K Number
K141210

Validate with FDA (Live)

Date Cleared
2015-01-29

(265 days)

Product Code
Regulation Number
866.5100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

QUANTA Flash SS-B is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-SS-B autoantibodies in human serum. The presence of anti-SS-B autoantibodies, in conjunction with clinical findings and other laboratory tests is an aid in the diagnosis of Sjögren's Syndrome and Systemic Lupus Erythematosus.

QUANTA Flash SS-B Calibrators are intended for use with the QUANTA Flash SS-B Reagents for the determination of IgG anti-SS-B autoantibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

QUANTA Flash SS-B Controls are intended for use with the QUANTA Flash SS-B reagents for quality control in the determination of IgG anti-SS-B autoantibodies in human serum.

Device Description

The QUANTA Flash SS-B assay is designed to run on the BIO-FLASH® instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash SS-B assay utilizes a reagent cartridge format, which is compatible with the BIO-FLASH instrument.

Purified recombinant SS-B antigen is coated onto paramagnetic beads. The bead suspension is lyophilized and stored in the bead tube. Prior to use in the BIO-FLASH system, the sealed reagent tubes are pierced with the reagent cartridge lid and the beads are rehydrated and resuspended using resuspension buffer by pipetting up and down with a transfer pipette. The reagent cartridge is then loaded onto the BIO-FLASH instrument. Samples are also loaded onto the instrument in sample racks. Serum samples are prediluted by the BIO-FLASH with system rinse in a small disposable plastic cuvette. Small amounts of the diluted patient serum, the beads, and assay buffer are all combined into a second cuvette, and mixed. This cuvette is then incubated at 37°C. The beads are magnetized and washed several times. Isoluminol conjugated anti-human IgG antibodies are then added to the cuvette, and again incubated at 37°C. The beads are magnetized and washed repeatedly. The isoluminol conjugate is oxidized when Trigger 1 (Fe(II)coproporphyrin in sodium hydroxide solution) and Trigger 2 (ureahydrogen peroxide in sodium chloride solution) are added to the cuvette, and the flash of light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. The RLU are proportional to the amount of isoluminol conjugate that is bound to the human IgG, which is in turn proportional to the amount of anti-SS-B antibodies bound to the corresponding beads.

For quantitation, the QUANTA Flash SS-B assay utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. Every new lot number of reagent cartridge must be calibrated before first use, with the QUANTA Flash SS-B Calibrators. Based on the results obtained with the two Calibrators included in the Calibrator Set (sold separately), an instrument specific Working Curve is created, which is used to calculate chemiluminescent units (CU) from the instrument signal (RLU) obtained for each sample.

The QUANTA Flash SS-B kit contains the following materials:

One (1) QUANTA Flash SS-B Reagent Cartridge One (1) vial of Resuspension buffer One (1) Transfer pipette

The QUANTA Flash SS-B reagent cartridge contains the following reagents for 50 determinations:

  • a. SS-B antigen coated paramagnetic beads, lyophilized.
  • Assay buffer colored pink, containing Tris-buffered saline, Tween 20, protein b. stabilizers and preservatives.
  • C. Tracer IgG - Isoluminol labeled anti-human IgG antibodies in buffer, containing protein stabilizers and preservative.

The QUANTA Flash SS-B Calibrators kit contains two vials of Calibrator 1 and two vials of Calibrator 2:

QUANTA Flash SS-B Calibrators:

  • । QUANTA Flash SS-B Calibrator 1: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human antibodies to SS-B in stabilizers and preservatives.
  • QUANTA Flash SS-B Calibrator 2: Two (2) barcode labeled tubes containing 0.3 mL - prediluted, ready to use reagent. Calibrators contain human antibodies to SS-B in stabilizers and preservatives.

The QUANTA Flash SS-B Controls kit contains two vials of Negative Control and two vials of Positive Control:

QUANTA Flash SS-B Controls:

  • । QUANTA Flash SS-B Negative Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to SS-B in stabilizers and preservatives.
  • QUANTA Flash SS-B Positive Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to SS-B in stabilizers and preservatives.
AI/ML Overview

The document describes the QUANTA Flash® SS-B assay and its performance characteristics. This device is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-SS-B autoantibodies in human serum, used as an aid in diagnosing Sjögren's Syndrome and Systemic Lupus Erythematosus.

Here's an analysis of the acceptance criteria and study data:


1. Table of Acceptance Criteria and Reported Device Performance

This section focuses on the analytical performance characteristics and clinical performance.

