(115 days)
The InFill® cervical interbody fusion device is intervertebral body fusion of the spine in skeletally mature patients. The InFill® cervical interbody fusion device is designed for use with autogenous bone grafi to facilitate fusion. InFill® is intended for use at one level in the cervical spine, from C2 to T1, for the treatment of degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The devices are to be used in patients who have had at least six weeks of non-operative treatment. Patients with previous non-fusion spinal surgery at the involved level may be treated with this device. The InFill® cervical interbody fusion device is intended to be used with supplemental internal fixation systems that are cleared by the FDA for use in the cervical spine.
The InFill® Cervical Interbody Fusion device are radiolucent implantable devices manufactured from PEEK Optima LT1 and tantalum ASTM F560 (marker material). The implants are available in various sizes to suit the individual pathology and anatomical conditions of the patient.
The InFill® Cervical Interbody Fusion device is provided sterile, for single use only.
This is an intervertebral body fusion device, and the information provided details its substantial equivalence to predicate devices, rather than a clinical study with AI or human readers. Therefore, many of the requested categories are not applicable.
Here's the breakdown of the available information:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (Performance Test) | Reported Device Performance |
|---|---|
| Static/dynamic axial compression met criteria | Met established acceptance criteria |
| Static/dynamic torsion met criteria | Met established acceptance criteria |
| Subsidence met criteria | Met established acceptance criteria |
| Expulsion met criteria | Met established acceptance criteria |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
Not applicable. The provided text describes mechanical performance testing of a physical device, not a study involving human subjects or data provenance in the clinical sense.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. Ground truth in this context refers to engineering specifications and performance limits for mechanical testing, not clinical diagnoses by experts.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods are relevant for clinical studies with subjective interpretations, not for objective mechanical performance tests.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical device for surgical implantation, not an AI-powered diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth for the mechanical performance tests would implicitly be the pre-defined engineering standards and performance specifications for intervertebral body fusion devices, as established by relevant regulatory bodies (e.g., FDA guidance documents, ASTM standards).
8. The sample size for the training set
Not applicable. There is no training set as this is not a machine learning model.
9. How the ground truth for the training set was established
Not applicable. There is no training set.
{0}------------------------------------------------
MAY 0 5 2014
510(k) Summary
This 510(k) summary information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT: | Pinnacle Spine Group, LLC |
|---|---|
| DATE PREPARED: | May 1, 2014 |
| CONTACT PERSON: | Rebecca K Pine1601 Elm Street, Suite 1930Dallas, TX 75201Phone: 760.809.5178Fax: 760.290.3216 |
| TRADE NAME: | InFill® Cervical |
| COMMON NAME: | Spinal Implant |
| CLASSIFICATIONNAME: | Intervertebral Body Fusion Device |
| DEVICECLASSIFICATION: | Class II |
| REGULATIONNUMBER | 888.3080 (product code: ODP) |
| PREDICATE DEVICES: | Calix- Cervical Spinal Implant System (K083637, K112036)Bengal Cervical System (K081917) |
Substantially Equivalent To:
The InFill® Cervical Interbody Fusion device are substantially equivalent in intended use, principal of operation and technological characteristics to the Calix Cervical Spinal Implant System and the Bengal Cervical System.
Description of the Device Subject to Premarket Notification:
The InFill® Cervical Interbody Fusion device are radiolucent implantable devices manufactured from PEEK Optima LT1 and tantalum ASTM F560 (marker material). The implants are available in various sizes to suit the individual pathology and anatomical conditions of the patient.
The InFill® Cervical Interbody Fusion device is provided sterile, for single use only.
Indication for Use:
InFill® cervical interbody fusion device is indicated for intervertebral body fusion of the spine in skeletally mature patients. InFill® cervical interbody fusion device is designed
ﺮ ﺍ
{1}------------------------------------------------
for use with autogenous bone graft to facilitate fusion. InFill® is intended for use at one level in the cervical spine, from C2-T1, for the treatment of degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The devices are to be used in patients who have had at least six weeks of non-operative treatment. Patients with previous non-fusion spinal surgery at the involved level may be treated with this device. InFill® cervical interbody fusion device is intended to be used with supplemental internal fixation systems that are cleared by the FDA for use in the cervical spine.
Technical Characteristics:
The InFill® 41-TLIF Convex Oblique device, InFill® 43-TLIF Contour Oblique, InFill® 44-TLIF Contour Oblique and the InFill® 60-ALIF have similar physical and technical characteristics to the predicate devices, as illustrated in the table below.
| TechnicalCharacteristics | InFill® cervicalintervertebralbody fusiondevices | Calix CervicalSpinal ImplantSystem (K083637,K112036) | Bengal CervicalSystem (K081917) |
|---|---|---|---|
| Shape | Box-shaped, bulletnose | SAME | SAME |
| Bone to implantsurface | Surface teeth | SAME | SAME |
| Bone graftsupport feature | Centralfenestration | SAME | SAME |
| Primary implantmaterial | PEEK OPTIMALT1 ® | SAME | Carbonfiberreinforced polymer(PEEK) |
| SurgicalApproach | Anterior | Anterior | Anterior |
Performance Data:
To establish substantial equivalence with predicate devices, testing was performed including static/dynamic axial compression, static/dynamic torsion, subsidence and expulsion. All performance test values met the established acceptance criteria..
Basis for Determination of Substantial Equivalence:
Upon reviewing the performance data provided in this submission and comparing intended use, design, materials, principle of operation and overall technological characteristics, the InFill® Cervical Interbody Fusion devices are determined by Pinnacle Spine Group. LLC, to be substantially equivalent to existing legally marketed devices.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administr 10903 New Hampshire Aven Document Control Center - WO6 Silver Spring, MD 20993-0002
May 5, 2014
Pinnacle Spine Group, LLC % Ms. Rebecca K. Pine Consultant 1601 Elm Street. Suite 1930 Dallas, Texas 75201
Re: K140066
Trade/Device Name: InFill® Cervical Interbody Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class H Product Code: ODP Dated: April 3, 2014 Received: April 4, 2014
Dear Ms. Pinc:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA `s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Rebecca K. Pine
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801) please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Ronald P. Jean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K140066
Device Name
InFill® Cervical Interbody Fusion Device
Indications for Use (Describe)
The InFill® cervical interbody fusion device is intervertebral body fusion of the spine in skeletally mature patients. The InFill® cervical interbody fusion device is designed for use with autogenous bone grafi to facilitate fusion. InFill® is intended for use at one level in the cervical spine, from C2 to T1, for the treatment of degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The devices are to be used in patients who have had at least six weeks of non-operative treatment. Patients with previous non-fusion spinal surgery at the involved level may be treated with this device. The InFill® cervical interbody fusion device is intended to be used with supplemental internal fixation systems that are cleared by the FDA for use in the cervical spine.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY CALL . . .
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
James P. Bertram -S 2014.05.05 15:28:35 -04'00
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
- 1-861120 E Serves (301) 433-6740
N/A