K Number
K140020
Date Cleared
2014-10-28

(298 days)

Product Code
Regulation Number
870.2340
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BSP HyperO™ Rest System is intended to be used during rest ECG tests, using the analysis of high frequency components present within the central portion of the QRS complex (HF-QRS). The device outputs a reduced amplitude zone (RAZ) value for the physician to interpret (based on reported clinical study results). The device is indicated for use in the ECG interpretation of patients with acute chest pain presenting to the Emergency Department, aged 18 years and above, as an additional adjunctive aid during initial patient presentation. The significance of the HF-QRS changes and interpretation of the HyperQTM Rest analysis must be determined by a physician, based on the reported clinical study results presented in the device labeling.

Device Description

The HyperQTM Rest System measures, processes, stores, and displays information derived from an electrocardiogram (ECG) with a high sampling rate (1000Hz). The device analyzes and records the High Frequency components of the QRS complex of standard ECG (HFQRS). The device displays the HFQRS to the user and also summarizes the analysis to generate an indication whether observed changes are within normal HyperO range, or constitute significant HyperO changes (Reduced Amplitude Zones, a.k.a RAZ); or represent a non-diagnostic test (unreadable). These findings are used as an aid in the clinical evaluation of acute chest pain on initial presentation to the Emergency Department.

AI/ML Overview

Here's a summary of the acceptance criteria and the study details for the HyperQ Rest System, based on the provided FDA submission document:

1. Table of Acceptance Criteria and Reported Device Performance

The FDA document focuses on demonstrating substantial equivalence to predicate devices rather than specific quantitative performance metrics as "acceptance criteria" in the way one might see for a diagnostic AI. The primary acceptance criterion appears to be the ability to identify Reduced Amplitude Zones (RAZ) in HF-QRS signals as an aid in diagnosing acute chest pain, with performance comparable to its predicate (HyperQ™ AD-100) and suitable for rest ECG analysis.

Acceptance Criteria (Implied)Reported Device Performance
Primary Indication: Identify Reduced Amplitude Zones (RAZ) in HF-QRS signals for patients with acute chest pain (as an adjunctive aid for physician interpretation).Effectiveness demonstrated in a clinical study: "Results demonstrated that the device can be used as an aid in the identification of Reduced Amplitude Zones (RAZ) values in patients presenting with acute chest pain to an Emergency Department, aged 18 years and above." The device delivers an "automatic interpretation indicating significant HyperQ changes or changes within normal range for all tests that were readable." This implies the device successfully outputs RAZ values that align with a gold standard, though specific metrics (e.g., sensitivity, specificity, AUC) are not provided in this summary.
Safety and Efficacy Equivalence to Predicate Device: HyperQ™ AD-100 system (K102579) and GE MAC 800 Resting ECG Analysis System (K090212).Substantially Equivalent: "The indications for use and technological characteristics of the BSP HyperQ™ Rest device are substantially equivalent to the indications for use and technological characteristics of the Predicate GE MAC 800 Resting ECG Analysis System and the HyperQ AD-100 System." "The new BSP HyperQTM Rest System is identical in its hardware to the HyperQ™ AD-100 System." "Software algorithm was modified, software validation testing was performed. The results... demonstrated that the device meets its specifications, which are similar to those of the predicate device. These performance tests demonstrated that the minor differences in the device specifications meet the system requirements and do not raise new safety or effectiveness concerns."

2. Sample Size for the Test Set and Data Provenance

  • Sample Size: The document does not explicitly state the sample size for the test set used in the clinical study. It only mentions that "Demographic data, ECG recordings and gold-standard diagnostics were used as a source for validating the diagnostic accuracy of the system."
  • Data Provenance: Not specified. It's unclear if the data was retrospective or prospective, or the country of origin.

3. Number of Experts and Qualifications

  • Number of Experts: Not specified.
  • Qualifications of Experts: Not specified. The document only mentions that the "gold standard diagnosis" was obtained from a "database."

4. Adjudication Method

  • Adjudication Method: Not specified. The document states the device's analysis was "compared to a gold standard diagnosis from the database," but it does not detail how that gold standard was established in terms of expert review or consensus.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? No, an MRMC comparative effectiveness study was not explicitly described or noted in the provided documents. The study described focuses on the device's standalone performance compared to a "gold standard diagnosis."
  • Effect size of human readers with AI vs. without AI assistance: Not applicable, as an MRMC study was not conducted or reported.

