K Number
K140583

Validate with FDA (Live)

Date Cleared
2014-06-18

(104 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Digital Automatic Blood Pressure Monitor BPM20 Scrics is for use by medical professional or home user. The BPM20 Series is intended to measure the systolic blood pressure, and pulse rate of an adult individual by using a non-invasive technique, in which an inflatable culf is wrapped around the upper arm of an individual. The inflaable cuff circumference is limited to 17cm - 44cm via 3 diffs. The 3 different cuff sizes are 17 - 22cm, 22 - 32cm and 32 - 44cm. For Bluetooth capable the measurement result can be transmitted to pre-approved mobile devices.

Device Description

Digital Automatic Blood Pressure Monitor BPM20 Series is a non-invasive blood pressure measurement system for use by medical professional or at home. It is designed to measure the systolic and diastolic blood pressure, and pulse rate of an individual in each measurement and then display the readings on a digital panel. The device utilizes the oscillometric methodology, in which an inflatable cuff is wrapped around the upper arm of an individual, for blood pressure measurement. This means the monitor detects your blood's movement through your brachial artery and converts the movement into a digital reading. Certain models of Digital Automatic Blood Pressure Monitor BPM20 Series equipment with Bluetooth transmission function, which enable user to transfer the measurement record from the device to a mobile platform through Bluetooth. User can manage the measurement record by using the mobile application.

AI/ML Overview

Here's the information about the acceptance criteria and the study that proves the device meets them, based on your provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (Predicate Device)Reported Device Performance (BPM20 Series)
Blood Pressure Measurement Accuracy± 3 mmHg± 3 mmHg or 2% of reading
Pulse Rate Measurement Accuracy± 4 % of the reading± 5 % of the reading
Resolution of Measurement1 mmHg1 mmHg
Operation Temperature+5 to +40 °C+5 to +40 °C
Operation Humidity15-93% R.H. max15-93% R.H. max
Storage Temperature-25 to 70 °C-25 to 70 °C
Storage HumidityUp to 95% R.H. maxUp to 95% R.H. max
Operation, Storage and Transportation Atmospheric Pressure700hPa - 1060hPa700hPa-1060hPa

Note: The reported device performance for "Blood Pressure Measurement Accuracy" and "Pulse Rate Measurement Accuracy" are stated as "Equivalent" and "Similar" respectively, implying they meet or are comparable to the predicate device's criteria. The other metrics are stated as "Identical."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 100 patients (52 males and 48 females).
  • Data Provenance: Not explicitly stated, but the study was conducted to comply with ISO 81060-2:2009, which is an international standard for non-invasive sphygmomanometers. This suggests the data was collected clinically, likely in a controlled environment. The country of origin is not mentioned. The study is prospective for the purpose of validating this device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document mentions that "Standard auscultation method was used as the reference blood pressure monitor," but it does not specify the number of experts who performed the auscultation or their specific qualifications (e.g., "radiologist with 10 years of experience").

4. Adjudication Method for the Test Set

The document states that "Blood pressure measurements were repeated alternatively with the device and auscultation in the same arm according to the sequence in ISO 81060-2:2009." This implies a direct comparison method against the auscultation standard. An explicit adjudication method like "2+1" or "3+1" is not mentioned.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no MRMC comparative effectiveness study mentioned. The device is a self-measuring blood pressure monitor, not an AI-assisted diagnostic tool that would involve human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, a standalone clinical test was performed. The device's performance was compared directly against the standard auscultation method, without human interpretation of the device's readings for adjustment or decision-making. The device, by its nature, is a standalone measurement system.

7. The Type of Ground Truth Used

The ground truth for the clinical test was established using the standard auscultation method.

8. The Sample Size for the Training Set

The document does not mention a training set or any machine learning algorithms that would require one. The device uses an oscillometric methodology, which is a traditional method for blood pressure measurement, not typically involving a "training set" in the context of AI.

9. How the Ground Truth for the Training Set was Established

As no training set is mentioned, this information is not applicable.

{0}------------------------------------------------

GRANDWAY TECHNOLOGY (SHENZHEN) LIMITED

Section 7 510(k) Summary

Re: K_

CHECK ONLY ONE

风 510(k) Summary. Attached is a summary of safety and effectiveness information upon which an equivalence determination could be based.

510(k) Statement. I certify that, in my capacity as (the position held in company by person required to submit the premarket notification, preferably the official correspondent) of (company name), I will make available all information included in this premarket notification on safety and effectiveness within 30 days of request by any person if the device described in the premarket notification submission is determined to be substantially equivalent. The information I agree to make available will be a duplicate of the premarket notification submission, including any adverse safety and effectiveness information, but excluding all patient identifiers, and trade secret and confidential commercial information, as defined in 21 CFR 20.61.

