(46 days)
The Pioneer Sternal Cable Plate System is intended for use in the stabilization and fixation of fractures of the anterior chest wall including Sternal fixation following Sternotomy and Sternal reconstructive surgical procedures.
The Pioneer Sternal Cable Plate System contains various configurations of plates, some with integrated cables and crimps, and Ø2.7mm and 3.0mm self-drilling screws (lengths 8-20mm) to allow for multiple anterior chest wall and sternal repair and reconstruction options. The implants are manufactured from medical grade ASTM F67 Grade IV commercially pure titanium and ASTM F136 Titanium alloy. The implants may be implanted via an open or minimally invasive approach using Class I (exempt) orthopedic manual surgical instruments. The purpose of this submission is to add additional implant configurations to the system.
The acceptance criteria and study proving the device meets them are described below:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Mechanical Strength (Static and Dynamic Tensile Tests) | Pioneer Sternal Cable Plate System performs in a manner substantially equivalent to predicate systems. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the specific sample sizes used for the static and dynamic tensile tests. It refers to "Static and Dynamic Tensile Tests" as the pre-clinical performance data. Since these are bench tests for mechanical strength, the "test set" would refer to the physical specimens of the device that underwent testing.
The data provenance is not explicitly stated in terms of country of origin. The study appears to be entirely pre-clinical bench testing.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
This information is not applicable as the study described is a pre-clinical bench test of mechanical properties, not a clinical study involving expert evaluation or ground truth based on medical diagnoses.
4. Adjudication Method for the Test Set
This information is not applicable as the study described is a pre-clinical bench test and does not involve human readers or adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Human Readers' Improvement with AI vs. Without AI Assistance
This information is not applicable as the study described is a pre-clinical bench test of a medical device (sternal cable plate system), not an AI-assisted diagnostic device.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
This information is not applicable as the device is a physical medical implant, not a software algorithm.
7. The Type of Ground Truth Used
The "ground truth" for this pre-clinical study effectively comes from established engineering standards and requirements for mechanical strength and durability for sternal fixation devices. The comparison is made against the performance of legally marketed predicate devices, implying that their performance serves as the benchmark or "ground truth" for acceptable mechanical properties.
8. The Sample Size for the Training Set
This information is not applicable as the study is a pre-clinical bench test of a physical medical device, not a machine learning model that requires a training set.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable for the same reasons as above.
{0}------------------------------------------------
| 510(k) Summary Pursuant to 21 CFR 807.92JAN : 2 | ||
|---|---|---|
| Sponsor: | Pioneer Surgical Technology, Inc.(RTI Surgical, Inc.)375 River Park CircleMarquette, MI 49855 USAContact: Emily Downs, Sarah McIntyrePh: (906) 225-5602Fx: (906) 226-4459Prepared: January 2, 2014 | |
| Name: | Pioneer Sternal Cable Plate System | |
| Trade names: | Tritium | |
| Common name: | Pioneer Sternal Cable Plate System | |
| Classifications: | §888.3010 Cerclage, Fixation, Metallic§888.3030 Plate. Fixation. Bone, Non-Spinal, Metallic§888.3040 Screw. Fixation. Bone. Non-Spinal, Metallic | |
| Product Codes: | JDQ, HRS, HWC | |
| Panel/ Branch: | Orthopaedic and Rehabilitation Devices Panel: Panel Code 87 | |
| Predicates: | Pioneer Sternal Cable System (K122293)Biomet SternaLock Blu Microfixation Sternal Closure System (K110574)Ethicon Wire (K946173) | |
| Description: | The Pioneer Sternal Cable Plate System contains various configurations ofplates, some with integrated cables and crimps, and Ø2.7mm and 3.0mmself-drilling screws (lengths 8-20mm) to allow for multiple anterior chestwall and sternal repair and reconstruction options. The implants aremanufactured from medical grade ASTM F67 Grade IV commercially puretitanium and ASTM F136 Titanium alloy. | |
| The implants may be implanted via an open or minimally invasiveapproach using Class I (exempt) orthopedic manual surgical instruments. | ||
| The purpose of this submission is to add additional implant configurationsto the system. |
it to 21 CFR 807.92 510/L) St Pursual
.
~ "
)14
{1}------------------------------------------------
- The Pioneer Sternal Cable Plate System is intended for use in the Intended Use: stabilization and fixation of fractures of the anterior chest wall including Sternal fixation following Sternotomy and Sternal reconstructive surgical procedures.
Static and Dynamic Tensile Tests were provided to support that the Pre-Clinical Pioneer Sternal Cable Plate System performs in a manner substantially Performance Data: equivalent to that of predicate systems. No new issues of safety or effectiveness were raised.
Substantial This submission supports the position that the subject system is Equivalence substantially equivalent to previously cleared systems. The subject and predicate systems are similar in terms of indications for use, material composition, sterilization, packaging, technological characteristics, design features, and mechanical strength.
There are no significant differences between the subject system and the predicates which would adversely affect the use of the product. Any differences were not considered significant based on mechanical bench testing.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002
January 27, 2014
Pioneer Surgical Technology, Incorporated (RTI Surgical, Incorporated) Ms. Emily Downs Director of Regulatory and Clinical Affairs 375 River Park Circle Marquette, Michigan 49855
Re: K133785
Trade/Device Name: Pioneer Sternal Cable Plate System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone fixation cerclage Regulatory Class: Class II Product Code: JDQ, HRS, HWC Date: January 2, 2014 Received: January 3, 2014
Dear Ms. Downs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Emily Downs
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Vincent Devlin -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K133785
Device Name
Pioneer Sternal Cable Plate System
Indications for Use (Describe)
The Pioneer Sternal Cable Plate System is intended for use in the stabilization and fixation of fractures of the anterior chest wall including Sternal fixation following Sternotomy and Sternal reconstructive surgical procedures.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
| SC | Indilishing Services ( 101) 443-6740
EF
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Elizabet阿姆Frank -S
Division of Orthopedic Devices
FORM FDA 3881 (9/13)
See PRA Statement on last page.
Form Approved: OMB No. 0910-0120
Expiration Date: December 31, 2013
N/A