AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For VIDAS H. pylori IgG:
VIDAS® H. pylori IgG (HPY) is an automated qualitative test for use on the instruments of the VIDAS family, for the detection of anti-Helicobacter pylori IgG antibodies in human serum or plasma (EDTA) using the ELFA technique (Enzyme Linked Fluorescent Assay). The VIDAS HPY assay is intended as an aid in diagnosis of H. pylori infection in an adult symptomatic population.
This device is an in vitro diagnostic medical device for professional use only.

For VIDAS 3:
The VIDAS 3 system is a complete standalone immunodiagnostic system intended for trained and qualified laboratory technicians (daily routine use) and laboratory administrators (application configuration). This device is an in vitro diagnostic medical device for professional use only.

For VIDAS Lyme IgG II:
The VIDAS Lyme IgG II (LYG) assay is an automated qualitative enzyme immunoassay intended for use on the instruments of the VIDAS family in the presumptive detection of human IgG antibodies to Borrelia burgdorferi in human serum (plain or separation gel) or plasma (sodium heparin). It should be used to test patients with a history and/or symptoms of infection with B. burgdorferi. All VIDAS Lyme IgG II positive specimens should be further tested with a Western Blot IgG assay to obtain supportive evidence of infection with B. burgdorfei. This device is an in vitro diagnostic medical device for professional use only.

For VIDAS RUB IgG:
The VIDAS® RUB IgG (RBG) assay uses Enzyme Linked Fluorescent Assay (ELFA) technology on the instruments of the VIDAS family for the in vitro quantitative measurement of IgG antibodies to rubella virus in human serum. The VIDAS RUB IgG (RBG) assay is intended as an aid in the determination of immune status to rubella. The performance of this device has not been established for screening of cord blood, or for neonatal samples. Likewise, performance characteristics of the assay have not been established for immunocompromised or immunosuppressed individuals.
This device is an in vitro diagnostic medical device for professional use only.

For VIDAS TOXO IgM:
The VIDAS® TOXO IgM (TXM) assay is intended for use on the instruments of the VIDAS family (VITEK ImmunoDiagnostic Assay System) as an automated enzyme-linked fluorescent immunoassay (ELF A) for the presumptive qualitative detection of anti-Toxoplasma gondii IgM antibodies in human serum, as an aid in the diagnosis of acute, recent, or reactivated Toxoplasma gondii infection. This assay must be performed in conjunction with an anti-Toxoplasma gondii lgG antibody assay. VIDAS TOXO IgM (TXM) assay performance has not been established for prenatal screening or newborn testing. This assay has not been cleared by the FDA for blood/plasma donor screening. This device is an in vitro diagnostic medical device for professional use only.

For VIDAS Human Chorionic Gonadotropin:
The VIDAS® HCG (HCG) assay is intended for use on the instruments of the VIDAS family as an automated quantitative enzyme linked fluorescent immunoassay (ELFA) for the determination of human Chorionic Gonadotropin (hCG) concentration in human serum or plasma. The VIDAS HCG (HCG) assay is intended to aid in the early detection of pregnancy.
This device is an in vitro diagnostic medical device for professional use only.

For VIDAS T4:
The VIDAS® T4 (T4) assay is intended for use on the instruments of the VIDAS family as an automated quantitative enzyme-linked fluorescent immunoassay for the determination of human thyroxine (T4) concentration in serum or plasma (heparin). It is intended for use as an aid in the diagnosis and treatment of thyroid disorders. This device is an in vitro diagnostic medical device for professional use only.

For VIDAS Testosterone:
The VIDAS Testosterone (TES) assay is an automated quantitative test for use on the instruments of the VIDAS family for the enzyme immunoassay measure of total testosterone in human serum or plasma (lithium heparin), using the ELFA technique (Enzyme Linked Fluorescent Assay). It is intended as an aid in the diagnosis and management of conditions involving excess or deficiency of this androgen.
This device is an in vitro diagnostic medical device for professional use only.

For VIDAS TSH:
The VIDAS® TSH (TSH) assay is intended for use on the instruments of the VIDAS family as an automated quantitative enzyne-linked fluorescent immunoassay (ELFA) for the determination of human thyroid stimulating hormone- (TSH) concentration in human serum or plasma (heparin). It is intended for use as an aid in the diagnosis of thyroid or pituitary disorders.
This device is an in vitro diagnostic medical device for professional use only.

For VIDAS D-Dimer Exclusion II:
VIDAS® D-Dimer Exclusion II™ is an automated quantitative test for use on the instruments of the VIDAS family for the immunoenzymatic determination of fibrin degradation products (FbDP) in human plasma (sodium citrate, CTAD) using the ELFA technique (Enzyme Linked Fluorescent Assay).
VIDAS D-Dimer Exclusion II is indicated for use in conjunction with a clinical pretest probability assessment model to exclude deep vein thrombosis (DVT) and pulmonary embolism (PE) disease in outpatients suspected of DVT or PE. This device is an in vitro diagnostic medical device for professional use only.

Device Description

The VIDAS® 3 instrument is an automated multiparametric immunoassay system, which uses ELFA (Enzyme Linked Fluorescent Assay) technology. The VIDAS 3 system offers primary tube sampling, automated sample dilution, reagent/sample detection and reagent traceability.
The technology used, which is adaptable to a wide range of assays, combines the EIA method with a final fluorescence reading: this technology is known as ELFA (Enzyme Linked Fluorescent Assay). The enzyme used in the VIDAS product range is alkaline phosphatase, which catalyzes the hydrolysis of the substrate 4-methyl umbelliferyl phosphate (4-MUP) into a fluorescent product 4-methyl umbelliferone (4-MU) the fluorescence of which is measured at 450nm. The immunological methods are either indirect ElA, immunocapture, sandwich or competition, all involving a conjugate using the alkaline phosphatase.

AI/ML Overview

This document describes the performance data for several VIDAS assays when used on the VIDAS 3 instrument, comparing them to their performance on the predicate VIDAS instrument. The tests are primarily for establishing substantial equivalence for the new VIDAS 3 instrument and do not typically include detailed acceptance criteria for the assays themselves, which are already established for the predicate devices. The studies focus on method comparison, precision, linearity, and detection limits.

Here's a breakdown of the requested information based on the provided text, focusing on the VIDAS H. pylori IgG assay as a primary example, and generalizing for others where appropriate:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state acceptance criteria in a quantitative format for method comparison. Instead, it demonstrates "correlation" and "equivalency" between the new device (VIDAS 3) and the predicate device (VIDAS). For precision, specific CV% ranges are reported.

Here's an example for the VIDAS H. pylori IgG assay's method comparison:

Performance MetricAcceptance Criteria (Implied - "Correlation/Equivalency")Reported Device Performance (VIDAS H. pylori IgG on VIDAS 3 vs. VIDAS)
Method Comparison (Qualitative)
Overall Agreement(Implied: High agreement with predicate)High, as demonstrated by contingency table and % agreements
Positive Agreement(Implied: High agreement for positive results)100% [96.9 ; 100.0] %
Negative Agreement(Implied: High agreement for negative results)96.6% [91.5 ; 98.7] %
Precision(Implied: Acceptable and reproducible CV%)
Within-Run CV%(Specific ranges not explicitly stated as criteria, but reported)Sample 1: 7.7%, Sample 2: 6.2%, Sample 3: 5.2%
Total Between-Calibration CV%Sample 1: 9.8%, Sample 2: 7.2%, Sample 3: 6.4%
Total Between-Instrument CV%Sample 1: 10.1%, Sample 2: 7.2%, Sample 3: 6.8%

Note: For quantitative assays like VIDAS RUB IgG, VIDAS HCG, VIDAS T4, VIDAS Testosterone, VIDAS TSH, and VIDAS D-Dimer Exclusion II, method comparison relies on slope, intercept, and correlation coefficient, implying acceptance criteria for these values (e.g., slope close to 1, intercept close to 0, high correlation coefficient). Precision for these assays also includes CV% for various components.

2. Sample Size Used for the Test Set and Data Provenance

  • VIDAS H. pylori IgG:

    • Test Set Size: 250 serum samples (positive, equivocal, and negative).
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). The study compares performance between two instruments, implying samples are run on both.
  • VIDAS Lyme IgG II:

    • Test Set Size: 220 serum samples (positive and negative).
    • Data Provenance: Not explicitly stated.
  • VIDAS RUB IgG:

    • Test Set Size (Quantitative Method Comparison): 112 serum samples (ranging from 0 to 225 IU/mL).
    • Test Set Size (Qualitative Method Comparison): 220 serum samples (positive, equivocal, and negative).
    • Test Set Size (CDC Reference Panel): 100 specimens (50 pairs of sera).
    • Data Provenance: Not explicitly stated for general samples. The CDC panel implies a curated and standardized set.
  • VIDAS TOXO IgM:

    • Test Set Size: 198 serum samples.
    • Data Provenance: Not explicitly stated.
  • VIDAS Human Chorionic Gonadotropin (hCG):

    • Test Set Size: 113 serum samples.
    • Data Provenance: Not explicitly stated.
  • VIDAS T4:

    • Test Set Size: 105 serum samples.
    • Data Provenance: Not explicitly stated.
  • VIDAS Testosterone:

    • Test Set Size: 172 serum samples.
    • Data Provenance: Not explicitly stated.
  • VIDAS TSH:

    • Test Set Size: 179 serum samples.
    • Data Provenance: Not explicitly stated.
  • VIDAS D-Dimer Exclusion II:

    • Test Set Size: 219 plasma samples.
    • Data Provenance: Not explicitly stated.

Across all assays, the studies are described as "Method Comparison" and "Precision" studies, which are typically retrospective analyses of patient samples to compare device performance to an established method.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. For in vitro diagnostic devices, ground truth is typically established by comparative methods (e.g., predicate device, reference methods, clinical diagnosis, or other laboratory gold standards) rather than expert consensus on individual cases. The document states that performance was evaluated against the predicate device (e.g., "VIDAS H. pylori IgG assay on the VIDAS 3 to the VIDAS H. pylori IgG assay on the VIDAS"). The "ground truth" for these studies is the result obtained from the predicate VIDAS instrument using its established methodology.

For the VIDAS RUB IgG, a "CDC reference panel" and "CDC low-titer rubella antibody standard" are mentioned, where the reference panel sera were "titered by Hemagglutination Inhibition." This implies that the ground truth for this specific part of the study was established by a recognized reference method (Hemagglutination Inhibition) and certified reference materials from the CDC.

4. Adjudication Method for the Test Set

This information is not explicitly provided. For method comparison studies, typically, discordant results between the new device and the predicate device (or reference method) are investigated. However, the exact adjudication process (e.g., by a third, more definitive test, or expert review of patient clinical history) is not detailed. The phrase "results were evaluated according to CLSI EP12-A2" or CLSI EP9 suggests standard statistical methods for agreement or correlation, which do not necessarily involve expert adjudication of individual discrepancies beyond reporting them.

For quantitative assays where method comparison statistics (slope, intercept, correlation coefficient) are used, "outliers" were removed in some cases (e.g., VIDAS RUB IgG quantitative comparison), implying some form of review or statistical exclusion, but not necessarily expert "adjudication" in the sense of clinical decision-making.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No. This document describes performance studies for in vitro diagnostic instruments and assays, not imaging or similar devices that would typically involve human readers interpreting results. Therefore, an MRMC comparative effectiveness study, which assesses improvements in human interpretation with AI assistance, is not applicable and was not performed.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The studies described are for the standalone in vitro diagnostic instruments and their associated assays. These are standalone tests, meaning the algorithm (or assay chemistry in this case) processes the sample and provides a result without direct human interpretation of raw data for diagnosis. The "human-in-the-loop" here refers to trained laboratory technicians operating the instrument and interpreting the final quantitative or qualitative results according to established cut-offs/guidelines, rather than interpreting complex images or signals. The purpose of these studies is to confirm that the new instrument (VIDAS 3) produces equivalent results to the predicate instrument (VIDAS) for these assays.

7. The Type of Ground Truth Used

The primary type of "ground truth" used in these studies is the results obtained from the predicate device (VIDAS instrument) for the same assays. The goal is to demonstrate "substantial equivalence" of the new instrument (VIDAS 3) to the predicate.

For the VIDAS RUB IgG assay, a CDC reference panel where samples were "titered by Hemagglutination Inhibition" served as an additional, external reference for ground truth in a specific subset of testing. This is a form of reference method/standardized panel data.

