(42 days)
The X-ray Attenuating Cream has the same intended use, and indications for use as before, with the addition of indications for use with poly-isoprene surgeon's gloves as below:
The X-ray Attenuating Cream is intended for use as a radiation shield. It is intended to be applied to the user's hand before donning gloves, or it may be applied on a glove on the hand, followed by donning a second glove. The X-ray Attenuating Cream is intended to be used during medical procedures where hands are necessarily exposed to radiation to offer some degree of protection from radiation exposure in the diagnostic imaging range of up to 130 kVp. This may include surgical procedures that require the use of fluoroscopy or radiography or other procedures.
NOTE: For use with natural rubber latex and latex-free poly-isoprene Surgeon's Gloves only.
The X-ray Attenuation Cream is a sterile cream intended for use as a radiation shield. It is intended to be applied to the user's hand before donning gloves, or it may be applied on a glove on the hand, followed by donning a second glove. The X-ray Attenuation Cream is intended for use during medical procedures in which hands are necessarily exposed to radiation to offer some degree of protection from radiation exposure in the diagnostic imaging range of up to 130 KVp. This may include surgical procedures that require the use of fluoroscopy or radiography or other procedures. The X-ray Attenuation Cream is not intended to be used in the primary beam or the transmitted beam and should not be used in lieu of a Radiographic Procedure Glove, which is used in radiography for those studies requiring the physician's hand or forearm be in the direct path of the primary x-ray beam.
Acceptance Criteria and Study for X-ray Attenuating Cream (K133684)
This submission (K133684) is a 510(k) for an X-ray Attenuating Cream (Ultrablox) seeking to expand its indications for use to include compatibility with latex-free poly-isoprene surgeon's gloves, in addition to its previously cleared use with natural rubber latex surgical gloves. The original device was cleared under De Novo 510(k) K123422.
Given that this 510(k) is for an expansion of use for an already cleared device, and the primary change is compatibility with a different type of glove, the performance testing focuses on demonstrating this compatibility. The document explicitly states: "The performance data presented in this 510(k) application demonstrate the X-ray Attenuating Cream can be used with latex-free poly-isoprene surgeon's gloves in addition to the previously cleared use with natural rubber latex surgeon's gloves."
Therefore, the acceptance criteria and study design are centered around confirming that this new compatibility does not negatively impact the device's function or safety, implicitly relying on the original clearance (K123422) for the core performance aspects (radiation attenuation and biocompatibility).
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state quantitative acceptance criteria or detailed performance results in the format of a table for this 510(k) submission (K133684). This is because the device itself (its composition, radiation attenuation, and biocompatibility) has not changed from the predicate device (K123422). The "performance testing" described is to demonstrate compatibility with a new glove material.
However, based on the principle of substantial equivalence for this type of submission, the implicit acceptance criteria and reported performance are:
| Acceptance Criteria Category | Specific Acceptance Criteria (Implicit) | Reported Device Performance (as stated or implied) |
|---|---|---|
| Compatibility with Latex-Free Poly-isoprene Gloves | The X-ray Attenuating Cream, when used with latex-free poly-isoprene surgeon's gloves, must maintain its intended function as a radiation shield. | "The performance data presented in this 510(k) application demonstrate the X-ray Attenuating Cream can be used with latex-free poly-isoprene surgeon's gloves..." |
| Material Degradation / Impact on Glove Integrity | The cream should not degrade the integrity or performance of the latex-free poly-isoprene gloves. | (Implied by the statement of "compatibility"; no issues reported) |
| Radiation Attenuation | The radiation attenuation characteristics of the cream remain unchanged, regardless of the glove material. | "Radiation attenuation: None [change from predicate]" |
| Biocompatibility | The biocompatibility of the cream remains unchanged. | "Bio-compatibility: None [change from predicate]" |
| Packaging and Sterilization | Packaging and sterilization methods remain unchanged and effective. | "Packaging and sterilization: None [change from predicate]" |
| Intended Use | The intended use and indications for use remain consistent with the predicate, with the addition of poly-isoprene glove compatibility. | "The X-ray Attenuating Cream has the same intended use, and indications for use as before, with the addition of indications for use with poly-isoprene surgeon's gloves..." |
2. Sample Size Used for the Test Set and Data Provenance
The document for K133684 does not specify a separate "test set" with a sample size in the conventional sense (e.g., patient data, image dataset). The "performance testing" referenced appears to be focused on material compatibility testing rather than a clinical study. The details of the "performance data" are not provided in this summary but are stated to demonstrate compatibility.
