(15 days)
This product is intended to be used in displaying and viewing digital images including those of digital mammography for review and analysis by trained medical practitioners.
RadiForce RX850 is a color LCD monitor for viewing medical images including those of mammography. The color panel employs in-plane switching (IPS) technology allowing wide viewing angles. With the matrix size (or resolution) of 4.096 x 2.160 pixels (8MP), the RX850 is an alternate replacement for traditional dual head 5MP display installations. Since factory calibrated display modes, each of which is characterized by a specific tone curve (including DICOM GSDF), a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor, the tone curve is e.g. DICOM compliant regardless of the display controller used. RadiCS is application software to be installed in each workstation offering worry-free quality control of the diagnostic monitors including the RadiForce RX850 based on the OC standards and guidelines and is capable of quantitative tests and visual tests defined by them. The RadiCS and its subset, RadiCS LE, are included in this 510(k) submission as an accessory to the RadiForce RX850.
Here's an analysis of the provided text regarding the EIZO RadiForce RX850 monitor's acceptance criteria and studies:
Product: EIZO RadiForce RX850 8MP Color LCD Monitor
Study Reference: K140702 (510(k) Summary)
Acceptance Criteria and Reported Device Performance
The document does not explicitly list acceptance criteria in a quantitative table. Instead, it states that "the display characteristics of the RadiForce RX850 meet the pre-defined criteria when criteria are set" and that the "test results showed that the RadiForce RX850 has display characteristics equivalent to those of the predicate device, RadiForce RX840-MG except 2 items, each of which was determined that it would not affect observer's performance."
Based on the performance testing section, the acceptance criteria are implicitly tied to the performance metrics outlined by the AAPM Task Group 18 (TG18 guideline) and specific measurements related to display characteristics. The "reported device performance" is essentially that it met these implicit criteria and demonstrated equivalence to the predicate.
While quantitative acceptance criteria are not presented in a table with reported performance, the document lists various performance aspects that were tested. These can be inferred as the targets for acceptance.
| Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|
| Conformance to DICOM GSDF (as per TG18) | Met (The display characteristics meet the pre-defined criteria) |
| Acceptable angular dependency of luminance response (horizontal, vertical, diagonal) | Met (The display characteristics meet the pre-defined criteria) |
| Acceptable luminance non-uniformity characteristics (as specified in TG18) | Met (The display characteristics meet the pre-defined criteria) |
| Acceptable chromaticity non-uniformity characteristics (as specified in TG18) | Met (The display characteristics meet the pre-defined criteria) |
| Acceptable chromaticity at center of screen (5%, 50%, 95% max luminance) | Met (The display characteristics meet the pre-defined criteria) |
| Acceptable display reflections (specular, diffuse, haze) | Met (The display characteristics meet the pre-defined criteria) |
| Acceptable small-spot contrast ratio | Met (The display characteristics meet the pre-defined criteria) |
| Acceptable spatial resolution (MTF) | Met (The display characteristics meet the pre-defined criteria) |
| Acceptable noise (NPS) | Met (The display characteristics meet the pre-defined criteria) |
| Acceptable pixel aperture ratio | Met (The display characteristics meet the pre-defined criteria) |
| Absence of miscellaneous artifacts on the display screen (as specified in TG18) | Met (The display characteristics meet the pre-defined criteria) |
| Acceptable temporal response | Met (The display characteristics meet the pre-defined criteria) |
| Acceptable luminance stability | Met (The display characteristics meet the pre-defined criteria) |
| Maximum allowed pixel defects/faults | Met (The display characteristics meet the pre-defined criteria) |
| Equivalence to predicate device (RadiForce RX840-MG) for display characteristics functions | Demonstrated (except for 2 items determined not to affect observer performance, which are not specified in detail but imply other aspects were equivalent or superior, e.g., higher contrast, higher max luminance) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document describes bench tests performed on the RadiForce RX850 monitor. These tests assess the physical and photometric properties of the display device itself, not image data analysis from patients. Therefore, the concept of "test set" in terms of patient data or clinical cases is not applicable here. The "sample size" would refer to the number of RadiForce RX850 units tested, which is not specified but implicitly at least one unit.
The manufacturer, EIZO Corporation, is based in Japan, so the testing would have been conducted there or by their representatives. This is not a study involving patient data, so "retrospective or prospective" provenance is not relevant.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. As this is a technical bench test of a display device, clinical "ground truth" established by medical experts is not relevant. The ground truth for these tests would be the measured physical properties of the display against established technical standards (like DICOM GSDF or TG18 guidelines). There is no mention of experts establishing a ground truth for a test set of images.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. There is no test set of clinical cases requiring adjudication. The performance is measured against technical specifications and guidelines.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This submission is for a medical display monitor, not an AI-powered diagnostic algorithm. No MRMC study was performed or is relevant to this device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is a hardware monitor, not an algorithm. Standalone algorithm performance is not relevant.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the performance testing is based on established technical guidelines and standards for medical displays. Specifically, the AAPM Task Group 18 (TG18 guideline) for display performance assessment and DICOM GSDF (Grayscale Standard Display Function) are mentioned as references. These are objective, quantifiable standards for display characteristics.
