K Number
K140539

Validate with FDA (Live)

Date Cleared
2014-06-09

(97 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Aero Blue* Performance Surgical Gowns are sterile, single use surgical apparel intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter. The Aero Blue* Performance Surgical Gowns meet the Level 3 requirements of the AAMI PB70: 2012 Liquid Barrier classifications. The Aero Blue* Performance Surgical Gowns are also sold as bulk non-sterile, single use items, to repackager/relabeler establishments for further packaging and Ethylene Oxide (EtO) sterilization.

Device Description

A Spunbond/Film/Spunbond/Meltblown/Spunbond design (SFSMS) that provides AAMI Level 3 liquid barrier protection in the critical zone of the gown. The back of the gown is a SMS (spunbond/meltblown/spunbond fabric that is air-breathable and provides AAMI Level 1 liquid barrier protection.

AI/ML Overview

Acceptance Criteria and Device Performance Study for Aero Blue* Performance Surgical Gowns

This document outlines the acceptance criteria and the study that demonstrates the Aero Blue* Performance Surgical Gowns meet these criteria, based on the provided 510(k) summary (K140539).

1. Acceptance Criteria and Reported Device Performance

The Aero Blue* Performance Surgical Gowns are designed to meet the Level 3 requirements of the AAMI PB70: 2012 Liquid Barrier classifications in the critical zone, and Level 1 in the non-critical back zone. The non-clinical testing performed demonstrated compliance with these and other relevant standards.

Standard or ReferenceTest MethodAcceptance Criteria (Implied by standard and intended use)Reported Device Performance
Standard for the Flammability for Clothing Textiles16 CFR 1610Meet Class 1 Flammability requirements.Pass
ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5CytotoxicityNo significant cytotoxic effects.Pass
ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10Skin Irritation StudyNo significant skin irritation.Pass
ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10Sensitization TestNo significant sensitization.Pass
ISO 10993-7:2008, Ethylene Oxide sterilization residualsEO residualsEO residuals within acceptable limits for a surface device with limited contact with breached or compromised surfaces.Pass
Laser Ignition ResistanceISO 11810-1 (2005)Demonstrate resistance to laser ignition.Pass
Spray Impact (critical zones)AATCC 42:2007Demonstrate water resistance in the critical zones consistent with AAMI Level 3.Pass
Liquid Barrier PerformanceAAMI PB70:2012 Level 3Critical Zone: Meet Level 3 liquid barrier performance (SFSMS design). (Implied: Resist liquid penetration under specific conditions, including spray impact and hydrostatic pressure.) Non-Critical Zone (back): Meet Level 1 liquid barrier performance (SMS fabric). (Implied: Provide minimum liquid barrier protection.)Pass
Hydrostatic Head (critical zones)AATCC 127:2008Demonstrate water resistance under hydrostatic pressure in the critical zones consistent with AAMI Level 3.Pass
Spray Impact (non-critical zones)AATCC 42:2007Demonstrate water resistance in the non-critical zones consistent with AAMI Level 1.Pass
Grab Tensile, Peak Stretch, and Peak Energy - NonwovensASTM D5034 (2009)Possess adequate tensile strength for intended use.Pass
Abrasion Resistance and Surface Bonding of SMS LaminatesWSP 20.5 (2008)Demonstrate adequate abrasion resistance and surface bonding.Pass
Synthetic Blood PenetrationASTM-1670-08 (2008)Resist penetration by synthetic blood in the critical zone.Pass
Mass Per Area (Basis Weight) of MaterialsD3776 (2009)Maintain consistent fabric basis weight as specified.Pass
Water Vapor Transmission Rate Through Nonwovens and Plastic FilmsWSP 70.4 (2008)Demonstrate appropriate water vapor transmission rate for breathability.Pass
Degree Peel Strength of Laminated Nonwovens - Raw MaterialsSTM-00197(2010)Achieve adequate peel strength for laminated components.Pass
Resistance to Linting Dry Particle GenerationINDA WSP 160.1 (2009)Demonstrate resistance to linting and dry particle generation.Pass

Note: The document explicitly states "All results of testing met acceptance criteria." This implies that the device performed within the established limits for each test.

