K Number
K141128

Validate with FDA (Live)

Date Cleared
2014-08-13

(104 days)

Product Code
Regulation Number
870.2700
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Fingertip Pulse Oximeter MD300CL37 is a portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, adolescent, child and infant patients in hospitals, hospital-type facilities, and home environments.

Device Description

The proposed device of Fingertip Pulse Oximeter MD300CL37 is a battery powered fingertip device, which can detect, display the measured %SpO2 and pulse rate value. The device is normally applied to adult, adolescent, child and infant patient in hospitals, hospital-type facilities, and home environments.

The proposed device consists of power supply module, detector and emitter LED, signal collection and process module, display module and user interface.

The Pulse Oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light; the other is 940nm, which is ultra red light. Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photodetector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.

The power sources of the proposed device are 2 AAA alkaline batteries. The proposed device have low battery voltage indicator function and the proposed device will automatically power off when there is no signal for longer than 8 seconds. When the finger is placed into the device, the device will be powered on automatically.

The proposed device is not for life-supporting or life-sustaining, not for implant. The device or transducers are not sterile and the transducer is reusable and does not need sterilization or re-sterilization. The device is for prescription. The device does not contain drug or biological products.

The device is software-driven and the software validation is provided in Section of Software.

AI/ML Overview

The document provided describes the Beijing Choice Electronic Technology Co., Ltd. Fingertip Pulse Oximeter MD300CL37 and its substantial equivalence to a predicate device. Here's a breakdown of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

Feature/MetricAcceptance Criteria (Predicate Device)Reported Device Performance (MD300CL37)
SpO2 Accuracy70-99%: ±3%70-99%: ±3%
0-69%: Undefined0-69%: Undefined
SpO2 Display Range0-99%0-99%
SpO2 Measurement Range70-99%70-99%
SpO2 Resolution1%1%
PR Display Range0-254 BPM0-235 BPM
PR Measurement Range30-235 BPM30-235 BPM
PR Accuracy30-99bpm: ±2bpm30-99bpm: ±2bpm
100-235bpm: ±2%100-235bpm: ±2%
PR Resolution1bpm1bpm
Operating Temperature5-40°C5-40°C
Storage Temperature-20-55°C-25-70°C
Relative Humidity≤80% in operation15%-93% in operation
≤93% in storage≤93% in storage
Atmosphere Pressure86kPa-106kPa86kPa-106kPa
Low Power Beep Tip FunctionYesNo
Automatically Powered OnNoYes
Biocompatibility (Fingertip Cushion)No cytotoxic potential, No sensitization, No significant sensitization from test extract to rabbits (for both ABS and Black Medical Silicon gel)No cytotoxic potential, No sensitization, No significant sensitization from test extract to rabbits (for Laser etching medical silicone gel)

Note: While the document lists "Acceptance Criteria (Predicate Device)" as directly matching the predicate's performance, the primary acceptance criteria for the new device would be to demonstrate performance comparable to the predicate within accepted margins, often specified by standards like ISO 80601-2-61.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 12 healthy adult volunteer subjects.
  • Data Provenance: The study was conducted in China ("Yue Bei People's hospital" is mentioned as the location of the clinical test, which is a hospital in China). It was a prospective study where subjects were intentionally exposed to induced hypoxia.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number of experts or their qualifications directly involved in establishing the ground truth. It states that "Arterial blood samples were drawn during simultaneous data collection from the test devices. The blood was immediately analyzed on Reference CO-Oximetry providing functional SaO2 for the basis of the SpO2 accuracy comparison." This implies that the ground truth was established by direct physiological measurement using a reference CO-Oximetry device, which is a gold standard for blood oxygen saturation. The expertise would lie in the laboratory personnel operating the CO-Oximetry and the medical staff conducting the induced hypoxia and blood draws, though specific qualifications are not detailed.

4. Adjudication Method for the Test Set

The document does not explicitly describe an adjudication method for the test set. The ground truth was established by direct comparison to a reference CO-Oximetry device. This is a direct measurement comparison, rather than an expert consensus that would typically require adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The study focused on the device's standalone accuracy against a reference measurement.

6. Standalone Performance

Yes, a standalone (algorithm only without human-in-the-loop performance) study was done for the device's accuracy. The clinical test directly compared the device's SpO2 readings to functional SaO2 measured by reference CO-Oximetry.

7. Type of Ground Truth Used

The ground truth used was physiological measurement obtained from Reference CO-Oximetry of arterial blood samples.

8. Sample Size for the Training Set

The document does not specify a sample size for a "training set." This type of premarket notification (510(k)) primarily demonstrates substantial equivalence to a legally marketed predicate device, often relying on established principles and clinical validation rather than a deep learning model's training set. The clinical study described is for validation.

