K Number
K133816

Validate with FDA (Live)

Date Cleared
2014-06-24

(190 days)

Product Code
Regulation Number
862.1660
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IMMULITE® Free PSA Calibration Material (CVM) is intended for monitoring system performance of the IMMULITE Immunoassay system for the quantitative measurement of Free PSA antigen.

Device Description

IMMULITE® 2000 Free PSA Calibration Verification Material (CVM) contains one set of four vials (LPFCVM1-4), 2 mL each. LPFCVM1 contains bovine protein/buffer matrix with preservatives. LPFCVM2, LPFCVM3 and LPFCVM4 contain low, intermediate and high levels of Free PSA respectively, in human-source matrix with preservatives. The CVMs are supplied lyophilized form.

AI/ML Overview

This document describes the IMMULITE® 2000 Free PSA Calibration Verification Material (CVM) and the performance testing conducted to demonstrate its substantial equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

The device's performance is demonstrated through its stability testing. The acceptance criteria for stability state that the dose concentration for the stability CVM must be within ±10% of the assigned dose concentration. The study reports that acceptable results were observed, confirming it meets these criteria.

CVM LevelAssigned Dose (ng free PSA/mL)Acceptance Criteria (% difference to assigned dose)Acceptable Range (ng free PSA/mL)Reported Performance (meets criteria?)
LPFCVM10.00N/AN/AN/A (not applicable for 0.00 dose)
LPFCVM20.20+10%0.18 - 0.22Yes (reported as "acceptable results")
LPFCVM31.45+10%1.31 - 1.60Yes (reported as "acceptable results")
LPFCVM424.40+10%21.96 - 26.84Yes (reported as "acceptable results")

2. Sample Size Used for the Test Set and Data Provenance

The "test set" in this context refers to the samples used for the stability study and value assignment, as this device itself is a calibration verification material and not a diagnostic test with a patient-based test set.

  • Stability Study (test set for CVM stability):

    • Sample Size: "CVM lot 15," "lots 90 and 91." Specific numbers of vials/samples are not provided, but it states CVMs are run "in duplicate (as a minimum)."
    • Data Provenance: Not specified, but implied to be from Siemens' internal testing. It's a prospective study for stability monitoring.
  • Value Assignment Validation (part of the device's performance characterization):

    • Sample Size:
      • 3 levels of commercially available controls.
      • 20 normal male patient samples.
      • 10 spiked samples.
      • 10 PSA patient samples.
    • Data Provenance: Not specified, but implied to be from Siemens' internal testing. Retrospective or prospective nature not explicitly stated, but typically these are prospective studies for product development.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. The ground truth for this device (a calibration verification material) is established through traceability to a recognized standard (WHO 15 NIBSC IS (96/668)) and internal gravimetric preparation of reference calibrators, not through expert consensus on medical images or clinical outcomes.

4. Adjudication Method for the Test Set

Not applicable. This is not a diagnostic device requiring adjudication of results from different readers or interpretations. The performance is assessed against quantitative analytical criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a calibration verification material for an immunoassay system, not an AI-assisted diagnostic tool or a device that involves human readers interpreting cases.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is a laboratory reagent, not an algorithm. Its "standalone performance" is its ability to maintain its assigned concentration over time, which is what the stability study assesses.

7. The Type of Ground Truth Used

The ground truth for the IMMULITE® 2000 Free PSA CVM is established through:

  • Traceability to Reference Standard: The CVMs are traceable to WHO 15 NIBSC IS (96/668).
  • Gravimetric Preparation: The assigned reference calibrators, from which the CVMs are derived, are prepared using Free PSA antigen stock and traceable to internal material that has been gravimetrically prepared.
  • Value Assignment: CVM values are determined by running the CVMs in replicates on multiple instruments using reference internal calibration curves.

8. The Sample Size for the Training Set

Similar to "test set" described above, in the context of calibration verification materials, the concept of a "training set" doesn't directly apply as it would for an AI algorithm. However, for the initial value assignment of the CVM levels, which could be considered analogous to establishing "reference" data:

  • CVM Level Testing: The Free PSA CVM levels were tested on 18 replicates in total, comprised of 6 runs and 3 replicates per run, on 4 IMMULITE 2000 system analyzers and 2 different reagent kit lots.

9. How the Ground Truth for the Training Set Was Established

The "ground truth" (the assigned values and ranges for the CVMs) was established as follows:

  • Primary Traceability: Traceable to WHO 15 NIBSC IS (96/668).
  • Internal Reference: Using internal material which has been gravimetrically prepared.
  • Comprehensive Testing: The Free PSA CVMs are a subset of 11-level Free PSA calibrators. These calibrators and CVMs are value assigned using assigned reference calibrators. The CVM values are calculated based on recovered values from 18 replicates run on 4 different IMMULITE 2000 system analyzers with 2 different reagent kit lots. The values are then averaged across all systems.
  • Validation: The CVM value assignments were validated using 3 levels of commercially available controls, 20 normal male patient samples, 10 spiked samples, and 10 PSA patient samples.
  • Guideline Range: The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD) from the collected data.