8. The Sample Size for the Training Set

This information is not explicitly provided in the document. For in vitro diagnostic assays, "training sets" are usually involved in the initial development and optimization of the assay itself (e.g., establishing reagents, parameters, cut-offs). The studies described in this document are focused on the validation and verification of the new instrument's performance with existing, already developed assays, often referred to as "test sets" or "evaluation sets." The assays themselves were presumably developed and "trained" using various sample sets prior to these studies.

9. How the Ground Truth for the Training Set Was Established

Since information on a distinct "training set" for the new instrument's validation isn't provided (as the assays were pre-existing), details on its ground truth establishment are also not available in this document. For the development of the original assays, ground truth would have been established through a combination of clinical diagnoses, established reference methods, and correlation with disease status.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

BIOMERIEUX JOHN ALBRIGHT REGULATORY AFFAIRS MANAGER 595 ANGLUM ROAD HAZELWOOD MO 63042-2320

July 9, 2015

Re: K141133

Trade/Device Name: VIDAS H. pylori IgG, VIDAS 3, VIDAS Lyme IgG II, VIDAS RUB IgG, VIDAS TOXO M, VIDAS Human Chorionic Gonadotropin, VIDAS T4, VIDAS Testosterone, VIDAS TSH, and VIDAS D-Dimer Exclusion II

Regulation Number: 21 CFR 866.3110 Regulation Name: Campylobacter fetus serological reagents Regulatory Class: I Product Code: LYR, JJE, LSR, LFX, LGD, DHA, KLI, CDZ, JLW, DAP Dated: June 15, 2015 Received: June 16, 2015

Dear Mr. Albright:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Uwe Scherf -S for

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141133

Device Name

VIDAS H. pylori IgG, VIDAS Lyme IgG II, VIDAS RUB IgG, VIDAS TOXO IgM, VIDAS Human Chorionic Gonadotropin, VIDAS T4, VIDAS Testosterone, VIDAS TSH, and VIDAS D-Dimer Exclusion II

Indications for Use (Describe)

For VIDAS H. pylori IgG:

VIDAS® H. pylori IgG (HPY) is an automated qualitative test for use on the instruments of the VIDAS family, for the detection of anti-Helicobacter pylori IgG antibodies in human serum or plasma (EDTA) using the ELFA technique (Enzyme Linked Fluorescent Assay). The VIDAS HPY assay is intended as an aid in diagnosis of H. pylori infection in an adult symptomatic population.

This device is an in vitro diagnostic medical device for professional use only.

For VIDAS 3:

The VIDAS 3 system is a complete standalone immunodiagnostic system intended for trained and qualified laboratory technicians (daily routine use) and laboratory administrators (application configuration). This device is an in vitro diagnostic medical device for professional use only.

For VIDAS Lyme IgG II:

The VIDAS Lyme IgG II (LYG) assay is an automated qualitative enzyme immunoassay intended for use on the instruments of the VIDAS family in the presumptive detection of human IgG antibodies to Borrelia burgdorferi in human serum (plain or separation gel) or plasma (sodium heparin). It should be used to test patients with a history and/or symptoms of infection with B. burgdorferi. All VIDAS Lyme IgG II positive specimens should be further tested with a Western Blot IgG assay to obtain supportive evidence of infection with B. burgdorfei. This device is an in vitro diagnostic medical device for professional use only.

For VIDAS RUB IgG:

The VIDAS® RUB IgG (RBG) assay uses Enzyme Linked Fluorescent Assay (ELFA) technology on the instruments of the VIDAS family for the in vitro quantitative measurement of IgG antibodies to rubella virus in human serum. The VIDAS RUB IgG (RBG) assay is intended as an aid in the determination of immune status to rubella. The performance of this device has not been established for screening of cord blood, or for neonatal samples. Likewise, performance characteristics of the assay have not been established for immunocompromised or immunosuppressed individuals.

This device is an in vitro diagnostic medical device for professional use only.

For VIDAS TOXO IgM:

The VIDAS® TOXO IgM (TXM) assay is intended for use on the instruments of the VIDAS family (VITEK ImmunoDiagnostic Assay System) as an automated enzyme-linked fluorescent immunoassay (ELF A) for the presumptive qualitative detection of anti-Toxoplasma gondii IgM antibodies in human serum, as an aid in the diagnosis of acute, recent, or reactivated Toxoplasma gondii infection. This assay must be performed in conjunction with an anti-Toxoplasma gondii lgG antibody assay. VIDAS TOXO IgM (TXM) assay performance has not been established for prenatal screening or newborn testing. This assay has not been cleared by the FDA for blood/plasma donor screening. This device is an in vitro diagnostic medical device for professional use only.

For VIDAS Human Chorionic Gonadotropin:

The VIDAS® HCG (HCG) assay is intended for use on the instruments of the VIDAS family as an automated quantitative enzyme linked fluorescent immunoassay (ELFA) for the determination of human Chorionic Gonadotropin (hCG) concentration in human serum or plasma. The VIDAS HCG (HCG) assay is intended to aid in the early detection of pregnancy.

{3}------------------------------------------------

This device is an in vitro diagnostic medical device for professional use only.

For VIDAS T4:

The VIDAS® T4 (T4) assay is intended for use on the instruments of the VIDAS family as an automated quantitative enzyme-linked fluorescent immunoassay for the determination of human thyroxine (T4) concentration in serum or plasma (heparin). It is intended for use as an aid in the diagnosis and treatment of thyroid disorders. This device is an in vitro diagnostic medical device for professional use only.

For VIDAS Testosterone:

The VIDAS Testosterone (TES) assay is an automated quantitative test for use on the instruments of the VIDAS family for the enzyme immunoassay measure of total testosterone in human serum or plasma (lithium heparin), using the ELF A technique (Enzyme Linked Fluorescent Assay). It is intended as an aid in the diagnosis and management of conditions involving excess or deficiency of this androgen.

This device is an in vitro diagnostic medical device for professional use only.

For VIDAS TSH:

The VIDAS® TSH (TSH) assay is intended for use on the instruments of the VIDAS family as an automated quantitative enzyne-linked fluorescent immunoassay (ELFA) for the determination of human thyroid stimulating hormone- (TSH) concentration in human serum or plasma (heparin). It is intended for use as an aid in the diagnosis of thyroid or pituitary disorders.

This device is an in vitro diagnostic medical device for professional use only.

For VIDAS D-Dimer Exclusion II:

VIDAS® D-Dimer Exclusion II™ is an automated quantitative test for use on the instruments of the VIDAS family for the immunoenzymatic determination of fibrin degradation products (FbDP) in human plasma (sodium citrate, CTAD) using the ELFA technique (Enzyme Linked Fluorescent Assay).

VIDAS D-Dimer Exclusion II is indicated for use in conjunction with a clinical pretest probability assessment model to exclude deep vein thrombosis (DVT) and pulmonary embolism (PE) disease in outpatients suspected of DVT or PE. This device is an in vitro diagnostic medical device for professional use only.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for bioMérieux. The logo consists of a blue circle with a half-moon design on the left side. A green and yellow line runs vertically through the center of the circle. The word "BIOMÉ" is written in black, sans-serif font above the word "RIEUX".

510(k) SUMMARY

According to the 21 CFR 807.92, this 510(k) summary aims to provide an understanding of the basis for a determination of substantial equivalence.

Assigned 510(k) number: K141133

A1. Owner information

Submitter's Name:bioMérieux, Inc.
Address:595 Anglum RoadHazelwood, MO 63042
Contact Person:John Albright
Telephone number314-731-8546
Fax number314-731-8689
Date of Preparation:July 6, 2015

B1. Device Name

Trade Name:VIDAS® H.pylori IgG
Common Name:H.pylori IgG
Classification name:Campylobacter fetus serological reagents
Classification panel:Microbiology
Governing Regulation:21 CFR 866.3110
Device Classification:Class I
Product code:LYR

C1. Predicate Device

VIDAS H.pylori IgG (K001460)

D1. Device Description

The assay principle combines a 2-step enzyme immunoassay sandwich method with a final fluorescent detection (ELFA). All of the assay steps as well as the assay temperature are controlled automatically by the instrument. The Solid Phase Receptacle (SPR®) serves as the solid phase as well as the pipetting device for the assay. Reagents for the assay are ready-to-use and predispensed in the sealed reagent strips. After preliminary wash and sample dilution steps, the sample is cycled in and out of the SPR for a specified length of time. IgG antibodies to H. pylori present in the specimen will bind to the H. pylori antigen coating the interior of the SPR. Unbound sample components are washed away.

Anti-human IgG antibodies conjugated with alkaline phosphatase are cycled in and out of the SPR and will attach to any human IgG bound to the SPR wall. A final wash step removes unbound anti-human antibody conjugate. During the final detection step, the substrate (4 Methylumbellifery) phosphate) is cycled in and out of the SPR. The conjugate enzyme catalyzes the hydrolysis of this substrate into a

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for bioMérieux. The logo features a stylized globe with blue lines on the left side. A green and yellow line runs vertically through the center of the globe. The text "BIOMÉRIEUX" is written in a sans-serif font below the globe.

fluorescent product (4-Methylumbelliferone), the fluorescence of which is measured at 450 nm. The intensity of fluorescence is measured by the optical scanner in the instrument. At the assay, results are automatically calculated by the instrument, a test value is generated and a report is printed for each sample.

E1. Intended Use

VIDAS® H. pylori IgG (HPY) is an automated qualitative test for use on the instruments of the VIDAS family, for the detection of anti-Helicobacter pylori IgG antibodies in human serum or plasma (EDTA) using the ELFA technique (Enzyme Linked Fluorescent Assay). The VIDAS HPY assay is intended as an aid in diagnosis of H. pylori infection in an adult symptomatic population.

This device is an in vitro diagnostic medical device for professional use only.

F1. Technological Performances Summary

A summary of the technological characteristics of the new device in comparison to those of the predicate device is presented in the table below.

ItemNew DeviceVIDAS H. pyloriPredicate (K001460)VIDAS H. pylori
Intended UseVIDAS® H. pylori IgG (HPY) is an automatedqualitative test for use on the instruments of theVIDAS family, for the detection of anti-Helicobacterpylori IgG antibodies in human serum or plasma(EDTA) using the ELFA technique (Enzyme LinkedFluorescent Assay). The VIDAS HPY assay isintended as an aid in diagnosis of H. pylori infectionin an adult symptomatic population.Same
Sample typeSerum or plasmaSame
AnalyteH.pylori IgGSame
Assay TechniqueEnzyme-linked fluorescent immunoassay (ELFA)Same
Kit Composition30 Strips, 1x30 SPRs, 1 Standard (liquid), 1 PositiveControl (liquid), 1 Negative Control (liquid)Same(no change to assay).
AutomatedYesSame
InstrumentationVIDAS, miniVIDAS, VIDAS 3VIDAS, miniVIDAS

G1. Performance Data

Method Comparison

A study was conducted to verify the correlation of the VIDAS H. pylori IgG assay on the VIDAS 3 to the VIDAS H. pylori lgG assay on the VIDAS. One reagent lot, one of each instrument and 250 serum samples including positive, equivocal and negative samples were used, and results were evaluated according to CLSI EP12-A2 quideline "User Protocol for Evaluation of Qualitative Test Performance; Approved guideline".

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the logo for bioMérieux. The logo features the company name in a stylized, sans-serif font, with the "bio" portion in a smaller font size than "Mérieux". Above the name is a circular graphic that is half blue and half white with a green and yellow curved line running through the middle of the circle and extending above and below it.

Contingency Table:

VIDAS
PositiveEquivocalNegativeTotal
VIDAS 3Positive12230125
Equivocal06410
Negative02113115
Total12211117250

Associated percent agreements and their 95% two-sided score confidence intervals (CI) are calculated in the table below :

CategorySamples ofinterest/TotalPercent Agreement2-sided 95% CI
Negative113/11796.6%[ 91.5 ; 98.7 ] %
Positive122/122100%[ 96.9 ; 100.0 ] %

Precision

Three serum samples with samples close to the assay cut-off and moderate positive samples were tested in triplicate (3 replicates) twice a day (2 runs per day) over 6 days on 1 reagent lot using 3 instruments at 1 site (N = 108). The repeatability (within-run precision), between-day, between calibration, between instrument, total between calibration, and total between instrument precision were calculated according to CLSI EP5-A2 "Evaluation of Precision Performance of Quantitative Measurement Methods" and CLSI EP12-A2 "User Protocol for Evaluation of Qualitative Test Performance; Approved guideline - Second Edition". The results were as follows:

Sample 1N = 108Mean Test Value (TV)0.76Sample 2N = 108Mean Test Value (TV)1.26Sample 3N = 108Mean Test Value (TV)2.23
SDCV (%)SDCV (%)SDCV (%)
Within-RUN(Repeatability)0.067.70.086.20.125.2
Between-RUN0.045.70.043.00.041.7
Between-DAY0.000.00.000.00.052.4
Between-CALIBRATION0.012.00.032.00.052.1
Between-INSTRUMENT0.022.40.000.00.052.3

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for bioMérieux. The logo features the company name in a sans-serif font, with the letters spaced apart. Above the name is a blue sphere with curved lines on the left side, and a green line extending upwards from the sphere.