9. How the Ground Truth for the Training Set Was Established

As no specific training set is discussed for an AI/algorithm in the context of this device's submission, how its ground truth was established is not applicable or detailed in the provided document. The device is a traditional sensor-based medical device, not a machine learning algorithm requiring a separate training phase in the AI sense.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with three profiles facing right, suggesting a sense of forward movement and progress. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 13, 2014

Beijing Choice Electronic Technology Co., Ltd. Mr. Lei Chen North Building 3F, No.9 Shuangyuan Road, Badachu Hi-tech Zone, Shijingshan District, Beijing, P.R. China, 100041

Re: K141128

Trade/Device Name: Fingertip Pulse Oximeter MD300CL37 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: May 13, 2014 Received: May 15, 2014

Dear Mr. Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Chen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mary S. Runner -S

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Section II Indications for Use Statement

Indications for Use

510(k) Number (if known): Device Name: Fingertip Pulse Oximeter MD300CL37

Indications for Use:

The Fingertip Pulse Oximeter MD300CL37 is a portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, adolescent, child and infant patients in hospitals, hospital-type facilities, and home environments.

Prescription Use

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

Section III 510(k) Summary

This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92

There is no prior submission for the device.

3.1Submitter Information

● Manufacturer Name:

Establishment Registration Number: 3005569927 Beijing Choice Electronic Technology Co., Ltd. Room 320,West Building 4, No.83 FuXing Road Beijing 100039, P.R.China

● Contact Person:

Mr.Lei Chen Beijing Choice Electronic Technology Co., Ltd. North Building 3F, No. 9 Shuangyuan Road, Badachu Hi-tech Zone, Shijingshan District Beijing China 100041 Phone: +86-10-88798300 Ext 6020 Fax:215-4052545 Email: cc@choicemmed.com

● Date prepared: Mar.25, 2014

3.2Proposed Device Information

Device Common Name: Pulse Oximeter Device Trade/Proprietary Name: Fingertip Pulse Oximeter Model: MD300CL37 Classification Name: Oximeter Regulation Number: 870.2700 Product Code: DQA Class: II Panel: Anesthesiology

3.3Predicate Device

510(k) Number: K123871 Common Name: Pulse Oximeter

{4}------------------------------------------------

Page 2 of 10

Premarket Notification 510(k) Submission—Section III 510(k) Summary Device Trade/Proprietary Name: Fingertip Pulse Oximeter Model: MD300CF315 Classification Name: Oximeter Product Code: DQA Regulation Number: 870.2700 Device Class: II Panel: Anesthesiology Manufacturer: Beijing Choice Electronic Technology Co., Ltd.

3.4Device Description

The proposed device of Fingertip Pulse Oximeter MD300CL37 is a battery powered fingertip device, which can detect, display the measured %SpO2 and pulse rate value. The device is normally applied to adult, adolescent, child and infant patient in hospitals, hospital-type facilities, and home environments.

The proposed device consists of power supply module, detector and emitter LED, signal collection and process module, display module and user interface.

The Pulse Oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light; the other is 940nm, which is ultra red light. Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photodetector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.

The power sources of the proposed device are 2 AAA alkaline batteries. The proposed device have low battery voltage indicator function and the proposed device will automatically power off when there is no signal for longer than 8 seconds. When the finger is placed into the device, the device will be powered on automatically.

The proposed device is not for life-supporting or life-sustaining, not for implant. The device or transducers are not sterile and the transducer is reusable and does not need sterilization or re-sterilization. The device is for prescription. The device does not contain drug or biological products.

The device is software-driven and the software validation is provided in Section of Software.

