(72 days)
The MD302C4 color display is intended to be used for displaying and viewing of digital images for diagnosis by trained physicians.
To guarantee the display performance as specified, it must only be used for in conjunction with NEC approved display controllers.
MD302C4 cannot be used for a life-support system.
This device must not be used in digital mammography.
This device is designed for exclusive interconnection with IEC60601-1-1 certified equipment.
Medical Display, MD302C4 is a 29.8" Color LCD monitor that displays image for medical use. It provides 4 mega pixel (2560*1600) resolution with adjustable gamma gray scale for more precise diagnose use in CT, MRI, HIS, and PACS.
This document describes a 510(k) premarket notification for a diagnostic imaging LCD monitor (MD302C4). The submission aims to demonstrate substantial equivalence to a predicate device (MD242C2) rather than proving the device meets specific acceptance criteria through a clinical study. As such, most of the requested information (sample size, data provenance, expert qualification, adjudication method, MRMC study, standalone performance, training set details) is not applicable to this type of submission.
Here's the information that can be extracted:
1. Table of Acceptance Criteria and Reported Device Performance
The submission does not explicitly state acceptance criteria in the typical sense of a clinical study with performance metrics. Instead, it compares the technical specifications and intended use of the new device (MD302C4) to a predicate device (MD242C2) to establish substantial equivalence. The "acceptance criteria" here are implicitly that the new device's performance characteristics are comparable or superior to the predicate device, or fall within acceptable ranges for diagnostic display devices.
| Items | Predicate Device (MD242C2) | Reported Device Performance (MD302C4) |
|---|---|---|
| Panel Size and Type | 24.1" TFT Color LCD Monitor | 29.8" TFT Color LCD Monitor |
| Pixel Pitch | 0.270 mm x 0.270mm | 0.251mm x 0.251mm |
| Display Color | 1,073,741,824 | 16,777,216 |
| Viewing Angles (°) | H:178, V:178 | H:178, V:178 |
| Scanning Frequency(H, V) | 31.5-93.8, 118.4kHz, 50-85 Hz | 31.5-98.7 kHz, 50-85 Hz |
| Native Resolutions | 1920X1200 | 2560X1600 |
| Brightness | 180 cd/m² calibrated,350 cd/m² Max. | 180 cd/m² calibrated,340 cd/m² Max. |
| Contrast Ratio | 1000 : 1 (typical) | 1000 : 1 (typical) |
| DOT Clock | 202.5 MHz (Max) (Analog)162 MHz (Max) (Digital) | 269 MHz (Max) |
| Input Signals | Three connectors: one DVI port,one Display port, one HDMI port. | Three connectors: one DVI port,one Display port, one HDMIport. |
| Input Terminals | DVI-D, Display port, HDMI port | DVI-D, Display port, HDMI port |
| USB (option) / Standard | No | No |
| Active Display Size (H x V) | Landscape: 518.4mmX324mmPortrait: 324X518.4mm | Landscape: 641mmX401mmPortrait: 401X641mm |
| Viewable Image Size | 540 mm (diagonal) | 756 mm |
| Luminance Calibration | Software | Software |
| Default Gamma | 1.8,2.0,2.2 DICOM part 14 | 1.8,2.0,2.2 DICOM part 14 |
| Power | AC100-240V, 50/60Hz | AC100-240V, 50/60Hz |
| Power Consumption | 38.4W (Max) | 87W (Max) |
| Power Save Mode | <2W | <2W |
| Dimensions (W x H x D) | W:Landscape: 556.8mmPortrait: 362.4 mmH:Landscape: 378 - 528mmPortrait: 572.4-625.2mmD: 227.6 mm | W:Landscape: 688mmPortrait: 446.8 mmH:Landscape: 466.4 - 616.4mmPortrait: 707.1-737.0mmD: 301.6 mm |
| NET Weight | 10.2 kg | 17 kg |
| Intended of use | Displaying and viewing of digital images for diagnosis by trained physiciansThis device can not use for a life support system.This device must not be use in digital mammography.This device is designed for exclusive interconnection with IEC60601-1-1 certified equipment | Displaying and viewing of digital images for diagnosis by trained physiciansThis device can not use for a life support system.This device must not be use in digital mammography.This device is designed for exclusive interconnection with IEC60601-1-1 certified equipment |
| Certifications & Standards | CE ITE/Medical Device Directive, UL/cUL (ANSI/AAMI ES 60601-1:2005), FCC Class B, EN60601-1-2, DIN V 6868-57, DICOM | CE ITE/Medical Device Directive, UL/cUL (ANSI/AAMI ES 60601-1:2005), FCC Class B, EN60601-1-2, DIN V 6868-57, DICOM |
2. Sample size used for the test set and the data provenance
Not applicable. This is a 510(k) submission for a display monitor, not an AI/algorithm-based diagnostic device requiring a test set of medical data (images, patient records, etc.). The "test" in this context refers to engineering and performance verification of the display hardware against technical specifications and safety standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. Ground truth typically refers to clinical diagnoses or pathology results for AI/algorithm validation. For a display monitor, "ground truth" would be the expected performance based on engineering specifications and calibration.