Sample 1Sample 2Sample 3
N = 108N = 108N = 108
Mean Test Value (TV)0.76Mean Test Value (TV)1.26Mean Test Value (TV)2.23
SDCV (%)SDCV (%)SDCV (%)
Total Between-CALIBRATION0.079.80.097.20.146.4
Total Between-INSTRUMENT0.0810.10.097.20.156.8

H1. Conclusion

The results for the performance testing submitted in this premarket notification are complete and demonstrate that the VIDAS H.pylori Assay is substantially equivalent to the predicate device identified in Item C1 of this summary. The addition of the VIDAS 3 instrument to the VIDAS Family does not affect the safety and effectiveness of the Reagent kit.

{8}------------------------------------------------

Image /page/8/Picture/13 description: The image shows the logo for bioMérieux. The logo consists of the company name in a sans-serif font, with a stylized globe above it. A green line bisects the globe and the company name, adding a touch of visual interest to the logo.

A2. Owner information

Submitter's Name:bioMérieux, Inc.
Address:595 Anglum RoadHazelwood, MO 63042
Contact Person:John Albright
Telephone number314-731-8546
Fax number314-731-8689
Date of Preparation:July 6, 2015

B2. Device Name

Trade Name:VIDAS® 3
Common Name:VIDAS 3
Classification name:Analyzer, chemistry (photometric, discrete), for clinical use
Classification panel:Clinical Chemistry
Governing Regulation:21 CFR 862.2160
Device Classification:Class I
Product code:JJE

C2. Predicate Device Name

Trade Name:

VITEK IMMUNODIAGNOSTIC ASSAY SYSTEM (VIDAS)

D2. Device Description

The VIDAS® 3 instrument is an automated multiparametric immunoassay system, which uses ELFA (Enzyme Linked Fluorescent Assay) technology. The VIDAS 3 system offers primary tube sampling, automated sample dilution, reagent/sample detection and reagent traceability.

The technology used, which is adaptable to a wide range of assays, combines the EIA method with a final fluorescence reading: this technology is known as ELFA (Enzyme Linked Fluorescent Assay). The enzyme used in the VIDAS product range is alkaline phosphatase, which catalyzes the hydrolysis of the substrate 4-methyl umbelliferyl phosphate (4-MUP) into a fluorescent product 4-methyl umbelliferone (4-MU) the fluorescence of which is measured at 450nm. The immunological methods are either indirect ElA, immunocapture, sandwich or competition, all involving a conjugate using the alkaline phosphatase.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image contains the logo for bioMérieux. The logo features a blue globe-like design with curved lines on the right side, and a green blade of grass extending from the top of the globe. Below the globe, the word "BIOMÉRIEUX" is written in a sans-serif font.

E2. Intended Use

The VIDAS 3 system is a complete standalone immunodiagnostic system intended for trained and qualified laboratory technicians (daily routine use) and laboratory administrators (application configuration).

This device is an in vitro diagnostic medical device for professional use only.

F2. Technological Performances Summary

ComponentNew DeviceVIDAS 3Predicate (K891385)VIDAS Instrument
Trade NameVIDAS 3VIDAS
Indications forUseThe VIDAS 3 system is a complete standaloneimmunodiagnostic system intended for trainedand qualified laboratory technicians (dailyroutine use) and laboratory administrators(application configuration).This device is an in vitro diagnostic medicaldevice for professional use only.The VIDAS® is a compact automatedmultiparametric immunoanalyzer thatusing predispensed disposable reagentstrips and specially coated Solid PhaseReceptacles (SPRs), can pipette, mix,incubate, control and analyze samplesfor in vitro diagnostic purposes.The VIDAS is only dedicated to be usedin combination with VIDAS reagentsrange, designed and produced bybioMérieux.The technology used, which isadaptable to a wide range of assays,combines the EIA method with a finalfluorescence reading: this technology isknown as ELFA (Enzyme LinkedFluorescent Assay). The enzyme usedin the VIDAS product range is alkalinephosphatase, which catalyzes thehydrolysis of the substrate 4-methylumbelliferyl phosphate (4-MUP) into afluorescent product 4-methylumbelliferone (4-MU) the fluorescenceof which is measured at 450nm.The immunological methods are eitherindirect EIA, immunocapture, sandwichor competition, all involving a conjugateusing the alkaline phosphatase.This device is an in vitro diagnosticmedical device for professional useonly.
TechnologyAutomated multiparametric immunoassaysystem which uses ELFA (Enzyme LinkedFluorescent Assay) technology.Same
# Sections45
Reagent slotsper section36
ComponentNew DeviceVIDAS 3Predicate (K891385)VIDAS Instrument
Total # samplesthat can be runsimultaneously1230
MonitorPeripheral (Attached)Peripheral
ComputerPeripheralPeripheral
KeyboardPeripheralPeripheral
PrinterPeripheralPeripheral
InstrumentComponentsAll unique or updated components are verifiedto not affect reagent performance.All unique or updated components areverified to not affect reagentperformance.
Components(Scanner head)The scanner head is an instrument sub-systemthat's primary function is to perform the opticalreading of the fluorescence as generated by theimmunoassay reaction.Same
Components(Software)The user Software functions include: entry ofpatient data; run execution; data reduction andinterpretation through a computation engine;edition of results; system supervision;management of calibrations and controls;validation of results; management of patientrecords.While the VIDAS 3 software is uniqueto the VIDAS 3, it offers the same basicfunctions as the VIDAS software anduses the same computation engine.
Components(Reagents)VIDAS reagents are comprised of predispenseddisposable reagent strips and specially coatedSolid Phase Receptacles (SPRs).Same
Reagent(Principle ofOperation)Each VIDAS assay has been designed to berun on any of the three VIDAS familyinstruments. Each assay has a unique protocol(volumes, sequence of steps, incubation times,etc) that is independent of the instrument.Same
ReagentThe assay intended use, clinical utility, principleof operation, kit composition, kit stability, kitstorage conditions, calibration type, calibrationfrequency, sample type, sample volume,calculation of results, and interpretation ofresults are all independent of the instrument.Same
Reagent LoadingManualSame
SPR/StripsynchronizationAutomated confirmationManual
Sample PipettingManual or automatedManual
Sample dilutionsManual or automatedManual
Execution of theassay protocolEach assay has its own pre-defined protocolthat defines the sequence of assay steps (e.g.volumes, incubation times, order of steps).Same
EnzymaticReadingThe reading made by the scanner head isbased on the 4-methylumbelliferone (4-MU)fluorescent product located in the opticalcuvette of the reagent strip after the enzymaticreaction has occurred.Same
ComponentNew DeviceVIDAS 3Predicate (K891385)VIDAS Instrument
Calculation andInterpretation ofresultsData reduction of the fluorescencemeasurement is based upon computationengine and a knowledge base including assayand lot specific information.Same
Unload Strips /SPRsAfter completion of the run, the user manuallyremoves the reagent Strips and SPRs.Same
Unload otherreagents andwasteManualNot applicable

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for bioMérieux. The logo features the company name in a stylized font, with the letters spaced apart. Above the name is a circular graphic with blue lines and a green line running vertically through the center.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image contains the logo for bioMérieux. The logo features the company name in a sans-serif font, with a stylized graphic above it. The graphic includes a blue circle with curved lines and a green line running vertically through the center.

G2. Performance Data

The VIDAS 3 performance testing was demonstrated through assay performance testing, using VIDAS D-Dimer Exclusion II, VIDAS H. pylori, VIDAS Human Chorionic Gonadotropin, VIDAS Lyme IgG II, VIDAS RUB IgG, VIDAS T4, VIDAS Testosterone, VIDAS TOXO M, and VIDAS TSH assays as the representative assays. This assay performance testing, which included method comparison, precision, detection limits, and linearity, demonstrated equivalency of the VIDAS 3 to the VIDAS and that running a VIDAS assay on either the VIDAS instrument or the VIDAS 3 instrument has no significant impact on the assay results.

H2. Conclusion

The results for the performance testing submitted in this premarket notification are complete and demonstrate that the VIDAS 3 is substantially equivalent to the predicate device identified in Item C2 of this summary.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image contains the logo for bioMérieux. The logo features a blue sphere with white lines on the left side. The text "BIOMÉRIEUX" is written in all caps below the sphere. A green and yellow line runs vertically through the sphere and text.

A3. Owner information

Submitter's Name:bioMérieux, Inc.
Address:595 Anglum RoadHazelwood, MO 63042
Contact Person:John Albright
Telephone number314-731-8546
Fax number314-731-8689
Date of Preparation:July 6, 2015

B3. Device Name

Trade Name:VIDAS® Lyme IgG II
Common Name:Lyme IgG
Classification name:Treponema pallidum treponemal test reagent
Classification panel:Microbiology
Governing Regulation:21 CFR 866.3830
Device Classification:Class II
Product code:LSR

C3. Predicate Device

VIDAS® Lyme IgG (K122986)

D3. Device Description

The VIDAS Lyme IgG II assay principle combines a 2-step enzyme immunoassay sandwich method with a final fluorescent detection (ELFA) (see User's Manual). The Solid Phase Receptacle (SPR®) serves as the solid phase as well as the pipetting device for the assay. Reagents for the assay are ready-to-use and predispensed in the sealed reagent strips. All of the assay steps are performed automatically by the instrument. The reaction medium is cycled in and out of the SPR several times. After a preliminary wash step and a sample dilution step, the antibodies to B. burgdorferi present in the specimen will bind to the B. burgdorferi specific recombinant proteins coating the interior of the SPR.

Unbound sample components are washed away. Antihuman IgG antibodies conjugated with alkaline phosphatase, will attach to the immunocomplex bound to the SPR wall.

A final wash step removes unbound conjugate. During the final detection step, the substrate (4-Methylumbelliferyl phosphate) is cycled in and out of the SPR. The conjugate enzyme catalyzes the hydrolysis of this substrate into a fluorescent product (4-Methylumbelliferone) the fluorescence of which is measured at 450 nm. The intensity of the fluorescence is proportional to the quantity of anti-B. burgdorferi IgG antibody present in the sample.

At the end of the VIDAS Lyme IgG II assay, results are automatically calculated by the instrument. A test value is generated and a report is printed.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for bioMérieux. The logo consists of a blue circle with a series of curved lines on the left side, resembling a stylized representation of a cell or microorganism. Below the circle is the word "BIOMÉRIEUX" in a sans-serif font. A green and yellow curved line extends from the top of the circle down through the middle of the word.

E3. Intended Use

The VIDAS Lyme IgG II (LYG) assay is an automated qualitative enzyme immunoassay intended for use on the instruments of the VIDAS family in the presumptive detection of human IgG antibodies to Borrelia burgdorferi in human serum (plain or separation gel) or plasma (sodium heparin or lithium heparin). It should be used to test patients with a history and/or symptoms of infection with B. burgdorferi. All VIDAS Lyme IgG II positive specimens should be further tested with a Western Blot IgG assay to obtain supportive evidence of infection with B. burgdorferi.

This device is an in vitro diagnostic medical device for professional use only.

F3. Technological Performances Summary

A summary of the technological characteristics of the new device in comparison to those of the predicate device is presented in the table below.

ItemNew DeviceVIDAS Lyme IgG IIPredicate (K122986)VIDAS Lyme IgG
Intended UseThe VIDAS Lyme IgG II (LYG) assay is anautomated qualitative enzyme immunoassayintended for use on the instruments of the VIDASfamily in the presumptive detection of human IgGantibodies to Borrelia burgdorferi in human serum(plain or separation gel) or plasma (sodiumheparin or lithium heparin). It should be used totest patients with a history and/or symptoms ofinfection with B. burgdorferi. All VIDAS Lyme IgGII positive specimens should be further testedwith a Western Blot IgG assay to obtainsupportive evidence of infection with B.burgdorferi.Same
Sample typeSerum or plasmaSame
AnalyteLyme IgGSame
Assay TechniqueEnzyme-linked fluorescent immunoassay (ELFA)Same
Kit Composition60 Strips, 2x30 SPRs, 1 Calibrator (liquid), 1Control (liquid)Same
AutomatedYesSame
InstrumentationVIDAS, miniVIDAS, VIDAS 3VIDAS, miniVIDAS

G3. Performance Data

Method Comparison

A study was conducted to verify the correlation of the VIDAS Lyme IgG II assay on the VIDAS 3 to the VIDAS Lyme IgG II assay on the VIDAS. One reagent lot, one of each instrument and 220 serum samples including positive and negative samples were used, and results were evaluated according to CLSI EP12-A2 guideline "User Protocol for Evaluation of Qualitative Test Performance; Approved quideline".

