K Number
K140920

Validate with FDA (Live)

Date Cleared
2014-12-03

(237 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

syngo. CT Colonography is used for easy-to-perform and efficient inspection of the colonic surface. It facilitates the search and diagnosis of colon lesions. The workflow management ensures that the required data and tools are offered to you according to your role and task.

syngo. CT Colonography is a clinical post-processing workflow for basic virtual colonoscopy. It is designed to support the following image reconstruction techniques:

  • · Multiplanar Reconstruction (MPR)
  • · Volume Rendering Technique (VRT)
  • · Perspective surface shaded display (pSSD)
  • The following evaluation tools are provided with this workflow:
  • Virtual Flight
  • · Panoramic View
  • Polyp Lens
  • Stool Tagging Stool Subtraction
  • Polyp Enhanced Viewing (PEV)
  • Movie

syngo.CT Colonography supports reporting with appropriate reporting tools such as lesion location, lesion characterization, and key image creation. Commercially available digital image processing tools with an optimized workflow and reporting tools, the software is designed to support the physician on confirming the presence or absence of physician identified colon lesions (for example, polyps) in addition to evaluation, and follow-up of any such lesions using standard or low-dose spiral CT scanning.

Device Description

syngo.CT Colonography is a self-contained image analysis software package for evaluating CT volume data sets. The software is designed to support the physician in studying the inside, wall, and outside of the colon. It helps the physician to classify conspicuous regions of tissue with respect to their size, shape, and position.

Therefore, it combines tools for

  • visualization of the data: ●
    • Panoramic Endoluminal View a)
    • Stool Tagging b)
    • Workflow Improvements C)
    • Virtual Dissection d)
    • e) Polyp Lens
    • Stool Subtraction f)
  • navigation through the CT volume data sets ●
  • evaluation of potential findings: ●
    • a) Semi-automatic measurements
  • reporting of results:
    • a)
AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for syngo.CT Colonography based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state specific numerical acceptance criteria for clinical performance (e.g., sensitivity, specificity, accuracy for polyp detection). Instead, it relies on demonstrating that the device functions as intended and supports the physician in evaluating colon lesions.

The document refers to non-clinical and clinical performance tests to ensure the device fulfills requirements. The reported device performance is described in a general statement rather than specific metrics against a pre-defined threshold.

Acceptance Criteria (Implied)Reported Device Performance
Functionality of features for inspecting colonic surface and colon lesions"The result of all conducted testing was found acceptable to support the claim of substantial equivalence."
Ability to support image reconstruction techniques (MPR, VRT, pSSD)Device is designed to support these techniques.
Provision of evaluation tools (Virtual Flight, Panoramic View, Polyp Lens, Stool Tagging, Stool Subtraction, Polyp Enhanced Viewing, Movie)These tools are provided with the workflow.
Support for reporting tools (lesion location, characterization, key image creation)Device supports reporting with these tools.
Ability to assist physicians in confirming presence/absence of colon lesions, evaluation, and follow-up"The software is designed to support the physician on confirming the presence or absence of physician identified colon lesions... in addition to evaluation, and follow-up."
Conformance with safety and performance standards (DICOM, IEC 62304, ISO 14971, IEC 60601-1-6, IEC 60601-1-4)Device is designed to fulfill these requirements. Verification/validation testing supported modifications. Risk analysis completed and controls implemented.
Software specifications met acceptance criteria (Nonclinical Testing)"The testing results supports that all the software specifications have met the acceptance criteria."
Clinical performance acceptable across a variety of clinical situations"Testing was provided to cover a variety of clinical situations that would be seen in daily clinical use of the subject device."

2. Sample Size Used for the Test Set and Data Provenance

The document does not provide specific details on the sample size used for the clinical test set. It states:

  • "Clinical performance tests were conducted to demonstrate performance, safety, and effectiveness of syngo.CT Colonography."
  • "Testing was provided to cover a variety of clinical situations that would be seen in daily clinical use of the subject device."

The data provenance (country of origin, retrospective/prospective) is not specified.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not provide information on the number of experts used or their qualifications for establishing ground truth in the clinical testing.

4. Adjudication Method for the Test Set

The document does not describe any adjudication method used for the test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

The document does not mention or describe a Multi-Reader Multi-Case (MRMC) comparative effectiveness study. Therefore, no effect size of human reader improvement with AI assistance is provided.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The document notes that syngo.CT Colonography is "designed to support the physician" and assists in "confirming the presence or absence of physician identified colon lesions." This implies a human-in-the-loop scenario. While "Performance tests were conducted to test the functionality of the syngo.CT Colonography post processing application," which might involve some standalone functional testing of its features, the document does not explicitly present data from a standalone algorithm-only performance study. The overall context is that of a tool assisting a physician.

