K Number
K140705

Validate with FDA (Live)

Date Cleared
2014-12-05

(259 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BL Navigated Biopsy Needle is indicated for stereotactic biopsy of cranial tissue using BrainLab AG's VectorVision navigation system. The Navigated Biopsy Needle is a pre-sterilized, single-use, side-cutting needle where the cutting action is achieved by rotation of an inner cannula within an outer cannula.

Device Description

BL Navigated Biopsy Needle contains a calibrated biopsy cannula that is used with a compatible image guided surgery navigation system in stereotactic biopsy of cranial tissue. The device is compatible with specified BrainLab navigation systems (see intended use section below). Like the predicate device, the BL Navigated Biopsy Needle is a pre-sterilized, single-use, stainless steel device and uses a side-cutting cannula where the cutting action is achieved by rotation of an inner cannula within an outer cannula.

BL Navigated Biopsy Needle's components include a navigable biopsy cannula, a biopsy cannula stop with screw, a stop adjustment tool and an aspiration tube.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the "Navigable Brain Biopsy Cannula Set" (K140705). This document focuses on demonstrating substantial equivalence to predicate devices, rather than on a study proving the device meets specific acceptance criteria in the context of AI/algorithm performance. The device described is a physical medical instrument (a biopsy needle), not an AI or software-based device.

Therefore, many of the requested criteria about AI/algorithm performance studies (e.g., sample size for AI test/training sets, ground truth establishment for AI, MRMC studies for AI, standalone AI performance) are not applicable to the information contained in this 510(k) summary.

However, I can extract the relevant information regarding the performance testing conducted for this physical device.

Here's a breakdown of the available information:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state numerical "acceptance criteria" for the performance tests in a table format, nor does it provide quantitative performance metrics beyond confirming the tests were conducted and the device "performs as intended."

Reported Device Performance:

  • Tissue extraction testing: Verified the ability of the device to sample tissue.
  • Navigation testing: Verified compatibility and accuracy with specified navigation systems (BrainLab AG's VectorVision navigation system).
  • Real-time shelf life testing: Verified no change in visual attributes, dimensional characteristics, or tensile strength after 5 years of real-time storage.
  • Biocompatibility testing:
    • Cytotoxicity: Performed in accordance with ISO 10993-5.
    • Pyrogen Testing: Performed in accordance with ANSI/AAMI ST72:2011.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: Not specified for any of the performance tests (tissue extraction, navigation, shelf life, or biocompatibility).
  • Data Provenance: Not specified. These appear to be bench tests, not clinical studies involving patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. The ground truth for this physical device's performance tests would be based on objective measurements and established standards (e.g., tissue presence in a sample, accurate navigation tracking, material integrity, biocompatibility standards), not expert consensus on images or interpretations.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable for the type of bench testing described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC study was not done. This device is a physical biopsy cannula, not an AI or software diagnostic tool, so such a study is not relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No, a standalone algorithm performance study was not done. This device is a physical biopsy cannula.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Tissue extraction testing: The ground truth would likely be the physical presence and integrity of the tissue sample obtained.
  • Navigation testing: The ground truth would be the known, actual position or trajectory of the cannula compared to the system's displayed position, verified through metrology or other accurate measurement methods.
  • Shelf life testing: Ground truth would be objective measurements of material properties (visual, dimensional, tensile strength) against baseline or established specifications.
  • Biocompatibility testing: Ground truth is defined by the standards (ISO 10993-5, ANSI/AAMI ST72:2011) which specify thresholds for cytotoxicity and endotoxin levels.

8. The sample size for the training set

  • Not applicable. This is a physical device, not an AI or machine learning algorithm requiring a training set.

9. How the ground truth for the training set was established

  • Not applicable. As above, this is a physical device, not an AI/ML algorithm.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 5, 2014

IZI Medical Products Qiang Cao Manager of Ouality Assurance and Regulatory Affairs 5 Easter Court, Suite J Owings Mills, MD 21117

Re: K140705 Trade/Device Name: Navigable Brain Biopsy Cannula Set Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: November 4, 2014 Received: November 6, 2014

Dear Mr. Qiang Cao,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements

{1}------------------------------------------------

as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carlos L. Pena -S 同公

Carlos L. Peña, Ph.D., M.S. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K140705

Device Name

IZI Navigable Brain Biopsy Cannula Set

Indications for Use (Describe)

The BL Navigated Biopsy Needle is indicated for stereotactic biopsy of crainlab AG's VectorVision navigation system. The Navigated Biopsy Needle is a pre-sterilized, single-use, side-cutting needle where the cutting action is achieved by rotation of an inner cannula within an outer cannula.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

