K Number
K140326

Validate with FDA (Live)

Device Name
XGEO GF50
Date Cleared
2014-05-29

(108 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The XGEO GF50 Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

Device Description

This system is used to capture images by transmitting X-ray to a patient's body.

The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process device being sent to the XGEO Station (Operation Software) and saved in DICOM file, a standard for medical imaging. The captured images are sent to the Picture Archiving & Communication System (PACS) server, and can be used for reading images.

The XGEO GF50 is intended to be used for capturing digital X-ray images of various parts of the body including the head, chest, spine, abdomen, joints, hands, feet and other organs.

AI/ML Overview

The XGEO GF50 is a Digital X-ray Imaging System.

Here's an analysis of its acceptance criteria and the study that proves its performance:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document describes the XGEO GF50 as being "Substantially Equivalent" to a predicate device (XGEO GC80) rather than setting specific numeric acceptance criteria for diagnostic performance. Therefore, the "acceptance criteria" here refer to the demonstration of substantial equivalence.

Acceptance Criteria (Implied)Reported Device Performance
Imaging Performance EquivalenceImages provide equivalent diagnostic capability to predicate device.
Non-clinical: Resolution (MTF)Proposed detector performs better than predicate.
Non-clinical: Detective Quantum Efficiency (DQE)Proposed device has better DQE performance.
Safety and EffectivenessAll test results for electrical, mechanical, environmental, EMC, and wireless function were satisfied; device is safe and effective.
Intended Use EquivalenceSame as predicate device (generating radiographic images, not mammographic).
Technological Characteristics EquivalenceNo significant change in materials, energy source, or technological characteristics compared to predicate device. Differences explained and deemed equivalent.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated. The clinical study is referred to as a "concurrence study" but does not detail the number of images or patients used.
  • Data Provenance: Not explicitly stated regarding country of origin or specific institutions. It is a "clinical data" study. The document refers to "radiologists in the study" evaluating images of the XGEO GF50 against the predicate device. It is a prospective evaluation comparing the new device's images to a known, established device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: Not explicitly stated. The document mentions "qualified radiologists" without specifying their number.
  • Qualifications of Experts: "Qualified radiologists." Specific experience (e.g., "10 years of experience") is not provided.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not explicitly stated. The term "concurrence study" suggests that the radiologists evaluated the images to confirm they were diagnostically equivalent, likely reaching a consensus or agreement among themselves. However, the specific method (e.g., 2+1, 3+1, none) is not detailed.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

  • MRMC Study: The document describes a "concurrence study" where "qualified radiologists" evaluated images for "equivalent diagnostic capability." This implies a comparison against the predicate device, which is a form of comparative effectiveness. However, it's not explicitly labeled as a formal MRMC study, nor does it provide details on the statistical methodology (e.g., comparing reader performance with/without AI assistance, as the device itself is an X-ray system, not primarily an AI diagnostic tool in this context).
  • Effect Size: Not applicable/not provided. This device is a digital X-ray system, not an AI-powered diagnostic tool intended to assist or replace human readers. The study aims to show equivalence of the image quality and diagnostic capability produced by the new X-ray system compared to an existing one. Therefore, "how much human readers improve with AI vs. without AI assistance" is not a relevant metric for this type of device.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Standalone Performance: Not applicable. The XGEO GF50 is a digital X-ray system that produces images for human interpretation, not an algorithm that provides a standalone diagnostic output. The "performance" being evaluated is the diagnostic quality of the images produced by the system. Non-clinical data (MTF, DQE) does provide a form of "standalone" (system-only) technical performance metrics.

7. The Type of Ground Truth Used

  • Type of Ground Truth: The clinical study relies on expert consensus/radiologist interpretation for the diagnostic capability of the images. The radiologists "evaluate the image of XGEO GF50 is substantially equivalent to predicate device." This means their clinical assessment of the images from the new device, likely compared to images from the predicate device (or possibly compared to a reference standard diagnosis if provided, though not explicitly stated), formed the "clinical ground truth" for equivalence.

8. The Sample Size for the Training Set

  • Sample Size: Not applicable. As an X-ray imaging system, the device does not employ a "training set" in the context of machine learning algorithms for diagnostic output. The system is designed to acquire images, and its performance is evaluated based on the quality of those images as assessed by human experts, and technical specifications.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth Establishment: Not applicable, as there is no "training set" for a diagnostic algorithm in this context. The "ground truth" for the device's technical specifications and image quality is established through engineering standards, non-clinical tests (MTF, DQE), and clinical evaluation by radiologists on test images.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the Samsung logo, which is a black oval with the word "SAMSUNG" in white letters inside. There is also some handwritten text above the logo, which appears to say "Page 1". The handwriting is cursive and slightly slanted.