{14}------------------------------------------------

Image /page/14/Picture/1 description: The image shows the logo for bioMérieux. The logo features the company name in a sans-serif font, with the letters spaced out. Above the name is a blue globe-like design with curved lines, bisected by a green and yellow curved line that extends upwards.

Contingency Table

VIDAS
PositiveNegativeTotal
VIDAS 3Positive1090109
Negative1110111
Total110110220

Associated percent agreements and their 95% two-sided score confidence intervals are presented in the table below:

CategorySamples of Interest/TotalPercent Agreement with 2-sided 95% CI
Positive109/11099.09%[95.04;99.98] %
Negative110/110100.00%[96.70;100.00] %

Precision - VIDAS 3

Three serum samples were tested in triplicate (3 replicates) twice a day (2 runs per day) over 6 days on 1 reagent lot using 3 instruments at 1 site (N = 108). The results were calculated according to CLSI EP5-A2 and were as follows:

Panel MemberNMean IndexWithin-runBetween-runBetween-dayTotalBetween
SDCV (%)SDCV (%)SDCV (%)SDCV (%)
Sample 11080.170.015.40.000.00.001.80.018.2
l Sample 21080.230.014.00.000.90.001.00.026.9
Sample 31080.640.023.40.000.00.000.00.046.2

H3. Conclusion

The results for the performance testing submitted in this premarket notification are complete and demonstrate that the VIDAS Lyme IgG II Assay when used on VIDAS 3 instrument is substantially equivalent to the predicate device identified in Item C3 of this summary. The addition of the VIDAS 3 instrument to the VIDAS Family does not affect the safety and effectiveness of the Reagent kit.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image contains the logo for BIOMÉRIEUX. The logo features a blue globe-like symbol with curved lines on the top half and a green line running vertically through the center. Below the symbol, the company name "BIOMÉRIEUX" is written in capital letters with a sans-serif font.

A4. Owner information

Submitter's Name:bioMérieux, Inc.
Address:595 Anglum RoadHazelwood, MO 63042
Contact Person:John Albright
Telephone number314-731-8546
Fax number314-731-8689
Date of Preparation:July 6, 2015

B4. Device Name

Trade Name:VIDAS® RUB IgG
Common Name:Anti-Rubella IgG
Classification name:Rubella virus serological reagents
Classification panel:Microbiology
Governing Regulation:21 CFR 866.3510
Device Classification:Class II
Product code:LFX

C4. Predicate Device

VIDAS® RUB IgG (K080766)

D4. Device Description

The VIDAS® RUB IqG (RBG) assay is an enzyme-linked fluorescent immunoassay (ELFA) that consists of a two step enzyme immunoassay sandwich method and a fluorescent detection. A pipette tip-like disposable device, the Solid Phase Receptacle (SPR®), serves as the solid phase as well as a pipettor for the assay. Reagents for the assay are ready-to-use and are provided in sealed reagent strips. All assay steps are automated by the instrument. The assay reactions take place by pipetting up and down with the SPR through the wells of the reagent strip. The sample is initially diluted, and follows a first incubation step in the SPR, where anti-Rubella IgG antibodies present in the sample bind to the rubella antigen coating the interior of the SPR. After a wash step to eliminate any unbound component, a second incubation is performed using a monoclonal anti-human IgG alkaline phosphatase conjugate. Following another wash step, the detection substrate (4-methyl-umbellifery) phosphate) is added to the SPR.

The enzyme conjugate catalyzes the hydrolysis of the substrate into a fluorescent product (4-methylumbelliferone), and its fluorescence is measured by the intensity of the fluorescence is proportional to the concentration of rubella antibodies present in the sample. Results are automatically determined by the instrument based on the stored calibration curve and according to the CLSI® I/LA6-A recommended cut-off of 10 IU/mL. The assay is calibrated against the WHO 1st International Standard for anti Rubella Immunoglobulin, Human (1997).

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image contains the logo for bioMérieux. The logo features a blue circle with a half-moon design on the left side. A green and yellow curved line runs vertically through the center of the circle. The word "BIOMÉRIEUX" is written in a sans-serif font below the circle.

E4. Intended Use

The VIDAS® RUB IgG (RBG) assay uses Enzyme Linked Fluorescent Assay (ELFA) technology on the instruments of the VIDAS family for the in vitro quantitative measurement of IgG antibodies to rubella virus in human serum. The VIDAS® RUB IgG (RBG) assay is intended as an aid in the determination of immune status to rubella. The performance of this device has not been established for screening of cord blood, or for neonatal samples. Likewise, performance characteristics of the assay have not been established for immunocompromised or immunosuppressed individuals.

This device is an in vitro diagnostic medical device for professional use only.

F4. Technological Performances Summary

A summary of the technological characteristics of the new device in comparison to those of the predicate device is presented in the table below.

ItemNew DeviceVIDAS RUB IgGPredicate (K080766)VIDAS RUB IgG
Intended UseThe VIDAS® RUB IgG (RBG) assay uses EnzymeLinked Fluorescent Assay (ELFA) technology onthe instruments of the VIDAS family for the in vitroquantitative and qualitative measurement of IgGantibodies to rubella virus in human serum. TheVIDAS® RUB IgG (RBG) assay is intended as anaid in the determination of immune status to rubella.The performance of this device has not beenestablished for screening of cord blood, or forneonatalsamples. Likewise, performancecharacteristics of the assay have not beenestablished for immunocompromised orimmunosuppressed individualsSame
Sample typeSerumSame
AnalyteAnti-Rubella IgGSame
Assay TechniqueEnzyme-linked fluorescent immunoassay (ELFA)Same
Kit Composition60 Strips, 2x30 SPRs, 1 Calibrator (liquid), 1Positive Control (liquid), 1 Negative Control (liquid)Same(no change to assay).
AutomatedYesSame
InstrumentationVIDAS, miniVIDAS, VIDAS 3VIDAS, miniVIDAS

{17}------------------------------------------------

Image /page/17/Picture/1 description: The image shows the logo for bioMérieux. The logo consists of the company name in a sans-serif font, with a stylized globe above it. A green and yellow line bisects the globe vertically.

G4. Performance Data

Method Comparison - Quantitative

A study was conducted to verify the correlation of the VIDAS RUB IgG assay on the VIDAS 3 to the VIDAS RUB IgG assay on the VIDAS. One reagent lot, one of each instrument and 112 serum samples (ranging from 0 to 225 IU/mL) were used, and results were evaluated according to CLSI EP9. "Method Comparison and Bias Estimation Using Patient Samples". The results were as follows:

AnalysisNSlopeInterceptCorrelationcoefficient
Weighted Deming91*0.9791-0.22450.9776
Passing Bablock1121.000.000.9821
Weighted Deming89**0.9640-0.23430.9805
Passing Bablock111**1.000.000.9840
  • 21 samples had an average concentration equal to zero for VIDAS 3 and/or VIDAS, and were excluded from the Weighted Deming analysis.

** After removal of outliers.

Method Comparison - Qualitative

A study was conducted to verify the correlation of the VIDAS RUB IgG assay on the VIDAS 3 to the VIDAS RUB IgG assay on the VIDAS. One reagent lot, one of each instrument and 220 serum samples including positive, equivocal and negative samples were used, and results were evaluated according to CLSI EP12-A2 quideline "User Protocol for Evaluation of Qualitative Test Performance; Approved guideline".

Contingency Table:

VIDAS
PositiveEquivocalNegativeTotal
VIDAS 3Positive11100111
Equivocal38112
Negative019697
Total114997220

Associated percent agreements and their 95% two-sided exact confidence interval are calculated in the table below:

CategorySamples ofinterest/TotalPercent Agreement 2-sided 95% CI
Negative96/9799%[94.4;99.8] %
Positive111/11497.4%[92.5;99.1] %

{18}------------------------------------------------

Image /page/18/Picture/1 description: The image shows the logo for bioMérieux. The logo features the company name in a sans-serif font, with the "bio" portion in a darker, bolder font. Above the name is a blue globe-like graphic with curved lines, bisected by a green line that extends upward like a plant stem.

Precision

Five serum samples (with 2 samples close to the clinical decision points) were tested in duplicate (2 replicates) twice a day (2 runs per day) over 6 days on 1 reagent lot using 3 instruments at 1 site (N = 72). The repeatability (within-run precision), between-day, between calibration, between instrument, total between calibration, and total between instrument precision were calculated according to CLSI EP5-A2. "Evaluation of Precision Performance of Quantitative Measurement Methods". The results were as follows:

Sample 1Sample 2Sample 3Sample 4Sample 5
N = 72N = 72N = 72N = 72N = 72
Mean (IU/mL)Mean (IU/mL)Mean (IU/mL)Mean (IU/mL)Mean (IU/mL)
2.695.8112.8228.93132.88
SD%CVSD%CVSD%CVSD%CVSD%CV
Within-RUN(Repeatability)0.4115.20.55ರಿ.50.745.72.398.310.247.7
Between-RUN0.2910.70.000.00.000.00.000.00.000.0
Between-DAY0.000.00.000.00.191.50.431.50.000.0
Between-CALIBRATION0.000.00.101.80.332.60.752.65.254.0
Between-INSTRUMENT0.093.50.000.00.000.00.000.00.000.0
Total Between-CALIBRATION0.5018.60.569.70.836.52.548.811.518.7
Total Between-INSTRUMENT0.5118.90.569.70.836.52.548.811.518.7

Linearity

Two sample pools (VNH1 and VNH2) were serially diluted into a total of 11 samples and tested in triplicate in order to evaluate the linearity according to CLSI EP06-A "Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach". The VIDAS RUB IgGa assay on VIDAS 3 is linear across the measuring range 0 - 250 IU/mL.

Image /page/18/Figure/7 description: The figure is a scatter plot titled "VN H1" that shows the observed concentration of RBG (IU/mL) on the y-axis versus the expected concentration of RBG (IU/mL) on the x-axis. The plot includes data points for observed concentrations, observed concentration means, a straight regression line, and a third-order polynomial. The x and y axis range from 0 to 40, and the data points appear to follow a linear trend.

{19}------------------------------------------------

Image /page/19/Picture/1 description: The image shows the logo for bioMérieux. The logo consists of the company name in a stylized font, with the "bio" part in a larger font size than the "Mérieux" part. Above the company name is a blue globe-like symbol with curved lines, and a green line that extends from the top of the globe down through the middle of the logo.

Image /page/19/Figure/2 description: The image is a scatter plot showing the relationship between the expected concentration and observed concentration of RBG in IU/mL. The x-axis represents the expected concentration of RBG, ranging from 0 to 300 IU/mL. The y-axis represents the observed concentration of RBG, ranging from 0 to 400 IU/mL. The plot includes data points representing observed concentrations, with a straight regression line indicating the trend between the two variables.

Detection Limits

The detection limits of the VIDAS RUB IgG assay on the VIDAS 3 were evaluated per CLSI EP17-A2 "Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline, Second Edition''. For LoB determination, 6 blank samples were each tested in 4 replicates in a single run per day for 4 days with 1 VIDAS 3 instrument and 2 assay lots. For LoD/LoQ determinations, 6 low-level samples were each tested in 6 replicates, 2 runs per day, for 5 days, with 1 VIDAS 3 instrument and 2 assay lots. The LoB, LoD and LoQ determinations were based consequently on 912 results, including 192 results from blank samples and 720 results from low-level samples both assay lots combined. The LoB, LoD, LoQ of the VIDAS RUB IgG assay on the VIDAS 3 are as follows: LoB = 0.389 IU/mL; LoD = 0.595 IU/mL; LoQ = 0.608 IU/mL.

Assay Performance Testing using a CDC Panel

A CDC low-titer rubella antibody standard and a CDC reference panel (CDC RUB IgG serum panel) was tested using the VIDAS RUB IgG assay on the VIDAS 3 system and on the VIDAS system.

