Intended Use

The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.

With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. Additionally, the CD HORIZON® Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/spondylolysis and fracture caused by tumor and/or trauma. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

The CDHORIZON® SPIRE™ Plate is a posterior, single level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (TI-SI) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined); spondylolisthesis; trauma; and/or tumor.

In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.

Device Description

The CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® Plates, staples, and connecting components, as well as implant components from other Medtronic spinal systems which can be rigidly locked into a variety of configurations with each construct being tailor-made for the individual case.

A subset of CD HORIZON® Spinal System components may be used for posterior pedicle screw fixation in pediatric cases. These constructs may be comprised of a variety of shapes and sizes of rods (ranging in diameter from 3.5mm to 6.35mm), hooks, screws, CROSSLINK® Plates and connecting components. Similarly to the CD HORIZON® implants used in adult cases, these components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.

The purpose of this submission is to modify Medtronic's CD HORIZON® Spinal System to add additional components to the system, specifically, modified Side Loading/Closed Domino Connectors.

AI/ML Overview

The provided document is a 510(k) summary for the CD Horizon® Spinal System, a medical device for spinal fixation. This type of regulatory submission primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed clinical study results with acceptance criteria.

The document explicitly states: "No clinical testing was performed" (Section XI).

Therefore, it is not possible to provide the requested information regarding acceptance criteria and a study proving the device meets those criteria, as no such study was conducted for this specific 510(k) submission.

The manufacturer is submitting for a modification to an already marketed device, adding "modified Side Loading/Closed Domino Connectors." The regulatory strategy relies on non-clinical testing and comparison to previously cleared predicate devices.

Here's a breakdown of why the requested information cannot be provided based on the given text:

  1. A table of acceptance criteria and the reported device performance: Not available. Since no clinical testing was performed, there are no reported device performance metrics against acceptance criteria from a clinical study.
  2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable. No clinical test set was used.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. No clinical test set with ground truth established by experts was used.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. No clinical test set was used.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a mechanical spinal system, not an AI-assisted diagnostic tool.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a mechanical spinal system, not an algorithm.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. No clinical study requiring ground truth was performed for this 510(k).
  8. The sample size for the training set: Not applicable. This is a medical device, not a machine learning model requiring a training set.
  9. How the ground truth for the training set was established: Not applicable.

Instead of clinical testing, the submission relies on:

  • Risk analysis: "A risk analysis of the device modifications was completed in accordance with Medtronic design control procedures. The risk analysis, which included an engineering rationale, demonstrated that the subject CD HORIZON® Spinal System does not introduce new issues of safety or effectiveness." (Section X)
  • Substantial Equivalence to predicate devices: The modification is deemed substantially equivalent to "Medtronic's CD HORIZON® Spinal System K090390 (S.E. 05/15/2009) and CD HORIZON® Spinal System K101074 (S.E. 06/22/2010)." This means the new components are considered safe and effective because similar components already on the market have demonstrated safety and effectiveness.

For devices like the CD Horizon® Spinal System undergoing minor modifications, regulatory bodies often accept non-clinical evidence (like engineering rationale and comparison to predicate devices) if the changes do not introduce new safety or effectiveness concerns.

{0}------------------------------------------------

CD Horizon® Spinal System 510(k) Summary February 2014

I. COMPANY:Medtronic Sofamor Danek USA, Inc1800 Pyramid PlaceMemphis, Tennessee 38132
CONTACT:Gregory MaschekRegulatory Affairs SpecialistTelephone: (901) 396-3133Fax: (901) 346-9738
II. PROPRIETARY TRADE NAME:CD Horizon® Spinal System
III. COMMON OR USUAL NAME:Spinal Fixation Appliance, Spinal Fixation Orthosis
IV. CLASSIFICATION NAMES:Spinal Interlaminal Fixation Orthosis(21 CFR 888.3050)Spinal Intervertebral Body Fixation Orthosis(21 CFR 888.3060)Spondylolisthesis Spinal Fixation Orthosis(21 CFR 888.3070)Spinal Pedicle Fixation Orthosis(21 CFR 888.3070)Spinal Pedicle Fixation, For DegenerativeDisc Disease Orthosis (21 CFR 888.3070)Adolescent Idiopathic Scoliosis PedicleScrew Spinal System, (21 CFR 888.3070)
CLASS:III (Pre-Amendment)
V. PRODUCT CODE:KWP, KWQ, MNH, MNI, NKB, & OSH

VI. PRODUCT DESCRIPTION:

The CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® Plates, staples, and connecting components, as well as implant components from other Medtronic spinal systems which can be rigidly locked

{1}------------------------------------------------

into a variety of configurations with each construct being tailor-made for the individual case.