510(k) Premarket Notification - Traditional

SAMSUNG ELECTRONICS Co., Ltd.

MAY 2 9 2014

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted accordance with requirements of 21 CFR 807.92

    1. Date: 2014. Feb. 06
    1. Submitter
    • Company Name: SAMSUNG ELECTRONICS Co., Ltd. A.
    • B. Address: 129, Samsung-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 443-742, Republic of Korea

Primary Contact Person 3.

  • Name: Donald D. Fickett A.
  • Title: Vice President Regulatory & Quality Assurance B.
  • Phone Number: 978-564-8523 ﻥ
  • FAX Number: 978-750-6677 D.
  • E-Mail: dfickett@samsungneurologica.com E.

Secondary Contact Person 4.

  • Name: Ha, Aeyoung A.
  • B. Title: Regulatory Affairs Manager
  • C. Phone Number: +82-31-200-1537
  • FAX Number: +82-31-200-1199 D.
  • E-Mail: ay.ha@samsung.com E.

Device 5.

  • Trade Name: XGEO GF50 A.
  • B. Common Name: Digital Diagnostic X-ray System
  • Classification Name: System, X-Ray, Stationary ்.
  • D. Product Code: KPR

હ. Predicate Device

  • Manufacturer: SAMSUNG ELECTRONICS Co., Ltd. A.
  • B. Trade Name: XGEO GC80
  • C. 510(k) Number: K123098

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the Samsung logo. The logo is in black and white. The word "SAMSUNG" is written in bold, sans-serif letters.

510(k) Premarket Notification - Traditional

D. 510(k) Decision Date: 01/08/2013

7. Device Description

This system is used to capture images by transmitting X-ray to a patient's body.

The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process device being sent to the XGEO Station (Operation Software) and saved in DICOM file, a standard for medical imaging. The captured images are sent to the Picture Archiving & Communication System (PACS) server, and can be used for reading images.

The XGEO GF50 is intended to be used for capturing digital X-ray images of various parts of the body including the head, chest, spine, abdomen, joints, hands, feet and other organs.

Intended Use 8.

The XGEO GF50 Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

9. Summary of Technological characteristic of the proposed device compared with the predicate device

The XGEO GF50 does not have significant change in materials, energy source or technological characteristics compared to the predicate device. Comparisons of the following technological characteristics were assessed and the results demonstrate the substantial equivalence to the predicates.

SpecificationPredicate DeviceProposed DeviceDiscussion
Device NameXGEO GC80XGEO GF50
ManufacturerSAMSUNG ELECTRONICSSAMSUNG ELECTRONICS
510(k) NumberK123098N/A
AppearancesImage: Predicate Device AppearanceImage: Proposed Device AppearanceEquivalent(1)

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the word "SAMSUNG" in white letters against a black background. The word is in all capital letters and is in a bold, sans-serif font. The background is an oval shape, with the word "SAMSUNG" centered in the middle. The logo is simple and modern.

510(k) Premarket Notification - Traditional

Intended UseThe XGEO GC80 digital X-ray imaging system is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.The XGEO GF50 digital X-ray imaging system is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.Same
ConfigurationCeiling typeFloor-mounted or Floor-ceiling typeEquivalent (1)
Manufacturer ContentsXGEO GC80XGEO GF50Discussion
(1)High Voltage Generator
TypeHigh FrequencyHigh FrequencySame
Max. Power80kW52kW
Tube Voltage40-150kV40-150kVEquivalent
OutputRANGETube Current10-1000mA10-640mA(2)
Exposure Time1msec-2sec1msec-10sec
AEC(Automatic Exposure Control)YesYesSame
APR (AnatomicallyProgrammed Radiography)YesYesSame
(2)Ceiling Suspension
MovingRange(mm)Longitudinal1350 ~ 3350(Varies with room size)2100(Floor-mounted)2500(Floor ceiling)Equivalent(3)
Lateral770 - 2770(Varies with room size)220
Vertical18001330
Vertical TubeMoving methodMotorized/ManualManualEquivalent
Tube AssemblyRotation-157 ~ +183$\pm$ 135Equivalent(4)
Brake lockingMethodElectromagneticElectromagneticSame
Moving Rail TypeAl ExtrusionSteel plateEquivalent(5)
Image PreviewOO
Display TypeColor LCDColor LCDSame
Control Switch TypeButton+ Touch ScreenButton+ Touch ScreenSame

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the SAMSUNG logo. The logo is in white text on a black oval background. The text is in a bold, sans-serif font.