{5}------------------------------------------------

Page 3 of 10

3.5 Comparison list of the technological characteristics

Comparison ElementsProposed DevicePredicate Device
Product NameFingertip Pulse OximeterFingertip Pulse Oximeter(K123871)
ModelMD300CL37MD300CF315
Regulation No.21 CFR 870.270021 CFR 870.2700
ClassificationIIII
Classification NameOximeterOximeter
Product CodeDQADQA
Indications for UseThe fingertip pulse oximeter is a portable non-invasive device intended for spot-checking of arterial hemoglobin oxygen saturation(SpO2) and pulse rate of adult, adolescent, child and infant patient in hospitals, hospital-type facilities, and home environments.The Fingertip Pulse Oximeter is a portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patient at home, and hospital (including clinical use in internist/surgery, Anesthesia, intensive care etc).
Comparison StatementThe proposed device has the same indications for use and classification as the predicate device.
device.
ComponentsThe proposed device consists of detector and emitter LED, signal amplify unit, CPU, data display unit and power unitpower supply module, detector and emitter LED, signal collection and process module, display module, user interface, voice module and button control circuit.
Design PrincipleThe pulse oximeter works by applying aThe pulse oximeter works by applying a
sensor to a pulsating arteriolar vascular bed.The sensor contains a dual light source andphoto detector. The one wavelength of lightsource is 660 nm, which is red light; the otheris 940 nm, which is ultra red light. Skin, bone,tissue, and venous vessels normally absorb aconstant amount of light over time. The photodetector in finger sensor collects and convertsthe light into electronic signal which isproportional to the light intensity. Thearteriolar bed normally pulsates and absorbsvariable amounts of light during systole anddiastole, as blood volume increases anddecreases. The ratio of light absorbed atsystole and diastole is translated into anoxygen saturation measurement. Thismeasurement is referred to as SpO2.sensor to a pulsating arteriolar vascular bed.The sensor contains a dual light source andphoto detector. The one wavelength of lightsource is 660 nm, which is red light; theother is 940 nm, which is ultra red light. Skin, bone, tissue, and venous vesselsnormally absorb a constant amount of lightover time. The photo detector in fingersensor collects and converts the light intoelectronic signal which is proportional to thelight intensity. The arteriolar bed normallypulsates and absorbs variable amounts oflight during systole and diastole, as bloodvolume increases and decreases. The ratio oflight absorbed at systole and diastole istranslated into an oxygen saturationmeasurement. This measurement is referredto as SpO2.
MeasurementWavelengthRed660nm660nm
Infrared940nm940nm
Comparison StatementThe proposed device has the same design and similar components as the predicatedevice.
DevicespecificationDisplay TypeOLED
Power Supply2 AAA-size alkaline batteries
Display DataSpO2 , PR
SpO2display0~99%
range
Measurementrange70-99%70~99%
Accuracy70%~99%: ±3%;0%~69% no definition70%~99%: ±3%;0%~69% no definition
Resolution1%1%
PR display range0~235BPM0~254BPM
PR measurementrange30-235 BPM30-235 BPM
PR Accuracy3099bpm, ±2bpm; 100235bpm, ±2%3099bpm, ±2bpm; 100235bpm,±2%
Resolution1bpm1bpm
OperatingTemperature5~40°C5~40°C
StorageTemperature-25~+70°C-20~55°C
RelativeHumidity15%~93% in operation≤93% in storage≤80% in operation≤93% in storage
Atmospherepressure86kPa~106kPa86kPa~106kPa
Low PowerBeep TipFunctionNoYes
Automaticallypowered onYesNo
Comparison StatementThe applicant device has similar device specifications as the predicate device
ConstructionBattery CoverABSABS
MaterialsFingertipCushionLaser etching medical silicone gelBlack Medical Silicon gel
EncloseABSABS
Comparison StatementThe materials used in the applicant device are similar as the predicated device except forthe material used in the fingertip cushion. The test report was presented inBiocompatibility.
Performance TestingBench TestThe bench tests include Pulse Rate and SpO2accuracy test after 3000 cycles disinfection,Weak Perfusion Test, High and LowTemperature & Humidity Test andISO80601-2-61.Meet the requirements of FDA Guidance
Clinical TestConformed to ISO 80601-2-61Clinical test for device accuracy is conductedby the Yue Bei People's hospital. The clinicaltest report and protocol are provided inPerformance Testing-ClinicalConformed to ISO 80601-2-61
Electrical SafetyConformed to IEC60601-1Conformed to IEC60601-1
Electromagnetic compatibilityand safetyElectromagnetic CompatibilityConformed to IEC60601-1-2Conformed to IEC60601-1-2
SoftwareModerate Level of ConcernModerate Level of Concern

