K Number
K140717

Validate with FDA (Live)

Date Cleared
2014-11-17

(242 days)

Regulation Number
862.2900
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CLINITEK Novus® Automated Urine Chemistry Analyzer is a fully automated urinalysis instrument. The CLINITEK Novus analyzer is intended to read Siemens Healthcare Diagnostics CLINITEK Novus Cassettes, as well as determine urine specific gravity and urine clarity. The CLINITEK Novus 10 Urinalysis Cassette is intended for the seniquantitative measurement of the following parameters in urine: bilirubin, blood (occult), glucose, ketone (acetoacetic acid), leukocytes, nitrite (qualitative), pH, protein, color, and urobilinogen.

These measurements are used to assist diagnosis in the following areas:

  • · Carbohydrate metabolism (such as diabetes mellitus)
  • · Kidney function
  • · Liver function
  • · Metabolic disorders
  • · Urinary tract infection

For in vitro diagnostic use.

The CLINITEK Novus® Calibration Kit is intended to be used with the CLINITEK Novus Urinalysis Cassette to calibrate the CLINITEK Novus Automated Urine Chemistry analyzer. This product is for professional in vitro diagnostic use.

Device Description

The CLINITEK Novus® system is a fully automated urine chemistry analyzer that is designed for use with the CLINITEK Novus® 10 Urinalysis Cassette. The analyzer automates the process of urine strip testing by dispensing urine samples onto a test pad, and then by reading the color and intensity of light reflected from the reacted test pads, converts the results into clinically meaningful units.

Test results from the test pads are qualitative and semi-quantitative; specific gravity is measured by a refractometer assembly.

AI/ML Overview

Here's an analysis of the acceptance criteria and study detailed in the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria (e.g., "The device must achieve X% agreement"). However, it does present the results of comparison studies against a predicate device and states that "The results for overall percent agreement and within one level agreement met acceptance criteria." This implies that acceptable thresholds were achieved. Based on the presented data, we can infer the achieved performance. For specific analytes, the values vary, but they consistently show high exact match and very high agreement within +/- 1 block.

AnalytePerformance MetricCLINITEK Novus Result (All Sites Combined)Implied/Achieved Acceptance Criteria (based on "met acceptance criteria")
Bilirubin% exact match (range)64.4% - 99.8%High % (exact match)
% within 1 block (range)98.2% - 100.0%Very high % (within 1 block)
Blood% exact match (range)81.5% - 97.8%High % (exact match)
% within 1 block (range)100.0% (except one at 99.7%)Very high % (within 1 block)
Glucose% exact match (range)88.5% - 99.6%High % (exact match)
% within 1 block (range)100.0%Very high % (within 1 block)
Ketone% exact match (range)88.2% - 100.0%High % (exact match)
% within 1 block (range)99.2% - 100.0%Very high % (within 1 block)
Leukocyte% exact match (range)81.1% - 96.6%High % (exact match)
% within 1 block (range)99.9% - 100.0%Very high % (within 1 block)
Nitrite% exact match (range)99.3% - 100.0%High % (exact match)
% within 1 block (range)100.0%Very high % (within 1 block)
pH% exact match (range)45.4% - 99.8%High % (exact match)
% within 1 block (range)97.5% - 100.0%Very high % (within 1 block)
Protein% exact match (range)89.9% - 98.7%High % (exact match)
% within 1 block (range)99.9% - 100.0%Very high % (within 1 block)
Urobilinogen% exact match (range) (sites 1-3)85.6% - 99.9%High % (exact match)
% within 1 block (range) (sites 1-3)100.0%Very high % (within 1 block)
Urobilinogen% exact match (range) (site 4)93.8% - 100.0%High % (exact match)
% within 1 block (range) (site 4)100.0%Very high % (within 1 block)
Specific GravityR20.983High R2 (correlation)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size:
    • For the method comparison study (comparing CLINITEK Novus to the predicate CLINITEK Atlas), up to 2773 specimens were tested across 3 sites for most analytes.
    • For Urobilinogen, an additional fourth site was involved due to a high number of contrived samples in the initial three sites. This fourth site tested 261 samples (187 negative, and then 48, 10, 12, 4 for various positive levels).
  • Data Provenance: The document does not explicitly state the country of origin. The study was conducted "across 3 sites" (and later a fourth), implying a multi-center study within the scope of the device manufacturer's operations/clinical trials. It is a prospective comparison study as it compared the new device (CLINITEK Novus) against an existing predicate device (CLINITEK Atlas) using actual patient specimens.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The context of this device being an automated urine chemistry analyzer comparing its readings to a predicate automated urine chemistry analyzer suggests that the "ground truth" for the test set was established by the results from the predicate device (CLINITEK Atlas), not human experts. Therefore, the concept of a number of experts and their qualifications for establishing ground truth is not directly applicable in this context. The predicate device's established performance serves as the reference.