{0}------------------------------------------------

SIEMENS

510(k) Summary

Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence.

The assigned 510(k) Number: K133816

1.SubmitterMailing Address:Siemens Healthcare Diagnostics Inc.511 Benedict AvenueTarrytown, NY 10591
Contact Person:Ernest JosephSenior Manager, Regulatory Affairs
Phone Number:(914)-524-2431
Fax Number:(914)-524-2101
E-mail Address:Ernest.joseph@siemens.com
Date Prepared:May 21, 2014
    1. Device Name Proprietary Name: Measurand: Type of Test:
      Regulation Section: Classification: Products Code:

Panel:

    1. Predicate Device Name Predicate 510(k) No:
    1. Device Description:
    1. Intended Use: Indication for Use:

Special Conditions for Use Statement(s): Special Instrument Requirements:

IMMULITE® 2000 Free PSA Calibration Verification Material (CVM) 21 CFR 862.1660, Quality Control Material Class I Reserved JJX - Single (Specified) Analyte Controls (Assayed and Unassayed) Clinical Chemistry (75)

IMMULITE® 2000 PSA Calibration Verification Material

K131536

IMMULITE® 2000 Free PSA Calibration Verification Material (CVM) contains one set of four vials (LPFCVM1-4), 2 mL each. LPFCVM1 contains bovine protein/buffer matrix with preservatives. LPFCVM2, LPFCVM3 and LPFCVM4 contain low, intermediate and high levels of Free PSA respectively, in human-source matrix with preservatives. The CVMs are supplied lyophilized form.

See Indications for Use Statement below The IMMULITE® Free PSA Calibration Verification Material (CVM) is intended for monitoring system performance of the IMMULITE Immunoassay system for the quantitative measurement of Free PSA antigen.

For prescription use only

IMMULITE® 2000 Systems

{1}------------------------------------------------

    1. Technological Characteristics and Substantial Equivalence Comparison with Predicate:
      A comparison of the device features, intended use, and other information demonstrates that the IMMULITE® 2000 Free PSA CVM is substantially equivalent to the predicate device, IMMULITE® 2000 PSA CVM, as summarized in Table 1.
SIMILARITIES
Candidate DevicePredicate Device
IMMULITE 2000 Free PSACVMIMMULITE 2000 PSA CVM
Intended UseThe IMMULITE® Free PSACalibration Verification Material(CVM) is intended for monitoringsystem performance of theIMMULITE Immunoassay systemfor the quantitative measurementof Free PSA antigenThe IMMULITE® 2000 PSA CalibrationVerification Material (CVM) is intendedfor monitoring system performance of theIMMULITE Immunoassay system for thequantitative measurement of PSA antigen
Levels44
StabilityStable unopened until theexpiration dateStable unopened until theexpiration date
Storage≤-20°C≤-20°C
UseSingle Use OnlySingle Use Only
Table 1: Substantial Equivalence Comparison
DIFFERENCES
Candidate DevicePredicate Device
IMMULITE 2000 Free PSA CVMIMMULITE 2000 PSA CVM
AnalyteFree PSAPSA
FormLyophilizedLiquid
MatrixBovine Serum AlbuminProcessed (pH-treated) Chicken Serum

7. Non-Clinical Performance Testing

Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the candidate device.

7.1 Stability Summary:

The stability study was conducted to validate the shelf life claim for the IMMULITE® 2000 Free PSA Calibration Verification Material (CVM) to ensure that it maintains optimal product performance on IMMULITE 2000 platform throughout the established shelf life of the CVM. The IMMULITE® 2000 Free PSA Calibration Verification Material (CVM) stability study performed with CVM lot 15 shows acceptable results up to 3.5 years when stored at -20°C, and

{2}------------------------------------------------

SIEMENS

for on-going stability for lots 90 and 91 shows acceptable results up to 6 months at this time when stored at -20°C. Therefore the current shelf life stability claim is 6 months.

7.1.1 Stability Protocol Summary:

The CVMs are run as part of the calibrator stability testing. The stability calibrators/CVMs are run in duplicate (as a minimum) and the concentration value is determined from the reference internal calibration curve. Stability calibrators/CVMs are run at 6 months then annually up to the month after expiry. The stability will be tested at 24, 36 and 42 months.

