K Number
K140988

Validate with FDA (Live)

Date Cleared
2014-08-12

(117 days)

Product Code
Regulation Number
878.4460
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The powder free sterile latex surgical glove, yellow color (Brand Name: Titanfine), is a disposable device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

Device Description

Powder Free Sterile Latex Surgical Gloves, Yellow Color (Brand Name: Titanfine) are made of natural rubber, and are sterile that meets all of the requirements of ASTM standard D 3577-2009. They have yellow color.

AI/ML Overview

This document describes the premarket notification for Powder Free Sterile Latex Surgical Gloves, Yellow Color (Brand Name: Titanfine). The study proves the device meets the acceptance criteria by demonstrating its compliance with various ASTM standards and ISO 10993-10 for physical, biocompatibility, and performance testing.

1. Table of Acceptance Criteria and Reported Device Performance

DescriptionAcceptance Criteria (Standard)Reported Device Performance (Titanfine)
DimensionMeets ASTM D 3577-09Meets ASTM D 3577-09
Physical PropertyMeets ASTM D 3577-09Meets ASTM D 3577-09
Free of PinholeMeets ASTM D5151 (AQL 1.5)Meets ASTM D5151 (AQL 1.5)
Residue PowderMeets ASTM D6124Meets ASTM D6124
Water Soluble ProteinMeets ASTM D5712-10Meets ASTM D5712-10 (Limit: 200 µg/dm²)
Primary Skin IrritationISO 10993-10: 2010Not a primary skin irritant
Dermal SensitizationISO 10993-10: 2010Not a dermal sensitizer

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes used for each individual test or the specific data provenance (e.g., country of origin, retrospective or prospective nature) for the testing. It generally states that "Tests were conducted following the recommended procedures outlined in the respective consensus standards."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The ground truth for this device is based on compliance with established technical standards (ASTM and ISO), not expert consensus in a diagnostic sense.

4. Adjudication method for the test set

Not applicable. The test results are pass/fail based on meeting the specified criteria of the referenced standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical device (surgical gloves), not an AI/imaging diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is a medical device (surgical gloves), not an algorithmic device.

7. The type of ground truth used

The ground truth used for demonstrating compliance is based on established industry standards and physical/chemical testing results. Specifically:

  • Physical properties and dimensions: ASTM D3577-09
  • Freedom from pinholes: ASTM D5151 (AQL 1.5)
  • Residual powder: ASTM D6124
  • Water-soluble protein: ASTM D5712-10
  • Biocompatibility (primary skin irritation and dermal sensitization): ISO 10993-10: 2010

8. The sample size for the training set

Not applicable. This is a medical device (surgical gloves), not an AI/machine learning device that requires a training set. The "training" in this context refers to the manufacturing process ensuring consistency with quality standards.

9. How the ground truth for the training set was established

Not applicable. As this is not an AI/machine learning device, there is no "training set ground truth" in that sense. The manufacturing process is designed to produce gloves consistently meeting the established standards.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo features a stylized image of a human face in profile, with three overlapping profiles suggesting a sense of community and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 12, 2014

Hebei HongSen Plastics Technology Company, Limited C/O Mr. Charles Shen Official Correspondent Manton Business and Technology Services 5 Carey Street Pennington, New Jersey 08534

Re: K140988

Trade/Device Name: Powder Free Sterile Latex Surgical Gloves, Yellow Color (Brand Name: Titanfine) Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: I Product Code: KGO Dated: July 11, 2014 Received: July 15, 2014

Dear Mr. Shen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Shen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mary S. Runner -S

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K140988

Device Name

Powder Free Sterile Latex Surgical Gloves, Yellow Color (Brand Name: Titanfine)

Indications for Use (Describe)

The powder free sterile latex surgical glove, yellow color (Brand Name: Titanfine), is a disposable device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

Type of Use (Select one or both, as applicable)

  • Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5: 510(k) Summary:

This summary of 510k safety and effectiveness information is being submitted In accordance with the requirements of 21CFR 807.92

5.1 Submitter & Foreign Manufacture Identification

Hebei HongSen Plastics Technology Co., Ltd The Eastern Industrial Accumulation Area Nangong Town, Xingtai City, Hebei Province, China Tel: 86- 311-85656588 Submitter's FDA Registration Number: 3010582952

5.2 US Agent and Contact Person

Charles Shen Manton Business and Technology Services 5 Carey Street Pennington, NJ 08534 Tel: 608-217-9358 Email: cyshen@aol.com