7. The Type of Ground Truth Used

The document does not explicitly state the type of ground truth used (e.g., expert consensus, pathology, outcomes data). Given the nature of a colonography device, it is highly probable that a combination of expert review (e.g., endoscopist review, radiology consensus) and potentially pathology reports for confirmed lesions would be used, but this is not confirmed in the text.

8. The Sample Size for the Training Set

The document does not provide any information regarding the sample size of a training set. This is not surprising as the document primarily focuses on demonstrating substantial equivalence to a predicate device, and the details of model training are typically not elaborated in such 510(k) summaries unless there are novel AI capabilities requiring specific validation.

9. How the Ground Truth for the Training Set Was Established

Since no training set information is provided, the document does not describe how ground truth for a training set was established.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines extending from the bottom of the profiles.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Siemens Medical Solutions USA, Inc. % Ms. Kimberly Mangum Regulatory Affairs Specialist 51 Valley Stream Parkway MALVEN PA 19355

December 3, 2014

Re: K140920

Trade/Device Name: syngo.CT Colonography Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: NWE Dated: October 31, 2014 Received: November 04, 2014

Dear Ms. Mangum:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A Ocks

for

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K140920

Device Name syngo.CT Colonography

Indications for Use (Describe)

syngo. CT Colonography is used for easy-to-perform and efficient inspection of the colonic surface. It facilitates the search and diagnosis of colon lesions. The workflow management ensures that the required data and tools are offered to you according to your role and task.

syngo. CT Colonography is a clinical post-processing workflow for basic virtual colonoscopy. It is designed to support the following image reconstruction techniques:

  • · Multiplanar Reconstruction (MPR)
  • · Volume Rendering Technique (VRT)
  • · Perspective surface shaded display (pSSD)
  • The following evaluation tools are provided with this workflow:
  • Virtual Flight
  • · Panoramic View
  • Polyp Lens
  • Stool Tagging Stool Subtraction
  • Polyp Enhanced Viewing (PEV)
  • Movie

syngo.CT Colonography supports reporting with appropriate reporting tools such as lesion location, lesion characterization, and key image creation. Commercially available digital image processing tools with an optimized workflow and reporting tools, the software is designed to support the physician on confirming the presence or absence of physician identified colon lesions (for example, polyps) in addition to evaluation, and follow-up of any such lesions using standard or low-dose spiral CT scanning.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The letters are all capitalized and are a light blue color. The word is centered and takes up most of the frame.

510(K) SUMMARY FOR syngo.CT Colonography Submitted by: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355 Date Prepared: October 30, 2014

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

    1. General Information
      Importer/Distributor:

Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355

Establishment Registration Number: 2240869

Manufacturing Site: Siemens AG Medical Solutions Henkestraße 127 D-91052 Erlangen, Germany

Establishment Registration Number: 3002808157

2. Contact Person:

Mrs. Kimberly Mangum Technical Specialist, Requlatory Affairs Submissions Siemens Medical Solutions, Inc. USA 51 Valley Stream Parkway D02 Malvern, PA 19355-1406 Phone: (610) 448-1772 Fax: (610) 448-1778 Email: kimberly.mangum@siemens.com

3. Device Name and Classification

Product Name:syngo.CT Colonography
Propriety Trade Name:syngo.CT Colonography
Classification Name:Colon Computed Tomography System,
Computer Aided Detection
Classification Panel:Radiology
CFR Section:21 CFR §892.2050
Device Class:Class II

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The word is written in all capital letters. The color of the text is a light blue or teal color.

Product Code:NWE
Legally Marketed Predicate Devices
Trade Name:syngo Colonography – Software Packagewith Extended Functionality
510(k):K042605
Clearance Date:October 08, 2004
Classification Name:Colon Computed Tomography System,Computer Aided Detection
Classification Panel:Radiology
Classification Regulation:21 CFR § 892.2050
Device Class:II
Product Code:NWE

Substantial Equivalence 4.

The subject device syngo.CT Colonography is substantially equivalent to following medical devices in commercial distribution:

Predicate Device NameFDA ClearanceNumberFDA Clearance Date
Colonography Software Packagewith extended functionalityK04260510/08/2004

5. Device Description

syngo.CT Colonography is a self-contained image analysis software package for evaluating CT volume data sets. The software is designed to support the physician in studying the inside, wall, and outside of the colon. It helps the physician to classify conspicuous regions of tissue with respect to their size, shape, and position.

Therefore, it combines tools for

  • visualization of the data: ●
    • Panoramic Endoluminal View a)
    • Stool Tagging b)
    • Workflow Improvements C)
    • Virtual Dissection d)
    • e) Polyp Lens
    • Stool Subtraction f)
  • navigation through the CT volume data sets ●
  • evaluation of potential findings: ●
    • a) Semi-automatic measurements
  • reporting of results:
    • a)

{5}------------------------------------------------

SIEMENS

For the inspection of the colonic surface and diagnosis of colon lesions clinically, two workflows have been established:

  • 2-dimensional search and 3-dimensional diagnosis: . Multiplanar Reconstructions (MPRs) are used for the detection of lesions. If a suspicious structure has been detected, the virtual endoscope is located and the final diagnosis can be performed.
  • 3-dimensional search and 2-dimensional diagnosis: . A virtual flight is performed starting at the rectum position of one data set. If a suspicious structure has been detected, the final diagnosis is performed using the MPRs.

