K Number
K140122

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-05-09

(113 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GentleMAX Family of Laser Systems is indicated for the following at the specified wavelength:

755mm
Temporary hair reduction. Stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin.
Treatment of benign pigmented lesions.
Treatment of wrinkles.
The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).

1064nm
Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term. stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.
The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
Treatment of wrinkles.

1064nm
Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and/or yeast Candida Albicans, etc.)

Device Description

The Candela GentleMAX Family of Laser Systems contains two separate laser heads (Alexandrite and Nd: YAG), which produce laser light outputs of 755 nm and 1064 nm. respectively. The output of each laser head is optically combined on the laser rail, so that their beam paths are identical as they exit the laser system. This allows the use of a single delivery system which can output either 755 nm or 1064 nm wavelengths. The laser system creates a beam of high intensity light that penetrates deep into the skin tissue where it delivers a controlled amount of therapeutic heat. The Dynamic Cooling Device (DCD) protects the upper layers of the skin with a cooling burst of cryogen.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information based on the provided text, focused on the GentleMAX Family of Laser Systems for its primary indications (hair reduction, pigmented lesions, wrinkles, vascular lesions) and the new indication (onychomycosis) which is the focus of this 510(k) submission:

Acceptance Criteria and Device Performance

The provided document is a 510(k) summary for the GentleMAX Family of Laser Systems, specifically regarding a modification that includes a new indication for onychomycosis. For the existing indications (hair reduction, pigmented lesions, wrinkles, vascular lesions), the submission asserts substantial equivalence to previously cleared devices. For the new onychomycosis indication, the substantial equivalence is drawn against the Cutera GenesisPlus Laser (K122493).

The "acceptance criteria" in this context are primarily based on demonstrating substantial equivalence to existing predicate devices, meaning the device performs as safely and effectively as a legally marketed device without raising new questions of safety or effectiveness. For medical devices, this is often assessed through bench testing, device specifications, and sometimes clinical data.

For the existing indications (Hair Reduction, Benign Pigmented Lesions, Wrinkles, Vascular Lesions):

The acceptance criteria are implicitly met by demonstrating that the Modified GentleMAX Family of Laser Systems maintains the same technological characteristics and intended uses as the GentleMAX Family of Laser Systems (K133283). The "reported device performance" in this context is that its performance specifications (wavelength, pulse duration, fluence, spot size, repetition rate, etc.) are identical to the predicate.

Acceptance Criteria (based on Predicate K133283)Reported Device Performance (Modified GentleMAX Family)
Maintain 755nm Wavelength755nm
Maintain 1064nm Wavelength1064nm
Maintain Pulse duration (0.25 - 100 ms)0.25 - 100 ms
Maintain Maximum Fluence (53 J/cm2 for ALEX, 80 J/cm2 for YAG)53 J/cm2 (ALEX), 80 J/cm2 (YAG)
Maintain Spot Sizes (1.5, 3, 6, 8, 10, 12, 15, 18mm)1.5, 3, 6, 8, 10, 12, 15, 18mm
Maintain Pulse Repetition Rate (10 Hz, maximum)10 Hz, maximum
Maintain Pulsing Control (Fingerswitch or footswitch)Fingerswitch or footswitch
Same Intended Uses at 755nm and 1064nmIdentical Intended Uses for these wavelengths (e.g., hair reduction, pigmented lesions, vascular lesions, wrinkles)

For the new indication (Temporary increase of clear nail in patients with onychomycosis at 1064nm):

The acceptance criteria are established by comparing to the Cutera GenesisPlus Laser System (K122493).

Acceptance Criteria (based on Predicate K122493)Reported Device Performance (Modified GentleMAX Family)
1064nm Nd:YAG Wavelength1064nm Nd:YAG
0.3ms Pulse Duration0.3ms
15-18 J/cm2 Fluence15-18 J/cm2
5mm Spot Size5mm
2-3 Hz Repetition Rate2-3 Hz
Pulsed Output ModePulsed
Indications for Use (Temporary increase of clear nail in onychomycosis patients)Identical Indications for Use for onychomycosis

Study Information

The document explicitly states: "No clinical data was required." This is a key piece of information. The "study" proving the device meets acceptance criteria primarily relies on bench testing and the demonstration of substantial equivalence through comparison of technological characteristics and intended uses with predicate devices.