2. Sample Size Used for the Test Set and Data Provenance

The 510(k) summary does not specify the exact sample sizes used for each individual non-clinical test. It lists the standards and test methods applied. For instance, for liquid barrier and blood penetration, it refers to standards like AAMI PB70:2012 and ASTM F-1670-08. These standards typically define minimum sample sizes for testing to ensure statistical validity.

The data provenance is retrospective, as the tests were conducted on the manufactured Aero Blue* Performance Surgical Gowns to demonstrate compliance with established standards. The country of origin of the data is not explicitly stated but is implicitly from the manufacturer's testing facilities or accredited third-party labs where these standards are tested, likely within the US given the submission to the FDA.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This section is not applicable for this medical device. The "ground truth" for the performance of a surgical gown is established by objective, standardized test methods and criteria defined by recognized national and international standards organizations (e.g., AAMI, ASTM, ISO, CPSC). These standards specify the methodology, equipment, and acceptable limits for performance. There is no human expert "ground truth" to be established in the context of device performance testing for surgical gowns.

4. Adjudication Method for the Test Set

Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or image-based diagnostic device evaluations where expert consensus is needed to establish ground truth. As noted above, the performance of a surgical gown is evaluated against predefined objective criteria and standardized test results. Therefore, no human adjudication method was employed or is relevant for this type of device testing.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

An MRMC study is not applicable and was not done. This type of study assesses the impact of a device (often an AI algorithm) on human reader performance, typically in diagnostic imaging. The Aero Blue* Performance Surgical Gowns are barrier garments, not diagnostic tools, and do not involve human "readers" or AI assistance in their function.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

A standalone study of an algorithm is not applicable and was not done. The Aero Blue* Performance Surgical Gown is a physical medical device, not an algorithm or software. Its performance is evaluated directly through physical and material property tests.

7. The Type of Ground Truth Used

The ground truth used for the device's performance claims is based on objective, quantitative measurements and qualitative assessments derived from recognized industry standards and test methods. Specifically:

  • Standardized Test Data: The "Pass" results in the "Summary of Non-Clinical Testing" indicate that the device met the precise, objective criteria set forth by each referenced standard (e.g., AAMI PB70:2012 for liquid barrier, ASTM-1670-08 for synthetic blood penetration, ISO 10993 for biocompatibility).
  • Performance Metrics: The "ground truth" is defined by specific physical and material properties (e.g., water resistance, tensile strength, flammability, cytotoxicity) measured against predefined thresholds in the standards.

8. Sample Size for the Training Set

A "training set" is not applicable to this device. Training sets are used in the development of machine learning algorithms. The Aero Blue* Performance Surgical Gowns are physical products whose performance is determined by their material composition, design, and manufacturing processes, and subsequently validated through non-clinical laboratory testing.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable as there is no training set for this device. The physical and chemical properties of the materials, and thus the performance characteristics of the gown, are established through material science, engineering, and validated test methods, not through an iterative process of training an algorithm on a dataset with established ground truth.

{0}------------------------------------------------

JUN 0 9 2014

EB Kimberly-Clark Corporation

510K Summary

May 8, 2014 Date Summarv was Prepared:

510(k) David M. Lee. JD Associate Director of Regulatory Affairs Submitter: Kimberly-Clark Health Care 1400 Holcomb Bridge Road Bldg. 300/1 Roswell, GA 30076 770-587-7167 David.m.lee@kcc.com

Primary Contact David M. Lee, JD for this 510(k) Associate Director of Regulatory Affairs Submission: Kimberly-Clark Health Care 1400 Holcomb Bridge Road Bldg. 300/1 Roswell, GA 30076 770-587-7167 David.m.lee@kcc.com

Device Trade Aero Blue* Performance Surgical Gowns Name:

Device Common Surgical Gown Names

  • Device Product FYA Codes and Class II Classification Surgical Gown Names:
    Predicate The Aero Blue* Performance Surgical Gowns under submission are Devices substantially equivalent to the predicate device, the Microcool Breathable High Performance Surgical Gowns (K103406).

Device A Spunbond/Film/Spunbond/Meltblown/Spunbond design (SFSMS) that Description: provides AAMI Level 3 liquid barrier protection in the critical zone of the gown. The back of the gown is a SMS (spunbond/meltblown/spunbond fabric that is air-breathable and provides AAMI Level 1 liquid barrier protection.

