K Number
K140007

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-04-07

(95 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AbacusTM Spacer System is indicated for interverebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-L5. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radios. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The AbacusTM Spacer System is to be used with autogenous bone grafi and with supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral body fusion device.

Device Description

The Abacus™ Spacer System is an intervertebral body fusion device manufactured from PEEK-OPTIMA (ASTM F2026) and includes tantalum markers (ASTM 560/ISO 13782). The Abacus™ Spacer System is available in a variety of shapes and sizes to accommodate variations in anatomy. The Abacus™M Spacer System is a rectangular-shaped device with a textured "tooth" pattern on both the superior and inferior surfaces designed to resist migration of the device once it is surgically positioned. The device also incorporates an internal cavity that allows for the placement of autograft material.

AI/ML Overview

This document describes the Abacus™ Spacer System, an intervertebral body fusion device. The acceptance criteria and the study that proves the device meets the acceptance criteria are detailed below.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific TestAcceptance CriteriaReported Device Performance
Mechanical PerformanceStatic Axial Compression (per ASTM F2077)Must meet or exceed the static axial compression requirements for intervertebral body fusion devices as defined in ASTM F2077.Performed and reported as demonstrating substantial equivalence to predicate.
Mechanical PerformanceDynamic Axial Compression (per ASTM F2077)Must meet or exceed the dynamic axial compression requirements for intervertebral body fusion devices as defined in ASTM F2077.Performed and reported as demonstrating substantial equivalence to predicate.
Mechanical PerformanceStatic Compression Shear (per ASTM F2077)Must meet or exceed the static compression shear requirements for intervertebral body fusion devices as defined in ASTM F2077.Performed and reported as demonstrating substantial equivalence to predicate.
Mechanical PerformanceDynamic Compression Shear (per ASTM F2077)Must meet or exceed the dynamic compression shear requirements for intervertebral body fusion devices as defined in ASTM F2077.Performed and reported as demonstrating substantial equivalence to predicate.
Mechanical PerformanceSubsidence (per ASTM F2267)Must meet or exceed the subsidence requirements for intervertebral body fusion devices as defined in ASTM F2267.Performed and reported as demonstrating substantial equivalence to predicate.
BiocompatibilityMaterial (PEEK-OPTIMA, ASTM F2026; Tantalum markers, ASTM 560/ISO 13782)Biocompatibility of materials must be demonstrated for implantable devices.Manufactured from PEEK-OPTIMA (ASTM F2026) and includes tantalum markers (ASTM 560/ISO 13782), which are widely accepted biocompatible materials for such applications. No specific biocompatibility testing reported in this summary, but reliance on material standards.
Substantial EquivalenceIntended Use, Indications for Use, Materials, PerformanceMust demonstrate substantial equivalence to legally marketed predicate devices.The submission concludes "substantial equivalence...in intended use, indications for use, materials and performance" to the StaXx® IBL System and CoRoent® System.

2. Sample Size Used for the Test Set and the Data Provenance

The provided document does not specify a "test set" in the context of clinical or image-based data. The performance data presented refers to mechanical testing of the device.

  • Sample Size for Mechanical Testing: The document does not explicitly state the number of samples used for each mechanical test (Static and dynamic axial compression, Static and dynamic compression shear, Subsidence). Mechanical testing typically involves multiple samples to establish statistical validity, but the exact numbers are not present here.
  • Data Provenance: The mechanical tests were performed by Spine Wave, Inc. in support of the 510(k) submission. These are laboratory tests, not clinical data from patients or countries.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. This submission concerns a physical medical device (intervertebral spacer) and its mechanical performance, not a diagnostic algorithm that requires expert-established ground truth from a test set of images or clinical cases. The "ground truth" for the mechanical tests is derived from the established testing standards (ASTM F2077, ASTM F2267).

