K Number
K140041

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-11-28

(324 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ZEN-7000 is a mobile fluoroscopy system is designed to provide fluoroscopic and spot-film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, critical care and emergency room procedures. The system may be used for other imaging applications at the physician's discretion.

Device Description

ZEN-7000 mobile fluoroscopy system is an image intensified fluoroscopic system consisting of two mobile units: a Mainframe (C-Arm) and a Workstation. The Mainframe (C-Arm) is comprised of a high voltage generator, x-ray control, and a "C" shaped apparatus, which supports an X-ray tube and an image intensifier. The Mainframe is designed to perform linear and rotational motions that allow the user to position the x-ray imaging components at various angles and distances with respect to the patient. The Workstation is a mobile platform that supports image display monitors and recording devices.

AI/ML Overview

The document does not describe a clinical study or performance data based on specific acceptance criteria for the device's diagnostic performance, e.g., sensitivity, specificity, or accuracy.

Instead, this 510(k) submission relies on demonstrating substantial equivalence to predicate devices (GENORAY ZEN-7000, K103425, and GE Healthcare Surgery OEC 9900 Elite, K082781) by showing that the new device (GENORAY ZEN-7000, K140041) has similar technological characteristics and indications for use, and complies with relevant safety and performance standards.

The document indicates that the device meets acceptance criteria related to safety and effectiveness standards, not diagnostic accuracy for a specific clinical condition using a test set of patient data.

Here's a breakdown of the information provided, specifically addressing your points where data is available:

1. A table of acceptance criteria and the reported device performance

The document does not provide a table for diagnostic performance metrics (e.g., sensitivity, specificity) against specific acceptance criteria. Instead, the acceptance criteria are related to compliance with industry and regulatory standards for safety and performance of X-ray systems.

Acceptance Criteria CategoryReported Device Performance / Compliance
Electrical, Mechanical, Environmental SafetyComplies with IEC 60601-1 Series, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-43, and IEC 60601-2-54.
Electromagnetic Compatibility (EMC)Complies with IEC 60601-1-2.
X-ray Imaging Device GuidanceComplies with FDA Guidance "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices" (though the device is image-intensified, this guidance might overlap with general X-ray imaging).
Electronic Product Radiation Control (EPRC)Meets the EPRC standards (21 CFR 1020.30, 31).
Digital Imaging and Communications in Medicine (DICOM)Meets the provisions of NEMA PS 3.1-3.18, Digital Imaging and Communications in Medicine (DICOM) Set.
Quality Management SystemNew device modifications completed in accordance with GENORAY quality management system design controls.
Test Results"All test results were satisfactory and the result of bench and clinical evaluation indicates that the ZEN-7000 is as safe and effective as the predicate devices." (Note: "clinical evaluation" here likely refers to safety and foundational performance, not a diagnostic accuracy study with patient images).

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Not Applicable. The document does not describe a test set of patient cases or images for evaluating diagnostic performance. The evaluation focused on engineering performance and safety compliance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Not Applicable. As no diagnostic test set was used, no experts were required to establish ground truth for such a purpose.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not Applicable. No diagnostic test set or adjudication for ground truth of diagnostic findings was used.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This device is a Fluoroscopic X-ray system, not an AI-powered diagnostic tool. No MRMC study or AI assistance evaluation was conducted.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Not Applicable. This is a hardware imaging device, not an algorithm, and does not operate in a standalone diagnostic capacity separate from human operation and interpretation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not Applicable for diagnostic accuracy. The "ground truth" equivalent in this submission relates to established engineering standards, regulatory requirements, and the technical specifications of the predicate devices. For example, a "ground truth" for resolution would be the measurable resolution value achieved, compared against the specified minimums or predicate device values.

8. The sample size for the training set

  • Not Applicable. This is an X-ray hardware system, not a machine learning algorithm requiring a training set.

9. How the ground truth for the training set was established

  • Not Applicable. As no training set was used, this question is not relevant.

