K Number
K140266

Validate with FDA (Live)

Device Name
NMI PICC IV
Date Cleared
2014-03-05

(30 days)

Product Code
Regulation Number
880.5970
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NMI PICC IV is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, for central venous pressure monitoring and for power injection of contrast media.

Device Description

The NMI PICC IV is a Peripherally Inserted Central Catheter (PICC). It is a percutaneous, implanted, long-term intravascular catheter. It is available in single and multi-lumen configurations in a wide range of sizes from 3F to 6 F outside catheter diameter. It is rated for maximum power injector settings up to 325 psi and rated for maximum power injection flow rate up to 6 ml/second based on model. It is available kitted with a range of procedural accessories for user convenience and demonstrates resistance to blood components (platelet and thrombus) accumulation.

AI/ML Overview

The NMI PICC IV device's acceptance criteria and studies are described below:

1. A table of acceptance criteria and the reported device performance:

The provided document describes the NMI PICC IV device as substantially equivalent to predicate devices based on technological characteristics and non-clinical testing. Instead of specific acceptance criteria as thresholds for performance, the document refers to compliance with established standards and guidance documents. The "reported device performance" is essentially that it met these standards and performed similarly to its predicate.

Performance CharacteristicAcceptance Criteria (Compliance with)Reported Device Performance
General Requirements for IV CathetersEN ISO 10555-1:2009 (Sterile, Single use intravascular catheters Part 1: General Requirements)Compliant
Central Venous Catheter Specific RequirementsEN ISO 10555-3:1997 Corrigendum 1:2002 (Sterile, Single-Use Intravascular Catheters - Part 3: Central Venous Catheters)Compliant
FDA Guidance for IV Catheter SubmissionsFDA's "Guidance on Premarket Notification [510(K)] Submissions for Short-Term and Long-Term Intravascular Catheters dated March 16, 1995"Compliant
Power Injection Flow Rate (Maximum)Established by predicate device and based on modelRated for maximum flow rates up to 6 ml/second (specific values for different configurations listed in the "Intended Use" section)
Power Injection Pressure (Maximum)Established by predicate deviceRated for maximum pressure up to 325 psi
Resistance to Blood Components AccumulationNot explicitly stated as a numerical criterion, but implied by substantial equivalence to predicateDemonstrated resistance to platelet and thrombus accumulation
Gravity Flow RateNot explicitly stated, implied to meet functional requirementsTested and found acceptable
Priming VolumeNot explicitly stated, implied to meet functional requirementsTested and found acceptable
Design and MaterialsIdentical to predicate devicesIdentical materials, design, and components as predicate intravascular catheters (NMI PICC III - K133264, K131942)

2. Sample size used for the test set and the data provenance:

The document does not specify sample sizes for a "test set" in the context of clinical trials. The evaluation was based on non-clinical tests (bench testing) and comparison to a predicate device. There is no mention of human subject data, retrospective or prospective studies, or country of origin for such data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable as the study described is non-clinical (bench testing) and relies on compliance with engineering standards and comparison to a predicate device, rather than expert-established ground truth from clinical data.

4. Adjudication method for the test set:

This information is not applicable for the same reasons as point 3. There was no clinical test set requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

An MRMC comparative effectiveness study was not conducted. This device is a physical medical device (PICC line) and does not involve AI or human image interpretation.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

A standalone algorithm performance study was not conducted. This device is a physical medical device and does not involve an algorithm.

7. The type of ground truth used:

The "ground truth" for this device's performance is established by compliance with recognized industry standards (EN ISO 10555-1:2009, EN ISO 10555-3:1997), FDA guidance documents, and demonstrated equivalence to previously cleared predicate devices through non-clinical testing. No clinical "ground truth" (e.g., pathology, outcomes data) was used in the context of this 510(k) submission.

8. The sample size for the training set:

There is no mention of a training set as this is a non-AI, physical medical device. The term "training set" is not relevant to this submission.

9. How the ground truth for the training set was established:

This information is not applicable as there was no training set used for this device submission.

{0}------------------------------------------------

1/140266

MAR - 5 2014

510(K) SUMMARY FOR THE NMI PICC IV

Date prepared: 04-March-2014
A.SponsorNavilyst Medical, Inc26 Forest StreetMarlborough, MA 01752
B. Contact
Brandon BrackettSpecialist, Global Regulatory Affairs508-658-7984brandon.brackett@navilyst.comORWanda CarpinellaDirector, Global Regulatory Affairs508-658-7929wanda.carpinella@navilyst.com
C.Device NameTrade Name:Common/Usual name:Classification Name:Classification Panel:NMI PICC IVPeripherally Inserted Central Catheter (PICC)Percutaneous, implanted, long-term intravascularcatheter21CFR§880.5970, Class IIGeneral Hospital
D.Predicate DeviceCommon/Usual name:Classification Name:Classification Panel:Premarket NotificationPeripherally Inserted Central Catheter (PICC)Percutaneous, implanted, long-term intravascularcatheter21CFR&880.5970, Class IIGeneral HospitalK 133264, K 131942 (NMI PICC III)

E. Device Description

Intended Use

The NMI PICC IV is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, for central venous pressure monitoring and for power injection of contrast media.