4. Adjudication Method for the Test Set

Since the comparison is primarily between two automated devices, and the "ground truth" is derived from the predicate device's readings, a typical adjudication method involving human readers (like 2+1 or 3+1 consensus) is not applicable. The data presented is a direct comparison of the readings obtained from the CLINITEK Novus analyzer against the readings obtained from the CLINITEK Atlas analyzer.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance?

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study involving human readers and AI assistance was not performed. This device is an automated in vitro diagnostic (IVD) analyzer, not an AI-powered image analysis tool meant to assist human readers. Its primary function is to automatically measure chemical parameters in urine.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done?

Yes, the performance reported is a standalone performance of the CLINITEK Novus Automated Urine Chemistry Analyzer. It assesses the device's ability to accurately measure urine parameters independently, by comparing its results directly to those of a predicate automated device. There is no human-in-the-loop component in the reported performance evaluation.

7. The Type of Ground Truth Used

The primary "ground truth" used for the method comparison study was the results obtained from the predicate device, the CLINITEK Atlas Automated Urine Chemistry Analyzer. For the detection limit study, "ground truth" was established by spiking negative pooled human urine with known quantities of analytes.

8. The Sample Size for the Training Set

The document describes studies comparing the new device to a predicate device and analytical performance characteristics. It does not explicitly mention a "training set" or a machine learning/AI model training process for this IVD device. The description focuses on demonstrating substantial equivalence to a predicate device through analytical and method comparison studies. If the device uses algorithms derived from prior data, that information is not detailed in this summary.

9. How the Ground Truth for the Training Set Was Established

Since a "training set" is not explicitly mentioned in the context of machine learning/AI for this device, the method for establishing its ground truth is not applicable/not provided in this document. The device's operation appears to be based on established reflectance photometry and refractive index principles for chemical analysis, rather than a learned model from a training set.

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 17, 2014

SIEMENS HEALTHCARE DIAGNOSTICS, INC. SUSAN TIBEDO SR. MANAGER REGULATORY AFFAIRS 2 EDGEWATER DRIVE NORWOOD MA 02062

Re: K140717

Trade/Device Name: CLINITEK Novus Automated Urine Chemistry Analyzer CLINITEK Novus 10 Urinalysis Cassette, CLINITEK Novus Calibration Kit Regulation Number: 21 CFR 864.6550 Regulation Name: Occult Blood Test Regulatory Class: II Product Code: JIO, KQO, JJB, JIL, JIN, LJX, JMT, CEN, JIR, JRE, CDM, JIX Dated: September 29, 2014 Received: October 1, 2014

Dear Ms. Susan Tibedo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K140717

Device Name CLINITEK Novus® Automated Urine Chemistry Analyzer CLINITEK Novus® 10 Urinalysis Cassette CLINITEK Novus® Calibration Kit

Indications for Use (Describe)

The CLINITEK Novus® Automated Urine Chemistry Analyzer is a fully automated urinalysis instrument. The CLINITEK Novus analyzer is intended to read Siemens Healthcare Diagnostics CLINITEK Novus Cassettes, as well as determine urine specific gravity and urine clarity. The CLINITEK Novus 10 Urinalysis Cassette is intended for the seniquantitative measurement of the following parameters in urine: bilirubin, blood (occult), glucose, ketone (acetoacetic acid), leukocytes, nitrite (qualitative), pH, protein, color, and urobilinogen.

These measurements are used to assist diagnosis in the following areas:

  • · Carbohydrate metabolism (such as diabetes mellitus)
  • · Kidney function
  • · Liver function
  • · Metabolic disorders
  • · Urinary tract infection

For in vitro diagnostic use.

The CLINITEK Novus® Calibration Kit is intended to be used with the CLINITEK Novus Urinalysis Cassette to calibrate the CLINITEK Novus Automated Urine Chemistry analyzer. This product is for professional in vitro diagnostic use.

Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary (K140717)

This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.