8. The sample size for the training set
Not applicable. This is a hardware monitor, not a machine learning model. There is no "training set."
9. How the ground truth for the training set was established
Not applicable. As there is no training set for a machine learning model, the establishment of ground truth for such a set is not relevant.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for EIZO. The logo consists of a square shape with a checkered pattern on the left side and the word "EIZO" in bold, sans-serif font on the right side. There is a registered trademark symbol next to the "O" in "EIZO".
K140702
Page 1 of 6
APR 0 4 2014
EIZO Corporation, 153 Shimokashiwano, Hakusan, Ishikawa 924-8566 Japan
U.S. Food and Drug Administration Center for Devices and Radiological Health Document Mail Center - WO66-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002
Name ·Department Hiroaki Hashimoto Medical System Standards
Telephone Fax E-Mail
+81 (76) 274-2468 +81 (76) 274-2484 hiroaki.hashimoto@eizo.com
510(k) Summary (in accordance with 21 CFR 807.92)
1. Company
EIZO Corporation 153 Shimokashiwano, Hakusan Ishikawa 924-8566 Japan Tel: +81 (76) 274-2468 Fax: +81 (76) 274-2484
- Contact Person
Hiroaki Hashimoto
- Date of Summary March 18th, 2014
4. Device Information
- Trade Name/Model: · RadiForce RX850 �
- Common Name: 8MP Color LCD Monitor ●
- Classification Name: Display, Diagnostic Radiology .
- Product Code: PGY ●
- Regulation Number: 21 CFR 892.2050 .
5. Predicate Device
- 8MP Color LCD Monitor, RadiForce RX840-MG (K120451) .
{1}------------------------------------------------
6. Device Description
RadiForce RX850 is a color LCD monitor for viewing medical images including those of mammography. The color panel employs in-plane switching (IPS) technology allowing wide viewing angles. With the matrix size (or resolution) of 4.096 x 2.160 pixels (8MP), the RX850 is an alternate replacement for traditional dual head 5MP display installations.
Image /page/1/Picture/2 description: The image shows a computer monitor displaying two brain scans. The scans appear to be axial views, showing the brain from above. The scans are similar in appearance, with areas of varying brightness indicating different tissue densities or activity levels. The monitor is a standard desktop model with a stand.
RadiForce RX850
Image /page/1/Picture/4 description: The image shows two computer monitors displaying similar images of a brain scan. The scans appear to be cross-sectional views, with the brain's structure clearly visible. The monitors are black and have a standard desktop design with a base for support. The scans are likely being used for medical analysis or research purposes.
5MP Monochrome Monitors
Since factory calibrated display modes, each of which is characterized by a specific tone curve (including DICOM GSDF), a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor, the tone curve is e.g. DICOM compliant regardless of the display controller used.
RadiCS is application software to be installed in each workstation offering worry-free quality control of the diagnostic monitors including the RadiForce RX850 based on the OC standards and guidelines and is capable of quantitative tests and visual tests defined by them. The RadiCS and its subset, RadiCS LE, are included in this 510(k) submission as an accessory to the RadiForce RX850.
7. Intended Use
This product is intended to be used in displaying and viewing digital images including those of digital mammography for review and analysis by trained medical practitioners.
8. Comparison of Technological Characteristics
The comparison table below enumerates information derived from the product brochure of the each device and different technological characteristics are discussed in it:
{2}------------------------------------------------
| Attributes | RadiForce RX850 | RadiForce RX840-MG | Explanation ofDifferences |
|---|---|---|---|
| Display Performance/Specifications | |||
| Screentechnology | IPS TFT Color LCD | IPS TFT Color LCD | |
| Viewingangle (H, V) | H: 178°, V: 178° | H: 176°, V: 176° | Eizo uses typical data forvery low contrastprovided by the panelmanufacturers. |
| Resolution | 8MP (4,096 x 2,160) | 8MP (4,096 x 2,160) | |
| Aspect ratio | 17 : 9 | 17 : 9 | |
| Activescreen size | 697.9 mm x 368.0 mm | 817.1 mm x 430.9 mm | The smaller pixel pitch orpixel size means higher |
| Pixel pitch | 0.1704 mm x 0.1704 mm | 0.1995 x 0.1995 mm | density usually resultingin higher quality ofdisplayed images. If onecares about the smallerpixel size, the perceivedpixel size similar to thatof the predicate devicecan be realized easily byadjusting the viewingdistance. |
| Maximumluminance | 850 cd/m² | 700 cd/m² | |
| DICOMcalibratedluminance | 500 cd/m² | 500 cd/m² | |
| Contrastratio | 1450 : 1 | 1000 : 1 | Eizo uses typical contrastratio data provided bypanel manufacturers. |
| Backlighting | LED | LED | |
| DisplayColors | From a palette of 68 billioncolors:- 10-bit input (DisplayPort):1.07 billion colors(maximum)- 8-bit input: 16.77 millioncolors | From a palette of 68 billioncolors:- 10-bit input (DisplayPort):1.07 billion colors(maximum)- 8-bit input: 16.77 millioncolors | |
| Luminancenon-uniformitycompensation | Digital UniformityEqualizer | Digital UniformityEqualizer | |
