K Number
K140400

Validate with FDA (Live)

Date Cleared
2014-07-01

(133 days)

Product Code
Regulation Number
874.1820
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ES2® Awl, ES2® Taps, and ES2® Screwdriver can be used to assist in location of the spinal nerves by providing proximity information before, during or after bone preparation and placement of bone screws in open and perculaneous minimally invasive posterior surgical approaches of the non-cervical spine.

Device Description

The ES2® Neuromonitoring instruments (Awl, Taps, and Screwdriver) are accessory devices to be used with FDA cleared neuromonitoring systems to deliver electrical stimulation to assist in location of the spinal nerves during intraoperative neurological monitoring of the non-cervical spine in open and percutaneous minimally invasive posterior surgical approaches. The instruments are manufactured from surgical grade stainless steel and are provided non-sterile.

The neuromonitoring application is a surgical option that allows the surgeon to locate the spinal nerves by providing proximity information during bone preparation and placement/insertion of bone screws. The ES2® Awl and ES2® Taps facilitate bone preparation, and the ES2® Screwdriver facilitates bone screw placement/insertion. The surgical accessories are compatible with commercially available FDA cleared neuromonitoring consoles/systems and associated electrodes. The nerves are stimulated using electrodes attached to the subject accessory devices. The neuromonitoring accessory instrument can be used with or without a

AI/ML Overview

The Stryker Spine ES2® Neuromonitoring Accessory Instruments are intended to assist in locating spinal nerves by providing proximity information during bone preparation and placement of bone screws in posterior spinal surgical approaches.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicAcceptance Criteria (Predicate Device Performance)Reported Device Performance (ES2® Neuromonitoring Accessory Instruments)
Neuromonitoring AccessoriesTaps and Screwdriver (NuVasive® NVM5® System)Awl, Taps, and Screwdriver. Used the same types of accessory instruments for neuromonitoring applications during bone preparation/pilot hole starter and bone screw placement. Employed and assembled in a similar manner with comparable set-up configurations to deliver electrical signals. Testing confirmed that the powered ES2® Screwdriver configuration does not interfere with the neuromonitoring signals and is safe and effective.
Use of DilatorsDilators (NuVasive® NVM5® System)Dilators or Tap Sleeve.
Compatibility with Neuromonitoring Consoles & SoftwareCompatible with NuVasive® NVM5® SystemCompatible with FDA cleared neuromonitoring systems, including the NuVasive® NVM5® System.
Connection to Neuromonitoring UnitClip (NuVasive® NVM5® System)Clip or Probe (based on Neuromonitoring System used). The clip or probe are attached to the accessory instrument on the exposed instrument contact area above the dilator or tap sleeve.
MaterialsUnknown for NVM5® System. Assumed to be well-characterized and accepted biocompatible materials.Awl, Taps, & Screwdriver: Surgical Grade Stainless Steel. Dilators & Tap Sleeve: RADEL®. Instruments composed of well-characterized and accepted biocompatible materials. Stable chemical and mechanical properties not affected by aging or storage conditions.
Indication for UseIntraoperative neurophysiologic monitoring during spinal surgery; locating and evaluating spinal nerves (XLIF Detection, Screw Test); identifying spontaneous EMG activity (Free Run EMG); assessing neuromuscular block (Twitch Test); assessing acute dysfunction in corticospinal tract (TCMEP); assessing sensory spinal cord function (SSEP); remote access (Remote Reader); aid for pedicle cannulation (Guidance function).The ES2® Awl, ES2® Taps, and ES2® Screwdriver can be used to assist in location of the spinal nerves by providing proximity information before, during or after bone preparation and placement of bone screws in open and percutaneous posterior surgical approaches of the non-cervical spine. This is a subset of the predicate's indications, not resulting in a new or different intended use.
SterilizationValidated to assure a SAL of 10-6Instruments provided as reusable non-sterile devices with validated sterilization parameters to assure a SAL of 10-6.
Surgical ApproachOpen or Percutaneous/Minimally InvasiveOpen or Percutaneous/Minimally Invasive.
Electromagnetic Compatibility & Electrical SafetyIEC 60601-1-2 (NuVasive® NVM5® System)IEC 60601-1, IEC 60601-1-2, IEC 61000-3-2, IEC 61000-3-3.
Min. exposed surface area during tissue stimulationUnknown (NuVasive® NVM5® System)0.53mm². Test data demonstrated that the design, exposed surface area, and neuromonitoring instrument set-up/configurations do not impact or interfere with electrical signaling and are comparable to predicate devices.