1.0Submitter Information
OwnerSiemens Healthcare Diagnostics, Inc.2 Edgewater DriveNorwood, MA 02062
ContactPrimary: Susan TibedoSenior Manager Regulatory Affairssusan.tibedo@siemens.com781-269-3888 TEL
Secondary: Noor MalkiVP, Clinical, Regulatory and Quality - POC781-269-3401 TEL
Date Summary PreparedNovember 12, 2014
2.0Device Information
Proprietary NamesCLINITEK Novus® Automated Urine ChemistryAnalyzerCLINITEK Novus® 10 Urinalysis CassetteCLINITEK Novus® Calibration Kit
Common NameAutomated Urine Chemistry Analyzer
Main Classification NameAutomated urinalysis system21 CFR 862.2900, Class IProduct Code KQO

Subsequent Classifications

Classification
DeviceRegulationSectionDeviceClassProductCodeTest
Automatedurinalysissystem21 CFR 862.29001KQOAutomated Urine ChemistryAnalyzer
Bilirubin21 CFR 862.11151JJBUrinary bilirubin and its conjugates(non-quantitative) test system
Blood (occult)21 CFR 864.65502JIOOccult blood test
Glucose21 CFR 862.13402JILUrinary glucose (non-quantitative)test system
Ketone(acetoaceticacid)21 CFR 862.14351JINKetones (non-quantitative) testsystem
Leukocytes21 CFR 864.76751LJXLeukocyte peroxidase test

Page 1 of 19

{4}------------------------------------------------

Classification
DeviceRegulationSectionDeviceClassProductCodeTest
Nitrite21 CFR 862.15101JMTNitrite (non-quantitative) testsystem
pH21 CFR 862.15501CENUrinary pH (nonquantitative) testsystem
Protein21 CFR 862.16451JIRUrinary protein or albumin (non-quantitative) test system
Specificgravity21 CFR 862.28001JRERefractometer for clinical use
Urobilinogen21 CFR 862.17851CDMUrinary urobilinogen (non-quantitative) test system
Calibrator21 CFR 862.11502JIXCalibrator, multi-analyte mixture

3.0 Predicate Device

Predicate
Device NameCLINITEK Atlas
Common NameAutomated Urine Chemistry Analyzer
510(k) NumberK932674
ManufacturerSiemens Healthcare Diagnostics

4.0 Device Description

The CLINITEK Novus® system is a fully automated urine chemistry analyzer that is designed for use with the CLINITEK Novus® 10 Urinalysis Cassette. The analyzer automates the process of urine strip testing by dispensing urine samples onto a test pad, and then by reading the color and intensity of light reflected from the reacted test pads, converts the results into clinically meaningful units.

Test results from the test pads are qualitative and semi-quantitative; specific gravity is measured by a refractometer assembly.

AnalyteAbbreviationConventionalSemi-QuantitativeSISemi-QuantitativeQualitative
BilirubinBILNegativeNegativeNegative
SmallSmall1+
ModerateModerate2+
LargeLarge3+
AnalyteAbbreviationConventionalSemi-QuantitativeSISemi-QuantitativeQualitative
BloodBLONegativeNegativeNegative
TraceTraceTrace
NHTNHTNHT
SmallSmall1+
ModerateModerate2+
LargeLarge3+
GlucoseGLUNegativeNegativeNegative
100 mg/dL5.5 mmol/LTrace
250 mg/dL14 mmol/L1+
500 mg/dL28 mmol/L2+
1000 mg/dL>=55 mmol/L3+
KetoneKETNegativeNegativeNegative
TraceTraceTrace
15 mg/dL1.5 mmol/L1+
40 mg/dL3.9 mmol/L2+
80 mg/dL7.8 mmol/L3+
>=160 mg/dL>=15.6 mmol/L4+
LeukocyteLEUNegativeCa 15 cells/µLNegative
TraceCa 15 cells/µLTrace
SmallCa 70 cells/µL1+
ModerateCa 125 cells/µL2+
LargeCa 500 cells/µL3+
NitriteNITNegativeNegativeNegative
PositivePositivePositive
pHpH5.05.05.0
5.55.55.5
6.06.06.0
6.56.56.5
7.07.07.0
7.57.57.5
8.08.08.0
8.58.58.5
>=9.0>=9.0>=9.0
AnalyteAbbreviationConventionalSemi-QuantitativeSISemi-QuantitativeQualitative
ProteinPRONegativeNegativeNegative
TraceTraceTrace
30 mg/dL0.3 g/L1+
100 mg/dL1.0 g/L2+
300 mg/dL3.0 g/L3+
>=1000 mg/dL>=10.0 g/L4+
SpecificGravity(0.005 units)SG<=1.005<=1.005<=1.005
1.0101.0101.010
1.0151.0151.015
1.0201.0201.020
1.0251.0251.025
>=1.030>=1.030>=1.030
SpecificGravity(0.001 units)SG1.000 to <=1.099 in increments of 0.001
UrobilinogenUROUBG (SI)0.2 E.U./dL(mg/dL)3.2 µmol/L0.2 E.U./dL
1 E.U./dL16 µmol/L1 E.U./dL
2 E.U./dL33 µmol/L2 E.U./dL
4 E.U./dL66 µmol/L4 E.U./dL
>=8 E.U./dL>=131µmol/L>=8 E.U./dL
ColorCOLYellowYellowYellow
Dark YellowDark YellowDark Yellow
OrangeOrangeOrange
RedRedRed
GreenGreenGreen
OtherOtherOther
ClarityCLAClearClearClear
CloudyCloudyCloudy
TurbidTurbidTurbid
Sl. CloudySl. CloudySl. Cloudy

