K Number
K133952

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-08-19

(239 days)

Product Code
Regulation Number
866.1640
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VITEK® 2 Yeast Flucytosine is designed for antimicrobial susceptibility testing of clinically significant yeast species and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Yeast Flucytosine is a quantitative test. Flucytosine has been shown to be active against most isolates of the microorganisms listed below, according to the FDA label for this antifungal.

Active in vitro and in clinical infections Candida albicans

The following in vitro data are available, but their clinical significance is unknown. Candida dubliniensis Candida glabrata Candida guilliermondii Candida lusitaniae Candida parapsilosis Candida tropicalis

The VITEK ® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococus spp., Enterococcus spp. and clinically significant yeast.

Device Description

The VITEK® 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology. The isolate to be tested is diluted to a standardized concentration in 0.45 - 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling and sealing operation. The VITEK® 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.

AI/ML Overview

The VITEK® 2 AST-YS Flucytosine device is designed for antimicrobial susceptibility testing of clinically significant yeast species (Candida species) and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems. It provides a quantitative test to determine in vitro susceptibility to antimicrobial agents.

1. A table of acceptance criteria and the reported device performance

Performance MetricAcceptance Criteria (Implicit from Guidance)Reported Device Performance
Essential Agreement (EA)Not explicitly stated, but typically ≥ 90%98.8% Overall
Category Agreement (CA)Not explicitly stated, but typically ≥ 90%98.5% Overall
Reproducibility"acceptable results""acceptable results"
Quality Control"acceptable results""acceptable results"

Note on Acceptance Criteria: The document refers to the "FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA. Issued August 28, 2009." This guidance document would contain the specific performance acceptance criteria for Essential Agreement and Category Agreement. Without direct access to this document, the reported performance metrics (98.8% EA and 98.5% CA) are generally considered excellent and likely meet or exceed standard acceptance thresholds for AST devices.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: Not explicitly stated as a single number. The study used "fresh and stock clinical isolates, as well as a set of challenge strains." The exact number of isolates for the test set is not provided in this summary.
  • Data Provenance: Not specified regarding country of origin. The data includes "clinical isolates," implying real-world samples, and "challenge strains," which are likely laboratory-controlled strains used to test performance across the range of expected behaviors. The evaluation was an "external evaluation," suggesting independent testing beyond the manufacturer's internal labs. The study refers to "fresh and stock clinical isolates," which implies a prospective (fresh isolates) and retrospective (stock isolates) component.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. The ground truth for this device is established by a reference laboratory method, not expert human interpretation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Ground truth is established by a reference method, not a panel of human adjudicators.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is an automated antimicrobial susceptibility test system, not an AI-assisted diagnostic device requiring human reader improvement studies.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Yes, a standalone study was conducted. The VITEK® 2 AST-YS Flucytosine device's performance was compared directly against the CLSI broth microdilution reference method (the gold standard). The reported Essential Agreement and Category Agreement percentages directly reflect the device's standalone performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth was established by the CLSI broth microdilution reference method. This is a widely accepted, standardized laboratory procedure for determining the minimum inhibitory concentration (MIC) of an antimicrobial agent against a microorganism. The incubation period for this reference method was 48 hours.

8. The sample size for the training set

The document does not explicitly mention a "training set" or its size in the context of device development. Given that this is an AST system, the device's "training" would primarily involve the development and calibration of the photometric algorithms that detect growth and determine MIC values based on known responses of various microorganisms to different antimicrobial concentrations. This development process isn't typically referred to as a "training set" in the same way an AI model is trained.

