K Number
K140838

Validate with FDA (Live)

Date Cleared
2014-12-05

(247 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IV Administration Sets with Hand Pump are intravenous administration sets intended for delivery of fluids from a container into a patient's vascular system. When the hand pump component is activated to deliver only crystalloid and colloid resuscitative fluids. These devices may patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy.

Device Description

The IV Administration Sets with Hand Pump are single use, disposable, intravenous administration sets with a pressure pump and air guard filter used to deliver fluids from a container into a patient's vascular system rapidly through the use of the pressure pump and/or gravity flow. These sets may be comprised of various generic components which are broadly used throughout the industry such as a bag spike, drip chamber, slide clamp, roller clamp, luer access device, stopcocks, manifolds, tubing, and luer connections. Each of these needleless components have already been cleared through the 510(k) process either individually or as part of a B. Braun IV pump 510(k) which included sets. The unique components to these sets include the manual hand pump and a 15um filter integrated within the drip chamber to reduce the potential for bubble formation within the set during use. The addition of the hand pressure pump provides the capability for delivering IV fluids more rapidly by compressing the pump by hand.

AI/ML Overview

This document describes a 510(k) premarket notification for the B. Braun IV Administration Set with Hand Pump and outlines the testing conducted to demonstrate its substantial equivalence to predicate devices. However, the document does not present specific acceptance criteria or reported device performance in a quantitative manner that can be organized into a table. The performance testing section broadly states that functional performance testing was completed to demonstrate that the sets perform as intended and that results of testing demonstrate that the proposed device performs similarly to the predicate device and can be used safely and effectively according to its intended use. This is a qualitative conclusion rather than a presentation of specific metrics against predefined thresholds.

Based on the provided text, here is an attempt to address your request, acknowledging the limitations of the input:


1. Table of Acceptance Criteria and Reported Device Performance

As the document does not specify quantitative acceptance criteria or precise performance metrics, the table below reflects the qualitative statements provided.

Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred/Qualitative)Reported Device Performance (Inferred/Qualitative)
Functional PerformanceDevice performs as intended for fluid delivery, including gravity flow and increased flow with hand pump."Results of testing demonstrate that the proposed device performs similarly to the predicate device and can be use safety and effectively according to its intended use."
Material BiocompatibilityMaterials are safe for clinical application."Biocompatibility test results verify that the IV Administration Sets with Hand Pump materials of construction are safe for their clinical application."
Adherence to StandardsConformance to relevant ISO standards (e.g., fluid path integrity, luer connections, sterilization).The device was evaluated against ISO 594-1, ISO 594-2, ISO 8536-4, ISO 10993-1, and ISO 11135-1. (Implies conformance was met to achieve substantial equivalence).

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for the performance or biocompatibility testing. It only mentions that "Functional performance testing was completed" and "The materials of construction...were tested."

The data provenance is not specified in terms of country of origin or whether it was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

This information is not provided in the document. The testing described is primarily engineering/laboratory-based functional and biocompatibility testing, not clinical studies involving expert interpretation of results to establish a "ground truth" in the way one might for diagnostic devices.

4. Adjudication Method for the Test Set

This information is not provided. Given the nature of the testing (functional and biocompatibility), an adjudication method in the context of expert consensus on clinical findings is not applicable.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. The document focuses on demonstrating substantial equivalence through functional performance and biocompatibility testing against predicate devices and relevant industry standards, not on human reader performance with or without AI assistance.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

This question is not applicable as the device is an IV administration set with a hand pump, not an AI or algorithm-driven device.

7. The Type of Ground Truth Used

For the functional performance testing, the "ground truth" would be the expected physical and mechanical performance of an IV administration set as defined by engineering specifications and industry standards (e.g., flow rates, luer connection integrity, pressure resistance). For biocompatibility, the "ground truth" is that the materials do not elicit adverse biological responses, as determined by ISO 10993-1 standard tests.

8. The Sample Size for the Training Set

This question is not applicable as the device is not an AI/machine learning device that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable.


{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other to create a sense of depth and connection.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 5,2014

B. Braun Medical, Inc. Angela Caravella Regulatory Affairs Specialist 901 Marcon Boulevard ALLENTOWN, PA 18109-9341

Re: K140838

Trade/Device Name: IV Administration Set with Hand Pump Regulation Number: 880.5440 Regulation Name: Set, Administration, Intravascular Regulatory Class: II Product Code: FPA Dated: October 31, 2014 Received: November 4, 2014

Dear Ms. Caravella:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{1}------------------------------------------------

Page 2 - Ms. Caravella

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Image /page/1/Picture/8 description: The image shows a handwritten signature that reads "Susan Runno DDS, MA". The signature is written in black ink and appears to be on a white background. The letters are connected and flow together, giving the signature a cursive style. The letters "DDS, MA" are smaller than the rest of the signature.