Ch. Fanz

Patrick Chow, General Manager (00852)-2851-6789

05th March, 2014

{1}------------------------------------------------

510(k) Summary

Submitter Identification 1.

510(k) SubmitterGRANDWAY TECHNOLOGY (SHENZHEN) LIMITED
AddressBlock 6 and 7, Zhu Keng Industrial Zone, Ping Shan, Long Gang
District, Shenzhen, Guang Dong, People's Republic of China
Phone Number(00852)-2851-6789
Fax Number(00852)-2851-6278
Contact PersonMr. Patrick Chow
Date of Submission05th March, 2014

2. Device Identification

Trade NameDigital Automatic Blood Pressure Monitor BPM20 Series[Model No.: MD20xy]
x --- The first character (0, 1, 2, 3, 4, 5, 6, 7, 8 & 9) is for theminor change revision of device. The mentioned "minorchange" refers to those device changes not to be affecting theconformity test results of EMC & safety as well as deviceperformance, i.e. IEC 60601-1, EN 60601-1-2, ETSI EN 301489-1 V1.92 and ETSI EN 300 328 V1.7.1.
y --- The second character (0 & 1) is for the identification ofpresence of DC jack. Surely, the presence of DC jack doesnot affecting the conformity test results of EMC & safety aswell as device performance, i.e. IEC 60601-1, EN 60601-1-2,ETSI EN 301 489-1 V1.92 and ETSI EN 300 328 V1.7.1.
Common NameNon-invasive Blood Pressure Measurement System
Classification NameNon-invasive Blood Pressure Measurement System(CFR 870.1130, Class II, Product Code DYN)

3. Predicate Device

Predicate DeviceAVITA Bluetooth Blood Pressure Monitor (Model: BPM656ZB)
ManufacturerAVITA Corporation
510(k) NumberK072137

{2}------------------------------------------------

Device Description 4.

Digital Automatic Blood Pressure Monitor BPM20 Series is a non-invasive blood pressure measurement system for use by medical professional or at home. It is designed to measure the systolic and diastolic blood pressure, and pulse rate of an individual in each measurement and then display the readings on a digital panel.

The device utilizes the oscillometric methodology, in which an inflatable cuff is wrapped around the upper arm of an individual, for blood pressure measurement. This means the monitor detects your blood's movement through your brachial artery and converts the movement into a digital reading.

Certain models of Digital Automatic Blood Pressure Monitor BPM20 Series equipment with Bluetooth transmission function, which enable user to transfer the measurement record from the device to a mobile platform through Bluetooth. User can manage the measurement record by using the mobile application.

ModelBlood PressureMeasurementPulse RateMeasurementWHOIHBLCD TypeBacklightDCJackBLEMemory
MD2000LEDPositive ReflectiveXX4 × 120
MD2010LEDPositive TransmissiveBlueX4 × 240
MD2020LEDPositive TransmissiveWhite4.04 × 240
MD2030Positive TransmissiveBlue4.04 × 240
MD2040Positive ReflectiveXX4 × 240
MD2050Positive TransmissiveBlueX4 × 240
MD2060Positive TransmissiveWhite4.04 × 240
MD2070LEDPositive TransmissiveBlue4.04 × 240
MD2080LEDNegative TransmissiveWhite4.04 × 240
MD2090LEDNegative TransmissiveWhiteX4 × 240
MD2001LEDPositive ReflectiveXXX4 × 120
MD2011LEDPositive TransmissiveBlueXX4 × 240
MD2021LEDPositive TransmissiveWhiteX4.04 × 240
MD2031Positive TransmissiveBlueX4.04 × 240
MD2041Positive ReflectiveXXX4 × 120
MD2051Positive TransmissiveBlueXX4 × 240
MD2061Positive TransmissiveWhiteX4.04 × 240
MD2071LEDPositive TransmissiveBlueX4.04 × 240

The table below illustrate the feature presence in Digital Automatic Blood Pressure Monitor BPM20 Series.

{3}------------------------------------------------

5. Indication for Use

Digital Automatic Blood Pressure Monitor BPM20 Series is for use by medical professional or home user. The BPM20 Series is intended to measure the systolic and diastolic blood pressure, and pulse rate of an adult individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the upper arm of an individual. The inflatable cuff circumference is limited to 17cm - 44cm via 3 different size of cuffs. The 3 different cuff sizes are 17 - 22cm, 22 - 32cm and 32 - 44cm. For Bluetooth capable models, the measurement result can be transmitted to pre-approved mobile devices.