4. Adjudication method for the test set
Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a display monitor, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a display monitor, not an algorithm.
7. The type of ground truth used
Not applicable in the clinical sense. The "ground truth" for evaluating this device would be its adherence to established technical specifications, industry standards (like DICOM Part 14), and safety regulations, verified through engineering tests and calibrations.
8. The sample size for the training set
Not applicable. This is a hardware device (display monitor), not a machine learning algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable.
{0}------------------------------------------------
KB33708
Page 1 of 4
510(K) Summary of Safety and Effectiveness
As required by 807.92
FEB 1 4 2014
DEVICE ESTABLISHMENT AND CONTACT PERSON l.
Mr. Naohiko Shimazu Senior Manager NEC Display Solutions Ltd. 686-1 NISHIOI, OI-MACHI, ASHIGARAKAMI-GUN, KANAGAWA, 258-8533, Japan Ph: +81-465-85-2384 Fax: +81-465-85-2393
2. COMPANY REISTRATION NUMBER
3003623028
3. DATE SUMMARY PREPARED
09 October 2013
4, DEVICE NAME
| Trade Name: | MD302C4 29.8" Diagnostic Imaging LCD monitor |
|---|---|
| Model Name: | MD302C4 |
| Common Name: | Color LCD Monitor, Color Diagnostic Display, etc. |
| Classification Name: | System, Image Processing, Radiological (CLASS II CFR892.2050) |
PREDICATE DEVICE 4.
MD242C2 Color LCD Monitor by NEC Display Solutions Ltd. (K 132587)
DEVICE DESCRIPTION 5.
Medical Display, MD302C4 is a 29.8" Color LCD monitor that displays image for medical use. It provides 4 mega pixel (2560*1600) resolution with adjustable gamma gray scale for more precise diagnose use in CT, MRI, HIS, and PACS.
{1}------------------------------------------------
6. DEVICE OF INTENDED USE
The MD302C4 color display is intended to be used for displaying and viewing of digital images for diagnosis by trained physicians.
To guarantee the display performance as specified, it must only be used for in conjunction with NEC approved display controllers.
MD302C4 cannot be used for a life-support system.
This device must not be used in digital mammography.
This device is designed for exclusive interconnection with IEC60601-1-1 certified equipment.