7.1.2 Stability Acceptance Criteria Summary:

The Acceptance Criteria for the IMMULITE Free PSA Calibration Verification Material (CVM) requires that the dose concentration for the stability CVM is within 10% of assigned dose concentration.

CVM LevelAssigned Dose (ngfree PSA/mL)Acceptance Criteria(% difference toassigned dose)Acceptable Range(ng free PSA/mL)
LPFCVM10.00N/AN/A
LPFCVM20.20+100.18 - 0.22
LPFCVM31.45+101.31 - 1.60
LPFCVM424.40+1021.96 - 26.84

Table 2: Stability Acceptance Criteria for IMMULITE 2000 Free PSA CVM

7.2 Traceability:

The IMMULITE Free PSA CVMs are traceable to WHO 15 NIBSC IS (96/668). The CVMs are manufactured using qualified materials and measurement procedures.

7.3 Value Assignment, Expected Values and Reference Range:

Free PSA CVMs are 4 level materials which are subset of 11 level Free PSA calibrators. Calibrators are not commercialized but are used internally during manufacture and release testing of Free PSA reagents and two point adjustors. The IMMULITE calibrators and therefore CVMs are value assigned using assigned reference calibrators. The assigned reference calibrators are prepared using Free PSA antigen stock and are traceable to internal material which has been gravimetrically prepared. Three levels of commercially available controls and 20 normal male patient samples, 10 spiked samples and 10 PSA patient samples were used to validate CVM value assignments.

The Free PSA CVM level were tested on 18 replicates in total comprised of 6 runs and 3 replicates per run on 4 IMMULITE 2000 system analyzers and 2 different reagent kit lots. The CVM values are calculated based on the recovered values for each run on each instrument independently. CVM values are then averages across all systems. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The expected assay range is 0.07 - 25.0 ng/mL. The controls must fall within their target ranges and the target specifications are shown in Table 4.

{3}------------------------------------------------

AnalyteTargetLevelsLevelREF andLot numberTarget Mean(ng/mL)SDGuideline ± 2SD Range(ng/mL)
CVM1LPFCVM1 0150.000.000.07
CVM2LPFCVM2 0150.200.01250.170.22
CVM3LPFCVM3 0151.450.081.291.61
CVM4LPFCVM4 01524.401.222.026.8*
Assay Range0.07 - 25.0 ng/mL

Table 4: Analyte Target Range Levels

  • The upper limit for this CVM exceeds the assay range of 25.0 ng/mL. The range of the CVM is determined from the precision %CV around the target. The upper limit of this CVM is within +10% of the assay range which ensures the whole of the assay range is covered.

Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, total precision tabulated in the respective assay instructions for use may be considered as one factor when establishing local, acceptable ranges. The values provided above may be considered as guidelines. Value assignment is lot specific.

Standard/Guidance Documents Referenced:

  • CEN 13640 Stability Testing of In Vitro Diagnostic Reagents .
  • Guidance for Industry Abbreviated 510(k) Submissions for In Vitro Diagnostic ◆ Calibrators
  • Guidance for Industry and FDA Staff Assayed and Unassayed Quality Control � Material

Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10

8. Conclusion:

The IMMULITE® 2000 Free PSA Calibration Verification Material is substantially equivalent to other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent to the currently marketed, IMMULITE® 2000 PSA Calibration Verification Material. The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence, product safety and effectiveness. Based on the testing completed and the comparisons with predicate device, The IMMULITE® 2000 Free PSA Calibration Verification Material does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services (USA). The logo features a stylized depiction of an eagle or bird-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" is arranged in a circular pattern around the symbol. The logo is in black and white.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-Gritt Silver Spring, MD 20993-0002

June 24, 2014

SIEMENS HEALTHCARE DIAGNOSTICS INC. MR. ERNEST JOSEPH SENIOR MANAGER, REGULATORY AFFAIRS 511 BENEDICT AVENUE TARRYTOWN. NY 10591

Re: K133816

Trade/Device Name: IMMULITE 2000 Free PSA Calibration Verification Material (CVM) Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: I Product Code: JJX Dated: June 16. 2016 Received: June 17, 2014

Dear Mr. Joseph:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 331-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2-Mr. Joseph

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 80) and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fdaggov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.goy/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Elizabeth A. Stafford -S

for Maria M. Chan, Ph.D.

Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known) K133816

Device Name

IMMULITE® 2000 Free PSA Calibration Verification Material

Indications for Use (Describe)

The IMMULITE® Free PSA Calibration Material (CVM) is intended for monitoring system performance of the IMMULITE Immunoassay system for the quantitative measurement of Free PSA antigen.

Type of Use (Select one or both, as applicable)

[X] Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 807 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Elizabeth A. Stafford -S

FORM FDA 3881 (6/13)

PSC Publishing Services 13011 413-6740 EF

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.