1. Sample size used for the test set and the data provenance:
* Test Set Sample Size: Not applicable. The submission states "No clinical data was required," indicating there was no clinical test set of patients or anonymized data sets from patients used for this 510(k) submission.
* Data Provenance: Not applicable for a clinical test set. The data provenance would be from internal bench testing (laboratory-generated) and comparison of specifications from the predicate devices' publicly available 510(k) information.

2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
* Not applicable as no clinical test set was used. Ground truth for device specifications would be based on engineering and manufacturing standards.

3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
* Not applicable as no clinical test set requiring expert adjudication was used.

4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
* Not applicable. This device is a laser system for dermatology, not an AI-assisted diagnostic or imaging system that would involve human readers interpreting cases.

5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
* Not applicable. This is a physical medical device, not an algorithm or software-only device. Performance is evaluated based on its physical characteristics and functionality.

6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
* The "ground truth" for this submission is based on engineering specifications and performance data collected during bench testing (e.g., laser output measurements, pulse duration, fluence, spot size consistency). For the new onychomycosis indication, the ground truth for equivalence is the performance specifications of the predicate device (Cutera GenesisPlus Laser System K122493).

7. The sample size for the training set:
* Not applicable. This is a medical device clearance based on substantial equivalence and bench testing, not a machine learning or AI-driven device requiring training data.

8. How the ground truth for the training set was established:
* Not applicable as no training set was used.

{0}------------------------------------------------

510(k) Summary GentleMAX Family of Laser Systems

General Information:

This Traditional 510(k) is to provide notification of substantial equivalence for the GentleMAX Family of Laser Systems (GentleMax Pro Laser System) to the cleared Cutera GenesisPlus Laser (K122493) manufactured by Cutera, Inc. and the cleared GentleMAX Family of Laser Systems (K133283).

Submitted by:Candela Corporation530 Boston Post RoadWayland, MA 01778-1886
Contact Person:Sam Wade,Global Vice President, Regulatory AffairsTel: 508-358-7400 x330Fax: 508-358-5602
Date prepared:January 13, 2014
Trade Name:GentleMAX Family of Laser Systems
Common Name:Dermatology Laser System
Classification Name:Laser surgical instrument for use in general and plastic surgery andin dermatology (21 CFR 878.4810, Product Code GEX)
Predicate Devices:GentleMAX Family of Laser Systems (K133283)Cutera GenesisPlus Laser System (K122493)

Intended Use / Indications for Use:

The GentleMAX Family of Laser Systems is indicated for the following at the specified wavelength:

755mm

Temporary hair reduction. Stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin.

Treatment of benign pigmented lesions.

Treatment of wrinkles.

{1}------------------------------------------------

The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).

1064nm

Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term. stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.

The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.

Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.

Treatment of wrinkles.

1064nm

Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and/or yeast Candida Albicans, etc.)

Description:

The Candela GentleMAX Family of Laser Systems contains two separate laser heads (Alexandrite and Nd: YAG), which produce laser light outputs of 755 nm and 1064 nm. respectively. The output of each laser head is optically combined on the laser rail, so that their beam paths are identical as they exit the laser system. This allows the use of a single delivery system which can output either 755 nm or 1064 nm wavelengths. The laser system creates a beam of high intensity light that penetrates deep into the skin tissue where it delivers a controlled amount of therapeutic heat. The Dynamic Cooling Device (DCD) protects the upper layers of the skin with a cooling burst of cryogen.

Technological Characteristics:

The Candela GentleMAX Family of Laser Systems delivers laser energy through an optical fiber handpiece delivery system, which can output either 755 nm or 1064 nm wavelengths. The output of this laser is delivered to the area of treatment by means of a lens coupled user replaceable optical fiber with a treatment handpiece attached to its distal end. A trigger switch (fingerswitch or footswitch) is used to control the delivery of laser pulses. The Dynamic Cooling Device provides a short burst of cryogen spray prior to firing the laser pulse. The laser output energy is delivered via an optical fiber to a handpiece, which produces circular beams with diameters of

{2}------------------------------------------------

1.5, 3, 6, 8, 10, 12, 15 and 18 millimeters on the skin. The cryogen, which is housed within the laser enclosure, is delivered via a hose to a nozzle located in the handpiece.

