K Number
K140714

Validate with FDA (Live)

Date Cleared
2014-07-16

(117 days)

Product Code
Regulation Number
872.3930
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Porcine Anorganic Bone Mineral is intended for use in dental surgery. The products may be used in surgical procedures such as:

  • Augmentation or reconstructive treatment of alveolar ridge
  • Filling of infrabony periodontal defects
  • Filling of defects after root resection, apicoectorny, and cystectorny
  • Filling of extraction sockets to enhance preservation of the alveolar ridge
  • Elevation of maxillary sinus floor
  • Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
  • Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration.
Device Description

Porcine Anorganic Bone Mineral is a porous bone graft material consisting predominantly of calcium phosphate supplied in granular form for use in dental surgeries. The anorganic bone mineral is produced by removal of the organic components from porcine bone. The composition of Porcine Anorganic Bone Mineral meets the requirements of ASTM F1581 Standard Specification for Composition of Anorganic Bone for Surgical Implants. The anorganic bone mineral matrix is biocompatible, having interconnecting macro- and microscopic porous structure that supports the formation and ingrowth of new bone at the implantation site. The use of Porcine Anorganic Bone Mineral may be considered when autogenous bone is not indicated, or insufficient in quantity to fulfill the needs of the proposed surgical procedure. The granular product is supplied sterile, non-pyrogenic, and for single use only.

AI/ML Overview

The provided text is a 510(k) summary for a medical device (Porcine Anorganic Bone Mineral), not a study describing the acceptance criteria and performance of an AI/ML device. Therefore, much of the requested information, such as sample sizes for test sets, expert qualifications, and details about AI performance, is not present.

However, I can extract information related to the acceptance criteria for the physical properties of the Porcine Anorganic Bone Mineral device and the types of studies conducted to demonstrate its substantial equivalence to predicate devices.

Here's the analysis based on the provided text, focusing on what is available:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Specification)Reported Device Performance (as stated for "Porcine Anorganic Bone Mineral")
AppearanceWhite to off-white granules
Mineral StructureCarbonate Apatite
Material CompositionCalcium Phosphate
Particle Size0.25 – 1 mm and 1 – 2 mm
pH6 - 9.5
Nitrogen ContentMeets Specification (ASTM 1581)
BiocompatibilityDetermined safe as an implantable material (based on testing)
Viral InactivationPerformed to ensure viral safety
Composition (General)Meets requirements of ASTM F1581 Standard Specification for Composition of Anorganic Bone for Surgical Implants
Porous StructureInterconnecting macro- and microscopic porous structure

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not applicable in the context of an AI/ML device. For the material characterization and biocompatibility studies, specific sample sizes are not provided in this summary. For the animal efficacy studies, it mentions "two separate animal efficacy studies," but the number of animals used in each study is not stated.
  • Data Provenance:
    • Material Characterization & Biocompatibility: Performed by the applicant, Collagen Matrix, Inc. The location of testing is not specified, but the company address is Oakland, New Jersey, USA.
    • Animal Efficacy Studies: Conducted in "a dog dental intrabony defect model" and "a rabbit femoral condyle defect model." The geographical location of these studies is not specified.
  • Retrospective or Prospective: These non-clinical and animal studies are inherently prospective, as they are designed experiments to evaluate the device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable as this is not an AI/ML device requiring expert ground truth for classification. The "ground truth" here is established through objective physical/chemical measurements and biological assessment according to recognized standards and protocols.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable for a physical device evaluation. Adjudication methods are typically relevant for human interpretation tasks, especially in AI/ML performance evaluation where ground truth might be debatable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is not an AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable as this is not an AI/ML device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For Material Characterization: The "ground truth" is based on the specifications defined by ASTM F1581, general material science principles, and comparison to the predicate devices. This involves objective physical and chemical testing.
  • For Biocompatibility: The "ground truth" is established by adherence to ISO 10993 standards for biological evaluation of medical devices, which includes tests for cytotoxicity, genotoxicity, systemic toxicity, local effects after implantation, carcinogenicity, and reproductive toxicity. The outcome (safe/unsafe) is determined by the results of these standardized tests.
  • For Animal Efficacy Studies: The "ground truth" is assessed through the observed biological response in the animal models (e.g., new bone formation, integration) as compared to the predicate device. This would typically involve histological analysis, imaging, and other measures of tissue regeneration and healing.
  • For Viral Inactivation: The "ground truth" is established by adherence to ISO 22442 standards, demonstrating that the manufacturing process effectively eliminates or inactivates potential viruses and transmissible agents.

8. The sample size for the training set

  • Not applicable. This is not an AI/ML device that requires a training set. The development of the device itself would involve numerous R&D iterations, but not a formally defined "training set" in the AI sense.

9. How the ground truth for the training set was established

  • Not applicable. See point 8.

{0}------------------------------------------------

JUL 16 2014

510(k) SUMMARY

1.Applicant Information
Applicant Name:Address:Collagen Matrix, Inc.15 Thornton RoadOakland, New Jersey 07436
Telephone:Fax:Contact Person:(201) 405-1477(201) 405-1355Peggy Hansen, RACVP, Clinical, Regulatory, QA, and Marketing
Date Prepared:June 16, 2014
2.Name of the Device
Device Trade Name:Device Common Name:Device Classification Name:Porcine Anorganic Bone, MineralBone Grafting MaterialBone Grafting Material, Animal Source21 CFR 872.3930Product Code NPMDevice Class II
3.Legally Marketed Devices to Which Substantial Equivalence is Claimed
Predicate Device(s):OsteoGuide™ Anorganic Bone MineralK043034
Bio-Oss® GranulesK952617

4. Description of the Device

Porcine Anorganic Bone Mineral is a porous bone graft material consisting predominantly of calcium phosphate supplied in granular form for use in dental surgeries. The anorganic bone mineral is produced by removal of the organic components from porcine bone. The composition of Porcine Anorganic Bone Mineral meets the requirements of ASTM F1581 Standard Specification for Composition of Anorganic Bone for Surgical Implants. The anorganic bone mineral matrix is biocompatible, having interconnecting macro- and microscopic porous structure that supports the formation and ingrowth of new bone at the implantation site. The use of Porcine Anorganic Bone Mineral may be considered when autogenous bone is not indicated, or insufficient in quantity to fulfill the needs of the proposed surgical procedure.