{1}------------------------------------------------

IntendedUse:The Aero Blue* Performance Surgical Gowns are sterile, single use surgicalapparel intended to be worn by healthcare professionals to help protect boththe patient and the healthcare worker from the transfer of microorganisms,body fluids, and particulate matter. The Aero Blue* Performance SurgicalGowns meet the Level 3 requirements of the AAMI PB70: 2012 Liquid Barrierclassifications.
The Aero Blue* Performance Surgical Gowns are also sold as bulk non-sterile,single use items, to repackager/relabeler establishments for furtherpackaging and Ethylene Oxide (EtO) sterilization.
TechnologicalCharacteristicsand SubstantiaEquivalence:The Kimberly-Clark* Aero Blue* Performance Surgical Gowns aresubstantially equivalent to the predicate device, the Microcool Breathable HighPerformance Surgical Gowns (K103406), in intended use, design, performanceand manufacturing processes. Both gowns incorporate a film laminate fabric inthe gown critical zone. The primary differences are that the Aero Blue*Performance Surgical Gowns provide AAMI PB70:2012 Level 3 liquid barrierprotection in the critical zone, while the predicate gown provides AAMI Level 4protection, and the Aero Blue* Performance Surgical Gowns incorporate inthe back of the gown a breathable SMS fabric providing AAMI Level 1protection. The Aero Blue* Performance Surgical Gowns fabric in thecritical zone also complies with ASTM F-1670-08 Standard Test Method forResistance of Materials Used in Protective Clothing to Penetration by SyntheticBlood. The above differences in design and AAMI liquid protection level raiseno new issues of safety and efficacy since the Aero Blue* PerformanceSurgical Gowns are designed, tested, and labeled in compliance with theAAMI PB70: 2012 Liquid Barrier Level 3 liquid barrier requirements.
Summary ofNon-ClinicalTesting:The Aero Blue* Performance Surgical Gowns have been tested forcompliance in the critical zone with the requirements of Level 3 liquid barrierperformance requirements of ANSI/AAMI PB70: 2012 "Liquid barrierperformance and classification of protective apparel and drapes intended foruse in health care facilities". The back of the gown, which is in the non-criticalzone, provides ANSI/AAMI PB70: 2012 Level 1 liquid barrier protection. TheAero Blue* Performance Surgical Gowns also meet ASTM F1670-08;Standard Test Method for Resistance of Materials Used in Protective Clothingto Penetration by Synthetic Blood in the critical zone fabric. The Aero Blue*Performance Surgical Gowns also meet the requirements of Flame ResistantCPSC 1610 Class 1. The Aero Blue* Performance Surgical Gown has alsobeen tested in compliance with the biocompatibility requirements of ISO 10993for surface devices with limited contact with breached or compromisedsurfaces. The Aero Blue* Performance Surgical Gowns are also tested incompliance with ISO 11810 for resistance to laser ignition.

All results of testing met acceptance criteria

:

.

{2}------------------------------------------------

Summary of Non-Clinical Testing

Standard or ReferenceTest MethodData GeneratedMeetsRequirement
Standard for the Flammability forClothing Textiles16 CFR 1610FlammabilityPass
ISO 10993-5:2009 BiologicalEvaluation of Medical Devices-Part 5:CytotoxicityCytotoxicityPass
ISO 10993-10:2010 BiologicalEvaluation of Medical Devices-Part 10Skin Irritation StudyIrritationPass
ISO 10993-10:2010 BiologicalEvaluation of Medical Devices-Part 10Sensitization TestSensitizationPass
ISO 10993-7:2008, EthyleneOxide sterilization residualsEO residualsEO residualsPass
Laser Ignition ResistanceISO 11810-1 (2005)Laser resistancePass
Spray Impact(critical zones)AATCC 42:2007Water resistancePass
Liquid Barrier PerformanceAAMI PB70:2012Level 3Water ResistancePass
Hydrostatic Head(critical zones)AATCC 127:2008Water resistancePass
Spray Impact(non- critical zones)AATCC 42:2007Water resistancePass
Grab Tensile, Peak Stretch, andPeak Energy - NonwovensASTM D5034(2009)Tensile StrengthPass
Abrasion Resistance and SurfaceBonding of SMS LaminatesWSP 20.5 (2008)Abrasion resistancePass
Synthetic Blood PenetrationASTM-1670-08(2008)Resistance topenetrationPass
Mass Per Area (Basis Weight) ofMaterialsD3776 (2009)Fabric basis weightPass
Water Vapor Transmission RateThrough Nonwovens and PlasticFilmsWSP70.4 (2008)Water VaporTransmissionPass
Degree Peel Strength ofLaminated Nonwovens - RawMaterialsSTM-00197(2010)Peel StrengthPass
Resistance to Linting Dry ParticleGenerationINDA WSP 160.1(2009)ParticulatePass