CDC low-titer rubella antibody standard results:

Vials of CDC low-titer reference standard were reconstituted water as preconized by CDC and then pooled as MSN (Master Stock Neat). According to the CDC information, the theoretical dose of the neat solution was 21 IU/mL. A ½ dilution solution was then prepared from the MSN solution and had an expected dose at 10.5 IU/mL.

The neat and 1/2 dilutions solutions of the CDC low titer standard (MSN & MS1/2 solutions) were then tested in triplicate in the same run on the VIDAS and the VIDAS 3 instruments using the same VIDAS RUB IgG reagent lot.

The mean result (IU/mL) obtained with the neat and 1/2 dilutions of the CDC low-titer standard and the difference in % compared to the theoretical/expected concentration of CDC low-titer standard were calculated for the VIDAS and the VIDAS 3:

  • for VIDAS 3, the mean value of the neat and ½ dilution were respectively -11.0% and -11.4% compared to the theoretical concentration.
  • for VIDAS, the mean value of the neat and ½ dilution were respectively -15.7% and -7.6% compared to the theoretical concentration.

{20}------------------------------------------------

Image /page/20/Picture/0 description: The image contains the logo for bioMérieux. The logo features the company name in a sans-serif font, with the "bio" portion in a darker shade. Above the name is a circular graphic with blue lines on the left side and a green line that curves through the center. The overall design is clean and corporate.

CDC reference panel testing, data analysis and results:

The CDC reference panel consisted of 100 specimens, 50 pairs of sera titered by Hemagglutination Inhibition (9 negative sera resulting in 18 negative specimens and 41 positive sera resulting in 82 positive specimens). A single replicate of each of the 100 CDC reference panel samples was tested on the VIDAS and the VIDAS 3 instruments using the same VIDAS RUB IgG reagent lot.

The number of positive and negative sera detected are identical between VIDAS 3 systems: The VIDAS RUB IgG (RBG) assay identified 80/82 (97.6%) positive tests on 82 positive sera and 18/18 (100%) negative tests on 18 negative sera. One of the pairs of HI positive sera was reported as VIDAS equivocal (both results).

{21}------------------------------------------------

Image /page/21/Picture/0 description: The image contains the logo for bioMérieux. The logo features a blue and white globe-like symbol with vertical lines, bisected by a green and yellow line. Below the symbol, the company name "BIOMÉRIEUX" is written in a simple, sans-serif font, with each letter separated by a space.

A5. Owner information

Submitter's Name:bioMérieux, Inc.
Address:595 Anglum RoadHazelwood, MO 63042
Contact Person:John Albright
Telephone number314-731-8546
Fax number314-731-8689
Date of Preparation:July 6, 2015

B5. Device Name

Trade Name:VIDAS® TOXO M
Common Name:Anti-TOXO IgM
Classification name:Toxoplasma gondiiserological reagents
Classification panel:Microbiology
Governing Regulation:21 CFR 866.3780
Device Classification:Class II
Product code:LGD

C5. Predicate Device

VIDAS® TOXO M (K923166)

D5. Device Description

The VIDAS TOXO IgM (TXM) assay is an enzyme-linked fluorescent immunoassay (ELFA) that is performed in an automated instrument. All assay steps and assay temperature are controlled by the instrument. A pipette tip-like disposable device, the Solid Phase Receptacle (SPR®), serves as the solid phase as well as the pipettor for the assay. The SPR is coated with goat anti-u chain antibodies. The VIDAS TOXO IgM (TXM) assay configuration prevents nonspecific reactions with the SPR. Reagents for the assay are in the sealed TXM Reagent Strips. After a sample dilution step, the sample is cycled in and out of the SPR for a specified length of time. IgM antibodies present in the specimen will bind to the anti-u chain antibodies coating the interior of the SPR. Unbound sample components are washed away. An immunocomplex of T. gondii antigen and mouse monoclonal anti-P30 antibodies conjugated with alkaline phosphatase is cycled in and out of the SPR and will attach to the human anti-T. gondii IgM bound to the SPR wall (Note: P30 is a major surface protein of the T. gondii tachyzoite, with molecular weight of 30,000 (5)). A final wash step removes unbound conjugate. A fluorescent substrate, 4-methylumbelliferyl phosphate, is introduced into the SPR. Enzyme remaining on the wall of the SPR will catalyze the conversion of the fluorescent product, 4-methylumbelliferone. The intensity of the fluorescence is measured by the optical scanner in the instrument.

{22}------------------------------------------------

Image /page/22/Picture/0 description: The image shows the logo for bioMérieux. The logo consists of the company name in a stylized font, with a blue sphere above it. A green and yellow line bisects the sphere and the company name. The sphere has a series of curved lines on the left side.

When the VIDAS TOXO IgM (TXM) assay is completed, the results are analyzed automatically by the instrument, a test value is generated, and a report is printed for each sample.

E5. Intended Use

The VIDAS® TOXO IgM (TXM) assay is intended for use on the instruments of the VIDAS family ImmunoDiagnostic Assay System) as an automated enzyme-linked fluorescent (VITEK® immunoassay (ELFA) for the presumptive qualitative detection of anti-Toxoplasma gondii IgM antibodies in human serum, as an aid in the diagnosis of acute, recent, or reactivated Toxoplasma gondii infection. This assay must be performed in conjunction with an anti-Toxoplasma gondii IgG antibody assay. VIDAS TOXO IgM (TXM) assay performance has not been established for prenatal screening or newborn testing. This assay has not been cleared by the FDA for blood/plasma donor screening.

This device is an in vitro diagnostic medical device for professional use only.

F5. Technological Performances Summary

A summary of the technological characteristics of the new device in comparison to those of the predicate device is presented in the table below.

ItemNew DeviceVIDAS TOXO IgMPredicate (K923166)VIDAS TOXO IgM
Intended UseThe VIDAS® TOXO IgM (TXM) assay is intended for useon the instruments of the VIDAS family (VITEK®ImmunoDiagnostic Assay System) as an automatedenzyme-linked fluorescent immunoassay (ELFA) for thepresumptive qualitative detection of anti-Toxoplasmagondii IgM antibodies in human serum, as an aid in thediagnosis of acute, recent, or reactivated Toxoplasmagondii infection. This assay must be performed inconjunction with an anti-Toxoplasma gondii IgG antibodyassay. VIDAS TOXO IgM (TXM) assay performance hasnot been established for prenatal screening or newborntesting. This assay has not been cleared by the FDA forblood/plasma donor screening.Same
Sample typeSerumSame
AnalyteAnti-Toxo IgMSame
Assay TechniqueEnzyme-linked fluorescent immunoassay (ELFA)Same
Kit Composition60 Strips, 2x30 SPRs, 1 Calibrator (liquid), 1 Control(liquid)Same(no change to assay).
AutomatedYesSame
InstrumentationVIDAS, miniVIDAS, VIDAS 3VIDAS, miniVIDAS

G5. Performance Data

Method Comparison

A study was conducted to verify the correlation of the VIDAS TOXO IgM assay on the VIDAS 3 to the VIDAS TOXO IgM assay on the VIDAS. One reagent lot, one of each instrument and 198 serum

{23}------------------------------------------------

Image /page/23/Picture/1 description: The image shows the logo for bioMérieux. The logo features the company name in a sans-serif font, with the "BIO" portion in a bolder font. Above the name is a blue globe-like symbol with curved lines on the left side. A green line curves over the globe and extends downward, creating a visual element that suggests growth or movement.

samples were used, and results were evaluated according to CLSI EP9. "Method Comparison and Bias Estimation Using Patient Samples". The results were as follows:

Contingency table:

VIDAS
PositiveEquivocalNegativeTotal
VIDAS 3PositiveવેરૂO0ત્ત્વે ઉત્પત્તર તે ઉત્પત્તર તે જે જે જે જે જે જે જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયે
Equivocal206
Negative10ರಿಕ99
Total96র্বರಿ8198

Associated percent agreements and their 95% two-sided score confidence intervals are calculated in the table below :

CategorySamples of interest /TotalPercent Agreement 2-sided 95% CI
Negative98/98100%[96.2;100.0] %
Positive93/9696.9%[91.2;98.9] %

Precision

Four serum samples were tested in 3 replicates twice a day (2 runs per day) over 6 days on 1 reagent lot using 3 instruments at 1 site (N = 108). The repeatability (within-run precision), between-run, between-day, between calibration, between instrument, total between calibration, and total between instrument precision were calculated according to CLSI EP5-A2. "Evaluation of Precision Performance of Quantitative Measurement Methods". The results were as follows:

Sample 1N = 108Sample 2N = 108Sample 3N = 108Sample 4N = 108
Mean Index0.47Mean Index0.60Mean Index0.79Mean Index1.07
SD%CVSD%CVSD%CVSD%CV
Within-RUN(Repeatability)0.013.00.023.00.022.70.032.7
Between-RUN0.000.00.000.70.000.00.000.0
Between-DAY0.000.00.000.50.010.90.011.1
Between-CALIBRATION0.012.40.000.00.010.80.000.0
Between-INSTRUMENT0.000.00.000.80.010.70.021.7
Total Between-CALIBRATION0.023.90.023.20.023.00.032.9
Total Between-INSTRUMENT0.023.90.023.30.023.10.043.4

{24}------------------------------------------------

A6. Owner information

Submitter's Name:bioMérieux, Inc.
Address:595 Anglum RoadHazelwood, MO 63042
Contact Person:John Albright
Telephone number314-731-8546
Fax number314-731-8689
Date of Preparation:July 6, 2015

B6. Device Name

Trade Name:VIDAS® Human chorionic gonadotropin
Common Name:HCG
Classification name:Human chorionic gonadotropin (HCG) test system
Classification panel:Clinical Chemistry
Governing Regulation:21 CFR 862.1155
Device Classification:Class II
Product code:DHA

C6. Predicate Device

VIDAS® HCG (K921302)

D6. Device Description

The VIDAS HCG (HCG) assay is an enzyme-linked fluorescent immunoassay (ELFA) that is performed in an automated instrument. All assay steps and assay temperature are controlled by the instrument. A pipette tip-like disposable device, the Solid Phase Receptacle (SPR), serves as a solid phase for the assay as well as a pipetting device. The SPR is coated at the time of manufacture with mouse monoclonal anti-hCG antibodies. The VIDAS HCG (HCG) assay configuration prevents nonspecific reactions with the SPR. Reagents for the assay are located in the sealed Reagent Strips. The sample is transferred into the well containing anti-hCG antibody conjugated with alkaline phosphatase.

The sample/conjugate mixture is cycled in and out of the SPR and the hCG will bind to antibodies coated on the SPR and to the conjugate forming a "sandwich". Wash steps remove unbound conjugate. A fluorescent substrate, 4-methylumbelliferyl phosphate, is cycled through the SPR. Enzyme remaining on the SPR wall will catalyze the conversion of the substrate to the fluorescent product 4-methylumbelliferone. The intensity of fluorescence is measured by the optical scanner in the instrument ; it is proportional to the hCG concentration present in the sample. When the VIDAS HCG (HCG) assay is completed, the results are analyzed automatically by the instrument and a report is printed for each sample.

{25}------------------------------------------------

Image /page/25/Picture/0 description: The image shows the logo for bioMérieux. The logo consists of a blue circle with vertical lines on the left side, resembling a globe. A green and yellow curved line runs vertically through the center of the circle. Below the circle, the word "BIOMÉRIEUX" is written in a sans-serif font.

E6. Intended Use

The VIDAS® HCG (HCG) assay is intended for use on the instruments of the VIDAS family as an automated quantitative enzyme linked fluorescent immunoassay (ELFA) for the determination of human Chorionic Gonadotropin (hCG) concentration in human serum or plasma. The VIDAS HCG (HCG) assay is intended to aid in the early detection of pregnancy.

This device is an in vitro diagnostic medical device for professional use only.

F6. Technological Performances Summary

A summary of the technological characteristics of the new device in comparison to those of the predicate device is presented in the table below.

