K Number
K133750

Validate with FDA (Live)

Date Cleared
2014-01-08

(30 days)

Product Code
Regulation Number
876.5130
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Occluder™ Occlusion Balloon Catheters are indicated for use for temporary Ureteral occlusion and applications including, renal opacification, dislodgment of calculi and preventative calculi migration.

Device Description

Occluder Occlusion Balloon Catheters are designed for use for temporary occlusion of the ureter and applications including renal opacification, dislodgement of calculi and preventative calculi migration. The devices are provided sterile and are intended for single use. The Occluder Occlusion Balloon Catheters are constructed of a soft compliant latex balloon mounted on the tip of a catheter shaft. Catheter shafts are radiopaque, maximizing fluoroscopic visibility. The Occluder Occlusion Balloon Catheters have two lumens that are marked and color-coded. The balloon tubing, marked BALLOON is a balloon inflation lumen. The tubing marked DISTAL, is the essential lumen of the catheter, which terminates at the distal tip. This lumen is used to pass the catheter over a guidewire. This lumen can also be used for infusion of contrast medium.

AI/ML Overview

This document describes the Boston Scientific Occluder™ Occlusion Balloon Catheter, which is cleared through the 510(k) pathway as substantially equivalent to a predicate device. As such, the study focuses on demonstrating this substantial equivalence rather than establishing new acceptance criteria or conducting a multi-reader multi-case comparative effectiveness study in the way one might for an AI/ML diagnostic device.

Here's an analysis of the provided text in the context of your request:

1. Table of Acceptance Criteria and Reported Device Performance

Criterion TypeCriterionReported Performance (from provided text)
Substantial EquivalenceThe device must demonstrate substantial equivalence to a legally marketed predicate device (Van-Tec Occlusion Balloon Catheter, K841941/A) in terms of intended use, technological characteristics, types of materials, and performance characteristics, demonstrating it is as safe, as effective, and performs as well as the predicate device."A direct comparison of key characteristics has been performed and demonstrates that the proposed Occluder Occlusion Balloon Catheter is substantially equivalent to the predicate device in terms of intended use. The proposed device is substantially equivalent to the reference devices in terms of technological characteristics, types of materials and performance characteristics. The proposed Occluder Occlusion Balloon Catheter is as safe, as effective, and performs as well as the predicate device." "The results of the performance testing demonstrate equivalence of the Occluder Catheter to the predicate device. The Occluder Occlusion Balloon Catheter is considered safe and effective for its intended use."
Performance TestingPerformance testing (bench evaluation) on samples (aged at T=0 and 7-months Accelerated Aging) to support balloon material and shaft material changes."Boston Scientific has conducted performance testing with samples aged at T=0 and 7-months Accelerated Aging in support of the balloon material and shaft material changes. The results of the performance testing demonstrate equivalence of the Occluder Catheter to the predicate device."
Intended UseThe device's intended use must align with temporary ureteral occlusion and applications including renal opacification, dislodgment of calculi, and preventative calculi migration."The Occluder™ Occlusion Balloon Catheters are indicated for use for temporary Ureteral occlusion and applications including, renal opacification, dislodgment of calculi and preventative calculi migration." (This is both the stated indication and the demonstration of alignment with the predicate's implied use).
Technological CharacteristicsThe device must have similar technological characteristics and fundamental occlusion balloon catheter design to the predicate device. Minor differences (e.g., packaging) must not raise new questions of safety or effectiveness."The Occluder Occlusion Balloon Catheters have the same technological characteristics and fundamental occlusion balloon catheter design as the predicate device. The proposed Occluder Occlusion Balloon Catheters are packed using a thermoformed, multi-product tray and supplied with a stopcock and syringe. A tray lid is applied to help secure the device inside the tray cavities. The lidded tray is then placed into a heal-sealed poly/Tyvek pouch. The pouch is labeled and placed into a labeled shelf carton along with a DFU."
Material ChangesAny balloon and shaft material changes must be supported by performance testing demonstrating equivalence."Boston Scientific has conducted performance testing with samples aged at T=0 and 7-months Accelerated Aging in support of the balloon material and shaft material changes. The results of the performance testing demonstrate equivalence of the Occluder Catheter to the predicate device."

2. Sample size used for the test set and the data provenance

The document does not specify a "test set" in the context of clinical data or patient samples. The performance testing mentioned is a bench evaluation.

  • Sample Size: The document mentions "samples aged at T=0 and 7-months Accelerated Aging." The specific number of samples tested is not provided.
  • Data Provenance: The data is from "Boston Scientific" and is generated through "performance testing (Bench Evaluation)." There is no mention of country of origin of data or whether it is retrospective or prospective, as it does not involve human subjects or historical patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable to this submission. The "ground truth" for a substantial equivalence claim based on bench testing does not involve expert consensus on clinical findings. Instead, the "truth" is established by direct measurement against engineering specifications and comparison to the predicate device's known performance/specifications.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable. Adjudication methods are typically used in clinical studies or when interpreting complex images/data to establish ground truth for diagnostic devices, which is not the case here.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices involving human interpretation, not for a medical device like an occlusion balloon catheter, especially when seeking 510(k) clearance based on substantial equivalence.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No, a standalone algorithm-only performance study was not done. This device is a physical medical instrument, not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this submission is based on engineering specifications and direct measurements/observations from bench testing. The goal was to show that the new device's physical and functional characteristics (e.g., balloon inflation, shaft properties, radiopacity) meet predetermined criteria and are equivalent to the predicate device. This essentially means the "ground truth" is adherence to design specifications and performance metrics established from the predicate.