510(k) Premarket Notification - Traditional

Manufacturer ContentsXGEO GC80XGEO GF50Discussion
(3) Wall Stand
VerticalMovementMechanismMotorized/ManualManualEquivalent
Range(mm)400~1,800420~1770Equivalent
DetectorMechanismMotorizedNIAEquivalent(6)
TiltingRange-20-+90
AECConventionalConventionalSame
Lines/cm84.684.6Same
GridGrid mechanismStationaryStationarySame
RemovabilityRemovableRemovableSame
Detector Support MountingFloorFloorSame
Patient Support DevicePatient handgrips,lateral support barPatient handgrips,lateral support barSame
(4)Patient Table
Size(mm)2433 X 8062200 X 750(4way)2200 X 810(6way)Equivalent
Table TopRangeLateral±140+125(7)
(mm)Longitudinal±480+500
Table heightMechanismDC Motor,Ball screwDC Motor (6way)Equivalent
Range(mm)545 - 900565 ~ 850Equivalent
Horizontal range of detector(mm)590590Same
AECConventionalConventionalSame
Lines/cm84.684.6Same
GridGrid mechanismStationaryStationarySame
RemovabilityRemovableRemovableSame
Vertical Sync.00Same
Control Switch TypeFoot switchFoot switchSame
Maximum Patient Weight(kg)350(Static, Center load)250(Static, Center load)Equivalent
(5)Collimator
Overall Size(mm)H212 X W300X D179H185 X W213X D180Equivalent
Beam Limiting BladeMoving MethodMotorized/ManualManualEquivalent
Manual Operation MethodVolumeVolumeSame
Collimator Rotation+45400Equivalent

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the SAMSUNG logo. The logo is in black and white. The word "SAMSUNG" is written in white letters on a black oval background.

510(k) Premarket Notification - Traditional

Manufacturer ContentsXGEO GC80XGEO GF50Discussion
(8)
Beam Light SourceLEDHalogen LampEquivalent(9)
Light Field Indicator TimerOOSame
Side LampOOSame
Laser ModuleLaser ModuleSame
Field Size / SID DisplayColor LCDColor LCDSame
Manufacturer ContentsXGEO GC80XGEO GF50Discussion
(6) Detector
NameSDX-4336CPSDX-4343CSS4335-W
CsICsISame
Detector TypeIndirectIndirectIndirectSame
Detector Area14"X17"(351mmX423mm)17"X17"(429mmX429mm)14"X17"(345mmX425mm)Equivalent
Number of pixels2340X28203000X30002466X3040
Pixel Pitch(um)150143140Equivalent
High Contrast LimitingResolution (LP/mm)3.33.53.5(10)
CommunicationWiredWired / WirelessEquivalent(11)
No.DescriptionExplanation
(1)ConfigurationPredicate device is a ceiling type digital x-ray imaging system andthe proposed device is a Floor-mounted or Floor ceiling typedigital x-ray imaging system. The difference of THU locationdoesn't affect the fundamental scientific characteristic of x-rayoperation.
(2)High Voltage GeneratorThe proposed device has the lower output power and tube currentspecification than the predicate device. The wider time range ofproposed device makes the equivalent mAs range to predicatedevice.
(3)Moving rangeDue to the nature of Floor-mounted device, it has less flexibility ofmoving range than ceiling type device. This is a matter ofconvenience to use. Therefore proposed device is equivalent topredicate device.
(4)Tube Assembly RotationPredicate device has wider range of the tube assembly rotationthan proposed device. The rotation range of proposed device, ±135, can operate the most of procedure intended for use ingeneral radiography which makes equivalence of the proposeddevice to predicate device.
(5)Moving Rail TypeThe moving rail type of predicate and proposed device is designed suitable for each type of device, ceiling and floor mounted type. As the XGEO GF50 is complied with IEC60601-1, the fundamental mechanical safety is equivalent to the predicate device.
(6)Detector TiltProposed device doesn't provide detector tilt function. As detector tilt function is an optional function and not an essential requirement for safety and effectiveness of the system, it doesn't affect to the safety and effectiveness.
(7)Table size and moving rangeProposed device has 2 types of the patient tables provides option to physicians for their convenience. The patient tables are complied with IEC60601-1 for basic electrical and mechanical safety. Therefore, it is equivalent to the predicate device.
(8)Collimator RotationProposed device has more collimator rotation angle than predicate device which gives more flexibility to the collimator positioning for patient.
(9)Beam light sourcePredicate device used LED lamp and proposed device uses Halogen lamp for beam light source. Halogen lamp is the one generally used light source before LED lamp was used. It is equivalent to the predicate device.
(10)Resolution of detectorProposed detector has better resolution than predicated. In the specification sheet, to compare with the same size detector (SDX-4336CP), proposed detector shows more number of pixels, short pixel pitch and higher contrast limiting resolution.In non-clinical data, the propose detector shows better curves and measurements of MTF and DQE than predicate device.In clinical data, the radiologists evaluate the image of XGEO GF50 is substantially equivalent to predicate device.
(11)CommunicationProposed device supports wireless connection and wired connection by tether, while PD only supports wired connection.To ensure the safety and effectiveness of wireless communication between detector and workstation, the wireless function is tested by EN 301 489-1, EN 301 489-17, 47 CFR Part 15C, 47 CFR Part 15E, wireless specification and performance in testing lab and actual environment and evaluated under the risk management followed by a guidance, Radio frequency Wireless Technology in Medical Devices