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

5.0 Intended Use Statement

The CLINITEK Novus® Automated Urine Chemistry Analyzer is a fully automated urinalysis instrument for clinical laboratory use. The CLINITEK Novus analyzer is intended to read Siemens Healthcare Diagnostics CLINITEK Novus Cassettes, as well as determine urine specific gravity and urine clarity.

The CLINITEK Novus 10 Urinalysis Cassette is intended for the semi-quantitative measurement of the following parameters in urine: bilirubin, blood (occult), glucose, ketone (acetoacetic acid), leukocytes, nitrite (qualitative), pH, protein, color, and urobilinogen.

These measurements are used to assist diagnosis in the following areas:

  • · Carbohydrate metabolism (such as diabetes mellitus)
  • Kidney function
  • Liver function
  • Metabolic disorders
  • Urinary tract infection

For in vitro diagnostic use.

The CLINITEK Novus® Calibration Kit is intended to be used with the CLINITEK Novus Urinalysis Cassette to calibrate the CLINITEK Novus Automated Urine Chemistry Analyzer. This product is for professional in vitro diagnostic use.

6.0Summary Comparison of Technological Characteristics
----------------------------------------------------------------
FeatureCandidate DeviceCLINITEK Novus analyzerPredicate DeviceCLINITEK Atlas analyzer
Intended UseAutomated Urine Chemistry Analyzerintended for professional in vitrodiagnostic use in clinical laboratories.The analyzer is intended for themeasurement of the followingcomponents in urine: bilirubin, blood(occult), glucose, ketone (acetoaceticacid), leukocytes, nitrite, pH, protein,urobilinogen and specific gravity.Same
UrinalysisStripsAnalyticalMethodLight is reflected at specificwavelengths from the test pad "readarea". Reported results depend uponthe degree of color change andintensity in the pad which is directlyrelated to the concentration of theanalyte in the urine.Same
FeatureCandidate DeviceCLINITEK Novus analyzerPredicate DeviceCLINITEK Atlas analyzer
SpecificGravityFiber optic refractive index method.Same
InstrumentOpticalSystemTwo pairs of LEDs illuminate the testpads and a camera in the analyzerelectronically analyzes the color andintensity of the light reflected from thereacted test pads.Reflectance photometerconsisting of halogen lamp,fiber optics and lens anddetector.Light travels onto a lenswhere it is focused onto thedetector, and the intensity isconverted into electricalimpulses to generateclinically meaningful units.
ReagentFormatCassette contains reagent test padsorganized in strips on a test card format.Roll containing reagent testpads in strip format.
ReagentTest PadChemistryEach reagent test pad consists ofimpregnated paper with chemicalsspecific to measuring the analyte.Same
ReagentStorage andStabilityStorage temperature: 15-30°CShelf life: 12 monthsOpen shelf life: 14 daysSame
CalibrationMethodThe system performs a cleaning cycleand dry pad calibration. Four, liquid,ready-to-use calibrators are used tocalibrate the SG sensor and certain testpads.Same without dry padcalibration
QualityCheckConducts a quality check to assess if thetest pad has been exposed to highhumidity.None
Data Storage7,500 Patient results200 Calibration results400 Quality control results1000 Patient results200 QC / calibrator results
ConnectivityLIS, HIS and remote-diagnosis ready.Optional module enables hospitalnetwork (such as HIS/LIS) connectivity.This also allows copying a selectedsetup across multiple instruments via thenetwork.Same
BarcodeInternal barcode readerOptional handheld barcode reader -automatic entry of control lot andexpirationSame
UserInterfaceIntegrated color touch screenKeypad and 12-line display

{8}------------------------------------------------

6.0 Summary Comparison of Technological Characteristics

{9}------------------------------------------------

Calibrators

Similarities and Differences

FeatureCandidate DeviceCLINITEK Novus Calibration KitPredicate DeviceCLINITEK Atlas CalibrationKit (K932674)
Intended UseIn vitro diagnostic product for thecalibration of CLINITEK AutomatedUrine Chemistry AnalyzerSame
PreparationReady to useSame
MatrixLiquidSame
Number of calibratorsFour: Calibrator #1, #2, #3 and #4Same
Cleaning cycleIncludes a cleaning cycle beforecalibrationDoes not include a cleaning cycle before calibration
StabilityShelf-life: 18 months stored at 2-8°COpen bottle: 18 months stored at 2-8°CSame
TraceabilityCalibrators have been tested usingmaterials or calibrated devices whichare traceable to the National Instituteof Standards and Technology inconformance with standard EN17511:2001.Same
Value AssignmentCalibrators are prepared to achievetargeted analyte values. Thecalibrators have fixed values and arenot lot specific. Calibrators mustmeet internal specifications beforerelease.Same

