K Number
K140839

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-04-28

(26 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Caiman Seal and Cut Technology consists of dedicated bipolar electrosurgical instruments intended for use in general surgery and gynecologic surgical procedures where ligation and division of vessels is desired. The instruments create a seal by the application of bipolar electrosurgical RF energy (coagulation) to vascular structure (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue.

The Caiman 12 Plus (44cm) and the Caiman 5 are indicated for laparoscopic procedures and the Caiman 12 Plus (24cm) is indicated for open procedures. The indications for use include general surgical procedures, (including urologic, vascular, thoracic, and thoracoscopic), and gynecological procedures where ligation and division of vessels is performed. These procedures include: vaginal hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, bowel resection, and oophorectomy etc., or any procedure where vessel ligation (seal and cut), tissue grasping, and dissection is performed. The devices can be used on vessels up to and including 7mm and bundles as large as will fit in the jaws of the instrument.

Caiman Seal and Cut Technology has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the system for these procedures.

Device Description

Caiman Seal and Cut Technology consists of the Lektrafuse RF Generator and the sterile, single use Caiman devices. These devices are capable of vessel sealing, blunt dissection, grasping and dividing tissue enclosed within its jaws during open and laparoscopic procedures. The devices are designed to be used with the dedicated Lektrafuse RF Generator and create vessel ligation by the application of bipolar electrical RF energy and tissue division with a cutting blade.

AI/ML Overview

The provided text does not contain specific acceptance criteria or a detailed study proving device performance in a quantitative manner as typically found in medical device reports for AI/software products. Instead, it describes a traditional medical device (electrosurgical cutting and coagulation device).

Therefore, many of the requested sections (sample size, expert qualifications, adjudication, MRMC studies, standalone performance, training set details) are not applicable to the information provided.

However, I can extract the information that is present:

Acceptance Criteria and Device Performance

Acceptance Criteria CategoryReported Device Performance
Substantial EquivalenceSubstantially equivalent to Caiman Seal and Cut Technology cleared via K130596.
Safety and EffectivenessModifications made do not raise any new issues of safety and effectiveness.
Bench TestingFound to be substantially equivalent to predicate devices.
Standards ComplianceIn compliance with IEC 60601-1 3rd edition, IEC 60601-1-2:2007, IEC 60601-2-2:2009, and IEC 62304 1st edition.

Missing Information: The document does not provide specific numerical acceptance criteria (e.g., minimum seal strength, maximum thermal spread) or quantitative performance metrics for the modified device beyond stating "substantially equivalent" and "in compliance with standards."

Study Details for Device Performance

  1. Sample size used for the test set and the data provenance: Not specified. The document states "Bench testing was performed on the modified devices," but does not provide details on the sample size of tested vessels/tissues or their origin (e.g., in vitro, ex vivo, animal, human, country of origin, retrospective/prospective).

  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This device is an electrosurgical instrument, not an AI/diagnostic imaging device. Ground truth, in this context, would involve physical measurements rather than expert consensus on diagnostic interpretations.

  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable for this type of device testing.

  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/diagnostic device.

  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a surgical instrument that requires human operation.

  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Based on the general nature of electrosurgical device testing, the ground truth for bench testing would likely involve objective physical measurements such as burst pressure of sealed vessels, thermal spread measurements, and impedance measurements. The document does not explicitly state the types of ground truth for its "bench testing."

  7. The sample size for the training set: Not applicable. This device does not involve a "training set" in the context of machine learning.

  8. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

APR 2 8 2014

K140839 Page 1 of 2

510(k) SUMMARY (as required by 21 CFR 807.92) A.

Caiman® Seal and Cut Technology April 25, 2014

COMPANY:Aesculap ®, Inc.3773 Corporate ParkwayCenter Valley, PA 18034Establishment Registration Number: 2916714
CONTACT:Denise R. Adams610-984-9076 (phone)610-791-6882 (fax)
TRADE NAME:Caiman Seal and Cut Technology
COMMON NAME:Electrosurgical, Cutting & Coagulation & Accessories
CLASSIFICATION NAME:Electrosurgical Cutting and Coagulation Device andAccessories
REGULATION NUMBER:21 CFR 878.4400
PRODUCT CODE:GEI

SUBSTANTIAL EQUIVALENCE

Caiman Seal and Cut Technology is substantially equivalent to Caiman Seal and Cut Technology cleared via K130596.