Test CategoryAcceptance CriteriaReported Device Performance
PrecisionTotal %CV: < 10%Achieved. For 10 samples with various concentrations, total CV ranged from 5.1% to 8.9% (for samples 110684-04 and 110689-25 respectively), which is within the acceptance range. The majority of samples showed total CV below 8%.
ReproducibilityTotal %CV: < 10%Achieved. For 7 samples tested across different reagent lots, calibrator lots, and operators, total CV ranged from 3.4% to 7.9% (for Sample 2 and Sample A respectively), well within the 10% limit.
Limit of Blank (LoB)LoB determined at 95th percentile, using parametric method if data is normal. Consistent with CLSI EP17-A2 guideline.Achieved. LoB for lot 131009 was 269 RLU, and for lot 141010 was 294 RLU. The final LoB value is 294 RLU. These values are below the analytical measuring range.
Limit of Detection (LoD)LoD based on proportions of false positives (alpha) < 5% and false negatives (beta) < 5%, consistent with CLSI EP17-A2 guideline.Achieved. LoD for lot 131009 was 360 RLU, and for lot 141010 was 398 RLU. The final LoD value is 398 RLU. These values are below the analytical measuring range.
LinearityRecovery between 80-120%, or ± 4 CU, whichever is greater. For linear regression, slope is between 0.9-1.1, and R² is ≥ 0.95.Achieved. All four specimens showed dilution linearity individually, with slopes ranging from 0.95 to 1.04 and R² values of 0.99 for all. The combined data yielded a slope of 0.98 and R² of 0.99. The upper limit of the analytical measuring range was 1550 CU.
Interference85% - 115% recovery, or ± 4 CU difference, whichever is greater.Achieved. No interference detected with bilirubin (95.0% to 108.4% recovery), hemoglobin (96.0% to 105.8%), triglycerides (96.0% to 104.8%), cholesterol (96.0% to 104.8%), and RF IgM (90.7% to 111.5%) within tested concentrations.
Cross-reactivityNo specific numerical acceptance criteria mentioned, but implied expectation is minimal cross-reactivity with other autoantibodies or infection-induced antibodies.Achieved. Out of 273 control samples from various autoimmune diseases and infectious conditions, only 7 (2.6%) showed anti-SS-B positivity, suggesting minimal cross-reactivity. This includes zero positivity in Graves' Disease, Hashimoto Thyroiditis, HCV, HBV, HIV, Syphilis, Primary Antiphospholipid Syndrome, Vasculitis, and Autoimmune myositis cohorts.
Lot to Lot ComparisonWeighted r: ≥0.975 for linear regression. Intercept of the regression line (constant bias): ± 15% of cut-off (3 CU). Slope of the regression line (proportional bias): 0.9-1.1. Weighted Sy/x: ≤ 0.5. Predicted bias (difference) at cut-off: ±15% (3 CU).Achieved. All pair-wise comparisons between three reagent lots (131009 vs 14010, 131009 vs 14011, 141010 vs 141011) met the criteria. Weighted r values were 0.99-1.00, intercept biases were within ±15% of cutoff, and slopes were 1.0. Weighted Sy/x was 0.06-0.11. Predicted bias at cutoff was within limits.
Shelf Life StabilityBeads: Lower 95% CI of regression line > 85% at 2 weeks, no individual data point ≤ 75% recovery at 2 weeks. Controls/Calibrators: Lower 95% CI of regression line ≥ 90% at 2 weeks, no individual data point ≤ 80% recovery at 2 weeks.Achieved. All three lots of beads retained > 85% reactivity. All calibrators and controls maintained > 90% reactivity.
In-use (onboard) StabilityCalibrators: 5 successful calibrations over 8.5 hours, average RLU recovery 90-110%, all controls/patient samples within expected range. Controls: All replicates within established range, linear regression line of %recovery 85-115% at run 15. Reagent Cartridge: Stability claim established when 95% CI of regression line reaches 85% or 115% recovery, or when 2 data points or ≥2% of recovery data is ≤ 75% or ≥ 125%.Achieved. Calibrators: 5 successful calibrations over 8.5 hours, RLU values within 90-110%, controls/patient panel within expected range. Controls: All controls ran within acceptable ranges, regression line remained between 85% and 115% at run 15. Reagent Cartridge: In-use stability for reagent cartridge was set at 57 days based on the criteria.
Real Time StabilityControls: Results fall within acceptable ranges. Calibrators: % recovery of average triplicates between 85-115%, %CV < 10%. Reagent Cartridge: QC panel results fall within respective QC ranges.Achieved for 3, 6, 9, and 12 months. All results were within the acceptance limits for controls, calibrators, and reagent cartridges, supporting a one-year expiration.
Clinical Sensitivity (Sjögren's Syndrome)Not explicitly stated as acceptance criteria, but reported.35.0% (95% CI: 20.6-51.7%)
Clinical Specificity (Sjögren's Syndrome)Not explicitly stated as acceptance criteria, but reported.97.8% (95% CI: 95.9-99.0%)
Clinical Sensitivity (SLE)Not explicitly stated as acceptance criteria, but reported.13.1% (95% CI: 9.4-17.5%)
Clinical Specificity (SLE)Not explicitly stated as acceptance criteria, but reported.98.0% (95% CI: 96.0-99.1%)
Agreement with Predicate Device (All Samples)Not explicitly stated as acceptance criteria, but reported agreement.Positive Agreement = 81.4% (69.1 – 90.3%), Negative Agreement = 98.8% (97.5 – 99.5%), Total Agreement = 97.2% (95.6 – 98.3%).

Study Details:

This document primarily describes the analytical and clinical validation studies for the QUANTA Flash® SS-B assay, Calibrators, and Controls. There is no information about an AI component or expert readers in this specific submission. The device is an immunoassay, not an AI-driven diagnostic. Therefore, some requested information related to AI and human readers is not applicable.

2. Sample size used for the test set and the data provenance:

  • Precision Test Set: 10 samples (quantities for each are not specified beyond the number of replicates) for precision, and 7 samples for reproducibility.
  • LoB/LoD Test Set: 4 blank samples (System Rinse) and 4 low-level samples for LoB/LoD determination. 60 data points were generated for each lot tested.
  • Linearity Test Set: 5 serum samples with various SS-B antibody concentrations.
  • Interference Test Set: 3 specimens (near-cutoff negative, weak positive, high positive).
  • Cross-reactivity Test Set: 273 control samples from patients with various autoimmune diseases and infectious conditions.
  • Clinical Validation Set: A total of 761 characterized samples.
    • Sub-cohort for Clinical Sensitivity/Specificity: Sjögren's Syndrome (n=40), SLE (n=290), Controls (n=431 including various autoimmune/infectious diseases and healthy donors).
    • Sub-cohort for Expected Values: 138 apparently healthy blood donors (118 F, 20 M; ages 17-60).
    • Sub-cohort for Comparison with Predicate Device: 639 samples (from the 761 clinical validation samples) that had predicate ELISA results available. This cohort included Sjögren's syndrome (n=40), SLE (n=240), and relevant disease controls (n=359). No healthy controls were included in this specific comparison cohort.
  • Data Provenance: The document does not explicitly state the country of origin but implies the samples are from human patients/donors. The studies are prospective in the sense that the device was tested on these cohorts to establish performance, however, the samples themselves were "characterized samples" meaning they were retrospectively collected with known diagnoses.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable as the device is an immunoassay. The ground truth (diagnosis of Sjögren's Syndrome or SLE, or specific autoimmune/infectious conditions for control samples) was established clinically rather than by expert review of imaging or similar data. The characterization of the samples is the ground truth.

4. Adjudication method for the test set:

  • Not applicable as the device is an immunoassay measuring anti-SS-B autoantibodies. The "ground truth" for clinical samples was their established clinical diagnosis, not a subjective interpretation requiring adjudication. For analytical studies, internal reference standards and external guidelines (CLSI) were used.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is not an AI-driven device or an imaging diagnostic that involves human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Yes, implicitly. The QUANTA Flash SS-B assay is a diagnostic device that performs the tests and yields results semi-quantitatively. Its performance characteristics (precision, linearity, sensitivity, specificity, etc.) are determined in a standalone mode. Human involvement is in operating the instrument, collecting samples, and interpreting the numerical results in a clinical context, but not in directly influencing the assay's biochemical reaction or signal generation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • Clinical Ground Truth for Clinical Validation: Established clinical diagnoses of Sjögren's Syndrome, Systemic Lupus Erythematosus (SLE), and various other autoimmune/infectious diseases for the control groups. For the reference range and expected values, apparently healthy blood donors were used.
  • Analytical Ground Truth: For linearity, LoB/LoD, precision, and interference studies, the ground truth was based on known concentrations, dilutions, or characteristics of the samples, derived from accepted laboratory methods and standards (e.g., CDC ANA reference sera, in-house standards).

