K Number
K141385

Validate with FDA (Live)

Date Cleared
2015-03-27

(304 days)

Product Code
Regulation Number
882.5360
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RAPIDSORB Injectable Polymer System (IPS) is intended for use in non-load bearing fracture repair and reconstructive procedures of the craniofacial skeleton (excluding the upper and lower jaw) in pediatric and adult populations. RAPIDSORB IPS fasteners are designed to be used for the fixation of RAPIDSORB plates, and sheets.

In addition, RAPIDSORB IPS implants and instruments may be used with RAPIDSORB meshes and sheets in non-load bearing applications for maintaining the relative position of, and/or containing, bony fragments, bone grafts (autograft or allograft), or bone graft substitutes in craniofacial reconstruction (excluding the upper and lower jaw).

Device Description

DePuy Synthes RAPIDSORB IPS Injectable Polymer System consists of a sterile resorbable polymer, 85:15 Poly (L-lactide co-glycolide) and a battery powered delivery device. The polymer can be extruded from the delivery device into a predrilled pilot hole at the surgical site.

The polymer is presented as a rod in a sterile package (cartridge) that fits into the delivery device. The device then briefly heats the polymer at the implant site beyond the glass transition temperature of the polymer, allowing the polymer to be extruded in a controlled manner in a preselected length from the device through a plate, mesh, or sheet and into a predrilled hole at the site of the defect. The extruded polymer fills the predrilled hole. Retention is attained through friction and interdigitation with the surrounding bone.

The delivery device consists of three principal components: 1) a Power Drive Unit, which is nonsterile and reusable, and which fits into and is enclosed by 2) a disposable, sterile, single-use, rigid plastic shell and cap, and 3) a sterile, single-use, battery pack that attaches to the shell and serves as the power source. The delivery device is comfortably shaped and balanced to fit the left or right hand of the surgeon.

Enclosing the Power Drive Unit thus assures sterile delivery of the polymer at the implant site. The brevity of the increase in the temperature of the polymer combined with the controlled, minimized volume of polymer implanted prevents tissue injury or necrosis at the implant site.

AI/ML Overview

The given text describes the DePuy Synthes RAPIDSORB Injectable Polymer System (IPS) and the studies performed to demonstrate its substantial equivalence to predicate devices. However, the document does NOT include a table of acceptance criteria for specific performance metrics nor does it directly list "reported device performance" against such criteria. Instead, it summarizes the conclusions of various nonclinical, in vitro, and animal studies.

Here's an attempt to extract the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria with corresponding device performance values. The "Conclusion" sections for each test type indicate whether the device "Passed" or met a comparative standard (e.g., "identical to the cleared polymers," "significantly greater," "superior," "comparable").

Test TypeAcceptance Criteria (Implied)Reported Device Performance
Biocompatibility (various ISO tests)Compliance with ISO 10993 standards and FDA guidance; no significant adverse biological reactions.Pass. The IPS polymer implant and delivery device are biocompatible. (Specific tests: Guinea Pig Maximization Sensitization, Intra-cutaneous Reactivity, MEM Elution Assay, Acute Systemic Injection all passed; no significant dermal reactions, no cytotoxicity, etc.)
Electrical Safety (IEC 60601-1)Compliance with IEC 60601-1 medical device electrical standards.Pass.
Controlled Extraction StudyIdentity to previously cleared polymers.The polymer in the IPS submission was found to be identical to the cleared polymers.
Axial Pullout Test (in vitro)Sufficient pullout strength compared to predicate.At each time point, the pull-out strength was significantly greater for the Rapidsorb IPS Fastener compared to the predicate Lactosorb.
Shear Testing (in vitro)Sufficient shear strength compared to predicate.The mean peak load of the Rapidsorb IPS fastener was superior to the mean peak load for the Lactosorb fastener.
Immature Ovine Model (animal)Comparable biocompatibility, irritancy scores, and healing to predicate.Demonstrated comparable biocompatibility, irritancy scores, and healing to the predicate Sonic Weld Sonic Pin fasteners.