Maximum Power Injection Flow Rate
Description .Flow Rate
3F Single Lumen - 55cm length1 mL/sec
4F Single Lumen - 55cm length3.5 mL/sec
5F Single Lumen - 55cm length5 mL/sec
5F Dual Lumen - 55 cm length4 mL/sec
6F Dual Lumen - 55cm length5 mL/sec
6F Triple Lumen - 55cm length6 mL/sec

16

{1}------------------------------------------------

F. Summary of Similarities and Differences in Technological Characteristics and Performance

The proposed device has identical materials, design, components and technological characteristics as the predicate intravascular catheters.

Both the proposed and predicate devices are, in brief,

  • · intended for short- or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, for central venous pressure monitoring and power injection of contrast media.
  • · available in single and multi-lumen configurations in a wide range of sizes from 3F to 6 F outside catheter diameter;
  • · rated for maximum power injector settings up to 325 psi
  • · rated for maximum power injection flow rate up to 6 ml/second based on model; and
  • · available kitted with a range of procedural accessories for user convenience and,
  • · demonstrate resistance to blood components (platelet and thrombus) accumulation.

G. Performance Data

The NMI PICC IV is substantially equivalent to Navilyst predicate devices based on comparison of technological characteristics and the results of non-clinical tests which included the performance evaluation conducted in accordance with the following FDA guidance documents, international standards, and testing which included:

  • EN ISO 10555-1:2009, Sterile, Single use intravascular catheters Part 1: General . Requirements
  • EN ISO 10555-3:1997 Corrigendum 1:2002, Sterile, Single-Use Intravascular Catheters -. Part 3: Central Venous Catheters
  • FDA's "Guidance on Premarket Notification [510(K)] Submissions for Short-Term and . Long-Term Intravascular Catheters dated March 16, 1995"
  • . Gravity Flow Rate
  • Priming Volume .

H. Conclusion

The results of the non-clinical testing and a comparison of similarities and differences demonstrate that the proposed and predicate devices are substantially equivalent.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 5, 2014

Navilyst Medical, Incorporation Mr. Brandon Brackett Specialist, Global Regulatory Affairs 26 Forest Street Marlborough, MA 01752

Re: K140266

Trade/Device Name: NMI PICC IV Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: January 31, 2014 Received: February 3, 2014

Dear Mr. Brackett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Brackett

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerelv vours.

Image /page/3/Picture/8 description: The image shows the text "Kwame Ulmer's" in a stylized font. The text is arranged in two lines, with "Kwame" on the top line and "Ulmer's" on the bottom line. The letters are bold and have a slightly distressed appearance. The letters "EDA" are in the middle of the text, with the "O" and triangle symbol to the right of the letters.

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

for

Enclosure

{4}------------------------------------------------

Navilyst Medical, Inc. NMI PICC IV, Special 510(k) 31-January-2014

INDICATIONS FOR USE STATEMENT

510(k) Number (if Known):

K140266

Device Name: NMI PICC IV

Indications for Use:

The NMI PICC IV is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, for central venous pressure monitoring and for power injection of contrast media.

Maximum Power Injection Flow Rate
DescriptionFlow Rate
3F Single Lumen - 55cm length1 mL/sec
4F Single Lumen - 55cm length3.5 mL/sec
5F Single Lumen - 55cm length5 mL/sec
5F Dual Lumen - 55 cm length4 mL/sec
6F Dual Lumen - 55cm length5 mL/sec
6F Triple Lumen - 55cm length6 mL/sec

Prescription Use (21 CFR 801 Subpart D) ਲ And/Or AND/OR Over-The-Counter Use: (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Image /page/4/Picture/13 description: The image shows a stylized logo of the letters "FDA". The letters are formed by multiple parallel lines, giving them a layered or outlined appearance. The overall design is simple and geometric, with the letters being easily recognizable despite the abstract representation.

Digitally.signed by Richard C. Chapman Date: 2014.03.04 13:09:50 -05'00'

14

§ 880.5970 Percutaneous, implanted, long-term intravascular catheter.

(a)
Identification. A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”