Comparison of Technological Characteristics with the 6. Predicate Device

syngo.CT Colonography is a post processing application operating on the multi-user syngo.via client server platform. The subject device syngo.CT Colonography provides similar evaluation, processing, reporting and visualization tools, and functionality as the predicate device syngo Colonography with extended functionality (K042605, clearance date October 8, 2004). This includes various image processing and visualization. At a high level, the subject and predicate devices are based on the following same technical elements:

  • Polyp Enhanced Viewing
  • Segmentation in Global View
  • Size measurements of polyps ●
  • 3D Inspection of colon in Endoluminal view ●

The subject device contains additional automatic and semi-automatic visualization and evaluation tools which support assessment of colonic lesions. syngo.CT Colonography does not have significant changes in technological characteristics when compared to the predicate devices. The indication for use, intended use, operating principle, and fundamental scientific technology are similar; therefore, Siemens believes that syngo.CT Colonography is substantially equivalent to the predicate devices.

{6}------------------------------------------------

SIEMEN

Performance Data 7.

Nonclinical Testing:

syngo.CT Colonography is designed to fulfill the requirements of the following safety and performance standards:

RecognitionNumberProductAreaTitle of StandardReferenceNumber andDatePublicationDateStandardsDevelopmentOrganization
12-238RadiologyDigital Imaging andCommunications in Medicine(DICOM) SetPS 3.1 - 3.1803/16/2012NEMA
13-8SoftwareMedical device software -Software life cycle processes62304 Firstedition 2006-0508/20/2012IEC
5-40GeneralMedical devices - Application ofrisk management to medicaldevices14971 SecondEdition 2007-03-0108/20/2012ISO
5-85GeneralMedical electrical equipment --Part 1-6: General requirementsfor basic safety and essentialperformance -- CollateralStandard: Usability60601-1-6Edition 3.02010-011/30/2014IEC
5-41GeneralMedical electrical equipment -Part 1-4: General requirementsfor safety -- Collateral standard:Programmable electrical medicalsystems, edition 1.160601-1-4:2000, Consol.Ed. 1.1IEC

This submission contains performance data to demonstrate continued conformance with special controls for medical devices containing software. Non clinical tests (integration and functional) were conducted for the syngo.CT Colonography during product development. The modifications described in this Premarket Notification were supported with verification/validation testing.

The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results supports that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.

Software Verification and Validation

Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for

{7}------------------------------------------------

SIEMENS

Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission.

Clinical Testing:

Clinical performance tests were conducted to demonstrate performance. safety, and effectiveness of syngo.CT Colonography. Testing was provided to cover a variety of clinical situations that would be seen in daily clinical use of the subject device.

Summary:

Performance tests were conducted to test the functionality of the syngo.CT Colonography post processing application. These tests have been performed to test the ability of the included features of syngo.CT Colonography. The results of these tests demonstrate that this application is able to inspect the colonic surface and colon lesions. Furthermore, clinical performance testing information was provided to demonstrate the functionality of supported application features.

The result of all conducted testing was found acceptable to support the claim of substantial equivalence.

8. Indications for Use

syngo.CT Colonography is used for easy-to-perform and efficient inspection of the colonic surface. It facilitates the search and diagnosis of colon lesions. The workflow management ensures that the required data and tools are offered to you according to your role and task.

syngo.CT Colonography is a clinical post-processing workflow for basic virtual colonoscopy. It is designed to support the following image reconstruction techniques:

  • Multiplanar Reconstruction (MPR) ●
  • Volume Rendering Technique (VRT)
  • Perspective surface shaded display (pSSD)

The following evaluation tools are provided with this workflow:

  • Virtual Flight ●
  • · Stool Subtraction
  • . Panoramic View
  • · Polyp Enhanced Viewing
  • Polyp Lens
  • (PEV)
  • Stool Taqqing .
  • · Movie

syngo.CT Colonography supports reporting with appropriate reporting tools such as lesion location, lesion characterization, and key image creation. Combining enhanced commercially available digital image processing tools with an optimized workflow and reporting tools, the software is designed to support the physician on confirming the presence or absence of physician identified colon lesions (for example,

{8}------------------------------------------------

SIEMENS

polyps) in addition to evaluation, documentation, and follow-up of any such lesions using standard or low-dose spiral CT scanning.

General Safety and Effectiveness Concerns 9.

The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device. Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.

10. Conclusion as to Substantial Equivalence

In summary, Siemens is of the opinion that the syngo.CT Colonography does not introduce any new potential safety risk and is substantially equivalent to and performs as well as the predicate devices.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).