7.0 Test Principle

The CLINITEK Novus analyzer utilizes an optical system that has a digital imager. The concentration of each analyte is measured by the color change that develops when a sample is deposited on a test pad. The imager records the color and intensity of the light reflected off the reacted tested pads, which are illuminated by two pairs of LED lights. The optical system converts this data into clinically meaningful results.

Specific gravity (SG) is determined by the fiber optic refractive index method. The SG sensor measures the amount of light passing through the fiber optic at one end. Because the refractive index is proportional to specific gravity, the light measured correlates to the specific gravity of the sample yielding a quantitative result.

{10}------------------------------------------------

8.0 Performance Characteristics

Analytical Performance

a. Precision / Reproducibility

Precision was evaluated on 3 CLINITEK Novus analyzers over 10 days of testing. Two runs per day were completed with 2 replicates per run across 1 lot of CLINITEK Novus reagent urinalysis cassettes. Commercially available controls were used providing a high positive control, a low clinically significant positive control and a negative control for evaluating the percent agreement. The results are shown in the following tables:

AnalyteControl LimitsExact Block Agreement (%)Agreement within ±1 Block (%)
BilirubinNegative120/120 (100%)120/120 (100%)
BloodNegative119/120 (99.2%)120/120 (100%)
GlucoseNegative120/120 (100%)120/120 (100%)
KetoneNegative120/120 (100%)120/120 (100%)
LeukocyteNegative120/120 (100%)120/120 (100%)
NitriteNegative120/120 (100%)120/120 (100%)
pH< 6.5120/120 (100%)120/120 (100%)
ProteinNegative120/120 (100%)120/120 (100%)
SG< 1.005120/120 (100%)120/120 (100%)
Urobilinogen< 1.0 mg/dL114/120 (95%)120/120 (100%)

Table 1. Reproducibility (Within Device) for Low / Negative Control

Table 2. Reproducibility (Within Device) for Low Positive Control

AnalyteControl LimitsExact BlockAgreementAgreement within ±1 Block
BilirubinModerate120/120 (100%)120/120 (100%)
BloodTrace120/120 (100%)120/120 (100%)
Glucose100 mg/dL119/120 (99.2%)120/120 (100%)
Ketone15 mg/dL120/120 (100%)120/120 (100%)
LeukocyteSmall109/120 (90.8%)120/120 (100%)
рн7.0-8.0120/120 (100%)120/120 (100%)
Protein30 mg/dL120/120 (100%)120/120 (100%)
ട്ട1.006 — 1.019120/120 (100%)120/120 (100%)
Urobilinogen2 mg/dL120/120 (100%)120/120 (100%)

Table 3. Reproducibility (Within Device) High Positive Control

AnalyteControl LimitsExact BlockAgreementAgreement within ±1 Block
BilirubinLarge120/120 (100%)120/120 (100%)
BloodLarge120/120 (100%)120/120 (100%)
Glucose1000 mg/dL117/120 (97.5%)120/120 (100%)
Ketone80 mg/dL108/120 (90%)120/120 (100%)
LeukocyteLarge120/120 (100%)120/120 (100%)
NitritePositive120/120 (100%)120/120 (100%)

Siemens Healthcare Diagnostics CLINITEK Novus® Automated Urine Chemistry Analyzer / Traditional 510(k) Page 8 of 19

{11}------------------------------------------------

AnalyteControl LimitsExact BlockAgreementAgreement within ±1 Block
pH>9.0120/120 (100%)120/120 (100%)
Protein300 mg/dL120/120 (100%)120/120 (100%)
SG>/= 1.020120/120 (100%)120/120 (100%)
Urobilinogen4 mg/dL115/120 (95.8%)120/120 (100%)
Table 3. Reproducibility (Within Device) High Positive Control

b. Detection Limit

The sensitivity of the CLINITEK Novus urinalysis cassette reagents was evaluated by spiking negative pooled human urine with known quantity of analytes. The analytes were spiked at multiple concentrations around the transition value from negative to positive. Multiple samples were tested at each concentration on the CLINITEK Novus analyzer. Sensitivity was defined as the concentration in which ≥ 55% of the results were positive. The lower limits of analyte detection for each reagent test are shown in the following table:

