K Number
K140468

Validate with FDA (Live)

Date Cleared
2014-06-02

(97 days)

Product Code
Regulation Number
866.1645
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus and Streptococcus.

This premarket notification is for the antimicrobial agent ceftaroline at concentrations of 0.0625-4 ug/ml. to Gram-positive ID/AST or AST only Phoenix panels. Ceftaroline has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package inserts for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Staphylococcus aureus (including methicillin-susceptible and -resistant isolates)

Device Description

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • BD Phoenix instrument and software. .
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
  • BD Phoenix AST Broth used for performing AST tests only. .
  • BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth determination.

The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35° ± 1°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, R or N (susceptible, intermediate, resistant or not susceptible).

AI/ML Overview

Here's an analysis of the acceptance criteria and study as described in the provided 510(k) summary (K140468):

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as numerical targets in a table format within the provided document. However, the study aims to demonstrate "substantial equivalence" based on Essential Agreement (EA) and Category Agreement (CA) values derived from comparisons to a reference method. The implicit acceptance criteria are high percentages for these agreement metrics. The document states that overall reproducibility was "greater than 95% (+/- 1 dilution) agreement."

Metric / Agreement TypeAcceptance Criteria (Implicit)Reported Device Performance (Ceftaroline, Gram-Positive Organisms)
Essential Agreement (EA)High percentage (e.g., >90-95%)94.7% (n=866)
Category Agreement (CA)High percentage (e.g., >95-98%)98.2% (n=866)
Intra- and Inter-site Reproducibility>95% (+/- 1 dilution)>95% (+/- 1 dilution)

2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size for Test Set:
    • Clinical and Challenge Isolates (Combined): 866 (This is the 'n' value for both EA and CA in the provided table). The document refers to "all isolates tested" when summarizing this performance.
    • Reproducibility Study: A "panel of Gram-positive isolates" was tested at three sites. The specific number of isolates is not explicitly stated, but each was tested in triplicate on three different days.
  • Data Provenance: The isolates were tested "across multiple geographically diverse sites across the United States." This indicates a prospective collection of data from clinical, stock, and challenge isolates.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

The ground truth was established by the CLSI (Clinical and Laboratory Standards Institute) reference broth microdilution method. This is a standardized laboratory method, not dependent on individual expert interpretation. Therefore, there were no individual human experts establishing the ground truth in the traditional sense for the MIC values.

4. Adjudication Method for the Test Set

Not applicable. The ground truth was established by a standardized laboratory method (CLSI reference broth microdilution method), which does not involve human adjudication for reconciling differing interpretations among experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This device is an automated system for determining antimicrobial susceptibility, not an imaging or diagnostic device that assists human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, a standalone performance study was done. The BD Phoenix™ Automated Microbiology System is an automated system that performs identification and susceptibility testing. The study directly compares the Phoenix System's results (algorithm's output) to the CLSI reference method, which is a standalone performance assessment of the device.

7. The Type of Ground Truth Used

The type of ground truth used was a standardized laboratory reference method: the CLSI reference broth microdilution method (specifically CLSI M7). For challenge isolates, "expected results" were used, which would typically come from well-characterized strains with known susceptibility profiles as determined by the reference method.

8. The Sample Size for the Training Set

The document does not provide information regarding the sample size used for the training set. The focus of this 510(k) summary is on the validation or performance testing of the device rather than its development or training phase.

9. How the Ground Truth for the Training Set Was Established

The document does not provide information on how the ground truth for any training set might have been established. Given that this is a 510(k) submission for a specific antimicrobial agent (Ceftaroline) on an existing platform (BD Phoenix), the "training" (if applicable) likely occurred during the earlier development and validation of the core Phoenix system and methodologies, for which details are not included in this document. Any ground truth for training would also typically rely on the CLSI reference broth microdilution method.

{0}------------------------------------------------

K140468

510(k) SUMMARY

JUN 0 2 2014

SUBMITTED BY: Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 Phone: 410-316-4287 Fax: 410-316-4188 CONTACT NAME: Monica E. Giguere, RAC Regulatory Affairs Project Manager May 22, 2014 DATE PREPARED: BD Phoenix™ Automated Microbiology System -DEVICE TRADE NAME: Ceftaroline (0.0625-4 ug/mL) Antimicrobial susceptibility test system-short incubation DEVICE COMMON NAME: DEVICE CLASSIFICATION: 21 CFR 866.1645 Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System. (Product Code LON) VITEK® System (PMA No. N50510) PREDICATE DEVICES: The BD Phoenix™ Automated Microbiology System is INTENDED USE: intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus and Streptococcus.