8. The sample size for the training set:

  • The document mentions that the cut-off value was initially established using 187 subjects (162 apparently healthy blood donors, 10 viral hepatitis, 5 HIV, 5 Syphilis, 5 Rheumatoid arthritis patients). Then, "32 samples characterized as positive for SS-B antibodies by IIF, FIA, and ELISA" were used to "aid in the determination of the cutoff." It also states "One reference sample tested positive at this threshold."
  • For the Calibrators and Master Curve Standards, the "target CU is achieved through trial dilutions on small scale" and "Calibrator and Control values are directly traceable to in-house Standards that are used to create the Master Curves for the QUANTA Flash SS-B assay." No specific 'training set' sample size is explicitly defined beyond these descriptions, as the "training" here refers to the development and calibration of the assay itself, rather than training a machine learning model.

9. How the ground truth for the training set was established:

  • For Cut-off Establishment:
    • The "reference population for establishing the reference interval" (187 subjects) was characterized clinically. The distribution of results from these subjects, particularly the 99th percentile (4097 RLU), informed the initial cut-off.
    • Additionally, 32 samples "characterized as positive for SS-B antibodies by IIF, FIA, and ELISA" were used. This indicates that these samples had prior positive results from other established methods, forming a reference "positive" ground truth. The cut-off was adjusted to 12,000 RLU (corresponding to 20 CU) to optimize differentiation based on these known positive samples.
  • For Calibrators and Master Curves: The ground truth for these quantitative assignments is based on "in-house Standards," which are the fundamental reference materials used by the manufacturer to ensure consistency and accuracy of the assay. These standards are likely developed and characterized using highly controlled and validated methods, ensuring traceability where possible (though an international standard for SS-B antibodies is noted as not available). The process involves "trial dilutions," testing on instruments, and "final value assignment."

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three human profiles incorporated into its design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 29, 2015

INOVA DIAGNOSTICS, INC. C/O GABRIELLA LAKOS, MD, PhD DIRECTOR, RHEUMATOLOGY RESEARCH 9900 OLD GROVE ROAD SAN DIEGO, CA 92131

Re: K141210

Trade/Device Name: OUANTA Flash® SS-B QUANTA Flash® SS-B Calibrators QUANTA Flash® SS-B Controls

Regulation Number: 21 CFR § 866.5100 Regulation Name: Antinuclear antibody immunological test system Regulatory Class: Class II Product Codes: Dated: December 29, 2014 Received: December 30, 2014

Dear Dr. Lakos:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{1}------------------------------------------------

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Leonthena R. Carrington -A

Leonthena Carrington, MS, MBA, MT(ASCP) Director (Acting) Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141210

Device Name

QUANTA Flash SS-B, QUANTA Flash SS-B Calibrators, QUANTA Flash SS-B Controls

Indications for Use (Describe)

QUANTA Flash SS-B is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-SS-B autoantibodies in human serum. The presence of anti-SS-B autoantibodies, in conjunction with clinical findings and other laboratory tests is an aid in the diagnosis of Sjögren's Syndrome and Systemic Lupus Erythematosus.

QUANTA Flash SS-B Calibrators are intended for use with the QUANTA Flash SS-B Reagents for the determination of IgG anti-SS-B autoantibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

QUANTA Flash SS-B Controls are intended for use with the QUANTA Flash SS-B reagents for quality control in the determination of IgG anti-SS-B autoantibodies in human serum.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Inova Diagnostics, a Werfen Company. The logo consists of two overlapping shapes, one red and one blue, resembling stylized checkmarks or arrows. To the right of the shapes is the text "Inova Diagnostics" in a bold, gray font, with the tagline "A Werfen Company" in a smaller font below.

510(k) Summary

QUANTA Flash® SS-B QUANTA Flash® SS-B Calibrators QUANTA Flash® SS-B Controls

Table of Contents

Administrative data……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
Device description
Intended use(s)
Substantial equivalence
Comparison to predicate device
Analytical performance characteristics
Value assignment and traceability of Calibrators and Controls
Precision
Reproducibility
Limit of Blank (LoB) and Limit of Detection (LoD)
Analytical Measuring Range (AMR)
Auto-rerun function and reportable results
High concentration hook effect
Linearity
Interference
Cross-reactivity
Lot to lot comparison
Stability
Cut-off, reference range
Clinical performance characteristics
Clinical sensitivity, specificity
Expected values
Comparison with predicate device

{4}------------------------------------------------

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Administrative data

Submitter:INOVA Diagnostics, Inc9900 Old Grove Road,San Diego, CA, 92131
Purpose of submission:New device(s)
Devices in the submission:QUANTA Flash® SS-B
QUANTA Flash® SS-B Calibrators
QUANTA Flash® SS-B Controls
Scientific contact:Gabriella Lakos, Director of Research, Rheumatology
INOVA Diagnostics, Inc
9900 Old Grove Road, San Diego, CA, 92131
Phone: 858-586-9900/1393
Fax: 858-863-0025
email: glakos@inovadx.com
Quality Systems contact:Ronda Elliott, VP, Quality Systems and RA
INOVA Diagnostics, Inc
9900 Old Grove Road, San Diego, CA, 92131
Phone: 858-586-9900/1381
Fax: 858-863-0025
email: relliot@inovadx.com
Preparation date:05/05/2014
Device name (assay kit):Proprietary name:QUANTA Flash® SS-B
Common name:Anti-SS-B Chemiluminescent Immunoassay
Classification name:anti-SS-B antibody, antigen and control
Regulation DescriptionAntinuclear antibody immunological test system
Regulation Medical SpecialtyImmunology
Review PanelImmunology
Product CodeLLL, Extractable antinuclear antibody, antigen and control
Regulation Number866.5100
Device Class2
Device name (Calibrators):Proprietary name:Common name:Classification name:QUANTA Flash® SS-B CalibratorsSS-B CalibratorsCalibrator, secondary
Regulation DescriptionCalibrator
Regulation Medical SpecialtyClinical Chemistry
Product CodeJIT
Regulation Number862.1150
Device Class2
Device name (Controls):Proprietary name:Common name:Classification name:QUANTA Flash® SS-B ControlsSS-B Controlssingle (specified) analyte controls (assayed andunassayed)
Regulation DescriptionQuality control material (assayed and unassayed)
Regulation Medical SpecialtyClinical Chemistry
Product CodeJJX
Regulation Number862.1660
Device Class1 (reserved)
Predicate device:QUANTA Lite® SS-B ELISA, 510(k) number: K922832
Device description

