) was not conducted or presented in this 510(k) summary.

Instead, the submission relies on non-clinical performance data by testing the device against various medical device safety and performance standards (listed in Table 3 in the document) and by demonstrating substantial equivalence to predicate devices that have already been cleared.

Here's a breakdown of what can and cannot be answered from the provided text:


1. A table of acceptance criteria and the reported device performance

The document lists numerous standards that the Finapres NOVA was tested against (Table 3), indicating that compliance with these standards serves as the "acceptance criteria" for non-clinical performance. However, it does not provide specific numerical values for acceptance criteria or reported device performance for blood pressure accuracy (e.g., Mean Difference, Standard Deviation as per AAMI SP10 or ISO 81060-2:2009 for the Finapres NOVA itself). It only states that the device "met all applicable requirements."

The document mentions compliance with "AAMI SP10:2002 and its successor IEC 81060-2:2009" for blood pressure calibration, but it does not present the direct results of such a clinical validation for the Finapres NOVA. It implies that the ability to be calibrated allows for accurate data, rather than directly presenting the Finapres NOVA's accuracy results against those standards.

Therefore, a table with specific numerical acceptance criteria and reported performance cannot be fully constructed directly from this document for the Finapres NOVA's clinical accuracy.


2. Sample size used for the test set and the data provenance

Since clinical performance data was explicitly "not required to demonstrate substantial equivalence" for the Finapres NOVA device itself, there is no test set sample size or data provenance provided for a clinical study of the Finapres NOVA.

The non-clinical performance testing against standards doesn't typically involve a "test set" in the sense of clinical data from patients.


3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable, as no clinical test set for the Finapres NOVA's performance is described.


4. Adjudication method for the test set

Not applicable, as no clinical test set for the Finapres NOVA's performance is described.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The Finapres NOVA is a non-invasive blood pressure and hemodynamic monitor, not an AI-assisted diagnostic tool that would involve "human readers" or "AI assistance" in the context of interpretation.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The Finapres NOVA is described as a standalone device with embedded software that computes real-time blood pressure and hemodynamic parameters. Therefore, its "standalone" performance is implicitly what the non-clinical testing against standards evaluated. However, specific performance metrics for the algorithm itself (e.g., accuracy of cardiac output calculation against a gold standard) are not provided in numerical detail.


7. The type of ground truth used

For the non-clinical performance testing, the "ground truth" would be established by the specifications and measurement methods defined within the various international standards the device was tested against (e.g., IEC, ISO, AAMI standards for electrical safety, EMC, basic safety, essential performance, and specific requirements for sphygmomanometers, oximeters, and ECG equipment).

The document mentions: "Cardiac output derived from the flow signal requires a calibration with thermal dilution." This implies that for cardiac output measurements, thermal dilution is the clinical "gold standard" or "ground truth" for calibration. However, it's not stated that the Finapres NOVA's cardiac output was validated against this in a clinical study presented here.


8. The sample size for the training set

Not applicable. This device is not described as involving a machine learning or AI algorithm that requires a "training set" in the conventional sense for image analysis or classification. Its embedded software is based on established physiological models (e.g., arterial volume-clamp method, Modelflow method).


9. How the ground truth for the training set was established

Not applicable, as there's no machine learning "training set" described.


In summary, the provided 510(k) summary focuses on demonstrating substantial equivalence primarily through non-clinical performance data (compliance with device standards) and comparison to predicate devices, rather than through a new clinical study with specific performance acceptance criteria for the Finapres NOVA itself.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol consisting of three stylized human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 20, 2015

Finapres Medical Systems c/o Ms. Iris van Uitert Quality Manager Hogehilweg 8 NL - 1101 CC Amsterdam The Netherlands

Re: K141460 Trade/Device Name: Finapres Nova Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN, DSB, DOA, DRT Dated: February 2, 2015 Received: February 5, 2015

Dear Ms. Van Uitert:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might

{1}------------------------------------------------

Page 2 – Ms. Iris van Uitert

have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market, but it does not mean that FDA approves your device. Therefore, you may not promote or in any way represent your device or its labeling as being approved by FDA.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141460

Device Name Finapres NOVA

Indications for Use (Describe)

The Finapres NOVA is intended to be used with patients who have a need for a noninvasive blood pressure and hemodynamic monitor. The noninvasive blood pressure waveform is measured on the subject's finger. The Finapres NOVA provides a noninvasive characterization of the arterial circulation and its beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals. Cardiac output derived from the flow signal requires a calibration with thermal dilution.