| Video Signal Input | |||
| Input video signals | DVI-D (dual link) x 2, DisplayPort x 2(two inputs are required) | DVI-D (dual link) x 2, DisplayPort x 2 | |
| Scanning Frequency (H, V) | 31 - 140 kHz / 59 - 61 Hz (VGA Text: 69 - 71 Hz)Frame synchronous mode: 29.5 - 30.5 Hz, 59 - 61 Hz | 31 - 140 kHz / 29.5 - 30.5 Hz only for 2,048 x 2,160 and 1,920 x 2,160, 59 - 61 Hz for other matrix sizes(VGA Text: 69-71 Hz)Frame synchronous mode: 29.5 - 30.5 Hz, 59 - 61 Hz | |
| Power Related Specifications | |||
| Power Requirements | AC 100 - 120 V, 200 - 240 V: 50 / 60 Hz | AC 100 - 120 V, 200 - 240 V: 50 / 60 Hz | |
| Power Consumption / Save Mode | 227 W / Less than 6 W | 350 W / Less than 6 W | |
| Power Management | DVI DMPM, DisplayPort 1.1a | The proposed device consumes less power than the predicate device. | |
| Miscellaneous Features/Specifications | |||
| QC software | RadiCS | RadiCS | |
| Sensors | Backlight Sensor, Integrated Front Sensor, Presence Sensor, Ambient Light Sensor | Backlight Sensor, Integrated Front Sensor, Presence Sensor, Ambient Light Sensor | |
| USB Ports / Standard | 1 upstream, 2 downstream / Rev. 2.0 | 1 upstream, 2 downstream / Rev. 2.0 | |
| Dimensions w/o stand (W x H x D) | 747 x 430 x 130 mm | 896 x 527 x 157 mm |
·
Section 05
:
·
:
. :
:
.
·
.
.
{3}------------------------------------------------
It is clear that the technological characteristics differences discussed above do not affect the safety and the effectiveness of the RX850.
·
:
:
{4}------------------------------------------------
9. Performance Testing
The bench tests below were performed on the RadiForce RX850 following the instructions in Guidance for Industry and FDA Staff: Display Accessories for Full-Field Digital Mammography Systems-Premarket Notification (510(k)) Submissions:
- . Verification of the conformance to DICOM GSDF as specified in Assessment of Display Performance for Medical Imaging Systems by AAPM Task Group 18 (TG18 guideline)
- Measurement of the angular dependency of luminance response in horizontal, ● vertical and diagonal directions
- Measurement of the luminance non-uniformity characteristics of the display screen . as specified in TG18 guideline
- . Measurement of the chromaticity non-uniformity characteristics of the display screen as specified in TG18 guideline
- Measurement of the chromaticity at the center of the display screen at 5%. 50% and . 95% of the maximum luminance
- . Measurement of display reflections including specular, diffuse and haze components
- . Measurement of small-spot contrast ratio
- Measurement of spatial resolution expressed as modulation transfer function (MTF)
- . Measurement of noise expressed as noise power spectrum (NPS)
- Measurement of pixel aperture ratio .
- . Visual check of presence or absence of miscellaneous artifacts on the display screen as specified in TG18 guideline
- Measurement of temporal response .
- Performance data on luminance stability .
- . The maximum number allowed for each type of pixel defects/faults
The test results showed that the RadiForce RX850 has display characteristics equivalent to those of the predicate device, RadiForce RX840-MG except 2 items, each of which was determined that it would not affect observer's performance.
Besides, the display characteristics of the RadiForce RX850 meet the pre-defined criteria when criteria are set.
No animal or clinical testing was performed on the RadiForce RX850.
{5}------------------------------------------------
K140702
Page 6 of 6
10. Conclusion
The RadiForce RX850 was determined to be substantially equivalent to the predicate device due to the following reasons:
- The stated intended use is completely the same as that of the predicate device. .
- It was confirmed that the technological characteristics differences from those of the . predicate device do not affect the safety or the effectiveness.
- The bench tests demonstrated that the display characteristics are equivalent to those of . the predicate device.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three angled lines that resemble an abstract representation of a human figure or a bird in flight.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-GB09 Silver Spring, MD 20993-0002
April 4, 2014
EIZO Corporation % Mr. Hiroaki Hashimoto 153 Shimokashiwano. Hakusan ISHIKAWA 924-8566 JAPAN
Re: K140702
Trade/Device Name: 8MP Color LCD Monitor, RadilForce RX850 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications systems Regulatory Class: II Product Code: PGY Dated: March 18, 2014 Received: March 20, 2014
Dear Mr. Hashimoto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class 11 (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
Page 2-Mr. Hashimoto
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safctv/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K140702
Device Name
8MP Color LCD Monitor, RadiForce RX850
Indications for Use (Describe)
This product is intended to be used in displaying and viewing digital images including those of digital mammography for review and analysis by trained medical practitioners.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Smh. 7)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
N/A