Summary of Technological CharacteristicsProductModified GentleMAX Family ofLaser Systems (GentleMax Pro Laser System)Cutera GenesisPlus Laser System
510(k)K140122K122493
ManufacturerCandela CorporationCutera, Inc.
Product CodeGEXPDZ, GEX
Regulation Number21 CFR 878.481021 CFR 878.4810
Device ClassIIII
Indications for UseThe GentleMAX Family of Laser Systems is indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and/or yeast Candida Albicans, etc.).The Cutera GenesisPlus laser is indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and/or yeast Candida Albicans, etc.).
Laser MediaFlashlamp pumped solid state rodFlashlamp pumped solid state rod
Wavelength1064nm Nd:YAG1064nm Nd:YAG
Pulse duration0.3ms0.3ms
Fluence15-18 J/cm215-18 J/cm2
Spot size5mm5mm
Repetition rate2-3 Hz2-3 Hz
Laser activationFingerswitch or footswitchFootswitch
Output modePulsedPulsed
User InterfaceLCD color touchscreenLCD color touchscreen
Summary of Technological Characteristics
ProductModified GentleMAX Family ofLaser Systems (GentleMax ProLaser System)GentleMAX Family of LaserSystems (GentleMax Pro LaserSystem)
510(k)K140122K133283
ManufacturerCandela CorporationCandela Corporation
Product CodeGEXGEX
Regulation Number21 CFR 878.481021 CFR 878.4810
Device ClassIIII
Indications for Use755nm755nm
The GentleMAX Family of LaserSystems is indicated forThe GentleMAX Family of LaserSystems is indicated for

{3}------------------------------------------------

temporary hair reduction. Stable temporary hair reduction. Stable long-term or permanent reduction long-term or permanent reduction through selective targeting of through selective targeting of melanin in hair follicles. melanin in hair follicles. Permanent hair reduction is Permanent hair reduction is defined as long-term stable defined as long-term stable reduction in the number of hairs reduction in the number of hairs regrowing after a treatment regrowing after a treatment regime. On all skin types regime. On all skin types (Fitzpatrick I- VI) including (Fitzpatrick I- VI) including tanned skin. Permanent hair tanned skin. Permanent hair reduction is defined as the longreduction is defined as the longterm, stable reduction in the term, stable reduction in the number of hairs regrowing when number of hairs regrowing when measured at 6, 9, and 12 months measured at 6, 9, and 12 months after the completion of a after the completion of a treatment regime. treatment regime. Treatment of benign pigmented Treatment of benign pigmented lesions. lesions. Treatment of wrinkles. Treatment of wrinkles. The photocoagulation of The photocoagulation of dermatological vascular lesions dermatological vascular lesions (such as port-wine stains, (such as port-wine stains, hemangiomas, telangiectasias). hemangiomas, telangiectasias). 1064nm 1064nm The GentleMAX Family of Laser The GentleMAX Family of Laser Systems is indicated for the Systems is indicated for the removal of unwanted hair, for removal of unwanted hair, for stable long term or permanent stable long term or permanent hair reduction and for treatment hair reduction and for treatment of PFB. Permanent hair reduction of PFB. Permanent hair reduction is defined as the long-term, stable is defined as the long-term, stable reduction in the number of hairs reduction in the number of hairs regrowing when measured at 6, 9, regrowing when measured at 6, 9. and 12 months after the and 12 months after the completion of a treatment regime. completion of a treatment regime. The lasers are indicated on all The lasers are indicated on all skin types Fitzpatrick I-VI skin types Fitzpatrick I-VI including tanned skin. including tanned skin. Photocoagulation and hemostasis Photocoagulation and hemostasis of pigmented and vascular lesions of pigmented and vascular lesions such as but not limited to port such as but not limited to port wine stains, hemangioma, warts, wine stains, hemangioma, warts, telangiectasia, rosacea, venus telangiectasia, rosacea, venus lake, leg veins and spider veins. lake, leg veins and spider veins.