DEVICE DESCRIPTION:

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • BD Phoenix instrument and software. .
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
  • BD Phoenix AST Broth used for performing AST tests only. .
  • BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth � determination.

{1}------------------------------------------------

JURY TO WEIGH

። containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed into the instrument. Inoculum for use with the Phoenix system may be prepared either manually or may be automated using the BD Phoenix™ AP System.

The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35° ± 1°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, R or N (susceptible, intermediate, resistant or not susceptible).

DEVICE COMPARISON:

The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance for inoculum prepared manually and inoculum prepared with the BD Phoenix™ AP instrument when compared with the CLSI reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Gram positive ID/AST or AST only Phoenix panels with this antimicrobial agent.

SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:

The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to CLSI M7). The system has been evaluated as defined in the FDA guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems: Guidance for Industry and FDA". August 28, 2009. Shelf life (stability data) for the drug is being collected and will be maintained on file at BD as indicated in the guidance document.

Site Reproducibility

Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of Gram-positive isolates. Each site tested the isolates in triplicate on three different days using one lot of Gram Positive Phoenix panels containing this antimicrobial agent and associated reagents.

The results of the study demonstrate that for this antimicrobial agent and the Gram-positive organisms tested there was an overall reproducibility across test sites of greater than 95% (+/- 1 dilution) agreement when compared to the test mode.

{2}------------------------------------------------

Clinical Studies

Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with a Gram Positive Phoenix Panel containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.

The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within ± one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, resistant or not susceptible).

The following table summarizes the performance for Clinical and Challenge isolates tested in this study.

Performance of BD Phoenix System for Gram-Positive Organisms by Ceftaroline
AntimicrobialConcentrationEA (n)EA (%)CA (n)CA (%)
Ceftaroline0.0625-4 µg/mL86694.786698.2

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------D ~~*** く ens bi n.

Conclusions Drawn from Substantial Equivalence Studies

The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", August 28, 2009. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure, with three curved lines representing its wings or body. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002

June 2nd 2014

BECTON, DICKINSON AND COMPANY MONICA E. GIGUERE REGULATORY AFFAIRS PROJECT MANAGER 7 LOVETON CIRCLE-MC 694 SPARKS. MD 21152

Rc: K140468

Trade/Device Name: BD Phoenix™ Automated Microbiology System-Ceflaroline (0.0625-4 ug/ml)

Regulation Number: 21 CFR 866.1645

Regulation Name: Fully automated short-term incubation cycle antimicrobial susceptibility system

Regulatory Class: Class II Product Code: LON Dated: May 20, 2014 Received: May 22, 2014

Dear Ms. Giguere:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave leviewed your Section 310(t) promation is substantially equivalent (for the indications feterenced above and have acternined the actreed predicate devices marketed in interstate for use stated in the encry to tegany manologic of the Medical Device Amendments, or to commerce prior to May 26, 1976, the endomine with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). alle Cosment Act (7tel) market the device, subject to the general controls provisions of the Act. You may, therefore, market the de ried, shorted, shortements for annual registration, listing of The general controls provisions and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (Sec above) into time major regulations affecting your device can be may be subject to additions, Title 21. Parts 800 to 898. In addition, FDA may found in the Code of reading your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I DA has made a decermination an administered by other Federal agencies. You must or any Federal Statures and regulations annunbive, but not limited to: registration and listing (2 l
ess and the Act 's requirements, including, but not limited to: registrat comply with an the Act 3 requiremons, money, or ever reporting (reporting of medical

{4}------------------------------------------------

Page 2 - Ms. Giguere

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please It you active bivision of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part the regulations regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the r ou may obtain other general internation at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Sally A. Hojvat -S

Sally Hojvat, M.Sc., PhD. Director, Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

510(k) Number (if known)

K140468

Device Name

BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent Ceftaroline (0.0625-4 pg/ml.) -Gram-positive ID/AST or AST only Phoenix Panels.

Indications for Use (Describe)

The BD Phoenix™ Automated Microbiology System is intended for in viro quantitative devermination of animicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative and facultative anaerobic bacteria isslates from pure culture for Enterobacteriaceae and most Gram-positive hacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus and Streptococcus.

This premarket notification is for the antimicrobial agent ceftaroline at concentrations of 0.0625-4 ug/ml. to Gram-positive ID/AST or AST only Phoenix panels. Ceffaroline has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package inserts for this antimierobial agent.

Active In Vitro and in Clinical Infections Against:

Staphylococcus aurens (including methicillin-susceptible and -resistant isolates)

Type of Use (Select one or both, as applicable)

IXI Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Ribhi Shawar -S 2014.05.29 22:09:07 -04'00'

{6}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.

(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”