Therefore, based on the provided document:
- Sample Size for Test Set: Not specified.
- Data Provenance: Not specified. It is likely laboratory-based material compatibility testing rather than clinical data from a specific country.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This type of information is not applicable to this 510(k) submission. The "performance testing" detailed in this summary relates to material compatibility, not the assessment of imaging data or clinical outcomes that would require ground truth established by experts like radiologists or pathologists.
4. Adjudication Method for the Test Set
This information is not applicable to this 510(k) submission, as it does not involve assessment of imaging data or clinical scenarios requiring adjudication.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study
An MRMC study was not done. This type of study would be relevant for evaluating the performance of an AI algorithm in an imaging context, which is not the subject of this medical device submission.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
A standalone study was not done. This submission is for an X-ray attenuating cream, not a digital algorithm or AI-powered device.
7. Type of Ground Truth Used
This information is not applicable to this 510(k) submission. The "ground truth" for material compatibility would typically be based on established material science standards, physical testing, or chemical analysis, not expert consensus, pathology, or outcomes data in a clinical sense. The document implies that the "performance data" collected confirmed the physical and chemical compatibility of the cream with poly-isoprene gloves.
8. Sample Size for the Training Set
This information is not applicable to this 510(k) submission. There is no AI or machine learning component requiring a training set.
9. How the Ground Truth for the Training Set was Established
This information is not applicable to this 510(k) submission, as there is no training set.
{0}------------------------------------------------
K133684
Page 1 of 3
JAN 13 2014
X-ray Attenuating Cream Section 5 -- 510(k) Summary
1 GENERAL INFORMATION
1.1 Submitter and Owner of the 510(k) BLOXR Corporation 960 Levoy Drive Salt Lake City. Utah 84123
1.2 Official Correspondent
Rai Chowdhary Vice President of Operations and Engineering 960 Levoy Drive Salt Lake City, Utah 84123 Telephone: (512) 560-8326 Cell: (512) 560-8326 Fax: (801) 254-4888 E-mail: rchowdhary@bloxr.com
1.3 Date of Preparation Nov 25 2013
2 NAME OF THE DEVICE 2.1 Trade/Proprietary Name ULTRABLOX
2.2 Common/Usual Name
X-ray Attenuating Cream, or Cream for x-ray attenuation
2.3 Classification Information
Classification Name: Cream for x-ray attenuation Classification Regulation: 21 CFR § 892.6510 Class: II Product Code: PDK Panel: Radiology
X-Ray Attenuating Cream
1 Nov 25 2013
{1}------------------------------------------------
3 PREDICATE DEVICES
This device is the same X-ray Attenuating Cream that was cleared under the De Novo 510(k) K 123422, and classified as a Class II medical device. The device was earlier cleared for use with natural rubber latex surgical gloves only.
4 DESCRIPTION OF THE DEVICE
The X-ray Attenuation Cream is a sterile cream intended for use as a radiation shield. It is intended to be applied to the user's hand before donning gloves, or it may be applied on a glove on the hand, followed by donning a second glove. The X-ray Attenuation Cream is intended for use during medical procedures in which hands are necessarily exposed to radiation to offer some degree of protection from radiation exposure in the diagnostic imaging range of up to 130 KVp. This may include surgical procedures that require the use of fluoroscopy or radiography or other procedures. The X-ray Attenuation Cream is not intended to be used in the primary beam or the transmitted beam and should not be used in lieu of a Radiographic Procedure Glove, which is used in radiography for those studies requiring the physician's hand or forearm be in the direct path of the primary x-ray beam.