Table 3-1 Performance Specification Comparison

{6}------------------------------------------------

Page 4 of 10

Premarket Notification 510(k) Submission—Section III 510(k) Summary

{7}------------------------------------------------

Page 5 of 10

Premarket Notification 510(k) Submission—Section III 510(k) Summary

{8}------------------------------------------------

Page 6 of 10

Premarket Notification 510(k) Submission—Section III 510(k) Summary

{9}------------------------------------------------

Page 7 of 10


Premarket Notification 510(k) Submission—Section III 510(k) Summary

Compliance with FDA Guidance for theContent of Premarket Submissions forSoftware Contained in Medical Devices.Compliance with FDA Guidance for theContent of Premarket Submissions forSoftware Contained in Medical Devices.
Risk Management in Compliance withISO:14971:2007Risk Management in Compliance withISO:14971:2007
BiocompatibilityMedical Silicon gelIn Vitro CytotoxicityNo cytotoxic potential.In Vitro CytotoxicityNo cytotoxic potential.
Skin Irritation TestNo evidence of causing sensitization.Skin Irritation TestNo evidence of causing sensitization.
Animal Skin Sensitization TestNo evidence of significant sensitization from the test extract to rabbits.Animal Skin Sensitization TestNo evidence of significant sensitization from the test extract to rabbits.
ABS plastic EnclosureIn Vitro CytotoxicityNo cytotoxic potential.In Vitro CytotoxicityNo cytotoxic potential.
Skin Irritation TestNo evidence of causing sensitization.Skin Irritation TestNo evidence of causing sensitization.
Animal Skin Sensitization TestNo evidence of significant sensitization from the test extract to rabbits.Animal Skin Sensitization TestNo evidence of significant sensitization from the test extract to rabbits.
Label and LabelingCompliance with FDA guidanceCompliance with FDA guidance

{10}------------------------------------------------

Premarket Notification 510(k) Submission—Section III 510(k) Summary

3.6Intended use

The MD300CL37 Fingertip pulse oximeter is a portable non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, adolescent, child and infant patients in hospitals, hospital-type facilities, and home environments.

3.7 Test

Non-clinical Test

The Fingertip Pulse Oximeter MD300CL37 is designed and tested and will be manufactured in accordance with both mandatory and voluntary standards, including:

IEC 60601-1:2005 Medical Electrical Equipment-Part 1: General requirements for safety.

IEC60601-1-2:2007 Medical Electrical Equipment-Part 1-2:General requirements for safety, Collateral standard: Electromagnetic compatibility -Requirements and tests.

IEC60601-1-11:2010 Medical electrical equipment-Part1-11:General requirements for basic safety and essential performance-Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.

ISO 80601-2-61:2011 Medical Electrical Equipment-Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use.

The Software Validation is in compliance with FDA Guidance to Compliance with FDA

Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.

The compatibility of the skin-contact component material in the finished product meets the requirement of Biocompatibility. The Biological Evaluation Tests are in compliance with the standards of ISO 10993, "Biological Evaluation of Medical Devices".

No.Test Name
1Broad-band Vibration Test( The test result is presented in the test report ofISO80601-2-61)
2Pulse rate and SpO2 Accuracy Test after 3000 cycles' disinfection
3System Performance Test
4Shelf Life Test

The list of non-clinical test performed on the proposed device.

{11}------------------------------------------------

5Performance Test according to ISO 80601-2-61
6Electromagnetic Compatibility Test According to IEC 60601-1-2
7Electrical Safety Test According to IEC 60601-1
8Irritation ,Sensitization and Cytotoxicity Test according to ISO 10993
9Used in the Home Healthcare Environment Test According to IEC60601-1-11

Premarket Notification 510(k) Submission--Section III 510(k) Summary

The test results indicate that the safety and effectiveness of the proposed devices is identical to that of the predicate device.

Clinical Test

The Clinical Test reports was conducted in accordance to EN ISO 14155-1:2009, EN ISO 14155-2:2009, ISO 9919:2005, EN ISO 9919:2009, BS EN ISO 80601-2-61:2011, and the FDA Guidance Document for Pulse Oximeters.

Subjects:

After Institutional Review Board (IRB) approval, 12 healthy adult volunteer subjects (ages 21-33yr, 46-75kg, 158-183cm, with light to dark pigmentation) where included in the study conducted July.5-6, 2013 to evaluate the SpO2 accuracy performance of the MD300CL37 Finger Pulse Oximeter.

Methods:

Each system was evaluated during steady state/non-motion conditions with various levels of induced hypoxia resulting in stable oxygen saturation levels between 100% and 70% SaO2. Arterial blood samples were drawn during simultaneous data collection from the test devices. The blood was immediately analyzed on Reference CO-Oximetry providing functional SaO2 for the basis of the SpO2 accuracy comparison.

Adverse events and complications:

There are no adverse events during the clinical test.

Conclusions:

The results show the MD300CL37 Finger Pulse Oximeter to pass a SpO2 accuracy specification of 3 which Beijing Choice Electronic Tech Co., Ltd. claims during steady state conditions over the range of 70-100%.

3.8Determination of substantial equivalence

The proposed device of Fingertip Pulse Oximeter MD300CL37 has the same classification information, same indications and intended use, same design principle, similar product design and specifications, same performance performance effectiveness, performance

{12}------------------------------------------------

Premarket Notification 510(k) Submission—Section III 510(k) Summary

safety as the predicated device. The differences only exist audible indicator and low power beep tip function. These differences are slight and do not influence the effectiveness and safety of the device. According to the non-clinical test results, the proposed device is as safe as effective and perform as well as the predicate device. So the proposed device is Substantially Equivalent (SE) to the predicate device which is US legally market device.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).