6. Standalone Performance

  • Was standalone (algorithm-only) performance done? Yes, the described clinical study assessed the device's standalone performance. "High frequency ECG signals were analyzed to obtain an automatic interpretation... The analysis was based solely on the digital data from the ECG device and was compared to a gold standard diagnosis from the database." The device "outputs a reduced amplitude zone (RAZ) value for the physician to interpret."

7. Type of Ground Truth Used

  • Type of Ground Truth: The ground truth was referred to as "gold standard diagnosis from the database." The specific nature of this gold standard (e.g., expert consensus of ECGs, confirmed pathology, clinical outcomes, or a combination) is not explicitly detailed. Given the indication (acute chest pain), the gold standard likely reflects a clinical diagnosis, potentially supported by further diagnostics or outcomes, but this is an inference.

8. Sample Size for the Training Set

  • Sample Size: This information is not provided. The document describes a clinical study to demonstrate effectiveness but does not delve into the development or training of the algorithm itself, nor the data used for it.

9. How Ground Truth for the Training Set Was Established

  • How Ground Truth Was Established: This information is not provided. The document focuses on the validation study and its ground truth, not the ground truth used during the algorithm's development or training phase.

{0}------------------------------------------------

Public Health Service

WO66-G609

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -

Silver Spring, MD 20993-0002

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.

October 28, 2014

Biological Signal Processing, Ltd. % Ahava Stein A. Stein Regulatory Affairs Consulting Ltd. 20 Hata' As Str., Suite 102 Kfar Saba, 44425 Israel

Re: K140020 Trade/Device Name: HyperQ Rest System Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph; Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms) Regulatory Class: Class II Product Code: DPS, MWI Dated: August 3, 2014 Received: August 8, 2014

Dear Ahava Stein:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Melissa A. Torres -S

For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K140020

Device Name HyperQ Rest

Indications for Use (Describe)

The BSP HyperO™ Rest System is intended to be used during rest ECG tests, using the analysis of high frequency components present within the central portion of the QRS complex (HF-QRS). The device outputs a reduced amplitude zone (RAZ) value for the physician to interpret (based on reported clinical study results). The device is indicated for use in the ECG interpretation of patients with acute chest pain presenting to the Emergency Department, aged 18 years and above, as an additional adjunctive aid during initial patient presentation. The significance of the HF-QRS changes and interpretation of the HyperQTM Rest analysis must be determined by a physician, based on the reported clinical study results presented in the device labeling.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY BSP HYPERQ™ ACS-100 SYSTEM

510(k) Number K140020

Applicant Name:

Company Name:Biological Signal Processing Ltd.
Address:22a Raoul Wallenberg Street
Tel Aviv 69719
Israel
Tel: +972-3-647-4840
Fax: +972-3-647-1498
E-mail: yair@bsp.co.il

Contact Person:

Official Correspondent:Ahava Stein
Name:A. Stein - Regulatory Affairs Consulting Ltd.
Address:20 Hata'as Str., Suite 102Kfar Saba 44425Israel
Tel: +972-9-7670002
Fax: +972-9-7668534
E-mail: ahava@asteinrac.com
Date Prepared:October 12, 2014
Trade Name:BSP HyperQTM Rest System
Classification Name:CFR Classification section 870.2340; (Product code DPS)
Classification:Class II Medical Device

Predicate Device:

The BSP HyperQ™ Rest System is substantially equivalent to the following predicate devices.

ManufacturerDevice510(k) No.
GE Medical SystemsInformation TechnologiesMAC 800 Resting ECG Analysis SystemK090212
BSP Ltd.HyperQ™ AD-100 SystemK102579

{4}------------------------------------------------

Device Description:

The HyperQTM Rest System measures, processes, stores, and displays information derived from an electrocardiogram (ECG) with a high sampling rate (1000Hz). The device analyzes and records the High Frequency components of the QRS complex of standard ECG (HFQRS). The device displays the HFQRS to the user and also summarizes the analysis to generate an indication whether observed changes are within normal HyperO range, or constitute significant HyperO changes (Reduced Amplitude Zones, a.k.a RAZ); or represent a non-diagnostic test (unreadable). These findings are used as an aid in the clinical evaluation of acute chest pain on initial presentation to the Emergency Department.