ItemNew DeviceVIDAS HCGPredicate (K921302)VIDAS HCG
Intended UseThe VIDAS® HCG (HCG) assay isintended for use on the instruments of theVIDAS family as an automated quantitativeenzyme linked fluorescent immunoassay(ELFA) for the determination of humanChorionic Gonadotropin (hCG)concentration in human serum or plasmato aid in the early detection of pregnancy.The VIDAS HCG (HCG) assay is intendedfor use on the instruments of the VIDASfamily (VITEK ImmunoDiagnostic AssaySystem) as an automated quantitativeenzyme linked fluorescent immunoassay(ELFA) for the determination of humanChorionic Gonadotropin (hCG)concentration in human serum or plasmafor use in the early detection of pregnancy.
Sample typeSerum or plasmaSame
AnalyteHCGSame
Assay TechniqueEnzyme-linked fluorescent immunoassay(ELFA)Same
Kit Composition60 Strips, 2x30 SPRs, 1 Calibrator(lyophilized), 1 Control (lyophilized), 1Diluent Buffer (liquid)Same(no change to assay).
Measuring Range2 - 1500 mIU/mLSame
AutomatedYesSame
InstrumentationVIDAS, miniVIDAS, VIDAS 3VIDAS, miniVIDAS

{26}------------------------------------------------

Image /page/26/Picture/1 description: The image shows the logo for bioMérieux, a French multinational biotechnology company. The logo features a blue globe-like symbol with curved lines, bisected by a green and yellow line. Below the symbol is the company name, "BIOMÉRIEUX," in a sans-serif font.

G6. Performance Data

Method Comparison

A study was conducted to verify the correlation of the VIDAS HCG assay on the VIDAS 3 to the VIDAS HCG assay on the VIDAS. One reagent lot, one of each instrument and 113 serum samples were used, and results were evaluated according to CLSI EP9. "Method Comparison and Bias Estimation Using Patient Samples". The results, and the 95% Confidence Intervals (CI) were as follows:

AnalysisNSlopeInterceptCorrelationcoefficient
Weighted Deming1130.9265[0.9043, 0.9488]0.0828[-0.1242, 0.2898]0.9848[0.9779, 0.9895]

Precision

Six serum samples were tested in duplicate (2 replicates) twice a day (2 runs per day) over 6 days on 1 reagent lot using 3 instruments at 1 site (N = 72). The repeatability (within-run precision), betweenrun, between-day, between calibration, between instrument, total between calibration, and total between instrument precision were calculated according to CLSI EP5-A2. "Evaluation of Precision Performance of Quantitative Measurement Methods". The results were as follows:

Sample 1N = 72Sample 2N = 72Sample 3N = 72Sample 4N = 72Sample 5N = 72Sample 6N = 72
Mean (mIU/mL)4.46Mean (mIU/mL)6.46Mean (mIU/mL)9.48Mean (mIU/mL)74.43Mean (mIU/mL)311.52Mean (mIU/mL)1109.32
SD%CVSD%CVSD%CVSD%CVSD%CVSD%CV
Within-RUN(Repeatability)0.306.70.355.40.414.43.164.212.484.056.365.1
Between-RUN0.122.70.203.10.131.40.731.00.000.00.000.0
Between-DAY0.000.00.000.00.192.00.320.45.751.80.000.0
Between-CALIBRATION0.081.70.264.10.121.32.363.25.591.832.712.9
Between-INSTRUMENT0.081.70.000.00.232.40.000.00.000.00.000.0
Total Between-CALIBRATION0.337.40.487.50.495.24.025.414.844.865.165.9
Total Between-INSTRUMENT0.347.60.487.50.545.74.025.414.844.865.165.9

Linearity

Three sample pools were serially diluted into a total of 11 samples and tested in triplicate in order to evaluate the linearity according to CLSI EP06-A "Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach". The VIDAS HCG assay on VIDAS 3 is linear across the measuring range 2 - 1500 mIU/mL.

{27}------------------------------------------------

Image /page/27/Picture/0 description: The image shows the logo for bioMérieux. The logo consists of the company name in a sans-serif font, with a stylized globe above it. The globe is blue and has curved lines on the left side, and a green and yellow line runs vertically through the center of the globe and the company name.

Detection Limits

The detection limits of the VIDAS HCG assay on the VIDAS 3 were evaluated per CLSI EP17-A2 "Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures: Approved Guideline. Second Edition", For LoB determination, 6 blank samples were each tested in 4 replicates in a single run per day for 4 days with 1 VIDAS 3 instrument and 2 assay lots. For LoD/LoQ determinations, 6 low-level samples were each tested in 6 replicates, 2 runs per day, for 5 days, with 1 VIDAS 3 instrument and 2 assay lots. The LoB, LoD and LoQ determinations were based consequently on 912 results, including 192 results from blank samples and 720 results from Iow-level samples both assay lots combined. The LoB, LoQ of the VIDAS HCG assay on the VIDAS 3 are as follows: LoB = 0.242 mlU/mL; LoD = 0.571 mlU/mL; LoQ = 1.280 mlU/mL.

Carry-over

One high HCG positive sample was tested with one HCG batch analyte negative sample and one carry-over analyte negative sample during 7 runs on 3 VIDAS 3 instruments until 84 replicates of each sample was obtained. The average carry-over observed on VIDAS 3 and its 95% confidence interval were calculated and compared to a <5 PPM (parts per million) acceptance criteria. The capacity of VIDAS 3 to not generate Carry-Over was demonstrated.

Dilution

Four samples (human serum pool) with the following approximate HCG concentrations 6 000, 25 000, 75 000, and 150 000 mUllmL, were tested on two VIDAS 3 systems using an automated dilution sequence and a manual dilution sequence. For the automated dilutions, eleven (11) dilution levels ranging from 1/1 to 1/1000 were tested per sample. For the manual dilutions, two dilution levels were defined and tested per sample.

For each sample, each system and each dilution ratio, the average relative difference between the observed HCG concentration (automatic testing) and the estimated theoretical HCG concentration (manual testing) is estimated as follow:

Relative diff. (%) = 100 X (Observed dose - Theoretical dose) / Theoretical dose.

At a global level, the global average relative difference observed per dilution ratio ranges from -3.1% to +9.4%.

H6. Conclusion

The results for the performance testing submitted in this premarket notification are complete and demonstrate that the VIDAS HCG Assay is substantially equivalent to the predicate device identified in ltem C6 of this summary. The addition of the VIDAS 3 instrument to the VIDAS Family does not affect the safety and effectiveness of the Reagent kit.

{28}------------------------------------------------

Image /page/28/Picture/0 description: The image shows the logo for bioMérieux. The logo consists of the company name in a sans-serif font, with a stylized graphic above it. The graphic is a blue circle with curved lines on the left side, bisected by a green and yellow line.

A7. Owner information

Submitter's Name:bioMérieux, Inc.
Address:595 Anglum RoadHazelwood, MO 63042
Contact Person:John Albright
Telephone number:314-731-8546
Fax number314-731-8689
Date of Preparation:July 6, 2015

B7. Device Name

Trade Name:VIDAS® T4
Common Name:T4
Classification name:Total thyroxine test system
Classification panel:Clinical Chemistry
Governing Regulation:21 CFR 862.1700
Device Classification:Class II
Product code:KLI

C7. Predicate Device

VIDAS® T4 (K926393)

D7. Device Description

The VIDAS T4 (T4) assay is an enzyme-linked fluorescent immunoassay (ELFA) that is performed in an automated instrument. All assay steps and assay temperature are controlled by the instrument. A pipette tip-like disposable device, the Solid Phase Receptacle (SPR), serves as a solid phase for the assay as well as a pipetting device. At the time of manufacture, the SPRs are coated with mouse monoclonal anti-T4 antibodies. The VIDAS T4 (T4) assay configuration prevents nonspecific reactions with the SPR. Reagents for the assay are in the sealed T4 Reagent Strips. The sample is transferred into the well containing the T4 antigen conjugated with alkaline phosphatase. The conjugate solution also contains ANS and sodium salicylate, which liberate bound T4 from the carrier proteins in the sample.

The sample/conjugate mixture is cycled in and out of the SPR and the T4 in the sample competes with the T4-alkaline phosphatase conjugate for binding with the mouse monoclonal anti-T4 antibodies coated on the SPR. Wash steps remove unbound conjugate. A fluorescent substrate, 4methylumbelliferyl phosphate, is cycled in and out of the SPR. Enzyme remaining on the SPR well will catalyze the conversion of the substrate to the fluorescent product 4-methylumbelliferone (450 nm). The intensity of fluorescence is measured by the optical scanner in the instrument. It is inversely proportional to the T4 concentration present in the sample. When the VIDAS T4 (T4) is completed, the results are analyzed automatically by the instrument, and a report is printed for each sample.

{29}------------------------------------------------

Image /page/29/Picture/0 description: The image contains the logo for bioMérieux. The logo features the company name in a stylized, sans-serif font, with the "BIO" portion in a bolder weight. Above the name is a circular graphic that is half blue and half white, with a green and yellow curved line running vertically through the center of the circle and extending downwards.

E7. Intended Use

The VIDAS® T4 (T4) assay is intended for use on the instruments of the VIDAS family as an automated quantitative enzyme-linked fluorescent immunoassay for the determination of human thyroxine (T4) concentration in serum or plasma (heparin). It is intended for use as an aid in the diagnosis and treatment of thyroid disorders.

This device is an in vitro diagnostic medical device for professional use only.

F7. Technological Performances Summary

A summary of the technological characteristics of the new device in comparison to those of the predicate device is presented in the table below.

ItemNew DeviceVIDAS T4Predicate (K926393)VIDAS T4
Intended UseThe VIDAS® T4 (T4) assay is intended foruse on the instruments of the VIDASfamily as an automated quantitativeenzyme-linked fluorescent immunoassayfor the determination of human thyroxine(T4) concentration in serum or plasma(heparin). It is intended for use as an aidin the diagnosis and treatment of thyroiddisorders.The VIDAS T4 (T4) assay is intended foruse on the instruments of the VIDAS family(Vitek ImmunoDiagnostic Assay System) asan automated quantitative enzyme-linkedfluorescent immunoassay for thedetermination of human thyroxine (T4)concentration in serum or plasma (heparin).It is intended for use as an aid in thediagnosis and treatment of thyroid disorders.
Sample typeSerum or plasmaSame
AnalyteT4Same
Assay TechniqueEnzyme-linked fluorescent immunoassay(ELFA)Same
Kit Composition60 Strips, 2x30 SPRs, 1 Calibrator(liquid),1 Control (liquid)Same
Measuring Range6 - 320 nmol/LSame
AutomatedYesSame
InstrumentationVIDAS, miniVIDAS, VIDAS 3VIDAS, miniVIDAS

G7. Performance Data

Method Comparison

A study was conducted to verify the correlation of the VIDAS T4 assay on the VIDAS 3 to the VIDAS T4 assay on the VIDAS. One reagent lot, one of each instrument and 105 serum samples were used, and results were evaluated according to CLSI EP9. "Method Comparison and Bias Estimation Using Patient Samples". The results were as follows:

AnalysisNSlopeInterceptCorrelation coefficient
Weighted Deming1050.9547-0.68600.9866
Passing - Bablok1050.9523-0.43970.9866

{30}------------------------------------------------

Image /page/30/Picture/1 description: The image shows the logo for bioMérieux. The logo consists of a blue circle with a series of curved lines on the left side, resembling a globe. A green and yellow curved line extends from the top of the circle to the bottom. The word "BIOMÉRIEUX" is written in a sans-serif font below the circle, with each letter separated by a space.

Precision

Five serum samples were tested in duplicate (2 replicates) twice a day (2 runs per day) over 6 days on 1 reagent lot using 3 instruments at 1 site (N = 72). The repeatability (within-run precision), betweenrun, between-day, between calibration, between instrument, total between calibration, and total between instrument precision were calculated according to CLSI EP5-A2. "Evaluation of Precision Performance of Quantitative Measurement Methods". The results were as follows:

Sample 1Sample 2Sample 3Sample 4Sample 5
N = 72N = 72N = 72N = 72N = 72
Mean (nmol/L)Mean (nmol/L)Mean (nmol/L)Mean (nmol/L)Mean (nmol/L)
12.835.1463.60121.99227.17
SD%CVSD%CVSD%CVSD%CVSD%CV
Within-RUN(Repeatability)1.179.11.544.41.883.04.243.512.465.5
Between-RUN0.000.00.000.00.621.01.531.30.000.0
Between-DAY0.766.00.000.00.340.50.000.04.742.1
Between-CALIBRATION0.514.00.952.71.442.31.050.94.131.8
Between-INSTRUMENT0.000.00.000.00.000.00.480.40.000.0
Total Between-CALIBRATION1.4811.61.805.12.473.94.633.813.966.1
Total Between-INSTRUMENT1.4811.61.805.12.473.94.663.813.966.1

Linearity

One sample pool was serially diluted into a total of 11 samples and tested in triplicate in order to evaluate the linearity according to CLSI EP06-A "Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach", The VIDAS T4 assay on VIDAS 3 is linear across the measuring range 6 - 320 nmol/L.

