K Number
K140748

Validate with FDA (Live)

Date Cleared
2014-08-20

(148 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

One Step Single/Multi-drug Test Cup and One Step Single/Multi-drug Test Dipcard are lateral flow chromatographic immunoassays designed to qualitatively detect the presence of drug metabolites in human urine at the following cut-off concentrations:

TestCalibratorCutoff Level (ng/mL)
Buprenorphine (BUP)Buprenorphine10
2-ethylidene-1, 5-dimethyl-3,3-diphenylpyrrolidine (EDDP)2-ethylidene-1, 5-dimethyl-3,3-diphenylpyrrolidine300
Morphine (MOP300)Morphine300
Propoxyphene (PPX)Propoxyphene300
Tricyclic Antidepressants (TCA)Nortriptyline1000

There are two formats: 1) Test Cup, 2) Test Dipcard. Each format may have from 1 to 5 drugs in any combination. The assays are intended for in vitro diagnostic use. They are intended for prescription use including point of care sites and over-the-counter use.

The tests may yield preliminary positive results even when prescription drugs including Buprenorphine, or Tricyclic Antidepressants are ingested, at prescribed to distinguish between prescription use or abuse of these drugs. There are no uniformly recognized cutoff concentration levels for Buprenorphine, or Tricyclic Antidepressants in urine.

This assay provides only a preliminary analytical test result. Gas Chromatography/Mass spectrometry (GC/MS) or an equivalent analytical method is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

Device Description

One Step Single/Multi-drug Test Cup and One Step Single/Multi-drug Test Dipcard are competitive binding , lateral flow immunochromatographic assays for qualitatively the detection of Buprenorphine, 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine, Morphine, Propoxyphene, Tri-cyclic Antidepressants and their metabolites at or above the cut-off levels as indicated. The tests can be performed without the use of an instrument.

Test Cup and Test Dipcard use identical test strips made with same chemical formulation and manufacturing procedures.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the One Step Single/Multi-drug Test Cup and Dipcard, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a quantitative manner for overall device performance (e.g., global sensitivity/specificity targets). Instead, the performance data is presented against the established cutoff concentrations for each drug. The implicitly accepted performance is demonstrated by the agreement between the device's results and the confirmatory methods (GC/MS or HPLC) across different concentration ranges.

The tables below synthesize the performance of the device (both Cup and Dipcard formats, for Single and Multi-drug tests) by showing how results align with concentrations relative to the cutoff. The provided data focuses on the distribution of results across different concentration categories rather than overall accuracy metrics like sensitivity or specificity. However, we can infer performance by observing the agreement rates within each category.

Inferred Acceptance Criteria:
The device is expected to:

  • Consistently produce negative results for drug-free samples.
  • Consistently produce negative results for samples significantly below the cutoff concentration (less than half the cutoff).
  • Demonstrate appropriate discrimination around the cutoff, with increasing positive results as concentrations approach and exceed the cutoff.
  • Show agreement with confirmatory methods (GC/MS or HPLC) for samples within and outside the cutoff ranges.

Reported Device Performance (Sample Summary for BUP from 'Single drug Test' and 'Multi-drug Test' for both Cup and Dipcard):

Since the results are identical across all provided tables for BUP, EDDP, MOP, PPX, and TCA for both Single/Multi-drug Test Cup and Dipcard, a representative row is used.

Buprenorphine (BUP) Performance (Representative Example):

  • Calibrator: Buprenorphine
  • Cutoff Level: 10 ng/mL
Concentration Category (by GC/MS/HPLC)Co-Innovation Test ResultCount
Drug free+0
(n=35)-35
Less than half the cutoff ( < 5 ng/mL)+0
(n=0)-0
Near Cutoff Negative (5 ng/mL to < 10 ng/mL)+1
(n=5)-4
Near Cutoff Positive (10 ng/mL to 15 ng/mL)+5
(n=5)-0
High Positive (> 15 ng/mL)+35
(n=35)-0
Total (n=80)80

Discordant Results (Representative Example for BUP):

  • BUP Test: Positive result at 7.8 ng/mL (below 10 ng/mL cutoff). This is a false positive based on the cutoff but within the expected variability around the cutoff.
  • TCA Test: Negative result at 1138 ng/mL (above 1000 ng/mL cutoff). This is a false negative.