5 INDICATIONS FOR USE AND INTENDED USE
The X-ray Attenuating Cream has the same intended use, and indications for use as before, with the addition of indications for use with poly-isoprene surgeon's gloves as below:
The X-ray Attenuating Cream is intended for use as a radiation shield. It is intended to be applied to the user's hand before donning gloves, or it may be applied on a glove on the hand, followed by donning a second glove. The X-ray Attenuating Cream is intended to be used during medical procedures where hands are necessarily exposed to radiation to offer some degree of protection from radiation exposure in the diagnostic imaging range of up to 130 kVp. This may include surgical procedures that require the use of fluoroscopy or radiography or other procedures.
NOTE: For use with natural rubber latex and latex-free poly-isoprene Surgeon's Gloves only.
X-Ray Attenuating Cream
{2}------------------------------------------------
K133684
Page 3 of 3
6 TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE DEVICES
The X-ray Attenuating Cream is the same device that was cleared under De Novo 510(k) K123422. There is no change to packaging and sterilization. Since it is the same device, it has the same radiation attenuation and bio compatibility characteristics. The X-ray Attenuating Cream is now compatible for use with natural rubber latex, and latex-free poly-isoprene Surgeon's Gloves. The table below provides for a summary of what has changed in the X-ray Attenuating Cream.
| Characteristic | Change from Predicate(from / to) |
|---|---|
| Composition | None |
| Radiation attenuation | None |
| Bio-compatibility | None |
| Compatibility withgloves | Yes(compatible with natural rubberlatex surgeons gloves /compatible with natural rubber latexsurgeons gloves, and latex-freepoly-isoprene surgeon's gloves) |
| Packaging andsterilization | None |
| Type of proceduresused for | None |
| Intended use | ... not intended to be used in oradjacent to the primary beam.../...not intended to be used in theprimary beam... |
X-ray Attenuating Cream vs. predicate X-ray Attenuating Cream
7 PERFORMANCE TESTING
The performance data presented in this 510(k) application demonstrate the X-ray Attenuating Cream can be used with latex-free poly-isoprene surgeon's gloves in addition to the previously cleared use with natural rubber latex surgeon's gloves.
8 CONCLUSIONS
This 510(k) submission provides the evidence that the X-ray Attenuating Cream is substantially equivalent to the device cleared under 510(k) K 123422.
X-Ray Attenuating Cream
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G6 Silver Spring, MD 20993-0002
January 13, 2014
BLOXR Corporation % Mr. Prataprai (Rai) Chowdhary VP Operations and Engineering 960 Levoy Drive SALT LAKE CITY UT 84123
Re: K133684
Trade/Device Name: Ultrablox Regulation Number: 21 CFR 892.6510. Regulation Name: Cream for X-ray Attenuation Regulatory Class: II Product Code: PDK Dated: November 26, 2013 Received: December 2, 2013
Dear Mr. Chowdhary:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2-Mr. Chowdhary
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,

for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K133684
Device Name
UltraBLOX X-ray Attenuating Cream
Indications for Use (Describe)
The X-ray Attenuating Cream has the same indications for use as the device cleared under 510(k) K 123422, with the addition of indications for use with poly-isoprene surgeon's gloves as below:
Device Name: X-ray Attenuating Cream
Indications for use:
The UltraBLOX X-ray Attenuating Cream is intended for use as a radiation shield. It is intended to the user's hand before donning gloves, or it may be applied on a glove on the hand, followed by donning a second glove. The UltraBLOX X-ray Attenuating Cream is intended to be used during medical procedures where hands are necessarily exposed to radiation to offer some degree of protection from radiation exposure imaging range of up to 130 kVp. This may include surgical procedures that require the use of fluoroscopy or radiography or other procedures.
NOTE: For use with natural rubber latex and latex-free poly-isoprene Surgeon's Gloves only.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Smh?)
- Publishing Servers (10) | 441-6741
{6}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
N/A