510(k) Owner

IZI Medical Products LLC

5 Easter Court, Suite J

Owings Mills, MD 21131

Phone: (410) 594-9403

Fax: (410) 594-0540

Date Summary Prepared: July 25, 2014

Device Name

Trade Name: BL Navigated Biopsy Needle

Common or Usual Name: Stereotactic Biopsy Needle

Classification Name: Neurological Stereotactic Instrument (21 CFR 882.4560)

Product Code: HAW (Neurological Stereotactic Instrument)

Predicate Device

VectorVision Frameless Biopsy System (K012564)

Medtronic Biopsy Needle (K971247)

Device Description

BL Navigated Biopsy Needle contains a calibrated biopsy cannula that is used with a compatible image guided surgery navigation system in stereotactic biopsy of cranial tissue. The device is compatible with specified BrainLab navigation systems (see intended use section below). Like the predicate device, the BL Navigated Biopsy Needle is a pre-sterilized, single-use, stainless steel device and uses a side-cutting cannula where the cutting action is achieved by rotation of an inner cannula within an outer cannula.

{4}------------------------------------------------

BL Navigated Biopsy Needle's components include a navigable biopsy cannula, a biopsy cannula stop with screw, a stop adjustment tool and an aspiration tube.

Indications for Use

The BL Navigated Biopsy Needle is indicated for stereotactic biopsy of cranial tissue using BrainLab AG's VectorVision navigation system. The Navigated Biopsy Needle is a pre-sterilized, single-use, side-cutting needle where the cutting action is achieved by rotation of an inner cannula within an outer cannula.

Patient Population

The device is intended for use in patients undergoing stereotactic cranial tissue biopsy procedures.

Environment of Use

The device is intended for use by a trained healthcare professional in a healthcare facility.

Summary of Technological Characteristics

The BL Navigated Biopsy Needle contains a navigable biopsy cannula constructed with an inner and outer stainless steel needle with a side-cutting window, a plastic hub and a set of retro-reflective discs. The reflective discs allow the device to be registered with BrainLab navigation systems and to be visible during a biopsy procedure. The Set also contains a biopsy cannula stop with screw, a stop adjustment tool and an aspiration tube.

The BL Navigated Biopsy Needle and the predicated devices, the BrainLab Biopsy Needle Type A (K012564) and Medtronic Biopsy Needle (K971247), are based on the following same technological elements: material, construction, registration and tracking method and cutting actions. The table below summarizes the comparison of technological characteristics among the subject device and predicate devices.

Technical CharacteristicsSubject Device (K140705)BrainLab Biopsy Needle Type AMedtronic Biopsy Needle
Needle MaterialStainless steelStainless steelStainless steel
Biopsy Needle ConstructionInner and outer cannulaInner and outer cannulaInner and outer cannula
Biopsy Cutting ActionSide cuttingSide cuttingSide cutting

{5}------------------------------------------------

windowwindowwindow
Registration and Trackingmethodretro-reflectivematerialretro-reflectivematerialretro-reflectivematerial
Needle Dimension (diameter)2.1 mm1.8 mm2.1 mm

There are minor differences in needle dimension between the devices.

Summary of Performance Testing

Bench testing was conducted to demonstrate that the BL Navigated Biopsy Needle performs as intended and is substantially equivalent to the predicate device. Tissue extraction testing was conducted to verify the ability of the device to sample tissue. Navigation testing was conducted to verify the compatibility and accuracy of the device with the specified navigation systems. Real-time shelf life testing was conducted to verify that there was no change in visual attributes, dimensional characteristics, or tensile strength after 5 years real time storage. These studies verify that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device identified above.

The biocompatibility evaluation for the BL Navigated Biopsy Needle was conducted in accordance with ISO 10993-5 Biological evaluation of medical devices -- Part 5: Tests for In Vitro Cytotoxicity and ANSI/AAMI ST72:2011 Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing. The battery of testing included the following tests:

  • Cytotoxicity ●
  • Pyrogen Testing

The patient contacting components of the Navigated Biopsy Needle are the outer and inner cannula. Patient contact time of the device is less than 24 hours. The outer and inner cannula are both made from AISI 304 stainless steel.

Summary

In summary, the non-clinical data support the safety of the device and the hardware verification and validation testing demonstrate that the IZI Navigated Biopsy Needle, used with BrainLab Vector Vision navigation system, should perform as intended in the specified use conditions.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).