DEVICE DESCRIPTION

Caiman Seal and Cut Technology consists of the Lektrafuse RF Generator and the sterile, single use Caiman devices. These devices are capable of vessel sealing, blunt dissection, grasping and dividing tissue enclosed within its jaws during open and laparoscopic procedures. The devices are designed to be used with the dedicated Lektrafuse RF Generator and create vessel ligation by the application of bipolar electrical RF energy and tissue division with a cutting blade.

{1}------------------------------------------------

INDICATIONS FOR USE

Caiman Seal and Cut Technology consists of dedicated bipolar electrosurgical instruments intended for use in general surgery and gynecologic surgical procedures where ligation and division of vessels is desired. The instruments create a seal by the application of bipolar electrosurgical RF energy (coagulation) to vascular structure (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue.

The Caiman 12 Plus (44cm) and the Caiman 5 are indicated for laparoscopic procedures and the Caiman 12 Plus (24cm) is indicated for open procedures. The indications for use include general surgical procedures, (including urologic, vascular, thoracic, and thoracoscopic), and gynecological procedures where ligation and division of vessels is performed. These procedures include: vaginal hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, bowel resection, and oophorectomy etc., or any procedure where vessel ligation (seal and cut), tissue grasping, and dissection is performed. The devices can be used on vessels up to and including 7mm and bundles as large as will fit in the jaws of the instrument.

Caiman Seal and Cut Technology has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the system for these procedures.

TECHNOLOGICAL CHARACTERISTICS (compared to predicates)

The modifications made to the Caiman Seal and Cut Technology system do not affect the fundamental scientific technology. The design, materials, and principal of operation have not changed for these devices. The modifications made to these devices do not raise any new issues of safety and effectiveness.

PERFORMANCE DATA

Bench testing was performed on the modified devices and found them to be substantially equivalent to the predicate devices.

Caiman Seal and Cut Technology is in compliance with the following standards: IEC 60601-1 300 edition, IEC 60601-1-2:2007, IEC 60601-2-2:2009, and IEC 62304 1" edition.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized human figure with outstretched arms, representing care and protection. The figure is composed of three curved lines, giving it a sense of movement and dynamism. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 28, 2014

Aesculap® Incorporated Ms. Denise R. Adams Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034

Re: K140839

Trade/Device Name: Caiman Seal and Cut Technology Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: April 1, 2014 Received: April 2, 2014

Dear Ms. Adams:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Ms. Denise R. Adams

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

for

Sincerely yours,

David Krause -S

Binita S.

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K140839

Device Name Caiman Seal and Cut Technology

Indications for Use (Describe)

Caiman Seal and Cut Technology consists of dedicated bipolar electrosurgical instruments intended for use in general surgery and gynecologic surgical procedures where ligation of vessels is desired. The instruments create a seal by the application of bipolar electrosurgical RF energy (coagulation) to vascular structure (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue.

The Caiman 12 Plus (44cm) and the Caiman 5 are indicated for laparoscopic procedures and the Caiman 12 Plus (24cm) is indicated for open procedures. The include general surgical procedures, (including urologic, vascular, thoracic, and thoracoscopic), and gynecological procedures where ligation of vessels is performed. These procedures include: vacinal hysterectomies. Nissen fundoplication, colectorny, adhesiolysis, bowel resection, and ophorectorny etc., or any procedure where vessel ligation (seal and cut), tissue grasping, and dissection is performed. The devices can be used on vessels up to and including 7mm and bundles as large as will fit in the jaws of the instrument.

Caiman Seal and Cut Technology has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the system for these procedures.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FOR FDA USE ONLY - JUST - JUST - FOR FOR FOR FOR

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

:

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.