7. SE Comparison Table:
| Items | MD242C2 | MD302C4 |
|---|---|---|
| 510(k) Number | ||
| Panel Size and Type | 24.1" TFT Color LCD Monitor | 29.8" TFT Color LCD Monitor |
| Pixel Pitch | 0.270 mm x 0.270mm | 0.251mm x 0.251mm |
| Display Color | 1,073,741,824 | 16,777,216 |
| Viewing Angles (°) | H:178, V:178 | H:178, V:178 |
| Scanning Frequency(H, V) | 31.5-93.8, 118.4kHz, 50-85 Hz | 31.5-98.7 kHz, 50-85 Hz |
| Native Resolutions | 1920X1200 | 2560X1600 |
| Brightness | 180 cd/m² calibrated,350 cd/m² Max. | 180 cd/m² calibrated,340 cd/m² Max. |
| Contrast Ratio | 1000 : 1 (typical) | 1000 : 1 (typical) |
| DOT Clock | 202.5 MHz (Max) (Analog)162 MHz (Max) (Digital) | 269 MHz (Max) |
| Input Signals | Three connectors: one DVI port,one Display port, one HDMI port. | Three connectors: one DVI port,one Display port, one HDMIport. |
| Input Terminals | DVI-D, Display port, HDMI port | DVI-D, Display port, HDMI port |
| USB (option) / Standard | No | No |
| Active Display Size (H x V) | Landscape: 518.4mmX324mmPortrait: 324X518.4mm | Landscape: 641mmX401mmPortrait: 401X641mm |
| Viewable Image Size | 540 mm (diagonal) | 756 mm |
| Luminance Calibration | Software | Software |
| Default Gamma | 1.8,2.0,2.2 DICOM part 14 | 1.8,2.0,2.2 DICOM part 14 |
| Power | AC100-240V, 50/60Hz | AC100-240V, 50/60Hz |
| Power Consumption | 38.4W (Max) | 87W (Max) |
| Power Save Mode | <2W | <2W |
| Dimensions (W x H x D) | W:Landscape: 556.8mmPortrait: 362.4 mmH:Landscape: 378 - 528mmPortrait: 572.4-625.2mmD: 227.6 mm | W:Landscape: 688mmPortrait: 446.8 mmH:Landscape: 466.4 - 616.4mmPortrait: 707.1-737.0mmD: 301.6 mm |
| NET Weight | 10.2 kg | 17 kg |
| Intended of use | Displaying and viewing of digital images for diagnosis by trained physiciansThis device can not use for a life support system.This device must not be use in digital mammography.This device is designed for exclusive interconnection with IEC60601-1-1 certified equipment | Displaying and viewing of digital images for diagnosis by trained physiciansThis device can not use for a life support system.This device must not be use in digital mammography.This device is designed for exclusive interconnection with IEC60601-1-1 certified equipment |
| Certifications & Standards | CE ITE/Medical Device Directive, UL/cUL (ANSI/AAMI ES 60601-1:2005), FCC Class B, EN60601-1-2, DIN V 6868-57, DICOM | CE ITE/Medical Device Directive, UL/cUL (ANSI/AAMI ES 60601-1:2005), FCC Class B, EN60601-1-2, DIN V 6868-57, DICOM |
Comparison tables between MD302C4 & MD242C2
{2}------------------------------------------------
v - 3
:
{3}------------------------------------------------
K133708
Page 4 of 4
CONCLUSION
These two devices have the same target population of trained practitioner in hospital; it shares the same design. same performance and is the same in radiation safety (EN60601-1-2), mechanical safety, electrical safety (AAMI/ES 60601-1) human factors and DICOM conformance. It use similar material, and have same compatibility with environment and other device. Comparison table of the principal characteristics of two devices is shown in the Section 3, table 3.3. These two devices also have the same intended use; Therefore we concluded that it is substantially equivalent to MD242C2 by NEC Display Solutions Ltd. (K132587)
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Contro! Center - WO66-G609 Silver Spring, MD 20993-0002
February 14, 2014
NEC Display Solutions, Ltd. % Mr. Tony Hsu Safety Engineer Prodigy Technology Consultant Co., Ltd. 1F, No. 181, Sec. 2 Wunhua 1st Road. Linkou Township NEW TAIPEI CITY 24457 TAIWAN
Re: K133708
Trade/Device Name: MD302C4 29.8" Diagnostic Imaging LCD Monitor Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 23, 2013 Received: December 23, 2013
Dear Mr. Hsu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2-Mr. Hsu
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K133708
Device Name MD302C4
Indications for Use (Describe)
The MD302C4 color display is intended to be used for displaying of digital images for diagnosis by trained physicians.
To guarantee the display performance as specified, it must only be used for in conjunction with NEC approved display controllers.
MD302C4 cannot be used for a life-support system.
This device must not be used in digital mammography.
This device is designed for exclusive interconnection with IEC60601-1-1 certified equipment.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
N/A