6. Comparison of Technological Characteristics between New Device and Predicate Devices

Digital Automatic Blood Pressure Monitor BPM20 Series is compared to the predicate device, BPM656ZB (K072137) in the device comparison table below.

Comparison between Digital Automatic Blood Pressure Monitor BPM20 Series and predicate device
ItemPredicate Device (K072137)Digital Automatic Blood PressureMonitor BPM20 SeriesComment
Indication forUseThe device is arm type Blood PressureMonitor that applies oscillometric methodto measure human Systolic, Diastolicblood pressure and heart rate. Themeasurement results are displayed on theLCD and transmitted to Bluetooth enableddevices, such as a PC, a PDA or a printer.The device is designed for adult.Digital Automatic Blood Pressure MonitorBPM20 Series is for use by medicalprofessional or home user. The BPM20Series is intended to measure the systolicand diastolic blood pressure, and pulserate of an adult individual by using anon-invasive technique, in which aninflatable cuff is wrapped around theupper arm of an individual. The inflatablecuff circumference is limited to 17cm -44cm via 3 different size of cuffs. The 3different cuff sizes are 17 - 22cm,22 - 32cm and 32 - 44cm. ForBluetooth capable models, themeasurement result can be transmitted topre-approved mobile devices.Similar
MeasurementMethodNon-invasive, OscillometricNon-invasive, OscillometricIdentical
PatientPopulationAdultAge 16 or aboveEquivalent
Blood PressureMeasurementRangeCuff Pressure: 30 - 280 mmHgCuff Pressure: 0 - 300 mmHgSystolic Pressure: 50 - 250 mmHgDiastolic Pressure: 30 - 200 mmHgSimilar
Number of User2 independent users4 independent usersSimilar,improvedspecification
Memory Space2 users × 60 memory space4 users × 120 memory space; or4 users × 240 memory space(Depends on models identification)Similar,improvedspecification
Comparison between Digital Automatic Blood Pressure Monitor BPM20 Series and predicate device
ItemPredicate Device (K072137)Digital Automatic Blood PressureMonitor BPM20 SeriesComment
Resolution ofMeasurement1 mmHg1 mmHgIdentical
Blood PressureMeasurementAccuracy$\pm$ 3 mmHg$\pm$ 3 mmHg or 2% of readingEquivalent
Pulse RateMeasurementRange40 - 199 beats/min30 - 180 beats/minSimilar
Pulse RateMeasurementAccuracy$\pm$ 4 % of the reading$\pm$ 5 % of the readingSimilar
OperationTemperature+5 to +40 °C+5 to +40 °CIdentical
OperationHumidity15-93% R.H. max15-93% R.H. maxIdentical
StorageTemperature-25 to 70 °C-25 to 70 °CIdentical
StorageHumidityUp to 95% R.H. maxUp to 95% R.H. maxIdentical
Operation,Storage andTransportationAtmosphericPressure700hPa - 1060hPa700hPa-1060hPaIdentical
Display TypeLCDLCDIdentical
Power Source4 × 1.5 V AA-batteries; and/orAC adaptor (6V/600mA)4 × 1.5 V AA-batteries; and/orAC adaptor (6V/600mA)Equivalent
PressurizationModeAutomatic InflationAutomatic InflationIdentical
Deflation ModeAutomatic Exhaust/ DeflationAutomatic Exhaust/ DeflationIdentical
BluetoothfunctionBluetooth 4.0 (2402Hz-2480Hz)Bluetooth 4.0 (2402Hz - 2480Hz)Identical
Cuff Size23-33cm17-22cm, 22-32cm, 32-44cmSimilar
ApplicableStandard◇ EN 1060-1◇ EN 1060-3◇ ANSI/ AAMI SP-10◇ IEC 60601-1◇ IEC 60601-1-2◇ EN 1060-1:1995+A2:2009◇ EN 1060-3:1997+A2:2009◇ IEC 60601-1:2012◇ EN 60601-1-2:2007◇ FCC Part 15 Subpart B◇ FCC Part 15 Subpart C◇ ISO 10993-5:2009◇ ISO 10993-10:2010◇ IEC 62304:2006◇ IEC 81060-2:2009◇ ETSI EN 301 489-1 V1.9.2(2011-09) & ETSI EN 301 489-17V2.2.1 (2012-09)◇ ETSI EN 300 328 V1.7.1 (2006-10)& EN 62479:2010◇ FDA Radio-Frequency WirelessTechnology in Medical Device,Equivalent

{4}------------------------------------------------

510(k) PREMARKET NOTIFICATION

GRANDWAY TECHNOLOGY (SHENZHEN) LIMITED

Digital Automatic Blood Pressure Monitor BPM20 Series is a non-invasive measuring device and utilizes the oscillometric methodology to measure the blood pressure reading. The key

{5}------------------------------------------------

510(k) PREMARKET NOTIFICATION

GRANDWAY TECHNOLOGY (SHENZHEN) LIMITED

components of device are: a pressure sensor, a electric valve, an electronic control module, an electric pump and a Bluetooth module. The electric pump inflate (and deflate) the inflatable cuff automatically according to our designed architecture. The Bluetooth module responses to the Bluetooth Version 4.0 (BLE) connection. The predicate device adopts exactly same methodology and key components for measuring blood pressure.