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K140838

Device Name IV Administration Sets with Hand Pump

Indications for Use (Describe)

The IV Administration Sets with Hand Pump are intravenous administration sets intended for delivery of fluids from a container into a patient's vascular system. When the hand pump component is activated to deliver only crystalloid and colloid resuscitative fluids. These devices may patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

B. Braun Medical Inc. 510(k) Premarket Notification IV Administration Sets with Hand Pump

5. 510(k) SUMMARY

DATE:December 4, 2014
SUBMITTER:B. Braun Medical Inc.901 Marcon BoulevardAllentown, PA 18109-9341610-266-0500
Contact: Angela J. Caravella, Regulatory Affairs SpecialistPhone: (610) 596-2966Fax: (610) 266-4962E-mail: angela.caravella@bbraun.com
DEVICE NAME:B. Braun IV Administration Set with Hand Pump
COMMON ORNAME:IV Administration Set with Hand Pump
CLASSIFICATION:Class II, Product Code FPA, 880.5440
PREDICATE DEVICES:Infusomat Space Volumetric Infusion Pump System, BBraun Medical, Inc., K062700, Class II, FRN, 880.5725Hospira Infusion Blood Sets, Hospira Inc., K101677, ClassII, FPA, 880.5440Blood Hand Pump Administration Set, Cardinal Health,K050115, Class II, BRZ, 880.5440

DESCRIPTION:

The IV Administration Sets with Hand Pump are single use, disposable, intravenous administration sets with a pressure pump and air guard filter used to deliver fluids from a container into a patient's vascular system rapidly through the use of the pressure pump and/or gravity flow. These sets may be comprised of various generic components which are broadly used throughout the industry such as a bag spike, drip chamber, slide clamp, roller clamp, luer access device, stopcocks, manifolds, tubing, and luer connections. Each of these needleless components have already been cleared through the 510(k) process either individually or as part of a B. Braun IV pump 510(k) which included sets. The unique components to these sets include the manual hand pump and a 15um filter integrated within the drip chamber to reduce the potential for bubble formation within the set during use. The addition of the hand pressure pump provides the capability for delivering IV fluids more rapidly by compressing the pump by hand.

{5}------------------------------------------------

INTENDED USE:

The IV Administration Sets with Hand Pump are intravenous administration sets intended for delivery of fluids from a container into a patient's vascular system. When the hand pump component is activated. the device is intended to deliver only crystalloid and colloid resuscitative fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness for the solution being infused and duration of therapy.

SUBSTANTIAL EQUIVALENCE:

Technological Characteristics

Predicate Device - Infusomat Space Volumetric Infusion Pump System (K062700) The IV Administration Sets with Hand Pump have a similar intended use to the Infusomat Space Sets in that they are indicated for infusion therapy for the delivery of fluids to a patient by gravity or through an increased flow rate through the use of a manually activated hand pump (in the case of the proposed device) or with an electromechanical pump (in the case of the Infusomat Space Infusion Pump). The proposed device contains several of the same materials and components as the Infusomat Space Pump Administration Set predicate device.

Predicate Device - Infusion Blood Sets (K101677)

The IV Administration Sets with Hand Pump have a similar indication for use to the Infusion Blood sets marketed by Hospira for the delivery of fluids from a container to a patient's vascular system. Hospira's sets also allow for the delivery of blood to patients which is outside of the proposed device's indication for use. The hand pump described within Hospira's 510(k) K101677 identifies a cylindrical hand pump with ball check valves, this hand pump design with the ball check features are the same as the hand pump present on the proposed B. Braun IV Administration Sets with Hand Pump.

Predicate Device - Blood Hand Pump Administration Set (K050115)

The IV Administration Sets with Hand Pump have a similar indication for use to the Blood Hand Pump Administration set marketed by Cardinal Health in that they are intended for the delivery of fluids from a container to a patient's vascular system either by gravity or through an increased flow rate through the use of a manually activated hand pump. Additionally, the proposed device and Cardinal's Blood Hand Pump administration set indicates the use of a manually activated hand pump which is also present within the design of the proposed device.

{6}------------------------------------------------

Performance Testing

The following performance standards were utilized in evaluating the functionality of the IV Administration Sets with Hand Pump:

ISO 594-1:1986, "Conical fittings with a 6% (luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements"

ISO 594-2:1998, "Conical fittings with a 6% (luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings"

ISO 8536-4:2010, "Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed"

ISO 10993-1:2009, "Biological evaluation of medical devices - part 1: Evaluation and testing within a risk management process"

ISO 11135-1: 2007, "Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices"

Functional performance testing was completed with the proposed IV Administration Sets with Hand Pump to demonstrate that the sets perform as intended. Results of testing demonstrate that the proposed device performs similarly to the predicate device and can be use safety and effectively according to it's intended use.

Biocompatibility

The materials of construction of a fully assembled IV Administration Set with Hand Pump were tested according to ISO 10993-1:2009.

Biocompatibility test results verify that the IV Administration Sets with Hand Pump materials of construction are safe for their clinical application.

CONCLUSION:

Results of functional performance and biocompatibility testing conducted with the proposed device demonstrate that the IV Administration Sets with Hand Pump are substantially equivalent to the predicate devices and is safe and effective for its intended use.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.