The Finapres NOVA has the option to include additional modules to extend its functionality with ECG and SpO2 measurements and blood pressure calibration.

When the SpO2 module is present, the Finapres NOVA can additionally monitor the functional oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate.

When the ECG module is present, the Finapes NOVA can additionally monitor the ECG parameters of a patient and their pulse rate. Alarms concerning the pulse rate will be available from the monitor.

When the blood pressure calibration module is present, the Finapres NOVA can additionally provide an upper arm non-invasive blood pressure measurement to determine the blood pressure value for calibration.

The Finapres NOVA is intended to be used for subjects above 18 years of age.

The Finapres NOVA is intended for use in a professional medical environment.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

Finapres NOVA - K141460

This 510(k) summary is being submitted in accordance with the requirements of SMDA and 21CFR § 807.92.

1. Submitter's information

Submitter:Finapres Medical Systems B.V.Hogehilweg 8NL-1101 CC AmsterdamThe NetherlandsPhone number: +31 20 609 0974Fax Number: +31 20 609 0677Operator Number: 9051428Registration number: 3003803088
Contact person:Dr. Ir. Iris van UitertQuality ManagerPhone number: +31 20 609 0974E-mail:iris.van.uitert@finapres.com

Date of preparation: March 17, 2015

2. Device information

Trade name:Finapres NOVA Noninvasive Hemodynamic Monitor
Common name(s)Noninvasive Blood Pressure Monitor
Hemodynamic Monitor
Electrocardiograph
Oximeter
Classification nameSee Table 1
Device classificationClass II

Table 1. Classification name

Classification name21 CFR SectionProduct Code
Noninvasive Blood Pressure Monitor Measurement System870.1130DXN
Plethysmograph, Impedance870.2770DSB
Oximeter870.2700DQA
Cardiac monitor870.2300DRT

3. Predicate Devices

The Finapres NOVA is substantially equivalent in design (methodology) and indications for use to the devices shown in Table 2 that have previously been cleared:

{5}------------------------------------------------

Table 2. Predicate devices for the Finapres NOVA

Device nameManufacturer510(k)
Finometer® Model-1FMS, Finapres Medical Systems,Amsterdam, The NetherlandsK023723
Ohmeda 2350 Finapres ® Bloodpressure monitorOhmeda Medical, EnglewoodK880572
PortapresFMS, Finapres Medical Systems,Amsterdam, The NetherlandsK023338
NEXFIN_HD™BMEYE B.V., Amsterdam, The NetherlandsK072049
Nellcor Oximax N-600X PulseOximeterNellcor Puriton Bennet, Inc,PleasantonK060576
Zoll M Series with real CPR helpZoll Medical Corporation, 269 MillRoad, ChelmsfordK062537
MOVESTHORNHILL RESEARCH INC, Toronto,OntarioK093261

4. Description

The Finapres NOVA is an instrument to noninvasively monitor blood pressure and hemodynamic parameters. The Finapres NOVA provides a characterization of the arterial circulation and its beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals. The Finapres NOVA has the option to include four additional modules to extend its functionality with ECG and SpO2 measurements, blood pressure calibration and data transfer to and from the device.

The embedded software in the device provides computation of real-time and beat-tobeat blood pressure as well as hemodynamic parameters from the non-invasely measured blood pressure waveform. Hemodynamic parameters include cardiac output based on the modelflow method and total peripheral resistance.

The measurement of blood pressure in a finger is based on the arterial volume-clamp method of the Czech physiologist J. Peñáz, and the Physiocal - physiological calibration criteria for the proper unloading of the finger arteries of K.H. Wesseling. With this method, finger arterial pressure is measured using a finger cuff and an inflatable bladder in combination with an infrared plethysmograph, which consists of an infrared light source and detector.

The SpO2, upper arm calibration and ECG modules used in the Finapres NOVA are commercially available OEM modules that are used in FDA approved systems. The finger blood pressure measurement module used in the system is similar to other Finapres Medical Systems B.V. devices available on the market. The analog input/output module has been developed by Finapres during the Finapres NOVA development.

5. Indications for use

The Finapres NOVA is intended to be used with patients who have a need for a noninvasive blood pressure and hemodynamic monitor. The noninvasive blood pressure waveform is measured on the subject's finger. The Finapres NOVA provides a noninvasive characterization of the arterial circulation and its beat-to-beat variability in

{6}------------------------------------------------

pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals. Cardiac output derived from the flow signal requires a calibration with thermal dilution.

