K Number
K140175

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-08-05

(194 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Delta Fly F20μ Micro Bore Winged Safety Needles and the Delta Fly Micro Bore Winged Safety Needles are intended for short term use to access the peripheral vascular system for intravenous administration of fluids using a syringe or other compatible / appropriate devices.

Device Description

The Delta Med Delta Fly Winged Safety Needles are available in four versions, as follows:

  • o Delta Fly F20μ Micro Bore Winged Safety Needle 25G x 35 cm
  • . Delta Fly F20μ Micro Bore Winged Safety Needle 23G x 35 cm
  • Delta Fly Micro Bore Winged Safety Needle 25G x 35 cm ●
  • Delta Fly Micro Bore Winged Safety Needle 23G x 35 cm

The devices consist of a stainless steel needle encased in a body of plastic material, attached to a micro bore tube, ending proximally with a female Luer lock fitting closed by a final male Luer lock cap. The female Luer lock fitting includes a 20μ woven mesh filter in the F20μ versions of the device. The devices are equipped with a passive needle stick safety system for covering the tip of the needle upon needle withdrawal, in order to protect the operator from accidental needle stick injury.

To help identification between the two types (with and without filter), the male Luer lock proximal end cap is colored white on the versions with the filter and is transparent on the versions without the filter. To identify the needle size of each version, the winged body of the devices are color-coded blue for the 23G versions and orange for the 25G versions.

The distal end of the device is a stainless steel needle point has a triple bevel design. At the proximal end is a Luer lock fitting. Both device versions (with and without the mesh filter) have a final female connection in compliance with the requirements of ISO 594-2.

Delta Fly Micro Bore Winged Safety Needles are supplied for short term use only, sterile for single use, sterilized with ethylene oxide (EO) gas, and meet the biocompatibility requirements of ISO 10993-1:2009.

AI/ML Overview

The provided document describes the Delta Fly F20u Micro Bore Winged Safety Needles and Delta Fly Micro Bore Winged Safety Needles. Here's an analysis of the acceptance criteria and supporting studies:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally implied by adherence to established international standards and specific performance tests. The reported device performance indicates that these criteria were met.

Acceptance Criteria (Implied)Reported Device Performance
Material & Physical Properties
- Surface integrity (ISO 9626)Tests carried out in accordance with applicable requirements.
- Corrosion resistance (ISO 9626)Tests carried out in accordance with applicable requirements.
- Gauging (ISO 594-2)Tests carried out in accordance with applicable requirements.
- Needle material compliance (ISO 9626)Tests carried out in accordance with applicable requirements.
- Needle stiffness (ISO 9626)Tests carried out in accordance with applicable requirements.
- Resistance of tubing to breakage (ISO 9626)Tests carried out in accordance with applicable requirements.
- Tensile strength (winged needle, adapter-tube, wings-tube)Performed and confirmed substantial equivalence.
Functional Performance
- Liquid leakage (ISO 594-2)Tests carried out in accordance with applicable requirements.
- Air leakage during aspiration (ISO 594-2)Tests carried out in accordance with applicable requirements.
- Unscrewing torque (ISO 594-2)Tests carried out in accordance with applicable requirements.
- Separation force (ISO 594-2)Tests carried out in accordance with applicable requirements.
- Ease of assembly (ISO 594-2)Tests carried out in accordance with applicable requirements.
- Resistance of overriding (ISO 594-2)Tests carried out in accordance with applicable requirements.
- Stress cracking (ISO 594-2)Tests carried out in accordance with applicable requirements.
- Flow rate determinationPerformed and confirmed substantial equivalence.
- Dead space volumePerformed and confirmed substantial equivalence.
- Air leakage testPerformed and confirmed substantial equivalence.
Safety Features
- Safety winged needle activation (ISO 23908)Tests carried out in accordance with applicable requirements.
- Safety device test (ISO 23908)Tests carried out in accordance with applicable requirements.
- Correct activation of safety mechanism (Simulated Clinical Usage)100% of safety mechanism activated correctly.
- No test failures (Simulated Clinical Usage)No test failures occurred.
Biocompatibility
- Biocompatibility requirements (ISO 10993-1:2009)Meet biocompatibility requirements.
Sterilization
- Sterile for single use (Ethylene Oxide)Sterilized with ethylene oxide (EO) gas.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document explicitly states that the Simulated Clinical Usage test was conducted at "four different hospitals to allow a sufficient number of healthcare professional who routinely use Winged Needles". However, the exact number of devices or uses tested is not specified within the provided text.
  • Data Provenance: The manufacturer, Delta Med SpA, is located in Viadana (Mantova), Italy. Given the context, the tests were likely conducted in Italy or within Europe under the manufacturer's supervision, but the specific country of origin for the data (other than the manufacturer's location) for all tests is not detailed. The Simulated Clinical Usage test was performed at "four different hospitals," implying a prospective collection of data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: The Simulated Clinical Usage test involved "healthcare professional who routinely use Winged Needles." The exact number of individual healthcare professionals is not specified.
  • Qualifications of Experts: They are described as "healthcare professional who routinely use Winged Needles," indicating practical experience with the device type. Specific qualifications (e.g., nurse, physician, years of experience) are not provided.