(Similar tables and discordant results exist for EDDP, MOP, PPX, and TCA, exhibiting comparable patterns.)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 80 clinical urine specimens were used for each drug (BUP, EDDP, MOP, PPX, TCA) for each device type (Single drug Test Cup, Single drug Test Dipcard, Multi-drug Test Cup, Multi-drug Test Dipcard). This means:
    • Single drug Test Cup: 5 drugs * 80 specimens = 400 specimens
    • Single drug Test Dipcard: 5 drugs * 80 specimens = 400 specimens
    • Multi-drug Test Cup: 5 drugs * 80 specimens = 400 specimens
    • Multi-drug Test Dipcard: 5 drugs * 80 specimens = 400 specimens
    • Total specimens analyzed across all tests and drugs: 1600 individual drug/specimen analyses.
  • Data Provenance: The document states "80 clinical urine specimens". It does not specify the country of origin of the data or whether the study was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The ground truth was established by analytical methods, not human experts.

  • Number of Experts: Not applicable, as ground truth was not established by human experts.
  • Qualifications of Experts: Not applicable.

4. Adjudication Method for the Test Set

The ground truth was established by definitive analytical methods, not through an adjudication process involving human interpretation of the device results.

  • Adjudication Method: Not applicable. The reference method (GC/MS or HPLC) served as the definitive ground truth.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • MRMC Study: No, an MRMC comparative effectiveness study was not explicitly described or performed. This device is an in-vitro diagnostic assay for qualitative drug detection in urine, primarily evaluated against gold-standard analytical methods rather than against human readers or human readers with AI assistance.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Standalone Performance: Yes, the performance data presented is for the device's standalone performance. The "Co-Innovation Result" recorded for each specimen is directly from the device (visual interpretation of the test line on the Cup/Dipcard), without human intervention in interpreting the results. The comparison is made against laboratory-confirmed concentrations, representing the device's intrinsic analytical performance.

7. The Type of Ground Truth Used

  • Type of Ground Truth: Analytical confirmation methods:
    • Gas Chromatography/Mass Spectrometry (GC/MS)
    • High-Performance Liquid Chromatography (HPLC)

8. The Sample Size for the Training Set

The document describes performance studies (accuracy, precision, sensitivity, specificity/cross-reactivity, interference, stability, and home-use consumer study), but it does not specify a separate "training set" or "training data" used for algorithm development, as this device appears to be a lateral flow immunoassay, not a machine learning-based device.

9. How the Ground Truth for the Training Set Was Established

As there is no mention of a "training set" for an algorithm, this question is not applicable. For quality control during manufacturing and development, external controls and reference materials are typically used, but these are not equivalent to an algorithm's training set in the context of AI/ML.

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 20, 2014 CO-INNOVATION BIOTECH CO., LTD. HONG FENG NO. 13, YANYUAN ROAD, TIANHE DISTRICT, GUANGZHOU, P.R. CHINA

Re: K140748

Trade/Device Name: One Step Single/Multi-drug Test Cup One Step Single/Multi-drug Test Dipcard Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: II Product Code: DJG, DJR, DNK, JXN, LFG Dated: July 08, 2014 Received: July 09, 2014

Dear Hong Feng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Katherine Serrano -S

For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K140748

Device Name

One Step Single/Multi-drug Test Cup One Step Single/Multi-drug Test Dipcard

Indications for Use (Describe)

One Step Single/Multi-drug Test Cup and One Step Single/Multi-drug Test Dipcard are lateral flow chromatographic immunoassays designed to qualitatively detect the presence of drug metabolites in human urine at the following cut-off concentrations:

TestCalibratorCutoff Level (ng/mL)
Buprenorphine (BUP)Buprenorphine10
2-ethylidene-1, 5-dimethyl-3,3-diphenylpyrrolidine (EDDP)2-ethylidene-1, 5-dimethyl-3,3-diphenylpyrrolidine300
Morphine (MOP300)Morphine300
Propoxyphene (PPX)Propoxyphene300
Tricyclic Antidepressants (TCA)Nortriptyline1000

There are two formats: 1) Test Cup, 2) Test Dipcard. Each format may have from 1 to 5 drugs in any combination. The assays are intended for in vitro diagnostic use. They are intended for prescription use including point of care sites and over-the-counter use.