The Finapres NOVA has the option to include additional modules to extend its functionality with ECG and SpO2 measurements and blood pressure calibration.

When the SpO2 module is present, the Finapres NOVA can additionally monitor the functional oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate.

When the ECG module is present, the Finapres NOVA can additionally monitor the ECG parameters of a patient and their pulse rate. Alarms concerning the pulse rate will be available from the monitor.

When the blood pressure calibration module is present, the Finapres NOVA can additionally provide an upper arm non-invasive blood pressure measurement to determine the blood pressure value for calibration.

The Finapres NOVA is intended to be used for subjects above 18 years of age.

The Finapres NOVA is intended for use in a professional medical environment.

6. Intended Use

The Finapres NOVA is intended to be used with patients with a need for a noninvasive blood pressure and hemodynamic monitor. The noninvasive blood pressure waveform is measured on the subject's finger. The Finapres NOVA provides a noninvasive characterization of the arterial circulation and its beat-to-beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals.

The Finapres NOVA has the option to include four additional modules to extend its functionality with ECG and SpO₂ measurements, blood pressure calibration and data transfer to and from the device.

The Finapres NOVA is used in a professional medical environment such as hospitals, clinics and research institutions. The Finapres NOVA is a standalone device intended for desktop use. Measurements are to be performed under uninterrupted patient surveillance of the operator.

The user of the Finapres NOVA should be a qualified operator. The operator should have knowledge of the system and data interpretation, obtained via medical education, system manuals and/or specific courses. The device does not report any diagnosis but provides numerical values. It is the physician's responsibility to make proper judgments based on these numbers.

7. Summary of Technical Comparison with predicate devices

The Finapres NOVA is an updated model of the Ohmeda 2350, the Portapres and the Finometer® Model-1 device. As with predicate devices, the Finapres NOVA is indicated for use as a noninvasive blood pressure monitor. The Finapres NOVA is used to provide real-time and on-line monitoring and trending of hemodynamic parameters just as with the Finometer® Model-1 devices.

The Ohmeda 2350 device and the Finometer® Model-1 device have the same indications for use, the same design, and the same methodology as the Finapres NOVA Noninvasive

{7}------------------------------------------------

Hemodynamic Blood Pressure Monitor. The predicate Ohmeda 2350, the Finometer® Model-1 and the Nexfin HD provide the same noninvasive measurement and display of finger arterial blood pressure and pulse rate monitoring as the Finapres NOVA.

The Finapres NOVA consists of the same device architecture as the Finometer® Model-1 and the Nexfin HD.

The Finapres NOVA uses functionally the same embedded software for controlling the measurements as that used in the Finometer® Model-1 and Portapres device. The Finapres NOVA uses the proprietary NOVAScope/Modelflow software program for the measurement and trending of hemodynamic parameters, which is an updated version of the software that was used for data analysis in the Finometer and for the Portapres.

The Finapres NOVA can be equipped, equivalent to the Finometer® Model-1 and Nexfin_HD, with a blood pressure calibration method and does, therefore, allow for the measurement of accurate absolute blood pressure data complying with the AAMI SP10 :2002 and its successor IEC 81060-2:2009. The Finapres NOVA has, similar to the Nexfin_HD and the Finometer® Model-1, an optional Height correction unit (HCU) that allows the patient free hand movement during measurements with the device. The Finapres NOVA has, similar to the Finometer® Model-1, an optional ECG module that provides an analog ECG which is sampled and stored simultaneously with the blood pressure waveform. The Finapres NOVA has, same as the Ohmeda 2350, an optional SpO2 module that provides an SpO2 signal which is sampled and stored simultaneously with the blood pressure waveform. Similar to the Portapres, the Finapres NOVA enables 24 continuous measurements.

The three optional modules are available in the following predicate devices: MOVES, K093261; Nellcor Oximax N-600X Pulse Oximeter, K060576; Zoll M Series with real CPR help, K062537.

8. Non-clinical performance data for substantial equivalence determination

Testing of the Finapres NOVA was performed according to medical device safety standards shown in Table 3.