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: A "porcine animal study" was conducted. The document does not specify the exact number of animals or the number of data points collected during this study.
  • Data Provenance: The study was an animal study (porcine), which is a prospective, controlled experimental setting. The country of origin is not explicitly stated, but given Stryker's location (Allendale, NJ, USA) and the US FDA submission, it's highly probable the study was conducted in the US or under US regulatory guidelines.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • The document does not explicitly state the number of experts or their qualifications used to establish ground truth in the porcine study. However, for a study assessing neuromonitoring performance, it is implied that experienced veterinary surgeons or neurophysiologists would be involved in conducting the procedures and evaluating the nerve proximity information.

4. Adjudication method for the test set:

  • The document does not specify an adjudication method like 2+1 or 3+1. The assessment of performance, functionality, and safety was likely based on the observations and data collected by the surgical team and neuromonitoring specialists during the animal study.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is a surgical instrument (neuromonitoring accessory instruments) that directly assists in nerve location during surgery, not an AI-powered diagnostic imaging tool that would involve "human readers" or AI assistance in interpreting data in the traditional sense.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable in the context of this device. The ES2® Neuromonitoring Accessory Instruments are physical surgical tools. Their function is directly integrated with a human surgeon and a neuromonitoring system. There is no "algorithm only" or "standalone" performance to measure without human-in-the-loop, as the device's purpose is to provide real-time feedback to a human surgeon.

7. The type of ground truth used:

  • The ground truth in the porcine animal study would have been established through direct observation during surgery and potentially histological examination or direct nerve stimulation protocols used to confirm nerve location and proximity. It relates to the physiological response to electrical stimulation for nerve localization.

8. The sample size for the training set:

  • The document does not describe a "training set" in the context of machine learning. This device is a physical medical instrument, not a software algorithm that requires a training set. The performance evaluation was based on bench testing and an animal study.

9. How the ground truth for the training set was established:

  • Not applicable, as there was no training set in the machine learning sense. The ground truth for the performance evaluation was established through the methodologies of the porcine animal study and bench testing, as described above.