In summary: This 510(k) submission for the GENORAY ZEN-7000 (K140041) focuses on demonstrating substantial equivalence to previously cleared predicate devices through compliance with established safety, electrical, and performance standards relevant to X-ray imaging systems. It does not involve a clinical study to assess diagnostic accuracy or performance metrics using a patient-image test set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the department's name around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare. The caduceus is depicted with a staff entwined by two snakes and topped with wings.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 9, 2014

GENORAY Co., Ltd. % Ms. Kaitlynn Min Business Development Manager GENORAY America, Inc. 3002 Dow Avenue, Suite 420 TUSTIN CA 92780

Re: K140041

Trade/Device Name: Fluoroscopic X-ray System (Models: ZEN-7000) Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, OXO, JAA Dated: October 29, 2014 Received: October 30, 2014

Dear Ms. Min:

This letter corrects our substantially equivalent letter of November 28, 2014.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

Page 2 - Ms. Min

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D.'Hara

for

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K140041

Device Name

Fluoroscopic X-Ray System (Models: ZEN-7000)

Indications for Use (Describe)

ZEN-7000 is a mobile fluoroscopy system is designed to provide fluoroscopic and spot-film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, critical care and emergency room procedures. The system may be used for other imaging applications at the physician's discretion.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary Exhibit 5

Date of Summary Preparation: January 2, 2014

1.Submitter and US Official Correspondent:
1.1 Submitter
Submitter:GENORAY Co., Ltd.
Address:#512, Byucksan Technopia, 434-6, Sangdaewon 1-dong,
Jungwon-gu, Seongnam-city, Gyeonggi-do, 462-716, Korea
Telephone No.:+82-31-740-4100
Fax:+82-31-737-8025
1.2 Official Correspondent (U.S): Kaitlynn Min - Business Manager
Correspondent:GENORAY America Inc.
Address:3002 Dow Avenue, Suite 420, Tustin, CA 92780
Telephone No.:714-289-8020
Fax:714-786-8919
Email:kaitlynn@genorayamerica.com
2.Establishment Registration Number
3005843418
3.Device Information
Proprietary/Trade Name:ZEN-7000
Common/Usual Name:Fluoroscopic X-Ray System
Classification Name:System, X-Ray, Fluoroscopic, Image-Intensified
Product Code:Primary product code OWB
Secondary product code JAA, OXO
Device Class:Class II per regulation 21 CFR 892.1650
4.Legally Marketed Predicate Device
Manufacturer:GENORAY Co., Ltd.
Trade Name:ZEN-7000
510(k) Number:K103425 (Decision Date - Mar 29, 2011)
Classification Name:System, X-Ray, Fluoroscopic, Image-Intensified
Product Code:Primary product code OWB
Secondary product code JAA, OXO
Manufacturer:GE Healthcare Surgery
Trade Name:OEC 9900 Elite
510(k) Number:K082781 (Decision Date - May 01, 2009)
Classification Name:System, X-Ray, Fluoroscopic, Image-Intensified
Product Code:Primary product code OWB
Secondary product code JAA, OXO

{4}------------------------------------------------

5. Description of the Device

ZEN-7000 mobile fluoroscopy system is an image intensified fluoroscopic system consisting of two mobile units: a Mainframe (C-Arm) and a Workstation. The Mainframe (C-Arm) is comprised of a high voltage generator, x-ray control, and a "C" shaped apparatus, which supports an X-ray tube and an image intensifier. The Mainframe is designed to perform linear and rotational motions that allow the user to position the x-ray imaging components at various angles and distances with respect to the patient. The Workstation is a mobile platform that supports image display monitors and recording devices.

    1. Indications for use
      ZEN-7000 is a mobile fluoroscopy system is designed to provide fluoroscopic and spotfilm images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.

The system may be used for other imaging applications at the physician's discretion.