AnalyteLimit of Detection(Sensitivity)
Bilirubin0.5 mg/dL
Blood0.013 mg/dL
Glucose36 mg/dL
Ketone3.6 mg/dL
Leukocyte6.0 cells/µL
Nitrite0.06 mg/dL
Protein10.8 mg/dL
Urobilinogen0.24 mg/dL

c. Analytical Specificity

Urine samples were spiked with a high concentration of potentially interfering substances. Substances that resulted in a change in the reported clinical block output were serially diluted to determine the concentration at which the substance interfered. The results are in the following table:

{12}------------------------------------------------

InterferingSubstanceInitial ConcTestedBILBLOGLUKETLEUNITPROURO
Acetaminophen40 mg/dLNoneNoneNoneNoneNoneNoneNoneNone
Acetylcysteine270 mg/dLFalseNegative at135 mg/dLFalseNegative at67.5 mg/dLFalseNegative at67.5 mg/dLFalsePositive at67.5 mg/dLNoneNoneNoneFalseNegative at67.5 mg/dL
Ascorbic Acid100 mg/dLFalseNegative at25 mg/dLFalseNegative at25 mg/dLFalseNegative at25 mg/dLNoneNoneFalseNegative at100 mg/dLNoneNone
AmmoniumChloride200 mg/dLNoneNoneNoneNoneNoneNoneNoneNone
Boric Acid1000 mg/dLNoneNoneNoneFalseNegative at500 mg/dLNoneNoneNoneNone
CalciumChloride100 mg/dLNoneNoneNoneNoneFalseNegative at100 mg/dLNoneNoneNone
Captopril50 mg/dLNoneFalseNegative at12.5 mg/dLNoneFalsePositive at12.5 mg/dLNoneNoneNoneFalseNegative at25 mg/dL
Cefoxitin66 mg/dLNoneNoneNoneNoneNoneNoneNoneNone
Cephalosporin80 mg/dLNoneNoneNoneNoneNoneNoneNoneNone
Chlorhexidine150 mg/dLNoneFalseNegative at112.5 mg/dLNoneNoneFalseNegative at37.5 mg/dLNoneFalsePositive at37.5 mg/dLFalseNegative at75 mg/dL
Chloroquine40 mg/dLNoneFalseNegative at40 mg/dLNoneNoneFalseNegative at20 mg/dLNoneFalsePositive at20 mg/dLNone
InterferingSubstanceInitial ConcTestedBILBLOGLUKETLEUNITPROURO
Citric Acid80 mg/dLFalseNegative at40 mg/dLNoneNoneNoneNoneNoneNoneNone
Curcuma300 mg/dLFalseNegative at75 mg/dLFalseNegative at75 mg/dLNoneFalsePositive at75 mg/dLFalsePositive at75 mg/dLFalsePositive at75 mg/dLNoneFalseNegative/Positive at75 mg/dL
Formalin370 mg/dLNoneFalseNegative at185 mg/dLNoneFalseNegative at92.5 mg/dLFalsePositive at92.5 mg/dLFalseNegative at185 mg/dLNoneFalseNegative at92.5 mg/dL
Galactose20 mg/dLNoneNoneNoneNoneNoneNoneNoneNone
GentamicinSulfate6 mg/dLNoneNoneNoneNoneNoneNoneNoneNone
Glucose5 g/dLNoneNoneNANoneFalseNegative at1250 mg/dLNoneNoneFalseNegative at2500 mg/dL
Glycine400 mg/dLNoneNoneNoneNoneNoneNoneNoneNone
Glyburide15 ug/mLNoneNoneNoneNoneNoneNoneNoneNone
Hemoglobin5 mg/dLNoneNoneNoneNoneNoneNoneNoneNone
Hypochlorite2.60%FalseNegative at1.3%FalsePositive at0.65%FalsePositive at1.3%FalsePositive at1.3%FalseNegative at0.65%NoneNoneFalseNegative at0.65%
Hydrochloro-thiazide6 mg/mLNoneNoneNoneNoneNoneNoneNoneFalseNegative at6 mg/dL
Indican50 mg/dLNoneNoneNoneNoneNoneNoneNoneNone
Imipenem100 mg/dLNoneNoneNoneFalsePositive at50 mg/dLNoneNoneNoneNone
InterferingSubstanceInitial ConcTestedBILBLOGLUKETLEUNITPROURO
Keflex150 mg/dLNoneNoneNoneNoneNoneNoneNoneNone
Keflin50 mg/dLNoneNoneNoneNoneNoneNoneNoneNone
Lactose20 mg/dLNoneNoneNoneNoneNoneNoneNoneFalseNegative at20 mg/dL
Levadopa75 mg/dLNoneFalseNegative at18.8 mg/dLNoneFalsePositive at56 mg/dLNoneNoneNoneNone
Lodine100 mg/dLNoneNoneNoneNoneNoneNoneNoneNone
Meropenem100 mg/dLNoneNoneNoneNoneNoneNoneNoneFalseNegative at100 mg/dL
Mesna80 mg/dLFalseNegative at20 mg/dLFalseNegative at20 mg/dLFalseNegative at20 mg/dLFalsePositive at20 mg/dLNoneNoneNoneFalseNegative at20 mg/dL
Metformin4 mg/mLNoneNoneNoneNoneNoneNoneNoneNone
Nitrofurantoin120 mg/dLNoneNoneNoneNoneNoneNoneNoneFalseNegative at120 mg/dL
Oxalic Acid95 mg/dLNoneNoneNoneNoneNoneFalseNegative at71 mg/dLNoneNone
P-aminosalicyclicacid250 mg/dLNoneNoneNoneNoneNoneNoneNoneFalseNegative at62.5 mg/dL
Penicillin300 mg/dLNoneNoneNoneNoneNoneNoneNoneNone
Pyridium200 mg/dLNoneNoneNoneNoneNoneFalsePositive at50 mg/dLFalsePositive at150 mg/dLFalsePositive at100 mg/dL
InterferingSubstanceInitial ConcTestedBILBLOGLUKETLEUNITPROURO
PotassiumChloride1200 mg/dLNoneNoneNoneNoneNoneNoneNoneNone
Quinidine30 mg/dLNoneNoneNoneNoneNoneNoneNoneNone
Riboflavin120 mg/dLNoneNoneNoneNoneNoneFalsePositive at30 mg/dLNoneNone
Sodium Nitrite3 mg/dLNoneNoneNoneNoneNoneNANoneNone
Sulfamethoxazole70 mg/dLNoneNoneNoneNoneNoneNoneNoneFalseNegative at17.5 mg/dL
Tagamet10 mg/dLNoneNoneNoneNoneNoneNoneNoneFalseNegative at5 mg/dL
Tetracycline12 mg/dLNoneNoneNoneNoneNoneNoneNoneNone
pH > 9NANoneNoneNoneNoneNoneNoneNoneNone
MicrobialPeroxidaseNANoneFalsePositiveNoneNoneNoneNoneNoneNone