{0}------------------------------------------------

Traditional 510(k) #K140400

Stryker Spine ES2® Spinal System Neuromonitoring Accessory Instruments

Proprietary Name:ES2® Neuromonitoring Accessory Instruments
Common Name:Surgical Nerve Stimulator/Locator
Classification Name and Reference:21 CFR §874.1820: Surgical Nerve Stimulator/Locator
Device Product Code:ETN
Proposed Regulatory Class:Class II
For Information contact:Soraya King
Regulatory Affairs Specialist
2 Pearl Court
Allendale, NJ 07401
Telephone: (201) 760-8296
Fax: (201) 962-4296
Email: Soraya.King@Stryker.com
Date Summary Prepared:June 17, 2014
Predicate DevicesNuVasive® NVM5 System - K112718 NuVasive® NVM5 System - K123307 Stryker Spine ES2® Spinal System - K122845
Device DescriptionThe ES2® Neuromonitoring instruments (Awl, Taps, and Screwdriver) are accessory devices to be used with FDA cleared neuromonitoring systems to deliver electrical stimulation to assist in location of the spinal nerves during intraoperative neurological monitoring of the non-cervical spine in open and percutaneous minimally invasive posterior surgical approaches. The instruments are manufactured from surgical grade stainless steel and are provided non-sterile.The neuromonitoring application is a surgical option that allows the surgeon to locate the spinal nerves by providing proximity information during bone preparation and placement/insertion of bone screws. The ES2® Awl and ES2® Taps facilitate bone preparation, and the ES2® Screwdriver facilitates bone screw placement/insertion. The surgical accessories are compatible with commercially available FDA cleared neuromonitoring consoles/systems and associated electrodes. The nerves are stimulated using electrodes attached to the subject accessory devices. The neuromonitoring accessory instrument can be used with or without a
Indication for UseThe ES2® Awl, ES2® Taps, and ES2® Screwdriver can beused by the surgeon to assist in location of the spinal nerves byproviding proximity information before, during or after bonepreparation and placement of bone screws.
Intended UseThe ES2® Awl, ES2® Taps, and ES2® Screwdriver areintended to be used with the ES2® Dilators & ES2® TapSleeve during neuromonitoring applications. Theneuromonitoring accessory instrument set-up is also intendedto be used for bone screw insertion under power.
Summary of the TechnologicalCharacteristicsThe Stryker Spine ES2® Neuromonitoring Accessory Instruments aresubstantially equivalent to the predicate devices in terms of design,function, principals of operation, technological characteristics, andindications and intended uses. As compared to the NuVasive®NVM5® predicates, the ES2® Awl, ES2®Taps, and ES2®Screwdriver are cannulated to allow for K-Wire placement duringpercutaneous minimally invasive surgical approaches. The predicateNuVasive® NVM5® instruments and the ES2® NeuromonitoringAccessories are used with dilators and/or a tap sleeve to provideinsulation. The electrical signal for both the NuVasive® NVM5®and ES2® instruments are supplied via electrodes (clip or probe)that are attached to the neuromonitoring contact zone of the awls,taps, or screwdrivers.In addition, the ES2® Screwdriver can be used with or withoutpower for bone screw insertion. As FDA cleared in 510(k)#K122845, the power supply can be corded or cordless. Benchtesting demonstrated that the ES2® Screwdriver can deliver safe andeffective neuromonitoring signals using the non-powered orpowered (corded and cordless) bone screw placement/insertionoptions.

Section 008: 510(k) Summary

{1}------------------------------------------------

Traditional 510(k) #K140400

,

Stryker Spine ES2® Spinal System Neuromonitoring Accessory Instruments

{2}------------------------------------------------

raditional 510(k) #K140400

Stryker Spine ES2® Spinal System Neuromonitoring Accessory Instruments

OMPARISON OF ES2® NEUROMONITORING ACCESSORY INSTRUMENT AND THE PREDICATE DEVICE FOR ISSOCIETY

CharacteristicSubject Device:ES2® NeuromonitoringAccessory InstrumentsPredicate Device:NuVasive® NVM5® SystemSubstantial Equivalence
Neuromonitoring AccessoriesInstrumentsAwl, Taps, and ScrewdriverTaps and ScrewdriverYes - The ES2® and predicatesystem use the same types ofaccessory instruments forneuromonitoring applications duringbone preparation/pilot hole starterand bone screw placement. TheES2® Neuromonitoring AccessoryInstruments are employed andassembled in a similar manner withcomparable set-up configurations todeliver electrical signals. Testingconfirmed that the powered ES2®Screwdriver configuration does notinterfere with the neuromonitoringsignals and is safe and effective.
Use of DilatorsDilators or Tap SleeveDilatorsYes
Compatible with CommonNeuromonitoring Consoles &SoftwareCompatible with FDA clearedneuromonitoring systems toinclude the NuVasive® NVM5®SystemNuVasive® NVM5® SystemYes
Connection to NeuromonitoringunitClip or Probe (based onNeuromonitoring System used)ClipYes - the use of a clip or probe isdependent on the neuromonitoringconsole and/or surgeon'spreference. The clip or probe areattached to the accessory instrumenton the exposed instrument contactarea above the dilator or tap sleevefor both the ES2® and predicatedevices

Page 3 of 9

{3}------------------------------------------------

4000P-
ਮ:47
1ਮ
ી(l
్ర
aditionalTre

Stryker Spine ES2® Spinal System Neuromonitoring Accessory Instruments

.