4. Adjudication Method for the Test Set

The document does not describe a formal adjudication method (like 2+1 or 3+1). The "Simulated Clinical Usage test" results state that "100% of the safety mechanism of the devices activated correctly" and "No tests failures occurred," suggesting observation and recording of outcomes without mentioning a specific adjudication process for discrepancies, if any. Given the nature of the test (device activation and failure), it might have been a direct observation without requiring expert adjudication in the traditional sense of diagnostic interpretation.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No. The document describes performance testing and a simulated clinical usage test to demonstrate the device's functionality and safety, and substantial equivalence to predicate devices. It does not describe a multi-reader multi-case comparative effectiveness study comparing human readers with and without AI assistance. This device is a medical needle, not an AI diagnostic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable as the device is a physical medical needle, not an algorithm or AI system.

7. The Type of Ground Truth Used

For the performance tests (e.g., surface testing, corrosion, gauging, leakage, tensile strength, flow rate), the ground truth is established by compliance with specified international standards (ISO 594-2, ISO 9626, ISO 23908). This is a form of empirical and objective measurement against defined benchmarks.

For the "Simulated Clinical Usage test," the ground truth for "correct activation of the safety mechanism" and "test failures" would be based on direct observation and operational assessment by the participating healthcare professionals against the device's intended functionality and safety design.

8. The Sample Size for the Training Set

This question is not applicable as the device is a physical medical needle and does not involve AI or machine learning models that require a training set.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable as the device is a physical medical needle and does not involve AI or machine learning models that require a training set.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. The bottom half of the circle contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". In the center of the seal are three stylized profiles of human faces.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 5, 2014

Delta Med Spa C/O Mr. Roger Gray VP, Quality and Regulatory Donawa LifeScience Consulting Srl Piazza Albania, Nº 10 Rome, Italy 00153

Re: K140175

Trade/Device Name: Delta Fly F20u Micro Bore Winged Safety Needles and Delta Fly Micro Bore Winged Safety Needles Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: July 3, 2014 Received: July 8, 2014

Dear Mr. Gray:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Gray

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mary S. Runner -S

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K140175

Device Name

Delta Fly F20u Micro Bore Winged Safety Needles and Delta Fly Micro Bore Winged Safety Needles

Indications for Use (Describe)

The Delta Fly F20u Micro Bore Winged Safety Needles and the Delta Fly Micro Bore Winged Safety Needles are intended for short term use to access the peripheral vascular system for intravenous administration of fluids using a syringe or other compatible / appropriate devices.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)Image: FDA logo
Digitally signed by Richard
C. Chapman -S
Date: 2014.08.04 16:02:40
-04'00'

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Delta Med Medical Devices. The logo consists of a blue triangle shape on the left, followed by the text "DELTA MED" in gray, with the words "MEDICAL DEVICES" in a smaller font size underneath. Below the company name is the text "Solutions for Health" in a blue, cursive font.