9. How the ground truth for the training set was established

Not explicitly detailed for a "training set" as described above. The general principle would involve establishing ground truth via the CLSI broth microdilution reference method for a diverse set of isolates to calibrate the VITEK® 2 system's algorithms for accurate MIC determination.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 19, 2014

BIOMERIEUX, INC. JOCELYN JENNINGS, M.S., R.A.C. SENIOR MANAGER, REGULATORY AFFAIRS 100 RODOLPHE ST DURHAM NC 27712

Re: K133952

Trade/Device Name: VITEK 2 AST - Yeast Flucytosine Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial susceptibility test powder Regulatory Class: II Product Code: NGZ Dated: August 5, 2014 Received: August 6, 2014

Dear Ms. Jennings:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Sally A. Hojvat -S

Sally A. Hojvat. M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K133952

Device Name VITEK® 2 AST-YS Flucytosine ( ≤ 1 -> 64 ug/mL )

Indications for Use (Describe)

VITEK® 2 Yeast Flucytosine is designed for antimicrobial susceptibility testing of clinically significant yeast species and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Yeast Flucytosine is a quantitative test. Flucytosine has been shown to be active against most isolates of the microorganisms listed below, according to the FDA label for this antifungal.

Active in vitro and in clinical infections Candida albicans

The following in vitro data are available, but their clinical significance is unknown. Candida dubliniensis Candida glabrata Candida guilliermondii Candida lusitaniae Candida parapsilosis Candida tropicalis

The VITEK ® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococus spp., Enterococcus spp. and clinically significant yeast.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Sally A. Hojvat -S 2014.08.19 15:09:22 -04

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for bioMérieux. The logo features the company name in a sans-serif font, with the "BIO" portion in a slightly larger font size. Above the name is a stylized graphic of a sphere with curved lines and a curved line going through the sphere.

510(k) SUMMARY

VITEK® 2 AST-YS Flucytosine

510(k) Submission Information:

Submitter's Name:bioMérieux, Inc.
Address:100 Rodolphe StreetDurham, NC 27712
Contact Person:Jocelyn JenningsSenior Manager, Regulatory Affairs
Phone Number:919-620-2894
Fax Number:919-620-2548
Date of Preparation:December 19, 2013
B. Device Name:
Formal/Trade Name:VITEK® 2 AST-YS Flucytosine
Classification Name:21 CFR 866.1640Antimicrobial Susceptibility TestProduct Code: NGZ
Common Name:VITEK® 2 AST-YS Flucytosine
C. Predicate Device:VITEK® 2 AST-YS Voriconazole (K092454)

D. 510(k) Summary:

VITEK® 2 Yeast Flucytosine is designed for antimicrobial susceptibility testing of Candida species and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Yeast Flucytosine is a quantitative test. Flucytosine has been shown to be active against most isolates of the microorganisms listed below, according to the FDA label for this antifungal.

Active in vitro and in clinical infections Candida albicans

The following in vitro data are available, but their clinical significance is unknown. Candida dubliniensis Candida qlabrata Candida quilliermondii Candida Iusitaniae Candida parapsilosis Candida tropicalis

The antimicrobial presented in VITEK® 2 AST-YS Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/ml. The VITEK® 2 AST Cards are essentially miniaturized

bioMérieux, Inc.

{5}------------------------------------------------

2

versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.

The isolate to be tested is diluted to a standardized concentration in 0.45 - 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling and sealing operation. The VITEK® 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.

VITEK® 2 AST-YS Flucytosine demonstrated substantially equivalent performance when compared with the CLSI broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA. Issued August 28, 2009.

The 510(k) presents data in support of VITEK® 2 AST-YS Flucytosine. An external evaluation was conducted with fresh and stock clinical isolates, as well as a set of challenge strains. The external evaluations were designed to confirm the acceptability of VITEK® 2 AST-YS Flucytosine by comparing its performance with the CLSI broth microdilution reference method incubated at 48 hrs. The data is representative of performance on both the VITEK® 2 and VITEK® 2 Compact instrument platforms. VITEK® 2 AST-YS Flucitosine demonstrated acceptable performance of 98.8% overall Essential Agreement and 98.5% overall Category Agreement for Candida species. Reproducibility and Quality Control demonstrated acceptable results.

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).