Detection Limits

The detection limits of the VIDAS T4 assay on the VIDAS 3 were evaluated per CLSI EP17-A2 "Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline. Second Edition". For LoB determination, 6 blank samples were each tested in 4 replicates in a single run per day for 4 days with 1 VIDAS 3 instrument and 2 assay lots. For LoD/LoQ determinations, 6 low-level samples were each tested in 6 replicates, 2 runs per day, for 5 days, with 1 VIDAS 3 instrument and 2 assay lots. The LoB, LoD and LoQ determinations were based consequently on 912 results, including 192 results from blank samples and 720 results from low-level samples both assay lots combined. The LoB, LoD, LoQ of the VIDAS T4 assay on the VIDAS 3 are as follows: LoB = 1.596 nmol/L; LoD = 3.749 nmol/L; LoQ = 6.216 nmol/L.

H7. Conclusion

The results for the performance testing submitted in this premarket notification are complete and demonstrate that the VIDAS T4 Assay is substantially equivalent to the predicate device identified in ltem C7 of this summary. The addition of the VIDAS 3 instrument to the VIDAS Family does not affect the safety and effectiveness of the Reagent kit.

{31}------------------------------------------------

Image /page/31/Picture/0 description: The image shows the logo for bioMérieux. The logo consists of the company name in a sans-serif font, with a stylized globe above it. The globe is divided into two halves, one with blue lines and the other a solid dark blue. A green and yellow line runs vertically through the center of the globe and the company name.

A8. Owner information

Submitter's Name:bioMérieux, Inc.
Address:595 Anglum RoadHazelwood, MO 63042
Contact Person:John Albright
Telephone number:314-731-8546
Fax number314-731-8689
Date of Preparation:July 6, 2015

B8. Device Name

Trade Name:VIDAS® Testosterone
Common Name:Testosterone
Classification name:Testosterone test system
Classification panel:Clinical Chemistry
Governing Regulation:21 CFR 862.1680
Device Classification:Class I
Product code:CDZ

C8. Predicate Device

VIDAS® Testosterone (K021326)

D8. Device Description

The assay principle combines an enzyme immunoassay competition method with a final fluorescent detection (ELFA). The Solid Phase Receptacle (SPR®), serves as the solid phase as well as the pipetting device for the assay. Reagents for the assay are ready-to-use and predispensed in the sealed reagent strips. All of the assay steps are performed automatically by the instrument. The reaction medium is cycled in and out of the SPR several times. The sample is taken and transferred into the well containing the conjugate which is an alkaline phosphatase-labeled testosterone derivative. The testosterone present in the testosterone derivative in the conjugate compete for the antitestosterone specific antibody sites coated to the inner surface of the SPR. Unbound components are eliminated during the washing steps.

During the final detection step, the substrate (4-Methyl-umbelliferyl phosphate) is cycled in and out of the SPR. The conjugate enzyme catalyzes the hydrolysis of this substrate into a fluorescent product (4-Methyl-umbelliferone), the fluorescence of which is measured at 450 nm. The intensity of the fluorescence is inversely proportional to the concentration of testosterone present in the sample. At the end of the assay, results are automatically calculated by the instrument in relation to the calibration curve stored in memory, and then printed out.

{32}------------------------------------------------

Image /page/32/Picture/0 description: The image shows the logo for bioMérieux. The logo features a blue circle with a half-moon design on the left side. A green and yellow curved line runs vertically through the center of the circle and extends downwards. The word "BIOMÉRIEUX" is written in a stylized font below the circle.

E8. Intended Use

The VIDAS Testosterone (TES) assay is an automated quantitative test for use on the instruments of the VIDAS family for the enzyme immunoassay measure of total testosterone in human serum or plasma (lithium heparin), using the ELFA technique (Enzyme Linked Fluorescent Assay). It is intended as an aid in the diagnosis and management of conditions involving excess or deficiency of this androgen.

This device is an in vitro diagnostic medical device for professional use only.

F8. Technological Performances Summary

A summary of the technological characteristics of the new device in comparison to those of the predicate device is presented in the table below.

ItemNew DeviceVIDAS TestosteronePredicate (K021326)VIDAS Testosterone
Intended UseThe VIDAS Testosterone (TES) assay isan automated quantitative test for use onthe instruments of the VIDAS family for theenzyme immunoassay measure of totaltestosterone in human serum or plasma(lithium heparin), using the ELFAtechnique (Enzyme Linked FluorescentAssay). It is intended as an aid in thediagnosis and management of conditionsinvolving excess or deficiency of thisandrogen.VIDAS Testosterone is an automatedquantitative test for use on the instrumentsof the VIDAS family for the enzymeimmunoassay measure of totaltestosterone in human serum or plasma(lithium heparin), using the ELFA technique(Enzyme Linked Fluorescent Assay). It isintended as an aid in the diagnosis andmanagement of conditions involving excessor deficiency of this androgen.
Sample typeSerum or plasmaSame
AnalyteTestosteroneSame
Assay TechniqueEnzyme-linked fluorescent immunoassay(ELFA)Same
Kit Composition30 Strips, 1x30 SPRs, 1 Calibrator(lyophilized), 1 Control (lyophilized)Same(no change to assay).
Measuring Range0.1 - 13 ng/mlSame
AutomatedYesSame
InstrumentationVIDAS, miniVIDAS, VIDAS 3VIDAS, miniVIDAS

{33}------------------------------------------------

Image /page/33/Picture/1 description: The image shows the logo for the company bioMérieux. The logo features the company name in a sans-serif font, with the "bio" portion in a darker shade. Above the name is a stylized globe graphic with a green line running vertically through it. The globe is made up of curved lines, giving it a modern and scientific look.

G8. Performance Data

Method Comparison

A study was conducted to verify the correlation of the VIDAS Testosterone assay on the VIDAS 3 to the VIDAS Testosterone assay on the VIDAS. One reagent lot, one of each instrument and 172 serum samples were used, and results were evaluated according to CLSI EP9. "Method Comparison and Bias Estimation Using Patient Samples". The results were as follows:

AnalysisNSlopeInterceptCorrelationcoefficient
Weighted Deming1720.97990.00300.9957
Passing - Bablok1720.97510.00850.9957

Precision

Five serum samples were tested in duplicate (2 replicates) twice a day (2 runs per day) over 6 days on 1 reagent lot using 3 instruments at 1 site (N = 72). The repeatability (within-run precision), betweenrun, between-day, between calibration, between instrument, total between calibration, and total between instrument precision were calculated according to CLSI EP5-A2. "Evaluation of Precision Performance of Quantitative Measurement Methods". The results were as follows:

Sample 1Sample 2Sample 3Sample 4Sample 5
N = 72N = 72N = 72N = 72N = 72
Mean (ng/mL)Mean (ng/mL)Mean (ng/mL)Mean (ng/mL)Mean (ng/mL)
SD%CVSD%CVSD%CVSD%CVSD%CV
Within-RUN(Repeatability)0.0310.60.116.20.175.20.224.30.414.6
Between-RUN0.014.50.084.20.103.20.000.00.121.3
Between-DAY0.0000.000.00.000.00.071.40.000.0
Between-CALIBRATION0.038.10.126.90.154.50.224.30.394.3
Between-INSTRUMENT0.0411.40.000.00.051.40.173.20.151.7
Total Between-CALIBRATION0.0514.00.1810.20.257.60.326.20.586.4
Total Between-INSTRUMENT0.0618.10.1810.20.257.80.367.00.606.6

Linearity

Two sample pools were serially diluted into a total of 11 samples and tested in triplicate in order to evaluate the linearity according to CLSI EP06-A "Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach". The VIDAS Testosterone assay on VIDAS 3 is linear across the measuring range 0.1- 13 ng/mL.

{34}------------------------------------------------

Image /page/34/Picture/0 description: The image contains the logo for BIOMÉRIEUX. The logo features the company name in a stylized font, with the letters spaced apart. Above the name is a circular graphic that is bisected by a curved line, with one half of the circle being a dark blue color and the other half being a lighter shade. The curved line extends beyond the circle, adding a dynamic element to the logo.

Detection Limits

The detection limits of the VIDAS Testosterone assay on the VIDAS 3 were evaluated per CLSI EP17-A2 "Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures: Aproved Guideline, Second Edition". For LoB determination, 6 blank samples were each tested in 4 replicates in a single run per day for 4 days with 1 VIDAS 3 instrument and 2 assay lots. For LoD/LoQ determinations, 6 low-level samples were each tested in 6 replicates, 2 runs per day, for 5 days, with 1 VIDAS 3 instrument and 2 assay lots. The LoB, LoD and LoQ determinations were based consequently on 912 results, including 192 results from blank samples and 720 results from low-level samples both assay lots combined. The LoB, LoQ of the VIDAS Testosterone assay on the VIDAS 3 are as follows: LoB = 0.018 nq/mL: LoD = 0.035 ng/mL: LoQ = 0.088 ng/mL.

H8. Conclusion

The results for the performance testing submitted in this premarket notification are complete and demonstrate that the VIDAS Testosterone Assay is substantially equivalent to the predicate device identified in Item C8 of this summary. The addition of the VIDAS 3 instrument to the VIDAS Family does not affect the safety and effectiveness of the Reagent kit.

{35}------------------------------------------------

Image /page/35/Picture/0 description: The image contains the logo for bioMerieux. The logo consists of the company name in all caps, with a stylized globe above it. A green line runs vertically through the globe and the company name.

A9. Owner information

Submitter's Name:bioMérieux, Inc.
Address:595 Anglum RoadHazelwood, MO 63042
Contact Person:John Albright
Telephone number314-731-8546
Fax number314-731-8689
Date of Preparation:July 6, 2015

B9. Device Name

Trade Name:VIDAS® TSH
Common Name:TSH
Classification name:Thyroid stimulating hormone test system
Classification panel:Clinical Chemistry
Governing Regulation:21 CFR 862.1690
Device Classification:Class II
Product code:JLW

C9. Predicate Device

VIDAS® TSH (K921816)

D9. Device Description

The VIDAS TSH (TSH) assay is an enzyme-linked fluorescent immunoassay (ELFA) that is performed in an automated instrument. All assay steps and assay temperature are controlled by the instrument. A pipette tip-like disposable device, the Solid Phase Receptacle (SPR®), serves as a solid phase for the assay as well as a pipetting device. The SPR is coated at the time of manufacture with mouse monoclonal anti-TSH antibodies. The VIDAS TSH (TSH) assay configuration prevents nonspecific reactions with the SPR. Reagents for the assay are located in the sealed Reagent Strips. The sample is transferred into the well containing anti-TSH antibody conjugated with alkaline phosphatase. The sample/conjugate mixture is cycled in and out of the SPR and the TSH will bind to antibodies coated on the SPR and to the conjugate forming a "sandwich". Wash steps remove unbound conjugate. A fluorescent substrate, 4-methylumbelliferyl phosphate, is cycled through the SPR. Enzyme remaining on the SPR wall will catalyze the conversion of the substrate to the fluorescent product 4methylumbelliferone. The intensity of fluorescence is measured by the optical scanner in the instrument ; it is proportional to the TSH concentration present in the sample.

When the VIDAS TSH (TSH) assay is completed, the results are analyzed automatically by the instrument, and a report is printed for each sample.

{36}------------------------------------------------

E9. Intended Use

The VIDAS® TSH (TSH) assay is intended for use on the instruments of the VIDAS family as an automated quantitative enzyme-linked fluorescent immunoassay (ELFA) for the determination of human thyroid stimulating hormone- (TSH) concentration in human serum or plasma (heparin). It is intended for use as an aid in the diagnosis of thyroid or pituitary disorders.

This device is an in vitro diagnostic medical device for professional use only.

F9. Technological Performances Summary

A summary of the technological characteristics of the new device in comparison to those of the predicate device is presented in the table below.