The granular product is supplied sterile, non-pyrogenic, and for single use only.

{1}------------------------------------------------

ર. Intended Use

Porcine Anorganic Bone Mineral is intended for use in dental surgery. The products may be used in surgical procedures such as:

  • Augmentation or reconstructive treatment of alveolar ridge -
  • Filling of infrabony periodontal defects l
  • Filling of defects after root resection, apicoectorny, and cystectorny -
  • Filling of extraction sockets to enhance preservation of the alveolar ridge -
  • Elevation of maxillary sinus floor -
  • Filling of periodontal defects in conjunction with products intended for Guided -Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
  • l Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration

6. Summary of Technical Characteristics

Porcine Anorganic Bone Mineral has been determined to be substantially equivalent to the predicate devices having similar technological characteristics as follows:

Appearance:White to off-white granules
Mineral Structure:Carbonate Apatite
Material Composition:Calcium Phosphate
Particle Size:0.25 – 1 mm and 1 – 2 mm
pH:6 - 9.5
Nitrogen Content:Meets Specification (ASTM 1581)

The final composition of the anorganic bone mineral for both devices consists of a calcium phosphate mineral in a carbonate apatite that is produced by removal of organic components from the bone tissue structure.

7. Discussion of Non-clinical Testing

The substantial equivalence of Porcine Anorganic Bone Mineral and its predicates was demonstrated based on in vitro characterization studies, biocompatibility studies, and animal efficacy studies.

Non-clinical testing was performed in accordance with FDA recognized consensus standards and FDA quidelines as follows:

FDA Guidance Document entitled, "Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices," issued on April 28, 2005

ASTM F1581, Standard specifications for composition of anorganic bone for surgical implants.

{2}------------------------------------------------

ISO 22442-1 Animal Tissues and Their Derivatives Utilized in the Manufacture of Medical Devices - Part 1 Analysis and Risk Management

ISO 22442-2 Animal Tissues and Their Derivatives Utilized in the Manufacture of Medical Devices - Part 2 Controls on Sourcing, Collection, and Handling

ISO 22442-3 Animal Tissues and Their Derivatives Utilized in the Manufacture of Medical Devices - Part 3 Validation of the Elimination and/or Inactivation of Viruses and Transmissible Agents

ISO 10993-3:2009 Biological Evaluation of Medical Devices- Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity

ISO 10993-5:2009 Biological Evaluation of Medical Devices- Part 5: Tests for in vitro cytotoxicity

ISO 10993-6:2009 Biological Evaluation of Medical Devices- Part 6: Test for local effects after implantation

ISO 10993-10:2009 Biological Evaluation of Medical Devices- Part 10 Test for local effects after implantation

ISO 10993-11:2009 Biological Evaluation of Medical devices – Part 11 Tests for systemic toxicity

Non-clinical Testing Conducted

Material characterization testing was performed to demonstrate substantial equivalence of the subject device to its predicate devices. A series of bench tests were conducted to verify the design characteristics such as material composition. structure, physic-chemical properties and resorption profile as compared to the predicate devices.

A series of in vitro and in vivo biocompatibility testing was performed to assess safety of the Porcine Anorganic Bone Mineral as an implantable material.

Two separate animal efficacy studies were conducted to evaluate the device as compared to its predicate device; one in a dog dental intrabony defect model and the other in a rabbit femoral condyle defect model.

Viral inactivation studies were performed to ensure the viral safety of the product.

8. Conclusion of Non-clinical Studies

The results of the material characterization and biocompatibility testing, as well as animal efficacy studies show that Porcine Anorganic Bone Mineral is substantially equivalent to the identified predicate devices.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol consists of three stylized human figures or shapes that are interconnected. The text is arranged in a circular pattern around the symbol.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 16, 2014

Collagen Matrix, Incorporated Ms. Peggy Hansen, RAC Vice President, Clinical, Regulatory, QA and Marketing 15 Thornton Road Oakland, New Jersey 07436

Re: K140714

Trade/Device Name: Porcine Anorganic Bone Mineral Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: NPM Dated: June 16, 2014 Received: June 17, 2014

Dear Ms. Hansen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Hansen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Digitally signed
by Richard C.
Chapman -S
Date: 2014.07.16
08:15:10 -04'00'
for

Erin 1. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

K140714 510(k) Number (if known): __

Device Name: _ Porcine Anorganic Bone Mineral

Indications for Use:

Porcine Anorganic Bone Mineral is intended for use in dental surgery. The products may be used in surgical procedures such as:

  • Augmentation or reconstructive treatment of alveolar ridge ●
  • Filling of infrabony periodontal defects ●
  • Filling of defects after root resection, apicoectorny, and cystectorny ●
  • Filling of extraction sockets to enhance preservation of the alveolar ridge �
  • Elevation of maxillary sinus floor ●
  • Filling of periodontal defects in conjunction with products intended for Guided . Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
  • Filling of peri-implant defects in conjunction with products intended for Guided . Bone Regeneration.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sheena A. Green -S 2014.07.15 08:21:02 -04'00'

Page 1 of 1

§ 872.3930 Bone grafting material.

(a)
Identification. Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.(b)
Classification. (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.