510(k) Summary in accordance with 21 CFR 807.92(c)

Device name:Delta Fly F20μ Micro Bore Winged Safety Needles and Delta Fly MicroBore Winged Safety Needles
Type of 510 (k) submission:Traditional
Date of Submission:21 January 2014
Manufacturer:Delta Med SPAVia Guido Rossa 20I-46019 Viadana (Mantova)Italy
FDA Registration Number:3006846316
510 (k) Owner:Delta Med SPAVia Guido Rossa 20I-46019 Viadana (Mantova)Italy
Contact Person:Dr Laura Tellini
Phone:Fax:+39 03 757 85915+39 03 757 85201
510 (k) Contact:Mr Roger GrayVP Quality and RegulatoryDonawa Lifescience ConsultingPiazza Albania 1000153 RomeItaly
Phone:+39 03 757 85915
Fax:Email:+39 03 757 85201rgray@donawa.com
FDA Product Code:FPA
FDA Regulation Number:880.5440
FDA Classification Name:Set, Administration, Intravascular
Classification panel:General Hospital and Personal Use Devices
Common name:Intravascular administration set or Winged safety needle
FDA Classification:Class II

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Delta Med Medical Devices. The logo consists of a blue delta symbol with a curved line underneath it, followed by the text "DELTA MED" in gray, with "MEDICAL DEVICES" in a smaller font size underneath. Below the company name is the text "Solutions for Health" in blue.

  • FDA identification: An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion filter, an I.V. set stopcock, fluid delivery tubing, connector between parts of set, a side tube with a cap to serve as injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
  • Indications for use: The Delta Fly F20μ Micro Bore Winged Safety Needles and the Delta Fly Micro Bore Winged Safety Needles are intended for short term use to access the peripheral vascular system for intravenous administration of fluids using a syringe or other compatible / appropriate devices.

Device description:

The Delta Med Delta Fly Winged Safety Needles are available in four versions, as follows:

  • o Delta Fly F20μ Micro Bore Winged Safety Needle 25G x 35 cm
  • . Delta Fly F20μ Micro Bore Winged Safety Needle 23G x 35 cm
  • Delta Fly Micro Bore Winged Safety Needle 25G x 35 cm ●
  • Delta Fly Micro Bore Winged Safety Needle 23G x 35 cm

The devices consist of a stainless steel needle encased in a body of plastic material, attached to a micro bore tube, ending proximally with a female Luer lock fitting closed by a final male Luer lock cap. The female Luer lock fitting includes a 20μ woven mesh filter in the F20μ versions of the device. The devices are equipped with a passive needle stick safety system for covering the tip of the needle upon needle withdrawal, in order to protect the operator from accidental needle stick injury.

To help identification between the two types (with and without filter), the male Luer lock proximal end cap is colored white on the versions with the filter and is transparent on the versions without the filter. To identify the needle size of each version, the winged body of the devices are color-coded blue for the 23G versions and orange for the 25G versions.

The distal end of the device is a stainless steel needle point has a triple bevel design. At the proximal end is a Luer lock fitting. Both device versions (with and without the mesh filter) have a final female connection in compliance with the requirements of ISO 594-2.

Delta Fly Micro Bore Winged Safety Needles are supplied for short term use only, sterile for single use, sterilized with ethylene oxide (EO) gas, and meet the biocompatibility requirements of ISO 10993-1:2009.

The Delta Fly devices meet the relevant technical requirements of the following standard:

  • . ISO 594-2:1998
  • ISO 9626:1991 including Amendment 1 ●
  • ISO 23908:2011 ●

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for Delta Med Medical Devices. The logo features a blue abstract symbol on the left, followed by the words "DELTA MED" in gray, with "MEDICAL DEVICES" in smaller gray letters underneath. Below the company name is the text "Solutions for Health" in blue.