MaterialsIndication for UseUnknown
Awl, Taps, & Screwdriver:Surgical Grade Stainless SteelDilators & Tap Sleeve: RADEL®The ES2® Awls, ES2® Taps, and ES2® Screwdriver can be used to assist in location of the spinal nerves by providing proximity information before, during or after bone preparation and placement of bone screws in open and percutaneous posterior surgical approaches of the non-cervical spine.The NVM5® System is a medical device that is intended for intraoperative neurophysiologic monitoring during spinal surgery. The device provides information directly to the surgeon, to help assess a patient's neurophysiologic status. NVM5® provides this information by electrically stimulating nerves via electrodes located on surgical accessories and monitoring electromyography (EMG), transcranial motor evoked potential (TceMEP) or somatosensory evoked potential (SSEP) responses of nerves. XLIF (Detection) - The XLIF (Detection) function allows the surgeon to locate and evaluate spinal nerves, and is used as a nerve avoidance tool. Basic & Dynamic Screw Test – The Screw Test functions allow the surgeon to locate and evaluate the spinal nerves by providing proximity information before, during or after bone preparation and placement of bone screws. Free Run EMG – The Free Run EMG function identifies spontaneous EMG activity of spinal nerves by continually displaying a live stream waveform of any mechanically induced myotome contractions. Twitch Test (Train of Four) – The Twitch Test Function allows theYes – Instruments composed of well-characterized materials and accepted biocompatible materials.Yes – the indications of the ES2® Neuromonitoring Accessory Instruments are a subset of the indications of the predicate device. This does not result in a new or different intended use as compared to the predicate. The subject and predicate instruments are employed in the same manner when used as a tool to assist the surgeon in locating spinal nerves before, during, or after bone preparation and bone screw placement for open and percutaneous surgical approaches.

Page 4 of 9

.

:

·

.

{4}------------------------------------------------

raditional 510(k) #K140400

Stryker Spinc ES2® Spinal System Neuromonitoring Accessory Instruments

.

surgeon to assess moderatedegrees of neuromuscular blockin effect by evaluating musclecontraction following a train offour stimulation pulses.● TCMEP – Transcranialstimulation techniques for motorevoked potentials are used toassess for acute dysfunction inaxonal conduction of thecorticospinal tract. The TCMEPfunction provides an adjunctivemethod to allow the surgeon tomonitor spinal cord motorpathway integrity duringprocedures with a risk ofsurgically induced motor injury.● SSEP – The SSEP functionallows the surgeon to assesssensory spinal cord function insurgical procedures during whichthe spinal cord is at risk.● Remote Reader – The RemoteReader function provides realtime remote access to theNVM5® System for monitoringphysician outside of the operatingroom.● The Guidance function isintended as an aid for use in eitheropen or percutaneous pediclecannulation producedure in thelumbar spine of adult patients,and when used in conjunctionwith radiographic imaginganatomy for the creation of acannulation trajectory for bonescrew placement.
SterilizationInstruments provided as reusableYesAs selected for individual accessories.

Page 5 of 9

{5}------------------------------------------------

Fraditional 510(k) #K140400

Stryker Spine ES2® Spinal System Neuromonitoring Accessory Instruments

non-sterile devices with validatedsterilization parameters to assure aSAL of 10-6and validated to assure a SAL of 10-6
Surgical ApproachOpen or Percutaneous/MinimallyInvasiveOpen or Percutaneous/MinimallyInvasiveYes
Electromagnetic Compatibility &Electrical SafetyIEC 60601-1IEC 60601-1-2IEC 61000-3-2IEC 61000-3-3IEC 60601-1-2Yes
Min. exposed surface area duringtissue stimulation0.53mm²unknownYes - The test data demonstratedthat the design of the subject device,the exposed surface area duringtissue stimulation, andneuromonitoring instrument set-up/configurations do not impact orinterfere with the electricalsignaling and are comparable to thepredicate devices.