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

Comparison Studies

d. Method Comparison with Predicate Device

The method comparison study compared the clinical results of the CLINITEK Novus analyzer to the predicate device. The study was conducted across 3 sites with up to 2773 specimens tested. The results for overall percent agreement and within one level agreement met acceptance criteria. In the following tables, results are shown for all sites combined.

BILPredicate Result
NegativeSmallModerateLarge
CLINITEKNovusResultNegative223410411
Small5190151
Moderate11127
Large102
Total2239295128111
% exact match99.8%64.4%87.5%91.9%
% within 1 block100.0%100.0%99.2%98.2%
BLOPredicate Result
NegativeTraceSmallModerateLarge
CLINITEKNovusResultNegative148941
Trace/NHT349712
Small1812710
Moderate512514
Large1204
Total1523119145136218
% exact match97.8%81.5%87.6%91.9%93.6%
% within 1 block100.0%100.0%99.7%100.0%100.0%

{17}------------------------------------------------

GLUPredicate Result
Negative100mg/dL250mg/dL500mg/dL1000mg/dL
Negative1563
100mg/dL7147
CLINITEKNovusResult250mg/dL1995
500mg/dL8884
1000mg/dL4183
Total157016610392187
% exact match99.6%88.5%92.2%95.7%97.9%
% within 1 block100.0%100.0%100.0%100.0%100.0%
KETPredicate Result
NegativeTrace15 mg/dL40 mg/dL80 mg/dL160 mg/dL
Negative1670141*
CLINITEKNovusResultTrace21589
15 mg/dL2826
40 mg/dL21171
80 mg/dL66
160 mg/dL528
Total1672174931247228
% exact match99.9%90.8%88.2%94.4%91.7%100.0%
% within 1 block100.0%100.0%100.0%99.2%100.0%100.0%
  • This is a highly colored sample. Predicate labeling cautions testing of highly colored samples as they may yield a falsely elevated result.

