K Number
K141284

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-07-23

(68 days)

Product Code
Regulation Number
888.3050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CASPIAN OCT/MESA Mini/ DENALI Mini Spinal System is intended to provide stabilization as an adjunct to fusion of the cervical spine and occipito-cervico-thoracic junction (occiput-T3) when used with autograft or allograft and indicated for the following:

· DDD (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies)

  • · Spondylolisthesis
  • Spinal stenosis
  • · Fracture/dislocation
  • · Revision of previous cervical spine surgery
  • · Tumors
  • · Atlantoaxial fracture with instability
  • · Occipitocervical dislocation

The occipital bone screws are limited to occipital fixation only

The rod and hook components are in the cervical/upper thoracic (CI-T3) spine. The pedicle screws are limited to placement in the TI-T3 in treating thoracic conditions only. The pedicle screws are not intended to be placed in or treat conditions involving the cervical spine.

The CASPIAN OCT/MESA Mini/ DENALI Mini System can also be linked to the Range Spinal System using rod connectors or transitional rods.

Device Description

The CASPIAN OCT/MESA Mini/ DENALI Mini System is a top-loading, multiple component, posterior (occipital-cervical-thoracic) spinal fixation system. The purpose of the current submission is to add 03.5 contoured rods to expand the rod offerings for the system.

Materials: The subject devices are manufactured from Ti6Al4V Eli per ASTM and ISO standards.

Function: The system functions as an adjunct to fusion to provide immobilization of stabilization of the posterior cervical and thoracic spine.

AI/ML Overview

The provided document is a 510(k) summary for a spinal fixation system, not a study report for a medical device that uses AI or reports on acceptance criteria and performance data in the typical sense of a diagnostic or screening device.

Therefore, many of the requested categories for acceptance criteria and study design elements (such as sample size for test sets, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, training set details, etc.) are not applicable to this type of submission.

This 510(k) summary focuses on demonstrating substantial equivalence to previously marketed devices based on design, function, materials, and intended use, particularly for adding new rod sizes. The "acceptance criteria" here are implicitly related to mechanical testing and engineering rationale to ensure the new components maintain the same safety and effectiveness profile as the predicate devices.

Here's a breakdown of the relevant information from the document in relation to your request:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Acceptance Criteria (Implied by 510(k))Reported Device Performance (from document)
Mechanical PerformanceEquivalence to predicate devices in mechanical strength and stability under relevant biomechanical loads (e.g., as per ASTM F1717)."The worst case components of the CASPIAN OCT/MESA Mini/ DENALI Mini System were previously tested in accordance with ASTM F1717 and determined to be equivalent to predicate devices."
New components (03.5 contoured rods) do not represent a new worst-case scenario."The subject rods were determined not to represent a new worst case via finite element analysis and engineering rationale."
Material BiocompatibilityUse of biocompatible materials known to be safe for implantation."The subject devices are manufactured from Ti6Al4V Eli per ASTM and ISO standards."
FunctionalityAbility to provide immobilization and stabilization of the posterior cervical and thoracic spine as an adjunct to fusion."The system functions as an adjunct to fusion to provide immobilization of stabilization of the posterior cervical and thoracic spine."
Intended UseAlignment with established indications for use in spinal fixation.The Indications for Use statement in the document specifies conditions like DDD, Spondylolisthesis, Fracture/dislocation, etc. The device is deemed substantially equivalent for these uses.
Design CharacteristicsNo significant differences from predicate devices that would adversely affect use."There are no significant differences between the CASPIAN OCT/MESA Mini/ DENALI Mini System and other systems currently being marketed which would adversely affect the use of the product."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not Applicable: This submission relies on mechanical testing standards (ASTM F1717) and engineering rationale/finite element analysis, not clinical 'test sets' of patient data in the context of AI or diagnostic devices. There is no mention of "patient data" test sets.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not Applicable: "Ground truth" in this context would relate to defining mechanical testing parameters and interpreting engineering analyses, handled by biomechanical engineers and regulatory specialists, rather than clinical experts establishing ground truth for diagnostic imaging.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable: This relates to clinical data adjudication, which is not part of this 510(k) submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable: This is a submission for a spinal implant, not an AI-powered diagnostic or screening device, so MRMC studies are irrelevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable: This is a medical device (spinal implant), not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • For the mechanical performance, the "ground truth" is established by adherence to recognized biomechanical testing standards (e.g., ASTM F1717) and engineering principles/ finite element analysis (FEA), indicating that the device meets defined physical strength and durability criteria. The overall ground truth for regulatory clearance is substantial equivalence to existing, legally marketed devices.

8. The sample size for the training set

  • Not Applicable: This is a spinal implant, not a machine learning model, so there is no training set.

9. How the ground truth for the training set was established

  • Not Applicable: No training set exists for this type of device.