Page 6 of 9

,

.

.

)

:

{6}------------------------------------------------

raditional 510(k) #K140400

Stryker Spine ES2® Spinal System Neuromonitoring Accessory Instruments

:

100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000

:

·

:

.

{7}------------------------------------------------

Summary of the Performance DataThis 510(k) pre-market notification seeks expanded indications forthe ES2® Awl, ES2® Taps, and ES2® Screwdriver to be used asaccessory instruments during neuromonitoring applications. Nodesign modifications, or changes in the materials of constructionwere made to the ES2® Awl, ES2® Taps, or ES2® Screwdriver, aspreviously presented in 510(k) #K122845, to facilitateneuromonitoring. All instruments are manufactured from surgicalgrade stainless steel. The materials of construction have beenwell characterized and shown to have stable chemical andmechanical properties that are not affected by aging or storageconditions.Performance testing was performed to demonstrate that the subjectdevices are substantially equivalent to the identified predicateinstruments in terms of design, performance and intended use. TheES2® Neuromonitoring Accessory Instruments were tested forelectrical safety in accordance with IEC 60601-1-2 for EMC andSafety. The instruments were also evaluated as per IEC 60601-1. Aporcine animal study was conducted to assess the performance,functionality, and safety of the ES2® Neuromonitoring AccessoryInstruments and implants utilizing the powered screw insertionoption, corded and cordless. Bench test data and assessmentconfirmed substantial equivalence to the NuVasive® NVM5®predicates, and the safety and efficacy of the devices.Laboratory tests were conducted in compliance with applicableGood Laboratory Practices (GLP) requirements stipulated in 21 CFRPart 58.
ConclusionBased on the information and comparison tables included in this510(k), the ES2® Neuromonitoring Accessory Instruments aresubstantially equivalent to the NuVasive® NVM5® predicates. Theinstruments are employed in the same manner, have similar intended
and indications for use, principles of operation, and technologicalcharacteristics. The ES2® subject devices met all requiredacceptance criteria and did not create new safety or efficacyconcerns.

{8}------------------------------------------------

Traditional 510(k) #K140400

:

·

:

1

Stryker Spine ES2® Spinal System Neuromonitoring Accessory Instruments

ن

r

.

.

:

!

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains a logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol. The logo is in black and white.

Public Health Service

Food and Drue Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002

July 1, 2014

Stryker Spine Sorava King Regulatory Affairs Specialist 2 Pearl Court Allendale. NJ 07401

Re: K140400

Trade/Device Name: ES2® Neuromonitoring Accessory Instruments Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical nerve stimulator/locator Regulatory Class: Class II Product Code: ETN Dated: April 1. 2014 Received: April 2. 2014

Dear Ms. King:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that I'DA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{10}------------------------------------------------

Page 2 - Ms. Soraya King

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Felipe Aguel -S

Carlos L. Peña, Ph.D., M.S. for

Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) KI40400

Device Name

ES2® Neuromonitoring Accessory Instruments: ES20 Awl, ES20 Taps, and ES2® Screwdriver

Indications for Use (Describe)

The ES2® Awl, ES2® Taps, and ES2® Screwdriver can be used to assist in location of the spinal nerves by providing proximity information before, during or after bone preparation and placement of bone screws in open and perculaneous minimally invasive posterior surgical approaches of the non-cervical spine,

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

  • Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Felipe Aquel -S

Date: 2014.07.01 14:20:03 -04'00'

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 874.1820 Surgical nerve stimulator/locator.

(a)
Identification. A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.(b)
Classification. Class II.