{5}------------------------------------------------

510(k) Summary QUANTA Flash® SS-B

The QUANTA Flash SS-B assay is designed to run on the BIO-FLASH® instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash SS-B assay utilizes a reagent cartridge format, which is compatible with the BIO-FLASH instrument.

Purified recombinant SS-B antigen is coated onto paramagnetic beads. The bead suspension is lyophilized and stored in the bead tube. Prior to use in the BIO-FLASH system, the sealed reagent tubes are pierced with the reagent cartridge lid and the beads are rehydrated and resuspended using

{6}------------------------------------------------

resuspension buffer by pipetting up and down with a transfer pipette. The reagent cartridge is then loaded onto the BIO-FLASH instrument. Samples are also loaded onto the instrument in sample racks. Serum samples are prediluted by the BIO-FLASH with system rinse in a small disposable plastic cuvette. Small amounts of the diluted patient serum, the beads, and assay buffer are all combined into a second cuvette, and mixed. This cuvette is then incubated at 37°C. The beads are magnetized and washed several times. Isoluminol conjugated anti-human IgG antibodies are then added to the cuvette, and again incubated at 37°C. The beads are magnetized and washed repeatedly. The isoluminol conjugate is oxidized when Trigger 1 (Fe(II)coproporphyrin in sodium hydroxide solution) and Trigger 2 (ureahydrogen peroxide in sodium chloride solution) are added to the cuvette, and the flash of light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. The RLU are proportional to the amount of isoluminol conjugate that is bound to the human IgG, which is in turn proportional to the amount of anti-SS-B antibodies bound to the corresponding beads.

For quantitation, the QUANTA Flash SS-B assay utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. Every new lot number of reagent cartridge must be calibrated before first use, with the QUANTA Flash SS-B Calibrators. Based on the results obtained with the two Calibrators included in the Calibrator Set (sold separately), an instrument specific Working Curve is created, which is used to calculate chemiluminescent units (CU) from the instrument signal (RLU) obtained for each sample.

The QUANTA Flash SS-B kit contains the following materials:

One (1) QUANTA Flash SS-B Reagent Cartridge One (1) vial of Resuspension buffer One (1) Transfer pipette

The QUANTA Flash SS-B reagent cartridge contains the following reagents for 50 determinations:

  • a. SS-B antigen coated paramagnetic beads, lyophilized.
  • Assay buffer colored pink, containing Tris-buffered saline, Tween 20, protein b. stabilizers and preservatives.
  • C. Tracer IgG - Isoluminol labeled anti-human IgG antibodies in buffer, containing protein stabilizers and preservative.

The QUANTA Flash SS-B Calibrators kit contains two vials of Calibrator 1 and two vials of Calibrator 2:

QUANTA Flash SS-B Calibrators:

  • । QUANTA Flash SS-B Calibrator 1: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human antibodies to SS-B in stabilizers and preservatives.
  • QUANTA Flash SS-B Calibrator 2: Two (2) barcode labeled tubes containing 0.3 mL -

{7}------------------------------------------------

prediluted, ready to use reagent. Calibrators contain human antibodies to SS-B in stabilizers and preservatives.

The QUANTA Flash SS-B Controls kit contains two vials of Negative Control and two vials of Positive Control:

QUANTA Flash SS-B Controls:

  • । QUANTA Flash SS-B Negative Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to SS-B in stabilizers and preservatives.
  • QUANTA Flash SS-B Positive Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to SS-B in stabilizers and preservatives.

Intended use(s)

QUANTA Flash SS-B is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-SS-B autoantibodies in human serum. The presence of anti-SS-B autoantibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of Sjögren's Syndrome and Systemic Lupus Erythematosus.

QUANTA Flash SS-B Calibrators are intended for use with the QUANTA Flash SS-B Reagents for the determination of IgG anti-SS-B autoantibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

QUANTA Flash SS-B Controls are intended for use with the QUANTA Flash SS-B reagents for quality control in the determination of IgG anti-SS-B autoantibodies in human serum.

Substantial equivalence

The QUANTA Flash SS-B, the QUANTA Flash SS-B Calibrators and the QUANTA Flash SS-B Controls have the same intended use and assay principle as the predicate device.

Comparison to predicate device

Similarities
ltemQUANTA Flash SS-BPredicate Device
Intended useSemi-quantitative determination ofSemi-quantitative detection of anti-SS-
anti-SS-B antibodies in human serumB antibodies in human serum
Assay methodologySolid phase (heterogenous)Solid phase (heterogenous)

QUANTA Flash SS-B reagent kit

{8}------------------------------------------------

immunoassayimmunoassay
TraceabilityInternational Reference Preparationis not availableResults are traceable to in-houseStandardsInternational Reference Preparation isnot available
Sample typeSerumSerum
Shelf lifeOne yearOne year
Differences
ltemQUANTA Flash SS-BPredicate Device
Detection/Chemiluminescent immunoassayEnzyme-linked immunosorbent assay
Operating principle
Solid phase96-well plateParamagnetic microparticles (beads)
AntigenPurified recombinant SS-B antigenNative SS-B antigen, purified from
bovine thymus
Conjugatelsoluminol conjugated anti-humanHRP conjugated anti-human IgG
lgG
CalibrationLot specific Master Curve + twoSS-B ELISA Low Positive
calibrators (Sold separately)(Included in the kit)