2. Sample Sizes and Data Provenance

  • Test Set Sample Sizes:
    • Biocompatibility (Guinea Pig, Intra-cutaneous Reactivity, Acute Systemic Injection): Guinea pigs (number not specified, but stated as "Guinea pigs were injected...", "Test animals were topically patched...", implying multiple). Rabbits (three male NZW rabbits). Mice (number not specified, but "Mice were treated...").
    • In Vitro Axial Pullout Test: "two different IPS fasteners lengths" compared to "a 1.5 x 3mm (length) Direct Drive Lactosorb Screw." (Specific number of samples per group not specified).
    • In Vitro Shear Testing: IPS fastener compared to predicate Lactosorb Fastener. (Specific number of samples per group not specified).
    • Immature Ovine Model Animal Test: Number of animals used for the study is not specified, but the study was conducted in an "immature ovine model."
  • Data Provenance: The studies are nonclinical (in vitro bench tests, animal tests, and laboratory tests) and are likely prospective studies conducted specifically for this 510(k) submission. The country of origin for the data is not specified, but the submitting company is based in West Chester, Pennsylvania, USA.

3. Number of Experts and Qualifications

This document does not describe the use of human experts to establish ground truth for a test set in the context of diagnostic accuracy. The studies described are engineering, material science, and animal studies, not interpretive diagnostic studies.

4. Adjudication Method

Not applicable for the types of studies described (nonclinical, in vitro, and animal studies).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. The document describes the evaluation of a medical device (a polymer system for craniofacial repair), not an AI-assisted diagnostic tool. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant or described.

6. Standalone (Algorithm Only) Performance Study

No. This device is a polymer system, not an algorithm or AI.

7. Type of Ground Truth Used

The ground truth or reference standard for these studies is based on:

  • Standard biological responses (e.g., absence of toxicity, sensitization, irritation for biocompatibility tests).
  • Quantitative mechanical measurements (e.g., Newtons for pullout strength, peak load for shear strength).
  • Histopathological evaluation of bone and soft tissues in animal models for biocompatibility and healing.
  • Chemical identity comparison for controlled extraction studies.
  • Compliance with electrical safety standards.

8. Sample Size for the Training Set

Not applicable. This is a medical device, not an AI model requiring a training set.

9. How the Ground Truth for the Training Set was Established

Not applicable. This is a medical device, not an AI model.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol of a human figure in profile, with three overlapping heads suggesting community and support.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 27, 2015

Depuy Synthes Biomaterials Jeffrey L. Dow, JD Senior Principal Regulatory Affairs Strategy Advisor 1230 Wilson Drive West Chester, Pennsylvania 19380

Re: K141385

Trade/Device Name: Depuy Synthes RAPIDSORB Injectable Polymer System (IPS) Regulation Number: 21 CFR 882.5360 Regulation Name: Cranioplasty Plate Fastener Regulatory Class: Class II Product Code: HBW Dated: February 23, 2015 Received: February 26, 2015

Dear Mr. Dow,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carlos L. Pena -SFD/△

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141385

Device Name RAPIDSORB Injectable Polymer System (IPS)

Indications for Use (Describe)

The RAPIDSORB Injectable Polymer System (IPS) is intended for use in non-load bearing fracture repair and reconstructive procedures of the craniofacial skeleton (excluding the upper and lower jaw) in pediatric and adult populations. RAPIDSORB IPS fasteners are designed to be used for the fixation of RAPIDSORB plates, and sheets.

In addition, RAPIDSORB IPS implants and instruments may be used with RAPIDSORB meshes and sheets in non-load bearing applications for maintaining the relative position of, and/or containing, bony fragments, bone grafts (autograft or allograft), or bone graft substitutes in craniofacial reconstruction (excluding the upper and lower jaw).

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for DePuy Synthes Biomaterials. The logo consists of a circular symbol with overlapping shapes on the left, followed by the text "DePuy Synthes" in a bold, sans-serif font. Below "DePuy Synthes" is the word "BIOMATERIALS" in a smaller font. Underneath that is the text "companies of Johnson & Johnson" in a smaller, italicized font.