A subset of CD HORIZON® Spinal System components may be used for posterior pedicle screw fixation in pediatric cases. These constructs may be comprised of a variety of shapes and sizes of rods (ranging in diameter from 3.5mm to 6.35mm), hooks, screws, CROSSLINK® Plates and connecting components. Similarly to the CD HORIZON® implants used in adult cases, these components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.

The purpose of this submission is to modify Medtronic's CD HORIZON® Spinal System to add additional components to the system, specifically, modified Side Loading/Closed Domino Connectors.

VII. INDICATIONS FOR USE:

The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.

With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. Additionally, the CD HORIZON® Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/spondylolysis and fracture caused by tumor and/or trauma. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

{2}------------------------------------------------

The CDHORIZON® SPIRE™ Plate is a posterior, single level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (TI-SI) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined); spondylolisthesis; trauma; and/or tumor.

In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.

VIII. IDENTIFICATION OF LEGALLY MARKETED DEVICES USED TO CLAIM SUBSTANTIAL EQUIVALENCE:

The subject CD HORIZON® Spinal System indications, intended use, fundamental scientific technology, design features, sterilization, and materials are similar to other legally marketed devices, specifically Medtronic's CD HORIZON® Spinal System. These predicates include CD HORIZON® Spinal System K090390 (S.E. 05/15/2009) and CD HORIZON® Spinal System K101074 (S.E. 06/22/2010)

IX. SUMMARY OF THE TECHNILOGICAL CHRACTERISTICS:

The subject CD HORIZON® Spinal System has identical indications, intended use, fundamental scientific technology, materials, and similar design features as the previously FDA cleared predicates; CD HORIZON® Spinal System K090390 (S.E. 05/15/2009) and CD HORIZON® Spinal System K101074 (S.E. 06/22/2010). Like the predicate CD HORIZON® Spinal System domino connectors, the subject domino connectors are available in titanium alloy or stainless steel. The subject domino connectors are used for an offset/angled connection of two rods to extend a construct. Additionally, the sizes of the domino connectors are within the size range of the predicate domino connectors.

X. DISCUSSION OF NON-CLINICAL TESTING:

A risk analysis of the device modifications was completed in accordance with Medtronic design control procedures. The risk analysis, which included an engineering rationale, demonstrated that the subject CD HORIZON® Spinal System does not introduce new issues of safety or effectiveness. No additional non-clinical testing was performed.

XI. DISCUSSION OF CLINICAL TESTING:

No clinical testing was performed

{3}------------------------------------------------

XII. CONCLUSION:

A risk analysis was completed for the modifications to the subject devices. Based on the results and additional supporting documentation provided in this submission, the subject devices demonstrated substantial equivalence to the previously listed predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font. The logo is in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 25, 2014

Medtronic Sofamor Danek USA. Incorporated Mr. Gregory Maschek Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132

Re: K140449

Trade/Device Name: CD HORIZON® Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MNI, MNH, KWP, KWQ Dated: February 21, 2014 Received: February 24, 2014

Dear Mr. Maschek:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2 - Mr. Gregory Maschek

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Ronald P. Jean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K140449

Device Name CD HORIZON® Spinal System

Indications for Use (Describe)

The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fosion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.

With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated componentioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.

When used for posterior non-cervical pedicle screw fixation in pediatic patients, the CD HORUZON® System implants are indicated as an adjunct to freat adolescent idiopathic scoliosis. Additionally, the CD HORIZON® Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/ spondylolysis and fracture caused by tumor and/or trauma. These devices are to be used with autograft. Pediatric pedicle screw fixation is limited to a posterior approach.

The CD HORIZON® SPIRE™ Plate is a posterior, single level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T)-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/ attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined); spondylolisthesis; trauma; and/or tumor.

In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.

REAL STORE FOR FOR FOR FOR FOR FOR FOR CHARM CHANGE ON FREE OF FREE OF FREE OF Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

James P. Bertram-S
2014.08.24 15:55:21 -04'00'

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

{7}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.