QUANTA Flash SS-B Calibrators

ItemQUANTA Flash SS-B CalibratorsPredicate Device
Intended useQUANTA Flash SS-B Calibrators areintended for use with the QUANTAFlash SS-B Reagents for thedetermination of IgG anti-SS-Bautoantibodies in human serum.Each calibrator establishes a point ofreference for the working curve thatis used to calculate unit values.No separate intended use; calibratoris part of the kit.
AnalyteAnti-SS-B antibodiesAnti-SS-B antibodies
MethodQUANTA Flash SS-Bchemiluminescent immunoassayQUANTA Lite SS-B ELISA
MatrixHuman serum, stabilizers, andpreservativeHuman serum, buffer, stabilizers,preservative
UnitCU (Chemiluminescent units)(arbitrary)Units (arbitrary)
Physico-chemicalcharacteristicsLiquid, prediluted, ready to useLiquid, prediluted, ready to use

{9}------------------------------------------------

Storage2-8 °C2-8 °C
Shelf lifeOne yearOne year

QUANTA Flash SS-B Controls

ItemQUANTA Flash SS-B ControlsPredicate Device
Intended useQUANTA Flash SS-B Controls areintended for use with the QUANTAFlash SS-B reagents for qualitycontrol in the determination of IgGanti-SS-B autoantibodies in humanserum.No separate intended use; controlsare part of the kit.
AnalyteAnti-SS-B antibodiesAnti-SS-B antibodies
MethodQUANTA Flash SS-Bchemiluminescent immunoassayQUANTA Lite SS-B ELISA
MatrixHuman serum, stabilizers, andpreservativeHuman serum, buffer, stabilizers,preservative
UnitCU (Chemiluminescent units)(arbitrary)Units (arbitrary)
Physico-chemicalcharacteristicsLiquid, ready to useLiquid, prediluted, ready to use
Levels2 (negative and positive)2 (ELISA negative, high positive)
Storage2-8 °C2-8 °C
Shelf lifeOne yearOne year

Analytical performance characteristics

Value assignment and traceability of Calibrators and Controls

There is currently no recognized international standard for the measurement of SS-B antibodies. The CDC ANA reference sera #2 (REFERENCE SERUM FOR HUMAN ANTIBODIES TO SS-B/La) and #3 (REFERENCE SERUM, FLUORESCENCE ANTINUCLEAR ANTIBODY, SPECKLED PATTERN) were tested for SS-B and produced the following results: CDC ANA #2: >1550.0 CU CDC ANA #3: 284.1 CU

The QUANTA Flash SS-B Calibrators and Controls are manufactured by diluting human serum that contains high titer of antibodies with commercial antibody stabilizer, containing preservative. The human serum is obtained from commercial sources and it is tested for markers of infectious substances.

{10}------------------------------------------------

510(k) Summary QUANTA Flash® SS-B

The target CU is achieved through trial dilutions on small scale. Once a dilution is selected, the Calibrators and Control are bulked, tested, and adjusted. Upon completion of the manufacturing process, the Calibrators and Controls are tested on at least two instruments, on at least two lots of reagent cartridge, in replicates of 10 to determine final value assignment.

Calibrator and Control values are directly traceable to in-house Standards that are used to create the Master Curves for the QUANTA Flash SS-B assay.

MaterialAssigned Value
SS-B Master Curve Standard 13.3 CU
SS-B Master Curve Standard 215.8 CU
SS-B Master Curve Standard 398.8 CU
SS-B Master Curve Standard 4682.1 CU
SS-B Master Curve Standard 51241.6 CU
SS-B Master Curve Standard 61706.8 CU

List of SS-B Standards, Calibrators and Controls:

MaterialManufacturingTarget ValueManufacturingTarget Range
SS-B Calibrator 116 CU14-18 CU
SS-B Calibrator 2600 CU540-660 CU
SS-B Negative Control10 CU8-12 CU
SS-B Positive Control50 CU40-60 CU

Precision

The precision of the QUANTA Flash SS-B assay was evaluated first on 10 samples containing various concentrations of SS-B antibodies in accordance with CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Procedures - Approved Guideline: samples were run in duplicates, twice a day, for 21 days. Production reagent lot 131009 was used for the studies.

Data were analyzed with the Analyse-it for Excel method evaluation software, and within run, between run, between day and total precision were calculated.

Acceptance criteria: Total %CV: < 10%

Results are summarized in the Table below.

QUANTA Flash SS-BWithin RunBetweenRunsBetweenDaysTotal
Sample IDMean(CU)Number ofreplicatesSD(CU)CV (%)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV (%)
120814-7012.4840.43.50.00.00.64.60.75.8

{11}------------------------------------------------

510(k) Summary QUANTA Flash® SS-B

110689-2524.3841.45.80.00.01.66.72.28.9
110687-12025.0841.24.70.31.10.83.21.55.8
110686-4027.9841.03.60.20.71.03.61.45.1
110684-20132.7847.65.80.00.06.14.69.87.4
120814-02383.58414.23.77.92.116.54.323.26.0
110684-04552.38415.72.818.33.314.72.728.25.1
110687-02883.38440.34.630.23.424.32.755.96.3
000674-121356.88492.46.80.00.060.64.5110.58.1
000674-101539.68499.26.436.92.434.52.2111.37.2

Reproducibility

Seven samples containing various concentrations of SS-B antibodies were tested on two different reagent lots, using two different lots of Calibrators, by two operators. Samples were run in quadruplicates, two times a day, for 10 days, to generate 80 data points.

Production reagent lots 131009 and 141010, and Calibrator and Control lots 131006 and 131007 were used for the studies.

Data were analyzed with the Analyse-it for Excel method evaluation software, and within run, between reagent lots, between calibrator lots, between operators and total precision were calculated.

Acceptance criteria: Total %CV: < 10%

Results are summarized in the Table below.

Within RunBetween Reagent LotsBetweenCalibratorLotsBetweenOperatorsTotal
SampleIDMean(CU)NumberofreplicatesSD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)
111.7800.32.80.54.60.32.80.43.00.53.9%
226.1800.72.60.62.11.03.60.93.30.93.4%
3124.2802.82.32.82.37.15.74.33.55.34.3%
B252.7805.12.015.96.313.15.29.63.813.45.3
A640.88023.93.750.47.950.97.943.86.850.47.9
D981.98035.33.656.45.778.48.057.25.867.96.9
C1388.78054.63.989.86.5111.78.0103.67.5106.97.7

Limit of Blank (LoB) and Limit of Detection (LoD)

{12}------------------------------------------------

The LoD of the QUANTA Flash SS-B assay is 398 RLU, which is below the analytical measuring range of the assay. It was determined consistent with CLSI EP17-A2 guideline with proportions of false positives (alpha) less than 5% and false negatives (beta) less than 5%; based on 120 determinations, with 60 measurements on blank samples and 60 measurements of low level samples.