In summary, the provided document is a regulatory submission for a spinal fixation system, cleared based on substantial equivalence demonstrated through mechanical testing and engineering rationale, not clinical study data in the context of diagnostic performance.

{0}------------------------------------------------

JIII 2 3 2014

510(k) Summary

For the

CASPIAN OCT/MESA Mini/ DENALI Mini System

This 510(k) summary for the CASPIAN OCT/MESA Mini/ DENALI Mini System is provided as required per Section 513(i)(3) of the Food, Drug and Cosmetic Act.

1. Submitter:

K2M, Inc. 751 Miller Drive SE Leesburg, VA 20175

Contact Person:

Nancy Giezen K2M, Inc. Telephone: 703-777-3155

Date Prepared: 07/22/2014

2. Tradename:CASPIAN OCT/MESA Mini/ DENALI Mini System
Common Name:Spinal Fixation System
Classification Name:Spinal Interlaminal Fixation Orthosis (21 CFR 888.3050)
Device Product Code:KWP
Regulatory Class:Class II

3. Predicate or legally marketed devices which are substantially equivalent:

4. Description of the device:

The CASPIAN OCT/MESA Mini/ DENALI Mini System is a top-loading, multiple component, posterior (occipital-cervical-thoracic) spinal fixation system. The purpose of the current submission is to add 03.5 contoured rods to expand the rod offerings for the system.

Materials: The subject devices are manufactured from Ti6Al4V Eli per ASTM and ISO standards.

Function: The system functions as an adjunct to fusion to provide immobilization of stabilization of the posterior cervical and thoracic spine.

5. Intended Use:

The CASPIAN OCT/MESA Mini/ DENALI Mini Spinal System is intended to provide stabilization as an adjunct to fusion of the cervical spine and occipito-cervico-thoracic junction (occiput-T3) when used with autograft or allograft and indicated for the following:

  • . DDD (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies)

{1}------------------------------------------------

  • Spondylolisthesis .
  • Spinal stenosis .
  • Fracture/dislocation .
  • Revision of previous cervical spine surgery ●
  • Tumors �
  • Atlantoaxial fracture with instability .
  • Occipitocervical dislocation .

The occipital bone screws are limited to occipital fixation only

The rod and hook components are intended for use in the cervical/upper thoracic (C1-T3) spine. The pedicle screws are limited to placement in the TI-T3 in treating thoracic conditions only. The pedicle screws are not intended to be placed in or treat conditions involving the cervical spine.

The CASPIAN OCT/MESA Mini/ DENALI Mini System can also be linked to the Range Spinal System using rod connectors or transitional rods.

    1. Comparison of the technology characteristics of the device to predicate and legally marketed devices:
      The worst case components of the CASPIAN OCT/MESA Mini/ DENALI Mini System were previously tested in accordance with ASTM F1717 and determined to be equivalent to predicate devices. The subject rods were determined not to represent a new worst case via finite element analysis and engineering rationale. There are no significant differences between the CASPIAN OCT/MESA Mini/ DENALI Mini System and other systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to these other devices in design, function, material and intended use,

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 23, 2014

K2M. Incorporated Ms. Nancy Giezen Manager, Regulatory Affairs 751 Miller Drive Southeast Leesburg. Virginia 20175

Re: K141284

Trade/Device Name: CASPIAN OCT/MESA Mini/DENALI Mini Spinal System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: KWP Dated: June 23, 2014 Received: June 24, 2014

Dear Ms. Giezen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Nancy Giezen

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Ronald P. Jean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K141284

Device Name

CASPIAN OCT/MESA Mini/ DENALI Mini Spinal System

Indications for Use (Describe)

The CASPIAN OCT/MESA Mini/ DENALI Mini Spinal System is intended to provide stabilization as an adjunct to fusion of the cervical spine and occipito-cervico-thoracic junction (occiput-T3) when used with autograft or allograft and indicated for the following:

· DDD (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies)

  • · Spondylolisthesis
  • Spinal stenosis
  • · Fracture/dislocation
  • · Revision of previous cervical spine surgery
  • · Tumors
  • · Atlantoaxial fracture with instability
  • · Occipitocervical dislocation

The occipital bone screws are limited to occipital fixation only

The rod and hook components are in the cervical/upper thoracic (CI-T3) spine. The pedicle screws are limited to placement in the TI-T3 in treating thoracic conditions only. The pedicle screws are not intended to be placed in or treat conditions involving the cervical spine.

The CASPIAN OCT/MESA Mini/ DENALI Mini System can also be linked to the Range Spinal System using rod connectors or transitional rods.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Jonathan H. Peck -S

{5}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@lda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 888.3050 Spinal interlaminal fixation orthosis.

(a)
Identification. A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.(b)
Classification. Class II.