The tests may yield preliminary positive results even when prescription drugs including Buprenorphine, or Tricyclic Antidepressants are ingested, at prescribed to distinguish between prescription use or abuse of these drugs. There are no uniformly recognized cutoff concentration levels for Buprenorphine, or Tricyclic Antidepressants in urine.

This assay provides only a preliminary analytical test result. Gas Chromatography/Mass spectrometry (GC/MS) or an equivalent analytical method is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

മ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Section 5 - 510(k) Summary

Date of Summary Preparation: 8/14/2014

1. Submitter's Identifications

Submitter: Co-Innovation Biotech Co.,Ltd. Address: No.13, Yanyuan Road, Tianhe District, Guangzhou, P.R. China Contact Person: Hong Feng Contact Email Address: fenghongfda@126.com Telephone: + 86 -20-62867285 Fax: + 86 -20-62867285

2. Correspondent's Identifications

Correspondent's Name: Co-Innovation Biotech Co.,Ltd. Address: No.13, Yanyuan Road, Tianhe District, Guangzhou, P.R. China Contact Person: Hong Feng Contact Email Address: fenghongfda@126.com Telephone: + 86 -20-62867285 Fax: + 86 -20-62867285

3. Name of the Device

Recommended classification regulation:

21 CFR 862.3650 Opiate test system 21 CFR 862.3620 Methadone test system 21 CFR 862.3640 Morphine Test System 21 CFR 862.3700 Propoxyphene test system 21 CFR 862.3910 Tri-cyclic Antidepressants drug test system

Device class: Class II Panel: Toxicology (91) Product code: DJG,DJR,DNK,JXN,LFG Common Name: Buprenorphine(BUP) Test System 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine(EDDP) Test System Morphine(MOP) Test System Propoxyphene(PPX) Test System Tri-cyclic Antidepressants (TCA)Test System

Proprietary names:

One Step Single/Multi-drug Test Cup One Step Single/Multi-drug Test Dipcard

{5}------------------------------------------------

4. The Predicate Devices

K122809Advin Multi-Drug Screen Test Dip Card
Advin Multi-Drug Screen Test Cup
Advin Multi-Drug Screen Test Cassette

5. Device Description

One Step Single/Multi-drug Test Cup and One Step Single/Multi-drug Test Dipcard are competitive binding , lateral flow immunochromatographic assays for qualitatively the detection of Buprenorphine, 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine, Morphine, Propoxyphene, Tri-cyclic Antidepressants and their metabolites at or above the cut-off levels as indicated. The tests can be performed without the use of an instrument.

Test Cup and Test Dipcard use identical test strips made with same chemical formulation and manufacturing procedures.

6. Intended Use of Device

One Step Single/Multi-drug Test Cup and One Step Single/Multi-drug Test Dipcard are lateral flow chromatographic immunoassays designed to qualitatively detect the presence of drugs and drug metabolites in human urine at the following cut-off concentrations:

TestCalibratorCut-off level
Buprenorphine(BUP)Buprenorphine10ng/mL
2-ethylidene-1, 5-dimethyl-3,3-diphenylpyrrolidine(EDDP)2-ethylidene-1, 5-dimethyl-3,3-diphenylpyrrolidine300 ng/mL
Morphine(MOP300)Morphine300 ng/mL
Propoxyphene(PPX)Propoxyphene300 ng/mL
Tri-cyclic Antidepressants (TCA)Nortriptyline1000 ng/mL

There are two formats: 1) Test Cup, 2) Test Dipcard. Each format may have from 1 to 5 drugs in any combination. The assays are intended for in vitro diagnostic use. They are intended for prescription use including point of care sites and over-the-counter use.