Clinical and Non-clinical Tests 7.

Clinical Test Summary

Testing to insure clinical accuracy of the device in accordance with ISO 81060-222009 as documented in Clinical Test report.

One hundred patients (52 males and 48 females) were invited for the study. Standard auscultation method was used as the reference blood pressure monitor measuring in the left upper arm. Blood pressure measurements were repeated alternatively with the device and auscultation in the same arm according to the sequence in ISO 81060-2:2009.

Non-Clinical Test Summary

Digital Automatic Blood Pressure Monitor BPM20 Series has performed several non-clinical tests to show that all requirement specifications and standard requirements are met. The tests includes the follows:

  • � EN 1060-1:1995+A2:2009
  • � EN 1060-3:1997+A2:2009
  • � .IEC 60601-1:2012
  • EN 60601-1-2:2007 �
  • � FCC Part 15 Subpart B
  • � FCC Part 15 Subpart C
  • ゃ ISO 10993-5:2009
  • ISO 10993-10:2010 �
  • � IEC 62304:2006
  • ETSI EN 301 489-1 V1.9.2 (2011-09) & ETSI EN 301 489-17 V2.2.1 (2012-09) �
  • � ETSI EN 300 328 V1.7.1 (2006-10) & EN 62479:2010
  • FDA Radio-Frequency Wireless Technology in Medical Device, Wireless coexistence �

As all of the clinical and non-clinical testing performed on Digital Automatic Blood Pressure

Digital Automatic Blood Pressure Monitor BPM20 Series

{6}------------------------------------------------

510(k) PREMARKET NOTIFICATION

Monitor BPM20 Series are same as the predicate device. Therefore, no bench test is conducted to show the performance of Digital Automatic Blood Pressure Monitor BPM20 Series is equivalent to the predicate device.

Conclusion 8.

Digital Automatic Blood Pressure Monitor BPM20 Series has the similar intended use and same technological characteristics as the predicate device, BPM656ZB (K072137). Moreover clinical testing has demonstrated that no differences in the technological characteristics and questioning on safety or effectiveness to be raised. Thus, Digital Automatic Blood Pressure Monitor BPM20 Series is substantially equivalent to the predicate device.

{7}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/7/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem consisting of three curved lines that resemble a bird or a stylized human figure. The image is in black and white and appears to be a scan or photocopy due to its slightly grainy texture. The seal is likely used to represent the U.S. Department of Health and Human Services.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 18, 2014

Grandway Technology (Shenzhen) Limited Patrick Chow General Manager Zhu Keng Industrial Zone Ping Shan, Long Gang District Shenshen, Guang Dong, 518118 CH

Re: K140583

Trade/Device Name: Digital Automatic Blood Pressure Monitor BPM20 Series Regulation Number: 21 CFR 870.1130 Regulation Name: Non-Invasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: May 9, 2014 Received: May 12, 2014

Dear Patrick Chow,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{8}------------------------------------------------

Page 2 - Patrick Chow

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYowIndustry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/8/Picture/7 description: The image shows a logo with the letters FDA. The letters are stylized and overlapping. The logo appears to be a black and white image. The image is somewhat distorted.

forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

{9}------------------------------------------------

K140583 p.1 of 1

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K 140583

Device Name

Digital Automatic Blood Pressure Monitor BPM20 Scries

Indications for Use (Describe)

Digital Automatic Blood Pressure Monitor BPM20 Scrics is for use by medical professional or home user. The BPM20 Series is intended to measure the systolic blood pressure, and pulse rate of an adult individual by using a non-invasive technique, in which an inflatable culf is wrapped around the upper arm of an individual. The inflaable cuff circumference is limited to 17cm - 44cm via 3 diffs. The 3 different cuff sizes are 17 - 22cm, 22 - 32cm and 32 - 44cm. For Bluetooth capable the measurement result can be transmitted to pre-approved mobile devices.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpan D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED,

SE ST PHONE OF FOR FDAUSE ONLY MOSE ONLY MOSE ONLY MASSE ONLY MANAGE ONLY WANT Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Date: 2014.06.18
08:17:21 -04'00'

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).