For determining the LoB, 4 blank samples (System Rinse) from two different lots were run in replicates of five on two reagent lots, once per day, for 3 days. Production reagent lots 131009 and 141010 were used for the studies. 60 data points were generated on each lot.

The LoB was determined on each lot separately with the Analyse-it for Excel software's Reference Interval function, at the 95° percentile, using the parametric method, as the dataset showed normal distribution. The LoB for lot 131009 was determined as 269 RLU, and for lot 141010 as 294 RLU. The final LoB value is 294 RLU.

For determining the LoD, 4 low level samples (prepared by diluting anti-SS-B positive samples with System Rinse) were run in replicates of five on two reagent lots, once per day, for 3 days. Production reagent lots 131009 and 141010 were used for the studies. 60 data points were generated on each lot. The LoD was determined separately on each lot according to CLSI EP17-A2 guideline. The limit of detection for lot 131009 was determined as 360 RLU, and for lot 141010 as 398 RLU. The final LoD value is 398 RLU.

These values are below the value of the lowest QUANTA Flash SSB Master Curve standard, i.e. below the Analytical Measuring Range.

Analytical Measuring Range (AMR)

QUANTA Flash SS-B: 3. 3 CU - 1550.0 CU

Auto-rerun function and reportable results

The BIO-FLASH software has an Auto-rerun option available. If this option is selected, the instrument will automatically rerun any sample that has a result of >1550.0 by further diluting it by 10 fold, thereby bringing the measured value within the AMR. The final result will be calculated by the software by taking into account the additional dilution factor. As the highest value that can be measured is 1550.0, the highest value that can be reported is 15,500.

High concentration hook effect

To assess hook effect, measurement signal (relative light units, RLU) was examined for three high positive samples, with results above the AMR, before and after automatic or manual dilution. All sera produced significantly higher RLU values (above the AMR) when used "as is" compared to the manually or automatically diluted ones (that were within the AMR), thereby confirming that high positive specimens above the analytical measuring range do not show hook effect up to 5,004 CU in the SS-B assay (the highest concentration that was tested).

{13}------------------------------------------------

Linearity

The linearity of the AMR was evaluated by a study according to CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approved Guideline. Five serum samples with various SS-B antibody concentrations were diluted in 10% increments (from 0% to 90% negative base matrix) to obtain values that cover the AMR. The dilutions were assayed in duplicates. Percent recovery of obtained mean results was calculated compared to the expected mean results (based on the dilution factor). Moreover, obtained values were plotted against expected values, and linear regression analysis was performed.

Acceptance criteria:

  • Recovery is between 80-120%, or ± 4 CU, whichever is greater.

  • For linear regression analysis, slope is between 0.9-1.1, and R-is ≥ 0.95.

All four specimens showed dilution linearity individually.

SampleTest Range (CU)Slope (95% CI)R2
1204.4 - 1507.01.03 (0.98 to 1.08)0.99
2155.3 - 1546.81.04 (0.99 to 1.09)0.98
399.0 - 1048.60.96 (0.92 to 1.00)0.99
417.4 - 163.81.01 (0.97 to 1.05)0.99
53.6 - 29.40.95 (0.91 to 0.99)0.99

The combined data yielded the following results with linear regression:

SampleTest Range (CU)Slope (95% CI)R2
All3.6 - 1546.80.98 (0.96 to 1.00)0.99

The upper limit of the analytical measuring range was limited to 1550 CU.

Interference

The interference study was performed according to CLSI EPO7-A2, Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition. Three specimens were tested (near-the-cutoff negative: 15.4 CU; weak positive: 25.5 CU; high positive: 200.3). Interfering substances were spiked into every specimen at three different concentrations in 10% of total specimen volume, and the resulting samples were assessed in triplicates with the SS-B assay. Recovery of the unit values was calculated compared to control samples spiked with the same volume of diluent (10% of total).

{14}------------------------------------------------

Acceptance criteria: 85% - 115% recovery, or ± 4 CU difference, whichever is greater.

No interference was detected with bilirubin up to 10 mg/dL (recovery: 95.0% to 108.4%), hemoglobin up to 200 mg/dL (recovery: 96.0% to 105.8%), triglycerides up to 1000 mg/dL (recovery: 96.0% to 104.8%), cholesterol up to 224.3 mg/dL (recovery: 96.0% to 104.8%), and RF IgM up to 500 IU/mL (recovery: 90.7% to 111.5%).

Cross-reactivity

To test potential cross-reactivity with autoantibodies and infection-induced antibodies, results obtained on altogether 273 of the total 431 control samples that were included in the clinical validation study were assessed. These samples were from patients with autoimmune diseases that are characterized with disease specific autoantibodies, or from patients with positive infectious disease serology. The composition of the cohort and the anti-SS-B positivity rate is shown in the Table below:

DiagnosisNumber of samples# pos% pos
Graves' Disease1900.0%
Hashimoto Thyroiditis2100.0%
HCV1000.0%
HBV1000.0%
HIV500.0%
Syphilis500.0%
Primary Antiphospholipid Syndrome1500.0%
Vasculitis100.0%
Systemic sclerosis8911.1%
Autoimmune myositis400.0%
Rheumatoid arthritis7045.7%
Autoimmune liver disease group#12150.0%
Autoimmune liver disease group#2**2214.5%
Total controls27372.6%

** Samples contain autoimmune liver disease specific antibodies (SLA, F-actin, M2)

Based on the results, the QUANTA Flash SS-B assay does not show cross-reactivity with autoantibodies that are present in various autoimmune diseases, or antibodies against infectious agents.

Lot to lot comparison

Twenty-two unique samples and the Positive Controls (altogether 24 specimens) with various reactivity levels were tested in triplicates with three different reagent lots: 131009, 14010 and

{15}------------------------------------------------

  1. The samples covered the total analytical measuring range of the assay. Results were processed by linear regression analysis and bias calculation according to CLSI EP09-A2, Method Comparison and Bias Calculation Using Patient Samples; Approved Guideline - Second Edition.

Pair-wise comparisons were performed between lot 131009 vs 14010, lot 131009 vs 14011 and lot 141010 vs 141011, considering individual replicates instead of the mean of replicates.