:

{5}------------------------------------------------

510(k) Premarket Notification - Traditional

10. Safety, EMC and Performance Data:

Electrical, mechanical, environmental safety and performance testing according to standard ES 60601-1(2012), 21CFR1020.30 and 21CFR1020.31 were performed, and EMC testing was conducted in accordance with standard IEC 60601-1-2(2007). Wireless function was tested and verified followed by guidance, Radio frequency Wireless Technology in Medical Devices. All test results were satisfied.

{6}------------------------------------------------

K140322
Page 7 of

510(k) Premarket Notification - Traditional

11. Non-clinical data

Non clinical test was conducted for imaging performance of the proposed detector. Both of the predicate device and proposed device are based on a-Si TFT photodiode arrays with a scintillator designed for general radiography. MTF and DQE were tested and measured by IEC 62220-1. MTF curves and measurements from the two detectors were compared and it shows that the propose detector perform better than the predicate in terms of the resolution at all spatial frequencies tested. In DQE, the propose device also has better DQE performance at all spatial frequencies tested because scintillator deposition method is different. Therefore, in non-clinical data review, the XGEO GF50 is substantially equivalent with the predicate device.

12. Clinical data

According to the CDRH guidance for the submission of 510(k)'s for Solid State X-ray Imaging Devices, concurrence study was conducted. The evaluation of the qualified radiologists in the study shows that XGEO GF50 provides images of equivalent diagnostic capability to the predicate device. It demonstrates that XGEO GF50 is substantially equivalent with the predicate device.

13. Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Samsung Electronics Co., Ltd. concludes that The XGEO GF50 is safe and effective and substantially equivalent to predicate devices as described herein.

    1. Samsung Electronics Co., Ltd. will update and include in this summary any other information deemed seasonably necessary by the FDA.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three legs and a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the circular border of the logo.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 29, 2014

SAMSUNG ELECTRONICS Co., Ltd. % Jeongpyo Hong Assistant Manager, Regulatory Affairs 129, Samsung-ro, Yeongtong-gu Suwon-si. Gyeonggi-do. 443-742 REPUBLIC OF KOREA

Re: K140326

Trade/Device Name: XGEO GF50 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: April 28, 2014 Received: April 29, 2014

Dear Jeongpyo Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{8}------------------------------------------------

Page 2-Jeongpyo Hong

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Sm.7)

for

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the Samsung logo. The logo is a black oval with the word "SAMSUNG" written in white, sans-serif letters. The letters are bold and slightly slanted. The logo is simple and modern.

510(k) Premarket Notification - Traditional

510(k) Number (if known):

Device Name: XGEO GF50

Indications for Use:

The XGEO GF50 Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

× Prescription Use (Part 21 CFR 801 Subpart D)

ــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ

AND/OR

Over-The-Counter Use_ (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Imgur

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.