5.3 Date of Summary: March 15, 2014

5.4Device Names:
--------------------
Proprietary Name:Powder Free Sterile Latex Surgical Gloves, Yellow Color(Brand Name: Titanfine)
Common Name:Surgeon's glove
Classification Name:Surgeon's glove
Device Classification:I
Regulation Number:21 CFR 878.4460
Panel: GeneralGeneral & Plastic Surgery
Product Code:KGO

ર્સ્ટ Predicate Device Information:

(1) K121897, "Protexis' Latex Basic, Sterile Latex Powder-Free Surgical Glove with Protein Content Label Claim of 50ug/dm² or less (Cream)", manufactured by "Cardinal Health"

5.6 Device description:

Powder Free Sterile Latex Surgical Gloves, Yellow Color (Brand Name: Titanfine) are made of natural rubber, and are sterile that meets all of the requirements of ASTM standard D 3577-2009. They have yellow color.

{4}------------------------------------------------

5.7 Intended Use:

The powder free sterile latex surgical glove, yellow color (Brand Name: Titanfine) is a disposable device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

న్ని ని Comparison to Predicate Devices

The powder free sterile latex surgical gloves, yellow, are compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance.

K121897, "Protexis' Latex Basic, Sterile Latex Powder-Free Surgical Glove with Protein Content Label Claim of 50ug/dm² or less (Cream)", manufactured by "Cardinal Health"

The following table shows similarities and differences of use, design, and material between our device and the predicate devices.

DescriptionOur DevicePredicate Device (K121897)
Indication for UseThe powder free sterile latex surgical glove, yellow color (Brand Name: Titanfine) is a disposable device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.A powder-free sterile surgeon's glove is a disposable device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
Basic DesignA garment covering the hand and waist area. Gloves have separate sheaths or openings for each finger and the thumb.Same
MaterialsNatural rubberSame
Size6, 6.5, 7, 7.5, 8, 8.5, 9Information not available
Single UseYesYes
ColorYellowCream
SterileSterile (Irradiation)Sterile

Table 5.1: Comparison of Intended Use, Design, and Material

Our device is essentially identical to the predicate device in terms of indications for use, design, and material between our device and the predicate devices. The only minor difference is that the predicate device has cream color, while our device in submission has yellow color.

{5}------------------------------------------------

The following table shows similarities and differences of the performance between our device and the predicate devices. One minor difference is that predicate device has an extractable protein limit of 50 µg/dm², where our device has a limit of 200 µg/dm². This minor difference does not impact the safety of the device as 200 µg/dm² is still a well accepted safe limit.

Tests were conducted following the recommended procedures outlined in the respective consensus standards, and results for Powder Free Sterile Latex Surgical Gloves, Yellow Color (Brand Name: Titanfine), manufactured by "Hebei HongSen Plastics Technology Co., Ltd." met all relevant requirements in the test standards, and are comparable to the predicate device.

DescriptionOur DevicePredicate Device (K121897)
DimensionMeets ASTM D 3577-09(Published 02/1/2009)Meets ASTM D 3577(Publication date unknown)
Physical PropertyMeets ASTM D 3577-09(Published 02/1/2009)Meets ASTM D 3577(Publication date unknown)
Free of PinholeMeets ASTM D5151 (AQL1.5)(Published 08/20/2012)Meets ASTM D5151(Publication date unknown)
Residue PowderMeets ASTM D6124(Published 08/20/2012)Meets ASTM D6124(Publication date unknown)
Water Soluble ProteinMeets ASTM D5712-10(Published 02/01/2009)Meets ASTM D5712(Publication date unknown)
Primary Skin Irritation(ISO 10993-10: 2010)Not a primary skin irritant underthe conditions of the studySame
Dermal sensitization(ISO 10993-10: 2010)Not a dermal sensitizer under theconditions of the studySame
Table 5.2: Comparison of Physical, Biocompatibility and Performance Testing

5.9 A brief discussion of the non-clinical testing data that was submitted, referenced or relied on to demonstrate that the Subject Device is safe and effective, and whose performance meets the requirements of its user-defined acceptance criteria and intended uses:

Powder Free Sterile Latex Surgical Gloves, Yellow Color (Brand Name: Titanfine) meet requirements per ASTM D3577-09, ASTM D6124-06, ASTM D 5151-06, ASTM D5712-10 and ISO 10993-10. It is safe and effective, and its performance meets the requirements of its pre-defined acceptance criteria and intended uses.

{6}------------------------------------------------

A brief discussion of the clinical submitted, reference, or relied on in the 5.10 premarket notification submission for a determination of substantial equivalence.

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

Substantial Equivalent Conclusions 5.11

Based on the comparison of intended use, design, materials, and performance, our Powder Free Sterile Latex Surgical Gloves, Yellow Color (Brand Name: Titanfine) are substantial equivalent to its predicate devices.

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).