510(k) SummaryK141385
Submitter:DePuy Synthes Biomaterials1230 Wilson DriveWest Chester, PA 19380
Date of creation:February 23, 2015
Company Contact:Jeffrey L. Dow, JDSenior Principal R/A Strategy AdvisorDePuy Synthes Biomaterials484 467 4174Fax: 484 356 9636dow.jeff@synthes.com
Name of Device:DePuy Synthes RAPIDSORB® Injectable Polymer System (IPS)
Device Classification:Class II, 21 CFR § 882.5360
Product Code:HBW
Common Name:Cranioplasty Plate Fastener
Predicate Devices:(Please see Table 1, below)Synthes (USA) Rapid Resorbable Tack System (K050204)Synthes (USA) Rapid Resorbable Fixation System (K062789)(Principal Predicate) KLS-Martin, L.P., SonicWeld RX (Resorb-X) (K080862)Biomet Inc., Lactosorb Trauma Plating System (K992355)
Indications for Use:The RAPIDSORB Injectable Polymer System (IPS) is indicatedfor use in non-load bearing fracture repair and reconstructiveprocedures of the craniofacial skeleton (excluding the upper andlower jaw) in pediatric and adult populations. RAPIDSORB IPSfasteners are designed to be used for the fixation of RAPIDSORBplates, meshes, and sheets.
In addition, RAPIDSORB IPS implants and instruments may beused with RAPIDSORB meshes and sheets in non-load bearingapplications for maintaining the relative position of, and/orcontaining, bony fragments, bone grafts, (autograft or allograft), orbone graft substitutes in craniofacial reconstruction (excluding theupper and lower jaw)."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for DePuy Synthes BIOMATERIALS. The logo consists of a circular graphic to the left of the text "DePuy Synthes" in a dark blue, sans-serif font. Below "DePuy Synthes" is the word "BIOMATERIALS" in a smaller, sans-serif font. Underneath that is the text "COMPANIES of Johnson & Johnson" in a small font.

Radiolucent

Radiolucent

Radiolucent

DePuy Synthes Synthes (USA) Rapid Synthes (USA) Rapid (Principal Predicate) KLS-Martin, Biomet Inc., Lactosorb Trauma Plating System RAPIDSORB® Iniectable Resorbable Tack System Resorbable Fixation System LP., SonicWeldRX (Resorb X) (K992355) Polymer System (IPS) (K050204) (K062789) (K080862) (K141385) Indications. DePuy Synthes Indication. Synthes Rapid Indication. The Synthes (USA) Indication.The KLS Martin Indication: trauma procedures of the midface or RAPIDSORB Injectable Resorbable Tack System is Rapid Resorbable Fixation SonicWeld RX (Resorb X) is craniofacial skeleton Polymer System (IPS) is intended for use in fracture System is intended for use in intended for use in fracture repair Specific Indications: repair and reconstructions of the intended for use in non-load fracture repair in reconstructive and reconstructive procedures of 1. Comminuted fractures of the naso-ethmoidal bearing fracture repair and craniofacial skeleton. In procedures of the craniofacial the craniofacial skeleton in infraorbital areas reconstructive procedures of the addition, Rapid Resorbable skeleton in pediatric and adult pediatric and adult populations. In 2. Comminuted fractures of the frontal sinus wall craniofacial skeleton (excluding Tacks may be used in non-load populations. In addition. addition, resorbable meshes, plates. 3. Pediatric midface or craniofacial trauma the upper and lower jaw) in bearing applications for resorbable meshes, sheets, screws and pins may be used in 4. LeFort (I,II,III) fractures pediatric and adult populations. maintaining the relative position screws and tacks may be used in non-load bearing applications for 5. Orbital floor fractures RAPIDSORB IPS fasteners are and/or containing bony non-load bearing applications maintaining the relative position of. 6. Fractures of the maxilla, zygoma, zygomatic arch, designed to be used for the fragments, bone grafts (autograft for maintaining the relative and/or containing, bony fragments, orbital rim, nasal,ethmoid, and lacrimal bones position and/or containing, bony bone grafts (autograft or allograft), 7. Trauma of the craniofacial skeleton including: frontal, fixation of DePuy Synthes or allograft) or bone graft RAPIDSORB plates, meshes, substitutes in reconstruction of fragments, bone grafts, or bone graft substitutes in parietal, temporal, sphenoid, and occipital bones and sheets. the craniofacial or mandibular (autograft or allograft) or bone reconstruction of the craniofacial or General Indication: reconstructive procedures of the mandibular areas. midface or craniofacial skeleton areas. graft substitutes in In addition, RAPIDSORB IPS reconstruction of the Specific Indications: implants and instruments may be craniofacial or mandibular areas. 1. Infant craniofacial surgery (i.e.craniosynostosis, used with RAPIDSORB meshes congenital malformation, trauma, etc.) and sheets in non-load bearing 2. LeFort (I,II,III) osteotomies applications for maintaining the 3. Tumor reconstruction in midsface or craniofacial relative position of, and/or procedures 4. Bone graft procedures in the midface or craniofacial containing, bony fragments. bone grafts, (autograft or skeleton allograft), or bone graft 5. Pediatric reconstructive procedures 6. Reconstructive procedures of the craniofacial skeleton substitutes in craniofacial reconstruction (excluding the including: frontal, parietal, temporal, sphenoid, and upper and lower jaw). occipital bones 7. Craniotomy flap fixation Procode JEY 872.4760 Procode JEY 872.4760 Procode: HBW, 882.5360 Procode JEY 872.4760 Procode HRS, HWC 888.3030 Material: 85:15 Poly (L-lactide Material: 85:15 Poly (L-lactide Material: 85:15 Poly (L-lactide Material: Poly (D, L)-Lactide-Acid Material: 82 poly L-lactide acid and 18% poly glycolic co-glycolyde) co-glycolyde) co-glycolyde) acid Biocompatibility: Established Biocompatibility: Established Biocompatibility: Established Biocompatibility: Established Biocompatibility: Established Resorption time: Approximately Resorption time: Approximately Resorption time: Approximately Resorption time: unknown Resorbtion time: Approximately 12 months 12 months 12 months 12 months Sterile components: Single use Sterile components: Single use Sterile components: Single use Sterile components: Single use Sterile components: Single use only; Gamma or EtO only; Gamma or EtO sterilized only; Gamma or EtO sterilized only; Gamma or EtO sterilized only; Gamma or EtO sterilized sterilized