The tests may yield preliminary positive results even when prescription drugs including Buprenorphine, Propoxyphene, or Tricyclic Antidepressants are ingested, at prescribed doses; it is not intended to distinguish between prescription use or abuse of these drugs. There are no uniformly recognized cutoff concentration levels for Buprenorphine. or Tricyclic Antidepressants in urine.

This assay provides only a preliminary analytical test result. Gas Chromatography/Mass spectrometry (GC/MS) or an equivalent analytical method is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

{6}------------------------------------------------

7. Comparison to Predicate Devices:

A summary comparison of features of the One Step Single/Multi-drug Test Cup and One Step Single/Multi-drug Test Dipcard and the predicate devices is provided in the following Table:

ItemDevicePredicate (K122809)
Indication for useQualitative detection of drugs-of-abusein urine (Buprenorphine,2-ethylidene-1, 5-dimethyl-3,3-diphenylpyrrolidine, Morphine,Propoxyphene, Tri-cyclicAntidepressants)Same (but the number ofdrugs detected different)
Intended UsersOver the Counter (OTC) Use andPrescription UseOver the Counter (OTC)Use and Prescription Use
SpecimenUrineSame
CutoffBuprenorphine:10 ng/mLEDDP:300 ng/mLMorphine:300 ng/mLPropoxyphene:300 ng/mLTri-cyclic Antidepressants:1000 ng/mLSame
Read time5 minutesSame
ResultsQualitativeSame
MethodologyCompetitive binding, Lateral flowimmunochromatographic assay basedon the principle of antigen antibodyimmunochemistrySame
ConfigurationDipcard and CupCassette Dip Card and Cup

Remark:

  • 1、The subject devices have all features of the predicate device except the number of drugs detected . This difference do not affect the performance characteristics of the subject devices.

8. Performance Data:

Accuracy

Single drug Test:

80 clinical urine specimens for each drug were analyzed by GC/MS, or HPLC and by one lot of the corresponding One Step Single drug Test Cup. Samples were divided by concentration into five categories: drug free, less than half the cutoff negative, near cutoff positive, and high positive. Results were as follows:

DrugTestCo-InnovationResultDrug free byGC/MS analysisLess than halfthe cutoffconcentrationby GC/MSanalysisNear CutoffNegative (Between50% below thecutoff and the cutoffconcentration)Near CutoffPositive (Betweenthe cutoff and50% above thecutoffconcentration)High Positive(greater than50% above thecutoffconcentration)Total
BUP+00153580

{7}------------------------------------------------

-350400
EDDP+00173380
-331500
MOP+00153580
-312600
PPX+00163480
-331500
TCA+00053480
-350510

Analysis of Discordant Results with One Step Single drug Test Cup

One Step Single drug Test CupGC/MS Analysis
Drug TestCutoff(ng/mL)TestResultDrugConcentration(ng/mL)Drug in Urine
BUP**10Positive7.8Buprenorphine
EDDP300Positive2852-ethylidene-1, 5-dimethyl-3,3-diphenylpyrrolidine
MOP300Positive275Morphine
PPX300Positive269Propoxyphene
TCA*1000Negative1138Nortriptyline

(TCA*:TCA was based on HPLC data.BUP**:BUP was based on LC/MS data.)

80 clinical urine specimens for each drug were analyzed by GC/MS, or HPLC and by one lot of the corresponding One Step Single drug Test Dipcard. Samples were divided by concentration into five categories: drug free, less than half the cutoff negative, near cutoff positive, and high positive. Results were as follows:

DrugTestCo-InnovationResultDrug free byGC/MSanalysisLess than halfthe cutoffconcentrationby GC/MSanalysisNear CutoffNegative (Between50% below thecutoff and thecutoffconcentration)Near CutoffPositive (Betweenthe cutoff and50% above thecutoffconcentration)High Positive(greater than50% above thecutoffconcentration)Total
BUP+00153580
BUP-350400
EDDP+00173380
EDDP-331500
MOP+00153580
MOP-312600
PPX+00163480
PPX-331500
TCA+00053480
TCA-350510

Analysis of Discordant Results with One Step Single drug Test Dipcard

{8}------------------------------------------------

One Step Single drug Test DipcardGC/MS Analysis
Drug TestCutoff(ng/mL)Test ResultDrugConcentration(ng/mL)Drug in Urine
BUP**10Positive7.8Buprenorphine
EDDP300Positive2852-ethylidene-1, 5-dimethyl-3,3-diphenylpyrrolidine
MOP300Positive275Morphine
PPX300Positive269Propoxyphene
TCA*1000Negative1138Nortriptyline

(TCA*:TCA was based on HPLC data.BUP**:BUP was based on LC/MS data.)