{4}------------------------------------------------

Coagulation and hemostasis ofCoagulation and hemostasis of
soft tissue. Benign pigmentedsoft tissue. Benign pigmented
lesions such as, but not limited to,lesions such as, but not limited to,
lentigos (age spots), solar lentigoslentigos (age spots), solar lentigos
(sun spots), café au lait macules,(sun spots), café au lait macules,
seborrheic keratosis, nevi,seborrheic keratosis, nevi,
chloasma, verrucae, skin tags.chloasma, verrucae, skin tags,
keratosis, tattoos (significantkeratosis, tattoos (significant
reduction in the intensity of blackreduction in the intensity of black
and/or blue-black tattoos) andand/or blue-black tattoos) and
plaques.plaques.
The laser is indicated forThe laser is indicated for
pigmented lesions to reducepigmented lesions to reduce
lesion size, for patients withlesion size, for patients with
lesions that would potentiallylesions that would potentially
benefit from aggressive treatment,benefit from aggressive treatment,
and for patients with lesions thatand for patients with lesions that
have not responded to other laserhave not responded to other laser
treatments.treatments.
The laser is also indicated for theThe laser is also indicated for the
reduction of red pigmentation inreduction of red pigmentation in
hypertrophic and keloid scarshypertrophic and keloid scars
where vascularity is an integralwhere vascularity is an integral
part of the scar.part of the scar.
Treatment of wrinkles.Treatment of wrinkles.
Laser typeFlashlamp-excited, Solid stateFlashlamp-excited, Solid state
Alexandrite and Nd: Y AG laserAlexandrite and Nd: Y AG laser
Wavelength755nm/1064nm755nm/1064nm
Pulse duration0.25 - 100 ms0.25 - 100 ms
Maximum fluence53 J/cm2 (ALEX)53 J/cm2 (ALEX)
80 J/cm2 (YAG)80 J/cm2 (YAG)
Spot size1.5, 3, 6, 8, 10, 12, 15, 18mm1.5, 3, 6, 8, 10, 12, 15, 18mm
Pulse repetition rate10 Hz, maximum10 Hz, maximum
Pulsing controlFingerswitch or footswitchFingerswitch or footswitch
Product dimensions42" x 18" x 27"42" x 18" x 27"
(HxWxL)
Product Weight260 lbs260 Ibs

·

{5}------------------------------------------------

Performance Data:

Testing to the third edition of IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-22 has been conducted on the GentleMax Pro Laser System. Bench testing has also been performed to show that the performance specifications of our product are identical to the predicate device. All performance testing demonstrated that the GentleMax Pro Laser System performs according to specifications and functions as intended.

Clinical Data:

No clinical data was required.

Summary of Substantial Equivalence:

The GentleMAX Family of Laser Systems have the same intended uses, utilize similar operating principles, and match key design aspects, including similar spot size, the same wavelengths and the variable delivered fluence, in comparison to the predicate devices. The addition of a new indication to the GentleMAX Family of Laser Systems raises no new issues of safety or effectiveness. On the basis of similarities in method of operation, intended uses, and key technical characteristics, Candela Corporation believes that the GentleMAX Family of Laser Systems is substantially equivalent to the predicate devices.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 9, 2014

Candela Corporation Mr. Sam Wade Global Vice President, Regulatory Affairs 530 Boston Post Road Wayland, Massachusetts 01788

Re: K140122

Trade/Device Name: GentleMAX Family of Laser Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX, PDZ Dated: April 28. 2014 Received: April 29, 2014

Dear Mr. Wade:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warrantics. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{7}------------------------------------------------

Page 2 - Mr. Sam Wade

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours, Binita S. Ashar -S 2014.05.09 15:13:43 -04'00'

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510{k) Number (if known)

K140122

Device Name GentleMAX Family of Laser Systems

Indications for Use (Describe)

The GentleMAX Family of Laser Systems is indicated for the specified wavelength:

755mm

Temporary hair reduction. Stable long-term or permanent reductive targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I-VI) including tanned skin.

Treatment of benign pigmented lesions.

Treatment of wrinkles.

The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectas)

1064nm

Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, sebortheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.

The laser is indicated for pigmented lesion size. for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.

Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.

Treatment of wrinkles.

1064nm

Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and /or yeast Candida Albicans, etc.)

Type of Use (Select one or both, as applicable)

8 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FORM FDA 3881 (1/14)

Page 1 of 2

PSC Publishing Services (301) 443-6746 EF

GentleMAX Family of Laser Systems

6-1 A1-1

{9}------------------------------------------------

FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Neil R Ogden -S For BSA 2014.05.08 09:15:37 -04'00' This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)

Page 2 of 2

Image /page/9/Picture/10 description: The image shows the text "6-2 A1-2" stacked vertically. The number 6 is on top of the number 2, and the letter A is on top of the number 1, which is on top of the number 2. The text is black and the background is white.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.