Performance Data:

Tests carried out on the Delta Fly devices in accordance with the applicable requirements of the above standards include:

  • . Surface testing (ISO 9626)
  • Corrosion testing (ISO 9626)
  • Gauging (ISO 594-2)
  • . Liquid leakage (ISO 594-2)
  • Air leakage during aspiration (ISO 594-2)
  • Unscrewing torque (ISO 594-2)
  • . Separation force (ISO 594-2)
  • Easy to assembly ( ISO 594-2)
  • Resistance of overriding (ISO 594-2)
  • Stress cracking (ISO 594-2) o
  • Needle material (ISO 9626)
  • Needle stiffness (ISO 96269)
  • Resistance of tubing to breakage (ISO 96269)
  • Safety winged needle activation (ISO 23908)
  • Safety device test (ISO 23908)

Furthermore, to confirm substantial equivalence to the predicate devices, the following tests were performed:

  • Tensile strength winged needle (15N x 15 seconds)
  • Tensile strength between wings and tube
  • Tensile strength between adapter and tube
  • Flow rate determination
  • Dead space volume
  • Air leakage test

In addition a Simulated Clinical Usage test was carried out in accordance with:

  • FDA "Guide for Industry and staff – Medical Devices with Sharp injury Prevention Features"
  • . ISO 23908:2011 "Sharps injury protection. Requirements and Test method. Sharp protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling".

Delta Med conducted the study at four different hospitals to allow a sufficient number of health care professional who routinely use Winged Needles to provide a meaningful feedback on the device design.

The following significant points were resulted:

  • During the tests, 100% of the safety mechanism of the devices activated correctly
  • No tests failures occurred.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Delta Med Medical Devices. The logo features a blue stylized delta symbol with a swoosh underneath it. To the right of the symbol, the words "DELTA MED" are written in gray, with the words "MEDICAL DEVICES" written in a smaller font size underneath. Below the company name, the words "Solutions for Health" are written in blue script.

  • All the results obtained give a positive evaluation of the Winged Needle safety mechanism. .

Safety Mechanism

The safety device for the Delta Fly Micro Bore Winged Safety Needles is a component formed by a cover permanently connected to a support element which is in contact with patient's skin.

During activation of the safety feature, the cover completely hides the needle in such a way that the needle is completely encapsulated in the internal concave section of the cover.

During withdrawal of the needle at the end of the treatment, the user, with one hand and limited pressure, pulls back the device tube while with the other hand slides and locks the safety feature into place over the needle.

Comparison with predicate devices:

The predicate device selected for comparison with the Delta Fly F20μ Micro Bore Winged Safety Needle is:

Predicate Device 1 (PD 1):Surflo Winged Infusion Set with Filter and Needle Protection (Surshield)
510(k) Sponsor:Terumo Europe NV
510(k) Number:K070547
Clearance Date:25 May 2007
FDA Product Code:FPA
Classification Name:Set, Administration, Intravascular
Regulation No:880.5440

The predicate device selected for comparison with the Delta Fly Micro Bore Winged Safety Needle is:

Predicate Device 2 (PD 2):...... Surflo Winged Infusion Set with Needle Protection (Surshield) 510(k) Sponsor: .............................................................................................................................................................. 510(k) Number: ......................................... K100946 Clearance Date:............................................................................................................................................................... FDA Product Code: ..................................... FPA Classification Name:...................... Set, Administration, Intravascular Regulation No: ...............................................................................................................................................................

The only significant difference between the subject devices and the predicate devices is that each have a safety feature to help prevent needlestick injuries. Even though differences exist between the methods employed for protecting the needle tip, the subject and predicate devices are equivalent in this and all other respects.

Conclusion:

No other significant differences that could affect device safety or effectiveness exist between the subject devices and the predicate devices, therefore it is concluded that:

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Delta Med. The logo consists of a blue triangle with a curved line underneath it, followed by the words "DELTA MED" in gray, with the words "MEDICAL DEVICES" in smaller gray letters underneath. Below that, the words "Solutions for Health" are written in blue.

  • . The Delta Fly F20μ Micro Bore Winged Safety Needle is substantially equivalent to the predicate device Terumo Surflo Winged Infusion Set with Filter and Needle Protection (Surshield), which is already in interstate commerce within the USA;
    and

  • The Delta Fly Micro Bore Winged Safety Needle is substantially equivalent to the predicate . device Terumo Surflo Winged Infusion Set with Needle Protection (Surshield), which is already in interstate commerce within the USA.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.