Multi-drug Test:

80 clinical urine specimens for each drug were analyzed by GC/MS, or HPLC and by one lot of the corresponding One Step Multi-drug Test Cup. Samples were divided by concentration into five categories: drug free, less than half the cutoff negative, near cutoff positive, and high positive. Results were as follows:

DrugTestCo-InnovationResultDrug free byGC/MSanalysisLess than halfthe cutoffconcentrationby GC/MSanalysisNear CutoffNegative (Between50% below thecutoff and thecutoffconcentration)Near CutoffPositive (Betweenthe cutoff and50% above thecutoffconcentration)High Positive(greater than50% above thecutoffconcentration)Total
BUP+00153580
BUP-350400
EDDP+00173380
EDDP-331500
MOP+00153580
MOP-312600
PPX+00163480
PPX-331500
TCA+00053480
TCA-350510

Analysis of Discordant Results with One Step Multi-drug Test Cup

One Step Multi-drug Test CupGC/MS Analysis
Drug TestCutoff(ng/mL)Test ResultDrug Concentration (ng/mL)Drug in Urine
BUP**10Positive7.8Buprenorphine
EDDP300Positive2852-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine
MOP300Positive275Morphine
PPX300Positive269Propoxyphene
TCA*1000Negative1138Nortriptyline

{9}------------------------------------------------

(TCA*:TCA was based on HPLC data.BUP**:BUP was based on LC/MS data.)

80 clinical urine specimens for each drug were analyzed by GC/MS, or HPLC and by one lot of the corresponding One Step Multi-drug Test Dipcard. Samples were divided by concentration into five categories: drug free, less than half the cutoff, near cutoff negative, near cutoff positive, and high positive. Results were as follows:

DrugTestCo-InnovationResultDrug free byGC/MSanalysisLess than halfthe cutoffconcentrationby GC/MSanalysisNear CutoffNegative (Between50% below thecutoff and thecutoffconcentration)Near CutoffPositive (Betweenthe cutoff and50% above thecutoffconcentration)High Positive(greater than50% above thecutoffconcentration)Total
BUP+00153580
-350400
EDDP+00173380
-331500
MOP+00153580
-312600
PPX+00163480
-331500
TCA+00053480
-350510

Analysis of Discordant Results with One Step Multi-drug Test Dipcard

One Step Multi-drug Test DipcardGC/MS Analysis
Drug TestCutoff(ng/mL)Test ResultDrug Concentration (ng/mL)Drug in Urine
BUP**10Positive7.8Buprenorphine
EDDP300Positive2852-ethylidene-1, 5-dimethyl-3,3-diphenylpyrrolidine
MOP300Positive275Morphine
PPX300Positive269Propoxyphene
TCA*1000Negative1138Nortriptyline

(TCA*:TCA was based on HPLC data.BUP**:BUP was based on LC/MS data.)

Other Information About Performance Characteristics:

The performance characteristics of One Step Single/Multi-drug Test Cup and One Step Single/Multi-drug Test Dipcard were evaluated by precision study, sensitivity study, specificity and cross reactivity study, interference study, stability study and home use consumer study. The study results indicate that One Step Single/Multi-drug Test Cup and One Step Single/Multi-drug Test Dipcard perform satisfactorily when used according to the package inserts.

10. Conclusion:

{10}------------------------------------------------

One Step Single/Multi-drug Test Cup and One Step Single/Multi-drug Test Dipcard are substantially equivalent to Advin Multi-Drug Screen Test Dip Card, Advin Multi-Drug Screen Test Cup and Advin Multi-Drug Screen Test Cassette.

--- End of this section ---

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).