4. Adjudication Method for the Test Set

Not applicable, as there is no "test set" in the context of clinical or image-based data requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a submission for an intervertebral body fusion device, not an AI-powered diagnostic tool. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was performed.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This device is a physical implant, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the performance data in this submission is based on established mechanical testing standards (ASTM F2077 and ASTM F2267). The device's performance is compared against the requirements and specifications outlined in these industry standards to demonstrate mechanical suitability and substantial equivalence.

8. The Sample Size for the Training Set

Not applicable. There is no "training set" as this is a physical medical device, not a machine learning model.

9. How the Ground Truth for the Training Set was Established

Not applicable. There is no "training set."

{0}------------------------------------------------

K140007 Page 1 of 2

APR 0 7 2014

Image /page/0/Picture/2 description: The image shows the word "SpineWave" in a bold, sans-serif font. The word is all capitalized except for the "W", which is larger than the other letters. A curved line runs underneath the word, starting below the "E" in "Spine" and ending below the "E" in "Wave".

510(k) Summary Abacus™ Spacer System

1. Submitter Information

Submitter:Spine Wave, Inc.
Address:Three Enterprise DriveSuite 210Shelton, CT 06484
Telephone:203-712-1839
Telefax:203-944-9493
Contact:Joseph Mercado
Date Prepared:March 25, 2014

2. Device Information

Trade Name:Abacus™ Spacer System
Common Name:Intervertebral Body Fusion Device
Classification:Class II (special controls) per 21 CFR 888.3080
Classification Name:Intervertebral Fusion Device with Bone Graft, Lumbar
Product Code:MAX

3. Purpose of Submission

The purpose of this submission is to gain clearance for a new intervertebral body fusion device.

4. Predicate Device Information

The Abacus™ Spacer System described in this submission is substantially equivalent to the following predicates:

Predicate DeviceManufacturer510(k) No.
StaXx® IBL SystemSpine Wave, Inc.K131071, K132719
CoRoent® SystemNuVasive, Inc.K071795

{1}------------------------------------------------

5. Device Description

The Abacus™ Spacer System is an intervertebral body fusion device manufactured from PEEK-OPTIMA (ASTM F2026) and includes tantalum markers (ASTM 560/ISO 13782). The Abacus™ Spacer System is available in a variety of shapes and sizes to accommodate variations in anatomy. The Abacus™M Spacer System is a rectangular-shaped device with a textured "tooth" pattern on both the superior and inferior surfaces designed to resist migration of the device once it is surgically positioned. The device also incorporates an internal cavity that allows for the placement of autograft material.

6. Intended Use

The Abacus "10 Spacer System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-L5. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Abacus™ Spacer System is to be used with autogenous bone graft and with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral body fusion device.

7. Comparison of Technological Characteristics

The substantial equivalence of the Abacus™ Spacer System to the predicates is shown by similarity in intended use, indications for use, materials and performance.

8. Performance Data

The following mechanical tests were performed to demonstrate the substantial equivalence of the Abacus The Spacer System to its predicate:

  • Static and dynamic axial compression (per ASTM F2077) .
  • Static and dynamic compression shear (per ASTM F2077) .
  • Subsidence (per ASTM F2267)

9. Conclusion

Based on the indications for use, technological characteristics, performance testing and comparison to the predicates, the Abacus™ Spacer System has been shown to be substantially equivalent to the predicate devices identified in this submission, and does not present any new issues of safety or effectiveness.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tails, representing the three levels of government: federal, state, and local. The eagle is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 7, 2014

Spine Wave, Incorporated Mr. Joseph Mercado Regulatory Affairs Specialist Three Enterprise Drive, Suite 210 Shelton, Connecticut 06484

Re: K140007

Trade/Device Name: Abacus™ Spacer System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: January 16, 2014 Received: January 17, 2014

Dear Mr. Mercado:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Joseph Mercado

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note

the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Ronald: P.Jean -S for

  • Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K140007

Device Name AbacusTM Spacer System

Indications for Use (Describe)

The AbacusTM Spacer System is indicated for interverebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-L5. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radios. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The AbacusTM Spacer System is to be used with autogenous bone grafi and with supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral body fusion device.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page,

N/A