8. The sample size for the training set

This information is not applicable. There is no "training set" as this is not an AI/ML device.

9. How the ground truth for the training set was established

This information is not applicable as there is no training set.

{0}------------------------------------------------

Occluder™ Occlusion Balloon Catheter Traditional 510(k) .

Boston Scientific

Acres of Acres of Acres2014
SECTION 5510(k) SUMMARYIN IN AN R LEWINDS SHEENDERS ANNUAL BOOK OF ALL - I BELLENIN - A LA

510(k) Summary for Occluder™ Occlusion Balloon Catheter

A. Sponsor

Boston Scientific Corporation Urology and Women's Health Division 100 Boston Scientific Way Marlborough, MA 01756

B. Contact

Christine Shoemaker Specialist II, Regulatory Affairs 508-683-4214 christine.shoemaker@bsci.com

Or

Lisa Sullivan Manager, Regulatory Affairs 508-683-4745 lisa.sullivan@bsci.com

C. Device Name

Occluder™ Occlusion Balloon Catheter Trade name: Common/usual name: Catheter, Ureteral, Gastro-Urology Classification Name: EYB - Catheter, Ureteral, Gastro-Urology

I). Predicate Device

Occluder™ Occlusion Balloon Catheter (Boston Scientific) Trade name: Common/usual name: Catheter, Urological Classification Name: KOD - Catheter, Urological

Premarket Notification: Van-Tec Occlusion Balloon Catheter (submitted by Van-Tec. Inc.), K841941/A, August 8, 1984.

E. Device Description

Occluder Occlusion Balloon Catheters are designed for use for temporary occlusion of the ureter and applications including renal opacification, dislodgement of calculi and preventative calculi migration. The devices are provided sterile and are intended for single use.

The Occluder Occlusion Balloon Catheters are constructed of a soft compliant latex balloon mounted on the tip of a catheter shaft. Catheter shafts are radiopaque, maximizing fluoroscopic visibility.

{1}------------------------------------------------

SECTION 5

510(k) SUMMARY

The Occluder Occlusion Balloon Catheters have two lumens that are marked and color-coded. The balloon tubing, marked BALLOON is a balloon inflation lumen. The tubing marked DISTAL, is the essential lumen of the catheter, which terminates at the distal tip. This lumen is used to pass the catheter over a guidewire. This lumen can also be used for infusion of contrast medium.

F. Intended Use

The Occluder™ Occlusion Balloon Catheters are indicated for use for temporary Ureteral occlusion and applications including, renal opacification, dislodgment of calculi and preventative calculi migration.

G. Technological Characteristics

The Occluder Occlusion Balloon Catheters have the same technological characteristics and fundamental occlusion balloon catheter design as the predicate device. The proposed Occluder Occlusion Balloon Catheters are packed using a thermoformed, multi-product tray and supplied with a stopcock and syringe. A tray lid is applied to help secure the device inside the tray cavities. The lidded tray is then placed into a heal-sealed poly/Tyvek pouch. The pouch is labeled and placed into a labeled shelf carton along with a DFU.

H. Substantial Equivalence

A direct comparison of key characteristics has been performed and demonstrates that the proposed Occluder Occlusion Balloon Catheter is substantially equivalent to the predicate device in terms of intended use. The proposed device is substantially equivalent to the reference devices in terms of technological characteristics, types of materials and performance characteristics. The proposed Occluder Occlusion Balloon Catheter is as safe, as effective, and performs as well as the predicate device.

I. Performance Testing (Bench Evaluation)

Boston Scientific has conducted performance testing with samples aged at T=0 and 7-months Accelerated Aging in support of the balloon material and shaft material changes.

The results of the performance testing demonstrate equivalence of the Occluder Catheter to the predicate device. The Occluder Occlusion Balloon Catheter is considered safe and effective for its intended use.

Boston Scientific

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of an eagle or bird-like figure with three curved lines representing its wings or body.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-00002

January 8, 2014

Boston Scientific Corporation Urology and Women's Health Christine Shoemaker, MS, RAC Specialist II, Regulatory Affairs 100 Boston Scientific Way Mariborough, MA 01752

K133750 Re:

Trade/Device Name: Occluder™ Occlusion Balloon Catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EYB Dated: December 11, 2013 Received: December 12, 2013

Dear Christine Shoemaker,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{3}------------------------------------------------

Page 2 - Christine Shoemaker, MS, RAC

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/3/Picture/7 description: The image shows the name "Herbert P. Lerner -S" in a bold, sans-serif font. The text is black against a white background. The letters are evenly spaced and easily readable.

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

SECTION 4 1 INDICATIONS FOR USE

Indications for Use Statement

510(k) Number : K133750

Device Name Occluder™ Occlusion Balloon Catheter

Indications For Use

The Occluder™ Occlusion Balloon Catheters are indicated for use for temporary Ureteral occlusion and applications including, renal opacification, dislodgment of calculi and preventative calculi migration.

Prescription Use X (21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert P. Lerner 2014.01.08 15:46:09

Proprietary and Confidential Information of Boston Scientific Corporation

§ 876.5130 Urological catheter and accessories.

(a)
Identification. A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).(b)
Classification. (1) Class II (performance standards).(2) Class I for the ureteral stylet (guidewire), stylet for gastrourological catheter, ureteral catheter adapter, ureteral catheter connector, and ureteral catheter holder. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.