Intended Use/Indication for Use:

The BSP HyperQTM Rest System is intended to be used during rest ECG tests, using the analysis of high frequency components present within the central portion of the QRS complex (HF-QRS). The device outputs a reduced amplitude zone (RAZ) value for the physician to interpret (based on reported clinical study results). The device is indicated for use in the ECG interpretation of patients with acute chest pain presenting to the Emergency Department, aged 18 years and above, as an additional adjunctive aid during initial patient presentation. The significance of the HF-QRS changes and interpretation of the HyperQTM Rest analysis must be determined by a physician, based on the reported clinical study results presented in the device labeling.

Performance Standards:

The BSP HyperQTM Rest System is identical in hardware to the BSP HyperQTM AD-100, which was cleared under K102579. The Rest therefore maintains compliance with the recognized consensus standards previously described for the AD-100 device.

Non-Clinical (Bench) Performance Data:

Not applicable.

Pre-Clinical (Animal Study) Performance Data:

Not Applicable

{5}------------------------------------------------

Clinical Performance Data:

The effectiveness of the BSP HyperQTM Rest system was demonstrated in a clinical study. High frequency ECG signals were analyzed to obtain an automatic interpretation indicating significant HyperQ changes or changes within normal range for all tests that were readable. The analysis was based solely on the digital data from the ECG device and was compared to a gold standard diagnosis from the database.

Demographic data, ECG recordings and gold-standard diagnostics were used as a source for validating the diagnostic accuracy of the system. Results demonstrated that the device can be used as an aid in the identification of Reduced Amplitude Zones (RAZ) values in patients presenting with acute chest pain to an Emergency Department, aged 18 years and above.

Technological Characteristics:

The technology of the BSP devices involves analysis of High Frequency QRS (HFQRS) components of the ECG signal (HFQRS), made possible by the high frequency sampling rate. The high frequency range is represented by the HyperQ™ signal. The morphology of the HyperO envelope is examined for existence of reduced amplitude zones (RAZ) and High Frequency Morphological Indexes (HFMI) are extracted. The level of RAZ of all 12 leads is quantified by the HFMI, providing the user with a quantitative value.

Substantial Equivalence:

The indications for use and technological characteristics of the BSP HyperQ™ Rest device are substantially equivalent to the indications for use and technological characteristics of the Predicate GE MAC 800 Resting ECG Analysis System and the HyperQ AD-100 System.

The primary difference between the HyperQ AD-100 and the HyperQ Rest is in the indication for use for rest ECG analysis from the acquired ECG. The predicate device GE MAC 800 Resting ECG Analysis System (K090212) is specifically indicated for

K140020

{6}------------------------------------------------

rest ECG Testing. Based upon the validation results, BSP believes this change does not raise additional safety or efficacy concerns.

The design and components in the BSP HyperQ™ Rest system, including the hardware and software, are similar to the design and components found in the predicate HyperQ™ AD-100 System. The algorithm was modified to allow for analysis of rest ECG signals. The minor differences in the software user interface are insignificant and do not influence the safety or efficacy of the device. The performance specifications (including display of High Frequency QRS signals, and morphological assessment of them) of the BSP HyperQ™ Rest system are substantially equivalent to those in the HyperQ™ AD-100 System. There are no patient contact materials. Any minor differences in the technological characteristics do not raise new safety or effectiveness concerns. Furthermore, the new BSP HyperQTM Rest System is identical in its hardware to the HyperQ™ AD-100 System. The AD-100 system underwent performance testing, including electrical and mechanical safety testing according to IEC 60601-1 and electromagnetic compatibility testing according to IEC 60601-1-2. These were provided in the original 510(k) submission of the HyperQ™ AD-100 System, and are not repeated in this submission. As the software algorithm was modified, software validation testing was performed. The results of the software validation tests demonstrated that the device meets its specifications, which are similar to those of the predicate device. These performance tests demonstrated that the minor differences in the device specifications meet the system requirements and do not raise new safety or effectiveness concerns. Furthermore, the results of the clinical study performed demonstrate that the device provides an effective tool as an aid in the diagnostic process of acute chest pain in patients aged 18 years and above, for the identification of RAZ in the ECG signal.

Conclusions:

Based on the performance testing and comparison to predicate devices, the BSP HyperQTM Rest System is substantially equivalent to the predicate devices.

§ 870.2340 Electrocardiograph.

(a)
Identification. An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.(b)
Classification. Class II (performance standards).