Substantial Equivalence Table

AttributePredicate Device:Kimberly-ClarkMicroCoolBreathable High PerformanceSurgical Gown K103406, (AAMILiquid Barrier Level 4)Subject of this 510(k):K140539 Aero Blue* PerformanceSurgical Gown, (AAMI LiquidBarrier Level 3)
Indicationsfor UseKimberly-Clark* MicroCool* SurgicalGowns, are sterile, single use surgicalapparel intended to be worn byhealthcare professionals to helpAero Blue* Surgical Gowns, aresterile, single use surgical apparelintended to be worn by healthcareprofessionals to help protect both the
AttributePredicate Device:Kimberly-ClarkMicroCoolBreathable High PerformanceSurgical Gown K103406, (AAMILiquid Barrier Level 4)Subject of this 510(k):K140539 Aero Blue* PerformanceSurgical Gown, (AAMI LiquidBarrier Level 3)
protect both the patient and thehealthcare worker from the transfer ofmicroorganisms, body fluids, andparticulate matter.The MicroCool* Surgical Gown meetthe Level 4 requirements of the AAMILiquid Barrier classifications.The Kimberly-Clark* MicroCool*Surgical Gowns, are also sold as bulknon-sterile, single use items, torepackager/relabeler establishmentsfor further packaging and EthyleneOxide (EtO) sterilization.patient and the healthcare workerfrom the transfer of microorganisms,body fluids, and particulate matter.The Aero Blue* PerformanceSurgical Gowns meet the Level 3requirements of the AAMI PB70:2012 Liquid Barrier classifications.The Aero Blue* Surgical Gowns, arealso sold as bulk non-sterile, singleuse items, to repackager/relabelerestablishments for further packagingand Ethylene Oxide (EtO)sterilization.
HowsuppliedSterile (10-6) or bulk non-sterileSame
SterilizationMethodEthylene OxideSame
SAL10-6Same
Gown colorBlueSame
Gown sizesSmall, Large, X-Large, XX-Large,XXX-LargeSame
ConstructionOverviewThe MicroCool* Breathable HighPerformance Surgical Gown ismanufactured from a breathable,repellent, non-woven, polypropylenefabric. The gown is constructed from aspunbond/film/spunbond/meltblown/spunbond (SFSMS) design that isadhesively laminated. The gownmeets AAMI-4 liquid barrierrequirementsThe Aero Blue* PerformanceSurgical Gown under submission ismanufactured from a moisture-vaporbreathable, repellent, non-wovenfabric using a polymer blend ofpolypropylene and polyethylene.The front body and sleeve fabric is athree layer film laminate. This fabricis an SFSMS design Spunbond/Film/Spunbond-Meltblown-Spunbond that is adhesively bondedtogether. Front of Gown meetsmeet AAMI-3 liquid barrierrequirements, while back of Gown inthe non-critical zone is composed ofa breathable SMS fabric with anAAMI level 1 liquid barrier protection
Does notcontainNaturalRubberLatexYesSame
PerformanceTestingANSI/AAMI PB70: 2003 Level 4 LiquidBarrier Requirements - PassN/A
ANSI/AAMI PB70: 2012 Level 3 LiquidBarrier Requirements - N/AANSI/AAMI PB70: 2012 Level 3Liquid Barrier Requirements - Pass
RTM - 6405/STM00204/WSP 70.42008 -Water Vapor Transmission Rateof Materials - PassSame
RTM 6176/WSP 160.1 2009 - TestMethod for Resistance to Linting -PassSame
ASTM D 5034 - 2009 Standard TestMethod For Breaking Strength andElongation of Textile Fabrics (GrabTest) - PassSame
Peel Strength - EQ-STM5671/STM00197 2010 - PassSame
Hydrohead Low Surface Tension EQ-STM 4507-2012 PassSame
STM- 00149 (ASTM-4966/WSP 20.52008, Abrasion Test - PassSame
Biocompatibility - PassSame
16 CFR, Chapter II--ConsumerProduct Safety Commission Part 1610-Standard For The Flammability ofClothing Textiles Class I - PassSame

.