7. Substantial Equivalence Chart
Proposed devicePredicate devicePredicate device
GENORAYFluoroscopic X-Ray System,ZEN-7000GENORAYFluoroscopic X-Ray System,ZEN-7000K103425GE Healthcare SurgeryFluoroscopic X-Ray SystemOEC 9900 EliteK082781
Name
ManufacturerGENORAY Co., LtdGENORAY Co., LtdGE Healthcare Surgery
Indications for useZEN-7000 is a mobile fluoroscopy system is designed to provide fluoroscopic and spot-film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The system may be used for other imaging applications at the physician's discretion.ZEN-7000 is a mobile fluoroscopy system is designed to provide fluoroscopic and spot-film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The system may be used for other imaging applications at the physician's discretion.The OEC 9900 Elite mobile fluoroscopy system is designed to provide fluoroscopic and spot-film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The system may be used for other imaging applications at the physician's discretion.
GeneratorHigh Frequency InverterHigh Frequency InverterHigh Frequency Inverter
Max. output power5 kW15kW(Optional)5 kW15kW
X-ray tubeRotating tubeLarge: 0.6 mmSmall: 0.3 mmRotating tubeLarge: 0.6 mmSmall: 0.3 mmRotating tubeLarge: 0.6 mmSmall: 0.3 mm
Fluoroscopy40-120 kV / 0.2-6.0 mA40-120 kV / 0.2-6.0 mA40-120 kV / 0.2- 10 mA
PulsedFluoroscopy1 mA to 20 mA(5kW)1 mA to 48 mA(15kW)1 mA to 20 mA0.2 mA to 40 mA
Radiography40-120 kV / 1-100 mAs40-120 kV / 0.4-100 mAs40~120 kV / 1-300mAs
TVCameraCCD TypeCCD TypeCCD Type
9"(9"/6"/4.5")9"(9"/6"/4.5")
• Minimum central resolution(at the monitor):-9" (23cm): 2.2 lp/mm• Minimum central resolution(at the monitor):-9" (23cm): 2.1 lp/mm
-6" (15cm): 2.8 lp/mm9"(9"/6"/4.5")-6" (15cm): 2.9 lp/mm
-4.5" (11cm): 3.0 lp/mm• Minimum central resolution-4.5" (11cm): 3.4 lp/mm
ImageIntensif• DQE: 65% (typical)(at the monitor):-9" (23cm): 2.2 lp/mm• DQE: 65% (typical)
ier12"(12"/9"/6")-6" (15cm): 2.8 lp/mm12"(12"/9"/6")
• Minimum central resolution(at the monitor):-12" (31cm): 1.6 lp/mm-4.5" (11cm): 3.0 lp/mm• DQE: 65% (typical)• Minim m central resolution(at monitor):-12" (31cm): 1.5 lp/mm
-9" (23cm): 2.0 lp/mm-9" (23cm): 2.1 lp/mm
-6" (15cm): 2.5 lp/mm-6" (15cm): 2.6 lp/mm
• DQE: 65% (typical)• DQE: 65% (typical)
SID: 1000mmSID: 1000mmSID: 1000mm
Free space: 786 mmFree space: 750mmFree space: 787mm
Depth: 730 mmDepth: 700 mmDepth: 660mm
DimensionsC-arm wig wag: ±12.5°C-arm wig wag: ±12.5°C-arm wig wag: ±20°
Orbital Rotation: 135°Orbital Rotation: 135°Orbital Rotation: 115°
Vert. Travel: 500mmVert. Travel: 450mmVert. Travel: 457mm
Horiz. Travel: 200mmHoriz. Travel: 200mmHoriz. Travel: 203mm
  • Substantial Equivalongo Chart ー

{5}------------------------------------------------

The proposed ZEN-7000 is based on the predicate device, ZEN-7000(K103425). The generation is similar to predicate devices with optional output power, 15kW. And Image intensifier has 9" and 12" sizes as predicate device OEC 9900 Elite (K082781). The ZEN-7000 is substantially equivalent to the predicate devices, ZEN-7000(K103425) and OEC 9900 Elite (K082781).

    1. Safety and Effectiveness, comparison to Predicate
      ZEN-7000 complies with industry standards such as IEC 60601-1 Series and 21 CFR 1020.30 and 21 CFR 1020.31 to minimize electrical, mechanical and radiation hazards. The new device modified from the predicate device was completed in accordance with GENORAY quality management system design controls.

{6}------------------------------------------------

  • Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-43, and IEC 60601-2-54 were performed.
  • EMC testing was conducted in accordance with standard IEC 60601-1-2.
  • FDA Guidance "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices" was performed.
  • ZEN-7000 meets the EPRC standards (21 CFR 1020.30, 31).
  • ZEN-7000 also meets the provisions of NEMA PS 3.1-3.18, Digital Imaging and Communications in Medicine (DICOM) Set.

All test results were satisfactory and the result of bench and clinical evaluation indicates that the ZEN-7000 is as safe and effective as the predicate devices.

    1. Conclusion
      ZEN-7000 has the same indication for use as the predicate devices. And there have been no changes implemented in the modifications to the predicate ZEN-7000 and OEC 9900 Elite that impact either the fundamental technology or the indications for use. The New device outlined in this Premarket Notification is substantially equivalent to the currently commercially available predicate devices.

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.