Acceptance criteria and results are in the Table below. All results were within the acceptance limits.

131009 vs14010131009 vs14011141010 vs141011
Acceptance criteria
Weighted r: ≥0.975 for linear regression1.000.990.99
Intercept of the regression line (constant bias):0.5-0.6-1.0
± 15% of cut-off (3 CU)
Slope of the regression line (proportional bias): 0.9-1.01.01.0
1.1
Weighted S y/x: ≤ 0.50.060.110.11
Predicted bias (difference) at cut-off: ±15 (3 CU)0.4-0.6-1.0

Stability

Shelf life

To establish the initial claim for shelf life, accelerated stability studies were performed for 4 weeks at 37°C ± 3°C, where one week is equal to six months at 5 ± 3°C.

Accelerated stability testing was performed on each of the following sealed components of the QUANTA Flash SS-B to establish initial stability claim: the beads, the two Calibrators, and the negative and positive Controls. Each week a new sealed component was placed in the incubator, and all components were tested at the end of the experiment together with the one that was stored at 5 ± 3℃. The recovery of the measured values was calculated for each time point (compared to those obtained with 5 ± 3°C stored reagent). All calculations were performed by comparing results of sealed components stored at 5 ± 3℃ (control) to those stored at 37 ± 3℃ (test) for 1, 2, 3, and 4 weeks, where one week is equal to six months at 5 ± 3°C. Linear regression analysis was performed between recovery values and the number of days.

Acceptance criteria for one year preliminary expiration dating:

  • Beads:

With regression analysis, the lower 95% Cl interval of the regression line is > 85% at 2 weeks, and no individual data point has ≤ 75% recovery at 2 weeks.

  • Controls and Calibrators:

With regression analysis, the lower 95% Cl interval of the regression line is ≥ 90% at 2 weeks, and no individual data point has ≤ 80% recovery at 2 weeks.

{16}------------------------------------------------

Beads

Testing was performed on three lots of SS-B coupled beads using up to 7 characterized samples with various reactivity levels.

All three lots of beads retained > 85% reactivity (considering the 95% Cl) after two weeks at 37 ± 3ºC, and therefore pass the acceptance criteria for one year expiration date.

Calibrators and Controls

Testing was performed on three lots of SS-B Calibrators and Controls. All Calibrators and Controls maintained > 90% reactivity (considering the 95% Cl) when sored at 37 + 3ºC for 2 weeks, and therefore pass the acceptance criteria for one year expiration dating.

In-use (onboard) stability

Calibrators

Onboard stability claim: 4 calibrations, or 8 hours onboard

During assessing on-board stability, Calibrators were placed uncapped, onboard the instrument, and calibration was performed altogether five times over 8.5 hours. Controls and a panel of characterized patient specimens were run on each calibration curve.

Calibrators are considered stable if all five calibrations performed in the 8.5 hour period are successful, and average Calibrator RLU recovery values are between 90% and 110% compared to the first use. A total of 5 successful calibrations were performed over an 8.5 hour period. Calibrator RLU values remained within the 90-110% range. Moreover, all Controls and patient panel samples ran within their expected range. This supports the claim that calibrators can be used for up to 4 calibrations over an 8 hour period.

Controls

Onboard stability claim: up to 15 uses, at 10 minutes onboard per use

During assessing on-board stability, 2 vials of each Control were assayed twice a day for a total of 21 runs. The first run was used to establish baseline value, and then additional 20 runs were performed. During runs, the Controls were left uncapped, onboard the instrument for 15 minutes per run. When not in use, the controls were capped, and stored at 5º ± 3°.

Controls are considered stable when all replicates run within their established range, and the linear regression line obtained by plotting %recovery values against the number of runs stays between 85% and 115% at run 15.

All controls ran within their respective acceptable ranges for all runs. Moreover, the regression line remained between 85% and 115% at run 15 for both Controls. These results support the claim that controls can be used for up to 15 times, at 10 minutes per use.

Reagent Cartridge

To establish the in-use stability of the QUANTA Flash SS-B reagent cartridge, three lots of cartridges were tested with up to 6 serum specimens (with different reactivity levels) along with the Negative and

{17}------------------------------------------------

Positive Controls. The specimens were tested periodically up to 69 days. Percent recoveries were calculated compared to the day zero average values, and linear regression analysis was performed by plotting %recovery against the number of days. The claim was established using the following criteria (using the one that is fulfilled first):

  • The stability claim is established at the actual measurement day proceeding the day when the 95% confidence interval of the regression line reaches 85% or 115% recovery, or

  • At the actual measurement day preceding the day when 2 data points or ≥2% of the recovery data (whichever is greater) is ≤ 75% or ≥ 125% recovery.

The onboard stability results of the three lots are as follows:

RP0008: 68 days

121006: 57 days

131009: 60 days

Using these criteria, the in-use (onboard) stability of SS-B reagent cartridge was set at 57 days.

Real time stability

Real time stability testing was performed at 3, 6, 9 and 12 months on Calibrators, Controls and reagent cartridge to support the one year expiration.

For Controls, each control was tested in triplicates at each time point.

  • Acceptance criteria: results should fall within their acceptable ranges as it was established at the release of the controls.

Calibrators were tested in triplicates at each time point as it is done during calibration. Averages of the triplicates were compared to the value that was assigned to the Calibrators at release.

  • Acceptance criteria: % recovery of the average of the triplicates is between 85 and 115%, and %CV of the triplicates is < 10%

For reagent cartridge, the panel of QC samples was tested at each time point. This QC panel is used by the QC Department for reagent release and QC.

  • Acceptance criteria: results should fall within their respective QC ranges.

All results were within the acceptance limits.

Cut-off, reference range

QUANTA Flash SS-SB:Negative<20 CU
Positive>20 CU

The reference population for establishing the reference interval for the SS-B assay consisted of 187 subjects:

{18}------------------------------------------------

Apparently healthy blood donors162
Viral hepatitis positive samples10
HIV positive samples5
Syphilis positive samples5
Rheumatoid arthritis patients5

All specimens were the same matrix (serum) as specified in the Intended Use. All specimens were unaltered. The cut off was established in accordance with CLSI C28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition. The Analyseit for Excel software was used to make the calculations. The distribution of the results was non-normal (Saphiro-Wilk p<0.0001), so the non-parametric percentile method was used. One sample from a blood donor was excluded, as it was considered outlier. The RLU value of this sample was 138,062, and it was also positive with the predicate ELISA. The 99th percentile Reference Interval was calculated as 4097 RLU. One reference sample tested positive at this threshold.