ItemNew DeviceVIDAS TSHPredicate (K921816)VIDAS TSH
Intended UseThe VIDAS® TSH (TSH) assay isintended for use on the instruments ofthe VIDAS family as an automatedquantitative enzyme-linked fluorescentimmunoassay (ELFA) for thedetermination of human thyroidstimulating hormone- (TSH)concentration in human serum orplasma (heparin). It is intended for useas an aid in the diagnosis of thyroid orpituitary disorders.The VIDAS TSH (TSH) assay is intended foruse on the instruments of the VIDAS family(Vitek® ImmunoDiagnostic Assay System) asan automated quantitative enzyme-linkedfluorescent immunoassay (ELFA) for thedetermination of human thyroid stimulatinghormone- (TSH) concentration in humanserum or plasma (heparin). It is intended foruse as an aid in the diagnosis of thyroid orpituitary disorders.
Sample typeSerum or plasmaSame
AnalyteTSHSame
Assay TechniqueEnzyme-linked fluorescentimmunoassay (ELFA)Same
Kit Composition60 Strips, 2x30 SPRs, 1 Calibrator(liquid), 1 Control (liquid), 1 DiluentBuffer (liquid)Same(no change to assay).
Measuring Range0.05 – 57 µIU/ml0.05 – 60 µIU/ml
AutomatedYesSame
InstrumentationVIDAS, miniVIDAS, VIDAS 3VIDAS, miniVIDAS

{37}------------------------------------------------

Image /page/37/Picture/1 description: The image shows the logo for bioMérieux. The logo features the company name in a sans-serif font, with the "BIO" portion in a larger size. Above the name is a blue globe-like symbol with curved lines, bisected by a green and yellow curved line that resembles a plant stem. The overall design is clean and modern.

G9. Performance Data

Method Comparison

A study was conducted to verify the correlation of the VIDAS TSH assay on the VIDAS 3 to the VIDAS TSH assay on the VIDAS. One reagent lot, one of each instrument and 179 serum samples were used, and results were evaluated according to CLSI EP9. "Method Comparison and Bias Estimation Using Patient Samples". The results were as follows:

AnalysisNSlopeInterceptCorrelationcoefficient
Weighted Deming1791.00340.00420.9984
Passing Bablok1791.0028-0.00010.9984

Precision

Six serum samples were tested in duplicate (2 replicates) twice a day (2 runs per day) over 6 days on 1 reagent lot using 3 instruments at 1 site (N = 72). The repeatability (within-run precision), betweenrun, between-day, between calibration, between instrument, total between calibration, and total between instrument precision were calculated according to CLSI EP5-A2. "Evaluation of Precision Performance of Quantitative Measurement Methods". The results were as follows:

Sample 1N = 72Mean(µIU/mL)0.1Sample 2N = 72Mean(µIU/mL)0.26Sample 3N = 72Mean(µIU/mL)2.45Sample 4N = 72Mean(µIU/mL)5.1Sample 5N = 72Mean(µIU/mL)10.32Sample 6N = 72Mean(µIU/mL)35.01
SD%CVSD%CVSD%CVSD%CVSD%CVSD%CV
Within-RUN(Repeatability)0.005.10.012.50.052.00.102.00.202.00.702.0
Between-RUN0.001.70.001.20.021.00.081.60.131.30.000.0
Between-DAY0.000.00.001.70.020.80.000.00.000.00.190.5
Between-CALIBRATION0.000.00.000.60.041.50.081.60.161.60.752.1
Between-INSTRUMENT0.001.40.001.00.000.00.000.00.000.00.000.0
Total Between-CALIBRATION0.015.40.013.30.072.80.153.00.292.81.043.0
Total Between-INSTRUMENT0.015.50.013.40.072.80.153.00.292.81.043.0

Linearitv

Four sample pools were serially diluted into a total of 8 samples and tested in triplicate in order to evaluate the linearity according to CLSI EP06-A "Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach". The VIDAS TSH assay on VIDAS 3 is linear across the measuring range 0.05 – 57 µIU/mL.

{38}------------------------------------------------

Image /page/38/Picture/0 description: The image shows the logo for bioMérieux. The logo features a stylized blue and white sphere with curved lines, bisected by a green and yellow line. Below the sphere, the word "BIOMÉRIEUX" is written in a modern, sans-serif font, with the letters slightly spaced apart.

Detection Limits

The detection limits of the VIDAS TSH assay on the VIDAS 3 were evaluated per CLSI EP17-A2 "Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures: Approved Guideline, Second Edition". For LoB determination, 6 blank samples were each tested in 4 replicates in a single run per day for 4 days with 1 VIDAS 3 instrument and 2 assay lots. For LoD/LoQ determinations, 6 low-level samples were each tested in 6 replicates, 2 runs per day, for 5 days, with 1 VIDAS 3 instrument and 2 assay lots. The LoB, LoD and LoQ determinations were based consequently on 912 results, including 192 results from blank samples and 720 results from low-level samples both assay lots combined. The LoB, LoQ of the VIDAS TSH assay on the VIDAS 3 are as follows: LoB = 0.019 ulU/mL; LoD/LoQ = 0.05 µlU/mL.

H9. Conclusion

The results for the performance testing submitted in this premarket notification are complete and demonstrate that the VIDAS TSH Assay is substantially equivalent to the predicate device identified in Item C9 of this summary. The VIDAS 3 instrument to the VIDAS Family does not affect the safety and effectiveness of the Reagent kit.

{39}------------------------------------------------

Image /page/39/Picture/1 description: The image shows the logo for bioMérieux. The logo features the company name in a sans-serif font, with the letters spaced apart. Above the name is a blue circle with curved lines on the left side, and a green and yellow line running vertically through the circle and the company name.

A10. Owner information

Submitter's Name:bioMérieux, Inc.
Address:595 Anglum RoadHazelwood, MO 63042
Contact Person:John Albright
Telephone number314-731-8546
Fax number314-731-8689
Date of Preparation:July 6, 2015

B10. Device Name

Trade Name:VIDAS® D-Dimer Exclusion II
Common Name:D-Dimer
Classification name:Fibrinogen and fibrin split products, antigen, antiserum, control
Classification panel:Hematology
Governing Regulation:21 CFR 864.7320
Device Classification:Class II
Product code:DAP

C10. Predicate Device

VIDAS® D-Dimer Exclusion II (K112818)

D10. Device Description

The VIDAS D-Dimer Exclusion II (DEXII) assay is an enzyme-linked fluorescent immunoassay (ELFA) performed in an automated instrument. All assay steps and assay temperature are controlled by the instrument. A pipette tip-like disposable device, the Solid Phase Receptacle (SPR), serves as a solid phase for the assay as well as a pipetting device. The SPR is coated at the time of manufacture with anti-Fibrin degradation product (FbDP) monoclonal antibodies. The VIDAS D-Dimer Exclusion II assay configuration prevents nonspecific reactions with the SPR. Reagents for the assay are located in the sealed Reagent Strips. The sample is transferred into the well containing an alkaline phosphataselabeled anti-FbDP monoclonal antibody.

The sample/conjugate mixture is cycled in and out of the SPR and the antibodies coated on the SPR and to the conjugate forming a "sandwich". Wash steps remove unbound conjugate. A fluorescent substrate, 4-methylumbelliferyl phosphate, is cycled through the SPR. Enzyme remaining on the SPR wall will catalyze the conversion of the substrate to the fluorescent product 4-methylumbelliferone. The intensity of fluorescence is measured by the optical scanner in the instrument; it is proportional to the D-Dimer concentration present in the sample. When the VIDAS D-Dimer Exclusion II assay is completed, the results are analyzed automatically by the instrument, and a report is printed for each sample.

{40}------------------------------------------------

Image /page/40/Picture/0 description: The image contains the logo for bioMérieux. The logo features a blue globe with curved lines on the left side, and a green and yellow line extending upwards from the center. The text "BIOMÉRIEUX" is written in a sans-serif font below the globe.

E10. Intended Use

VIDAS® D-Dimer Exclusion II™ is an automated quantitative test for use on the instruments of the VIDAS family for the immunoenzymatic determination of fibrin degradation products (FbDP) in human plasma (sodium citrate, CTAD) using the ELFA technique (Enzyme Linked Fluorescent Assay). VIDAS D-Dimer Exclusion II is indicated for use in conjunction with a clinical pretest probability assessment model to exclude deep vein thrombosis (DVT) and pulmonary embolism (PE) disease in outpatients suspected of DVT or PE.

This device is an in vitro diagnostic medical device for professional use only.

F10. Technological Performances Summary

A summary of the technological characteristics of the new device in comparison to those of the predicate device is presented in the table below.

ltemNew DevicePredicate (K112818)
VIDAS D-Dimer Exclusion IIVIDAS D-Dimer Exclusion II
Intended UseVIDAS® D-Dimer Exclusion IIT™ is an automatedquantitative test for use on the instruments of thefamily for the immunoenzymaticVIDASdetermination of fibrin degradation products (FbDP)in human plasma (sodium citrate, CTAD) using theELFA technique (Enzyme Linked FluorescentAssay). VIDAS D-Dimer Exclusion II is indicated foruse in conjunction with a clinical pretest probabilityassessment model to exclude deep vein thrombosis(DVT) and pulmonary embolism (PE) disease inoutpatients suspected of DVT or PE.Same
Sample typePlasmaSame
AnalyteD-DimerSame
Assay TechniqueEnzyme-linked fluorescent immunoassay (ELFA)Same
Kit Composition60 Strips, 2x30 SPRs, 1 Calibrator (lyophilized),2 Controls (lyophilized), 1 Diluent (liquid)Same
AutomatedYesSame
InstrumentationVIDAS, miniVIDAS, VIDAS 3VIDAS, miniVIDAS

G10. Performance Data

Method Comparison

A study was conducted to verify the correlation of the VIDAS D-Dimer Exclusion II assay on the VIDAS 3 to the VIDAS D-Dimer Exclusion II assay on the VIDAS. One reagent lot, one of each instrument and 219 plasma samples were used, and results were evaluated according to CLSI EP9. "Method Comparison and Bias Estimation Using Patient Samples". The results were as follows:

AnalysisNSlopeInterceptCorrelation coefficient
Weighted Deming2190.94812.09740.9954
Passing Bablock2190.93945.11900.9954

{41}------------------------------------------------

Image /page/41/Picture/1 description: The image shows the logo for bioMérieux. The logo consists of the company name in a sans-serif font, with a stylized globe above it. A green and yellow line bisects the globe vertically. The globe is blue and has a series of curved lines on the left side.

Precision

Four plasma samples were tested in duplicate (2 replicates) twice a day (2 runs per day) over 6 days on 1 reagent lot using 3 instruments at 1 site (N = 72). The repeatability (within-run precision), between-run, between-day, between calibration, between calibration, and total between instrument precision were calculated according to CLSI EP5-A2. "Evaluation of Precision Performance of Quantitative Measurement Methods". The results were as follows:

Sample 1N = 72Mean (ng/mL)73.73Sample 2N = 72Mean (ng/mL)243.06Sample 3N = 72Mean (ng/mL)472.14Sample 4N = 72Mean (ng/mL)6939.36
SD%CVSD%CVSD%CVSD%CV
Within-RUN (Repeatability)2.363.26.662.710.042.1114.201.6
Between-RUN1.081.50.000.00.000.059.970.9
Between-DAY0.000.00.450.20.000.00.000.0
Between-CALIBRATION1.411.92.681.16.251.3139.702.0
Between-INSTRUMENT0.000.02.010.84.581.00.000.0
Total Between-CALIBRATION2.964.07.203.011.832.5190.142.7
Total Between-INSTRUMENT2.964.07.473.112.682.7190.142.7

Linearity

Two sample pools were serially diluted into a total of 11 and 15 samples and tested in triplicate in order to evaluate the linearity according to CLSI EP06-A "Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach". The VIDAS D-Dimer Exclusion II assay on VIDAS 3 is linear across the measuring range 45 - 10000 ng/mL.

Detection Limits

The detection limits of the VIDAS D-Dimer Exclusion II assay on the VIDAS 3 were evaluated per CLSI EP17-A2 "Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline, Second Edition". For LoB determination, 6 blank samples were each tested in 4 replicates in a single run per day for 4 days with 1 VIDAS 3 instrument and 2 assay lots. For LoD/LoQ determinations, 6 low-level samples were each tested in 6 replicates, 2 runs per day, for 5 days, with 1 VIDAS 3 instrument and 2 assay lots. The LoB, LoD and LoQ determinations were based consequently on 912 results, including 192 results from blank samples and 720 results from low-level samples both assay lots combined. The LoB. LoD. LoQ of the VIDAS D-Dimer Exclusion II assay on the VIDAS 3 are as follows: LoB = 10.562 ng/mL; LoD = 16.251 ng/mL; LoQ = 17.229 ng/mL.

H10. Conclusion

The results for the performance testing submitted in this premarket notification are complete and demonstrate that the VIDAS D-Dimer Exclusion II Assay is substantially equivalent to the predicate device identified in Item C10 of this summary. The addition of the VIDAS 3 instrument to the VIDAS Family does not affect the safety and effectiveness of the Reagent kit.

§ 866.3110

Campylobacter fetus serological reagents.(a)
Identification. Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identifyCampylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases.Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.(b)
Classification. Class I (general controls).