{18}------------------------------------------------

LEUCLINITEK Atlas Result
NegativeTraceSmallModerateLarge
CLINITEKNovusResultNegative160051
Trace798013
Small1101293
Moderate16855
Large6157
Total16809515994162
% exact match95.0%85.7%81.1%90.4%96.6%
% within 1 block99.9%100.0%99.7%100.0%100.0%
NITPredicate Result
NegativePositive
CLINITEKNovusResult1707
12399
Total1719399
% exact match99.3%100.0%
% within 1 block100.0%100.0%
pH*Predicate Result
55.566.577.588.5
CLINITEKNovusResult585121
5.51203213
6142619
6.52639910
7333762
7.5106178
8164441
8.515149
>=9.02361
Total8632566645149224597411
% exact match98.8%62.5%64.0%88.5%76.4%72.7%45.4%99.8%
% within 1 block100.0%100.0%99.9%100.0%100.0%99.6%97.5%100.0%

*Results on CLINITEK Novus of ≥8.5 were considered a match to a pH of 8.5 on the predicate.

{19}------------------------------------------------

PROPredicate Result
NegativeTrace30mg/dL100mg/dL300mg/dL1000mg/dL
Negative15112
CLINITEKNovusResultTrace242074
30mg/dL201843
100mg/dL171451
300mg/dL7778
1000mg/dL71
Total15362291951557879
% exact match98.4%90.4%94.4%93.5%98.7%89.9%
% within 1 block99.9%100.0%100.0%100.0%100.0%100.0%
UROPredicate Result
0.2 mg/dL1 mg/dL2 mg/dL4 mg/dL8mg/dL
CLINITEKNovusResult0.2mg/dL198253
1 mg/dL232715
2 mg/dL220910
4 mg/dL51163
8 mg/dL445
Total198438222913048
% exact match99.9%85.6%91.3%89.2%93.8%
% within 1 block100.0%100.0%100.0%100.0%100.0%

{20}------------------------------------------------

As the number of contrived samples for URO exceeded 15%, testing was also performed at a fourth site. For this fourth site, less than 5% of the samples were contrived. Results are in the following table:

UROPredicate Result
0.2mg/dL1 mg/dL2 mg/dL4 mg/dL8 mg/dL
CLINITEKNovusResult0.2 mg/dL1872
1 mg/dL46
2 mg/dL10
4 mg/dL12
8 mg/dL4
Total1874810124
% exact match100.0%95.8%100.0%100.0%100.0%
% within 1 level100.0%100.0%100.0%100.0%100.0%

Overall Specific Gravity Slope and Intercept

SiteNR2SlopeIntercept
Overall19930.9831.020-0.019
  • c. Matrix Comparison
    Not applicable. The device is for urine samples only.

Clinical Studies

a. Clinical Sensitivity Not applicable

b. Clinical Specificity Not applicable

Clinical Cut-off

Not applicable

Expected Values / Reference Range

Complete details of the expected analyte values are provided in the labeling.

9.0 Instrument Name:

CLINITEK Novus Automated Urine Chemistry Analyzer

{21}------------------------------------------------

10.0 System Descriptions

Modes of Operation

Single and continuous testing.

Specimen Identification

An internal barcode reader scans the barcode labels on the sample tubes in the racks or sample identification may be entered manually using the touch screen keyboard.

Specimen Sampling and Handling

The CLINITEK Novus system can analyze multiple sample pipette aspirates a sample from the sample tube and dispenses sample onto each test pad.

Calibration

The CLINITEK Novus system uses four liquid calibration solutions to perform a calibration reflectance check. The calibrators have a fixed value with no lot specific values. A calibration must be performed when a new lot of reagent cassette is loaded onto the system or a new cassette of the same lot is loaded and the calibration is older than 24 hours.

The calibrators are identical to those cleared in K932674.

Quality Control

Recommendations for testing quality control are in the labeling. Routinely test at least 2 levels (negative/low and positive) of an appropriate commercially available control to confirm performance of the CLINITEK Novus urinalysis cassette. Controls may also be randomly placed in each batch of samples tested. Each laboratory should establish its own qoals for acceptable standards of performance, and should question handling and testing procedures if these standards are not met.

The CLINITEK Atlas controls, cleared in K932674, may be used with the CLINITEK Novus automated urine chemistry analyzer.

Other Supportive Instrument Performance Characteristics Data Not Covered in the "Performance Characteristics" Section above:

None

11.0 Conclusion

The results of these studies demonstrate that the CLINITEK Novus automated urine chemistry analyzer and CLINITEK Novus 10 urinalysis cassette is similar to the predicate in both Technological Characteristics and Intended Use. The data presented is a summary of external clinical evaluation, internal laboratory evaluation, and software development information. The CLINITEK Novus performance was shown to be substantially equivalent to the predicate urinalysis device.

§ 862.2900 Automated urinalysis system.

(a)
Identification. An automated urinalysis system is a device intended to measure certain of the physical properties and chemical constituents of urine by procedures that duplicate manual urinalysis systems. This device is used in conjunction with certain materials to measure a variety of urinary analytes.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.