Radiolucent

Radiolucent

TABLE 1: COMPARISON OF INDICATIONS AND TECHNOLOGIES BETWEEN IPS AND PREDICATES

{5}------------------------------------------------

Device Description: DePuy Synthes RAPIDSORB IPS Injectable Polymer System consists of a sterile resorbable polymer, 85:15 Poly (L-lactide co-glycolide) and a battery powered delivery device. The polymer can be extruded from the delivery device into a predrilled pilot hole at the surgical site.

The polymer is presented as a rod in a sterile package (cartridge) that fits into the delivery device. The device then briefly heats the polymer at the implant site beyond the glass transition temperature of the polymer, allowing the polymer to be extruded in a controlled manner in a preselected length from the device through a plate, mesh, or sheet and into a predrilled hole at the site of the defect. The extruded polymer fills the predrilled hole. Retention is attained through friction and interdigitation with the surrounding bone.

The delivery device consists of three principal components: 1) a Power Drive Unit, which is nonsterile and reusable, and which fits into and is enclosed by 2) a disposable, sterile, single-use, rigid plastic shell and cap, and 3) a sterile, single-use, battery pack that attaches to the shell and serves as the power source. The delivery device is comfortably shaped and balanced to fit the left or right hand of the surgeon.

Enclosing the Power Drive Unit thus assures sterile delivery of the polymer at the implant site. The brevity of the increase in the temperature of the polymer combined with the controlled, minimized volume of polymer implanted prevents tissue injury or necrosis at the implant site.

Nonclinical Tests (Please see Tables 2 And 3 Below)

Biocompatibility testing of the IPS polymer implant and delivery device was performed in accordance with the applicable FDA guidance "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications" July 28, 2014, and applicable International Standards Organization requirements of ISO 10993 for implants of greater than 30 days duration, as well as those standards for brief patient contact. The results of all those tests showed that the IPS polymer implant and delivery device are biocompatible. '

1 The term "substantial equivalence" as used in this 510(k) notification of substantial equivalence found in the Federal Food, Drug and Cosmetic Act 21 USC $301 et seq., as amended, and as applied under 21 CFR Part 807, Subpart E, under which a device can be marketed without pre-market approval or reclassification. A determination of substantial equivalence under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in substantial equivalence herein may be construed as an admission against interest under the US patent laws or their application by the courts.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for DePuy Synthes Biomaterials. The logo consists of a geometric shape to the left of the text "DePuy Synthes". Below that is the word "BIOMATERIALS". Underneath that is the text "COMPANIES OF Johnson & Johnson".