{3}------------------------------------------------

:

'

:

{4}------------------------------------------------

  • Conclusion The performance testing submitted for the Aero Blue* Performance Surgical Gowns demonstrates substantial equivalence to the predicate Kimberly Clark* Microcool* Breathable High Performance Surgical Gown (K103406) in intended use, design, materials, performance and biocompatibility attributes.
    *Registered Trademark or Trademark of Kimberly-Clark Worldwide, Inc. or its affiliates.

·

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MID 20993-000

June 9, 2014

Kimberly-Clark Corporation David M. Lee. JD Associate Director of Regulatory Affairs Bldg. 300/l 1400 Holcomb Bridge Road Roswell. Georgia 30076

Re: K140539

Trade/Device Name: Acro Blue Performance Surgical Gowns Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Gown Regulatory Class: Class II Product Code: FYA Dated: May 8, 2014 Received: May 9, 2014

Dear Mr. Lee,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/5/Picture/10 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.

{6}------------------------------------------------

Page 2 - Mr. Lee

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejastri Purobit-Sheth, M.D. Toiashri Provisit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

510(k) Number (if known)

K140539

Device Name

Aero Blue* Performance Surgical Gowns

Indications for Use (Describe)

Intended Use: The Aero Blue* Performance Surgical Gowns are sterile, single use surgical apparel intended to be worn by healthcare professionals to help protect both the patient and the healthcare worker from the transfer of microorganisms, body fluids, and particulate matter.

The Aero Blue* Performance Surgical Gowns meet the Level 3 requirements of the AAMI PB70: 2012 Liquid Barrier classifications.

The Aero Blue* Performance Surgical Gowns are also sold as bulk non-sterile, single use items, to repackager/relabeler establishments for further packaging and Ethylene Oxide (EtO) sterilization.

Type of Use (Select one or both, as applicable)

□ Prescription Use (Part 21 CFR 801 Subpart D)

7% Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)Elizabeth F.Claverie -SDigitally signed by Elizabeth F. Claverie -SDN: c=US, o=U.S. Government, ou=HHS,ou=FDA, ou=People,0.9.2342.19200300.100.1.1=1300055864,cn=Elizabeth F. Claverie -SDate: 2014.06.06 12:04:26 -04'00'

FORM FDA 3881 (1/14)

{8}------------------------------------------------

The Aero Blue* Performance Surgical Gowns product codes, sizes, and a brief product description are provided below.

STERILE PRODUCT

Code

Product Description
41732Aero Blue Performance Surgical Gown, Small/Medium, Sterile
41733Aero Blue Performance Surgical Gown, Large, Sterile
41734Aero Blue Performance Surgical Gown, X-Large, Sterile
41735Aero Blue Performance Surgical Gown, XX-Large, Sterile
41736Aero Blue Performance Surgical Gown, XXX-Large, Sterile
41724Aero Blue Performance Surgical Gown, X-Long, Large, Sterile
41726Aero Blue Performance Surgical Gown, X-Long, X-Large, Sterile

NON-STERILE PRODUCT

Product Code

Product Description

41739Aero Blue Performance Surgical Gown, Large
41737Aero Blue Performance Surgical Gown, Large Handi-Bin
41738Aero Blue Performance Surgical Gown, X-Large, Handi-Bin
41740Aero Blue Performance Surgical Gown, X-Large
41741Aero Blue Performance Surgical Gown, XX-Large
41722Aero Blue Performance Surgical Gown in Overwrap, Large, Handi-Bin
41723Aero Blue Performance Surgical Gown in Overwrap, X-Large, Handi-Bin
41725Aero Blue Performance Surgical Gown, X-Long Large
41727Aero Blue Performance Surgical Gown, X-Long, X-Large

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.