Standard numberStandard name
ANSI/AAMI ES60601-1:2005 /A2:2010Medical electrical equipment - Part 1: General requirements forbasic safety and essential performance
IEC 60601-1:1988 +A1:1991 +A2:1995 (2nd edition)Medical electrical equipment - Part 1: General requirements forbasic safety and essential performance
IEC 60601-1-2:2007Medical electrical equipment - Part 1-2: General requirementsfor safety -Electromagnetic compatibility - Requirements andtests
IEC 60601-1-4 1996 + Am.11999Medical electrical equipment - Part 1-4 - Collateral Standard:Programmable electrical medical systems
IEC 60601-1-8: 2006Medical electrical equipment -- Part 1-8: General requirementsfor basic safety and essential performance -- Collateralstandard: General requirements, tests and quidance for alarmsystems in medical electrical equipment and medical electricalsystems
IEC 60601-1-8: 2003 + A1:Medical electrical equipment -Part 1-8: General requirements

Table 3.Medical device standards that the Finapres NOVA is tested against

{8}------------------------------------------------

2006for safety - Collateral standard: General requirements, testsand guidance for alarm systems in medical electrical equipmentand medical electrical systems
IEC 60601-2-27:2005Medical electrical equipment - Part 2-27: Particularrequirements for the safety, including essential performance, ofelectrocardiographic monitoring equipment
IEC 80601-2-30:2009Medical electrical equipment - Part 2-30: Particularrequirements for the basic safety and essential performance ofautomated non-invasive sphygmomanometers
ISO 9919:2005Medical electrical equipment – Particular requirements for thebasic safety and essential performance of pulse oximeterequipment for medical use
ISO 80601-2-61:2011Medical electrical equipment - Part 2-61: Particularrequirements for basic safety and essential performance ofpulse oximeter equipment.
ISO 81060-2:2009Non-invasive sphygmomanometers - Part 2: Clinical validationof automated measurement type
ISO 14971:2007Medical devices – Risk management - Application of risk-management to medical devices
IEC 62304:2006Medical device software - software life cycle processes
IEC 62366:2009Medical devices - Application of usability engineering to medicaldevices.
ISO 10993-1:2009Biological evaluation of medical devices - Part 1: Evaluationand testing
AAMI / ANSI / ISO 10993-5:2009Biological evaluation of medical devices -- Part 5: Tests for InVitro cytotoxicity
ISO 10993-10:2010Biological evaluation of medical devices - Part 10: Tests forirritation and skin sensitization

Software performance was established according to FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, consistent with moderate level of concern.

The data provided demonstrates that the Finapres NOVA met all applicable requirements and that it is substantially equivalent to its predecessors Finometer® Model-1 and Ohmeda 2350 and to the Nexfin_HD™.

9. Clinical performance data for substantial equivalence determination

Clinical performance data was not required to demonstrate substantial equivalence.

10. Conclusion

On basis of the information above, it is concluded that the device is as safe, as effective, and performs as well as, or better than, the predicate devices.

{9}------------------------------------------------

510(k) Summary

Finapres NOVA - K141460

This 510(k) summary is being submitted in accordance with the requirements of SMDA and 21CFR § 807.92.

1. Submitter's information

Submitter:Finapres Medical Systems B.V.Hogehilweg 8NL-1101 CC AmsterdamThe NetherlandsPhone number: +31 20 609 0974Fax Number: +31 20 609 0677Operator Number: 9051428Registration number: 3003803088
Contact person:Dr. Ir. Iris van UitertQuality ManagerPhone number: +31 20 609 0974E-mail:iris.van.uitert@finapres.com

Date of preparation: March 17, 2015

2. Device information

Trade name:Finapres NOVA Noninvasive Hemodynamic Monitor
Common name(s)Noninvasive Blood Pressure Monitor
Hemodynamic Monitor
Electrocardiograph
Oximeter
Classification nameSee Table 1
Device classificationClass II

Table 1. Classification name

Classification name21 CFR SectionProduct Code
Noninvasive Blood Pressure Monitor Measurement System870.1130DXN
Plethysmograph, Impedance870.2770DSB
Oximeter870.2700DQA
Cardiac monitor870.2300DRT

3. Predicate Devices

The Finapres NOVA is substantially equivalent in design (methodology) and indications for use to the devices shown in Table 2 that have previously been cleared:

{10}------------------------------------------------

Table 2. Predicate devices for the Finapres NOVA

Device nameManufacturer510(k)
Finometer® Model-1FMS, Finapres Medical Systems,Amsterdam, The NetherlandsK023723
Ohmeda 2350 Finapres ® Bloodpressure monitorOhmeda Medical, EnglewoodK880572
PortapresFMS, Finapres Medical Systems,Amsterdam, The NetherlandsK023338
NEXFIN_HD™BMEYE B.V., Amsterdam, The NetherlandsK072049
Nellcor Oximax N-600X PulseOximeterNellcor Puriton Bennet, Inc,PleasantonK060576
Zoll M Series with real CPR helpZoll Medical Corporation, 269 MillRoad, ChelmsfordK062537
MOVESTHORNHILL RESEARCH INC, Toronto,OntarioK093261

4. Description

The Finapres NOVA is an instrument to noninvasively monitor blood pressure and hemodynamic parameters. The Finapres NOVA provides a characterization of the arterial circulation and its beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals. The Finapres NOVA has the option to include four additional modules to extend its functionality with ECG and SpO2 measurements, blood pressure calibration and data transfer to and from the device.