Additionally, 32 samples characterized as positive for SS-B antibodies by IIF, FIA, and ELISA were assayed to aid in the determination of the cutoff. Based on the distribution of RLU values in these (known) positive samples (37,743-942,882 RLU), the cutoff was increased to 12,000 RLU to ensure optimal differentiation between negatives and positives, and 20 CU value was assigned to this RLU value.

Clinical performance characteristics

Clinical sensitivity, specificity

A cohort of characterized samples, none of which were used for establishing the reference range, was used to validate the clinical performance of the QUANTA Flash SS-B. A total of 761 characterized samples were included in the Validation Set for the QUANTA Flash SS-B. All samples were run on the QUANTA Flash SS-B. The distribution of the cohort and the SS-B positivity rate is in the Table below:

Patient groupNNumber positive% positive
Ulcerative colitis2000.0%
Graves' Disease1900.0%
Hashimoto Thyroiditis2100.0%
Non-autoimmune thyroid disease4300.0%
Crohn's disease2000.0%
HCV1000.0%
HBV1000.0%
HIV500.0%
Syphilis500.0%
Osteoarthritis2015.0%

{19}------------------------------------------------

Primary Antiphospholipid Syndrome1500.0%
Secondary Antiphospholipid Syndrome*1500.0%
Other rheumatic diseases4012.5%
Vasculitis100.0%
Systemic sclerosis8911.1%
Autoimmune myositis400.0%
Rheumatoid arthritis7045.7%
Autoimmune liver disease group#12150.0%
Autoimmune liver disease group#2**2214.5%
Sjögren's Syndrome401435.0%
SLE2903813.1%
Total761
Total controls43192.1%

*Patients may have SLE

** Samples contain autoimmune liver disease specific antibodies (SLA, F-actin, M2)

The results were analyzed to calculate sensitivity and specificity for SLE (n=290) and Sjögren's syndrome (SS) (n=40) separately, and SLE and SS combined.

Clinical sensitivity and specificity of the QUANTA Flash SS-B in Sjögren's syndrome

n=456DiagnosisAnalysis(95% confidence)
SSControls(excludingSLE)Total
QUANTA FlashSS-BPositive14923Sensitivity = 35.0% (20.6-51.7%)
Negative26407433Specificity = 97.8% (95.9-99.0%)
Total40416456

Clinical sensitivity and specificity of the QUANTA Flash SS-B in SLE

n=706Diagnosis
SLEControls(excludingSjögren's)TotalAnalysis(95% confidence)
QUANTA FlashSS-BPositive38947Sensitivity = 13.1% (9.4-17.5%)
Negative252407659Specificity = 98.0% (96.0-99.1%)
Total290416706

Clinical sensitivity and specificity of the QUANTA Flash SS-B in SLE+SS

{20}------------------------------------------------

SS orSLEControlsTotal
QUANTA FlashSS-BPositive52961Sensitivity = 15.8% (12.0-20.1%)
Negative278407685Specificity = 97.8% (95.9-99.0%)
Total330416746

Expected values

The expected value in the normal population is "negative". Anti-SS-B autoantibody levels were analyzed in a cohort of 138 apparently healthy blood donors (118 females and 20 males, ages 17 to 60 years, with an average age of 32.8 years and median age of 31 years) using the QUANTA Flash SS-B. This patient population was different from the one that was used to establish the cutoff, and was only used to assess expected values. With the cut-off of 20 CU, 1 (0.7 %) of the samples was positive on the QUANTA Flash SS-B. The mean concentration was < 3.3 CU, and the values ranged from <3.3 to 28.7 CU.

Comparison with predicate device

Samples for method comparison analysis included 639 samples from the total of 761 samples that were used in the clinical validation studies, and had predicate ELISA results available. The cohort consisted of Sjögren's syndrome (n=40) and SLE patients (n=240) and relevant disease controls (359). No healthy controls were included in this cohort.

Patient groupNNumber positive% positive
Ulcerative colitis2000%
Graves' Disease1900%
Hashimoto Thyroiditis2100%
Non-autoimmune thyroid disease4300%
Crohn's disease2000%
HCV1000%
HBV1000%
HIV500%
Syphilis500%
Osteoarthritis20210%
Primary Antiphospholipid Syndrome1500%
Secondary Antiphospholipid Syndrome1517%
Other rheumatic diseases and arthropathies4000%
Vasculitis100%

Results obtained with the predicate device are shown in the Table below:

{21}------------------------------------------------

Patient groupNNumber positive% positive
Systemic sclerosis41615%
Autoimmune myositis200%
Rheumatoid arthritis7011%
Autoimmune liver disease group 12150%
Autoimmune liver disease group 2**0n/an/a
Sjogren's Syndrome401538%
SLE2403314%

** Samples contain autoimmune liver disease specific antibodies (SLA, F-actin, M2). For distribution, please see table in Section #9.

Agreement between the QUANTA Flash SS-B and the predicate ELISA was calculated three ways:

Method comparison, all samples

Method Comparison (N=639)SS-B ELISAPercent Agreement(95% Confidence)
QUANTA FlashSS-BNegative57311584Pos. Agree = 81.4% (69.1 – 90.3%)
Positive74855Neg. Agree = 98.8% (97.5 – 99.5%)
Total58059639Total Agree = 97.2% (95.6 – 98.3%)

Method comparison, data within reportable range

Method Comparison (N=142)SS-B ELISAPercent Agreement(95% Confidence)
NegativePositiveTotal
QUANTA FlashSS-BNegative81687Pos. Agree = 88.9% (77.4 – 95.8%)
Positive74855Neg. Agree = 92.0% (84.3 – 96.7%)
Total8854142Total Agree = 90.8% (84.9-95.0%)

Method comparison, data around the cutoff

Method Comparison (N=10)SS-B ELISAPercent Agreement(95% Confidence)
NegativePositiveTotal
QUANTA Flash®SS-BNegative527Pos. Agree = 33.3% (0.8 – 90.6%)
Positive213Neg. Agree = 71.4% (29.0 – 96.3%)
Total7310Total Agree = 60.0% (26.2 – 87.8%)

§ 866.5100 Antinuclear antibody immunological test system.

(a)
Identification. An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).(b)
Classification. Class II (performance standards).