TABLE 2: NONCLINICAL TESTS SUMMARY,

TitleTest ArticleMethodsConclusion
ISO Guinea Pig MaximizationSensitization TestRAPIDSORB IPS DeliveryDevice Shell, Battery ShellGuinea pigs were injected with test article extract and FCA,and the same number guinea pigs were injected with controlblank and FCA. Test animals were topically patched withtest extract and control animals were patched with controlblank. Patches were removed after 48 hours. After 2 weeks,the animals were patched with test extract and control blank.This test was conducted in accordance to ISO 10993-10:2010part 10: Test for irritation and skin sensitization.Pass.
IEC 60601-1(ANSI/AAMI ES60601-1:2005/(R)2012 and C1:2009/(R)2012and A2:2010/(R)2012)Accredited laboratory tests ofcompliance with the medicaldevice electrical standards ofIEC 60601including software ofthe Power Drive unit, shell, capand batterySuch tests as are required for certification underANSI/AAMI ES60601-1:2005/(R)2012 andC1:2009/(R)2012 and A2:2010/(R)2012Pass.
Controlled Extraction Study85: 15 poly (L-lactide- co-glycolide)The polymer cleared in K030069, K050204, and K062789was compared to the polymer described in the IPSsubmission to establish its identity to the cleared polymer.This test followed test guidelines described in ISO 10993-18:2005.The polymer in the IPS submissionwas found to be identical to the clearedpolymers.
ISO Intra-cutaneous Reactivity TestRAPIDSORB IPS DeliveryDevice Shell, Battery ShellTest article was extracted at a ratio of 4 g to 20 mL in normalsaline and sesame oil for 72 hours at 50°C. Test articles andvehicle controls were injected into three male NZW rabbits.Rabbits were observed at 24, 48, and 72 hours. This test wasperformed in accordance with ISO 10993-10:2010Pass.There were no significant dermalreactions observed.
ISO MEM Elution Assay with L-929Mouse Fibroblast CellsRAPIDSORB IPS DeliveryDevice Shell, Battery ShellThe test article was extracted in E-MEM + 5% FBS for 24hours at 37°C. The L-929 cells were incubated with the extractfor 72 hours. The cultures were evaluated for cytotoxic effectsby microscopic examination at 24, 48, and 72 hours.. This testwas performed in accordance with ISO 10993-5:2009Pass.The test article did not inducecytotoxicity.
ISO Acute systemic Injection TestRAPIDSORB IPS DeliveryDevice Shell, Battery ShellMice were treated by intravenous or intraperitoneal injectionto screen for potential toxic effects as a result of a single-dosesystemic injection as recommended in ISO 10993-11:2006.Pass.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for DePuy Synthes BIOMATERIALS. The logo consists of a circular design on the left, followed by the text "DePuy Synthes" in a bold, sans-serif font. Below "DePuy Synthes" is the word "BIOMATERIALS" in a smaller, sans-serif font. Underneath that is the text "COMPANIES of Johnson & Johnson" in a smaller font.

TABLE 3: IN VITRO AND ANIMAL TESTS DEMONSTRATING SUBSTANTIAL
EQUIVALENCE.
TestTest Method SummaryResults
Axial Pullout TestThe object of this in vitrobench test was to determinethe pullout strength of 2different IPS fasteners lengthscompared to the predicate, a1.5 x 3mm (length) DirectDrive Lactosorb Screw.Measurements were made attime points as the testmaterials degraded.At each time point, the pullout strength, measured inNewtons, was significantlygreater for the Rapidsorb IPSFastener compared to thepredicate Lactosorb.
Shear TestingThe object of this in vitrobench test was to compare theinitial and degraded strengthof the IPS fastener to that ofthe predicate LactosorbFastener at comparable timepoints.Based on statisticalcomparison of data, the meanpeak load of the RapidsorbIPS fastener was superior tothe mean peak load for theLactosorb fastener.
Immature ovine model animaltestAn animal test in an immatureovine model was undertakento show the thermal effects ofthe IPS fastener compared tothat of the KLS Sonic WeldSonic Pin.Evaluation of bone and softtissues (dura matter and brain)associated with cranialimplant sites treated withRapidsorb IPS fasteners andthe predicate Sonic WeldSonic Pin fastenersdemonstrated comparablebiocompatibility, irritancyscores, and healing

Conclusion.

Performance testing of the RAPIDSORB IPS has demonstrated that the subject device is substantially equivalent to the predicate devices.

§ 882.5360 Cranioplasty plate fastener.

(a)
Identification. A cranioplasty plate fastener is a screw, wire, or other article made of tantalum, vitallium, or stainless steel used to secure a plate to the patient's skull to repair a skull defect.(b)
Classification. Class II (performance standards).