The embedded software in the device provides computation of real-time and beat-tobeat blood pressure as well as hemodynamic parameters from the non-invasely measured blood pressure waveform. Hemodynamic parameters include cardiac output based on the modelflow method and total peripheral resistance.

The measurement of blood pressure in a finger is based on the arterial volume-clamp method of the Czech physiologist J. Peñáz, and the Physiocal - physiological calibration criteria for the proper unloading of the finger arteries of K.H. Wesseling. With this method, finger arterial pressure is measured using a finger cuff and an inflatable bladder in combination with an infrared plethysmograph, which consists of an infrared light source and detector.

The SpO2, upper arm calibration and ECG modules used in the Finapres NOVA are commercially available OEM modules that are used in FDA approved systems. The finger blood pressure measurement module used in the system is similar to other Finapres Medical Systems B.V. devices available on the market. The analog input/output module has been developed by Finapres during the Finapres NOVA development.

5. Indications for use

The Finapres NOVA is intended to be used with patients who have a need for a noninvasive blood pressure and hemodynamic monitor. The noninvasive blood pressure waveform is measured on the subject's finger. The Finapres NOVA provides a noninvasive characterization of the arterial circulation and its beat-to-beat variability in

{11}------------------------------------------------

pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals. Cardiac output derived from the flow signal requires a calibration with thermal dilution.

The Finapres NOVA has the option to include additional modules to extend its functionality with ECG and SpO2 measurements and blood pressure calibration.

When the SpO2 module is present, the Finapres NOVA can additionally monitor the functional oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate.

When the ECG module is present, the Finapres NOVA can additionally monitor the ECG parameters of a patient and their pulse rate. Alarms concerning the pulse rate will be available from the monitor.

When the blood pressure calibration module is present, the Finapres NOVA can additionally provide an upper arm non-invasive blood pressure measurement to determine the blood pressure value for calibration.

The Finapres NOVA is intended to be used for subjects above 18 years of age.

The Finapres NOVA is intended for use in a professional medical environment.

6. Intended Use

The Finapres NOVA is intended to be used with patients with a need for a noninvasive blood pressure and hemodynamic monitor. The noninvasive blood pressure waveform is measured on the subject's finger. The Finapres NOVA provides a noninvasive characterization of the arterial circulation and its beat-to-beat variability in pressure and flow and in various hemodynamic parameters derived from these pressure and flow signals.

The Finapres NOVA has the option to include four additional modules to extend its functionality with ECG and SpO₂ measurements, blood pressure calibration and data transfer to and from the device.

The Finapres NOVA is used in a professional medical environment such as hospitals, clinics and research institutions. The Finapres NOVA is a standalone device intended for desktop use. Measurements are to be performed under uninterrupted patient surveillance of the operator.

The user of the Finapres NOVA should be a qualified operator. The operator should have knowledge of the system and data interpretation, obtained via medical education, system manuals and/or specific courses. The device does not report any diagnosis but provides numerical values. It is the physician's responsibility to make proper judgments based on these numbers.

7. Summary of Technical Comparison with predicate devices

The Finapres NOVA is an updated model of the Ohmeda 2350, the Portapres and the Finometer® Model-1 device. As with predicate devices, the Finapres NOVA is indicated for use as a noninvasive blood pressure monitor. The Finapres NOVA is used to provide real-time and on-line monitoring and trending of hemodynamic parameters just as with the Finometer® Model-1 devices.

The Ohmeda 2350 device and the Finometer® Model-1 device have the same indications for use, the same design, and the same methodology as the Finapres NOVA Noninvasive

{12}------------------------------------------------

Hemodynamic Blood Pressure Monitor. The predicate Ohmeda 2350, the Finometer® Model-1 and the Nexfin HD provide the same noninvasive measurement and display of finger arterial blood pressure and pulse rate monitoring as the Finapres NOVA.

The Finapres NOVA consists of the same device architecture as the Finometer® Model-1 and the Nexfin HD.

The Finapres NOVA uses functionally the same embedded software for controlling the measurements as that used in the Finometer® Model-1 and Portapres device. The Finapres NOVA uses the proprietary NOVAScope/Modelflow software program for the measurement and trending of hemodynamic parameters, which is an updated version of the software that was used for data analysis in the Finometer and for the Portapres.

The Finapres NOVA can be equipped, equivalent to the Finometer® Model-1 and Nexfin_HD, with a blood pressure calibration method and does, therefore, allow for the measurement of accurate absolute blood pressure data complying with the AAMI SP10 :2002 and its successor IEC 81060-2:2009. The Finapres NOVA has, similar to the Nexfin_HD and the Finometer® Model-1, an optional Height correction unit (HCU) that allows the patient free hand movement during measurements with the device. The Finapres NOVA has, similar to the Finometer® Model-1, an optional ECG module that provides an analog ECG which is sampled and stored simultaneously with the blood pressure waveform. The Finapres NOVA has, same as the Ohmeda 2350, an optional SpO2 module that provides an SpO2 signal which is sampled and stored simultaneously with the blood pressure waveform. Similar to the Portapres, the Finapres NOVA enables 24 continuous measurements.

The three optional modules are available in the following predicate devices: MOVES, K093261; Nellcor Oximax N-600X Pulse Oximeter, K060576; Zoll M Series with real CPR help, K062537.

8. Non-clinical performance data for substantial equivalence determination

Testing of the Finapres NOVA was performed according to medical device safety standards shown in Table 3.

Standard numberStandard name
ANSI/AAMI ES60601-1:2005 /A2:2010Medical electrical equipment - Part 1: General requirements forbasic safety and essential performance
IEC 60601-1:1988 +A1:1991 +A2:1995 (2nd edition)Medical electrical equipment - Part 1: General requirements forbasic safety and essential performance
IEC 60601-1-2:2007Medical electrical equipment - Part 1-2: General requirementsfor safety -Electromagnetic compatibility - Requirements andtests
IEC 60601-1-4 1996 + Am.11999Medical electrical equipment - Part 1-4 - Collateral Standard:Programmable electrical medical systems
IEC 60601-1-8: 2006Medical electrical equipment -- Part 1-8: General requirementsfor basic safety and essential performance -- Collateralstandard: General requirements, tests and guidance for alarmsystems in medical electrical equipment and medical electricalsystems
IEC 60601-1-8: 2003 + A1:Medical electrical equipment — Part 1-8: General requirements

Table 3.Medical device standards that the Finapres NOVA is tested against

{13}------------------------------------------------

2006for safety – Collateral standard: General requirements, testsand guidance for alarm systems in medical electrical equipmentand medical electrical systems
IEC 60601-2-27:2005Medical electrical equipment – Part 2-27: Particularrequirements for the safety, including essential performance, ofelectrocardiographic monitoring equipment
IEC 80601-2-30:2009Medical electrical equipment – Part 2-30: Particularrequirements for the basic safety and essential performance ofautomated non-invasive sphygmomanometers
ISO 9919:2005Medical electrical equipment – Particular requirements for thebasic safety and essential performance of pulse oximeterequipment for medical use
ISO 80601-2-61:2011Medical electrical equipment – Part 2-61: Particularrequirements for basic safety and essential performance ofpulse oximeter equipment.
ISO 81060-2:2009Non-invasive sphygmomanometers – Part 2: Clinical validationof automated measurement type
ISO 14971:2007Medical devices – Risk management – Application of riskmanagement to medical devices
IEC 62304:2006Medical device software – software life cycle processes
IEC 62366:2009Medical devices – Application of usability engineering to medicaldevices.
ISO 10993-1:2009Biological evaluation of medical devices – Part 1: Evaluationand testing
AAMI / ANSI / ISO 10993-5:2009Biological evaluation of medical devices – Part 5: Tests for InVitro cytotoxicity
ISO 10993-10:2010Biological evaluation of medical devices – Part 10: Tests forirritation and skin sensitization

Software performance was established according to FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, consistent with moderate level of concern.

The data provided demonstrates that the Finapres NOVA met all applicable requirements and that it is substantially equivalent to its predecessors Finometer® Model-1 and Ohmeda 2350 and to the Nexfin_HD™.

9. Clinical performance data for substantial equivalence determination

Clinical performance data was not required to demonstrate substantial equivalence.

10. Conclusion

On basis of the information above, it is concluded that the device is as safe, as effective, and performs as well as, or better than, the predicate devices.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).